Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Ceramics
CERAMICS
1
Forest Biopolymers Science and Engineering, USDA Forest Service, Forest Products Laboratory, One Gifford Pinchot Drive,
Madison, WI 53726, USA
https://doi.org/10.1007/s10853-017-1922-8
J Mater Sci (2018) 53:4814–4827 4815
affect the accuracy of the calibrations and subsequent indentation size effect is different than the intrinsic
experimental results obtained using the calibrations. material size effects often studied using nanoinden-
During nanoindentation, the displacement mea- tation [13]. Fused silica surfaces also have roughness
sured by the instrument includes both the depth of and modified surface chemical properties that can
penetration into the material surface and the dis- result in inhomogeneous surface mechanical proper-
placement from the deflection of the instrument ties. Additionally, at high loads, cracking may occur
during loading. This instrument deflection must be during the experiment. Therefore, a user must iden-
calculated and subtracted from the measured dis- tify and exclude nanoindentations affected by probe
placement to accurately assess the depth of penetra- tip imperfections, fused silica surface effects, and
tion. Nanoindenters are typically designed so the cracking in order to meet the assumption of size-in-
instrument deflection depends linearly on load. This dependent mechanical properties in the calculation of
deflection is quantified using a constant called the machine compliance. Cracking can typically be
machine compliance, which is defined as the instru- identified by a pop-in on the load–depth trace or by
ment deflection divided by applied load. Using microscopy of the residual nanoindentation impres-
machine compliance, displacement arising from the sion. Unfortunately, the choice for the threshold
instrument deflection at a given load is calculated nanoindentation size below which the data are
and subtracted from measured displacement to affected by tip imperfections and surface properties is
determine penetration depth. However, directly subjective, and the machine compliance and subse-
measuring instrument deflection to calculate machine quent probe area function can be substantially affec-
compliance is difficult for an operator. Therefore, ted by user bias.
machine compliance is typically determined using The size of the tip imperfections will vary between
results from nanoindentation experiments in a cali- Berkovich probes, and the utility of a probe for a
bration material with known properties [1, 4, 5, 7]. given application is often dictated by the size of the
For the common three-sided pyramidal Berkovich probe’s tip imperfections. For example, when testing
probe [8], the machine compliance is often deter- thin films, small tip imperfections are desired
mined using a series of quasistatic nanoindentations because the fully formed Berkovich plastic zone can
or a continuous stiffness nanoindentation in a fused be developed at smaller nanoindentation sizes, which
silica calibration standard. This same set of calibra- can reduce substrate effects [10]. Similarly, when
tion data is also used to calculate the probe area using the structural compliance method to remove
function. Fused silica is considered an accept- edge effects on a material like wood cell walls, it is
able calibration material because it is assumed to be a necessary to use only the range of nanoindentation
homogeneous material with known elastic modulus sizes over which elastic modulus and hardness are
and hardness. An ideal Berkovich probe is geomet- independent of nanoindentation size and not affected
rically self-similar, which means when testing a by tip imperfections [14–16]. Minimal tip imperfec-
homogeneous material such as fused silica the tions are also desired in the deconvolution of geo-
assessed mechanical properties can be assumed to be metric indentation size effects caused by tip
independent of nanoindentation size. Many of the imperfections, and the intrinsic material indentation
commonly used analyses for machine compliance size effects that are often of interest in nanoindenta-
calibration rely on this assumption [1, 6, 7]. However, tion studies [13]. Tip imperfections in Berkovich
in practice, this assumption is violated at low loads probes are commonly conceptualized by assuming
because of tip imperfections and fused silica surface the tip of the probe is spherical, and estimates of the
properties. Tip imperfections refer to any deviation of radius of the spherical tip are used as a metric of the
the Berkovich probe from its ideal shape near its tip, probe quality. The tip radius can be estimated
which may include rounding, flattening, or some directly using a scanning probe microscopy image of
other irregularity. It is well known that for real Ber- the tip or indirectly using the truncation depth esti-
kovich probes, the geometrically self-similar mated from a series of calibration nanoindentations
assumption is violated below a threshold nanoin- [11]. However, in practice, whether a Berkovich
dentation size, and measured properties, especially probe tip is spherical, flat, or exhibits some other
hardness, have an indentation size effect caused by geometric irregularity is inconsequential. The essen-
tip imperfections [1, 9–12]. This geometric tial parameter is the threshold nanoindentation size
4816 J Mater Sci (2018) 53:4814–4827
material parameter J1/2 0 is independent of A0 and b. If Experiments were performed on the fused silica cal-
both H and Eeff are independent of nanoindentation ibration standard provided by Hysitron. The fused
size, then J1/2
0 is also independent of P0. For Berko- silica was stored in ambient conditions and cleaned
vich nanoindentations on fused silica larger than before experiments by gently wiping the surface with
those affected by tip imperfections or fused silica dry tissue paper. For Probe #1, two sets of data were
surface properties, J1/2 0 is assumed to be independent collected. The first set of Probe #1 data was from
of P0, and CtP1/20 plotted as a function of P01/2 forms a when the probe was clean, and this set of data was
straight line of slope Cm and y-intercept of J1/2 0 . used to demonstrate acceptable fused silica calibra-
After correcting the fused silica calibration tions. The second set of Probe #1 data was from when
nanoindentation P–h traces for Cm, the Oliver–Pharr the probe was dirty. The probe became dirty while
method to calculate a probe area function can then be performing experiments on a research material, and
utilized [1, 5]. The Oliver–Pharr contact depth hc [5] is the series of fused silica nanoindentations was orig-
calculated for a Berkovich probe using inally performed to diagnose suspect experimental
hc ¼ h0 eCp P0 ð5Þ results. However, the data was used to demonstrate
how the proposed analyses can be used to detect a
where h0 is the depth immediately prior to unload- dirty probe. Probe #2 was found to have much larger
ing, and e is a constant that depends on the probe tip imperfections than Probe #1, and the Probe #2
geometry. Based on results of Sneddon, e can be data were included to show how the proposed
determined to be 0.72, 0.75, and 1 for conical, para- analyses can be used to determine probe quality.
boloid of revolution, and punch probe geometries, The sets of experiments consisted of series of
respectively [21]. Although of these probe geometries 80-120 load-control quasistatic nanoindentations with
a Berkovich probe is most similar to a cone, the P0 ranging from 0.01 to 12 mN. A 2-lN preload was
paraboloid of revolution value of e = 0.75 is found to used, and each nanoindentation was preceded by a
be more appropriate based on experimental obser- pre-nanoindentation liftoff that consisted of a 2-s
vations [5] and finite element analysis [22]. Therefore, segment during which the probe was lifted 20 nm
e = 0.75 is used here. and disengaged from the surface, followed by a 2-s
Using the known material and probe elastic prop- reapproach segment. As will be shown in the Results
erties and combining Eqs. 2 and 3, the calculated area section, an analysis of the pre-nanoindentation liftoff
Acalc
0 is determined for each calibration nanoinden- was used to define the initial surface contact and as a
tation using check for probe tip cleanliness. The load function
1 1 p1=2 1 m2s 1 m2d consisted of a 5-s load, 5-s hold at P0, 2-s unload to
Acalc
0 ¼ þ ð6Þ
b2 C2p 2 Es Ed 40% P0, 60-s thermal drift segment at 40% P0, and
finally a 1-s unload. The thermal drift was measured
The probe area function is then calculated by fitting from the final 40 s of the thermal drift segment and
the Acalc
0 –hc data from the series of nanoindentations used to correct the depth of penetration h. The initial
to a function, such as unloading stiffness dP/dh at P0 of each unloading
X
5
i1
segment was assessed by fitting the Oliver–Pharr
A0 ðhc Þ ¼ c0 h2c þ ci h1=2
c ð7Þ power law function P = A (h - hf)m, where A, hf, and
i¼1
m are fitting parameters [5], to 40–95% of P0 in the
where ci are fitting parameters. The c0 can either be a unloading segment. The Es of fused silica was cal-
fitting parameter or be set the ideal value of 24.5 for culated using Eqs. 2 and 3, where Ed and md for the
an ideal Berkovich probe. Here, the ideal c0 = 24.5 diamond probe were assumed to be 1137 GPa and
value is used. 0.07, respectively, and ms for fused silica was assumed
to be 0.17. For the calculation of the area function
using Eq. 6, Es = 72 GPa was assumed for the fused
Experimental procedure silica calibration standard.
A Quesant (Agoura Hills, California, USA) atomic
A Hysitron (Minneapolis, Minnesota, USA) TI 900 force microscope (AFM) incorporated in the Tri-
TriboIndenter upgraded with a performech controller boIndenter was used to image the Berkovich probe
was used with two different Berkovich probes. tips and residual nanoindentation impressions in the
4818 J Mater Sci (2018) 53:4814–4827
Results
Pre-nanoindentation liftoff analysis
J1/2
0 = 1.22 lm/N1/2 for Berkovich probe nanoin-
dentations on fused silica [1, 14] to decide whether or
not the calibration data are acceptable. A problem
with this method is that visually choosing a range of
data that looks like a straight line is subjective. For
damaged or dirty probes, it may also be possible to
inadvertently choose a range of data that just hap-
pens by chance to result in the expected value of SYS
plot J1/2
0 = 1.22 lm/N1/2.
A new analysis was needed to aid in identifying
the low-load nanoindentations to exclude in the
determination of Cm and detecting dirty or damaged
Berkovich probes. The systematic SYS plot analysis
was developed for these reasons. The analysis was
based on the idea that for data not affected by frac-
ture events, a threshold nanoindentation size exists
above which the effects of tip imperfections and
fused silica surface properties become negligible on
the assessed mechanical properties. Therefore, data
from nanoindentations larger than this threshold size
in the SYS plot form a straight line because J1/2 0 =
H1/2/Eeff will be a constant. Experimentally, the
threshold nanoindentation size and the straightness
of the SYS plot above this threshold size were
determined by systematically increasing the mini-
mum P1/2 0 used for the straightline fit in the SYS plot
datum by datum. Then, the resulting Cm and SYS
plot J1/2
0 are plotted as a function of the minimum SYS
plot P1/20 . If a threshold nanoindentation size exists
above which the SYS plot is a straight line, then in the
systematic SYS plot analysis there will be a minimum
SYS plot P1/2 0 above which both Cm and J01/2 are
constant values. The systematic SYS plot analyses are
shown in Fig. 4 for the SYS plots in Fig. 3. When the
minimum SYS plot P1/2 0 included the smallest
nanoindents, the slope of the straightline fit was
higher and the y-intercept was smaller, which results
in larger values of Cm and smaller values of J1/2 0 ,
respectively. As the minimum SYS plot P1/2 0 Figure 4 Systematic SYS plot analyses from series of fused silica
increased and fewer of the affected lower load calibration nanoindentations using a Probe #1 when it was clean, b Probe
nanoindentations were included in the fits, Cm #1 when it was dirty, and c Probe #2, which was damaged. The SYS plot
decreased and J01/2 increased. square root Joslin-Oliver parameters (J1/20 ) and machine compliances (Cm)
In the clean Berkovich Probe #1 systematic SYS plot were calculated using the corresponding SYS plots in Fig. 3 and the SYS
correlation (Eq. 4) by systematically removing the smallest nanoinden-
analysis (Fig. 4a), a threshold nanoindentation size
tations and fitting a straight line. The horizontal dashed line represents the
was evident at a minimum SYS plot P1/2 0 of about
expected J1/2
0 = 1.22 lm/N
1/2
for Berkovich probe nanoindentations on
0.025 N1/2. Above this threshold, both Cm and SYS fused silica [1, 14]. The vertical dashed line in a corresponds to the
plot J1/2
0 were fairly constant values. Above a mini- threshold minimum SYS plot P1/2 1/2
0 = 0.025 N identified for this data as
mum SYS plot P1/2 0 value of about 0.065 N1/2, the described in the text. The error bars represent uncertainties based on a
variability in the Cm and SYS plot J1/2 0 values least squares analysis for corresponding straightline fits.
increased because fewer data points were being used
J Mater Sci (2018) 53:4814–4827 4821
in the fits. A further quality check of the calibration was depth in the fused silica calibration experiments. It
done at this point by comparing the value of SYS plot was therefore decided that Probe #2 had too large of
J1/2
0 to the expected value of 1.22 lm/N1/2 for Berko- tip imperfections and could not be properly cali-
vich probe nanoindentations on fused silica [1, 14]. The brated over the range of nanoindentation sizes
J1/2
0 value for the clean Berkovich Probe #1 was very obtained in these calibration experiments. The probe
close to this expected value (Fig. 4a) above the may be useful for larger nanoindentations above
threshold size. Therefore, it was determined that these which the tip imperfections may become negligible.
fused silica calibration data were acceptable and could The necessity of the systematic SYS plot analysis was
be used to determine the Cm and probe area function. clearly demonstrated in the analysis of the data from
The systematic SYS plot analysis was also useful as Probe #2. The previous method commonly used to
a sensitive tool to detect dirty and damaged Berko- calculate Cm from the SYS plot is to visually choose the
vich probes. Figure 4b shows the systematic SYS plot range of data that looks like a straight line, fit the data
analysis of the SYS plot in Fig. 3b. It would appear to a straight line, and compare the value of the SYS plot
that above a minimum SYS plot P1/2 0 value of J1/2
0 to the expected value of J01/2 = 1.22 lm/N1/2. In
0.04 N1/2, there were nearly constant values of Cm the SYS plot (Fig. 3c), the data above P1/2
0 = 0.06 N1/2
and J1/2
0 . However, the average value of J0
1/2
was looked like a straight line. When those data were fitted
1/2
about 1.13 lm/N , which was much lower than the to a straight line, the value of SYS plot J1/2
0 was very
expected value of 1.22 lm/N1/2. Additionally, the near the expected 1.22 lm/N1/2 value as seen in the
value for Cm was about 1.3 lm/N, which was much systematic SYS plot analysis (Fig. 4c). This previous
higher than the value anticipated based on prior analysis would have erroneously led the user to con-
experience with this probe and transducer. These clude that above P1/20 = 0.06 N1/2, the effects of probe
observations led to the conclusion that the probe was tip imperfections and fused silica surface properties
either dirty or had become damaged. Visual inspec- became negligible and the probe was behaving like a
tion under a dissecting microscope revealed debris characteristic Berkovich probe. Only after the system-
near the probe tip, suggesting the probe was just atic SYS plot analysis, it was observed that the probe
dirty. This was confirmed after cleaning the probe was not actually behaving like a Berkovich probe. This
and the repeated fused silica calibration experiments observation led to further investigation using AFM
resulted in an acceptable systematic SYS plot analysis imaging and the conclusion that the probe tip imper-
similar to that in Fig. 3a (data not shown). fections for Probe #2 were so large that the probe must
The systematic SYS plot analysis in Fig. 4c is of the be considered damaged and not reliable as a Berkovich
data in Fig. 3c and demonstrates how the analysis probe over this range of nanoindentations sizes.
can be used to detect a damaged probe. Although J1/2 0
was near the expected 1.22 lm/N1/2 value for mini- Fused silica calibration
mum SYS plot P1/2 0 between 0.04 and 0.08 N1/2, nei-
1/2
ther Cm nor J0 reached a constant value in this After determining a set of Berkovich probe fused
region. In this region, the SYS plot J1/20 and Cm sys- silica calibration data is acceptable, the Cm and probe
tematically increased and decreased, respectively. area function calibration procedures can proceed as
This was suspicious behavior that did not change in demonstrated here using the clean Probe #1 data
repeated calibration experiments even after cleaning (Figs. 3a and 4a). In the systematic SYS plot analysis
the probe. To further investigate the probe cleanliness (Fig. 3a), Cm was observed to vary some even above
and tip imperfections, an AFM height image of Probe the threshold minimum SYS plot P1/2 0 value of
1/2
#2 was obtained (Fig. 5b). For comparison, an image 0.025 N . Looking at the reasonable range of values
was also obtained of Probe #1 (Fig. 5a). Probe #1 had above the threshold nanoindentation size and below
the anticipated pyramidal shape (Fig. 5a), and in the the values where the fit uncertainties started to
contour map, a triangular shape could be observed at noticeably increase (between 0.025 and 0.065 N1/2),
depths as shallow as 20 nm (Fig. 5c). In contrast, the calculated value of Cm ranged between 0.48 and
Probe #2 had numerous irregularities near its tip 0.64 lm/N. The average value of Cm = 0.55 lm/N
(Fig. 5b). These irregularities manifested as irregular was chosen. Although there was an uncertainty on
shapes in the contour plots at depths up to 200 nm the order of 10% in Cm, this uncertainty had little
(Fig. 5d), which covered nearly the entire range of effect on the Acalc
0 and hc used in the probe area
4822 J Mater Sci (2018) 53:4814–4827
calibration for this set of fused silica data. The effect average value of 1.217 ± 0.002 lm/N1/2 (uncertainty
of Cm uncertainties would be the largest for the big- standard error). The hc = 35 nm directly corre-
gest nanoindentations. For the biggest nanoindenta- sponded to the minimum SYS plot P1/2 0 = 0.025 N1/2
1/2
tions in this data set, an uncertainty of 10% in Cm threshold in Fig. 4a. The constant J0 above the
resulted in uncertainties of less than 1 and 0.2% for threshold size was a good check that the Cm was
Acalc
0 and hc, respectively. The effect was even smaller correct. The range of hc over which J1/2
0 was near the
for the smaller nanoindentations. However, this expected value of 1.22 lm/N1/2 was also a useful
uncertainty in Cm must be taken into consideration indication of the range of nanoindentation sizes over
when correcting experimental P–h traces of other which the probe exhibited the expected self-similar
materials, especially for experiments with stiffer behavior of a Berkovich probe testing fused silica.
contact compliances. To calculate the probe area function, Eq. 6 was
After correcting the fused silica calibration used to calculate Acalc
0 for each nanoindent, and then
P–h traces for Cm, Cp and hc were calculated for each Eq. 7 was fit to the Acalc
0 –hc data. After that, Eqs. 2 and
nanoindentation using the Oliver–Pharr method [5]. 3 were used to calculate Es, and Eq. 1 was used to
Figure 6a shows J1/20 = CpP1/2
0 plotted as a function calculate H. The Es = 72 GPa in Fig. 6b was expected
of hc. A threshold was observed at hc = 35 nm. because that was the value assumed for Acalc 0 .
Above this threshold J1/2 0 is constant and had an Checking that Es was the constant 72 GPa over the
J Mater Sci (2018) 53:4814–4827 4823
entire range of nanoindentation sizes was useful only The comparison of J1/2 0 to the expected value of
for confirming that the area function equation (Eq. 7) 1.22 lm/N1/2 [1, 14] is one of the criteria used to
was fit well to the entire range of experimental Acalc
0 –
decide whether or not fused silica calibration data are
hc data. It did not provide information about the acceptable for Cm and probe area function calibra-
accuracy of the Cm calculation, probe cleanliness, tions. There are many factors that may cause J1/20 to
probe tip imperfections, or overall quality of the deviate from its expected value. As shown in this
fused silica calibration nanoindentations. Because paper, J1/2
0 can be affected by a dirty probe, probe tip
H1/2 = J1/2
0 /Eeff and the probe area function was fit
imperfections, or fused silica surface properties. In
well over the entire range of experimental Acalc0 –hc
addition, the J1/2
0 value can also deviate from the
data, it was expected that the trend in H with expected 1.22 lm/N1/2 value if the properties of the
nanoindentation size would be similar to the J1/2
0 –hc
fused silica standard are different, the fused silica
4824 J Mater Sci (2018) 53:4814–4827
In Eq. 8, Acalc
0 depends on the squares of b and the The A0calc was calculated using Eq. 8 and was normalized to the
experimental J1/2 value of Acalc at J01/2 = 1.22 lm/N1/2 and b = 1. The b values of
0 . Those dependencies on a normal-
0
calc
ized A0 are shown in Fig. 8. The Oliver and Pharr 1.034 and 1.055 were from King [24] and Strader et al. [19],
respectively. The chosen value of b also had a large effect on the
suggestion that a proper value of J1/2
0 should be in the
normalized Acalc
0 . For example, the values of 1.034 and 1.055
range of 1.22 ± 0.04 lm/N1/2 can result in a differ-
resulted in a 7 and 11% decrease in normalized Acalc 0 at J01/2 =
ence in Acalc
0 of ± 7%. This means that two ‘‘proper’’ 1/2
1.22 lm/N , respectively.
J Mater Sci (2018) 53:4814–4827 4825
3. The accuracy of the probe area function equation when using Hcalc to check the calibration data. The
fit to the Acalc
0 –hc data. chosen b does not have an effect on fused silica Es,
4. The quality of the instrument calibrations to because the effect of the b choice on Acalc
0 is exactly
measure loads and displacements. compensated when the same value of b is used to
calculate Es using Eq. 3.
All these elements can be addressed in a short
The lower end of the reported range of hc over
statement. For example, this is achieved for clean
which constant J1/2 values are measured (for exam-
Probe #1 with the statement: ‘‘Constant values of J1/2
0 = 0
ple, hc = 35 nm for clean Probe #1) could also be used
1.217 ± 0.002 lm/N1/2, Es = 72.0 ± 0.2 GPa, and
as a metric of the size of the probe tip imperfections.
H = 9.13 ± 0.03 GPa (uncertainties are standard
However, this would require assuming that the
errors) were assessed for fused silica calibration
dominate factor affecting the smallest nanoindenta-
nanoindentations with hc between 35 and 211 nm; no
tions is tip imperfections and not fused silica surface
systematic variations of Cm or J1/20 were observed in
properties. For clean Probe #1, surface roughness
the systematic SYS plot analysis over this range of hc.’’
may have been the dominating factor affecting the
Element 1 is addressed by including the range of hc
smallest nanoindentations (Fig. 7). Therefore, a fused
over which a constant J1/20 , Es, and H are observed
silica surface with lowered roughness would be
and the statement that there is no systematic varia-
needed to more accurately characterize the sizes of
tion of J1/2
0 in the systematic SYS plot analysis over
probe tip imperfections. Nevertheless, by testing
the reported range of nanoindentation sizes. Element
multiple probes in the same fused silica specimen,
2 is addressed by stating that J1/2
0 is a constant value
relative comparisons could be made, and recording
and there is no systematic variation of the Cm in the
this value of hc would also be a useful way to track
systematic SYS plot analysis. Element 3 is addressed
the size of the probe tip imperfections during the
by the statements that Es and H were constants over
probe’s lifetime.
the same range of nanoindentations as J1/2 0 , and that
If the probe area function is used in experiments
Es is the assumed value of 72 GPa used in Acalc 0 over
for nanoindentations outside the reported range of hc
the range of nanoindentation sizes. Element 4 is
over which constant mechanical property values are
addressed by stating that the average constant value
assessed in fused silica, then the additional
of J1/2
0 is near the expected value of 1.22 lm/N1/2.
Including the uncertainties in the reporting sentence
is useful for conveying the variation in the calibration
data.
Another useful check for the calibration is com-
paring the fused silica H to the expected calculated H
(Hcalc) determined using the experimental J1/2 0 and
the equation
P0 2 p1=2 1 m2 1 m2 2
1=2
H calc ¼ calc ¼ b2 J0 s
þ d
A0 2 Es Ed
ð9Þ
Similar to Acalc
0 in Eq. 8, Hcalc depends on the
squares of b and the experimental J1/2 0 , and those
dependences on Hcalc are shown in Fig. 9. By input-
ting J1/2
0 = 1.217 lm/N1/2 and b = 1 into Eq. 9, it Figure 9 The calculated hardness (Hcalc) plotted as a function of
was calculated that Hcalc = 9.14 GPa for clean Probe the square root of the Joslin-Oliver parameter (J1/2 0 ) based on
#1 at hc between 35 and 211 nm. This Hcalc compared Eq. 9. The b values of 1.034 and 1.055 were from King [24] and
well to H = 9.13 ± 0.03 GPa over this range. This Strader et al. [19], respectively. The chosen value of b also had a
large effect on the Hcalc. For example, the values of 1.034 and
check further verified that the calibration calculations
1.055 resulted in a 7 and 11% increase in Hcalc at J1/2
0 = 1.22 lm/
were done correctly, such as the calculation of Acalc0
N1/2, respectively. Similar increases in experimental fused silica
using Eq. 6. The chosen b has a substantial effect on H would be expected if the corresponding values of b were used
fused silica H and must be taken into consideration in the fused silica calibration analyses.
4826 J Mater Sci (2018) 53:4814–4827
assumptions being employed in making such mea- they rely on a straightline fit of calibration data cho-
surements must also be stated. For example, for lar- sen by the user that are assumed to exhibit size-in-
ger nanoindents, the common assumptions are that dependent mechanical properties. Therefore, the
the Berkovich probe effectively has an ideal geometry basic construct behind the systematic SYS plot anal-
and that using the coefficient c0 = 24.5 in fitting the ysis could be used to develop analogous systematic
probe area function (Eq. 7) accurately captures the analyses of these other area-independent calibration
shape. Assumptions for smaller nanoindentations are methods.
more problematic. One of the assumptions to calcu-
late Acalc
0 is that Es is a constant 72 GPa. It is unknown Compliance with ethical standards
how valid that assumption is for the smallest
Confict of interest The author declares that he has
nanoindents, which are likely affected by the fused
no conflict of interest.
silica surface properties and probe tip imperfections.
If constant Es is assumed, then the calibration
reporting should also include this information. References
Additionally, the full hc range over which a constant
Es is calculated should be included to communicate [1] Oliver WC, Pharr GM (2004) Measurement of hardness and
the accuracy of the probe area function equation fit to elastic modulus by instrumented indentation: advances in
the Acalc
0 –hc data that include the smaller nanoin- understanding and refinements to methodology. J Mater Res
dentations. Furthermore, simulations using Berko- 19:3–20
vich probes modeled with spherical tip imperfections [2] Fischer-Cripps AC (2006) Critical review of analysis and
show that below a threshold nanoindentation size, interpretation of nanoindentation test data. Surf Coat Technol
H may be higher or lower than H that would have 200:4153–4165
been assessed with an ideal Berkovich probe, and the [3] Pethica JB, Hutchings R, Oliver WC (1983) Hardness
trends depend on the tested material’s ratio of yield measurement at penetration depths as small as 20 nm. Philos
stress to elastic modulus and strain hardening index Mag A 48:593–606
[9, 12]. Therefore, caution must be exercised when [4] Doerner MF, Nix WD (1986) A method for interpreting the
using the probe area function to measure properties, data from depth-sensing indentation instruments. J Mater
especially H, in experimental materials near or below Res 1:601–609
the lower range of hc over which constant mechanical [5] Oliver WC, Pharr GM (1992) Improved technique for
property values are assessed in fused silica. determining hardness and elastic modulus using load and
displacement sensing indentation experiments. J Mater Res
7:1564–1580
Concluding remarks [6] Fischer-Cripps AC (2004) Nanoindentation, 2nd edn.
Springer, New York
The amount of user subjectivity in Berkovich probe [7] Stone DS, Yoder KB, Sproul WD (1991) Hardness and
Cm and probe area function calibrations can be elastic modulus of TiN based on continuous indentation
decreased by using the demonstrated pre-nanoin- technique and new correlation. J Vac Sci Technol A
dentation liftoff and systematic SYS plot analyses. 9:2543–2547
Additionally, because some user bias and uncertain- [8] Berkovich E (1951) Three-faceted diamond pyramid for
ties in calibration results will always exist, the fused micro-hardness testing. Ind Diam Rev 11:129–133
silica calibration results should be reported to facili- [9] Chen W, Li M, Zhang T, Cheng Y-T, Cheng C-M (2007)
tate better comparisons of nanoindentation experi- Influence of indenter tip roundness on hardness behavior in
ments performed with different Berkovich probes. nanoindentation. Mater Sci Eng A 445–446:323–327
Although any set of Berkovich probe fused silica [10] Chen J, Bull SJ (2009) On the factors affecting the critical
calibration data could be analyzed using the SYS indenter penetration for measurement of coating hardness.
correlation and systematic SYS plot analysis as Vacuum 83:911–920
demonstrated in this paper, operators may also [11] Fischer-Cripps AC (2010) The sharpness of a Berkovich
choose to utilize other area-independent correlations indenter. J Mater Res 25:927–934
to assess the Cm [1, 6]. These other methods are
similar to the SYS correlation method. To estimate Cm
J Mater Sci (2018) 53:4814–4827 4827
[12] Li M, Chen W, Cheng Y-T, Cheng C-M (2009) Influence of [18] Chudoba T, Jennett NM (2008) Higher accuracy analysis of
contact geometry on hardness behavior in nano-indentation. instrumented indentation data obtained with pointed inden-
Vacuum 84:315–320 ters. J Phys D Appl Phys 41:215407
[13] Nix WD, Gao H (1998) Indentation size effects in crystalline [19] Strader JH, Shim S, Bei H, Oliver WC, Pharr GM (2006) An
materials: a law for strain gradient plasticity. J Mech Phys experimental evaluation of the constant b relating the contact
Solids 46:411–425 stiffness to the contact area in nanoindentation. Philos Mag
[14] Jakes JE, Frihart CR, Beecher JF, Moon RJ, Stone DS (2008) 86:5285–5298
Experimental method to account for structural compliance in [20] Joslin DL, Oliver WC (1990) New method for analyzing
nanoindentation measurements. J Mater Res 23:1113–1127 data from continuous depth-sensing microindentation tests.
[15] Jakes JE, Frihart CR, Beecher JF, Moon RJ, Resto PJ, J Mater Res 5:123–126
Melgarejo ZH, Surez OM, Baumgart H, Elmustafa AA, [21] Sneddon IN (1965) Relation between load and penetration in
Stone DS (2009) Nanoindentation near the edge. J Mater Res axisymmetric Boussinesq problem for punch of arbitrary
24:1016–1031 profile. Int J Eng Sci 3:47–57
[16] Jakes JE, Hunt CG, Yelle DJ, Lorenz LF, Hirth K, Gleber [22] Pharr GM, Bolshakov A (2002) Understanding nanoinden-
S-C, Vogt S, Grigsby W, Frihart CR (2015) Synchrotron- tation unloading curves. J Mater Res 17:2660–2671
based x-ray fluorescence microscopy in conjunction with [23] Harding DS, Oliver WC, Pharr GM (1994) Cracking during
nanoindentation to study molecular-scale interactions of nanoindentation and its use in the measurement of fracture
phenol-formaldehyde in wood cell walls. ACS Appl Mater toughness. In: Demczyk BG, Williams ED, Garfunkel E,
Interfaces 7:6584–6589 et al (eds) Evolution of thin-film and surface structure and
[17] Bolshakov A, Pharr GM (1997) Inaccuracies in Sneddon’s morphology. Mater Res Soc Symp Proc 356, Pittsburgh, PA,
solution for elastic indentation by a rigid cone and their p 663
implications for nanoindentation data analysis. In: Gerberich [24] King RB (1987) Elastic analysis of some punch problems for
WW, Gao H, Sundgren J-E, Baker SP (eds) Thin films: a layered medium. Int J Solids Struct 23:1657–1664
stresses and mechanical properties VI. Mater Res Soc Symp
Proc 436, San Francisco, CA, USA, p 189