CERTIFICATE OF CALIBRATION

Calibration Cert #1411.01

Client Address Calibration Location


(if different from client address)
THE SAMPLE COMPANY
1234 MAIN STREET
SAMPLETOWN, MA 12345

Date of Calibration: 6/27/2017 Sensor Type: TC= J ITS 90


Manufacturer: WATLOW Tolerance: ± 2°F
Model Number: F4 SERIES Tol. Reported: Client
Serial Number: 55555 Resolution: 1°F
Client I.D Number: Temp Control-01 Smallest Grad.: 1°F

Description: Temperature Controller (0 to 1000°F)


Method of Calibration: YCS-PT-1001 As Found: IN TOLERANCE
Location: Oven 2 As Left: IN TOLERANCE

Seq. Description Standard Tolerance - Tolerance + As Found As Left Unit


1 Applied Nominal 100.0 98.0 102.0 101.0 101.0 °F
2 Applied Nominal 300.0 298.0 302.0 301.0 301.0 °F
3 Applied Nominal 500.0 498.0 502.0 501.0 501.0 °F
4 Applied Nominal 700.0 698.0 702.0 700.0 700.0 °F
5 Applied Nominal 900.0 898.0 902.0 900.0 900.0 °F

Standard(s) Used To Calibrate The Above Instrument:


Company I.D. Description Cal. Due Date
MA METROLOGY QEN 189 Fluke 743B Calibrator 12/20/2017

Cal. Interval: 6 MONTHS Next Cal. Due: 12/27/2017


Ambient Temperature: 74 F Relative Humidity: 52
Uncertainty: ± 0.36 + 0.6R °F

The calibration results reported above relate only to the instrument(s) calibrated. Measurement uncertainties were budgeted and calculated using
the ISO Guide to the Expression of Uncertainty in Measurement (GUM). Reported uncertainities represent expanded uncertainties expressed at
approximately the 95% confidence level using a coverage factor of k=2. Unless otherwise stated, a Test Uncertainty Ratio (TUR) of 4:1 or greater
was used. If .6R is used as a floor value in the uncertainty listed, R represents the resolution of the instrument under test. Measurement
uncertainty was not taken into account when determining if the instrument under test was within the tolerance limits specified above.

Standards used for this calibration are traceable to NIST, or another N.M.I. This calibration is in accordance with ISO/IEC 17025, ISO Guide 25,
ANSI/NCSL Z540-1, ISO 9000 and MIL. Std. 45662A. This calibration certificate shall not be reproduced except in full without written permission
from Massachusetts Metrology and Instrument Service.

Metrologist: SAMPLE CERTIFICATE Signature:

Page 1 of 1
One Liberty Square - Rockland, MA 02370 Phone: 781-982-7125 Fax: 781-982-7152 Certificate Number: 235206