Sei sulla pagina 1di 26

Special Article

Guidelines for the use of an insulin infusion for the management


of hyperglycemia in critically ill patients
Judith Jacobi, PharmD, FCCM; Nicholas Bircher, MD, FCCM; James Krinsley, MD, FCCM;
Michael Agus, MD; Susan S. Braithwaite, MD; Clifford Deutschman, MD, FCCM;
Amado X. Freire, MD, MPH, FCCM; Douglas Geehan, MD, FCCM; Benjamin Kohl, MD, FCCM;
Stanley A. Nasraway, MD, FCCM; Mark Rigby, MD, PhD, FCCM; Karen Sands, APRN-BC, ANP, MSN, CCRN;
Lynn Schallom, RN, MSN, CCNS, FCCM; Beth Taylor, MS, RD, CNSD, FCCM; Guillermo Umpierrez, MD;
John Mazuski, MD, PhD, FCCM; Holger Schunemann, MD

Objective: To evaluate the literature and identify important monitoring system be designed to avoid and detect hypoglycemia
aspects of insulin therapy that facilitate safe and effective infusion (blood glucose ≤70 mg/dL) and to minimize glycemic variability.
therapy for a defined glycemic end point. Important processes of care for insulin therapy include use of a
Methods: Where available, the literature was evaluated using reliable insulin infusion protocol, frequent blood glucose monitor-
Grades of Recommendation, Assessment, Development, and Evalu- ing, and avoidance of finger-stick glucose testing through the use
ation (GRADE) methodology to assess the impact of insulin infu- of arterial or venous glucose samples. The essential components of
sions on outcome for general intensive care unit patients and an insulin infusion system include use of a validated insulin titration
those in specific subsets of neurologic injury, traumatic injury, and program, availability of appropriate staffing resources, accurate mon-
cardiovascular surgery. Elements that contribute to safe and effec- itoring technology, and standardized approaches to infusion prepa-
tive insulin infusion therapy were determined through literature ration, provision of consistent carbohydrate calories and nutritional
review and expert opinion. The majority of the literature supporting support, and dextrose replacement for hypoglycemia prevention and
the use of insulin infusion therapy for critically ill patients lacks treatment. Quality improvement of glycemic management programs
adequate strength to support more than weak recommendations, should include analysis of hypoglycemia rates, run charts of glucose
termed suggestions, such that the difference between desirable values <150 and 180 mg/dL. The literature is inadequate to support
and undesirable effect of a given intervention is not always clear. recommendations regarding glycemic control in pediatric patients.
Recommendations: The article is focused on a suggested glyce- Conclusions: While the benefits of tight glycemic control have
mic control end point such that a blood glucose ≥150 mg/dL trig- not been definitive, there are patients who will receive insulin infu-
gers interventions to maintain blood glucose below that level and sion therapy, and the suggestions in this article provide the struc-
absolutely <180 mg/dL. There is a slight reduction in mortality with ture for safe and effective use of this therapy. (Crit Care Med 2012;
this treatment end point for general intensive care unit patients 40:3251–3276)
and reductions in morbidity for perioperative patients, postopera- Key Words: critical care; glycemic control; glucose meter;
tive cardiac surgery patients, post-traumatic injury patients, and glucose monitoring; guideline; hyperglycemia; insulin; protocol;
neurologic injury patients. We suggest that the insulin regimen and stress hyperglycemia

From the Indiana University Health Methodist Supplemental digital content is available for this Glycemic Control. The remaining authors have not dis-
Hospital, Indianapolis, IN (JJ); University of Pittsburgh article. Direct URL citations appear in the printed text closed any potential conflicts of interest.
Medical Center, Pittsburgh, PA (NB); Stamford Hospital, and are provided in the HTML and PDF versions of this Listen to the iCritical Care podcasts for an
Stamford, CT (JK); Intermediate Care Unit, Children’s article on the journal’s Web site (http://journals.lww. in-depth interview on this article. Visit www.sccm.
Hospital Boston, Boston, MA (MA); Endocrine Consults com/ccmjournal). org/iCriticalCare or search “SCCM” at iTunes.
and Care, Evanston, IL (SSB); Critical Care and Surgery, Dr. Agus has consulted for the Diabetes Technology For information regarding this article, E-mail:
Perelman School of Medicine at the University of Society. He also has a pending pat­ent on an ECMO- jjacobi@iuhealth.org
Pennsylvania, Philadelphia, PA (CD); PCC and Sleep based glucose, sensor, which is not connected to idea Copyright © 2012 by the Society of Critical Care
Medicine, University of Tennessee Health Science discussed in this article. Dr. Braithwaite has a U.S. pat- Medicine and Lippincott Williams and Wilkins
Center, Memphis, TN (AXF); University of Missouri ent. Dr. Kohl has received grant support from Amylin and The American College of Critical Care Medicine
Kansas City, Kansas City, MO (DG); University of Eli Lilly. Dr. Krinsley has performed consulting work for (ACCM), which honors individuals for their achievements
Pennsylvania, Philadelphia, PA (BK); SICU, Tufts/New Medtronic Inc., Edwards Life Sciences, Baxter, Roche and contributions to multidisciplinary critical care medi-
England Medical Center, Boston, MA (SAN); Pediatric Diagnostics, and Optiscan Biomedical and has received cine, is the consultative body of the Society of Critical
Critical Care, Riley Hospital For Children, Indianapolis, IN speaker’s fees from Edwards Life Sciences, Roche Care Medicine (SCCM) that possesses recognized exper-
(MR); Novant Health, Winston Salem, NC (KS); Barnes- Diagnostics and Sanofi-Aventis. Dr. Nasraway has con- tise in the practice of critical care. The College has devel-
Jewish Hosptial, St. Louis, MO (LS); Barnes Jewish sulted for Optiscan, Echo Therapeutics. Dr. Geehan has oped administrative guidelines and clinical practice pa-
Hospital, St. Louis, MO (BT); Emory University School received grant support from the Department of Defense rameters for the critical care practitioner. New guidelines
of Medicine, Atlanta, GA (GU); Washington University Research. Dr. Rigby has received consulting fees from and practice parameters are continually developed, and
St. Louis, St. Louis, MO (JM); McMaster University, Medtronic. Dr. Schallom has received honoraria/speak- current ones are systematically reviewed and revised.
Hamilton, Ontario, Canada (HS). ing fees from Roche Laboratories Speakers Bureau on DOI: 10.1097/CCM.0b013e3182653269

Crit Care Med 2012 Vol. 40, No. 123251


T he notion of tight glycemic
control (GC) became more
prominent in the critical care
literature in 2001 when a land-
mark study by Van den Berghe and col-
leagues (1) demonstrated a significant
mortality benefit when maintaining blood
hypoglycemic drugs, and other antidia-
betic agents may be continued or restarted
in selected patients, but will not be dis-
cussed in this article. Studies evaluating
insulin as a component of other therapies
(such as glucose–insulin–potassium) were
unpublished studies or data were not
included in the analysis. The Grading of
Recommendations Assessment, Develop-
ment and Evaluation (GRADE) system
was used to rate the quality of evidence
and strength of the recommendation
for each clinical practice question (4).
not evaluated.
glucose (BG) between 80 and 110 mg/dL. A member of the GRADE group was avail-
Prior to that publication, GC was not a TARGET PATIENT POPULATION able to provide input and answer meth-
high priority in most intensive care unit FOR GUIDELINE odologic questions.
(ICU) patients. Data have confirmed the Meta-analyses using RevMan and
observation that hyperglycemia is associ- These guidelines are targeted to adult GRADEPro software were applied to orga-
ated with an increase in death and infec- medical and surgical ICU patients as a nize evidence tables, create forest and
tion, seemingly across the board among group, but individual population differ- funnel plots, and draw conclusions about
many case types in the ICU (2, 3). Many ences regarding therapy or monitoring the overall treatment effects or specific
centers have attempted to assess the fea- will be discussed. Data on the glycemic outcomes applicable to a particular rec-
sibility of maintaining normoglycemia management of pediatric ICU patients ommendation (5, 6).
in critically ill patients and to further are limited, but will be described where Recommendations are classified as
establish the potential risk or benefit of available. strong (Grade 1) or weak (Grade 2) and
this approach in a variety of ICU patient are focused on specific populations
subsets. While there have been conflicting METHODOLOGY where possible. Strong recommenda-
results from numerous studies, the ques- tions are listed as “recommendations”
The Guideline Task Force was com- and weak recommendations as “sugges-
tion is no longer whether GC is beneficial
posed of volunteers from the Society of tions.” Throughout the development of
or not, but rather what is the appropriate
Critical Care Medicine with a specific the guidelines, there was an emphasis on
degree of GC that can be accomplished
interest in the topic and the guideline patient safety and whether the benefit of
safely and with justifiable utilization of
process. The Task Force members devel- adherence to the recommendation would
resources.
oped a list of clinical questions regard- outweigh the potential risk, the burden
This Clinical Practice Guideline will
ing the appropriate utilization of insulin on staff, and when possible, the cost. If
evaluate the available literature and
infusions to achieve GC, considering the risk associated with an intervention
address aspects of implementation that
patient/populations, interventions, com- limited the potential for benefit, or if the
permit safe and effective insulin infusion
parisons, and outcomes. Applicable lit- literature was not strong, the statement
therapy. Methodology and assessment will
erature was compiled using a variety of was weakened to a suggestion. Individual
be emphasized to help clinicians achieve
search engines (PubMed, OVID, Google patient or ICU circumstances may influ-
the BG goal that is considered to have
Scholar, reference lists from other pub- ence the applicability of a recommenda-
the greatest benefit and safety for their
lications, search of Clinicaltrials.gov, tion. It is important to recognize that
patient population while avoiding clini-
cally significant hypoglycemia. and the expertise and experience of the strong recommendations do not neces-
authors). Searches were performed peri- sarily represent standards of care.
odically until the end of 2010 using the Numerous discussions among the
GUIDELINE LIMITATIONS
following terms: acute stroke, BG, car- authors led to consensus regarding the
Guidelines are limited by the available diac surgery, critical care, critical illness, recommendations. Individual members
literature and the expertise of the writing critically ill patients, dextrose, glucose, or subgroups drafted the recommenda-
panel and reviewers. The recommenda- glucose control, glucose metabolism, tions and justifications. Subsequently,
tions are not absolute requirements, and glucose meters, glucose toxicity, glyce- each recommendation was reviewed by
therapy should be tailored to individual mic control, glycemic variability, hyper- the Task Force members who were pro-
patients and the expertise and equipment glycemia, hypoglycemia, ICU, insulin, vided the opportunity to comment, pro-
available in a particular ICU. The use of insulin infusion, insulin protocols, insu- pose changes, and approve or disapprove
an insulin infusion requires an appro- lin resistance, insulin therapy, inten- each statement. Once compiled, each
priate protocol and point-of-care (POC) sive care, intensive insulin therapy, member was again asked to review the
monitoring equipment with frequent BG mortality, myocardial infarction, neu- article and provide input. Consensus was
monitoring to avoid hypoglycemia. Rec- rocognitive function, neuroprotection, sought for recommendation statements,
ommendations may not be applicable to outcomes, pediatric, pediatric intensive and controversial statements were repeat-
all ICU populations, and limitations will care, point-of-care, point-of-care testing, edly edited and feedback provided through
be discussed when applicable. Future lit- sepsis, sternal wound infection, stress secret ballots until there was consensus.
erature may alter the recommendations hyperglycemia, stress, stress hormones, Actual or potential conflicts of interest
and should be considered when applying stroke, subarachnoid hemorrhage, sur- were disclosed annually, and transpar-
the recommendations within this article. gery, tight glycemic control protocols, ency of discussion was essential. External
Intravenous (IV) insulin will be the pri- and traumatic brain injury (TBI). peer review was provided through the
mary therapy discussed, but subcutane- Published clinical trials were used as Critical Care Medicine editorial process,
ous (SQ) administration may also have a the primary support for guideline state- and approval was obtained by the govern-
role for GC in stable ICU patients. Other ments, with each study evaluated and ing board of the Society of Critical Care
agents and approaches, including oral given a level of evidence. Abstracts and Medicine.

3252 Crit Care Med 2012 Vol. 40, No. 12


RECOMMENDATIONS when BG >150 mg/dL (12). Insulin infu- interval (CI) [0.71, 0.99] (p = .04), but
sion therapy is recommended for most does not suggest an impact on ICU
While the initial goal was to suggest critically ill patients, although selected mortality OR 0.99, 95% CI [0.86, 1.15];
glycemic targets for critically ill patients, patients may be managed on SQ therapy (p = .92) (Fig. 1A and B). The data dem-
the limited available literature has nar- as discussed later in the article. onstrate a high level of heterogeneity,
rowed the scope of this article and the Several large randomized controlled I2 = 80%, that led to selection of the ran-
ability to make recommendations for spe- trials (RCTs) have addressed the impact dom-effects model for analysis. Sensitiv-
cific populations. An overriding focus is of GC on mortality with variable results, ity testing was performed excluding each
on the safe use of insulin infusions. The although the ability to compare results of the large randomized trials (1, 16),
glycemic goal range of 100–150 mg/dL is hampered by differing populations, but this did not substantially change the
is a consensus goal, and while it differs methodology, and end points (Table 1) (1, results (see Supplemental Digital Con-
slightly from the more stringent goal of 13–16). Small randomized trials, defined tent 1, http://links.lww.com/CCM/A589).
110–140 mg/dL for selected populations, as <1,000 patients, are also included Our selection of 150 mg/dL as a trig-
recently published by the American Dia- (17–22). Several large cohort trials have ger for intervention is a consensus deci-
betes Association, and the overall glucose also been reported, although use of remote sion to reflect the various treatment goals
goal of 140–180 mg/dL, this difference is historical controls, inconsistent or volun- reported in the literature. Using a higher
not likely to be clinically significant (7). tary utilization of insulin therapy and pro- trigger value could allow excursion of BG
tocols, and concurrent changes in clinical >180 mg/dL, which is undesirable with
1. In adult critically ill patients, does practice complicate the interpretation of respect to the immunosuppressive effects
achievement of a BG < 150 mg/dL with outcome (23–28). One small cohort trial and potential to exceed the renal thresh-
an insulin infusion reduce mortality, was also evaluated for the impact of insu- old for glucosuria. Our recommendation
compared with the use of an insulin lin therapy on patient outcome (29). is similar to the American Diabetes Asso-
infusion targeting higher BG ranges? Limitations in these trials are sig- ciation guidelines for initiation of insulin
nificant. Many are single-center trials, for a glucose threshold no higher than
We suggest that a BG ≥ 150 mg/dL and the influence of local practices (e.g., 180 mg/dL, and that a more stringent
should trigger initiation of insulin ther- nutrition, fluid therapy, available technol- goal of 110–140 mg/dL may be used if
apy, titrated to keep BG < 150 mg/dL for ogy, nursing expertise with insulin titra- there is a documented low rate of severe
most adult ICU patients and to maintain tion) cannot be adequately factored into hypoglycemia (7).
BG values absolutely <180 mg/dL using a the results. Several trials failed to achieve In contrast, there are at least three pub-
protocol that achieves a low rate of hypo- the glycemic target or had protocol viola- lished meta-analysis reviews published in
glycemia (BG ≤ 70 mg/dL) despite limited tions, or voluntary use of an insulin infu- the peer-review literature that have sug-
impact on patient mortality. sion protocol in the cohort studies might gested no significant mortality benefits
[Quality of evidence: very low] have biased the results. The small RCTs from insulin infusion therapy to maintain
Numerous reports have associated were inadequately powered to assess mor- “tight” GC (BG <150  mg/dL). The first
hyperglycemia with a poor patient out- tality. Most studies used glucose meters, review from Wiener et al (31) included
come (1–3, 8–11). Retrospective analysis and BG values were checked at varying abstracts and unpublished data, which we
of 259,040 admissions demonstrated a frequencies (30 mins–4 hrs), influenc- have excluded from our analysis, but did
significant association between hypergly- ing the risk for hypoglycemia detection. not include cohort studies. These authors
cemia and higher adjusted mortality in The data provided on actual BG values concluded that there was no significant
unstable angina, acute myocardial infarc- are variable, ranging from inclusion of impact on mortality when comparing insu-
tion, congestive heart failure, arrhythmia, one daily BG to a mean daily BG, or a lin infusions to achieve tight GC compared
ischemic and hemorrhagic stroke, gastro- time-weighted mean. Thus, effectiveness with usual care, OR 0.93, 95% CI [0.85,
intestinal bleeding, acute renal failure, of the protocols at minimizing glucose 1.03]. A more recent review following the
pneumonia, pulmonary embolism, and variability and hypoglycemia cannot be completion of the largest multicenter trial
sepsis (3). The mortality risk was signifi- thoroughly assessed. Nursing compli- found similar results with a mortality OR
cantly greater at each higher BG range in ance with intensive insulin protocols is 0.93, 95% CI [0.83, 1.04] (32). A third meta-
patients without a history of diabetes in typically unmeasured and unquantified. analysis evalu­ated only the seven largest
this large Veterans Affairs database. The Cohort studies could not control for trials and had a similar conclusion with a
intensity of the stress response, preexist- practice changes that occurred during mortality OR 0.95, 95% CI [0.87, 1.05] (33).
ing diabetes, and concurrent treatment the course of data collection or incon- The different methodologies employed
will influence the degree of hyperglyce- sistent protocol utilization. Importantly, and inclusion of different literature likely
mia. The impact of hyperglycemia on out- the standard of care likely influenced the explain results that are slightly differ-
come may be related to the presence of control population in several studies, as ent from the findings in this article. Fur-
preexisting diabetes, the intensity of the mean BG in the control group has fallen ther analysis of our data is available in the
hyperglycemic response, the diagnosis, throughout the last decade (30). supplemental materials (see Supplemental
and the risk for infection. Our meta-analysis included the larg- Digital Content 1, http://links.lww.com/
A simple intervention for slightly est clinical trials and large and small CCM/A589), with a subset analysis that sep-
elevated BG values is to avoid or mini- cohort trials. indicates a small but sig- arates observational trials from RCTs.
mize dextrose infusions when patients nificant, 16% reduction in the odds ratio
are receiving other sources of nutri- (OR) for hospital mortality with the use 2. In adult critically ill patients, what are
tional support; however, the majority of of insulin infusion therapy, targeting BG the morbidity benefits of maintaining
critically ill patients will require insulin < 150 mg/dL, OR 0.84, 95% confidence BG < 150 mg/dL?

Crit Care Med 2012 Vol. 40, No. 123253


Table 1.   Summary of key clinical trials used to evaluate the impact of glycemic controla

Study Quality

Intensive Care Unit


Author Population Design/End Point Design Assessment

Large randomized controlled trials


Van den Berghe Surgical, mechanical Randomized Single center, Evaluated mean morning
et al (1) ventilation 80–110 mg/dL vs. 180–200 mg/dL glucose
Research RN titrated insulin per protocol
Van den Berghe Medical, expected intensive Randomized Single center
et al (14) care unit stay >72 hrs Bedside RN titrated per paper protocol Evaluated mean morning glucose
Preiser et al (15) Medical and surgical Randomized Multicenter, Evaluated all glucose
80–110 mg/dL vs. 140–180 mg/dL values, also median morning value
Bedside RN titrated per protocol
The NICE-SUGAR Medical and surgical Randomized Multicenter Evaluated mean time-
Investigators (16) 80–110 mg/dL vs. 140–180 mg/dL weighted glucose
Adjusted via computerized algorithm Outcome based on 90-day mortality
Small randomized controlled trials
Brunkhorst Sepsis Randomized Multicenter, Evaluated mean morning
et al (17) 80–110 mg/dL vs. 180–200 mg/dL glucose
Bedside RN titrated per Van den Berghe protocol

De La Rosa Medical and surgical Randomized 80–110 vs. 180–200 mg/dL Single center
et al (18) Bedside RN titrated per protocol Evaluated mean morning glucose, also
daily minimum and maximum values
Arabi et al (19) Medical and surgical Randomized 80–110 mg/dL vs. 180–200 mg/dL Single center, Evaluated daily average
Bedside RN titrated per paper protocol glucose
Farah et al (20) Medical with >3- day Randomized 110–140 mg/dL vs. 140–200 mg/dL Single center, reported overall average
length of stay glucose
Grey and Surgical, excluded patients Randomized 80–120 mg/dL vs. 180–220 mg/dL Single center, reported daily average
Perdrizet (21) with diabetes and overall average glucose

Mackenzie et al (22) Medical and surgical Randomized 72– 108 mg/dL vs. 180–198 mg/dL Two centers, multiple glucose end
points reported
Large cohort
Krinsley (23) Medical and surgical Observational cohort Single center
<140 mg/dL vs. historical control Evaluated all glucose values
Bedside RN titrated
Furnary et al (24) CV surgical, diabetic Observational Single center
patients IV–various goals (final <150 mg/dL) vs. SQ control Historical control
Bedside RN titrated per protocol Variable end points
Long study timeline
Evaluated average daily glucose for
three postoperative days
Treggiari et al (25) Medical, surgical, and Observational cohort Single center
trauma 80–110 mg/dL vs. 80–130 mg/dL vs. historical control Protocol utilization was optional
Bedside RN titrated per protocol Evaluated all glucose values
Krinsley (26) Medical, surgical, and Observational cohort Single center, includes patients
trauma <140 or <125 mg/dL vs. historical control in reference 12
Bedside RN titrated
Scalea et al (27) Trauma Prospective cohort, post-protocol goal <150 mg/dL Single center, Evaluated highest single
Bedside RN titrated, no dosing guidelines daily glucose and pattern of response
Furnary CV surgical, diabetics Observational Single center
and Wu (28) IV–various goals (final <110 mg/dL) vs. SQ control Historical control
Bedside RN titrated per protocol Variable end points
Long study timeline
Evaluated average daily glucose for
three postoperative days
Small cohort
Toft et al (29) Medical–surgical and no Prospective cohort, post-protocol goal Single center, Evaluated mean morning
CV surgical 80–110 mg/dL glucose
Bedside RN titrated per Van den Berghe protocol

IV, intravenous; IQR, interquartile range; CV, cardiovascular; SQ, subcutaneous; 3-blood glucose = 3-day average postoperative blood glucose.
Small trials included <1,000 patients; bhospital mortality unless otherwise specified; ctreatment not blinded.
a

3254 Crit Care Med 2012 Vol. 40, No. 12


Findings

Actual Glycemic End Points


No. of Patients Mean ± sd (mg/dL)
Hospital Mortality
Odds Ratio [95%
Control Glycemic Control Control Glycemic Control Confidence Interval]b Commentc

783 765 Daily 153 ± 33 Daily 103 ± 19 0.64 [0.45, 0.91] Control glucose elevated with IV dextrose,
stopped early for benefit

605 595 Daily 153 ± 31 Daily 111 ± 29 0.89 [0.71, 1.13] Control glucose elevated with IV dextrose

542 536 144 (IQR 128–162) 117 (IQR 108–130) 1.27 [0.94, 1.7] Stopped early for hypoglycemia, many
median–all values median–all values protocol violations

3050 3054 144 ± 23 115 ± 18 28-day Glycemic control group did not achieve
1.09 [0.96, 1.23] target
90-day1.14 [1.02, 1.28]

290 247 Median daily 138 Median daily 130 28-day mortality Control target artificially elevated with IV
(IQR 111–184) (IQR 108–167) 0.94 [0.63, 1.38] dextrose. Study stopped early for hypo-
glycemia risk. Inadequate size to detect
difference in mortality
250 254 Median–all values Median–all values 28-day mortality Did not achieve target glucose, in-
149 (IQR 124–180) 120 (IQR 110–134) 1.08 [0.75, 1.54] adequate sample to detect mortality
difference
257 266 171 ± 34 115 ± 18 0.78 [0.53, 1.13] Small trial, inadequate size to show
mortality impact
48 41 174 ± 20 142 ± 14 1.37 [0.59, 3.16] Baseline imbalance in diabetes incidence
and admission glucose
27 34 179 ± 61 125 ± 36 mg/dL, 0.47 [0.12, 1.86] Lower nosocomial infection incidence
Daily mean value lower with glycemic control
on each day
119 121 8.4 ± 2.4 7.0 ± 2.4 0.73 [0.43, 1.24] Inadequate size to show mortality impact

800 800 152 ± 93 131 ± 55 0.45 [0.34, 0.60] High protocol adherence, subcutaneous
and IV insulin

942 2612 214 ± 41 177 ± 30 0.27 [0.19, 0.39] Remote historical controls on SQ insulin
p < .001 only

2366 <130 mg/dL, n = 3322 Mean all values Mean all values 1.07 [0.9, −1.21] Low protocol utilization, overlap in
<110 mg/dL n = 4786 147 ± 42 Goal < 130: 142 ± 37 glucose values actually achieved
Goal < 110:133 ± 31
2666 2699 Mean overall Mean overall 0.72 [0.62, 0.83] Includes patients in Krinsley above
154 ± 88 124 ± 51

1021 1108 Data not provided Data not provided 0.68 [0.52, 0.89] Changes in standard of care likely during
trial
1065 4469 Data not provided Data not provided 0.39 [0.28, 0.54] Includes patients in Furnary above

135 136 Median daily Median daily 0.77 [0.38, 1.55] Small trial, but trend to benefit with
133 (IQR 121–150) 110 (IQR 104–117) glycemic control

Crit Care Med 2012 Vol. 40, No. 123255


Figure 1.  Forest plots of (A) hospital or 28-day mortality and (B) intensive care unit mortality (1, 14–21, 25–29). CI, confidence interval; MH, Mantel-Haenszel.

A. We suggest that there is no consis- was consensus that too many confound- surgery to achieve a reduced risk of
tently demonstrated difference in ing variables existed for this outcome. A deep sternal wound infection and
several morbidity measures (renal reduction in critical illness polyneurop- mortality.
failure, transfusion, bacteremia, athy was not analyzed as this potential
polyneuropathy, and ICU length of benefit was reported in only one study. [Quality of evidence: very low]
stay [LOS]) when evaluated in the Our analysis suggests that no evidence The only large-scale RCT to date evalu-
general adult ICU population. of benefit was found in ICU LOS with OR ating the impact of tight GC on morbid-
−0.05, 95% CI [−0.14, 0.05]; prevention ity and mortality in a population weighted
[Quality of evidence: very low] of bacteremia OR 0.81, 95% CI [0.58, with postoperative cardiac surgical
The following were considered as 1.11]; need for transfusion OR 1.06, 95% patients was published in 2001 (1). Almost
morbidity outcomes for evaluation, CI [0.90, 1.26]; or need for renal replace- two thirds of this study population under-
acute renal replacement therapy, inci- ment therapy OR 0.90, 95% CI [0.7, went cardiac surgery. Patients in the GC
dence of transfusion, bacteremia, criti- 1.16], but variable study design, popula- group (80–110 mg/dL) had lower ICU and
cal illness polyneuropathy, and ICU LOS. tions, and end points limit the analysis. hospital mortality rates compared with
To analyze ICU LOS in those studies in The forest and funnel plots are avail- conventional therapy (BG 180–200  mg/
which data were reported nonpara- able in Supplemental Digital Content 1 dL). Morbidity benefits for the GC group
metrically, the median value was used (http://links.lww.com/CCM/A589). included a reduced need for renal replace-
and interquartile range (IQR, 1.35) was ment therapy, less chance of hyperbiliru-
used as an estimate of sample standard B. We suggest implementation of mod- binemia, earlier cumulative likelihood of
deviation (sd). Duration of mechanical erate GC (BG < 150 mg/dL) in the weaning from mechanical ventilation, and
ventilation was not analyzed as there postoperative period following cardiac ICU and hospital discharge. A follow-up

3256 Crit Care Med 2012 Vol. 40, No. 12


preplanned subanalysis of the 970 high- are needed to confirm this finding. Thus, the injured brain (76). Severe hypoglyce-
risk cardiac surgery patients from the orig- at this time we recommend that trauma mia (SH) can produce or exacerbate focal
inal study confirmed a survival benefit due ICU patients should be managed in the neurological deficits, encephalopathy,
to GC up to 2 yrs after hospital discharge same fashion as other ICU patients. seizures or status epilepticus, permanent
and longer for the subset treated for at cognitive dysfunction, and death. Further,
least 3 days (34). Additionally, a series of D. We suggest that a BG ≥ 150 mg/dL tight GC may induce regional neurogly-
reports from a clinical database of diabetic triggers initiation of insulin therapy copenia in TBI (77). Clinical trials are
cardiac surgery patients suggested that for most patients admitted to an ICU urgently needed to determine the opti-
maintenance of BG < 150 mg/dL is asso- with the diagnoses of ischemic stroke, mum degree of GC and a safe minimum
ciated with a reduction of sternal wound intraparenchymal hemorrhage, aneu- BG goal in neurologic injury populations
infection and an incremental decrease in rysmal subarachnoid hemorrhage, or with respect to mortality and morbid-
hospital mortality compared with remote TBI, titrated to achieve BG values abso- ity. Trials will require careful design as a
historical control patients treated with lutely < 180 mg/dL with minimal BG result of the following three confounders:
sliding-scale insulin (24, 35–37). Another excursions <100 mg/dL, to minimize 1) extreme hypoglycemia and hypergly-
retrospective review of patients treated the adverse effects of hyperglycemia. cemia on admission are associated with
with a combination of IV and SQ insulin in increased severity of underlying disease
the postoperative period showed a strong [Quality of evidence: very low] (i.e., a U-shaped mortality curve indepen-
association between GC and reduction in There is abundant experimental and dent of therapy); 2) current therapeutic
morbidity and mortality (38). observational evidence to show that interventions carry risks of both creating
hyperglycemia at the time of the neu- hypoglycemia (both global and regional)
C. In the population of critically ill injured rologic event is associated with adverse and allowing hyperglycemia to persist
(trauma) ICU patients, we suggest that outcomes in stroke and TBI, but no pro- (i.e., a U-shaped mortality curve as a direct
BG ≥ 150 mg/dL should trigger initia- spective interventional trial has shown consequence of therapy); and 3) response
tion of insulin therapy, titrated to keep that control of hyperglycemia with insu- to therapy may also be determined in part
BG < 150 mg/dL for most adult trauma lin reduces mortality, as demonstrated by by the severity of the underlying injury.
patients and to maintain BG values our meta-analysis OR 0.97, 95% CI [0.81, Case reports of neuroglycopenia and
absolutely < 180 mg/dL, using a pro- 1.16] (Fig. 2). Hyperglycemia is both a cerebral distress (altered lactate/pyruvate
tocol that achieves a low rate of hypo- common problem (49–53) and strongly ratios) during insulin infusion therapy
glycemia (BG ≤ 70 mg/dL) to achieve associated with greater mortality and have been reported independent of low
lower rates of infection and shorter worse functional outcome following isch- BG (77). The clinical significance of this
ICU stays in trauma patients. emic stroke (54–57), intraparenchymal finding remains unknown and is further
hemorrhage (58, 59), aneurysmal sub- complicated by data suggesting that the
[Quality of evidence: very low] arachnoid hemorrhage (60–62), and TBI rate of glucose change may be more
A hypermetabolic stress response (63–65). Patients who are responsive to important than the hypoglycemic event
resulting in hyperglycemia is common insulin therapy have a better prognosis itself (78).
in the trauma population (39). Hypergly- than those with persistent hyperglyce-
cemia on admission or within the first 2 mia (66, 67). Four small feasibility trials 3. What is the impact of hypoglycemia in
ICU days may be predictive of poor out- of insulin infusion have been undertaken the general ICU population?
come (longer LOS, more infection) and (68–71), but none was designed to evalu-
higher mortality (40–42). Additionally, ate outcome, and none is sufficiently We suggest that BG ≤ 70 mg/dL are
persistence of hyperglycemia is associated powerful to guarantee safety (Table 3). associated with an increase in mortality,
with poor outcome (43–45). A pre-trauma The Glucose Insulin in Stroke Trial was and that even brief SH (BG ≤ 40 mg/dL)
diagnosis of insulin-dependent diabetes stopped prematurely due to slow enroll- is independently associated with a greater
was not associated with higher mortality ment (72). Three more recent studies all risk of mortality and that the risk increases
or hospital LOS (46). failed to demonstrate decreased mortality with prolonged or frequent episodes.
The benefit of insulin therapy on with tight GC but confirmed substantial [Quality of evidence: low]
improving trauma patient outcome has increases in the rate of hypoglycemia The practice of GC in critically ill
not been clearly demonstrated (Table 2) with tight control (73–75). Thiele et al patients is associated with a higher inci-
(16, 27, 47, 48). In the Normoglycemia in (75) demonstrated that hypoglycemia dence of hypoglycemia (BG < 70 mg/dL)
Intensive Care Evaluation–Survival Using was an independent risk factor for mor- and a five-fold increase in the risk of SH
Glucose Algorithm Regulation (NICE- tality in multivariate analysis (OR 3.818). (BG < 40 mg/dL) OR 5.18, 95% CI [2.91,
SUGAR) multicenter trial of 6,104 patients, The NICE-SUGAR study has a TBI sub- 9.22] (Fig. 3). The percentage of adult
trauma patients represented 15.5% of the group, the results of which have yet to be patients sustaining one or more episodes
conventional therapy group (BG goal 140– reported. of SH in the interventional arms of three
180 mg/dL) and 14% of the GC group (goal major prospective randomized trials of
80–110 mg/dL) (16). Subset analysis indi- E. We further suggest that BG < 100 mg/ intensive insulin therapy has ranged from
cated a trend toward lower mortality in dL be avoided during insulin infusion 5.1% to 18.7% (1, 14, 16). Attempts to
the GC group (OR 0.77, 95% CI [0.5, 1.18]; for patients with brain injury. achieve tight GC (goal 80–110 mg/L) have
p = .07). Although these data are hypoth- not uniformly created the highest risk
esis-generating and that trauma patients [Quality of evidence: very low] of severe hypoglycemia, suggesting that
may benefit more from GC than the other Hypoglycemia carries specific risks for the protocol employed or the population
ICU patients, additional prospective trials the normal brain and a greater risk for studied might have influenced the risk.

Crit Care Med 2012 Vol. 40, No. 123257


Table 2.   Summary of clinical trials evaluating impact of insulin therapy on patient outcome after trauma

Study Quality No. of Patients Findings

Actual Glycemic End Points Hospital


Mean ± sd (mg/dL) Mortality
Odds Ratio
Intensive [95%
Care Unit Design/Glucose Design Glycemic Glycemic Confidence
Author Population End Point Assessment Control Control Control Control Interval] Comments

Scalea Trauma Prospective data Single center, 1021 1108 NA NA Adjusted Reduced
et al (27) intensive collection, two reported outcomes: vent days
care unit patient series patterns Pre vs. post; and length
before and of glucose mortality of stay with
after protocol control in 1.4 [1.1, 10] improved
100–150 mg/dL week 1. pattern of
Over 51% glucose
had glucose control
>150 mg/dL
in first week
(poor protocol
effect)
Reed Surgical and Retrospective Single center, Not Not Reported by Reported by Estimated Lower mean
et al (47) trauma ICU, query of uncontrolled reported reported study year study year mortality glucose not
n = 7261 prospective protocol by group by group 2003: 141 2005: 129 ratio correlated
database, compliance. 2004: 134 2006: 125 measured with
pre- and Measured p < 0.01 as actual/ estimated
post-protocol glucose estimated mortality
implementation control by mortality risk
year. End point unchanged reduction.
was estimated Other factors
mortality ratio changed
measured during
as actual/ observational
estimated period (key
mortality personnel,
population,
quality
emphasis)
Collier Trauma, on Prospective Single center, 383 435 130 ± 11 124 ± 13 ≥1 Glucose Preprotocol,
et al (48) mechanical postprotocol Reported days above historical
ventilation (80–110 mg/dL) mean glucose 150 mg/dL: control, no
vs. historical Pre- vs. post- 2.16 [1.0, significant
control mortality not 4.6] reduction
reported p = .049 in mean
glucose pre
to post, did
not achieve
glucose
target
The NICE- Trauma Randomized, Multicenter, 465 421 Not reported Not reported 0.77 [0.5, Hypothesis-
SUGAR subset Tight 81–108 mg/ Reported mean for trauma for trauma 1.18] generating
Investiga- dL vs. control glucose and subset subset p = .07 for subset
tors (16) < 180 mg/dL time-weighted heterogene- analysis
mean overall ity 90-day
mortality

NA, not applicable, not available.

The impact of insulin-induced hypo- of vasoactive infusions, insulin therapy, liver disease, immune compromise, and
glycemia has varied among populations, and the use of continuous renal replace- medical or nonelective admissions are
and in some reports, hypoglycemia was ment therapy with a bicarbonate-based noted as potential risk factors for the
thought to be a marker for more serious replacement fluid (81). Some authors also occurrence of low BG (79). Physiologic
underlying illness (79, 80). Risk factors found that diabetes, mechanical ventila- changes increase the effect of insulin as
for SH include renal failure, interruption tion, female sex, greater severity of illness, renal failure prolongs the half-life of insu-
of caloric intake without adjustments in and longer ICU stays are associated with lin, leading to insulin accumulation, while
the insulin infusion, sepsis with the use increased risk of SH (80, 82). Additionally, also attenuating renal gluconeogenesis.

3258 Crit Care Med 2012 Vol. 40, No. 12


Figure 2.  Forest plot of neurological mortality (hospital or 28-day) (1, 14, 16, 19, 23, 60, 65, 68, 71–75). CI, confidence interval; MH, Mantel-Haenszel.

Hepatic failure can also lead to reduced trials to analyze the independent effects value to avoid overcorrection. The BG
hepatic gluconeogenesis. The reliability of of hypoglycemia and glycemic variability should be repeated in 15 mins with fur-
the insulin infusion therapy protocol and (GV) on the risk of mortality (88). The ther dextrose administration as needed
frequency of BG monitoring also appear to occurrence of one or more episodes of to achieve BG > 70 mg/dL with a goal to
influence the frequency of hypoglycemia. SH was independently associated with a avoid iatrogenic hyperglycemia.
Multivariate regression models dem- higher risk of mortality (OR 3.233, 95% [Quality of data: very low]
onstrate that even a single episode of SH CI [2.251, 4.644]; p < .0001). Although prevention of hypoglycemia
is independently associated with higher Morbidity impact of SH is difficult to is important during insulin therapy, epi-
risk of mortality (80–85). The OR for quantitate on critically ill patients as con- sodes of low BG may occur despite rea-
mortality associated with one or more current illness and sepsis may increase sonable precautions, and steps should be
episodes was 2.28, 95% CI [1.41, 3.70]; the risk of cognitive impairment, and it is taken to recognize and treat it promptly.
(p = .0008) among a cohort of 5,365 unknown how hypoglycemia may inter- With severe hypoglycemia, interruption of
patients admitted to a single mixed medi- act with other risk factors. Low BG levels the insulin infusion is a prudent first step.
cal–surgical ICU (82). Most other reports lead to nonspecific neurologic symptoms, This interruption may be adequate for a
similarly indicate a higher risk of mortal- although severe or prolonged glycopenia patient receiving exogenous dextrose, but
ity with hypoglycemia of varying severity may produce neurocognitive impairment, treatment with additional IV dextrose is
(Table 4). Early hypoglycemia has been seizures, loss of consciousness, perma- typical, although there is no adequate data
associated with longer adjusted ICU LOS nent brain damage, depression, and death to dictate the optimal dose. While the first
and greater hospital mortality, especially (89–91). A number of factors including priority is patient safety through restora-
with recurrent episodes (86). Further- sedation, medication, or underlying dis- tion of normoglycemia, rebound hyper-
more, patients with more severe degrees ease may mask symptoms of neuroglyco- glycemia due to excessive replacement
of hypoglycemia sustained higher ICU penia. To further complicate the analysis, should also be avoided, especially because
and hospital mortality (85, 86). A greater hyperglycemia has also been associated the resulting increase in GV may contrib-
risk of mortality (RR 2.18, 95% CI [1.87, with adverse effects on the brain (92). Fur- ute to adverse outcomes (82, 83, 88, 93).
2.53]; p < .0001) was similarly reported ther, the risk for neurologic injury may be An IV dextrose dose of 15–20 g has been
with mild to moderate hypoglycemia compounded by additional oxidative stress recommended by the American Diabetes
(BG 55–69 mg/dL) in a post hoc analy- associated with rapid correction of hypo- Association, with instructions to recheck
sis of prospective data collected in a glycemia with IV dextrose (93). BG in 5–15 mins and repeat as needed
randomized trial and two large cohorts (7). A dose of 25-g IV dextrose adminis-
(87). These data confirmed the results of 4. How should insulin-induced tered to nondiabetic volunteers produced
another cohort study that demonstrated hypoglycemia be treated in adult ICU significant but variable BG increases of
that mild–moderate hypoglycemia, BG patients? 162 ± 31 mg/dL and 63.5 ± 38.8 mg/dL
54–63  mg/dL, was independently asso- when measured 5 and 15 mins postinjec-
ciated with increased risk of mortality We suggest that BG < 70  mg/dL tion, respectively (94). BG returned to
(85). In each of these studies, the mor- (<100 mg/dL in neurologic injury patients) baseline by 30 mins, but the duration may
tality risk was greater with more severe be treated immediately by stopping be different in patients receiving exog-
hypoglycemia (85, 87). Finally, the Leu- the insulin infusion and administering enous insulin.
ven investigators have recently published 10–20  g of hypertonic (50%) dextrose, A formula to calculate a patient-
data pooling the two interventional adult titrated based on the initial hypoglycemic specific dose of dextrose has been used

Crit Care Med 2012 Vol. 40, No. 123259


Table 3.  Summary of clinical trials in neurological patients

Study Quality No of Patients Findings

Neuro-intensive Hospital Mortalitya


Care Unit Design Tight Odds Ratio [95%
Author Population Design/End Point Assessment Control Glycemic Control Confidence Interval] Comments

Subsets of RCT
Van den Mixed, Randomized Single center 30 33 0.73 [0.21, 2.48] Control glucose
Berghe surgical, 80–110 mg/dL vs. Evaluated elevated with IV
et al (1) mechanical 180–200 mg/dL mean morning dextrose, stopped
ventilation Research RN titrated glucose early for benefit
insulin per protocol
Van den Mixed, Randomized Single center 31 30 1.05 [0.35, 3.15] Control glucose
Berghe medical Bedside RN titrated per Evaluated elevated with IV
et al (14) paper protocol mean morning dextrose
glucose
Small RCT or subset of small RCTb
Scott et al CVA Randomized Single center, 28 25 28-day mortality No difference in
(68) Fixed dose glucose- Evaluated 0.97 [0.29, 3.22] serum glucose at
potassium-insulin vs. glucose any point studied
saline infusion trajectory over
for 24 hrs treatment
period
Walters CVA Randomized Single center 12 13 3.00 [0.11, 80.95] AUC reduced
et al (71) Target 90–140 mg/dL vs. Evaluated
standard management glucose–time
curve AUC
Bilotta Aneurysmal Randomized Single center, 38 40 Six-month 83% of control
et al (60) subarachnoid Target 80–120 mg/dL vs. evaluated mortality and 69% of
hemorrhage 80–220 mg/dL percentage of 0.78 [0.24, 2.58] intensive therapy
glucose values in target range
in target range
Gray CVA Randomized, glucose- Multicenter, 469 464 90-day mortality Average difference
et al(72) potassium-insulin evaluated glu- 1.14 [0.86, 1.51] in glucose 10 mg/
infused to target cose every 8 hrs dL (p < .001)
72–126 mg/dL vs. using repeated
saline control measures analy-
sis of variance
Arabi et al Traumatic Randomized Single center, 39 55 1.43 [0.13, 16.39]
(19) brain injury 80–110 mg/dL vs. Evaluated aver-
180–200 mg/dL age glucose level
Bedside RN titrated per
paper protocol
Bilotta Traumatic Randomized, Target Single center, 49 48 1.02 [0.24, 4.35] Mean glucose
et al (73) brain injury 80–120 mg/dL vs. evaluated mean values 97 vs.
80–220 mg/dL glucose values 147 mg/dL
(p < .0001)
Bilotta Mixed, neuro- Randomized, Target Single center, 242 241 0.91 [0.61, 1.35] Difference in day
et al (74) surgical 80–110 mg/dL vs. evaluated mean 1 to day 14 mean
<215 mg/dL daily glucose glucoses: 92 mg/
values dL vs. 143 mg/dL
(p < .0001)
Cohort Studies
Krinsley (23) Mixed Observational co- Single center, 119 142 0.35 [0.17, 0.73] High protocol
hort<140 mg/dL vs. evaluated all adherence, sub-
historical control. glucose values cutaneous and IV
Bedside RN titrated insulin
Thiele et al Aneurysmal Retrospective postpro- Single center, 343 491 1.03 [0.67, 1.59] Median average
(75) subarachnoid tocol target 90–120 mg/ Evaluated glucose
hemorrhage dL. Bedside RN titrated median average 121 vs. 116 mg/dL
protocol glucose (p < .001)

RCT, randomized clinical trial; CVA, acute ischemic stroke; AUC, area under the curve.
a
Hospital mortality un­less otherwise specified; bsmall trials included <1,000 patients.
Aside from Krinsley (23), every trial had an inadequate sample size to detect mortality differences. Treatment was not blinded in any study.

3260 Crit Care Med 2012 Vol. 40, No. 12


Figure 3.  Forest plot of severe hypoglycemia (1, 14–19, 25, 26, 29). CI, confidence interval; MH, Mantel-Haenszel.

in several reports (50% dextrose dose in although additional testing of this inter- BG monitored at least every hour to allow
grams = [100 − BG] × 0.2 g), and it typi- vention appears warranted. rapid recognition of BG outside the goal
cally advises administration of 10–20 g Oral dextrose replacement (15  g) is range. More frequent reassessment is
of IV dextrose, an amount lower than that used in ambulatory patients with hypogly- needed after treatment of hypoglycemia,
in traditional dosing methods (95, 96). cemia, but is not tested for ICU patients. every 15 mins until stable.
This approach corrected the BG into the Fifteen grams of oral carbohydrate pro- A retrospective evaluation of data
target range in 98% within 30 mins for duced a BG increase of approximately from 6,069 insulin infusion episodes in
patients who had received IV insulin infu- 38 g/dL within 20 mins and provided 4,588 ICU patients suggested that delays
sions (95, 97). Similarly, titrated replace- adequate symptom relief in 14 ± 0.8 mins in measuring BG contributed to the risk
ment has been advocated for treatment of in hypoglycemic adult outpatients (101). of severe hypoglycemia. When a hypo-
adults in the prehospital setting. Adminis- If oral replacement is used, dextrose or glycemic episode occurred, the median
tration of 5-g aliquots of dextrose repeated sucrose tablets or solutions are preferred delay past the next hourly measurement
every minute, using either 10% (50 mL) for a more rapid or consistent response was 21.8 mins (IQR 12.2–29 mins) (97).
or 50% (10 mL) dextrose, restored mental compared with viscous gels or orange Modeling suggested SH was likely with as
status to normal in approximately 8 mins juice due to variable carbohydrate content little as a 12-min delay in the majority of
with both agents (IQR 5–15 and 4–11, in commercial juice (101). The impact of patients who developed hypoglycemia.
respectively), but the 50% dextrose group abnormal gastric emptying has not been Glucose checks every 4 hrs have been
received a larger median dose of dextrose, studied but may alter the response to used in some protocols; however, there
25 g (IQR 15–25) vs. 10 g (IQR 10–15), and therapy, especially in an ICU population. is a risk of unrecognized hypoglycemia
developed a higher median posttreatment with prolonged measurement intervals;
BG (169 mg/dL vs. 112 mg/dL [p = 003]), 5. How often should BG be monitored in so these intervals are not recommended
respectively (98). The authors recom- adult ICU patients? as a routine component of insulin infu-
mended titrating 10% dextrose in 50-mL sion protocols. The rates of hypoglyce-
IV (5-g) aliquots to treat the symptoms of We suggest that BG be monitored mia are above 10% for many protocols
hypoglycemia and to avoid overcorrec- every 1–2 hrs for most patients receiving using BG checks every 4 hrs (1, 14, 15,
tion of BG. The rate of administration of an insulin infusion. 17). One exception was reported with a
concentrated dextrose solutions may also [Quality of evidence: very low] computerized protocol that tested an
be important, as a report of cardiac arrest This is a consensus recommendation average of approximately six BG values
and hyperkalemia was associated with based on limited data, as this question has per day but produced SH in only 1%
rapid and repeated administration of 50% not been tested in a prospective fashion. of patients (102). With the higher rate
dextrose (99). The optimal frequency of BG testing has of hypoglycemia reported with every
A prehospital study comparing an not been established. Published protocols 4-hourly BG testing, this frequency is
intramuscular 1-mg injection of gluca- generally initiate insulin therapy with not suggested unless a low hypoglycemia
gon to a 25-g IV dose of dextrose demon- hourly BG testing, and then may liber- rate is demonstrated with the insulin
strated a rapid and potentially excessive alize the testing to every 4 hrs based on protocol in use.
BG response with dextrose, achieving the stability of the BG values within the
14–170 mg/dL increase in BG in the first desired range, as well as an assessment of 6. Are POC glucose meters accurate for
10 mins (100). The glucagon response was patient clinical stability. The personnel BG testing during insulin infusion
slower, achieving a final BG concentration time required for BG monitoring is the therapy in adult ICU patients?
of 167 mg/dL after 140 mins. Because vir- primary barrier to more frequent moni-
tually all ICU patients have venous access, toring. We suggest that unstable patients We suggest that most POC glucose
IV dextrose is preferred over glucagon, (e.g., titrating catecholamines, steroids, meters are acceptable but not optimal for
due to the delay in glucagon response, changing dextrose intake) should have routine BG testing during insulin infusion

Crit Care Med 2012 Vol. 40, No. 123261


Table 4.  Clinical trials reporting rate and impact of hypoglycemia on outcome of critically ill patients

Author/Reference Design n Population Results

Vriesendorp Retrospective 156 (245 events) Glucose <45 mg/dL, closed med- Risk factors: OR [95% CI]
et al (81) cohort 155 control surg ICU, University, teaching Nutrition interruption 6.6 [1.9, 23]
Diabetes mellitus 2.6 [1.5, 4.]
Sepsis 2.2 [1.2, 4.1]
Shock 1.8 [1.1, 2.9]
Renal replacement therapy with
bicarbonate fluids 14 [1.8, 106]
Insulin 5.4 [2.8, 10]
Vriesendorp Retrospective 156 Glucose <45 mg/dL, closed med- Cumulative in-hospital mortality:
et al (83) cohort 146 control surg ICU, University, teaching hazard ratio 1.03 [0.68, 1.56]; p = .88
Van den Berghe Post hoc 154 intensive Glucose ≤40 mg/dL, single Hypoglycemia frequency: control 1.8%,
et al (84) analysis of 25 control center, med-surg ICU intensive 11.3% (p < .0001)
two RCTs (2,748 total n) Hospital mortality: control 13 (52%),
intensive 78 (50.6%) (p = .9)
Mortality in 24 hrs: control 3 (12%),
intensive 6 (3.9%) (p < .0004)
Krinsley and Retrospective 102 cases Glucose <40 mg/dL, single- Hospital mortality: 55.9% vs. 39.5%
Grover (82) case-control 306 control center. Community med-surg (cases vs. controls) (p = .0057)
cohort (5,365 total in ICU. Insulin given to 72.5% of For the entire cohort, a single episode of
database) patients severe hypoglycemia
OR 2.28 [1.41–3.7]; p = .0008
(risk of hospital mortality)
Risk factors: OR [95% CI]
Diabetes 3.07 [2.03, 4.63]
Septic shock 2.03 [1.19, 3.48]
Mechanical ventilation 2.11 [1.28, 3.48]
Acute Physiology, Age and Chronic
Health Evaluation system II 1.07 [1.05,
1.10]Treatment in intensive insulin era
1.59 [1.05, 2.41]
Wiener et al (31) Meta-analysis 14 of 34 trials Intensive care patients, Hypoglycemia relative risk: 5.13
Tight glycemic International, glucose ≤40 mg/ (4.09, 6.43)
control vs. dL
control
Kosiborod Retrospective n = 7820, 482 Study of patients admitted with Higher mortality seen in patients with
et al (79) cohort hypoglycemia acute myocardial infarction; spontaneous hypoglycemia (OR 2.32
database from 40 U.S. medical [CI 1.31, 4.12]), but not in patients with
centers; hypoglycemia defined as insulin-related hypoglycemia (OR 0.92
glucose <60 mg/dL [CI 0.58, 1.45])
Arabi et al (80) Nested co- n = 523, 84 Med-surg ICU, RCT insulin Adjusted mortality hazard ratio, 1.31;
hort in RCT hypoglycemia infusion 80–110 mg/dL vs. con- 95% CI [0.70, 2.46]; p =.40
ventional 180–200 mg/dL Hypoglycemia rate 3.6 per 100 treat-
ment days
Risk factors: older age, higher Acute
Physiology, Age and Chronic Health
Evaluation system II score, longer LOS,
females, admitted for nonoperative
reasons, diabetics with higher admission
blood glucose, septic, m
­ echanically ven-
tilated, had received renal replacement
therapy intensive insulin protocol
Griesdale Meta-analysis 14 of 26 trials; Intensive care patients, Inter- Relative risk for hypoglycemia: 5.99
et al (32) tight glycemic national, glucose ≤40 mg/dL, (4.47, 8.03)
control vs. including NICE-SUGAR
control
Egi et al (85) Retrospective n = 4946, 1,109 Intensive care patients, two Higher unadjusted mortality: seen in
cohort hypoglycemia hospitals, 2000–2004 patients even with mild hypoglycemia,
54–80 mg/dL

CI, confidence interval; ICU, intensive care unit; med-surg, medical–surgical unit; OR, odds ratio; RCT, randomized controlled trial.

3262 Crit Care Med 2012 Vol. 40, No. 12


therapy. Clinicians must be aware of to delays that could add significant risk to We suggest arterial or venous whole
potential limitations in accuracy of glu- efficient insulin titration. blood sampling instead of finger-stick
cose meters for patients with concurrent The methodology used by a POC meter capillary BG testing for patients in shock,
anemia, hypoxia, and interfering drugs. (glucose oxidase vs. glucose-1-dehydro- on vasopressor therapy, or with severe
[Quality of evidence: very low] genase) will impact the accuracy and the peripheral edema, and for any patient on
The use of glucose meters has become potential for interference by patient phys- a prolonged insulin infusion.
common in hospitals due to their ease iology, other circulating substances, and [Quality of evidence: moderate]
of use, availability, and ability to pro- sample source. These have been reviewed Finger-stick capillary BG measurement
vide rapid results. Unfortunately, in the elsewhere, but some specific factors are is typical when using a meter, although as
limited testing that has been reported, pertinent to the ICU (107). For example, discussed, meters may introduce error
many of these devices lack accuracy high Po2 (>100 mm Hg) can falsely lower and bias in the BG value. Studies (Table 5)
when used in critically ill patients. How- BG readings on POC meters that use glu- have compared BG in simultaneous sam-
ever, insulin infusion therapy would be cose oxidase methods (108, 109). ples drawn from different sites in critically
impossible without some type of POC Hematocrit (Hct) is an important ill patients (105, 123–135). These are dif-
testing methodology. The initial study variable for POC glucose testing in criti- ficult to compare due to the differences in
by Van den Berghe et al (1) on intensive cally ill patients. Most POC meters are reporting, testing methodology, and com-
insulin therapy used a precise arterial approved for BG measurement within parators. Of importance to clinicians is
blood gas instrument for BG testing. a Hct range of 25%–55%, but low Hct that meter performance deviated from lab-
Later trials have used a variety of POC has repeatedly been shown to alter the oratory control by >20% in some reports,
devices. One possible explanation for the accuracy of BG results with a POC meter. regardless of the blood source (130).
generally unfavorable results in subse- Lower Hct values generally allow meters Samples from an arterial site are most
quent trials may be due to inappropriate to overestimate BG values, potentially similar to laboratory plasma or blood
insulin dosing in response to inaccurate masking hypoglycemia (110–114). There gas analyzer BG values in paired sam-
BG results. are no real-time alerts on meters to direct ples. Venous specimens are also gener-
Studies examining the accuracy of clinicians to use other methodologies ally acceptable, as long as care is taken
POC glucose meters compared with a in the face of low Hct, although newer to avoid contamination of the specimen
reference laboratory methodology of meters minimize Hct interference by cor- from IV fluid infusing through a multilu-
plasma glucose measurement reported recting abnormal values (115, 116). A for- men catheter.
significant variability and bias between mula may be applied to correct a meter Finger-stick capillary glucose levels
these testing methods (103). Clinicians BG value with low Hct (117). Newer glu- may provide significantly different results
must be aware of the limitations with the cose meters appear to have addressed the compared with arterial or venous speci-
specific device used. Comparing data on limitations of older meters (118). mens when patients have low perfusion
specific meters may be confounded by a Drugs such as acetaminophen, ascor- with hypotension, edema, vasopressor
lack of consensus on the limits of accept- bic acid, dopamine, or mannitol, along infusion, or mottled appearance of the
able error between the Food and Drug with endogenous substances such as uric skin (105, 124, 127–130, 132). Hypoper-
Administration (allows up to 20% error) acid or bilirubin, may interfere with the fusion may increase glucose extraction
and the American Diabetes Association accuracy of POC meters, especially those and increase the difference between cap-
(up to 5% error) standards. The Clini- meters using glucose oxidase methodol- illary whole blood and venous or arterial
cal and Laboratory Standards Institute ogy (119). The direction of interference on plasma glucose. Unfortunately, there is
and International Organization for Stan- BG values depends on the device and the no consistent pattern to the variability, as
dardization 15197 guidelines allow up interfering substance. Glucose–dehydro- finger-stick testing BG results might be
to 15 mg/dL variance for BG < 75 mg/dL genase-based assays are sensitive to inter- lower or higher than arterial or venous
and up to 20% of the laboratory analyzer ference and false elevation of results if the samples. Each institution should evaluate
value for BG ≥ 75 mg/dL (104). The Clini- patient receives medications containing the performance of their selected meter in
cal and Laboratory Standards Institute maltose (e.g., immune globulins) or ico- a variety of patient groups.
suggests that a correlation above .9751 dextrin (e.g., peritoneal dialysis solutions). A sampling site hierarchy that priori-
is indicative of equivalence to the labora- An alternative POC method with a tizes arterial or venous sampling should
tory standard (105). Simulation has sug- cartridge-based amperometric method be established for BG monitoring of criti-
gested that meter error exceeding 17% is available for whole blood testing and cally ill patients. Devices that minimize
may double the number of potentially sig- has been tested in critically ill popula- blood waste with catheter sampling are
nificant errors in insulin administration tions (120). There are few limitations to important to minimize the risk of ane-
and result in a higher risk of hypoglyce- these cartridge-type devices using glucose mia induced by frequent phlebotomy.
mia (106). While meters have generally oxidase technology with the exception of Finger-stick testing is invasive and often
been considered acceptable within the known interference from hydroxyurea painful for patients who need frequent BG
usual ranges of BG testing (80–200 mg/ and thiocyanate (121). A checklist for measurements, and thus it should be the
dL), additional laboratory testing of blood evaluation of POC glucose devices has site of last resort or avoided completely if
samples at the extremes of BG concentra- been published to improve the quality of the patient is on vasopressors or exhibits
tion is needed to detect potential errors device evaluation (122). hypoperfusion.
and avoid over- or under-treatment with
insulin. The logistics of obtaining timely 7. When should alternatives to finger- 8. Can continuous glucose monitoring
central laboratory measurement and stick capillary sampling be used in replace POC methods for critically ill
reporting can be overwhelming––leading adult ICU patients? patients?

Crit Care Med 2012 Vol. 40, No. 123263


Table 5.  Summary of clinical trials evaluating the use of glucose meters on blood from multiple sites for comparison of accuracy in various patient populations

Author Device Methodology Population Arterial POC vs. Laboratory

Cook et al (124) SureStepFlexxa 67 ICU patients NA


Single channel 67 samples
GO Glucose 62–218 mg/dL
vs. serum in laboratory Hct 22%–46.2%
Peripheral edema rated
Finkielman et al SureStepFlexxa 197 ICU patients Arterial and venous POC
(125) Single channel 816 samples Mean difference 7.9 ± 17.6 mg/dL
GO Retrospective data analysis LOA +43.1, −27.2
vs. plasma in laboratory
Lacara et al SureStepProa 49 ICU patients Arterial and venous POC
(126) GO 49 samples Bias 0.6
vs. laboratory (plasma or whole blood not Glucose 58–265 Precision 11.0 (p = .69)
specified) Hct 31.7 ± 0.8 sem
Atkin et al (127) Accu-Chek II 25 hypotensive patients NA
GD 39 normotensive patients
vs. serum in laboratory Glucose 52–485

Desachy Accu-Chek 103 patients Arterial and venous POC7% differ-


et al (135) GD 273 samples ent from laboratory by >20%
vs. laboratory assay (plasma or whole blood Glucose 56–675 mg/dL LOA 42.4, −39.5
not specified)
Kulkarni Accu-Chek Advantage 54 ICU patients NA
et al (128) GD 493 samples
vs. arterial whole blood gas analyzer Glucose 37.7–42.5 mg/dL
Capillary vs. arterial blood gas analyzer

Karon Accu-Cheka comfort curve 20 coronary artery bypass grafts patients Bias 14 mg/dL (p = .02)
et al (129) GD 14 on pressors, none with systolic blood 56% of POC samples were within
Result is factored to agree with plasma results pressure <80 mm Hg 10% of laboratory
vs. plasma in laboratory No temperatures recorded
Kanji et al (130) Accu-Cheka Inform 30 ICU patients Overall: 69.9% agreement
GD 36 samples Vasopressor: 67.6% agreement over-
Result is factored to agree with plasma results Poor peripheral perfusion or vasopressor, all, 50% with glucose <80 mg/dL
vs. plasma in laboratory significant peripheral edema, postoperative Edema: 71.4% agreement, 55% with
glucose <80 mg/dL
Meex Accu-Cheka Inform 20 ICU samples Arterial and venous: ICU subset, no
et al (131) GD (plasma strips) difference between POC and blood
vs. whole blood gas analyzer gas analyzer (p > .05; r = .98)
Critchell Accu-Cheka Inform 80 consecutive ICU patients NA
et al (105) GD Pressors 25%
by trained technician vs. plasma in laboratory Edema 42%
Pressors and edema 48% 277 samples
Meynaar Accu-Chek Inform 32 ICU patients Bias 11 mg/dL
et al (123) GD 239 samples 90% of samples <75 mg/dL were
vs. serum in laboratory Glucose 25–288 mg/dL within 15 mg/dL of laboratory
90.4% of samples >75 mg/dL were
within 20% of laboratory
Ray et al (132) One Touch Profile, LifeScan 10 ICU patients Bias 0.7 mg/dL (95% confidence
GO Three in shock interval [−41, 40]), intraclass
vs. plasma in laboratory 105 samples correlation coefficient = 0.86,
Glucose 86–256 mg/dL p < .0001
Corstjens Precision PCx 19 ICU patients 93.7% values within the 95%
et al (133) GO 145 samples confidence interval (values
vs. blood gas analyzer not reported)
Few hypoglycemic values
Boyd et al (134) Medisense Precision Plus 20 ED patients NA
GO 20 samples
vs. whole blood in laboratory

POC, point-of-care; GO, glucose oxidase; ICU, intensive care unit; Hct, hematocrit; NA, not applicable/not available; LOA, limits of agreement (2 sd);
GD, glucose dehydrogenase.
Studies illustrate the variability of glucose meters in clinical use when measuring arterial or venous blood specimens compared with capillary specimens.
Error is increased in patients with peripheral edema, poor skin perfusion, or receiving vasopressors. Trials with exogenously spiked blood samples
were excluded.
a
Meters display plasma-equivalent glucose results.
3264 Crit Care Med 2012 Vol. 40, No. 12
Venous POC vs.
Venous POC vs. Laboratory Capillary POC vs. Laboratory Capillary POC Confounders

Bias 9.51 mg/dL Bias 9.54 mg/dL Bias 0.03 mg/dL Venous vs. finger stick No
Precision 8.44 mg/dL Precision 11.96 mg/dL No significant significant difference
21% samples >20 mg/dL difference 15% samples >20 mg/dL difference difference between Low Hct contributed to
LOA +26.5, −10.3 LOA +31.5, −12.5 samples difference between POC and
R2 = .288, p < .001 R2 = .280, p = .02 LOA +24.1, −24.0 laboratory
NA NA NA Overall agreement, but
potential error for individual
samples

NA Bias 2.1 mg/dL NA Low Hct and Pco2 contributed


Precision 12.3 mg/dL to glucose over prediction
p = .23

Control: mean value 95.8% ± 1.1% of Control: mean value 91.8% ± 1.6% of NA Mean value from different
laboratory value laboratory methods were different
Hypotension: 99.2% ± 2.5% Hypotension: 67.5% ± 5.7%, p < .001 vs. (p < .05)
p < .05 vs. laboratory value laboratory
32% incorrectly diagnosed as hypoglycemic
NA 15% different from laboratory by >20% NA Perfusion index from Phillips
LOA 58.3, −55.3 monitor identified patients
with poor correlation

NA Bias 2.15 mg/dL NA Adequate agreement unless


Precision 13.8 mg/dL patient has systolic blood
LOA 29.8, −2.5 pressure <90 mm Hg or on
Hypoperfusion: subset 75 samples vasopressors
Bias 4.0Precision 16.2 mg/dLLOA −36.9, 28.4
Bias 12 mg/dL (p = .001) Bias −1 mg/dL NA Bias became greater at glucose
63% of samples within 10% of laboratory 74% of samples within 10% of laboratory >160 mg/dL with all methods
More potential insulin dosing (p < .001). No report on vaso-
discrepancies pressor effect
NA Overall: 56.8% agreement NA Agreement = same insulin dose
Vasopressor: 61.1% agreement overall, 25% based on glucose
with glucose < 80 mg/dL Agreement significantly lower
Edema: 55.8% agreement, 23.8% with when glucose <80 mg/dL vs.
glucose <80 mg/dL >80 mg/dL in all
NA NA NA Overall POC results higher
(−3% to 24%) but correlation
with laboratory good
Bias 8.6 ± 18.6 mg/dL NA NA Finger-stick overestimated
LOA: 45.8, −28.6 mg/dL glucose more than underesti-
mated. Vasopressor predicted
disagreement in results
NA Hct 20%–44% did not influence results NA Correction Art Accu-chek ×
1.086 = plasma glucose

NA NA NA Did not assess impact of pH


or Hct

NA NA NA ABL715 blood gas analyzer


highly correlated with labora-
tory plasma assay

Peripheral cannula NA NA 10 samples outside 95% CI


Bias 18 mg/dL (95% CI [11, 25] mg/dL) Significant difference between
mean values (p < .001)

Crit Care Med 2012 Vol. 40, No. 123265


In the absence of compelling data, regular insulin, although 0.5 unit/mL treatment with SQ insulin therapy may
no recommendation can be made for or solutions may also be found in the lit- be adequate to maintain BG < 180 mg/
against the use of continuous glucose erature. Insulin may be mixed with 0.9% dL in select patients with low insulin
sensors in critical care patients. sodium chloride, lactated Ringer’s injec- requirements who are clinically stable.
[Quality of evidence: very low] tion, Ringer’s injection, or 5% dextrose. Krinsley (26) reported a mean BG level
The safety and potentially the effective- Insulin may be prepared in glass or plas- of approximately 122 mg/dL using a pro-
ness of insulin infusion therapy could be tic containers (polyvinyl chloride [PVC], tocol that used titrated doses of short-
improved with more frequent or continu- ethylene vinyl acetate, polyethylene, and acting insulin given via SQ injection
ous glucose measurement. Ultimately, a other polyolefin plastics), although loss every 3 hrs. However, patients with sig-
closed-loop system (artificial pancreas) will occur through adsorption to contain- nificant hyperglycemia at baseline, type
could be used to titrate insulin infusion ers and to IV tubing and filters. Adsorption 1 diabetes mellitus, or two consecutive
therapy and minimize glucose variability, is immediate upon contact, producing a BG > 200 mg/dL triggered initiation of an
as it has been demonstrated to be feasible bioavailability of approximately 50–60% insulin infusion according to the nurse-
(136). Continuous glucose sensors have in PVC with sustained stability for 168 managed protocol. Long-acting insulin
been developed to measure interstitial hrs (146). Factors such as storage tem- was added to the SQ regimen when fea-
and intravascular glucose concentrations, perature, concentration, and infusion rate sible and appropriate. The patient-specific
and this technology has been reviewed influence the extent of adsorption. A trial treatment protocol combining SQ and
(137–139). However, intravascular devices of various priming volumes of 10–50 mL IV insulin regimens demonstrated safety
remain in preclinical and limited clinical concluded that a 20-mL prime from a and efficacy in maintaining the BG con-
testing (136). 100-mL polyvinyl chloride bag contain- centration predominately within the goal
Interstitial measurement devices ing regular insulin, 1 unit/mL, produced range with excursions of BG > 180 mg/
may be subject to the same limitations insulin delivery through a 100-inch latex- dL in <10% and BG < 40 mg/dL in only
as finger-stick BG testing, related to free polypropylene IV infusion set that was 1.9% of patients. While this approach may
variable tissue perfusion, temperature, not statistically different from a 50-mL not be feasible in all settings, and patient
and local humoral factors in addition to priming volume (147). This maneuver outcome has not been compared with
delays related to glucose equilibration, should be repeated each time new tubing insulin–infusion-only protocols, it has
and need for calibration. Initial reports is initiated to maintain consistent insulin the potential to reduce the number of BG
of continuous interstitial glucose sensors delivery rates. The optimal priming vol- measurements and associated workload.
have demonstrated acceptable accuracy ume for syringe pump systems has not
in select patients (133, 140–143). Con- been reported. 11. How should adult ICU patients be
current norepinephrine infusion did not Accurate insulin administration strat- transitioned off IV insulin infusions?
alter the accuracy of continuous SQ glu- egies include use of a reliable infusion A.  W
e suggest that stable ICU
cose monitoring (140). Additional evalua- pump for insulin administration, ideally patients should be transitioned
tion of accuracy and utility of continuous with safety software that prevents inad- to a protocol-driven basal/bolus
monitoring in broad patient populations vertent overdosing. The pump must be insulin regimen before the insu-
is needed before these devices can be rec- able to deliver insulin dose increments of lin infusion is stopped to avoid a
ommended for routine use. In studies of <1 unit/hr for insulin-sensitive patients significant loss of GC.
pediatric postoperative cardiac surgical (148). While most insulin infusion pro-
patients and pediatric medical/surgical tocols employ regular human insulin, [Quality of evidence: very low]
ICU patients, correlation of continuous rapid-acting insulin aspart and glulisine Specific patient groups have been
interstitial glucose monitors with BG are also compatible with 0.9% sodium shown to benefit from transition to a
readings is acceptable (i.e., mean absolute chloride in IV admixture and are labeled scheduled SQ insulin regimen, including
relative difference of 17.6% and 15.2%) and studied for IV use (149–151). type 1 diabetes patients, type 2 diabetes
and unaffected by inotrope use, body tem- patients on insulin as outpatients, type 2
perature, body wall edema, patient size, or 10. What is the role for SQ insulin in diabetes patients receiving insulin infu-
insulin use (144, 145). adult ICU patients? sion at a rate of >0.5 unit/hr, or stress
hyperglycemia patients receiving insulin
9. How should IV insulin be prepared and Subcutaneous insulin may be an alter- infusion at a rate of >1 unit/hr (152–156).
administered? native treatment for selected ICU patients. However, transition to SQ insulin
[Quality of evidence: very low] should be delayed until there are no
We suggest continuous insulin infu- Intravenous insulin infusion is pre- planned interruptions of nutrition for
sion (1 unit/mL) therapy be initiated after ferred for patients with type 1 diabetes mel- procedures, until peripheral edema has
priming new tubing with a 20-mL waste litus, hemodynamically unstable patients resolved, and until off vasopressors. A
volume. with hyperglycemia, and also patients in protocol for transition leads to better glu-
[Quality of evidence: moderate] whom long-acting basal insulin should cose control than nonprotocol therapy
Titration of insulin therapy to an not be initiated due to changing clini- (157). Failure of SQ regimens to pro­
end point of tight GC requires the rapid cal status (hypothermia, edema, fre- duce or maintain GC (BG < 180 mg/dL)
response and immediate flexibility of a quent interruption of dextrose intake, should trigger redesign of the regimen or
continuous infusion. These infusions etc.). Subcutaneous insulin regimens resumption of insulin infusion therapy.
should be prepared in a standardized con- with basal and rapid-acting insulin are A retrospective review of 614 cardio-
centration, with most protocols report- frequently initiated after stabilization of thoracic patients determined the effec­
ing use of a 1 unit/mL solution of human BG with IV insulin. However, initiating tiveness of an IV (in the ICU) followed by

3266 Crit Care Med 2012 Vol. 40, No. 12


SQ (outside the ICU) regimen on mor­ stabilized. Additional correction doses can support in the ICU setting when possible
bidity and mortality (158). The authors be given to fine-tune GC every 3–6 hrs. (165). However, due to several factors
found the SQ regimen to be less nursing- common to the ICU (e.g., gastric stasis,
intensive and less costly in all patients, 12. What are the nutritional consider- interruption of enteral nutrition for tests/
but only those with a preexisting diagno- ations with IV insulin therapy in procedures, and anatomical anomalies),
sis of diabetes demonstrated significantly adult ICU patients? the amount of feeding that can be deliv-
lower rates of postoperative mortality. A. 
We suggest that the amount and ered enterally is generally less than the
Protocolized transition to an SQ regi- timing of carbohydrate intake amount delivered parenterally. Interrup-
men has been shown to decrease rebound should be evaluated when calcu- tion of enteral feeds was found to cause
hyperglycemia after infusion discontinu- lating insulin requirements. the majority of the hypoglycemic events
ation (159). B. 
We also suggest that GC proto- (62%) in the Leuven MICU trial, and simi-
cols should include instructions lar results were noted elsewhere (13, 167).
B. 
We suggest that calculation of to address unplanned discontinu- Initiation of a 5% dextrose-containing IV
basal and bolus insulin dosing ance of any form of carbohydrate solution at the same rate as the discon-
requirements should be based on infusion. tinued enteral feeding solution appears
the patient’s IV insulin infusion to prevent hypoglycemia (168). Dextrose
history and carbohydrate intake. [Quality of evidence: low] (10%) solutions may be used to minimize
Nutritional support requirements of the volume of free water.
[Quality of evidence: very low] critically ill patients vary and are beyond Integration of an insulin protocol
Several models have been proposed the scope of this discussion. Guidelines with nutritional intervention has been
for transition from insulin infusion to for nutritional support of critically ill suggested to achieve a high level of GC.
SQ insulin therapy (156, 158–161). The patients are available (165). The Specialized Relative Insulin Nutrition
majority of these models include a three- Consistent intake of nutrition appears Tables protocol titrates both feeding and
component approach to insulin replace- to simplify glycemic management dur- insulin doses to achieve tight glucose
ment: basal insulin, nutritional insulin, ing an insulin infusion. Overfeeding may control and was more effective at achiev-
and correction insulin. Basal insulin is produce hyperglycemia that necessitates ing the BG target than a retrospective
provided as an injection of long-acting insulin infusion therapy, and should be control (169, 170). Insulin was admin-
insulin given every 24 hrs (e.g., glargine) avoided. istered with hourly bolus injections and
or intermediate-acting insulin given every Provision of 200–300 g of dextrose per could be supplemented by an infusion of
6–12 hrs (e.g., NPH). Basal insulin will day was a component of the initial trial up to 6 units/hr. The rate of enteral feed-
be needed in many diabetic patients on by Van den Berghe et al (1) in surgical ing was also adjusted to facilitate GC, but
enteral feedings to achieve the desired BG ICU patients. The reduction of mortality resulted in delivery of only 50% of the
goal (162). The initial basal insulin dose reported with achievement of BG values predicted caloric requirement, and thus
is recommended at least 2–4 hrs before of 80–110 mg/dL has been suggested to may not be an optimal long-term nutri-
stopping the insulin infusion when pos- reflect minimization of complications tional strategy.
sible to prevent rebound hyperglycemia from PN, although similar calories were Bolus doses of IV insulin may be
(28, 163). If this overlap is not feasible, provided in the medical ICU study, with- administered for nutritional insulin ther-
a simultaneous injection of rapid-acting out the same impact on outcome (14). apy during an insulin infusion when car-
insulin (approximately 10% of the basal While a meta-analysis of clinical trials, bohydrates are delivered intermittently,
dose) may be given with the basal insu- stratified by source of calories, suggested based on carbohydrate ratio, as previously
lin injection when stopping the infusion that tight GC is potentially more ben- discussed. Consistent oral intake should
(156). One group suggests calculating a eficial during PN regimen compared with trigger transition to SQ insulin therapy
total daily dose (TDD) of IV insulin from enteral feeding, this was not confirmed in and a consistent carbohydrate diet plan.
the mean hourly dose for at least the prior a prospective trial comparing early vs. late Glucose monitoring should be scheduled
6 hrs as a guide to the basal insulin dose PN (33, 166). Tight GC (mean BG 100–110 to avoid measurement of postprandial BG
(28). As IV insulin delivery is reduced mg/dL) was similarly achieved in patients concentrations.
by adsorption to the container and tub- who received 3–4 g/kg/d of carbohy-
ing, the authors reduced the initial basal drate (early) compared with 0.5–2 g/kg/d 13. What factors should be considered
dose to 80% of the estimated TDD and (late) over the first 7 ICU days (166). The for safe insulin therapy programs in
achieved their target for glucose control patients on early PN required higher total the adult ICU?
more readily than using smaller percent- insulin doses per day but fewer patients
ages of the TDD, although others have had SH (2% vs. 3.5%, p = .001). Neverthe- We suggest that insulin is a high-risk
shown acceptable glucose control using less, the patients on late PN (who received medication, and that a systems-based
60%–70% of the TDD (156, 158, 159, carbohydrates from enteral nutrition and approach is needed to reduce errors.
164). It is important to consider concur- 5% dextrose infusion for the first week) [Quality of evidence: very low]
rent changes in other drug therapy or had better overall outcomes. Thus, insu- Insulin is a high-alert, high-risk medi-
nutritional regimens when planning a lin infusion appears to be suitable for cation due to the risk of hypoglycemia,
transition regimen. patients regardless of the source of car- complexity of therapeutic regimens,
Mixing insulin in a parenteral nutri- bohydrates, and GC alone is not enough and availability of multiple products in
tion (PN) solution can replace a separate to reduce the apparent risks associated patient-care areas. It is in the top five
insulin infusion or basal insulin injec- with PN. The enteral route is preferred “high-risk” medications that account for
tions once the daily requirements are over the parenteral route for nutrition about one third of all major drug-related,

Crit Care Med 2012 Vol. 40, No. 123267


injurious medication errors. One analy- A standard insulin infusion protocol doses, and others require multiple steps
sis indicated that 33% of errors caus- should include a requirement for con- to alter insulin dosing, which can lead to
ing death within 48 hrs involved insulin tinuous glucose intake, standardized IV markedly different insulin doses in a sim-
therapy (171). Strategies to reduce such insulin infusion preparation, a dosing ulated patient model (183).
errors have been suggested and should format requiring minimal bedside deci- The original protocol published by Van
be applied to the ICU setting (172). These sion-making, frequent BG monitoring, den Berghe et al (1) (Leuven protocol) was
include standardized protocols for insu- provisions for dextrose replacement if relatively unstructured, although it was
lin dosing and monitoring, computer- feedings are interrupted, and protocolized successfully administered in a research
ized provider order entry, minimizing dextrose dosing for prompt treatment of setting with trained providers. Subse-
available insulin products, avoidance of hypoglycemia. quent use by bedside providers in other
abbreviations such as “U” for units, stor- [Quality of evidence: very low] ICU settings has produced hypoglycemia
ing insulin away from other medications, A standard protocol for insulin admin- rates that were deemed to be excessive
and detailed multiprofessional analysis of istration and monitoring is essential for (15, 17).
actual errors and near-miss events. Strat- consistency and safety. Comparison of Advantages of paper-based protocols
egies to improve insulin safety include existing protocols is difficult due to sig- include easy bedside access, insulin rate
mandating an independent double-check nificant differences in processes and out- changes are made only when outside of
of doses, frequent BG monitoring, and come measures, but key features will be goal BG ranges, and sometimes separate
prominent product labeling. discussed. scales for differing levels of insulin sen-
The limitations of BG monitoring Computerized decision-support sys- sitivity. The major disadvantages of these
equipment and methodology may also tems achieved better glucose control than protocols include their complexity (with
increase the risk of error. For example, that achieved with paper-based systems multiple recommendations on the same
factitious elevations in BG occur when using “if–then” decision model (175). page), a lack of flexibility with major
icodextrin peritoneal dialysis solutions Although paper-based systems may be clinical changes, and lag time to respond
or maltodextrin-containing medications adequate, they may be more complex and to BG trends (may recommend a dose
(selected immune globulin products) are time-consuming and lack a reminder sys- increase for a persistently high BG, even
administered and monitored with a glu- tem to ensure timely BG measurement. if the BG level has actually declined).
cose dehydrogenase monitoring system Most of the studies comparing protocols A more straightforward approach is to
(173). Also, a dextrose solution adminis- employed pre- and post-intervention use an algebraic formula to calculate the
tered via a pressurized flush system pro­ cohort design, limiting the ability to con- insulin rate based on the BG and a multi-
duced factitious elevations in BG values clude if the new protocol was the cause of plier (M) that relates to insulin sensitivity
drawn through an arterial line, and sub­ improved results. However, several RCTs (insulin dose [unit/hr] = [BG − 60] × M)
sequent inappropriate insulin administra­ demonstrated favorable features of com- (95–97, 184). This calculation can be
tion led to fatal neuroglycopenia (174). puterized insulin infusion protocols vs. computerized, assisted by a tabular for-
Safety in insulin administration meth- paper-based systems (148, 176–178). Gly- mat, or calculated manually (185, 186).
odology is also important, and a systems- cemic control metrics and hypoglycemia The multiplier increases for BG above
based approach is needed to reduce rates have been consistently better with the target range and decreases when
insulin errors. Complex insulin therapy computerized protocols. Reminder alerts the BG is below the goal. Advantages to
protocols with multiple patient-specific lead to more consistent and timely BG this approach include rapid determina-
exceptions and the need for a high-level assessments. Commercial systems have tion of the new insulin dose without the
training for accurate use are common. licensing fees that may be a barrier to need for extensive judgment or training
A standardized protocol should be utilized utilization, although several institutions of the bedside caregiver and constant
only after adequate education and pro- have developed custom computer-based titration based on the BG trend. It has
cesses are implemented to monitor out- systems (96, 97). The largest trial, NICE- resulted in some of the lowest reported
comes. Routine and frequent assessment SUGAR, had a computer-assisted proto- rates of severe hypoglycemia (177, 182,
of glucose metrics, as will be described, col, but dosing was based on a complex 183). Disadvantages include the need for
should be performed. Failure to achieve decision tree, rather than a specific set of a bedside computer and the potential for
adequate glucose control or frequent formulas (16). It should be noted that this exaggerated increases in insulin infusion
episodes of hypoglycemia should trigger protocol failed to achieve an average BG rate in response to an elevated BG value,
rapid reassessment of the protocol and level within the goal range of 80–110 mg/ especially with a high multiplier. The
monitoring system. dL. multiplier may need to be reset to a lower
Numerous cohort reports describe the value, especially following a significant
14. What are the characteristics of an utility and effectiveness of paper-based change in nutritional intake or change in
optimal insulin dosing protocol for protocols as they evolve over time, com- clinical status. More sophisticated com-
the adult ICU population? pared with historical controls (23, 151, puterized protocols have also been devel-
We suggest that ICUs develop a pro- 152, 179–182). The reports are of low qual- oped and have similarly been shown to
tocolized approach to manage GC. ity due to small study size, single-center perform better than conventional proto-
Components include a validated insu- experience, use of historical controls, and cols (169, 170, 178, 187, 188). Computer-
lin administration protocol, appropri- variable outcome measures (including ized programs can also collect data on the
ate staffing resources, use of accurate surrogate measures such as BG results performance of the program and calculate
monitoring technologies, and a robust rather than patient outcomes). These pro- a variety of metrics.
data platform to monitor protocol perfor- tocols vary in insulin dosing intensity and A source of error with virtually all insu-
mance and clinical outcome measures. complexity. Some contain insulin bolus lin protocols is incorrect transcription of

3268 Crit Care Med 2012 Vol. 40, No. 12


BG values into a freestanding computer Glycemic variability has been indepen- charts of percentage BG < 150 mg/dL and
program, which may occur approximately dently associated with mortality in several 180 mg/dL. We suggest that hypoglycemic
5% of the time (189). Similarly, protocol cohorts of critically ill patients; however, events should be monitored regularly and
violations are reported with paper-based there is no consensus regarding the appro- reported as events per patient, as a per-
systems (190). The amount of practitioner priate metric for mathematically defining centage of all BG values, and events per
latitude in deviating from the protocol GV. We suggest that the simplest tools––sd 100 hrs of insulin infusion.
recommendations should be predefined of each patient’s mean BG and coefficient [Quality of evidence: very low]
and evaluated as a component of quality of variation (sd/mean)––be reported in all This is a consensus suggestion to
assurance programs. published interventional studies. improve the safety and efficacy of GC and
With many published protocols avail- [Quality of evidence: very low] insulin therapy. Data on the performance
able, there is no need to reinvent the Glucose metrics are important to of an insulin infusion protocol should
wheel to implement an insulin infusion evaluate the overall results of a GC pro- be assessed multiple times throughout
protocol. The local barriers to safe insulin gram. In clinical trials, glucose variability the year (e.g., at least quarterly). Poten-
has been suggested as a better end point tial measures of protocol effectiveness
therapy must be identified and addressed,
to assess the impact of blood sugar on include global measures of BG control,
including availability of adequate and
patient outcome during insulin infusion such as mean and median BG per patient,
appropriate testing equipment, consid-
compared with other measures, such as measures of glucose variability, and time
eration of workforce impact, and a team
mean morning BG, mean of all BG values, to specific end points, including mean and
approach to education and implemen- or time-weighted average value. Higher median time required to reach the des-
tation (191). Tight levels of BG control levels of GV have been independently ignated glycemic target as well as mean
should not be attempted when a new pro- associated with mortality in adult cohorts and median time spent within the desired
tocol is initiated, to minimize hypoglyce- of mixed medical–surgical patients (83, glycemic range, reported as a percentage
mia risk during the initial learning curve. 86), surgical ICU patients (192), patients of total time in range (200–202). Patients
Systematic and frequent assessment of admitted with sepsis (193), as well as with diabetic ketoacidosis and hyper-
results is needed. Feedback to providers in critically ill pediatric patients (194). glycemic hyperosmolar coma should be
is essential when protocol violations or However, the most appropriate metric excluded from this analysis.
adverse events occur. In addition, a proto- to describe GV has not yet been defined. Protocol safety should be regularly
col is only effective if used in a consistent Relatively simple measures to calculate assessed through metrics relating to
fashion. Automatic triggers for protocol variability include sd, coefficient of varia- hypoglycemia, which should be defined as
initiation are more efficient than waiting tion, and mean daily delta (maximum − severe (<40 mg/dL), moderate (40–59 mg/
for prescriber recognition of hyperglyce- minimum BG). More complex measures dL), or mild (60–69 mg/dL). A system to
mia and appropriate response through that have been evaluated in different stud- evaluate patients with SH should ana-
patient-specific orders. ies include mean amplitude of glycemic lyze precipitating events and plan for
Other keys to a successful glycemic excursion, the glycemic lability index, prevention. The hypoglycemia event rate
management program include the avail- maximal glucose change, and the vari- could include patients with hypoglycemia
ability of a reliable methodology for BG ability index (195, 196). related to other treatments, such as oral
testing, with an adequate number of A recent review summarized the hypoglycemic agents or disease states
devices to minimize delays and wasted biologic basis for the deleterious effect such as hepatic failure or sepsis. There
time obtaining the device. The data should of increased GV (196). One purported are no existing benchmarks to establish a
be recorded in the electronic medical mechanism is the “oxidative stress” that goal, other than the lowest rate possible.
record promptly and be displayed along occurs at the cellular level induced by Although a hypoglycemia rate is
with insulin dosing adjustments to rapid changes in the BG level (93, 197, important for the overall assessment of
assess protocol performance and allow 198). Fluctuations in BG levels may lead a protocol, the impact of a single, severe
to changes in serum osmolality that cause hypoglycemic event cannot be overlooked
evaluation of variances. In addition,
injury at the cellular and organ levels or minimized by metrics that compress
the glycemic management program
(199). Finally, wide excursions may mask the GC measure into one global variable
should be coordinated with nutrition
occult hypoglycemia, which has been or BG averaging method.
support interventions to minimize the
recognized as a risk factor for mortality Other measures of glycemic perfor-
risk of hyperglycemia or hypoglycemia in the critically ill (82). It is not known mance have been studied in select popu-
with addition or interruption of nutri- yet whether efforts to minimize GV will lations. The percentage of patients with a
tional intake. Concurrent medications decrease the mortality rate in critically morning BG <200 mg/dL for the 3 days
dosed intermittently should be mixed ill patients, but this remains a promising after cardiovascular surgery is a compo-
in sodium chloride solutions to reduce avenue for future research. nent of the Surgical Care Improvement
glucose variation induced by episodic Project Measures based on the association
dextrose administration. While patients 16. What metrics are needed to evaluate of improved glucose control with fewer
should receive a consistent carbohydrate the quality and safety of an insulin deep sternal wound infections (203).
intake, the need for insulin may be mini- infusion protocol and GC program in Time-weighted mean BG, as used in
mized by limiting the infusion of exces- the adult ICU? NICE-SUGAR, may provide a more accu-
sive quantities of dextrose solutions. rate assessment of overall per-patient BG
Measures of overall glucose control control, but is more complex to calculate
15. What is the impact of GV on out- should include mean (sd) and median than a simple mean BG measurement (16).
comes of critically ill patients? (IQR) BG levels as well as ICU-level run The Glycemic Penalty Index is another

Crit Care Med 2012 Vol. 40, No. 123269


measure of the consistency of glucose BG, which extrapolated to nearly 2 hrs of multi-ICU glycemic management pro-
control (88). This tool scores glucose val- nursing time each day for insulin infusion gram and hospital and patient outcome
ues based on the degree of excursion from management. The design of this observa- variables. The participating ICUs dem-
the goal, making it a more dynamic mea- tional study did not include calculation of onstrated a reduction in mean BG com-
sure of the variability of glucose values in total paid nursing hours. Another time– pared with nonparticipating units in
a single patient. A higher value indicates motion study noted a marked difference the hospital. Outcomes were compared
fewer values within the goal range. This in the time required for GC activities with in the groups to address the impact of
tool has been used to compare insulin a paper protocol, depending on clinical secular time trends and patient char-
infusion protocols, but not to evaluate urgency. Malesker et al (208) reported acteristics that might have altered the
patient outcome. The Hyperglycemic a mean of 2.24 (±1.67) mins from BG results in this before and after study.
Index measures the area under the curve to therapeutic action and 10.55 (±3.24) The glycemic management protocol was
of BG values above the upper limit of the mins for hyperglycemia, although multi- associated with an average reduction
goal range vs. time (204). This method has tasking by nurses makes discreet evalua- of 1.19 days of ICU care per admission
shown a significant association with mor- tion of this activity more challenging. The (p ≤ .05) and a trend toward lower mor-
tality when used for retrospective analy- complete time from meter acquisition to tality and resource use including a reduc-
sis of BG values for surgical ICU patients. completion of documentation might have tion of $4,746 in total costs per patient
This metric is most meaningful when the been as long as 33 mins for adjustment of (−$10,509 to $1,832).
daily number of BG values is consistent infusion therapy, and longer for infusion
from patient to patient. initiation. 18. What are the implications of hyper-
There are few published studies of the glycemia in pediatric critically ill
17. What are the economic and work- effect of tight GC implementation on ICU patients?
force impacts of a GC program in the costs. Van den Berghe et al (209) per-
adult ICU? formed an analysis of the 1,548-patient In the absence of compelling data, no
A. 
We recommend that programs to cohort from their landmark surgical recommendations could be made for or
monitor and treat hyperglycemia ICU study. The methodology consisted against the use of tight GC in pediatric
in critically ill patients be imple- of a cost accounting of the components critical care patients.
mented to reduce hospital costs. of care found to change significantly as Hyperglycemia is highly prevalent
a result of intensive insulin therapy: the in pediatric critical care. While studies
[Quality of evidence: moderate] direct cost of insulin administration, ICU show an independent association between
days, mechanical ventilation, and the use hyperglycemia and morbidity and mortal-
B. 
We suggest implementation of of vasopressors, inotropes, IV antibiotics, ity rates, the paucity of data has resulted
programs to monitor and treat and blood transfusion. The total savings in practice variability (194). As in adults,
hyperglycemia in diabetic patients per patient associated with the intensive children develop critical illness hypergly­
following cardiovascular surgery insulin protocol was $2,638 per patient. cemia with no history of premorbid diabe­
to reduce hospital costs. The mean LOS in the conventional treat- tes or insulin resistance related to severity
ment and intensive treatment groups was of illness. Although most pediatric inten­
[Quality of evidence: low] 8.6 and 6.6 days, respectively, accounting sivists believe that hyperglycemia may
The cost implications of implemen- for over 80% of the cost per patient. cause harm in their patients and support
tation of programs to monitor and treat The cost implications of a 1,600-patient the concept of avoiding hyperglycemia,
hyperglycemia in hospitalized patients pre- and post-intervention cohort study most are reluctant to practice routine
have been studied in a variety of different of tight GC were implemented in a mixed GC (212, 213). An RCT of 700 critically
patient populations. Complications asso- medical–surgical ICU of a university- ill pediatric patients was completed in a
ciated with poor GC have the potential to affiliated community teaching hospital single center in Leuven, Belgium, which
increase total hospital costs. A reduction (210). These investigators attempted to established that insulin infusion titrated
in sternal wound infections was associ- quantify all major components of the cost to a goal of 50–80 mg/dL in infants and
ated with improved GC and produced of care: ICU days, mechanical ventilation 70–100 mg/dL in children, compared
lower costs (205). This single-center time, laboratory testing, pharmacy, diag- with insulin infusion only to prevent BG
investigation estimated that each 50 mg/ nostic imaging, and days in the hospital >215 mg/dL, improved short-term out­
dL increase in mean BG level was associ­ on the regular wards after discharge from comes (214). The absolute risk of mor-
ated with an excess of $2,824 in the cost of the ICU. The net savings per patient was tality was reduced by 3% (conventional
hospitalization. Promulgation of a hospi­ $1,580. The 17% decrease in ICU mean 5.7% vs. interventional 2.6%, p = .038),
tal-wide inpatient diabetes management LOS (from 4.1 to 3.4 days) accounted and insulin therapy also reduced the ICU
program produced a reduction in LOS for 28% of the savings, but there were LOS and C-reactive protein (the primary
that resulted in over $2 million in savings also substantial savings associated with outcome variable). The study was notable
to another facility (206). However, total decreased use of mechanical ventilation, for its first proof of principle that tighter
cost is not the only important measure of diagnostic imaging, laboratory testing, levels of GC produce clinical benefit. It
the impact of GC programs. Aragon eval­ and days in the hospital after discharge was also remarkable for its low target BG
uated the nursing work burden imposed from the ICU. ranges in the intervention groups, which
by an IV insulin protocol on four differ­ A third report from Sadhu and col- were described as “age-adjusted normo-
ent ICUs within a single academic insti­ leagues (211) used a difference-in-dif- glycemia” (50–80 mg/dL in those <1 yr
tution (207). A mean of 4.7 (±1.1) mins ferences (quasi-experimental) design old, 70–100 mg/dL in those >1 yr old).
was needed for each hourly analysis of to measure an association between a Although several outcomes in this trial

3270 Crit Care Med 2012 Vol. 40, No. 12


were favorable, there were extremely high infants, meticulous BG monitoring will as ICU LOS, rate of infection, or organ
rates of SH (<40 mg/dL): 44% in those <1 be crucial in pediatric insulin infusion dysfunction score.
yr old and 25% overall. In light of this, the protocols. Frequent BG monitoring, and 3. Methodology in proposed trials should
protocol is unlikely to be replicated out- ideally continuous glucose monitoring be as safe as possible and replicable
side Leuven, and the findings of clinical (145, 219), combined with explicit, pref­ in a naïve ICU setting. Study design
benefit cannot be widely applied. Of note, erably computer-assisted, algorithms will should target a range that may be
the importance of the hypoglycemia rates likely augment the safety and acceptance safely achieved without excessive (>5%)
will ultimately need to be reinterpreted in of these protocols. rates of severe hypoglycemia.
light of the neurocognitive outcomes in Stronger practice recommendations
these subjects, which is being assessed as and optimal glycemic targets in pediatric
a follow-up study. critical care can only come with the pub­
REFERENCES
Despite the inherent flaws of the ret- lication and confirmation of clinical trials
rospective pediatric literature and the with explicit methodologies in critically ill 1. Van den Berghe G, Wouters P, Weekers F,
single prospective RCT, many pediatric children (220, 221). An RCT of insulin infu- et al: Intensive insulin therapy in critically ill
intensivists believe hyperglycemia should sion (target BG 80–110 mg/dL) vs. standard patients. N Engl J Med 2001; 345:1359–1367
be avoided, and some pediatric ICUs have care produced improved glycemic control 2. Bagshaw SM, Egi M, George C, et al; Australia
New Zealand Intensive Care Society Database
implemented GC measures into their but did not reduce nosocomial infections,
Management Committee: Early blood glucose
standard clinical care. Recent studies have mortality, length of stay, or other morbid-
control and mortality in critically ill patients
shown that pediatric-specific GC proto- ity measures (222). Insulin infusion was in Australia. Crit Care Med 2009; 37:463–470
cols can be implemented in different ICU accomplished safely with SH reported in 3. Falciglia M, Freyberg RW, Almenoff PL, et al:
settings and afford seemingly reasonable only 3% of tight GC patients. This study Hyperglycemia-related mortality in critically
control with low rates of hypoglycemia does not alter the recommendation. ill patients varies with admission diagnosis.
(215–218). However, no formal recom- Crit Care Med 2009; 37:3001–3009
mendation can be made in favor of broad FUTURE RESEARCH 4. Grading of Recommendations Assessment,
implementation of GC to a low range. Development and Evaluation (GRADE) Work-
Regular internal “quality” evaluations of Although data have been generated ing Group Web site. Available at: http://www.
gradeworkinggroup.org. Accessed ­January 7,
GC in individual practices will likely assist in numerous subpopulations of criti-
2011
in refining and improving practice. cally ill patients, not all populations have
5. Review Manager (RevMan) [computer
In recognition of the distinct physi- been adequately studied and a “one size program]. Version 5.0. Copenhagen: The
ology and pathophysiology of children, fits all” treatment approach may not be Nordic Cochrane Centre, The Cochrane Col-
more clinical trials evaluating pediatric- appropriate for different institutions and laboration, 2008. Available at: http://www.
specific GC protocols in the different criti- patients. Furthermore, a critical reading ims.cochrane.org/revman/about-revman-5.
cal care disciplines (i.e., medical, surgical, of the published literature indicates that Accessed January 7, 2011
trauma, and cardiac), with an emphasis on these different populations have variable 6. GRADEprofiler [computer program]. Avail-
safety, are urgently needed. End points of responses; thus, survival benefit in one able at: http://www.ims.cochrane.org/revman/
any pediatric GC study will ideally include population may not be extrapolated to gradepro. Accessed January 7, 2011
7. American Diabetes Association: Standards
safety (hypoglycemia rates) and efficacy another. As such, more prospective RCTs
of medical care in diabetes – 2012. Diabetes
(time in goal BG range), length of ICU/ are needed in populations that have yet
Care 2012; 35(Suppl 1):S11–S63
hospital stay, ventilator and pressor/ino- to be adequately studied. Trials should be 8. Umpierrez GE, Isaacs SD, Bazargan N, et al:
trope days, rates of nosocomial infection designed to include several key features: Hyperglycemia: An independent marker of
and mortality, as well as rehabilitation and in-hospital mortality in patients with undi-
long-term neurodevelopmental outcome. 1. Inclusion and exclusion criteria should agnosed diabetes. J Clin Endocrinol Metab
The strongest recommendation that can reasonably define a unique population. 2002; 87:978–982
be made at this time is that it is reason- In light of the unique benefits to post- 9. Finney SJ, Zekveld C, Elia A, et al: Glucose
able to incorporate approaches to control operative cardiac surgical patients, for control and mortality in critically ill patients.
persistent significant hyperglycemia (i.e., example, trials should not mix cardiac JAMA 2003; 290:2041–2047
BG levels >180–220 mg/dL) into practice. and noncardiac patients unless the 10. Freire AX, Bridges L, Umpierrez GE, et al:
Admission hyperglycemia and other risk
An optimal glycemic range currently can- study design provides adequate power
factors as predictors of hospital mortality
not be recommended due to lack of pedi- to measure outcomes in each group. in a medical ICU population. Chest 2005;
atric-specific data. Yet, for those opting 2. Studies should be adequately powered 128:3109–3116
to practice GC in line with adult efforts, to detect clinically significant out- 11. Krinsley JS: Association between hyper-

choosing a target BG that is in the range comes. In adults, 28-day and in-hos- glycemia and increased hospital mortality
of 100–180 mg/dL may be a reasonable pital mortality should be considered in a heterogeneous population of criti-
goal. This suggestion should not preclude primary outcomes in most popula- cally ill patients. Mayo Clin Proc 2003; 78:
alternative glycemic targets, depending tions; additionally, in surgical patients 1471–1478
on the practice group’s comfort and expe- (i.e., coronary artery bypass graft) with 12. de Azevedo JRA, de Araujo LE, da Silva WS,
et al: A carbohydrate-restrictive strategy is
rience. Although children do have lower a low mortality rate, hospital com-
safer and as efficient as intensive insulin
basal BG levels than adults, levels <60 mg/ plications and costs may represent
therapy in critically ill patients. J Crit Care
dL should be minimized, and BG levels important secondary outcomes. Simi- 2010; 25:84–89
<40 mg/dL should be treated emergently. larly, in pediatrics, with very low ICU 13. Van den Berghe G, Wouters PJ, Bouillon

Due to the sensitivity of the developing mortality, surrogate outcomes may R, et al: Outcome benefit of intensive insu-
central nervous systems of neonates and need to be the primary outcomes, such lin therapy in the critically ill: Insulin dose

Crit Care Med 2012 Vol. 40, No. 123271


versus glycemic control. Crit Care Med 2003; 29. Toft P, Jørgensen HS, Toennesen E, et al:
44. Gale SC, Sicoutris C, Reilly PM, et al: Poor
31:359–366 Intensive insulin therapy to non-cardiac ICU glycemic control is associated with increased
14. Van den Berghe G, Wilmer A, Hermans G, patients: A prospective study. Eur J Anaesthesiol mortality in critically ill trauma patients. Am
et al: Intensive insulin therapy in the medical 2006; 23:705–709 Surg 2007; 73:454–460
ICU. N Engl J Med 2006; 354:449–461 30. Schultz MJ, Harmsen RE, Spronk PE: Clini- 45. Bochicchio GV, Sung J, Joshi M, et al: Persistent
15. Preiser JC, Devos P, Ruiz-Santana S, et al: cal review: Strict or loose glycemic control hyperglycemia is predictive of outcome in
A prospective randomised multi-centre con- in critically ill patients—Implementing best critically ill trauma patients. J Trauma 2005;
trolled trial on tight glucose control by inten- available evidence from randomized con- 58:921–924
sive insulin therapy in adult intensive care trolled trials. Crit Care 2010; 14:223 46. Kao LS, Todd SR, Moore FA: The impact of
units: The Glucontrol study. Intensive Care 31. Wiener RS, Wiener DC, Larson RJ: ­Benefits diabetes on outcome in traumatically injured
Med 2009; 35:1738–1748 and risks of tight glucose control in critically patients: An analysis of the National Trauma
16. The NICE-SUGAR Investigators: Intensive
ill adults: A meta-analysis. JAMA 2008; 300: Data Bank. Am J Surg 2006; 192:710–714
versus conventional glucose control in 933–944 47. Reed CC, Stewart RM, Sherman M, et al:

critically ill patients. N Engl J Med 2009; 32. Griesdale DE, de Souza RJ, van Dam RM,
Intensive insulin protocol improves glucose
360:1283–1297 et al: Intensive insulin therapy and mortality control and is associated with a reduction in
17. Brunkhorst FM, Engel C, Bloos F, et al;
among critically ill patients: A meta-analysis intensive care unit mortality. J Am Coll Surg
German Competence Network Sepsis
­ including NICE-SUGAR study data. CMAJ 2007; 204:1048–1054; discussion 1054
(SepNet): Intensive insulin therapy and pen- 2009; 180:821–827 48. Collier B, Diaz J Jr, Forbes R, et al: The impact
tastarch resuscitation in severe sepsis. N Engl J 33. Marik PE, Preiser JC: Toward understanding of a normoglycemic management protocol
Med 2008; 358:125–139 tight glycemic control in the ICU: A system- on clinical outcomes in the trauma inten-
18. De La Rosa G, Donado JH, Restrepo AH, et al: atic review and metaanalysis. Chest 2010; sive care unit. JPEN J Parenter Enteral Nutr
Strict glycemic control in patients hospital- 137:544–551 2005; 29:353–358; discussion 359
ized in a mixed medical and surgical intensive 34. Ingels C, Debaveye Y, Milants I, et al: Strict 49. Melamed E: Reactive hyperglycaemia in

care unit: A randomized clinical trial. Crit blood glucose control with insulin during patients with acute stroke. J Neurol Sci 1976;
Care 2008; 12:R120 intensive care after cardiac surgery: Impact 29:267–275
19. Arabi YM, Dabbagh OC, Tamim HM, et al: on 4-years survival, dependency on medical 50. Pentelényi T, Kammerer L, Stützel M, et al:
Intensive versus conventional insulin ther- care, and quality-of-life. Eur Heart J 2006; Alterations of the basal serum insulin and
apy: A randomized controlled trial in medical 27:2716–2724 blood glucose in brain-injured patients.
and surgical critically ill patients. Crit Care 35. Zerr KJ, Furnary AP, Grunkemeier GL,
Injury 1979; 10:201–208
Med 2008; 36:3190–3197 et al: Glucose control lowers the risk of wound 51. Scott JF, Robinson GM, French JM, et al:

20. Farah R, Samokhvalov A, Zviebel F, et al:
infection in diabetics after open heart opera- Prevalence of admission hyperglycaemia
Insulin therapy of hyperglycemia in intensive tions. Ann Thorac Surg 1997; 63:356–361 across clinical subtypes of acute stroke.
care. Isr Med Assoc J 2007; 9:140–142 36. Furnary AP, Zerr KJ, Grunkemeier GL,
­Lancet 1999; 353:376–377
21. Grey NJ, Perdrizet GA: Reduction of nosoco- et al: Continuous intravenous insulin infu- 52. Kernan WN, Viscoli CM, Inzucchi SE,

mial infections in the surgical intensive-care sion reduces the incidence of deep sternal et al: Prevalence of abnormal glucose toler-
unit by strict glycemic control. Endocr Pract wound infection in diabetic patients after car- ance following a transient ischemic attack
2004; 10(Suppl 2):46–52 diac surgical procedures. Ann Thorac Surg or ischemic stroke. Arch Intern Med 2005;
22. Mackenzie IM, Ercole A, Blunt M, et al:
1999; 67:352–360; discussion 360 165:227–233
Glycaemic control and outcome in general
­ 37. Leibowitz G, Raizman E, Brezis M, et al:
53. Allport L, Baird T, Butcher K, et al: Frequency
intensive care: The East Anglian GLYCOGENIC Effects of moderate intensity glycemic con- and temporal profile of poststroke hypergly-
study. Br J Intensive Care 2008; 18:121–126 trol after cardiac surgery. Ann Thorac Surg cemia using continuous glucose monitoring.
23. Krinsley JS: Effect of an intensive glucose 2010; 90:1825–1832 Diabetes Care 2006; 29:1839–1844
management protocol on the mortality of 38. Schmeltz LR, DeSantis AJ, Thiyagarajan V, 54. Capes SE, Hunt D, Malmberg K, et al: Stress
critically ill adult patients. Mayo Clin Proc et al: Reduction of surgical mortality and mor- hyperglycemia and prognosis of stroke in
2004; 79:992–1000 bidity in diabetic patients undergoing cardiac nondiabetic and diabetic patients: A system-
24. Furnary AP, Gao G, Grunkemeier GL, et al: surgery with a combined intravenous and atic overview. Stroke 2001; 32:2426–2432
Continuous insulin infusion reduces mor- subcutaneous insulin glucose management 55. Bruno A, Levine SR, Frankel MR, et al;

tality in patients with diabetes undergoing strategy. Diabetes Care 2007; 30:823–828 NINDS rt-PA Stroke Study Group: Admission
coronary artery bypass grafting. J Thorac 39. Heath DF: Glucose, insulin and other plasma glucose level and clinical outcomes in the
Cardiovasc Surg 2003; 125:1007–1021 metabolites shortly after injury. J Accid NINDS rt-PA Stroke Trial. Neurology 2002;
25. Treggiari MM, Karir V, Yanez ND, et al: Inten- Emerg Med 1994; 11:67–77 59:669–674
sive insulin therapy and mortality in critically 40. Yendamuri S, Fulda GJ, Tinkoff GH: Admis- 56. Dimopoulou I, Kouyialis AT, Orfanos S, et al:
ill patients. Crit Care 2008; 12:R29 sion hyperglycemia as a prognostic indicator Endocrine alterations in critically ill patients
26. Krinsley JS: Glycemic control, diabetic status, in trauma. J Trauma 2003; 55:33–38 with stroke during the early recovery period.
and mortality in a heterogeneous population 41. Bochicchio GV, Bochicchio KM, Joshi M,
Neurocrit Care 2005; 3:224–229
of critically ill patients before and during the et al: Acute glucose elevation is highly pre- 57. Garg R, Chaudhuri A, Munschauer F, et al:
era of intensive glycemic management: Six dictive of infection and outcome in critically Hyperglycemia, insulin, and acute ischemic
and one-half years experience at a university- injured trauma patients. Ann Surg 2010; stroke: A mechanistic justification for a trial
affiliated community hospital. Semin Thorac 252:597–602 of insulin infusion therapy. Stroke 2006;
Cardiovasc Surg 2006; 18:317–325 42. Sung J, Bochicchio GV, Joshi M, et al: Admis- 37:267–273
27. Scalea TM, Bochicchio GV, Bochicchio KM, sion hyperglycemia is predictive of outcome 58. Demchuk AM, Morgenstern LB, Krieger

et al: Tight glycemic control in critically in critically ill trauma patients. J Trauma DW, et al: Serum glucose level and diabetes
injured trauma patients. Ann Surg 2007; 2005; 59:80–83 predict tissue plasminogen activator-related
246:605–610; discussion 610 43. Vogelzang M, Nijboer JM, van der Horst IC, intracerebral hemorrhage in acute ischemic
28. Furnary AP, Wu Y: Eliminating the diabetic et al: Hyperglycemia has a stronger relation stroke. Stroke 1999; 30:34–39
disadvantage: The Portland Diabetic Proj- with outcome in trauma patients than in 59. Fogelholm R, Murros K, Rissanen A, et al:
ect. Semin Thorac Cardiovasc Surg 2006; other critically ill patients. J Trauma 2006; Admission blood glucose and short term sur-
18:302–308 60:873–877; discussion 878 vival in primary intracerebral haemorrhage:

3272 Crit Care Med 2012 Vol. 40, No. 12


A population based study. J Neurol Neurosurg 73. Bilotta F, Caramia R, Cernak I, et al: Inten- and relative association with mortality. Crit
Psychiatr 2005; 76:349–353 sive insulin therapy after severe traumatic Care Med 2010; 38:1021–1029
60. Bilotta F, Spinelli A, Giovannini F, et al: The brain injury: A randomized clinical trial. 89. Malouf R, Brust JC: Hypoglycemia: Causes,
effect of intensive insulin therapy on infec- Neurocrit Care 2008; 9:159–166 neurological manifestations, and outcome.
tion rate, vasospasm, neurologic outcome, 74. Bilotta F, Caramia R, Paoloni FP, et al: Safety Ann Neurol 1985; 17:421–430
and mortality in neurointensive care unit and efficacy of intensive insulin therapy in 90. Ben-Ami H, Nagachandran P, Mendelson A,
after intracranial aneurysm clipping in critical neurosurgical patients. Anesthesiology et al: Drug-induced hypoglycemic coma in
patients with acute subarachnoid hemor- 2009; 110:611–619 102 diabetic patients. Arch Intern Med 1999;
rhage: A randomized prospective pilot trial. 75. Thiele RH, Pouratian N, Zuo Z, et al: Strict 159:281–284
J Neurosurg Anesthesiol 2007; 19:156–160 glucose control does not affect mortality 91. Duning T, Ellger B: Is hypoglycaemia dan-
61. Pasternak JJ, McGregor DG, Schroeder
after aneurysmal subarachnoid hemorrhage. gerous? Best Pract Res Clin Anaesthesiol
DR, et al; IHAST Investigators: Hyper- Anesthesiology 2009; 110:603–610 2009; 23:473–485
glycemia in patients undergoing cerebral 76. Zhu CZ, Auer RN: Optimal blood glucose 92. Malone JI, Hanna S, Saporta S, et al: Hyper-
aneurysm surgery: Its association with levels while using insulin to minimize the glycemia not hypoglycemia alters neuro-
long-term gross neurologic and neuropsy- size of infarction in focal cerebral ischemia. nal dendrites and impairs spatial memory.
chological function. Mayo Clin Proc 2008; J Neurosurg 2004; 101:664–668 ­Pediatr Diabetes 2008; 9:531–539
83:406–417 77. Oddo M, Schmidt JM, Carrera E, et al:
93. Suh SW, Gum ET, Hamby AM, et al:

62. Kruyt ND, Biessels GJ, de Haan RJ, et al: Impact of tight glycemic control on cere- ­Hypoglycemic neuronal death is triggered
Hyperglycemia and clinical outcome in bral glucose metabolism after severe brain by glucose reperfusion and activation of
aneurysmal subarachnoid hemorrhage: A injury: A microdialysis study. Crit Care Med neuronal NADPH oxidase. J Clin Invest
meta-analysis. Stroke 2009; 40:e424–e430 2008; 36:3233–3238 2007; 117:910–918
63. Rovlias A, Kotsou S: The influence of
78. Helbok R, Schmidt JM, Kurtz P, et al:
94. Balentine JR, Gaeta TJ, Kessler D, et al:

hyperglycemia on neurological outcome in ­Systemic glucose and brain energy metab- Effect of 50 milliliters of 50% dextrose in
patients with severe head injury. Neurosur- olism after subarachnoid hemorrhage. water administration on the blood sugar of
gery 2000; 46:335–342; discussion 342 ­Neurocrit Care 2010; 12:317–323 euglycemic volunteers. Acad Emerg Med
64. Jeremitsky E, Omert LA, Dunham CM,
79. Kosiborod M, Inzucchi SE, Goyal A, et al: 1998; 5:691–694
et al: The impact of hyperglycemia on Relationship between spontaneous and 95. Davidson PC, Steed RD, Bode BW: Glucom-
patients with severe brain injury. J Trauma iatrogenic hypoglycemia and mortality in mander: A computer-directed intravenous
2005; 58:47–50 patients hospitalized with acute myocardial insulin system shown to be safe, simple, and
65. Coester A, Neumann CR, Schmidt MI:
infarction. JAMA 2009; 301:1556–1564 effective in 120,618 h of operation. Diabetes
Intensive insulin therapy in severe trau- 80. Arabi YM, Tamim HM, Rishu AH: Hypo-
Care 2005; 28:2418–2423
matic brain injury: A randomized trial. glycemia with intensive insulin therapy in 96. Dortch MJ, Mowery NT, Ozdas A, et al: A com-
J Trauma 2010; 68:904–911 critically ill patients: Predisposing factors puterized insulin infusion titration protocol
66. Baird TA, Parsons MW, Phanh T, et al:
and association with mortality. Crit Care improves glucose control with less hypogly-
­Persistent poststroke hyperglycemia is inde- Med 2009; 37:2536–2544 cemia compared to a manual titration proto-
pendently associated with infarct expansion 81. Vriesendorp TM, van Santen S, DeVries JH, col in a trauma intensive care unit. JPEN J
and worse clinical outcome. Stroke 2003; et al: Predisposing factors for hypoglycemia Parenter Enteral Nutr 2008; 32:18–27
34:2208–2214 in the intensive care unit. Crit Care Med 97. Juneja R, Roudebush CP, Nasraway SA, et al:
67. Gentile NT, Seftchick MW, Huynh T, et al: 2006; 34:96–101 Computerized intensive insulin dosing can
Decreased mortality by normalizing blood 82. Krinsley JS, Grover A: Severe hypoglyce-
mitigate hypoglycemia and achieve tight
glucose after acute ischemic stroke. Acad mia in critically ill patients: Risk factors and glycemic control when glucose measure-
Emerg Med 2006; 13:174–180 outcomes. Crit Care Med 2007; 35: ment is performed frequently and on time.
68. Scott JF, Robinson GM, French JM, et al: 2262–2267 Crit Care 2009; 13:R163
Glucose potassium insulin infusions in the 83. Vriesendorp TM, DeVries JH, van Santen 98. Moore C, Woollard M: Dextrose 10% or

treatment of acute stroke patients with mild S, et al: Evaluation of short-term con- 50% in the treatment of hypoglycaemia out
to moderate hyperglycemia: The Glucose sequences of hypoglycemia in an inten- of hospital? A randomised controlled trial.
Insulin in Stroke Trial (GIST). Stroke 1999; sive care unit. Crit Care Med 2006; 34: Emerg Med J 2005; 22:512–515
30:793–799 2714–2718 99. Bhatia A, Cadman B, Mackenzie I: Hypogly-
69. Bruno A, Saha C, Williams LS, et al: IV insulin 84. Van den Berghe G, Wilmer A, Milants I, et al: cemia and cardiac arrest in a critically ill
during acute cerebral infarction in diabetic Intensive insulin therapy in mixed medical/ patient on strict glycemic control. Anesth
patients. Neurology 2004; 62:1441–1442 surgical intensive care units: Benefit versus Analg 2006; 102:549–551
70. Bell DA, Strong AJ: Glucose/insulin infu- harm. Diabetes 2006; 55:3151–3159 100. Carstens S, Sprehn M: Prehospital treat-

sions in the treatment of subarachnoid 85. Egi M, Bellomo R, Stachowski E, et al: ment of severe hypoglycaemia: A com-
haemorrhage: A feasibility study. Br J Neu- Hypoglycemia and outcome in critically parison of intramuscular glucagon and
rosurg 2005; 19:21–24 ill patients. Mayo Clin Proc 2010; 85: intravenous glucose. Prehosp Disaster Med
71. Walters MR, Weir CJ, Lees KR: A ran-
217–224 1998; 13:44–50
domised, controlled pilot study to investi- 86. Bagshaw SM, Bellomo R, Jacka MJ, et al: 101. Slama G, Traynard PY, Desplanque N,

gate the potential benefit of intervention The impact of early hypoglycemia and blood et al: The search for an optimized treat-
with insulin in hyperglycaemic acute isch- glucose variability on the outcome of criti- ment of hypoglycemia. Carbohydrates in
aemic stroke patients. Cerebrovasc Dis cal illness. Crit Care 2009; 13:R91 tablets, solutin, or gel for the correction of
2006; 22:116–122 87. Krinsley JS, Schultz MJ, Spronk PE, et al: insulin reactions. Arch Intern Med 1990;
72. Gray CS, Hildreth AJ, Sandercock PA,
Mild hypoglycemia is independently associ- 150:589–593
et  al; GIST Trialists Collaboration: Glu- ated with increased mortality in the criti- 102. Vogelzang M, Loef BG, Regtien JG, et al:
cose-potassium-insulin infusions in the cally ill. Crit Care 2011; 15:R173 Computer-assisted glucose control in criti-
management of post-stroke hyperglycae- 88. Meyfroidt G, Keenan DM, Wang X, et al: cally ill patients. Intensive Care Med 2008;
mia: The UK Glucose Insulin in Stroke Dynamic characteristics of blood glucose 34:1421–1427
Trial (GIST-UK). Lancet Neurol 2007; 6: time series during the course of critical 103. Scott MG, Bruns DE, Boyd JC, et al: Tight
397–406 illness: Effects of intensive insulin therapy glucose control in the intensive care unit:

Crit Care Med 2012 Vol. 40, No. 123273


Are glucose meters up to the task? Clin 118. Lopez C, Serrano LdL, Rodriquez R, et al: point-of-care and continuous blood glucose
Chem 2009; 55:18–20 Validation of a glucose meters at an intensive analysis in critically ill ICU patients. Crit
104. Clinical and Laboratory Standards Institute care unit. Endocrinol Nutr 2012; 59:28–34 Care 2006; 10:R135
(formerly National Committee for Clinical 119. Tang Z, Du X, Louie RF, et al: Effects of 134. Boyd R, Leigh B, Stuart P: Capillary versus
Laboratory Standards) Web site. Available drugs on glucose measurements with hand- venous bedside blood glucose estimations.
at: http://www.clsi.org. Accessed January 7, held glucose meters and a portable glucose Emerg Med J 2005; 22:177–179
2011 analyzer. Am J Clin Pathol 2000; 113:75–86 135. Desachy A, Vuagnat AC, Ghazali AD, et al:
105. Critchell CD, Savarese V, Callahan A, et al: 120. Scott RJ, Deobald G, Griesmann L, et al: Accuracy of bedside glucometry in critically
Accuracy of bedside capillary blood glucose Evaluation of multiple whole blood glu- ill patients: Influence of clinical character-
measurements in critically ill patients. cose methods compared with a laboratory istics and perfusion index. Mayo Clin Proc
Intensive Care Med 2007; 33:2079–2084 plasma hexokinase reference assay. Point of 2008; 83:400–405
106. Karon BS, Boyd JC, Klee GG: Glucose meter Care 2008; 7:43–46 136. Yatabe T, Yamazaki R, Kitagawa H, et al:
performance criteria for tight glycemic con- 121. Abbott Laboratories Web site: Product
The evaluation of the ability of closed-loop
trol estimated by simulation modeling. Clin info: cartridge and test information sheets. glycemic control device to maintain the
Chem 2010; 56:1091–1097 ­Available at: http://www.i-stat.com/products/ blood glucose concentration in intensive
107. Dungan K, Chapman J, Braithwaite SS,
ctisheets/714177-01H.pdf. Accessed July 15, care unit patients. Crit Care Med 2011;
et  al: Glucose measurement: Confounding 2012 39:575–578
issues in setting targets for inpatient man- 122. Mahoney JJ, Ellison JM: Assessing glu-
137. Kondepati VR, Heise HM: Recent progress
agement. Diabetes Care 2007; 30:403–409 cose monitor performance–A standardized in analytical instrumentation for glycemic
108. Tang Z, Louie RF, Payes M, et al: Oxygen approach. Diabetes Technol Ther 2007; control in diabetic and critically ill patients.
effects on glucose measurements with a ref- 9:545–552 Anal Bioanal Chem 2007; 388:545–563
erence analyzer and three handheld meters. 123. Meynaar IA, van Spreuwel M, Tangkau PL, 138. Aye T, Block J, Buckingham B: Toward clos-
Diabetes Technol Ther 2000; 2:349–362 et  al: Accuracy of AccuChek glucose mea- ing the loop: An update on insulin pumps
109. Tang Z, Louie RF, Lee JH, et al: Oxygen surement in intensive care patients. Crit and continuous glucose monitoring sys-
effects on glucose meter measurements Care Med 2009; 37:2691–2696 tems. Endocrinol Metab Clin North Am
with glucose dehydrogenase- and oxidase- 124. Cook A, Laughlin D, Moore M, et al: Dif- 2010; 39:609–624
based test strips for point-of-care testing. ferences in glucose values obtained from 139. Klonoff DC, Buckingham B, Christiansen

Crit Care Med 2001; 29:1062–1070 point-of-care glucose meters and laboratory JS, et al; Endocrine Society: Continuous
110. Mann EA, Salinas J, Pidcoke HF, et al: Error analysis in critically ill patients. Am J Crit glucose monitoring: An Endocrine Society
rates resulting from anemia can be cor- Care 2009; 18:65–71; quiz 72 Clinical Practice Guideline. J Clin Endocri-
rected in multiple commonly used point-of- 125. Finkielman JD, Oyen LJ, Afessa B: Agree- nol Metab 2011; 96:2968–2979
care glucometers. J Trauma 2008; 64:15–20; ment between bedside blood and plasma 140. Holzinger U, Warszawska J, Kitzberger R,
discussion 20 glucose measurement in the ICU setting. et al: Impact of shock requiring norepineph-
111. Hoedemaekers CW, Klein Gunnewiek JM, Chest 2005; 127:1749–1751 rine on the accuracy and reliability of sub-
Prinsen MA, et al: Accuracy of bedside glu- 126. Lacara T, Domagtoy C, Lickliter D, et al: cutaneous continuous glucose monitoring.
cose measurement from three glucometers Comparison of point-of-care and laboratory Intensive Care Med 2009; 35:1383–1389
in critically ill patients. Crit Care Med 2008; glucose analysis in critically ill patients. Am 141. Siegelaar SE, Barwari T, Hermanides J, et al:
36:3062–3066 J Crit Care 2007; 16:336–346; quiz 347 Accuracy and reliability of continuous glu-
112. Hoedemaekers CW, Klein Gunnewiek JM, 127. Atkin SH, Dasmahapatra A, Jaker MA,
cose monitoring in the intensive care unit:
Van der Hoeven JG: Point-of-care glucose et al: Fingerstick glucose determina- A head-to-head comparison of two subcu-
measurement systems should be used with tion in shock. Ann Intern Med 1991; 114: taneous glucose sensors in cardiac surgery
great caution in critically ill intensive care 1020–1024 patients. Diabetes Care 2011; 34:e31
unit patients. Crit Care Med 2010; 38:339; 128. Kulkarni A, Saxena M, Price G, et al: Analy- 142. Ellmerer M, Haluzik M, Blaha J, et al: Clinical
author reply 339–339; author reply 340 sis of blood glucose measurements using evaluation of alternative-site glucose mea-
113. Tang Z, Lee JH, Louie RF, et al: Effects of capillary and arterial blood samples in surements in patients after major cardiac
different hematocrit levels on glucose mea- intensive care patients. Intensive Care Med surgery. Diabetes Care 2006; 29:1275–1281
surements with handheld meters for point- 2005; 31:142–145 143. Yamashita K, Okabayashi T, Yokoyama T,

of-care testing. Arch Pathol Lab Med 2000; 129. Karon BS, Gandhi GY, Nuttall GA, et al: et al: The accuracy of a continuous blood
124:1135–1140 Accuracy of roche accu-chek inform whole glucose monitor during surgery. Anesth
114. Louie RF, Tang Z, Sutton DV, et al: Point- blood capillary, arterial, and venous glu- Analg 2008; 106:160–163, table of contents
of-care glucose testing: Effects of critical cose values in patients receiving intensive 144. Piper HG, Alexander JL, Shukla A, et al:
care variables, influence of reference instru- intravenous insulin therapy after car- Real-time continuous glucose monitoring
ments, and a modular glucose meter design. diac surgery. Am J Clin Pathol 2007; 127: in pediatric patients during and after cardiac
Arch Pathol Lab Med 2000; 124:257–266 919–926 surgery. Pediatrics 2006; 118:1176–1184
115. Holtzinger C, Szelag E, DuBois JA, et al: 130. Kanji S, Buffie J, Hutton B, et al: Reliability 145. Bridges BC, Preissig CM, Maher KO, et al:
Evaluation of a new POCT bedside glucose of point-of-care testing for glucose mea- Continuous glucose monitors prove highly
meter and strip with hematocrit and inter- surement in critically ill adults. Crit Care accurate in critically ill children. Crit Care
ference corrections. Point of Care 2008; Med 2005; 33:2778–2785 2010; 14:R176
7:1–6 131. Meex C, Poncin J, Chapelle JP, et al: Analyti- 146. Greenwood BC, Chesnick MA, Szumita PM,
116. Biljak VR, Božičević S, Lovrenčić MV, et al: cal validation of the new plasma calibrated et al: Stability of regular human insulin
Performance of the Statstrip glucose meter Accu-Chek Test Strips (Roche Diagnostics). extemporaneously prepared in 0.9% sodium
in inpatient management of diabetes melli- Clin Chem Lab Med 2006; 44:1376–1378 chloride in a polyvinyl chloride bag. Hosp
tus. Diabetol Croat 2010; 39:105–110 132. Ray JG, Hamielec C, Mastracci T: Pilot study Pharm 2012; 47:367–370
117. Pidcoke HF, Wade CE, Mann EA, et al: Ane- of the accuracy of bedside glucometry in 147. Goldberg PA, Kedves A, Walter K, et al:

mia causes hypoglycemia in intensive care the intensive care unit. Crit Care Med 2001; “Waste not, want not”: Determining the
unit patients due to error in single-channel 29:2205–2207 optimal priming volume for intravenous
glucometers: Methods of reducing patient 133. Corstjens AM, Ligtenberg JJ, van der
insulin infusions. Diabetes Technol Ther
risk. Crit Care Med 2010; 38:471–476 Horst IC, et al: Accuracy and feasibility of 2006; 8:598–601

3274 Crit Care Med 2012 Vol. 40, No. 12


148. Hellman R: Patient safety and inpatient
cardiovascular surgery. Diabetes Technol 174. Sinha S, Jayaram R, Hargreaves CG: Fatal
glycemic control: Translating concepts Ther 2006; 8:609–616 neuroglycopaenia after accidental use of a
into action. Endocr Pract 2006; 12(Suppl 162. Korytkowski MT, Salata RJ, Koerbel GL,
glucose 5% solution in a peripheral arterial
3):49–55 et al: Insulin therapy and glycemic control cannula flush system. Anaesthesia 2007;
149. Newton CA, Smiley D, Bode BW, et al: A in hospitalized patients with diabetes dur- 62:615–620
comparison study of continuous insulin ing enteral nutrition therapy: A randomized 175. Eslami SAbu-Hanna A, de Jonge E, de

infusion protocols in the medical intensive controlled clinical trial. Diabetes Care 2009; Keizer NF: Tight glycemic control and
care unit: Computer-guided vs. standard 32:594–596 computerized decision-support systems: A
column-based algorithms. J Hosp Med 163. Lepore M, Pampanelli S, Fanelli C, et al: systematic review. Intensive Care Med 2009;
2010; 5:432–437 Pharmacokinetics and pharmacodynamics 35:1505–1517
150. Prescribing information, Apidra: Available of subcutaneous injection of long-acting 176. Blaha J, Kopecky P, Matias M, et al: Com-
at: http://products.sanofi.us/apidra/apidra. human insulin analog glargine, NPH insu- parison of three protocols for tight glycemic
pdf. Accessed March 18, 2012 lin, and ultralente human insulin and con- control in cardiac surgery patients. Diabetes
151. Nerenberg KA, Goyal AA, Xavier D, et al: tinuous subcutaneous infusion of insulin Care 2009; 32:757–761
Piloting a novel algorithm for glucose con- lispro. Diabetes 2000; 49:2142–2148 177. Cavalcanti AB, Silva E, Pereira AJ, et al: A
trol in the coronary care unit. Diab Care 164. Weant KA, Ladha A: Conversion from con- randomized controlled trial comparing a
2012; 35:19–24 tinuous insulin infusions to subcutaneous computer-assisted insulin infusion protocol
152. Taylor BE, Schallom ME, Sona CS, et al: insulin in critically ill patients. Ann Phar- with a strict and a conventional protocol
Efficacy and safety of an insulin infusion macother 2009; 43:629–634 for glucose control in critically ill patients.
protocol in a surgical ICU. J Am Coll Surg 165. Martindale RG, McClave SA, Vanek VW, et al; J Crit Care 2009; 24:371–378
2006; 202:1–9 American College of Critical Care Medicine; 178. Plank J, Blaha J, Cordingley J, et al: Mul-
153. Eigsti J, Henke K: Innovative solutions:
A.S.P.E.N. Board of Directors: Guidelines ticentric, randomized, controlled trial to
Development and implementation of a tight for the provision and assessment of nutri- evaluate blood glucose control by the model
blood glucose management protocol: One tion support therapy in the adult critically predictive control algorithm versus routine
community hospital’s experience. Dimens ill patient: Society of Critical Care Medicine glucose management protocols in inten-
Crit Care Nurs 2006; 25:62–65 and American Society for Parenteral and sive care unit patients. Diabetes Care 2006;
154. Furnary AP, Braithwaite SS: Effects of out- Enteral Nutrition: Executive Summary. Crit 29:271–276
come on in-hospital transition from intra- Care Med 2009; 37:1757–1761 179. Brown G, Dodek P: Intravenous insulin

venous insulin infusion to subcutaneous 166. Casaer MP, Mesotten D, Hermans G,
nomogram improves blood glucose con-
therapy. Am J Cardiol 2006; 98:557–564 et al: Early versus late parenteral nutrition trol in the critically ill. Crit Care Med 2001;
155. Grainger A, Eiden K, Kemper J, et al: A in critically ill adults. N Engl J Med 2011; 29:1714–1719
pilot study to evaluate the effectiveness of 365:506–517 180. Kanji S, Singh A, Tierney M, et al: Stan-
glargine and multiple injections of lispro in 167. Clayton SB, Mazur JE, Condren S, et al: dardization of intravenous insulin therapy
patients with type 2 diabetes receiving tube Evaluation of an intensive insulin protocol improves the efficiency and safety of blood
feedings in a cardiovascular intensive care for septic patients in a medical intensive care glucose control in critically ill adults. Inten-
unit. Nutr Clin Pract 2007; 22:545–552 unit. Crit Care Med 2006; 34:2974–2978 sive Care Med 2004; 30:804–810
156. O’Malley CW, Emanuele M, Halasyamani L, 168. Dickerson RN, Swiggart CE, Morgan LM, 181. Goldberg PA, Siegel MD, Sherwin RS, et al:
et al: Bridge over troubled waters: Safe and et al: Safety and efficacy of a graduated Implementation of a safe and effective insu-
effective transitions of the inpatient with intravenous insulin infusion protocol in lin infusion protocol in a medical intensive
hyperglycemia. J Hosp Med 2008; 3(Suppl critically ill trauma patients receiving spe- care unit. Diabetes Care 2004; 27:461–467
5):S55–S65 cialized nutritional support. Nutrition 2008; 182. Quinn JA, Snyder SL, Berghoff JL, et al:
157. Ramos P, Childers D, Maynard G, et al:
24:536–545 A practical approach to hyperglycemia
Maintaining glycemic control when transi- 169. Lonergan T, Compte AL, Willacy M, et al: A management in the intensive care unit:
tioning from infusion insulin: A protocol- pilot study of the SPRINT protocol for tight Evaluation of an intensive insulin infu-
driven, multidisciplinary approach. J Hosp glycemic control in critically ill patients. sion protocol. Pharmacotherapy 2006;
Med 2010; 5:446–451 Diabetes Technol Ther 2006; 8:449–462 26:1410–1420
158. Schmeltz LR, DeSantis AJ, Schmidt K,
170. Chase JG, Shaw G, Le Compte A, et al:
183. Wilson M, Weinreb J, Hoo GW: Intensive

et al: Conversion of intravenous insulin Implementation and evaluation of the insulin therapy in critical care: A review of 12
infusions to subcutaneously administered SPRINT protocol for tight glycaemic con- protocols. Diabetes Care 2007; 30:1005–1011
insulin glargine in patients with hypergly- trol in critically ill patients: A clinical prac- 184. Juneja R, Roudebush C, Kumar N, et al:
cemia. Endocr Pract 2006; 12:641–650 tice change. Crit Care 2008; 12:R49 Utilization of a computerized intravenous
159. Donaldson S, Villanuueva G, Rondinelli L, 171. Hellman R: A systems approach to reduc- insulin infusion program to control blood
et al: Rush University guidelines and proto- ing errors in insulin therapy in the inpa- glucose in the intensive care unit. Diabetes
cols for the management of hyperglycemia tient setting. Endocr Pract 2004; 10(Suppl Technol Ther 2007; 9:232–240
in hospitalized patients: Elimination of the 2):100–108 185. Osburne RC, Cook CB, Stockton L, et al:
sliding scale and improvement of glycemic 172. Bates D, Clark NG, Cook RI, et al; Writing Improving hyperglycemia management in
control throughout the hospital. Diabetes Committee on Patient Safety and Medical the intensive care unit: Preliminary report
Educ 2006; 32:954–962 System Errors in Diabetes and Endocrinol- of a nurse-driven quality improvement proj-
160. Datta S, Qaadir A, Villanueva G, et al: Once- ogy: American College of Endocrinology and ect using a redesigned insulin infusion algo-
daily insulin glargine versus 6-hour sliding American Association of Clinical Endocri- rithm. Diabetes Educ 2006; 32:394–403
scale regular insulin for control of hypergly- nologists position statement on patient 186. Anabtawi A, Hurst M, Titi M, et al: Incidence
cemia after a bariatric surgical procedure: safety and medical system errors in diabe- of hypoglycemia with tight glycemic control
A randomized clinical trial. Endocr Pract tes and endocrinology. Endocr Pract 2005; protocols: A comparative study. Diabetes
2007; 13:225–231 11:197–202 Technol Ther 2010; 12:635–639
161. Yeldandi RR, Lurie A, Baldwin D: Com-
173. Schleis TG: Interference of maltose, icodex- 187. Pielmeier U, Andreassen S, Juliussen B,

parison of once-daily glargine insulin with trin, galactose, or xylose with some blood et al: The Glucosafe system for tight glyce-
twice-daily NPH/regular insulin for con- glucose monitoring systems. Pharmaco- mic control in critical care: A pilot evalua-
trol of hyperglycemia in inpatients after therapy 2007; 27:1313–1321 tion study. J Crit Care 2010; 25:97–104

Crit Care Med 2012 Vol. 40, No. 123275


188. Cordingley JJ, Vlasselaers D, Dormand NC, 199. Otto NM, Schindler R, Lun A, et al: Hyperos- insulin therapy to improve hospital out-
et al: Intensive insulin therapy: Enhanced motic stress enhances cytokine production comes (TRIUMPH) project. Diabetes Care
Model Predictive Control algorithm versus and decreases phagocytosis in vitro. Crit 2008; 31:1556–1561
standard care. Intensive Care Med 2009; Care 2008; 12:R107 212. Hirshberg E, Lacroix J, Sward K, et al: Blood
35:123–128 200. Goldberg PA, Bozzo JE, Thomas PG, et al: glucose control in critically ill adults and
189. Campion TR Jr, May AK, Waitman LR,
“Glucometrics”—Assessing the quality of children: A survey on stated practice. Chest
et al: Effects of blood glucose transcription inpatient glucose management. Diabetes 2008; 133:1328–1335
mismatches on a computer-based intensive Technol Ther 2006; 8:560–569 213. Preissig CM, Rigby MR: A disparity between
insulin therapy protocol. Intensive Care 201.
Braithwaite SS: Patient-level glucose physician attitudes and practice regard-
Med 2010; 36:1566–1570 reporting: Averages, episodes, or something ing hyperglycemia in pediatric intensive
190. Cyrus RM, Szumita PM, Greenwood BC,
in between? Crit Care 2008; 12:133 care units in the United States: A survey
et al: Evaluation of compliance with a paper- 202. Schnipper JL, Magee M,Larsen K, et al:
on actual practice habits. Crit Care 2010;
based, multiplication-factor, intravenous Society of Hospital Medicine Glycemic 14:R11
insulin protocol. Ann Pharmacother 2009; Control Task Force Summary: Practical rec- 214. Vlasselaers D, Milants I, Desmet L, et al:
43:1413–1418 ommendations for assessing the impact of Intensive insulin therapy for patients in
191. Anger KE, Szumita PM: Barriers to glucose glycemic control efforts. J Hosp Med 2008; paediatric intensive care: A prospective,
control in the intensive care unit. Pharma- 3 (Suppl 5):s66–s83 randomised controlled study. Lancet 2009;
cotherapy 2006; 26:214–228 203. Furnary AP, Wu Y, Bookin SO: Effect of
373:547–556
192. Dossett LA, Cao H, Mowery NT, et al: Blood hyperglycemia and continuous intravenous 215. Preissig CM, Hansen I, Roerig PL, et al:

glucose variability is associated with mor- insulin infusions on outcomes of cardiac sur- A protocolized approach to identify and
tality in the surgical intensive care unit. Am gical procedures: The Portland Diabetic Proj- manage hyperglycemia in a pediatric criti-
Surg 2008; 74:679–85; discussion 685 ect. Endocr Pract 2004; 10(Suppl 2):21–33 cal care unit. Pediatr Crit Care Med 2008;
193. Ali NA, O’Brien JM Jr, Dungan K,
204. Vogelzang M, van der Horst IC, Nijsten MW: 9:581–588
et al: Glucose variability and mortality in Hyperglycaemic index as a tool to assess 216. Verhoeven JJ, Brand JB, van de Polder

patients with sepsis. Crit Care Med 2008; glucose control: A retrospective study. Crit MM, et al: Management of hyperglyce-
36:2316–2321 Care 2004; 8:R122–R127 mia in the pediatric intensive care unit;
194. Hirshberg E, Larsen G, Van Duker H: Alter- 205. Estrada CA, Young JA, Nifong LW, et al: Out- implementation of a glucose control pro-
ations in glucose homeostasis in the pedi- comes and perioperative hyperglycemia in tocol. Pediatr Crit Care Med 2009; 10:
atric intensive care unit: Hyperglycemia patients with or without diabetes mellitus 648–652
and glucose variability are associated with undergoing coronary artery bypass grafting. 217. Preissig CM, Rigby MR, Maher KO: Glycemic
increased mortality and morbidity. Pediatr Ann Thorac Surg 2003; 75:1392–1399 control for postoperative pediatric cardiac
Crit Care Med 2008; 9:361–366 206. Newton CA, Young S: Financial implications patients. Pediatr Cardiol 2009; 30:1098–1104
195. Eslami S, de Keizer NF, de Jonge E, et al: of glycemic control: Results of an inpatient 218. Faraon-Pogaceanu C, Banasiak KJ, Hirsh-
A systematic review on quality indicators diabetes management program. Endocr berg EL, et al: Management of hypergly-
for tight glycaemic control in critically ill Pract 2006; 12(Suppl 3):43–48 cemia in the pediatric intensive care unit;
patients: Need for an unambiguous indicator 207. Aragon D: Evaluation of nursing work effort implementation of a glucose control proto-
reference subset. Crit Care 2008; 12:R139 and perceptions about blood glucose testing col. Pediatr Crit Care Med 2010; 11:741–749
196. Ali NA, Krinsley JS, Preiser JC: Glucose vari- in tight glycemic control. Am J Crit Care 219. Branco RG, Chavan A, Tasker RC: Pilot eval-
ability in critically ill patients. In: Yearbook 2006; 15:370–377 uation of continuous subcutaneous glucose
of Intensive Care and Emergency Medi- 208. Malesker MA, Foral PA, McPhillips AC,
monitoring in children with multiple organ
cine. Vincent JL (Ed.). Berlin, Heidelberg, et al: An efficiency evaluation of protocols dysfunction syndrome. Pediatr Crit Care
Springer-Verlag, 2009, pp 728–737 for tight glycemic control in intensive care Med 2010; 11:415–419
197. Risso A, Mercuri F, Quagliaro L, et al: Inter- units. Am J Crit Care 2007; 16:589–598 220. Steil GM, Deiss D, Shih J, et al: ICU insulin
mittent high glucose enhances apoptosis 209. Van den Berghe G, Wouters PJ, Kesteloot delivery algorithms: Why so many? How to
in human umbilical vein endothelial cells K, et al: Analysis of healthcare resource choose? Diab Sci Tech 2009; 3:125–140
in culture. Am J Physiol Endocrinol Metab utilization with intensive insulin therapy in 221. Steil GM, Langer M, Jaeger K, et al: Value
2001; 281:E924–E930 critically ill patients. Crit Care Med 2006; of continuous glucose monitoring for mini-
198. Quagliaro L, Piconi L, Assaloni R, et
34:612–616 mizing severe hypoglycemia during tight
al: Intermittent high glucose enhances 210. Krinsley JS, Jones RL: Cost analysis of
glycemic control. Pediatr Crit Care Med
apoptosis related to oxidative stress in intensive glycemic control in critically ill 2011; 12:643–648
human umbilical vein endothelial cells: adult patients. Chest 2006; 129:644–650 222. Agus MSD, Steil GM, Wypij D, et al: Tight
The role of protein kinase C and NAD(P) 211. Sadhu AR, Ang AC, Ingram-Drake LA, et al: glycemic control versus standard care after
H-oxidase activation. Diabetes 2003; 52: Economic benefits of intensive insulin ther- pediatric cardiac surgery. N Engl J Med
2795–2804 apy in critically ill patients: The targeted 2012; 367:1208−1219

3276 Crit Care Med 2012 Vol. 40, No. 12

Potrebbero piacerti anche