Sei sulla pagina 1di 9

Analytica Chimica Acta 915 (2016) 102e110

Contents lists available at ScienceDirect

Analytica Chimica Acta


journal homepage: www.elsevier.com/locate/aca

Fast and sensitive supercritical fluid chromatography e tandem mass


spectrometry multi-class screening method for the determination of
doping agents in urine
Lucie Novakova
 a, Vincent Desfontaine b, Federico Ponzetto c, Raul Nicoli c,
Martial Saugy , Jean-Luc Veuthey b, Davy Guillarme b, *
c

a
Department of Analytical Chemistry, Faculty of Pharmacy, Charles University in Prague, Heyrovskeho 1203, 500 05 Hradec Kra lov
e, Czech Republic
b
School of Pharmaceutical Sciences, University of Geneva, University of Lausanne, Boulevard d'Yvoy 20, 1211 Geneva 4, Switzerland
c
Swiss Laboratory for Doping Analyses, University Center of Legal Medicine Lausanne-Geneva, Centre Hospitalier Universitaire Vaudois and University of
Lausanne, Chemin des Croisettes 22, 1066 Epalinges, Switzerland

h i g h l i g h t s g r a p h i c a l a b s t r a c t

 A high throughput UHPSFC-MS/MS


method was developed for
screening several classes of doping
agents in urine.
 Supported liquid extraction in 48-
well plate format was successfully
applied to extract the doping agents
from urine samples.
 Good extraction recoveries and
reasonable matrix effects were
observed for the whole set of doping
agents.
 The method exhibited very low LODs,
below the Minimum Required Per-
formance Levels, for most of the 100
compounds.

a r t i c l e i n f o a b s t r a c t

Article history: This study shows the possibility offered by modern ultra-high performance supercritical fluid chroma-
Received 11 November 2015 tography combined with tandem mass spectrometry in doping control analysis. A high throughput
Received in revised form screening method was developed for 100 substances belonging to the challenging classes of anabolic
3 February 2016
agents, hormones and metabolic modulators, synthetic cannabinoids and glucocorticoids, which should
Accepted 7 February 2016
be detected at low concentrations in urine. To selectively extract these doping agents from urine, a
Available online 12 February 2016
supported liquid extraction procedure was implemented in a 48-well plate format. At the tested con-
centration levels ranging from 0.5 to 5 ng/mL, the recoveries were better than 70% for 48e68% of the
Keywords:
Doping agents
compounds and higher than 50% for 83e87% of the tested substances. Due to the numerous interferences
Urine related to isomers of steroids and ions produced by the loss of water in the electrospray source, the
Supercritical fluid chromatography choice of SFC separation conditions was very challenging. After careful optimization, a Diol stationary
Tandem mass spectrometry phase was employed. The total analysis time for the screening assay was only 8 min, and interferences as
Supported liquid extraction well as susceptibility to matrix effect (ME) were minimized. With the developed method, about 70% of
Steroids the compounds had relative ME within the range ±20%, at a concentration of 1 and 5 ng/mL. Finally,
limits of detection achieved with the above-described strategy including 5-fold preconcentration were

* Corresponding author.
E-mail address: davy.guillarme@unige.ch (D. Guillarme).

http://dx.doi.org/10.1016/j.aca.2016.02.010
0003-2670/© 2016 Elsevier B.V. All rights reserved.
kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110 103

below 0.1 ng/mL for the majority of the tested compounds. Therefore, LODs were systematically better
than the minimum required performance levels established by the World anti-doping agency, except for
very few metabolites.
© 2016 Elsevier B.V. All rights reserved.

1. Introduction remains the gold standard for analyzing these substances in anti-
doping laboratories, but often requires laborious liquideliquid
The list of prohibited substances published by the World anti- extraction procedures as well as time-consuming derivatization
doping agency (WADA) is yearly updated. Today, there are about steps prior to injection, to make the substances volatile and
250 illicit compounds covering a wide range of physicochemical improve their detectability. Surprisingly, SFC is known to be a
properties that need to be tracked in the different specimens (urine reference strategy for the analytical identification and character-
and blood) collected in-and-out of competition [1]. The declaration ization of various types of steroids [10e12], but has rarely been
of a presence or an absence of a doping agent is a multistep process applied so far for the determination of steroids and derivatives in
including an initial screening approach followed by a confirmation doping control analysis [11], and never in routine analyses.
procedure, if applicable. To tackle the huge chemical diversity of the The goal of this work was to evaluate the performance of a
banned substances, various analytical methods are implemented in combination of supported liquid extraction (SLE) and UHPSFC-MS/
anti-doping laboratories and among them, GCeMS(/MS) and LC- MS analysis, for the high throughput screening of 100 substances
MS/MS are considered as the methods of choice offering high (parent compounds and phase I metabolites) in urine. All the
selectivity, sensitivity and fast turnaround. These methods focus on selected substances belong to the problematic classes of anabolic
the direct detection of prohibited substances, but also on the androgenic steroids, hormones and metabolic modulators, syn-
determination of their major phase I and phase II metabolites, to thetic cannabinoids and glucocorticoids, for which there are a lot of
further improve the detection windows capabilities in the matrix of isobaric compounds that should be detected at very low concen-
interest [2]. To ensure homogeneous performance between labo- tration levels in urine. The SLE recoveries, matrix effects and limits
ratories, Minimum Required Performance Levels (MRPLs) for of detection achieved in UHPSFC-MS/MS for the 100 substances will
detection of non-threshold substances are established by WADA, be shown and critically discussed.
and correspond to the minimum concentration levels that should
be attained for routine analyses. 2. Experimental method
To date, supercritical fluid chromatography (SFC) has been
scarcely used for doping control purposes, despite some obvious 2.1. Reagents and analytes
advantages of this separation technique [3]: i) the low fluid vis-
cosity and high diffusion coefficients under SFC conditions provide All doping agents were kindly provided by the Swiss Laboratory
excellent kinetic performance, ii) the organic solvent consumption for Doping Analysis (Epalinges, Switzerland). The exhaustive list of
is limited compared to LC, despite the fact that a significant pro- these target analytes is reported in Table S-1 and their structure in
portion of organic modifier (up to 40%) may be used in modern Fig. S-1. Methanol (MeOH), ethanol (EtOH), isopropanol and
packed column SFC, iii) a wide range of compounds, from relatively acetonitrile (ACN) of ULC/MS grade were provided by Biosolve
polar to highly non-polar, can be analyzed in SFC, with the same (Dieuze, France). Ammonium formate (AmF), ammonium acetate,
mobile phase components (CO2 and methanol). Still, these benefits heptane extra dry 99%þ and methyl tert-butyl ether (MtBE) were
were restricted by the poor quality of old-generation SFC in- provided by SigmaeFluka (Buchs, Switzerland). Diethylether was
struments providing low sensitivity, unacceptable quantitative purchased from Acros Organics (Geel, Belgium). Pressurized gas
performance and above all lack of reliability [4]. However, in the CO2 N48 (>99.998%) was purchased from Air Liquide (Dusseldorf,
last few years, providers of chromatographic instrumentation have Germany). Ultra-pure water was provided by a Milli-Q system from
launched new generation of SFC systems which undertake the Millipore (Bedford, MA, USA). b-glucuronidase from Escherichia coli
above-mentioned shortcomings. These new SFC instruments are was purchased from Roche Diagnostics GmbH (Mannhein,
described as UHPSFC which stands for ultra-high performance su- Germany).
percritical fluid chromatography [5]. UHPSFC systems offer an
improved compatibility with modern stationary phases, such as
2.2. Sample preparation of biological samples
columns packed with fully porous sub-2 mm particles as well as a
full compatibility with ESI-MS(/MS) devices thanks to commercial
A pool of blank urines was prepared by mixing 6 different urine
interfaces [6]. Considering all the beneficial features of modern
samples obtained from healthy volunteers. Each urine aliquot of
UHPSFC-MS/MS instrumentation, a screening method was previ-
1 mL was first spiked with doping agent standards in ACN to obtain
ously successfully developed in our laboratory for the determina-
5 different levels of concentration, namely 0.1, 0.5, 1, 5 and 10 ng/
tion of 110 doping agents in urine, including diuretics, b-blockers,
mL. Then, they were extracted on Isolute SLEþ 48-well plates
stimulants and narcotics, using a simple dilute-and-shoot proce-
(Biotage, Uppsala, Sweden). Urine was forced through the plate
dure [7,8].
sorbent using Biotage PRESSUREþ 96 positive pressure manifold at
Among the WADA list of banned substances, the determination
3 psi. After a waiting time of 5 min, the elution was made by
of the numerous anabolic agents, in particular androgenic steroids,
percolating 3 mL of MtBE through the wells into a 48-well collec-
is particularly challenging. Indeed, these substances are excreted in
tion plate. The SLE wells were dried by operating positive pressure
urine with very diverse concentrations, and MRPLs are extremely
again for few seconds. Then, the extracted samples in the collection
demanding. In addition, there is a large number of isomers and
plate were evaporated until dryness using UNIVAPO Rotational
metabolites, among this particular class, which are difficult to
Vacuum Concentrator 150 ECH (Biolabo Scientific Instruments,
separate and identify in a satisfactory manner [9]. GCeMS(/MS)
Ch^atel-Saint-Denis, Switzerland) at 1250 rpm, ambient
104 kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110

temperature and 30 Torr for 45 min. Then, they were reconstituted 3. Results and discussion
in 200 mL of a mixture of water and ACN (1:1). After 5 min of
agitation, the contents of each well were finally transferred into 3.1. Optimization of the chromatographic conditions
full-recovery vials (Waters, Milford, MA, USA) for injection into
UHPSFC-MS/MS. Because SFC is known for being a complementary method to LC
The same procedure was also used to obtain post-spiked blank and GC in terms of selectivity, this separation mode was evaluated
urine extracts. After SLE extraction of blank urine and evaporation for the screening of doping agents. It has to be noted, though, that,
until dryness, these samples were reconstituted in the mixture of taking into account the high proportion of MeOH in the mobile
water and ACN spiked with the compounds of interest at different phase at the end of the gradient, it is very likely that the fluid is no
concentrations. The results obtained with these samples were used longer under a supercritical state but rather in a subcritical state.
to calculate recoveries (RE) and to measure matrix effects (ME) for This is not a drawback as there is continuity in the properties of
each target compound: recoveries were calculated by dividing the supercritical and subcritical fluids. One of the crucial step of
signal of pre-extraction spiked urine by the signal of post- method development in SFC is the optimization of separation
extraction spiked urine, whereas the matrix effect was calculated conditions, to i) facilitate the identification of individual com-
making the ratio of signals of post-extraction spiked urine and pounds by separating isomers and interferences, ii) to ensure
standards in solvent. Relative matrix effects (MErel) were calculated elution of all compounds of interest during the gradient and iii) to
as MErel ¼ ME-100. obtain narrow and symmetrical peaks.
The first step of the study was to define SRM transitions of in-
dividual analytes. This must have been performed using individual
2.3. UHPSFC-MS/MS instrumentation and method compounds to define the transition correctly. As no positive criteria
have been set for screening methods, only one SRM transition was
All experiments were performed on a Waters Acquity UPC2 selected for each compound. All the analytes were ionized using ESI
system equipped with a binary solvent delivery pump, an auto- positive mode providing [MþH]þ ions for many tested analytes.
sampler which included a 10 mL loop for partial loop injection, a However, for steroid analysis, it is very common that protonated
column oven and a two-step (passive þ active) backpressure molecules provide lower signal intensity than fragments obtained
regulator (BPR). The chromatographic system was coupled to a by loss of one or two molecules of water. Therefore, [MþHeH2O]þ
Waters Xevo TQ-S triple quadrupole detector thanks to a dedicated or [MþHe2H2O]þ were monitored in many cases. Optimized SRM
double-T splitter interface from Waters. Additional make-up sol- transitions, cone voltages and collision energies were reported in
vent was brought by a Waters Isocratic Solvent Manager (ISM) Table S-1 (supporting information). Dwell times were set-up
pump. Pure MeOH was selected as the make-up solvent and automatically using MassLynx software calculation and defined
delivered at 0.3 mL/min. The hyphenation interface and splitter for peak width was 2.5 s.
UHPSFC-MS/MS are detailed elsewhere [6]. High separation efficiency in SFC was obtained by using state-
Eight different columns were tested for the method develop- of-the-art columns packed with sub-2 mm particles. To ensure the
ment, namely Waters Acquity UPC2 Torus Diol, 2-Picolylamine (2- highest possible selectivity, eight different sub-2 mm SFC stationary
PIC), Diethylamine (DEA) and 1-Aminoanthracene (1-AA), Acquity phases including hybrid silica (BEH), hybrid silica modified with 2-
UPC2 BEH and BEH 2-Ethylpyridine (2-EP), Acquity UPC2 CSH EP group, charged surface hybrid (CSH) phase modified with pen-
Fluoro-Phenyl (PFP) and Acquity UPC2 HSS C18 SB. All columns have tafluorophenyl (PFP) group, non-endcapped silica based C18,
dimensions of 100  3.0 mm and particle sizes of 1.7 mm, except the hybrid silica modified with diol group, hybrid silica modified with
latter which had 1.8 mm particle sizes. When injecting urine sam- 2-picolylamine group (2-PIC), hybrid silica modified with 1-
ples, the selected column was protected with adequate pre-column aminoanthracene group (1-AA) and hybrid silica modified with
Acquity UPC2 Torus Diol VanGuard, 1.7 mm and 5  2.1 mm pur- diethylamine group (DEA) were tested. The comparison of these
chased from Waters. stationary phases was carried out under the generic mobile phase
The mobile phase was a mixture of CO2 and 10 mM AmF dis- conditions reported in Refs. [7,8]. A 4 min generic gradient elution
solved in mixture of 98% MeOH and 2% water, as the organic run from 2 to 40% of methanol with addition of ammonium formate
modifier. The gradient program lasted 6 min and consisted in a was used, with a 1 min isocratic step at the beginning and at the
1 min isocratic step at 2% organic modifier followed by a gradient end of the program. Such generic conditions allow covering a wide
from 2 to 40% in 4 min and another 1 min isocratic step at 40% range of analyte polarities, together with a reasonable analysis time
organic modifier. The column was then reequilibrated for 2 min of 8 min including 2 min re-equilibration step. Three criteria were
prior to the next injection. The flow rate was 1.3 mL/min, column selected for the choice of the best stationary phase: (1) number of
temperature was fixed at 40  C, BPR pressure at 150 bar, and in- eluted compounds, (2) ability to separate isobaric compounds and
jection volume was equal to 1 mL. A mixture of heptane and iso- (3) peak shapes. As reported in Fig. 1, Torus Diol and Torus 2-PIC
propanol (7:3) was used as the weak solvent wash and pure columns clearly outperformed the other stationary phases in
isopropanol as the strong solvent wash. The volumes of weak and terms of peak shape and selectivity. Among these two phases, the
strong wash solvents were 600 and 200 mL, respectively. Torus Diol was finally selected for further experiments since it had
The Xevo TQ-S detector was exclusively operating in positive ESI slightly better peak shape than Torus 2-PIC. On the other hand,
mode and the different parameters were finely tuned as follows to Torus 1-AA was found less suitable for the analysis of selected
attain the highest possible sensitivity: source temperature at doping agents, due to very strong retention leading to non-elution
150  C, capillary voltage at 3.2 kV and ion source offset at 10 V. of several compounds. Also the C18 stationary phase provided
Nitrogen was used as desolvation gas at 850 L/h and 550  C, cone lower selectivity (many isobaric compounds were co-eluted) and
gas at 145 L/h and nebulizer gas at 7 bar. Argon was used as a not enough symmetrical peaks. Very good distinction of isobaric
collision gas at 0.17 mL/min. The SRM (selected reaction moni- compounds and various isomeric species was observed on CSH PFP
toring) transitions, the cone voltages and collision energies for all stationary phase, however, at the cost of many distorted peaks (see
target analytes are reported in Table S-1. MassLynx 4.1 software was Fig. S-2). Remaining three stationary phases (BEH, Torus DEA and
used for instrument control and data acquisition. TargetLynx soft- BEH 2-EP) also provided interesting selectivity for the separation of
ware was used for data processing and peak integration. selected doping agents (Fig. S-2), but the best peak shapes in overall
kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110 105

Fig. 1. Selectivity charts for four of the tested columns. Each compound is plotted according to its retention time (X axis) and mass (Y axis) for the Diol, 2-PIC, 1-AA and HSS C18 SB
columns. The chromatograms of 2 compounds (30 OH-stanozolol in red and raloxifen in blue) are shown to highlight peak shape performance on each column. (For interpretation of
the references to colour in this figure legend, the reader is referred to the web version of this article.)

comparison were obtained on Torus Diol. confirmation step. In fact, for an unequivocal identification of the
Fig. 2 illustrates the complexity of steroid isomers separation, suspected substance and related metabolites, experimental condi-
which remained one of the critical parts of this screening approach. tions for the confirmation procedure could be further optimized
In the first example (Fig. 2A), a typical case of isobaric compounds is using SFC-MS, including the use of second column chemistry with
shown. Three compounds, i. e. mibolerone, methyl-1-testosterone complementary selectivity (e.g. 2-PIC chemistry), chromatographic
and metenolone have the same chemical formula C20H30O2 and gradients with improved peak capacity and additional SRM tran-
therefore the same molecular weight of 302.2. Due to their struc- sitions. Furthermore, especially in the case of steroids, the use of an
tural similarity, interferences in the SRM channel 303.2 > 187.2 orthogonal technique such as GC-EI-MS/MS could also be envis-
were observed, and confirmed by the injections of individual aged for confirmation purposes, providing more informative MS/
standard solutions. Thanks to the high resolution offered by MS spectra through an additional derivatization procedure of the
UHPSFC, an easy distinction of these isomers was possible. How- compounds of interest. For confirmation, at least two SRM transi-
ever, the presence of isobaric compounds was not the only issue in tions should be monitored, as required by WADA. Then, unequiv-
method development. Indeed, most of the steroid structures pro- ocal identification of compounds should be obtained by comparing
vide a loss of one or two molecules of water in their fragmentation chromatographic retention times (±1% or ±0.1 min) and relative ion
pathway and these may also contribute to SRM channel of the other intensities of the two monitored transitions in real samples with
compounds, such as in case of one metabolite of danabol, which those of certified reference material and/or administration studies
was monitored as [MþHeH2O]þ to have higher sensitivity. There- [24].
fore, in the SRM channel of the metabolite of exemestane at After having selected the best stationary phase, fine method
299.2 > 121.1, [MþHeH2O]þ ion of the metabolite of danabol was tuning was performed. Based on previous findings [7,8], the influ-
also observed (Fig. 2B). Due to many interferences between isobaric ence of additive on peak shape, separation selectivity and especially
compounds and also [MþHeH2O]þ ions, the 100 compounds to be MS response was monitored. According to the provider, the Torus
analyzed were divided into two mixtures, as reported in Table S-1, columns are reported to provide good peak shapes even without
and specific acquisition windows centered on the retention time of additives in the mobile phase. Various types of mobile phase
each compound were created in the SRM method. Otherwise, the components were tested, including i) pure methanol, ii) methanol
quantitative evaluation of the method, including calculation of re- with 2% of water, iii) methanol þ10 mM ammonium acetate, iv)
covery and matrix effects would not be possible. In addition, the methanol þ10 mM ammonium formate and finally v)
presence of interfering compounds in the generic screening step methanol þ10 mM ammonium formate þ2% water. Symmetrical
may cause problems in the case of suspicious doping samples, peak shapes were observed for most of the investigated analytes
making the correct identification of doping agent more challenging. that are neutral or weakly acidic, when using methanol or meth-
This problem may be addressed with higher success rate during the anol with 2% water. However, using these organic modifiers, MS
106 kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110

Fig. 2. UHPSFC-MS/MS chromatograms for SRM transitions of metenolone (blue) and metabolite of exemestane (red). Conditions: column Torus Diol (100 mm  3.0 mm, 1.7 mm),
mobile phase A: CO2, mobile phase B: methanol with 10 mM ammonium formate þ 2% water, 2% B for 1 min then gradient from 2 to 40% in 4 min and 40% B for 1 min, flow rate:
1.3 mL/min, temperature: 40  C, backpressure: 150 bar, make-up pump: methanol at 0.3 mL/min, injection volume: 1 mL. (For interpretation of the references to colour in this figure
legend, the reader is referred to the web version of this article.)

sensitivity was systematically decreased, often by more than 70% immiscible organic solvent is passed through the column to extract
compared to the reference, which was methanol þ10 mM ammo- the desired components, typically 2e3 times of the sample volume.
nium formate. The performance of the two buffers, ammonium The residue is subsequently evaporated to dryness, reconstituted
formate and acetate in methanol, was very similar for our selected and injected onto the chromatographic system. Due to the low
group of compounds. In agreement with our previous findings [7,8], polarity of most of the tested compounds, SLE is a very interesting
the MS signal was systematically increased when 2% water was strategy providing urine preconcentration and sample clean-up
added to the mobile phase containing methanol þ ammonium (elimination of polar interfering compounds present in urine
formate. The increase in signal was about 30e70% in most cases, samples which do not participate in non-polar organic solvents
reaching even more than 200% for a few substances. Therefore, partitioning). A 48 well-plate format was selected in this research
water was added to the mobile phase. Finally, some experiments to ensure high sample throughput, which is required in routine
were also performed to verify the influence of the make-up solvent doping control analysis. Even if this study was carried out using
on the MS response using methanol, ethanol and isopropanol in only spiked urines, selected compounds are mainly excreted as
presence or absence of 2 and 5% of water. While ethanol and iso- glucuronides in real urine samples. Therefore, preliminary extrac-
propanol decreased the MS signal compared to the reference tion experiments have been performed to evaluate a possible in-
response of pure methanol, the addition of water did not change fluence of the enzymatic hydrolysis step on SLE procedure. For
the response significantly or lead to some slight decrease in the these tests, the pool of blank urines (pH determined at 7) was
case of 2% of water addition. Therefore, pure methanol was used as spiked with selected substances and was then subjected to hy-
make-up solvent. drolysis with b-glucuronidase from E. coli for 1 h at 50  C, followed
by SLE procedure as described above. No significant difference in
3.2. Optimization of the sample preparation step peak area was observed on the chromatograms after hydrolysis
compared to non-hydrolysed spiked samples (data not shown);
Using optimized conditions, developed UHPSFC-MS/MS method therefore this step was omitted for practical reasons in the
was tested with standard solutions to determine its linearity and continuation of this work. Finally, performance of SLE was also
sensitivity. The method demonstrated very good sensitivity and evaluated with two of the most widely used solvents in the
linearity (data on standard solutions not shown), which was a good extraction of steroids, namely methyl tert-butyl ether (MtBE) and
starting point for further development of sample preparation pro- diethylether (DEE). MtBE was selected since the overall recoveries
cedure in urine. For this purpose, SLE was selected as a simplified were comparable or even better when using MtBE vs. DEE.
variant of LLE. SLE columns contain specially processed wide-pore
diatomaceous earth, a chemically inert matrix, which only acts as
a holder for the sample [13]. After loading the sample, an
kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110 107

3.3. Evaluation of extraction recoveries metabolite 1 of budesonide.

Urines spiked with mixture 1 or mixture 2 (see Table S-1) were 3.4. Evaluation of matrix effects
tested at five concentration levels with the final protocol to obtain
extraction recovery values. The summary of results when using Matrix effects (ME) are sometimes an important issue of LC-MS
MtBE as an extraction solvent is shown in Table S-2 (supporting analytical methods, as they can negatively influence method ac-
information) for all compounds. The suitability of the SLE proce- curacy, precision and sensitivity. Due to the co-eluting compounds
dure for the sample treatment is shown in Fig. 3. For most of the from the matrix (or other sources), the signal response may be
compounds, the extraction recovery was >70% (48%, 59% and 68% of altered leading to ion suppression or enhancement [14,15]. There-
compounds at selected concentration levels of 0.5, 1 and 5 ng/ml, fore, it is important to evaluate the amount of ME when developing
respectively), while only nine compounds provided recovery lower new bioanalytical methods [16]. While the phenomenon of ME has
than 50% at low concentration levels equal to 0.5 and 1 ng/ml. Such been widely explored in LC-MS bioanalytical methods [15,17e19],
recoveries are perfectly adapted for the routine screening analysis only few reports deal with the determination of matrix effects in
of selected doping agents, provided that MRPLs of all compounds SFC-MS [8,20,21]. SFC might be a powerful strategy to reduce the
are reached. The capability of the method to reach MRPLs is further amount of ME [8,20], due to its very different separation mecha-
influenced by the amount of matrix effects, besides method pre- nism mostly based on H-bond and dipoleedipole interactions.
cision, sensitivity and recovery. The lowest recoveries were sys- Experimentally, blank urine samples were extracted using the final
tematically observed for zilpaterol (~5%), due to its high polarity protocol and then spiked with both standard mixtures (see Table S-
and tamoxifen (~20%). However, this issue was compensated by a 1) at five concentration levels to obtain matrix effect values for each
good response of both compounds in UHPSFC-MS/MS and thus compound. The calculation of ME was based on the post-extraction
sufficient method sensitivity. The MS responses of the investigated addition approach described elsewhere [22]. Since relative matrix
compounds were indeed quite different depending on their effects (MErel) were calculated, positive values implicitly indicated
physico-chemical properties, as illustrated at 5 ng/ml. At this con- signal enhancement, while negative values corresponded to signal
centration level, there were still three compounds (metabolites of suppression. The amount of ME observed in our study was very low,
norethisterone, norbolethone and turinabol) which provided very when using SLE as sample preparation step and UHPSFC-MS/MS for
low signal in UHPSFC-MS/MS and thus cannot be detected, while the analysis. The results of MErel for all analytes at five concentra-
six other compounds cannot be quantified anymore due to the tion levels are shown in Table S-2 (supporting information), while
saturation of the detector. This is typically the case of synthetic Fig. 4 shows the comparison of the amount of ME at three inter-
cannabinoids, which provide very good response in ESI positive mediate concentration levels (0.5, 1 and 5 ng/ml) as a function of
mode. Method precision was tested by performing quadruplicate retention times. For most of the compounds, typically 60e70% of
extractions of urine at all five concentration levels. Calculated RSD tested doping agents, matrix effects were negligible (MErel < þ20%
values were in most cases lower than 10e15%, with the large ma- and > 20%) using SLE as a sample preparation procedure. As
jority well below 10%, which is considered to be excellent for the shown, slightly more positive MEs were observed for 23% of com-
extraction of small compounds from biological material, without pounds at 0.5 ng/ml, for 17% of compounds at 1.0 ng/ml and for 12%
the use of internal standards in the case of MS detection (see of compounds at 5.0 ng/ml, while negative MEs were observed for
Table S-3). Worse results (RSD > 20%) were only observed for a few 7e10% of compounds at all three concentration levels. For most of
compounds including zilpaterol, toremifene, triamcinolone and the analytes, ME remained reasonable, except for the metabolite 2

Fig. 3. Recovery [%] of doping agents using SLE procedure for the extraction of substances from urine at concentration levels of 0.5, 1.0 and 5.0 ng/ml. Individual values of recoveries
for each compound are shown in Table S-2 of the supporting information.
108 kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110

Fig. 4. Relative matrix effects [%] observed in UHPSFC-MS/MS method after SLE procedure as a function of retention times at three concentration levels: 0.5 ng/ml (blue diamonds),
1 ng/ml (red circles) and 5 ng/ml (grey triangles). All the experimental data were now shown, to keep the clarity of the figure, but Table S-2 provides an exhaustive list of relative
matrix effects at all concentrations levels. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

of prostanozol, demonstrating matrix effects of 206% at 0.5 ng/ml, requested MRPLs. However, this does not have to be necessarily
179% at 1 ng/ml and 113% at 5 ng/ml. Therefore, these values were done, as other metabolites (e.g. norbolethone met. 2) or the parent
not included in Fig. 4 and are only shown in Table S-2 (supporting compound itself (e.g turinabol) were monitored with sufficient
information). Bolasterone, 30 OH-stanozolol and beclomethasone sensitivity. Regarding norethisterone, additional metabolites
were also importantly influenced by ME (MErel > þ50% or < 50%). should be included in this screening method to improve detect-
Similarly to Fig. 3, at the highest concentration level of 5 ng/ml, six ability of this compound in routine analyses.
compounds could not be evaluated quantitatively, due to the de-
tector saturation. Also, in the case of ME, it is important to have
4. Conclusion
consistent values through the measurements of different samples
[16], therefore, RSD values were calculated for the individual con-
In the field of doping control analysis, GCeMS(/MS) and
centration levels, and obtained RSD values were always better than
LCeMS(/MS) are the gold standard techniques for screening and
10%, except for the metabolites of budesonide, triamcinolone and
confirmatory analysis of substances prohibited in sport practice.
toremifene (see Table S-3).
The objective of this work was to assess the potential of UHPSFC-
MS/MS, as an alternative to the above-mentioned analytical stra-
3.5. Method sensitivity and comparison with MRPLs tegies, for the determination of 100 compounds in urine, including
parent substances and their main phase I metabolites. These ana-
Obviously, the most important aspect of the developed lytes belong to the difficult classes of anabolic agents, hormone and
screening method is to reach MRPL required by WADA. According metabolic modulators, synthetic cannabinoids and glucocorticoids.
to dedicated technical document [23], substances should be As a sample preparation procedure, SLE was found to be particu-
unambiguously detected at 100% MRPL value, while the estimated larly well suited for these relatively lipophilic substances and pro-
LOD (limit of detection) should not be higher than 50% of MRPL. vided a high throughput extraction using 48-well plate format.
Only if these conditions are met, the method may be applicable for Depending on the tested concentration levels (0.5, 1.0 and 5.0 ng/
routine doping screening analysis. In Fig. 5, the MRPLs of 100 tested mL), recoveries were better than 70% for 48e68% of the target
substances were compared with the achievable sensitivities esti- compounds and higher than 50% for 83e87% of the tested sub-
mated for target analytes spiked in urine samples. This figure stances. The chromatographic conditions were finely optimized,
confirms that the developed UHPSFC-MS/MS method using high considering the presence of numerous interferences related to
throughput SLE as a sample preparation (including a 5-fold pre- steroid isomers and ions produced by the loss of water in the ESI
concentration factor) is applicable for doping screening analysis source. Various stationary phase chemistries were tested and, a
of the selected classes of substances. The 50% MRPLs were Torus diol 100  3 mm, 1.7 mm column was finally selected. Using a
comfortably reached for almost all of the 100 doping agents, since generic gradient from 2 to 40% MeOH in presence of 10 mM
the achieved LODs were in average between 5 and 50-fold lower ammonium formate þ2% water, the elution of all the substances
than the MRPLs. Because LODs were excellent, the detection time was possible in only 8 min, with narrow peaks and limited in-
window of these doping agents can be extended. As shown in Fig. 5, terferences issues (high chromatographic selectivity). The suscep-
there were only three exceptions, namely the metabolites of nor- tibility of the developed method towards ME was also evaluated at
ethisterone, norbolethone and turinabol. These compounds pro- different concentration levels, ranging from 0.1 to 10 ng/mL. The
vided low response in UHPSFC-MS/MS, thus higher relative ME was reasonable, as 60e70% of the tested substances
preconcentration during the SLE step would be needed to attain the have relative ME within the range ±20%. In addition, there were
kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110 109

Fig. 5. Concentration levels providing sufficient sensitivity and precision (blue lines) versus required MRPLs (red lines) for individual doping agents of mixture 1 (A) and mixture 2
(B). (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

generally slightly more signal enhancement rather than ion sup- Acknowledgements
pression. Finally, achieved LODs were compared to MRPLs fixed by
WADA for all these classes of compounds. Thanks to the 5-fold SLE The authors gratefully acknowledge Waters, and particularly
preconcentration step, and the high sensitivity provided by modern Marcus Winkler, Stephane Canarelli and Joel Fricker for the kind
UHPSFC-MS/MS device, LODs were systematically below the opportunity to use the Waters Acquity UPC2 and Xevo TQ-S in-
MRPLs, except for one metabolite of each of these substances: struments in their demo-lab in Eschborn, Germany. This project has
norethisterone, norbolethone and turinabol. been carried out with the support of the World Anti-Doping Agency
In order to conclude, the developed SLE-UHPSFC-MS/MS (14A21RN).
method provided excellent performance in terms of turnaround,
selectivity and sensitivity. It could potentially be employed in Appendix A. Supplementary data
doping control laboratories for screening purposes in the near
future, as a replacement strategy to the most widely used LCeMS/ Supplementary data related to this article can be found at http://
MS and GCeMS/MS platforms. Moreover, despite that the hydro- dx.doi.org/10.1016/j.aca.2016.02.010.
lysis step should be maintained, the rapid sample preparation
developed in this study using automated SLE could be of great value References
for routine analysis avoiding tedious manipulations during LLE,
evaporation, reconstitution and derivatization steps required for [1] M. Thevis, T. Kuurane, H. Geyer, W. Schanzer, Annual banned-substance re-
view: analytical approaches in human sports drug testing, Drug Test. Anal. 7
GCeMS/MS analysis.
(2015) 1e20.
[2] F. Badoud, D. Guillarme, J. Boccard, E. Grata, M. Saugy, S. Rudaz, J.-L. Veuthey,
Analytical aspects in doping control: challenges and perspectives, Forensic Sci.
Int. 213 (2011) 49e61.
[3] E. Lesellier, C. West, The many faces of packed column supercritical fluid
110 kova
L. Nova  et al. / Analytica Chimica Acta 915 (2016) 102e110

chromatography e a critical review, J. Chromatogr. A 1382 (2015) 2e46. [14] L. Tang, P. Kebarle, Dependence of ion intensity in electrospray mass spec-
[4] V. Desfontaine, D. Guillarme, E. Francotte, L. Nov , Supercritical fluid
akova trometry on the concentration of the analytes in the electrosprayed solution,
chromatography in pharmaceutical analysis, J. Pharm. Biomed. Anal. 113 Anal. Chem. 65 (1993) 3654e3668.
(2015) 56e71. [15] A. Cappiello, G. Famiglini, P. Palma, E. Pierini, V. Termopoli, H. Trufelli,
[5] A. Grand-Guillaume Perrenoud, J.-L. Veuthey, D. Guillarme, The use of col- Overcoming matrix effects in liquid chromatography-mass spectrometry,
umns packed with sub-2 mm particles in supercritical fluid chromatography, Anal. Chem. 80 (2008) 9343e9348.
Trends Anal. Chem. 63 (2014) 44e54. [16] Committee for medicinal products for human use, Guideline on Bioanalytical
[6] A. Grand-Guillaume Perrenoud, J.-L. Veuthey, D. Guillarme, Coupling state-of- Method Validation, EMA, London, UK, 2011.
the-art supercritical fluid chromatography and mass spectrometry: From [17] R. Dams, M.A. Huestis, W.E. Lambert, C.M. Murphy, Matrix effect in bio-
hyphenation interface optimization to high-sensitivity analysis of pharma- analysis of illicit drugs with LC-MS/MS: influence of ionization type, sample
ceutical compounds, J. Chromatogr. A 1339 (2014) 174e184. preparation, and biofluid, J. Am. Soc. Mass Spectrom. 14 (2003) 1290e1294.
[7] L. Novakova, A. Grand-Guillaume Perrenoud, R. Nicoli, M. Saugy, J.-L. Veuthey, [18] E. Chambers, D.M. Wagrowski-Diehl, Z. Lu, J.R. Mazzeo, Systematic and
D. Guillarme, Ultra high performance supercritical fluid chromatography comprehensive strategy for reducing matrix effects in LC/MS/MS analyses,
coupled with tandem mass spectrometry for screening of doping agents. I: J. Chromatogr. B 852 (2007) 22e34.
Investigation of mobile phase and MS conditions, Anal. Chim. Acta 853 (2015) [19] A. Van Eckhaut, K. Lanckmans, S. Sarre, I. Smolders, Y. Michotte, Validation of
637e646. bioanalytical LC-MS/MS assays: evaluation of matrix effects, J. Chromatogr. B
[8] L. Novakova, A. Grand-Guillaume Perrenoud, R. Nicoli, M. Saugy, J.-L. Veuthey, 877 (2009) 2198e2207.
D. Guillarme, Ultra high performance supercritical fluid chromatography [20] T. Berg, L. Kaur, A. Risnes, S.M. Havig, R. Karinen, Determination of a selection
coupled with tandem mass spectrometry for screening of doping agents. II: of synthetic cannabinoids and metabolites in urine by UHPSFC-MS/MS and by
analysis of biological samples, Anal. Chim. Acta 853 (2015) 647e659. UHPLC-MS/MS, Drug Test. Anal. (2015), http://dx.doi.org/10.1002/dta.1844 (in
[9] W. Schanzer, M. Thevis, Human sports drug testing by mass spectrometry, press).
Mass Spec. Rev. (2015), http://dx.doi.org/10.1002/mas.21479. [21] Y. Hsieh, F. Li, C.J. Duncan, Supercritical fluid chromatography and high-
[10] L. Nova kov
a, P. Chocholous, P. Solich, Ultra-fast separation of estrogen ste- performance liquid chromatography/tandem mass spectrometric methods
roids using subcritical fluid chromatography on sub-2-micron particles, for the determination of cytarabine in mouse plasma, Anal. Chem. 79 (2007)
Talanta 121 (2014) 178e186. 3856e3861.
[11] M. Doue, G. Dervilly-Pinel, K. Pouponneau, F. Monteau, B. Le Bizec, Analysis of [22] B.K. Matuszewski, M.L. Constanzer, C.M. Chavez-Eng, Strategies for the
glucuronide and sulfate steroids in urine by ultra-high-performance super- assessment of matrix effect in quantitative bioanalytical methods based on
critical-fluid chromatography hyphenated tandem mass spectrometry, Anal. HPLC-MS/MS, Anal. Chem. 75 (2003) 3019e3030.
Bioanal. Chem. 407 (2015) 4473e4484. [23] TD2015MRPL, World Anti-doping Agency, Montreal, 2015. https://wada-
[12] X. Xu, J.M. Roman, T.D. Veenstra, J. Van Anda, R.G. Ziegler, H.J. Issaq, Analysis main-prod.s3.amazonaws.com/resources/files/wada_td2015mrpl_minimum_
of fifteen estrogen metabolites using packed column supercritical fluid required_perf_levels_en.pdf (Last accessed on October 2015).
chromatography-mass spectrometry, Anal. Chem. 78 (2006) 1553e1558. [24] TD2015IDCRL, World Anti-doping Agency, Montreal, 2015. https://wada-
[13] L. Novakova, H. Vl , A review of current trends and advances in modern
ckova main-prod.s3.amazonaws.com/resources/files/wada_td2015idcr_minimum_
bio-analytical methods: chromatography and sample preparation, Anal. Chim. criteria_chromato-mass_spectro_conf_en.pdf (Last accessed on January 2016).
Acta 656 (2009) 8e35.

Potrebbero piacerti anche