Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
The DPD method is relatively stable but not as accurate as some electrochemical techniques and ranges are typically
limited to 0 - 5.0 mg/I as free or total chlorine. This method is used extensively and quite adequately as a monitoring
instrument however it suffers from a few drawbacks. Since the DPD forms a free radical, it oxidizes quite rapidly. As
a result, monochloramines are readily reacted and subsequently reported as free chlorine. Additionally, this method
has problems when attempting to control chlorine addition. This is because the output of these analyzers must be held
during the reagent reaction time. So in the case above, the output is updated every 2 1/2 minutes. Finally, most of
the on-line colorimetric analyzers for chlorine do not have temperature compensation. Some manufacturers offer a sample
heater to maintain the sample at a constant temperature. Figure 6 demonstrates a comparison of the output of a continuous
method vs. batch method.
Potentiometric/lodometric
The potentiometric method measures trace materials in a solution using an electrode with two elements. This method is
only applicable to the measurement of total residual chlorine and is based on the detection of iodine, which is liberated from a
stoichiometric reaction with the total chlorine in the sample.
-5- 010.3035.0
Formerly 210.9005.1
Figure 6 - Continuous vs. Batch Output
-7- 010.3035.0
Formerly 210.9005.1
The reaction that occurs at the gold cathode in a buffered amperometric analyzer is the following:
HOCI + H+ + 2e ----> CI- + H2O
The anode is copper. As electrons are exchanged, an oxide product remains on the anode. The electrodes are kept clean
through the use of cleaning balls which are agitated to remove build-up as it occurs. Current flow is affected by pH. The
most stable current is between 4 to 4.5 pH therefore a buffering solution is used to stabilize the cell current. Temperature
compensation circuitry is employed to counter the effects of accelerated diffusion. Temperature is compensated through
the use of a thermistor. If total chlorine is desired, potassium iodide can be added to the pH 4 buffer solution.
Most amperometric analyzers on the market today utilize this or a similar technology.
ORP
An alternative method of chlorine residual monitoring that uses redox chemistry is known as ORP (oxidation reduction
potential). This form of measurement is a reading of the mV (millivolt) potential across a pair of electrodes. Originally, it had
been hoped that this will be a satisfactory method of indicating chlorine and investigative work continues in this area. This
work has had its deficiencies over the past years encountering unpredictable readings and variations in accuracy and
reliability. All chemical reactions that involve an electron transfer are oxidation reduction reactions. The relative strength
of these reactions is determined by measuring the potential. ORP is the potential to transfer electrons from the oxidant to the
reductant and is used as a qualitative measure of the state of oxidation in process waters. ORP is measured with
a platinum electrode and a reference electrode. If follows the Nernst equation which is described as follows:
E = Eo +(RT/n) In oxidant/reductant
E = The measured potential in voltage Eo = The standard molar potential R = Constant
T = Temperature, Degrees Kelvin
n = Number of electrons involved in the reaction
Oxidant = Molar concentration (activities)
Reductant = Molar concentration (activities)
We see that the potential is based on the ratio of oxidants to reductants, temperature and the equilibrium in standard
conditions (where oxidants = reductants). This calculation is only useful in a relatively pure or constant system. In more complex
waters, the concentration of specific oxidants or reductants cannot be determined by the electrode potential alone.
The use of ORP as a chlorine control devise assumes that all the oxidants are chlorine and the reductants do not change. This
matrix effect can be very complex. At best, the results from ORP are qualitative. If a relationship can be established
between the ORP and the chlorine level, care should be taken to continually calibrate this system because of fouling or drift
due to sample composition variations.
The ORP sensing system is comprised of a reference and sensing electrode with a very pure platinum element.
Comparison Data
Figure 9 describes an evaluation study of several methods of chlorine detection as compared to laboratory amperometric
titration. We see that the two electrode amperometric analyzer is the closest to laboratory amperometric titration. This
chart describes a range of 0 to 10 mg/I chlorine. The data below 3 mg/I is inconclusive and was not examined in detail. This
chart shows how the different methods compare over a wide range of chlorine.
Figure 10 shows several months of data taken at a wastewater treatment plant with different analyzers. We see that if we
attempt to compare analyzers with each other, the data is meaningless. This is the kind of variability that can occur
in different analyzers. In this chart we do not know which instrument is right. A referee method must be used as the basis of all
on-line and laboratory instruments. To further illustrate, as trials are run comparing competitive equipment, there is always one
detection device that is viewed as the standard or the control. Frequently, it is another on-line analyzer or in many cases it is a
DPD test kit. Care should be given in placing confidence in one instrument over another. This is why the AWWA went
to such pains to study the accuracy of various methods. The need was recognized for a referee method that could act as
a universal control. The laboratory amperometric titrator was selected because this instrument provided the most accurate
and precise chlorine values. It would be a difficult task to pick the most accurate on-line instrument when comparing against
each other. Although ORP was not part of this study, we see that this form of chlorine control also needs a highly accurate
lab technique to verify accuracy, stability and calibration.
010.3035.0 -8-
Formerly 210.9005.1
Figure 9 - Comparison Test With On-Line Instruments
5. Case Study
Problem
Minimum Chlorine Control
Critical cooling towers often need to control the minimum chlorine level as low as 0.3 mg/I free chlorine. At these levels
the operators are assured that both microbial growth and slime is controlled. Left uncontrolled, biofilms can
form and adversely affect equipment performance by decreasing heat transfer and reducing water flow. Biofilms can also
promote metal corrosion that can perforate heat exchanger surfaces and can accelerate wood deterioration resulting in collapse
of the structure.
There are several problems to the control and measurement of chlorine at the 0.3 mg/I level. First is that there are other
measurements that affect the level of activity of these microbial organisms and the effectiveness of chlorine. Among these
are pH, temperature and organics. The critical parameter has been pH. In most of the systems approaches offered
by water treatment companies, pH is regarded as a parameter that must be controlled in order to maximize the efficiency
of the towers. For the case study described in this paper, this level is set and controlled at 6.8 to 7.2 pH. When pH is closely
controlled into a narrow margin, the effectiveness of chlorine is easily predictable based on what is known about the
-9- 010.3035.0
Formerly 210.9005.1
HOCI/OCI- dissociation curve (see Figure 1). This is why amperometric analyzers try to adjust the pH of the sample
to below 4.5 pH. The signal is much stronger because almost 100% of the CI2 is HOCI. When the pH is tightly controlled, it is
easy to predict the required chlorine residual. This is how the minimum of 0.3 mg/I chlorine was established and then
proven experimentally. The question to be answered is that given the accuracy of most chlorine instruments, what method or
technique can be used to determine this low level of chlorine accurately?
Maximum chlorine control
If we were just worried about disinfection we would simply keep the chlorine residual high and let the system
run. However the other factor in this equation is that high levels of chlorine will render water treatment systems ineffective
and pave the way for corrosion and scale. Selective leaching can occur when the pH is low (below 6.0 pH) and the chlorine
is high (>1.0 mg/I CI2). Also, when the Cl2 level increases above 1.0 mg/I, tower lumber delignification can occur. We
can see that in order to keep the corrosion inhibitors working effectively and keep the structure intact, we need to keep
the pH above 6.0 pH and the chlorine below 1.0.
Figure 11 describes this relationship between pH and
corrosion. Tight control of pH will enable the corrosion and
scale inhibitors to work much more effectively, reduce
fouling and increase the effectiveness of the disinfectant.
From the information presented, we know that this will occur
above a pH of 6.0 and below a pH of 10.0. Above pH of 10.0 the
dissociation curve indicates most of the active chlorine is in the
form of hypochlorite. Various treatment systems will all have an
optimum pH. Careful control of pH means a more accurate
understanding of disinfection ability.
With controls on pH we can now turn our attention to the
control and measurement of Cl2. The following case study
describes how to best monitor and control chlorine in a critical
cooling tower.
Case Study
In this example, we will examine how precise Cl2 control can
help maintain the cooling tower at maximum efficiency. Figure 11 - Corrosion Rate
The cooling tower in this discussion is a twin cell counterflow used
in the petrochemical industry. The tower runs with a ∆T of 15° F and a 45,000 gal/min recirculating rate using river water as
makeup. The chemical additives used are supplied by a major water treatment company and uses a polyphosphate and
phosphonate blend with two separate dispersion packages. This system requires the pH to be controlled at 7.0 ±0.2 The
operators also had to make sure that the chlorine residual was between 0.3 and 0.6 mg/I CI2. This was necessary in order
to keep the heat exchanger corrosion rate below the cutoff of 1.5 millimeters/year. This control of 0.3 to 0.6 mg/I was
necessary in order to meet the chemical additive manufacturers specifications. The system proved to run much better
when the residual chlorine could be targeted around 0.47 mg/I Cl2. The following bar chart (Figure 12) describes this.
Initially the control of Cl2 was accomplished through a manual method. Samples were collected continually and a DPD
chlorine test was performed. This method did not provide the continuous control and monitoring that was necessary.
The first on-line system that was attempted was ORP. ORP theory indicates that as pH drifts the ability to disinfect
changes and the only true way to know that disinfection capability is through the oxidation-reduction potential of the water.
The theory is that since HOCI is a more powerful oxidant than OCI -, as the pH increases or decreases the relationship of
HOCI vs. OCI - changes so the ORP changes and the effectiveness of the disinfection changes. When the ORP system
is applied, the chlorine level began to drift upward and out of the upper limit without the ORP indicating this. We
discussed earlier why this can occur due to the relationship of oxidants/reductants. Another reason is drifting of the
probe from its calibration points. When the DPD method was performed, the operators noticed a very high
chlorine level. The initial thought was that the ORP was correct and no change to disinfection was correct. The
discussions ended when corrosion coupons indicated that the rate exceeded the 1.5 mil/year cutoff. Confirmation
occurred when the corrosion inhibition system failed due to high CI2 levels and a failure occurred in the heat
exchanger. In fact what had happened is that the ORP had drifted and the DPD answer was right in terms of
its high chlorine indication.
ORP failed to indicate that the chlorine residual was too high. A more sophisticated ORP system was tried with
the same inaccuracies. This second system was pulled in 3 weeks. The operators switched back to a manual
control system in order to keep the tower running.
010.3035.0 - 10 -
Formerly 210.9005.1
They realized that the chemical additive system that
controls corrosion and scaling is based on the precise
control of chlorine residual around 0.45 to 0.49 mg/I C12,
therefore a direct measurement of residual of CI2 is better
than an inferred measurement system. About one year
ago the continuous control of chlorine was attempted
using an amperometric analyzer.
Solution
The decision was made to try a microprocessor-based
amperometric analyzer to measure and control chlorine
feed.
The analyzer used in this case study incorporates the
latest technology in chlorine measurement. Most of
these additions seek to control the variables in chlorine
measurement. Among these are the flow of the sample
through the cells, the temperature compensation scheme
and the pH of the sample. The following diagrams the basic Figure 12 - Chlorine Control Band
flow of the Model 1870E Chlorine Analyzer, Figure 13.
While the general electronics of the unit is the same as other units, the ability to accurately measure low
residual levels is further enhanced by a cell pH monitoring and PID control system. The cell pH monitoring system
controls the pH level at 4.3 pH through a buffer feed system. As the pH of the buffered sample begins to increase
or decrease, the pump increases or decreases the feed rate appropriately. When the pH is controlled at those levels,
any variability resulting from dissociation is eliminated. If the flow is not kept above a certain rate, the cleaning
action of the unit can diminish and flow of sample past the electrodes will be too low to achieve an accurate signal. The
analyzer used on this cooling tower utilizes an infrared flow meter to indicate a low flow rate. Additionally, temperature
compensation is critical in the precise measurement of chlorine so this unit has a solid state temperature device
to compensate for sample temperature changes. What makes this particular amperometric analyzer work for critical
cooling towers is the independent control of each potential variant that may affect the chlorine reading. This includes
pH, temperature and flow.
When the analyzer and the residual chlorine level was maintained within the 0.45 to 0.49 level as confirmed by the lab
method the corrosion rate dropped from 1.5 mils/year to 0.6 mils/year. This represents a 60% decrease in the corrosion
rate. The benefits to this critical heat exchanger include a drop in the corrosion rate, a more efficient water treatment
process, improved heat transfer rates, and an extended life for the system.
- 11 - 010.3035.0
Formerly 210.9005.1
Figure 14 - Chlorine Control in Cooling Towers