Sei sulla pagina 1di 8

BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE 311

Original Article

IMPORTANCE OF THE PRE-ANALYTICAL PHASE


IN BLOOD GLUCOSE ANALYSIS
K. Janssen, J. Delanghe

Department of Clinical Chemistry, Ghent University Hospital, Belgium

Correspondence and offprint requests to: Joris Delanghe, E-mail: Joris.Delanghe@UGent.be

Key words: blood glucose analysis, pre-analytical, point-of-care


ABSTRACT testing

Blood glucose levels are characterized by a relatively


large intra-individual biological variability due to food INTRODUCTION
intake, physical activity and the body’s homeostatic
response. Careful attention to the pre-analytical phase is Blood glucose analysis is the cornerstone of both the di-
essential to ensure accurate glucose measurements. Blood agnosis and monitoring of therapy in patients with diabetes
samples should be drawn in the morning after an overnight mellitus. The current diagnostic limits of diabetes mellitus are
fast. Proper sample processing after blood collection is narrow and increase the need for reliable results to classify
crucial. When fast separation of the cells is not possible, individuals correctly (1). Methods for blood glucose measure-
blood should be collected into a tube containing a glucose ment exhibit a low imprecision at the diagnostic decision limits
preservative. Glucose concentrations may also differ of 7.0 mmol/L (126 mg/dL, fasting) and 11.1 mmol/L (200 mg/
according to the blood sampling site (venous, arterial or dL, post glucose load). However, classification errors may occur
capillary blood). Plasma and whole blood glucose values due to the relatively large intra-individual biological variability
are not interchangeable. The International Federation of (CVs of ~ 5-7%) (2). Next to physiological variability, blood
Clinical Chemistry and Laboratory Medicine recommends glucose analysis depends on pre-analytical conditions (fasting
reporting the glucose concentration in plasma to avoid or post-prandial), specimen type (whole blood, capillary or ve-
clinical misinterpretations irrespective of the sample type nous blood, serum or plasma), presence of a glycolytic inhibi-
and method of measurement. Point-of-care testing (POCT) tor in the collection tube as well as time from sampling until
glucose meters are widely used by both health profession- cell separation. Careful attention to the pre-analytical phase
als and diabetic patients to monitor blood glucose levels. is essential to ensure accurate glucose measurements. Blood
However, one should take into account that the reliability glucose testing can be performed in a central laboratory or ob-
of POCT glucose measurements depends upon a variety of tained by point-of-care testing (POCT). Nowadays, POCT glu-
factors including underlying disease, patient drug regimens cose devices are increasingly used to complement glucose test-
and interfering substances as well as instrument analytical ing at the central laboratory as it provides glucose results to be
performance and user proficiency. It is recommended to immediately available to the medical staff for taking therapeu-
perform a laboratory blood glucose analysis if the POCT tically important decisions. However, the quality of POCT still
glucose value is in the critical hypoglycaemic or hypergly- remains a concern especially in critical care settings.
caemic range. In this review, we focus on the pre-analytical requirements
of blood glucose analysis and discuss the usefulness of POCT
glucose measurements.
Abbreviations: ADA, American Diabetes Association; CLIA,
Clinical Laboratory Improvement Amendments; CLSI, Clinical
Laboratory Standards Institute; ICU, intensive care unit; IFCC,
International Federation of Clinical Chemistry and Laboratory PRE-ANALYTICAL CONSIDERATIONS
Medicine; NACB, National Academy of Clinical Biochemistry;
NaF, sodium fluoride; NCCLS, National Committee on Clinical Despite their experience in diagnosis and management
Laboratory Standards; OGTT, oral glucose tolerance test; POCT, of diseases, physicians often show a lack of knowledge about
point-of-care testing; SMBG, self-monitoring of blood glucose; the role of pre-analytical variables on clinical laboratory tests.
WHO, World Health Organization Moreover, since diabetic patients and diabetic educators are

Acta Clinica Belgica, 2010; 65-5


312 BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE

primarily responsible for the proper management of their con- Glucose analysis in capillary blood is not recommended in
dition, they must be knowledgeable of the factors that might situations with a reduced peripheral blood flow as it may not
render data invalid when specimens are collected (3). reflect the true physiological state. Fingerstick blood testing
must be avoided in the following situations: severe dehydration
The pre-analytical phase is an important, but often under- caused by diabetic ketoacidosis or hyperglycaemic hyperosmo-
estimated part of a laboratory analysis. About 46-68% of the lar nonketotic state, hypotension, shock or a peripheral vascu-
laboratory errors occur in the pre-analytical phase and can be lar disease (18). Furthermore, Stahl et al. investigated whether
avoided (4, 5). The pre-analytical phase consists of an order for it was possible to perform glucose analysis on capillary whole
laboratory analysis, sample collection, transport, registration, blood for the diagnosis of diabetes mellitus in the individual
sample processing (e.g. centrifugation) and preservation (6, 7). patient. Use of mathematically converted capillary glucose val-
All of these steps can be a source of error. ues instead of the recommended venous plasma values may
result in a random, unpredictable misclassification of patients
Generally, pre-analytical variables are divided into three (19).
categories: physiological factors, factors related to specimen
collection and interferences. Physiological factors comprise
age, sex, race, circadian rhythm, season, altitude, menstrual Influence of short-term venous stasis during
cycle, lifestyle and pregnancy which can all affect to a greater phlebotomy
or lesser extent laboratory results (6, 7). Venous stasis from tourniquet application during phle-
botomy should be minimized. Ideally, the tourniquet should be
The ultimate glucose result is usually determined by indi- released as soon as the needle enters the vein and the blood
vidual biological and pre-analytical factors (8). Blood glucose sampling time should be as short as possible. However, in rou-
levels vary continuously according to food intake, physical tine clinical practice the tourniquet is rarely released before
activity and the body’s homeostatic response. Besides delay the blood collection is complete to avoid vein collapse. Fur-
between blood sampling and processing, storage time and thermore, several concurrent causes may prolong the blood
temperature, centrifugation and addition of stabilizers, these sampling time up to 3 minutes after tourniquet fastening (e.g.
biological factors have considerable effects, which are often location of the appropriate venous access, collection of many
underestimated (9). blood samples,…) (20, 21).

The venous stasis that results from extended tourniquet


Specimen types placement may influence the plasma concentration of a wide
Several studies have reported that differences between ranges of analytes depending on the length of stasis, size and
blood glucose measurements vary according to sample type, protein-binding characteristics of the analytes. Lippi et al. in-
methodology and timing of collection (10-13). Glucose can be vestigated the influence of 1- and 3-min venous stasis on the
measured in different matrices such as whole blood, serum or measurement of 12 common analytes including glucose. This
plasma. Glucose measurements in plasma are recommended study showed that 1-min stasis had no significant effect on the
for the diagnosis of diabetes mellitus to avoid therapeutic plasma glucose concentration in contrast with the 3-min sta-
misjudgements (1, 2, 14). Blood for glucose analysis should be sis which caused a statistically significant decrease in plasma
drawn in the morning after an overnight fast which includes no glucose concentration. However, in scientific literature there
caloric intake for at least 8 hours, during which the individual is inconclusive data on the effects of venous stasis on routine
may consume water ad libitum (2). laboratory testing. Nevertheless, it is advisable that the tourni-
quet application time is less than one minute to minimize the
Proper sample processing after blood collection is im- possible effects of venous stasis (20, 21).
portant to ensure accurate measurement of plasma glucose.
This requires rapid separation of plasma after blood collection
(within minutes) (15). If this is not possible, the sample should Reporting
be collected into a tube containing a glycolytic inhibitor to at- Plasma and whole blood glucose are used interchangeably
tenuate the rate of glycolysis. Because the commonly used gly- despite a physiological difference in glucose concentration
colytic inhibitors (e.g. sodium fluoride) have a delayed activity, (plasma > whole blood). This may lead to a clinical misinterpre-
a tube containing a glycolytic inhibitor should be placed in ice- tation of the glucose result (1, 22) due to different reference
water immediately after blood collection and the separation intervals and clinical decision limits (2).
should take place within 30 minutes (8, 15).
In order to avoid therapeutic misjudgements, the Interna-
Glucose concentrations may also differ depending on the tional Federation of Clinical Chemistry and Laboratory Medi-
blood sampling site (venous, arterial or capillary) especially cine (IFCC) recommends reporting the glucose concentra-
when the patient is not fasting. Fasting values for venous and tion in plasma (unit: mmol/L). Regardless of the sample type
capillary plasma glucose are almost identical. But during an or measurement principle, plasma glucose better reflects the
OGTT, capillary glucose concentrations are significantly higher relevant extracellular glucose concentration that cells are pre-
than those in venous blood (on average 1.7 mmol/L or 30 mg/ sented to. Moreover, plasma glucose is not influenced by hae-
dL) (16, 17). The difference between capillary and venous blood matocrit. This recommendation is also valid for POCT devices
glucose measurements can be explained by the rate of glucose that measure glucose in whole blood (1, 23).
consumption in the tissues (10).

Acta Clinica Belgica, 2010; 65-5


BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE 313

The IFCC advises the use of a conversion factor of 1.11 for tion in contrast with the 5% decline in healthy adults when
the conversion of a glucose concentration in whole blood to a blood specimens are stored at room temperature (30). In ad-
plasma equivalent. This adjustment factor is only applicable for dition, as much as 68% of glucose can be consumed in blood
samples from the same sampling site and in the case of normal samples collected without glycolytic inhibitors from neonates
haematocrit values (1, 23, 24). It does not apply to the calcu- and stored at room temperature for 5 hours as a result of the
lation of venous plasma glucose concentrations from glucose high haematocrit (30).
concentrations in capillary or arterial blood (1, 23). In practice,
conversion factors used in the various POCT meters differ upon Sodium fluoride (NaF) and lithium iodoacetate have both
firm. been used in blood collection tubes to preserve glucose. These
reagents can be used alone or in combination with anticoagu-
lants such as EDTA, potassium oxalate, citrate or lithium hepa-
Use of glycolytic inhibitors rin (2). The nominal concentration of sodium fluoride used to
Plasma glucose levels are expected to be stable if plasma is inhibit glycolysis is 2.5 mg/mL of blood in combination with an
immediately separated from cells by centrifugation after blood anticoagulant such as Na2EDTA (1 mg/mL of blood) or potas-
collection since the enzymes that metabolize glucose are in the sium oxalate (2 mg/mL of blood) (7).
cellular fraction (7). However, this is difficult to achieve in rou-
tine clinical practice. Table 1 gives an overview of the influence
of different glucose stabilizers on the blood glucose concentra- Although NaF is capable of preserving glucose concentra-
tion. tion in blood, the rates of decline of glucose in the first hour
Blood samples should be kept at 0-4 °C or centrifuged im- after sampling in tubes with and without fluoride are virtually
mediately if whole blood is used for glucose analysis (25). In identical (28). Plasma glucose concentration decreases up to
separated, non-haemolysed, sterile serum without glycolytic 0.5 mmol/L (9 mg/dL) over a period of 2 to 4 hours after sam-
inhibitors, the glucose concentration is stable for 8 hours at ple collection in tubes containing NaF. After 4 hours only, the
25 °C and 72 hours at 4 °C (26). Since it is impractical to cen- glucose concentration will be stable for 72 hours at room tem-
trifuge all laboratory glucose samples immediately, the use of perature (28). Fluoride takes approximately 3-4 hours to stabi-
glycolytic inhibitors is necessary to attenuate the rate of gly- lize glucose levels because it inhibits enolase which is the ninth
colysis. Nevertheless, it should be emphasized that these glu- and penultimate step of the glycolytic pathway. The upstream
cose preservatives do not totally prevent glycolysis (27). enzymes in the glycolysis pathway still continue to metabolize
glucose until substrates are consumed (31).
Glucose concentrations decrease ex vivo with time in whole
blood as a result of glycolysis which leads to an unpredictable
The combination of mannose and NaF has been suggested
underestimation of the true glucose concentration. The rate of
to overcome this problem. Mannose is a competitive inhibitor
glycolysis is on average 5-7 % (~0.6 mmol/L or 10 mg/dL) per
of the first enzyme of the glycolytic pathway (hexokinase). The
hour (28) and is influenced by glucose concentration, tempera-
inhibitory effect of mannose is short lasting (at most 4 hours),
ture, white and red blood cell count, platelet count and other
but in the meantime NaF has become active. A mixture of so-
factors (29). Because of the instability of glucose in blood sam-
dium fluoride (2 mg/mL of blood) and mannose (2 mg/mL of
ples, well-defined and standardized pre-analytical conditions
blood) can be effective in minimizing the loss of glucose and
are essential to ensure that the measured blood glucose con-
achieving a better glucose stabilization (9).
centration reflects the real blood glucose in the patient.

In absence of glycolytic inhibitors, glucose concentration of Another approach to improve blood glucose preservation is
newborns can decrease up to 24% 1 hour after blood collec- the acidification of blood combined with the addition of NaF

Table 1: Influence of different glycolytic inhibitors on glucose concentration


Sample type and glycolytic inhibitor(s) Characteristics
Whole blood Rate of glycolysis is 5-7% on average (~0.6 mmol/L or 10 mg/dL) per hour
Separated, non-haemolysed, sterile serum without glycolytic inhibitors Stable glucose concentration for 8 hours at 25°C and for 72 hours at 4°C
Sodium fluoride (NaF) (2.5 mg/mL of blood) with EDTA (1 mg/dL of blood) or Glucose concentration decreases up to 0.5 mmol/L 2 to 4 hours after collection
potassium oxalate (2 mg/dL of blood) After 4 hours: stable glucose concentration for 72 hours during storage at RT

A mixture of NaF (2 mg/mL of blood) and mannose (2 mg/mL of blood) Effective in minimizing the loss of glucose and achieving a better glucose
stabilization in comparison with NaF alone
Acidification of blood combined with the addition of NaF and EDTA Rapid inhibition of the glycolysis and more effective than NaF alone
A mixture of 119 mmol/L NaF and 11 mmol/L glyceraldehyde (plus 21.7 Prevents consumption of glucose completely even after incubation for 24 hours at
mmol/L potassium oxalate) RT prior to centrifugation
Lithium iodoacetate (0.5 mg/mL blood) or in combination with lithium Optimal working 3 hours after blood sampling
heparin (14.3 U/mL blood).
RT= room temperature

Acta Clinica Belgica, 2010; 65-5


314 BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE

and EDTA. Acidification of blood with a citrate buffer inhibits Generally, haemolysis has little effect on analytes that are
hexokinase and phosphofructokinase, early enzymes in the gly- at lower concentrations in erythrocytes than in plasma, but it
colytic pathway, which quickly stops glycolysis. The inhibitory may have a substantial effect on constituents that are present
effect of acidification is sustainable for approximately 10 hours at a much higher concentration within the red blood cells than
at 25 °C which can easily be prolonged by the addition of a in plasma. Therefore, it is important that doctors are aware of
small amount of NaF (31). the possible effects of haemolysis on the interpretation of cer-
An alternative strategy for efficient glycolytic inhibition is tain laboratory tests (7, 33).
the use of glyceraldehyde in addition to fluoride. le Roux et al.
demonstrated that the best glycostatic agent is a combination The effect of haemolysis on blood glucose analysis is unpre-
of 11 mmol/L glyceraldehyde and 119 mmol/L NaF (plus 21.7 dictable due to variations in methodology and due to the mul-
mmol/L potassium oxalate as anticoagulant). This combination tiplicity of instrumentation. Nevertheless, haemolysis should
prevents consumption of glucose completely even after incu- be avoided or at least not exceed levels that would make hae-
bation for 24 hours at room temperature prior to centrifuga- molysis visually detectable in serum or plasma (7, 33).
tion (25).
2) Icodextrin interference
Besides NaF, lithium iodoacetate is frequently used as a Most glucose dehydrogenase based methods can give false-
glucose preservative. It inhibits the fifth enzyme in the glyco- ly elevated glucose determinations in presence of icodextrin
lytic pathway (glyceraldehyde-3-phosphate dehydrogenase) metabolites. Icodextrin is a glucose polymer that is used world-
and can be used alone (0.5 mg/mL blood) or in combination wide as an osmotic agent in continuous ambulatory peritoneal
with lithium heparin (14.3 U/mL blood). Lithium iodoacetate dialysis and which can reach the blood circulation via the lym-
works optimal 3 hours after blood sampling (28, 32). phatic system (34, 35). In the systemic circulation, icodextrin
is hydrolyzed to maltose, maltotriose, maltotetraose and other
oligosaccharides. These icodextrin metabolites, principally
Interferences maltose, may act as a substrate for glucose dehydrogenase
Interferences (endogenous and exogenous) are factors which may lead to spuriously high glucose results depending
which may have an influence on the clinical interpretation of on the specificity of the enzyme used. In contrast, glucose mea-
laboratory data or on the measurement of an analyte. Endoge- surements based on glucose oxidase and hexokinase systems
nous interferences are caused by substances that are present in are not influenced by icodextrin (34).
a patient by nature such as turbidity due to hyperlipemia, hae-
molysis,… Exogenous interferences are substances which are 3) Influence of oxygen
present in the blood collection tubes such as anticoagulants, The oxygen dependency of glucose oxidase based glucose
stabilizing additives or substances which were administered to meters may cause inaccurate glucose measurements when ox-
the patient (e.g. drugs) (6). ygen tension varies. Glucose measurements based on glucose
dehydrogenase methods are oxygen-insensitive (36).
1) Influence of haemolysis
Haemolysis has been defined as the release of intracellular 4) Influence of certain drugs
components from erythrocytes, thrombocytes and leukocytes Ascorbic acid, acetaminophen, dopamine and mannitol can
into the extracellular fluid (i.e. the plasma or serum). It can oc- interfere with glucose measurements performed with handheld
cur in-vivo as well as in-vitro during the different steps of the glucose meters. These drugs are frequently used for the treat-
pre-analytical phase. Haemolysis can be caused by inappropri- ment of critically ill patients which can mask the presence of
ate blood sampling or by errors during transport, handling of hypoglycaemia or hyperglycaemia and may lead to inappropri-
specimens and storage (7, 33). ate insulin therapy (37).

Table 2: Advantages and disadvantages of POCT glucose measurements


POCT glucose measurements
Advantages Disadvantages
Minimal sample volume (1-20 μL) Adequate training of patients and medical staff
Rapid turnaround times (<30 s) Possible interferences
• Icodextrin
• Oxygen
• Drugs (ascorbic acid, dopamine, mannitol, …)
Limited pre-analytical phase Analytical performance
• No ex vivo glycolysis • No consensus about the analytical goals
• No transport • Inaccurate at very low and high glucose levels
• No centrifugation • Influence of hematocrit
Calibration
• Plasma or whole blood
• Conversion factors differ upon firm
CL= central laboratory

Acta Clinica Belgica, 2010; 65-5


BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE 315

CENTRAL LABORATORY VERSUS POCT consensus. The widely divergent criteria include expert opinion,
opinion of clinicians, legislation, state of the art and biological
Blood glucose analysis can be performed in a central labo- variation.
ratory or achieved by use of POCT glucose meters. However,
the imprecision of the current POCT glucose devices precludes In 1987, the ADA recommended for self-monitoring of
their use as a diagnostic tool for diabetes mellitus (2). Table 2 blood glucose a maximum allowable CV of 10% at glucose
highlights the advantages and disadvantages of POCT glucose concentrations of 1.7-22.2 mmol/L (30-400 mg/dL) and a max-
measurements. imum bias of 15% from a reference method value (41, 42). In
1996, the ADA revised the performance goal to 5% (43).
One of the major advantages of POCT glucose meters is their
practicality with rapid turnaround times because of the limited The Clinical Laboratory Improvement Amendments (CLIA)
pre-analytical phase (no transport, no centrifugation). Moreover ’88 goal is less stringent in comparison with the ADA per-
they require very small sample volumes (1-20 μL) (38). formance criteria. Glucose results achieved by POCT devices
should be within 10% of target values or ± 0.3 mmol/L (6 mg/
POCT also requires an adequate training of patients and dL) (2).
medical staff. Skeie et al. compared the analytical perfor-
mance of five POCT glucose meters obtained by patients and The National Committee on Clinical Laboratory Standards
by medical laboratory technicians. These authors found that (NCCLS) guidelines stipulate that POCT glucose results should
the imprecision of the evaluated systems in the hands of pa- be within ± 20% of laboratory glucose at >5.5 mmol/L (100
tients varied from 7 to 20 %, while the imprecision varied from mg/dL) and ± 0.83 mmol/L (15 mg/dL) of laboratory glucose
2.5-5.9 % when the meters were used by lab technicians (39). if the glucose concentration is ≤5.5 mmol/L (100 mg/dL) (44).
Furthermore, none of the glucose meters met the published
performance specifications proposed by the American Diabe- The new Clinical Laboratory Standards Institute (CLSI)
tes Association (ADA) (± 10%). The performance was better recommendations determine that for glucose concentrations
when compared with the quality specifications of the Inter- above 4.2 mmol/L (75 mg/dL) the discrepancy between POCT
national Organization for Standardization (ISO) which requires devices and the central laboratory should be < 20%. For a glu-
that 95% of glucose measurements lie between the target ± cose concentration ≤ 4.2 mmol/L (75 mg/dL), the discrepancy
20%. In this study, only 86% of the blood glucose measure- should not exceed 0.83 mmol/L (15 mg/dL) (45).
ments performed by patients fulfilled the ISO standard. The The National Academy of Clinical Biochemistry (NACB) has
authors stress the critical role of education in ensuring good proposed quality criteria for laboratory glucose analysis based
performance of POCT glucose meters (39). As far as training on intra-individual biological variation (46) with a maximum
is concerned nursing staff should be included. Adequate train- allowable imprecision (CV) of ≤ 3.3% and a bias of ≤ 2.5% re-
ing of nursing staff is very difficult to organize because of high sulting in a total error of ≤ 7.9% (2). However, the systematic
turnover. difference of 11% between whole blood and plasma already
exceeds the recommended allowable maximum error of 7.9%
In practice, POCT glucose measurements often do not ful- by which incorrect specimen type may lead to a misinterpreta-
fil the quality goals. Bias is mainly due to the use of different tion of the glucose result and a wrong diagnosis (1).
standardization and calibration techniques (whole blood versus
plasma) and great lot-to-lot variability of the different glucose
strips in comparison with the reagents used in the central lab
Impact of pre-analytical features on POCT glucose
(24, 39, 40). measurements
POCT glucose meters are used to assist patients with dia-
It is important that POCT results are integrated into a betes in achieving and maintaining blood glucose concentra-
laboratory information system which provides reliable glucose tions as close to those found in non-diabetic individuals. These
results through the use of quality controls and automatic cali- devices are also used in hospital settings as an alternative for
bration. Several manufacturers have introduced POCT glucose blood glucose analysis in the central laboratory.
meters with data management software which allow quali-
fied laboratory personnel to monitor instrument performance, Most of these POCT glucose systems measure the glucose
quality control and patient results. concentration directly in the plasma component of the blood
by filtering out the red blood cells. The obtained signal is cal-
ibrated to produce a result either as whole blood or plasma
Analytical performance of POCT glucose meters glucose. Correct calibration of POCT glucose devices is one of
Nowadays POCT glucose devices are used for taking ther- the most important pre-analytical factors. Many POCT glucose
apeutically important decisions. Therefore it is essential that meters are still calibrated to whole blood despite the IFCC rec-
these POCT glucose meters have an acceptable analytical per- ommendation that all glucose measuring devices should report
formance over the practically relevant range of concentrations in plasma values by which the difference between the various
and that their glucose results correlate well with those of labo- types of POCT glucose meters becomes even larger (1, 23).
ratory analyzers (18).
POCT glucose devices are also inaccurate at very low and
Numerous analytical goals have already been proposed for high glucose levels. A study of Khan et al. showed that signifi-
the performance of glucose meters but currently there is no cant biases occurred in the low and high glucose range between

Acta Clinica Belgica, 2010; 65-5


316 BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE

a POCT glucose measurement and a glucose determination in Recently, the multinational Normoglycemia in Intensive
the central lab (18). In the hypo- and hyperglycaemic range, Care Evaluation-Survival Using Glucose Algorithm Regulation
glucose results differed > 10% in more than 50% of the cases. (NICE-SUGAR) trial questioned the findings of Van den Berghe
In the hypoglycaemic range, differences > 20% occurred in et al. The NICE-SUGAR study showed an absolute increase in
57% of all cases (18). Therefore it is necessary to draw a blood death at 90 days and reported higher rates of severe hypogly-
sample for the central laboratory whenever a hypoglycaemic or cemia in the intensive-control group as compared to the con-
hyperglycaemic reading is observed from point-of-care glucose ventional-control group (51). Stringent glycaemic control in ICU
testing. patients remains controversial and further studies are necessary
to unravel the goals and benefits of tight glycaemic control.
Moreover, haematocrit differences can significantly affect In many critical care settings blood glucose is measured us-
POCT glucose measurements. Haematocrit effects are both ing capillary blood and POCT glucose analyzers. Critchell et al.
glucose and system dependent. At low haematocrit levels, most compared fingerstick glucose measurements to simultaneously
POCT systems yield spurious higher glucose measurements sampled laboratory glucose in ICU patients. This study showed
and conversely high haematocrit values are known to falsely that capillary blood glucose measured by fingerstick tend to
decrease POCT glucose measurements. Clinicians should be overestimate the actual blood glucose level in critically ill ICU
aware of the effect of haematocrit levels on the performance patients and does not meet the CLSI standard. Therefore, epi-
of POCT glucose devices as anaemia, polycythaemia and unex- sodes of hypoglycaemia may be missed in ICU patients on tight
pected changes in haematocrit values are common in a hospi- glycaemic control. This finding is important because the usual
tal setting (47). clinical warning signs and symptoms of hypoglycaemia in ICU
patients can be masked by sedation and critical illness. Finger-
stick glucose measurements in critically ill patients should be
Use of POCT glucose meters in an outpatient setting interpreted with caution as this type of glucose measurements
Self-monitoring of blood glucose (SMBG) is an impor- may result in episodes of undetected hypoglycaemia and inap-
tant component of diabetes therapy in order to achieve and propriate adjustments of insulin dose (52).
maintain specific glycaemic goals. Therefore, diabetic patients
should be educated in the correct use of POCT glucose meters
Impact of quality specifications for glucose meters
including quality control. Patients should also be instructed in
how to use the SMBG results to determine appropriate insulin on the adjustment of insulin doses
doses at different times of the day. Comparison between SMBG Clinicians and patients with diabetes both use POCT glu-
and concurrent laboratory glucose analysis should be per- cose meters to monitor blood glucose in order to improve clini-
formed at regular intervals to evaluate the accuracy of POCT cal outcomes. Therefore good analytical performance is crucial.
glucose results. Furthermore, a comparison between the mean Current quality specifications for glucose meters allow results
POCT glucose concentration and the HbA1c value can be used with an error by 5-10% or more in comparison with the true
as an internal control measure. In contrast to a hospital setting, glucose concentration.
SMBG in ambulatory patients with diabetes is not susceptible
to errors due to haematocrit differences (2). The simulation modelling study of Boyd and Bruns exam-
ined the quantitative effect of meter error on the ability to
identify the insulin dose appropriate for the true glucose con-
Use of POCT glucose meters in critical care settings centration. Dosage errors occurred in 8-23% of insulin doses
Accurate blood glucose measurements are essential to di- when POCT devices with a total analytical error of 5% were
agnose critically ill patients and to guide algorithmic insulin used. With a 10% total error criterion, 16-45% of doses were
regimens. Use of POCT glucose analyzers in a intensive care not correct. The error rates for insulin dose errors of ≥ 4 units
unit (ICU) remains controversial as fingerstick glucose mea- with either the 5% or the 10% total error criterion were less
surements are inaccurate in those patients. than 0.2% (53).

The single-centre study of Van den Berghe et al. in 2001 This study demonstrated that POCT glucose meters that
showed that intensive insulin therapy to maintain blood glu- fulfil the current quality specifications allow a large fraction of
cose levels at or below 110 mg/dL reduced morbidity and mor- administered insulin doses to differ from the intended doses.
tality among critically ill patients in the surgical intensive care Similar simulation studies have shown that glucose meters that
unit regardless of whether they had a history of diabetes. In achieve both a CV and bias <5-6% rarely lead to two-category
this study, blood glucose was frequently measured on whole errors (≥4 units) concerning insulin doses. CV as well as bias
undiluted arterial blood using a blood gas analyzer in order to must be <1-1.5% to keep rates of smaller errors (≤2 units) be-
maintain normoglycaemia (target glucose 80-110 mg/dL) and low 5% (53).
to prevent adverse events related to hypoglycaemia (48).

Based on this trial, many attempts have been made to CONCLUSION


achieve tight glycaemic control in critically ill patients with
varying results. There are raising concerns about the poten- Blood glucose analysis requires careful attention in order to
tially increased risk of severe hypoglycemic episodes with more ensure accurate glucose results. A number of important consid-
stringent glycaemic protocols and debate is ongoing regarding erations may influence glucose measurements. Blood should be
the most appropriate target level for blood glucose (49, 50). drawn preferably in the morning after an overnight fast. Proper

Acta Clinica Belgica, 2010; 65-5


BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE 317

sample processing after blood collection is essential. Blood Verlag GmbH. 2003: 34-5.
samples should be collected into a tube containing a glycolytic 10. Kuwa K, Nakayama T, Hoshino T, Tominaga M. Relationships of glucose
concentrations in capillary whole blood, venous whole blood and ve-
inhibitor when rapid separation of the cells is not possible.
nous plasma. Clin Chim Acta 2001; 307: 187-92.
11. Colagiuri S, Sandbæk A, Carstensen B, et al. Comparability of venous
Glucose concentrations may also differ depending on the and capillary glucose measurements in blood. Diabet Med 2003; 20:
blood sampling site (venous, arterial or capillary blood). In the 953-6.
fasting state, the glucose concentration in venous and capillary 12. Neely RDG, Kiwanuka JB, Hadden DR. Influence of sample type on the
whole blood is almost indistinguishable. During an OGTT, capil- interpretation of the oral glucose tolerance test for gestational diabe-
lary glucose concentrations are significant higher than venous tes mellitus. Diabet Med 1991; 8: 129-34.
13. Farrer M, Fulcher G, Albers CJ, Neil HA, Adams PC, Alberti KG. Patients
glucose values. undergoing coronary artery bypass graft surgery are at high risk of im-
paired glucose tolerance and diabetes mellitus during the first postop-
Plasma and whole blood glucose values are not inter- erative year. Metabolism 1995; 44: 1016-27.
changeable. The IFCC recommends reporting the glucose con- 14. American Diabetes Association. Report of the expert committee on the
centration in plasma using the conversion factor of 1.11. This diagnosis and classification of diabetes mellitus. Diabetes Care 1997;
conversion provides consistent harmonized results which facili- 20: 1183-97.
15. World Health Organisation. Definition and diagnosis of diabetes mel-
tates the classification of patients and leads to fewer therapeu-
litus and intermediate hyperglycaemia. Report of a WHO/IDF consul-
tic misjudgements. This conversion factor is only applicable in tation. 2006; 1-46.
case of normal haematocrit values. 16. Larsson-Cohn U. Differences between capillary and venous blood glu-
cose during oral glucose tolerance tests. Scand J Clin Lab Invest 1976;
POCT plays an important role in monitoring blood glucose 36: 805-8.
levels in diabetic patients and remains a time-efficient method 17. Lind T, De Groot HA, Brown G, Cheyne GA. Observations on blood glu-
cose and insulin determinations. BMJ 1972; 3: 320-3.
of measuring glucose in acute situations. However, one should
18. Khan AI, Vasquez Y, Gray J, Wians FH, Kroll MH. The variability of results
take into account that the reliability of POCT glucose measure- between Point-Of-Care testing glucose meters and the central labora-
ments depends on a variety of factors including underlying dis- tory analyzer. Arch Pathol Lab Med 2006; 130: 1527-32.
ease, patient drug regimens and interfering substances as well 19. Stahl M, Brandslund I, Jørgensen LGM, Hyltoft Petersen P, Borch-
as instrument analytical performance and user proficiency. Johnsen K, De Fine Olivarius N. Can capillary whole blood glucose and
venous plasma glucose measurements be used interchangeably in di-
Fingerstick measurements should be interpreted with great agnosis of diabetes mellitus? Scand J Clin Lab Invest 2002; 62: 159-66.
20. Effects of posture and tourniquet. In: Guder WG, Narayanan S, Wisser H,
caution in protocols of tight glycemic control because use of Zawta B, eds. Samples: From the patient to the laboratory: The impact
POCT glucose meters in the ICU may lead to spurious glucose of preanalytical variables on the quality of laboratory results. 3rd revised
results and misdiagnosis. It is recommended to perform a labo- edition. Darmstadt, Germany: Wiley-VCH Verlag GmbH. 2003: 19.
ratory blood glucose analysis if the POCT glucose value is in the 21. Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi GC. Influence of
critical hypoglycaemic or hyperglycaemic range. short-term venous stasis on clinical chemistry testing. Clin Chem Lab
Med 2005; 43: 869-75.
22. Burrin JM, Alberti KGMM. What is blood glucose: can it be measured?
Diabet Med 1990; 7: 199-206.
REFERENCES 23. D’Orazio P, Burnett RW, Fogh-Andersen N et al. Approved IFCC recom-
mendation on reporting results for blood glucose. Clin Chem Lab Med
1. D’Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recom- 2006; 44: 1486-90.
mendation on reporting results for blood glucose (abbreviated). Clin 24. Floré KM, Fiers T, Delanghe JR. Critical evaluation of connectivity-based
Chem 2005; 51: 1573-6. point of care testing systems of glucose in a hospital environment. Clin
2. Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Chem Lab Med 2008; 46: 1763-8.
Guidelines and recommendations for laboratory analysis in the diagno- 25. le Roux CW, Wilkinson SD, Pavitt DV, Muller BR, Alaghband-Zadeh J. A
sis and management of diabetes mellitus. Clin Chem 2002; 48: 436-72. new antiglycolytic agent. Ann Clin Biochem 2004; 41: 43-6.
3. Young DS. Conveying the importance of the preanalytical phase. Clin 26. Sacks DB. Carbohydrates. In: Burtis C, Ashwood E, eds. Tietz textbook
Chem Lab Med 2003; 41: 884-7. of clinical chemistry, 3rd ed. Philadelphia: WB Saunders. 1999: 750-808.
4. Kalra J. Medical errors: impact on clinical laboratories and other critical 27. World Health Organisation. Definition, diagnosis and classification of
areas. Clin Biochem 2004; 37: 1052-62. diabetes mellitus and its complications. Report of a WHO consultation.
5. Lippi G, Guidi GC, Mattiuzzi C, Plebani M. Preanalytical variability: the Part 1: Diagnosis and classification of diabetes mellitus. Geneva: World
dark side of the moon in laboratory testing. Clin Chem Lab Med 2006; Health Organization Department of Non-communicable Disease Sur-
44: 358-65. veillance, 1999; 1-59.
6. Berth M, Bafort K, Bosmans E, et al. Foutieve resultaten van veneuze 28. Chan AYW, Swaminathan R, Cockram CS. Effectiveness of sodium fluo-
bloedafname in het klinisch laboratorium. Tijdschrift voor Geneeskunde ride as a preservative of glucose in blood. Clin Chem 1989; 35: 315-7.
2007; 63: 797-803. 29. Ladenson JH. Nonanalytical sources of variation clinical chemistry re-
7. Narayanan S. The preanalytical phase: an important component of sults. In: Sonnenwirth A, Jarett L, eds. Clinical laboratory methods and
laboratory medicine. Am J Clin Pathol 2000; 113: 429-52. diagnosis. St. Louis, MO: CV Mosby. 1980: 149-92.
8. Stahl M , Jørgensen LGM, Hyltoft Petersen P, Brandslund I, De Fine Oli- 30. Meites S, Saniel-Banrey K. Preservation, distribution and assay of glu-
varius N, K. Borch-Johnsen. Optimization of preanalytical conditions cose in blood, with special reference to the newborn. Clin Chem 1979;
and analysis of plasma glucose. Impact of the new WHO and ADA rec- 25: 531-4.
ommendations on diagnosis of diabetes mellitus. Scan J Clin Lab Invest 31. Gambino R, Piscitelli J, Ackattupathil TA, et al. Acidification of blood
2001; 61: 169-80. is superior to sodium fluoride alone as an inhibitor of glycolysis. Clin
9. Take a lavender tube! Additives and colour codes. In: Guder WG, Na- Chem 2009; 55: 1019-21
rayanan S, Wisser H, Zawta B, eds. Samples: From the patient to the 32. Chan AYW, Cockram CS, Swaminthan R. Effect of delay in separating
laboratory: The impact of preanalytical variables on the quality of plasma for glucose measurement upon the interpretation of oral glu-
laboratory results. 3rd revised edition. Darmstadt, Germany: Wiley-VCH cose tolerance tests. Ann Clin Biochem 1990; 24: 73-4.

Acta Clinica Belgica, 2010; 65-5


318 BLOOD GLUCOSE ANALYSIS AND THE PRE-ANALYTICAL PHASE

33. The serum sample looks reddish. Effects of haemolysis. In: Guder WG, 43. American Diabetes Association. Self-monitoring of blood glucose. Dia-
Narayanan S, Wisser H, Zawta B, eds. Samples: From the patient to betes Care 1996; 19(Suppl 1): S62-6.
the laboratory: The impact of preanalytical variables on the quality of 44. National Committee for Clinical Laboratory Standards. Ancillary (bed-
laboratory results. 3rd revised edition. Darmstadt, Germany: Wiley-VCH side) blood glucose testing in acute and chronic care facilities; ap-
Verlag GmbH. 2003: 80-1. proved guideline C30-A. Villanova, PA: NCCLS, 1994; 14: 1-14.
34. Janssen W, Harff G, Caers M, Schellekens A. Positive interference of 45. CLSI. Point-of-care blood glucose testing in acute and chronic care
icodxtrin metabolites in some enzymatic glucose methods. Clin Chem facilities; Approved guideline. Second edition. 940 West Valley Road,
1998; 44: 2379-80. Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2002. CLSI docu-
35. Apperloo JJ, Vader HL. A quantitative appraisal of interference by ico- ment C30-A2.
dextrin metabolites in point-of-care glucose analyses. Clin Chem Lab 46. Ricos C, Alvarez V, Cava F, et al. Current databases on biological varia-
Med 2005;43:314-8. tion: pros, cons and progress. Scand J Clin Lab Invest 1999; 59: 491-500.
36. Kost GJ, Vu HT, Lee JH, et al. Multicenter study of oxygen-insensitive 47. Tang Z, Lee JH, Louie RF, Kost GJ. Effects of different hematocrit lev-
handheld glucose point-of-care testing in critical care/hospital/ambu- els on glucose measurements with handheld meters for point-of-care
latory patients in the United States and Canada. Crit Care Med 1998; testing. Arch Pathol Lab Med 2000; 124: 1135-40.
26: 581-90. 48. Van Den Berghe G, Wouters P, Weekers F, et al. Intensive insulin therapy
37. Tang Z, Du X, Louie RF, Kost GJ. Effects of drugs on glucose measure- in critically ill patients. N Engl J Med 2001; 345: 1359-67.
ments with handheld glucose meters and a portable glucose analyzer. 49. Meijering S, Corstjens AM, Tulleken JE, Meertens JH, Zijlstra JG, Ligten-
Am J Clin Pathol 2000; 113: 75-86. berg JJ. Towards a feasible algorithm for tight glycaemic control in criti-
38. Winkelman JW, Wybenga DR, Tanasijevic MJ. The fiscal consequences cally ill patients: a systematic review of the literature. Crit Care 2006;
of central vs distributed testing of glucose. Clin Chem 1994; 40: 1628- 10: R19.
30. 50. Brunkhorst FM, Engel C, Bloos F, et al. Intensive insulin therapy and
39. Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring pentastarch resuscitation in severe sepsis. N Engl J Med 2008; 358: 125-
of blood glucose: comparison of testing quality achieved by patients 39.
and a technician. Clin Chem 2002; 48: 994-1003. 51. Inzucchi SE, Siegel MD. Glucose control in the ICU-- How tight is too
40. Plessis du M, Ubbink JB, Vermaak WJH. Analytical quality of near-pa- tight? N Engl J Med 2009; 360: 1346-9.
tient blood cholesterol and glucose measurement. Clin Chem 2000; 46: 52. Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marnik P. Ac-
1085-90. curacy of bedside capillary blood glucose measurements in critically ill
41. American Diabetes Association. Consensus statement: self-monitoring patients. Intensive Care Med 2007; 33: 2079-84.
of blood glucose. Diabetes Care 1990; 13(Suppl 1): 41-6. 53. Boyd JC, Bruns DE. Quality specifications for glucose meters: assess-
42. American Diabetes Association. Self-monitoring of blood glucose. Dia- ment by simulation modeling of errors in insulin dose. Clin Chem 2001;
betes Care 1994; 17: 81-6. 47: 209-14.

Acta Clinica Belgica, 2010; 65-5

Potrebbero piacerti anche