DA-650/DA-645/DA-640 Quick Manual

Calibration 5 Sampling of pure water


 Adjust sample/drain lever to

1 Select “Calibration” 
“Sample.”
Adjust controller to slowest.
 Press “Calibration”.
 Press “PUMP” to start
 Press “Next”.
sampling of pure water.
 Adjust controller to optimal
speed.
 Check through cell window
to make sure there is no air
bubble in the cell.
 Press “PUMP” to stop pump
Check air bubbles. operation.

2 Clean the cell 6 Calibration of pure water


 Adjust sample/drain lever to  Press “START” to start
“Drain.” calibration of pure water.
 Wipe off tip of sampling
tube.
 Press “PUMP” and clean the
cell with rinse solution 2 for 7 Print out result
about 10 sec.  Calibration result and factor
 Press “PUMP” to stop pump are shown on LCD and
operation. printed out with beep.

3 Sampling Tube
Dry the cell 8 Clean the cell
 Wipe off sampling tube and  Adjust sample/drain lever to
put it in drying tube. “Drain.”
 Press “PURGE” twice. (Auto  Wipe off tip of sampling
Off; equipment automatically tube.
stops working when finished.)  Press “PUMP” and clean the
cell with rinse solution 2 for
about 10 sec.
Drying Tube
Put in.  Press “PUMP” to stop pump
operation.

4 Air calibration 9 Sampling Tube


Dry the cell
 Press “Start Calib” to start  Wipe off sampling tube and
air calibration. put it in drying tube.
 Press “PURGE” twice.
(Auto Off)

Put in.
Drying Tube

Ver. 01
See 【DA-650/-645/-640 Operation Manual】 for details.
A/N 59-00053-01
DA-650/DA-645/DA-640 Quick Manual
Measurement 5 Measurement ends
 Once oscillation frequency

1 Clean the cell


becomes stable, measurement
comes to an end and result
 Adjust sample/drain lever to
(density of sample) will be
“Drain.”
shown.
 Wipe off tip of sampling


tube.
Press “PUMP” and clean the
6 Drain sample
 Adjust sample/drain lever to
cell with rinse solution 2 for
“Drain.”
about 10 sec.
 Remove sampling tube from
 Press “PUMP” to stop pump
sample.
operation.
 Press “PUMP” to drain
sample.

2 Dry the cell 7 Clean the cell(Rinse1)


Sampling Tube  Adjust sample/drain lever to
 Wipe off sampling tube and
put it in drying tube. “Drain.”

 Press “PURGE” twice.  Wipe off tip of sampling

(Auto Off) tube.


 Press “PUMP” and clean the
cell with rinse solution 1 for
about 10 sec.
 Press “PUMP” to stop pump
Put in.
Drying Tube operation.

3 Sampling 8 Clean the cell(Rinse2)


 Adjust sample/drain lever to  Adjust sample/drain lever to
“Sample.” “Drain.”
 Adjust controller to slowest.  Wipe off tip of sampling
 Press “PUMP” to start tube.
sampling.  Press “PUMP” and clean the
 Adjust controller to optimal cell with rinse solution 2 for
speed. about 10 sec.
 Check through cell window  Press “PUMP” to stop pump
to make sure there is no air operation.


bubble in the cell.
Press “PUMP” to stop pump 9 Sampling Tube Dry the cell
Check air bubbles. operation.  Wipe off sampling tube and
put it in drying tube.
4 Start measurement  Press “PURGE” twice.
(Auto Off)
 Press “START”.

Put in.
Drying Tube

Ver. 01
See 【DA-650/-645/-640 Operation Manual】 for details.