CUSTOMER : M/S FONTUS WATER PVT. LTD.

D-4, FREEDOM FIGHTER ENCLAVE, NEB SARAI, IGNOU ROAD NEW DELHI
Calibration Certificate of : VERNIER CALIPER Certificate No. – HTCS/03166
Make : MITUTOYO Service Request No. : HTCS/03166
Range : 0-150 mm Service Request Date : 25.04.2012
L.C. : 0.02 mm Date of Calibration : 26.04.2012
I.D. No. : FWPL/VC-01 Suggested due Date : 26.04.2013
Sl. No. : 90010252 Page : 1 of 1
: :
Visual Inspection OK Calibration Method WI-01
:
Environmental Condition of Lab.- : Uncertainty of Measurement : ±0.016 mm
Temperature : (20+/-1) Deg. C (At 95% Confidence Level,
Humidity (50+/-10)% RH k=1.96)

DETAILS OF STANDARD EQUIPMENTS USED FOR CALIBRATION


Name of the Instrument : CALIPER CHECKER : :
Make : KCP : :
Certificate No. : D/09/90013 : :
Sl. No./ID No. : 563 : :
: 23/12/12 : :
Calibration Due Date
: ACCURATE GGN : :
Traceable By

Calibration Results
Sl. No. Standard Value Measured Value Measured Value
(mm) (OD) (ID)
01 000.00 000.00 000.00
02 020.00 020.00 020.00
03 050.00 050.00 050.00
04 100.00 100.00 100.00
05 150.00 150.00 150.00

Remarks:-

Notes :-
1. This calibration results reported in this certificate are valid at the time of & under the stated conditions of measurements.
2. This report should not be reproduced except in full without our prior permission in writing.
3. This certificate relates only to the item calibrated.
4. Format No.-F01(5.10/01).

Calibrated By Approved By