Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Francesca Rovalino
Craig Syme
CHEM 1251-046
11/13/2019
2
Introduction:
In this lab, concepts of concentration were used to compare standard solutions of copper (II)
sulfate pentahydrate to made solutions of copper (II) sulfate pentahydrate. Molarity is the unit
used to describe the concentration of a solute in a solution and is found by dividing the moles of
solute by the liters of solution. The solute, or dissolved substance, used in this lab was the
copper(II) sulfate, which was mixed with water, the solvent. In this lab, the solution was blue
because it absorbed certain wavelengths of light and in this lab, the greatest absorbance was in
the orange range, which is about 600nm, resulting in a visible blue. The spectrophotometer
(Spec-20) is used to read the percent transmittance of a solution and in order to do so it must be
electromagnetic energy that is converted into an electric current. The Beer-Lambert Law directly
relates the amount of light absorbed by a colored substance and the concentration. The
concentration and absorbance were used to create a calibration curve. Standard solutions were
used in this lab to compare with made solutions of the same solute and solvent. Both standard
and made solutions were useful as a reference for the unknown solution. The general purpose of
this lab was to observe and analyze the accuracy of the standard solutions and the difference
between the standard copper (II) solutions and the made solutions. Another objective was to
identify possible sources of error that could have affected the results.
3
Procedure:
First, ten small, clean test tubes were labeled for different concentrations of Copper (II) Sulfate,
Standard 0.50M, Standard 0.20M, Standard 0.10M, and Standard 0.05M, with tape as follows:
Blank, S1, S2, S3, S4, M1, M2, M3, M4, Unknown. Then the test tubes were filled with clean
Pasteur pipettes to about ¾ full with the appropriate solution, according to the label for the
standard solutions. A kim wipe was then used to clean any solution and fingerprints on the
outside of the test tube and the test tubes were placed on a rack. Next, 2.5g of Copper (II) Sulfate
Pentahydrate was mixed with water, which was gradually added using a graduated cylinder, in a
100mL beaker to create 20mL of a 0.50M solution. The mixture was transferred back and forth
between the graduated cylinder and the beaker to ensure that all the crystals dissolved, creating a
homogenous mixture. The beaker was then rinsed several times with small portions of distilled
water, keeping the volume in the cylinder less than 20mL. Then a Pasteur pipette was used to
carefully add water until the solution reached 20mL. Again, the mixture was transferred back and
forth, thus creating a homogenous mixture and some of the mixture was collected in the M1 test
tube. Next, the process was repeated using the same mixture and 12mL of water to create 20mL
of a 0.20M copper (II) sulfate solution. The amount of water needed was calculated by the
dilution equation and some of the mixture was collected into the M2 test tube. The process was
repeated again to create 20mL of a 0.10M copper (II) sulfate solution for the test tube labeled M3
and then again for 20mL of a 0.05M copper (II) sulfate solution for the test tube labeled M4.
Finally, the rack of test tubes was taken to the spectrophotometer. In order to calibrate the
spec-20, the left knob was slightly adjusted so that the percent transmittance read zero with the
spec-20 empty and with the lid closed. Then the test tube containing only distilled water (Blank)
4
was inserted and the lid was closed. The pointer was set to 100% transmittance with the right
knob. Lastly, each test tube was inserted one by one, with the blank test tube used between each
one to ensure that the spec-20 was still calibrated. The percent transmittance for each sample was
recorded and then used to calculate the absorbance, which was then compared with the
Results:
0.50 2.5 20 - 20
0.20 - 20 8 12
0.10 - 20 10 10
0.05 - 20 10 10
5
Table 12.4 Comparison Between the Theoretical and Actual Concentration of the Made Copper
(II) Solutions
Discussion:
Table 12.5 Percent Error on the Concentration of the Made Copper(II) Solution
Table 12.6 Percent Error on the Absorbance of the Made Copper (II) Solution
Both Table 12.5 and Table 12.6 prove that the technique was an accurate method of measuring
molarity and absorbance since the error was minimal. The percent error was calculated by taking
the absolute value of the actual value (made solution) minus theoretical value (standard solution)
divided by the theoretical. The number was then multiplied by 100, resulting in the percent error.
Since none of the results had a percent error greater than 10, the conclusion can be drawn that by
properly calibrating the spec-20 and with the use of Pasteur pipettes to ensure accuracy in
measurements, error can be minimized, making the experiment effective and accurate.
7
Conclusion:
In conclusion, the made copper (II) solutions had similar concentrations and absorbance in
comparison to the standard solutions. This was proven by the R² value of 0.99972, which
indicates that the data fit well compared to the line of best fit. The line of best fit provided a
reference for the expected absorbance at a certain concentration and vice versa. One error that
could have caused a slightly lower absorbance could have been that the beaker was not properly
rinsed, resulting in a lower concentration of copper (II) sulfate in the mixture. The difference was
not much therefore the percent error was low. Similarly, another error for the first and last
solutions could have been that not enough water was added, resulting in a higher concentration
of copper (II) sulfate. Ultimately, the results aligned with the original hypothesis that absorbance
Sample Calculations:
.20/10= 0.20M