Sei sulla pagina 1di 4

NEW GENERATION OF TORQUE TUBE

INTELLIGENT

FST-3000 Series
TORQUE TUBE, DISPLACE TYPE

GENERAL
FST-3000 is a torque tube, displacement type level indicator / trans-
mitter with intelligent data processing function. Adding to all features
of existing torque tube type level transmitters, highly advanced signal
processing unit offers high function and high accuracy as well as func- N-
tion of field adjustment of density, range etc. O SIO
X PL
The measured data is indicated on LCD display and output by DC4 ~ YE
20mA signal. 4 point level alarm contact is optionally available. Also, F ET
SA BLE
remote control, function check and monitoring etc. can be done by I N SIC AILA
R V
HART digital communication. NT EA
E X I TYP
Several different material selection is ready to meet various applica- AT OF
O
tions such as high temperature application of boilers as well as low PR
temperature applications up to minus 196˚C of liquefied gases.
Intrinsically safe version is newly added.

STANDARD SPECIFICATION Ra=46.3UB - 565 (Ω)


Detection Method : Torque Tube Type Displacer ex. at UB=24V, Ra=546Ω
Measuring Range : Min. 0 to 300mm at UB=40V, Ra=1287Ω
Max. 0 to 3000mm 2) FST-3 ■ ■ ■S2
Other measuring ranges on request. Consult factory for details. UB=DC11.0 ~ 26V
Liquid density : Ra=650Ω (FOR ISOLATION BARRIER)
1) For level measurement :0.2 to 1.5g/cm3 Ra=(UB-11)/ 22 [kΩ] (FOR ZENNER BARRIER)
2) For interface measurement ex. at UB=24V Ra=591Ω
Difference of density of 2 liquids : To be more than 0.1g/cm3 Communication
3) Density measurement : Min. density span 0.2g/cm3 Protocol : HART™
Other density range on request. Consult factory for details. Transmission : FKC (Frequently irregular system)
Available Pressure class : ANSI#150, #300, #600, #1500 and #2500 •Communication needs the load resistance beyond 250Ω
class Alarm Contact (Option) : SPST relay contact, Max. 4 points
Fluid Temp. Range : -196 to +400˚C Capacity AC 125V, 3A DC 30V, 3A
Installation : Tank Top Installation or Tank Side Installation •Optionally available only for Weather proof version (FST-
Connection Size : External chamber type: 1 1/ 2" or 2" flange 3 ■ ■ ■W) and Pressure tight ex-proof version (FST-3 ■ ■ ■E).
Internal chamber type: 4" flange NOT available for intrinsically safe version and HART communi-
Amb. Temp. : Weather Proof -20 to 60˚C cation Foundation.
Pressure Tight Ex-proof -20 to 55˚C •Separate power supply of DC24V is additionally required than sig-
Intrinsically safe (S2) -20 to 60˚C nal line. One cable entry will be added. Alarm action to be speci-
Intrinsically safe (S+H) -20 to 40˚C fied (Hi or Lo).
Available Material : Accuracy : ±0.5% F.S. (Based on the mass conversion correction at
1) Torque tube : SUS316L, NW0276 (Hastelloy C-276® equiva the time of delivery from factory)
lent), Inconel Enclosure :
2) Bonnet : SCPH2,SCS13A, SCS14A, SCS16A,Titanium 1) Weather Proof
3) Chamber : Carbon steel, SUS304, SUS316, SUS316L 2) Pressure Tight Explosion Proof
Titanium, NW0276 (Hastelloy C-276® equiva 3) Intrinsically Safe
lent), Monel, PVC Cable Entry : Std. 1 x G1/2
4) Displacer : SUS304, SUS316, SUS316L Titanium, Option 1 x 1/2NPT
NW0276 (Hastelloy C-276® equivalent), Monel, For alarm contact version, 1 cable entry (Max. G3/4 or 3/4NPT)
PVC will additionally be provided.
5) Indicator : AC2A Intelligent Data Setting Function :
Local Indication : By 2 line LCD indication (Height 12.7mm) Ex-proof Built-in Key Board provided for data setting (Data setting
Line 1: Function Indication (4 digit) can be conducted in Hazardous area during operation.)
Line 2: Data Indication (4 digit) Setting items :
Output : DC4 ~ 20mA, 2 wire Burn-out setting, Density, Alarm point (Max. 4 points), Alarm re-
Digital signal : Superimposed to 4 to 20mADC output signal set span, Alarm action, Analog output direction, Indication unit
Supply Voltage (UB) and max. load (Ra) (mm, % or density), Analog output test, Indication test, others
1) FST-3■■■W and FST-3■■■E
UB=DC 12.2 ~ 40V

HART is a registered trademark of


TG-L545-5E JUN. 2000K
the HART Communication Foundation.
TG-L545-7E JAN. 2011K
FST-3000 SERIES TORQUE TUBE, DISPLACER TYPE LEVEL INDICATOR / TRANSMITTER

MODEL CODE

FST-3 - - DESCRIPTION
1 upto 300mm
2 301 to 500mm
3 501 to 700mm
4 701 to 1000mm
MEASURING SPAN 5 1001 to 1500mm
6 1501 to 2000mm
7 2001 to 2500mm
8 2501 to 3000mm
9 Special
1 SIDE to SIDE
2 BOTTOM to SIDE
3 SIDE to TOP
CONNECTION 4 BOTTOM to TOP
5 TOP
6 SIDE
9 Special
1 Carbon steel / SUS304 / Carbon steel
2 SCS13A / SUS304 / SUS304
MATERIAL
3 SCS14A / SUS316 / SUS316
Bonnet / Displacer / Chamber
4 SCS16A / SUS316L / SUS316L
9 Special
W Weather Proof (IP65 equiv.)
E Pressure Tight Ex-proof (Exd ll CT6)
S2 Intrinsically safe (Exia ll CT4) *5
ENCLOSURE
W+H Weather Proof + HART (IP65 equiv.) *6
E+H Pressure Tight Ex-proof + HART (Exd ll CT6) *6
S+H Intrinsically safe + HART (Exia ll CT6) *6
- 0 Not provided
- 1 1 point provided
ALARM CONTACTS *1 - 2 2 point provided
- 3 3 point provided
- 4 4 point provided
1 ANSI (JPI) #150
2 ANSI (JPI) #300
3 ANSI (JPI) #600
4 JIS 10K
PRESS. RATING 5 JIS 30K
6 JIS 63K
7 ANSI (JPI) #1500
8 ANSI (JPI) #2500
9 Special
1 1-1/2" Flange
2 2" Flange
CONN. SIZE
3 4" Flange
9 Special
1 0˚C <=t< =100˚C Standard type + No Fin
2 -30˚C <= t <0˚C or 100˚C < t =< 150˚C Standard type + Single Fin
3 -196˚C < = t <-30˚C or 150˚C < t <= 200˚C Extension type + Single Fin
LIQUID TEMP. RANGE (t)
4 200˚C < t < = 300˚C Extension type + Double Fin
5 300˚C < t Extension type + Triple Fin
9 Special
1 SUS316L -196~200˚C
2 INCONEL -196~400˚C
TORQUE TUBE MATERIAL
3 NW0276 (Hastelloy C-276® equivalent) -196~200˚C
9 Special
- 1 G 1/2 (F)
- 2 1/2 NPT (F)
CABLE ENTRY
- 3 G 1/2 (F) With Ex-proof Cable Gland *3
- 9 Special
L1~L8 Left Hand Side
ROTATION ANGLE *4
R1~R8 Right Hand Side

*1 : Not available for Intrinsically safe versions, specify ■


c0 .
*2 : Applicable only for tank top mounting type FST-3 ■ 5 ■
*3 : Specify the cable diameter.
*4 : Refer to following figures for installation. ex.L1
*5 : Recommended Safety Barrier
Isolation Barriers
MTL4041B-SCC, MTL3046B (MTL)
KHD2-STC3-Ex1. P (P&F)
Zenner Barriers
MTL728+(MTL)
Z728, Z928 (P&F)
Acceptable inductance :<
= 0.3 mH
Acceptable capacitance :<
= 0.04 µF
*6 : HART communication is not available for alarm contact.

*7 Possible combinaton with HART communication


Intrinsically safe Ex-proof Weather proof
Non alarm contact O O O
With alarm contact X X X

2 TOKYO KEISO CO., LTD. TG-L545-7E


FST-3000 SERIES TORQUE TUBE, DISPLACER TYPE LEVEL INDICATOR / TRANSMITTER

DIMENSIONS
(Dimensions are for #150 and #300 class. They will be different for #600 and above.)

Dimensions
F : Standard of Connection
Flange
H : Measuring Range
FF : Installation Dimension
R : Flange surface to High-
est position of span.
(Min. 50mm)
Please keep us informed of the
above Dimensions

Remarks : *1 : They may change for High Pressure Gas Regulation certified products.
*2 : *2 indicate 5 minimum possible dimensions. Consult factory in case different dimensions are required.

TG-L545-7E TOKYO KEISO CO., LTD. 3


FST-3000 SERIES TORQUE TUBE, DISPLACER TYPE LEVEL INDICATOR / TRANSMITTER

MATERIAL OF GASKET, BOLT & NUT WIRING


The following material of Gasket and Bolt & Nut are available. 1) Weather proof version (FST-3 ■ ■ ■ W) and Pressure tight ex-proof
CODE GASKET CODE BOLT & NUT Remarks version (FST-3 ■ ■ ■ E)

1 — 1 SS400 Steel
2 — 2 SUS304 St.stl, Low temp.
3 V#7020 3 SNB7 / S45C (A193-B7 / A194-2H) Steel, High temp.
4 V#8591V 4 A320-L7 / A194-4 Steel, Low temp.
5 V#6591V 5 A193-B8CL2 / A194-8 St.stl, High temp.
6 V#7591V 6 A320-B8CL2 / A194-8 St.stl, Low temp
9 Special 9 Special

2) Intrinsically safe version (FST-3■ ■ ■S2)

Connection

No Polar

DC

FST-3000HART
Communication Load resistance 250Ωmin.
275HART
Communicator

ORDERING INFORMATION
Specify the following for order / inquiry :

1 MODEL FST-3■■■■-■■■■■-■
2 LIQUID NAME
3 Sp. GR.
4 VISCOSITY
5 PRESSURE NOR. MAX. ■MPa ■( )
6 TEMPERATURE NOR. MIN. MAX. ■˚C

* Specification is subject to change without notice.

Head Office : Shiba Toho Building, 1 – 7 – 24 Shibakoen, Minato-ku, Tokyo 105 – 8558
Tel : +81-3 – 3431 – 1625 (KEY) ; Fax : +81-3 – 3433 – 4922
e-mail : overseas.sales@tokyokeiso.co.jp ; URL : http://www.tokyokeiso.co.jp

4 TOKYO KEISO CO., LTD. TG-L545-7E