Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Operator Manual
EMBC-220
Physical Address: 148 Epsom Avenue, North Riding A/H, Randburg 2169
Postal Address: P.O. Box 1915, Randburg 2125, South Africa
Tel: +27 11 462 0222 Fax: +27 11 462 0615
www.process-auto.com
Operator Manual
Series 220 Belt Scale Systems
with UNIPROIV Transmitter
(Software Ver. EMBC 220.40.29)
Table of Contents
Description Page
SECTION 1: System Description 3
General 3
SECTION 2: Component Descriptions 4
UNIPROIV Transmitter 4
Weigh Frames 5
Belt Speed Sensors 7
Boom Angle Compensation 10
SECTION 3: Portable Terminal Unit (PTUII) 11
SECTION 4: UNIPROIV Operation 12
General 12
Normal Display Mode And Access To Functions 14
SECTION 5: Access Codes 21
SECTION 6: Scale Configuration & Calibration 22
Application Details 22
Weigh Frames 23
Initial Calibration (Set-Up) 24
SECTION 7: Special Functions 38
Function No. 0 “Simulation Of Input Signals" 38
Function No. 1 "Filtering Of Input/Output Signals" 38
Function No. 2 "Alarms Settings" 39
Function No. 3 “LCB 4-20 Ma Output Set-Up" 39
Function No. 4 “Real Time Clock Settings" 40
Function No. 5 “Totaliser Functions" 40
Function No. 6 “Scada Communications” 44
Function No. 7 “Compensations” 45
Function No. 8 “Sampler Set-Up” 46
Function No. 9 “Production Reports” 46
Description Page
SECTION 8: Special Low Temperature Applications 47
Low Temperature Operation 47
SECTION 9: Calibration Checks 48
Frequency Of Calibration Checks 48
Quick Check On Calibration 49
(Full) Calibration Checks And Tests 51
As-Found-Error (AFE) 56
SECTION 10: Typical Conveyor Scale Log Book 57
SECTION 11: Calibration & Application Data 58
Viewing Of Calibration Constants 58
Entry (Manual) Of Calibration Constants 58
Backup Calibration Constants 58
Recall Of Backed Up Cal Constants 59
Reports Of Cal/Application Data 59
Calibration Constants 59
Application Data 59
Allocated I/O (Expansion I/O Bus) 60
Map Of Allocated I/O (Expansion I/O Bus) 60
SECTION 12: Belt Load Statistics 61
General 61
(1=) Belt Loading 62
(2=) Print Stats Report 63
(3=) Reset Statistics 65
SECTION 13: Production Reports 66
Reports 66
Stored Report Configuration 67
SECTION 14: Routine Maintenance 69
Physical Maintenance 69
Load Cell 69
Special Tolerance Conveyor (Weigh Class) Idlers 70
Care Of Test Weight 70
Weigh Platform Alignment 71
Speed Sensor Drive Assembly 72
Lubrication Requirements 72
Electronic Maintenance 73
SECTION 15: EMBC-220 Software / Hardware Compatibility 74
SECTION 16: Software (Revision) List 75
SECTION 17: Drawings 90
System Description SECTION 1
System Description
General
The Process Automation EMBC-220 Belt Scale Systems are used to measure, and (optionally
to control), the mass throughput rate and total of material as transported on a conveyor
belt.
The mass of material on the belt is measured by a strain gauge type load cell which is
coupled to the conveyor system through a weigh frame, idlers and the conveyor belt.
The speed of the material is measured by a speed sensor (tachometer) which may be
coupled to the conveyor system via a speed sensor drive assembly or via any convenient
non-driven pulley. The precision Tachometer Drive Assembly (TDU) has been designed to
provide a rugged belt / tachometer interface while keeping speed measurement errors to a
minimum.
In the event that a speed sensor is not used, the load cell board has an input which can
allow connection of a digital input to act as a speed sensor replacement sensor. Whenever
the conveyor runs, the contact is closed. When the conveyor is stopped the contacts should
open.
In those applications where the inclination angle of the conveyor may change
(Stacker/Reclaimer applications), optional Boom Angle Compensation is possible using our
Model ICL-102 inclinometer.
The UNIPROIV transmitter performs the task of manipulating these measured variables in
order to obtain Rate, Total and other information.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 3
Component Descriptions SECTION 2
Component Descriptions
UNIPROIV Transmitter
The UNIPROIV transmitter is identical for all the scale models and must be mounted within
the cable distance of the load cell supplied.
UNIPROIV provides a single 4-20 mA (field programmable for either Feed Rate or PID
Control) output and single relay output (programmable to be either a Totaliser or Sampler
Control output). The max count rate from UNIPROIV totaliser is 90 000 per hour (25
counts/second). With a setting of 1 tonne per pulse this means a theoretical maximum of
90 000 t/h. The mass per pulse resolution is programmable. Kindly note that the width of
the totaliser pulses are not freely selectable. The higher the maximum count rate the
smaller must be the pulse width.
The totaliser (digital) output is in the form of either:
Standard - a volt free relay (dry contact) capable of switching up to 2A at 48 VDC
max. The lifespan of the relay is fixed (typically >5 million operations mechanical &
200 000 electrical - at full load.) The max switching rate for the relay output is 10
operations per second. However due to the lifetime limitations of the relay output, it
is recommended that the totaliser count rate be set not to exceed 1000 counts per
hour. At count rates higher than 1000 per hour, the lifetime of the relay will be less
than 2 years for 24 hour operation at full count rate. A good “rule of thumb” is never
use a mass per pulse setting less than 0.1% of the maximum hourly count rate. E.g.
For 1000 t/h, use a mass per pulse of 1000/1000 = 1 tonne. Likewise for 10 000 t/h
use 10000/1000 = 10 t per pulse. This will greatly extend the life of the totaliser
relay.
Optional - a solid state relay (SSR) capable of switching 250 mA at 24 V AC/DC (24
VAC max & 36 VDC max). The lifespan of the SSR is very much greater than that of a
mechanical relay (typically tens of millions of operations.). It is also capable of
switching much faster than a relay (<3 mS, ON and OFF times), and is able to switch
at the maximum rate that UNIPROIV can output the pulses.
A digital input (24 VDC) input is also provided for Speed Sensor Replacement. This allows
the Speed Sensor to be ignored (in the event a speed sensor is either not used; or in the
event the speed sensor is out of service for repair etc.).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 4
Component Descriptions SECTION 2
Weigh Frames
The Process Automation family of weigh frames are designed to cater for the wide
variations in application requirements. These range from the low cost single weigh idler
system through the super accurate multi weigh idler system viz.:-
EC-1 - single weigh idler weigh frame.
EC-2 - dual weigh idler weigh frame.
EC-3 - triple weigh idler weigh frame.
EC-4 - four weigh idler weigh frame.
EC-6 - six weigh idler weigh frame.
Easyweigh® - counter-balanced single weigh idler weigh frame.
Accuweigh® One - counter-balanced Single weigh idler weigh frame.
Accuweigh® Two - counter-balanced Dual weigh idler weigh frame.
Accuweigh® Three - counter-balanced Three weigh idler weigh frame.
Accuweigh® Four - counter-balanced Four weigh idler weigh frame.
Accuweigh® Six - counter-balanced Six weigh idler weigh frame.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 5
Component Descriptions SECTION 2
Load Cells
EMB-220 Belt Scale Systems employ high accuracy Strain Gauge Load Cells.
The load cells supplied in the Accuweigh® and Easyweigh® counter-balanced scale models
are fitted to a Load Cell Assembly which additionally provides protection to the load cell
against mechanical overload and mechanical shock. The load cells used are from stainless
steel.
The load cells supplied in the EC scale models are from Nickel plated mild steel.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 6
Component Descriptions SECTION 2
PT-500/PT-500LS
The model PT-500/PT-500LS are our premium model speed sensors, used where the best
possible accuracy is required.
The PT-500/PT-500LS speed sensors may be coupled, directly with a solid coupling, to any
(non-driven) pulley or similar drive mechanism, or it may be supplied with our proprietary
Drive Assembly (see below).
The PT-500/PT-500LS is supplied with a stainless steel shaft which is 150 mm long and 15
mm in diameter. The shaft length may be trimmed in length to suit the coupling used.
The PT-500 has an operational speed range of 25 to 2000 RPM.
The PT-500LS is a higher resolution device and is used for lower speed applications in the
range of 1.0 to 50 RPM.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 7
Component Descriptions SECTION 2
Tachometer Drive Assembly(s) (TDA) are available as an option to simplify installation of the
PT-500/PT-500LS speed sensors, into the conveyor system and to provide an accurate
conversion of linear conveyor speed to rotational velocity.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 8
Component Descriptions SECTION 2
PT-501/PT-501LS
The PT-501/PT-501LS speed sensor is a low cost alternative with integrated drive assembly
that allows bolting directly to the conveyor stringer (via stringer customer supplied
brackets).
This provides a lower cost solution of providing a speed sensor with a drive arrangement.
The PT-501 has an operational speed range of 25 to 2000 RPM. The PT-500LS is a higher
resolution device providing 15000 pulses per revolution for lower speed applications in the
range of 1.0 to 50 RPM.
Due to limitations of the power supply in UNIPROIV, the tachometer should be installed no
further than 100 metres from UNIPROIV. Longer distances are possible with an external
power supply for the tachometer.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 9
Component Descriptions SECTION 2
Where the inclination angle of an installed scale may be changed, compensation for this is
required. This is because the force applied load cell(s) in the weigh frame is dependent on
the conveyor inclination angle.
Our model ICL-102 is a precision inclinometer used to sense inclination angle of the
conveyor in order to provide Boom Angle Compensation.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 10
PTUII SECTION 3
The PTUII provides all the functionality of the model PTU, with the added feature of data
capturing.
The PTUII plugs into a quick connect (RJ type) socket on the 3000 CPU situated inside the
UNIPROIV enclosure.
The PTUII connection point may also be located remotely by making use of a Portable
Terminal Connection Unit (PTCU) located at a more convenient location.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 11
UNIPROIV Operation SECTION 4
UNIPROIV Operation
General
In order to provide a simple "man-machine interface", the EMBC-220 Belt Scale is
completely menu driven. For each operation (Set-up, Re-Cal, etc.,) a suite of messages
guides the user and provides instructions for the action required. The flow charts provided
will help one to understand the operation and all the functions available in the instrument.
The following conventions have been employed and are used throughout the gauge
software:
Auto-repeat:
An auto-repeat function for all keys is provided. The repeat action is accelerated the longer
the key is depressed.
Entering of variables:
If the instrument prompts for an entry of a variable the "ENT" key has to follow the input of
figures. The variable is otherwise discarded.
It is irrelevant where the cursor is situated within the variable; once the "ENT" key is
depressed the variable is accepted as displayed. The "DEL" key moves the cursor to the left
by one position and resets that character to zero.
For certain variables the entering of a value is restricted by a lower and upper limit. If an
attempt is made to exceed those limits the variable is discarded and the cursor is placed at
the beginning of the variable.
If a negative variable needs to be entered, use the +\- key which will toggle the sign.
Previous/next menu:
The software allows for the following:
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 12
UNIPROIV Operation SECTION 4
This convention is however not desirable in certain situations (e.g. time-out calibrations)
and hence in these cases not applied.
Selection procedure:
In a menu where one item out of several items has to be selected, the one currently
selected will flash.
ESC. Key:
The function of the "ESC" key serves to either abort a timed procedure (without alteration
to any calibration data) or to bring the programme back to the nearest sub-menu. It should,
however, be noted that any variable keyed-in under a specific display stays valid even if the
"ESC" key is used to continue.
Simultaneous depressing of all the odd keys on the numeric keypad will result in a Master
Reset (the gauge is re-initialised and all calibration constants are reset to their default
values and all totalises are reset to zero).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 13
UNIPROIV Operation SECTION 4
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 14
UNIPROIV Operation SECTION 4
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 15
UNIPROIV Operation SECTION 4
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 16
UNIPROIV Operation SECTION 4
Msg. 220 (Moisture compensation info - only available when the transmitter enclosure door
interlock is either not installed or when it is installed and the transmitter enclosure
door is open)
F2=MOISTURE: Depressing F2 allows then keypad entry of the moisture per cent
to be used (wet basis). This data is used by the transmitter in
order to compute Dry mass Loading, Rate and Total.
F1 = TOGGLE: Depressing F1 toggles the display between the Normal Operating
display screens of Scale Information (msgs. 1 & 2), Rate Control
(msg. 151), Secondary Control (msg 305), Moisture Correction
(msg. 220) and Boom Angle Compensation (msg. 241). Note:
Messages 305, 220 and 241 will only be available if and when
these features are enabled in SET-UP).
F4 = MENU: Selecting F4 from either of the above screens will call up the main
menu (msg. 3).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 17
UNIPROIV Operation SECTION 4
Msg 241 (Boom Angle Compensation - only available when the transmitter enclosure door
interlock is either not installed or when it is installed and the transmitter enclosure
door is open)
ANGLE: This displays the inclination angle of the conveyor in degrees. This
angle information is obtained from the model ICL-102 angle
transmitter.
FACTOR: This is the factor that needs be applied to the gauge calibration in
order to compensate for the inclination of the conveyor.
F1 = TOGGLE: Depressing F1 toggles the display between the Normal Operating
display screens of Scale Information (msgs. 1 & 2), Rate Control
(msg. 151), Secondary Control (msg 305), Moisture Correction
(msg. 220) and Boom Angle Compensation (msg. 241). Note:
Messages 305, 220 and 241 will only be available if and when
these features are enabled in SET-UP).
F4 = MENU: Selecting F4 from either of the above screens will call up the main
menu (msg. 3).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 18
UNIPROIV Operation SECTION 4
4 = Serial Nos: Selecting “4” results in the display of a message indicating “Busy
Printing” and printing of the equipment Serial numbers will
commence. Printing may be aborted by pressing the ESC key.
Note: Calibration, Application and other data, is stored in a RAM
IC. Since the NVRAM (battery backed up RAM) is another
Integrated circuit, a physical movement must be made to copy the
data from RAM to NVRAM. However, owing to limitations of the
memory backed up by battery, the Serial Numbers and
Application Data are not able to be stored in battery backed-up
RAM. Instead this information is now stored in EEPROM. This
means that after entering/editing Serial Numbers and Application
Data, the user should Store them to EEPROM. If the UNIPROIV loses
power before the data is saved, it will be lost. If the user wants to
view or Print-out the Serial Numbers or Application Data, it should
always be recalled from EEPROM first. The Store and Recall
options are available in the same menus which allow their entry.
5 = App. Data: Selecting “5” results in the display of a message indicating “Busy
Printing” and printing of the details of the application will
commence. Printing may be aborted by pressing the ESC key.
Note: Calibration, Application and other data, is stored in a RAM
IC. Since the NVRAM (battery backed up RAM) is another
Integrated circuit, a physical movement must be made to copy the
data from RAM to NVRAM. However, owing to limitations of the
memory backed up by battery, the Serial Numbers and
Application Data are not able to be stored in battery backed-up
RAM. Instead this information is now stored in EEPROM. This
means that after entering/editing Serial Numbers and Application
Data, the user should Store them to EEPROM. If the UNIPROIV loses
power before the data is saved, it will be lost. If the user wants to
view or Print-out the Serial Numbers or Application Data, it should
always be recalled from EEPROM first. The Store and Recall
options are available in the same menus which allow their entry.
6 = Belt Load Stats: Selecting “6” results in the display of a message indicating
“Busy Printing” and printing of a Belt Load Statistics report.
Printing may be aborted by pressing the ESC key.
7 = I/O Map: Selecting “7” results in the display of a message indicating “Busy
Printing ” and printing of I/O Allocation on the expanded I/O will
commence. The I/O allocation defines how the UNIPROIV has been
configured - which physical Input or Output is connected to which
Input or Output on UNIPROIV. The transmitter has a maximum of
32 slots available (#0 to #31). Each carrier board has provision for
4 slots. These slots are always numbered 1 to 4. The absolute
(computer identified) slot numbers are offset by the carrier board
address. E.g. Carrier board with address “0” will contain absolute
slot #’s 1 through 4, carrier board with address “1” will contain
absolute slot #’s 5 through 8, carrier board with address “4” will
contain absolute slot #’s 17 through 20 etc. Printed behind the
absolute slot number will be the UNIPROIV variable assigned to
that particular output. Note: While it may be prudent to have any
carrier boards installed set so as to have their addresses start from
“0”, this is not essential.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 19
UNIPROIV Operation SECTION 4
8 = Angle Comp: Selecting “8” results in the display of a message indicating “Busy
Printing” and printing of Angle Compensation calibration info.
9 = Print All: Selecting “9” results in the display of a message indicating “Busy
Printing” and the sequential printing of all of the printouts with
the exception of the stored reports.
Msg 3
Depressing the F4 key from the main display (msg. 1) brings up the MAIN MENU
display screen (msg. 3) where further options are presented as follows:-
1=Setup: Depressing “1” brings up msg. 4 prompting for the SET-UP1 access
code.
2=Cal Checks: Depressing “2” brings up msg. 5 prompting for the access code for
CALIBRATION CHECKS2.
3=Special Funct: Depressing “3” brings up msg. 6 prompting for the access code for
SPECIAL FUNCTIONS1.
4=Cal Const: Depressing “4” brings up msg. 7 prompting for the access code for
CALIBRATION CONSTANTS3.
5=Control: Depressing “5” brings up msg. 150 prompting for the access code
for SET POINT CONTROL1.
6=Statistics: Depressing “6” brings up msg. 280, the sub menu for BELT LOAD
STATISTICS (access code not required).
Notes:
1. Some menu items are not accessible when the door interlock is active. MENU ACCESS
DENIED – THE DOOR INTERLOCK ACTIVE is indicated when access is attempted.
2. Reduced functionality access is available to Calibration Checks when the door interlock is
active.
3. Calibration Constants can be viewed (but not entered) when the door interlock is active.
4. Messages 3, 4, 5, 6, 7, 150 and 280 time-out. If the gauge is left in any one of these
display screens for more than thirty seconds it will revert to the next (previous) higher
display. If this higher display too has a time-out, this will be repeated and so on. Once
the gauge has accepted an access code the time-out will be de-activated.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 20
Access Codes SECTION 5
Access Codes
UNIPROIV is protected by coded access. The following is a list of codes which are required in
order to access the gauge.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 21
Configuration & Calibration - UNIPROIV SECTION 6
Application Details
Prior to placing the system in operation, the design criteria submitted by the purchaser at
the time of order and the actual application details must be verified. These measurements,
adjustments and calculations should require attention only at initial start-up; when
troubleshooting; or whenever any of the following conditions should change:
Measurement (calibration) Range.
Belt on which the gauge is mounted.
The belt speed operating range is changed.
The belt loading operating range is changed.
Tachometer drive wheel is changed.
Change in I/O requirements.
The design criteria (Application Data) to which the system was originally designed is
available from the transmitter (see Report/Printout options msg. 125).
Test Weight
When equipped with an (optional) test weight, the weight sensor calibration may be quickly
and easily checked.
The test weight is provided as a means for “locking in” the calibration, once the scale has
been accurately calibrated (using tests involving bulk material).
The Test Weight may be operated by either;
man-handling or,
manually via lever mechanism for larger test weights or;
automatically using a single test weight and single electric motor driven
actuator.
The test weight must be positioned in one of two (or three in special cases) positions.
TARE position: On counter-balanced scale types, this is the normal or
operating position for the test weight as the test weight is
used as part of the counter-balance mass. For non-counter-
balanced scales the Tare position is with all the test weights
removed.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 22
Configuration & Calibration - UNIPROIV SECTION 6
SPAN position: In this position the test weight represents a specific belt
loading which is used as a reference for calibration checks.
The value of the belt loading represented by this test weight
may be determined in one of two ways.
A theoretical determination of the representative belt loading
is possible using the physical details of the scale and weight of
the test weight, as detailed in the Section covering Calibration
Calculations.
After the scale has been accurately calibrated using a bulk
material test, the test weight may be weighed (measured) by
the scale. This measured value of the test weight may then be
stored for future reference by the scale.
Calibration
Calibration entails the matching (scaling) of the measured & computed information from the
primary sensors, to that of the actual value of the measured variable. In most of the
calibration procedures the gauge samples the incoming signal for a period and then requires
that the user inform the gauge what value this input signal represents. The user is guided
through all the calibration processes with simple menu type displays. Belt scales normally
have two sensors namely Weight and Speed.
Weigh Frames
The weight sensor assembly has been engineered for the application as specified at the time
of order. This includes the computer optimisation of the weigh frame arrangement, load
cell capacities etc.
Many different weigh frame models are available - from the single weigh idler type, through
counter balanced types with up to six weighing idlers.
The UNIPROIV transmitter is suitable for use on all these scale models.
The integrated load cell amplifier requires no gain jumpers or switches to scale the amplifier
output. The gain of the amplifier is automatically set under software by the UNIPROIV
transmitter, during calibration.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 23
Configuration & Calibration - UNIPROIV SECTION 6
Note: The Function keys "F3" and "F4" will have the following meaning throughout the Set-
up programme: F3 will take one to the menu previous to the one currently being
displayed; F4 will take one to the next menu to the one being displayed.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 24
Configuration & Calibration - UNIPROIV SECTION 6
(2=) Resolution
Selecting “2” from the Set-up menu, guides the operator through the same message/input
screens as above in Full Calibration with the difference being that after completion it will
return to the SETUP menu whereas in Full Calibration, it will continue with the Set-up of the
Mass & Speed sensors as well as the Re-cal Set-up as listed below. The full calibration
involves the configuration & selection of variables essential for the gauge's proper operation
viz.:
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 25
Configuration & Calibration - UNIPROIV SECTION 6
Selecting “3” from the Set-up menu, brings up the Speed Sensor Calibration menu (msg. 24)
which in turn allows for the calibration of the speed sensor at two known points
(1=) A zero Calibration for a stationary belt and;
(2=) A high point Calibration; either at full speed of the belt or, at some constant speed
close to maximum speed.
Notes: 1) The transmitter will allow a 5 time constant delay before starting
measurements, in order to ensure the measured variables have settled to
their final values.
2) The filter time constant for the (raw) input from the speed sensor has a
default setting of 2 seconds. This may be altered under Special Functions.
Msg. 25 Speed Sensor Zero Calibration. The gauge will determine the input counts
from the tachometer which represent a stationary belt.
Note: The counts from a stationary belt should normally be zero.
Msg. 26 & 27 Speed Sensor Span Calibration. This entails the measurement, by the
transmitter, of the input counts from the speed sensor and assigning to it of the value of the
actual belt speed as determined by (simultaneous) accurate measurement. Actual belt
speed may be measured by measuring the distance travelled by the belt in
3 minutes. The distance travelled may be measured by an accurate Odometer (available
from Process Automation).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 26
Configuration & Calibration - UNIPROIV SECTION 6
Note: 1) The transmitter will allow a 5 time constant delay before starting
measurements, in order to ensure the measured variables have settled to their
final values.
2) The filter time constant for the (raw) input from the weight sensor has a default
setting of 2 seconds. This may be altered under Special Functions.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 27
Configuration & Calibration - UNIPROIV SECTION 6
Should the scale be set-up to allow for Automated Calibration checking via an Electrically
actuated Test weight Lifting mechanism (see msg. 21 from selection “8” under the main
Set-Up menu), then;-
Msg. 82 Allows the further selection/confirmation of whether an
Automated electrically actuated test weight operation or a
Manual operation of the test weight is required. Selecting “1” will
bring msg. 83 to allow initiation of the sequence using the ENT
key. Selecting “2” will allow initiation of the sequence using the
ENT key - after manually positioning the Test Weight - where after
msg. 32 will displayed (see below).
Msg. 83 Ensure the conveyor is empty and clean. Depress the ENT key to
initiate the Span Checking.
Msg. 85 Displays the Status of the automated test (measured count and
belt length left to go until completion of the test).
Msg. 32 Allows the (span) calibration of the weight sensor to the On Board
Test Weight. The measurement made in msg. 85 above, is made
over the Calibration Belt Length (normally complete revolution(s)
of the conveyor). This msg. (msg. 32) will prompt for the input of
the belt loading represented by the Test Weight (kg per metre).
This is the value UNIPROIV will assign the average of its
measurement to. The (theoretical) value of lineal belt loading
represented by the Test Weight may be determined from the
force calculations applicable to the weigh frame type/model.
These force calculations are shown on the General Arrangement
Drawing of the Weigh Frame.
Should the scale not be set-up to allow for Automated Calibration checking via an
Electrically actuated Test weight Lifting mechanism (see msg. 21 from selection “8” under
the main Set-Up menu), then;-
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 28
Configuration & Calibration - UNIPROIV SECTION 6
Msgs. 31 & 32 Allows the (Span) calibration of the weight sensor to the manually
operated On Board Test Weight. The measurement is made over
the entire calibration belt length (normally a complete revolution
of the conveyor), after which time it will prompt for the input of
the belt loading represented by the Test Weight (kg per metre).
This is the value UNIPROIV will assign the average of its
measurement to. The value of lineal belt loading represented by
the Test Weight may be determined from the force calculations
applicable to the weigh frame type/model. These force
calculations are shown on the General Arrangement Drawing of
the Weigh Frame.
Example 1: Where it is only possible to perform the calibration by using the "Inverse Belt-
Cut Method", the following procedure is required.
Firstly perform a "static" zero calibration on the clean section of belt in the
measurement area.
Next place a known pre-weighed amount of material (in the range 50-80 % of
the calibrated belt loading) as evenly as possible onto the belt, between the +1
and -1 idlers (one idler before, to one idler after, the weigh frame). Ensure that
the material is evenly distributed over this area. For example 90 kg spread
over 3 metres, for a single weigh idler weigh frame, giving a belt loading of 30
kg/m. Follow the prompts on the display to start and stop the measurement
and enter the value of the simulated load (e.g. 30, 0 kg/m).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 29
Configuration & Calibration - UNIPROIV SECTION 6
Example 2: Where it is only possible to perform the calibration by using the "Belt-Cut
Method", the following procedure is required.
With the belt running loaded at a uniform belt loading, of at least 50% of the
calibrated belt loading, stop the conveyor and note the belt loading displayed
by the transmitter. Now remove and accurately weigh the material from the
belt between the +1 and -1 idlers (one idler before, to one idler after, the
weigh frame). For example 90 kg over 3 metres, for a single weigh idler weigh
frame, gives a belt loading of 30 kg/m. Now note the belt loading displayed by
the transmitter after removal of the material, for this empty belt. Subtract the
empty indication from the loaded indication to obtain a nett indicated value of
belt load. Compare this nett indicated value against the actual (measured)
value and compute the error/correction required.
Now perform a "dynamic" zero calibration on the belt.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 30
Configuration & Calibration - UNIPROIV SECTION 6
Notes:
1) IMPORTANT: After any weight sensor calibration in which a static zero
calibration has been performed, it is essential to perform a “weight sensor
zero check" with correction (messages 42, 43 and 44) - from the "Calibration
Check” Menu (message 5), since the proper tare (weight sensor zero) of the
belt can only be obtained when performing a measurement over one
complete revolution of the belt. It is not legitimate to perform a dynamic
zero Set up calibration after a static mass calibration, since this will result in
a new (incorrect) calibration curve being generated!
2) During the calibration period, the input from the weight Sensor is
continuously displayed. This allows for verification that the signal from the
measuring assembly is arriving at, and being seen, by UNIPROIV.
3) In all the cases above it follows that a "Re-Calibration Set up" (message 39 &
40 - see below) must be performed after completing the weight sensor
calibrations.
In the event the counts from the mass sensor is measured or calculated to exceed 65000 at
maximum loading, a warning will be displayed- msg. 251
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 31
Configuration & Calibration - UNIPROIV SECTION 6
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 32
Configuration & Calibration - UNIPROIV SECTION 6
Digital Inputs
Test Weight Normal position Feedback
Test Weight Calibrate position Feedback
*Start electrically operated Test Weight calibrate sequence
*Abort electrically operated Test Weight calibrate sequence
Conveyor Drive running feedback for the Rate controller.
Auxiliary Inputs 1 and 2.
Rate controller Set-point operation selection - Local/Remote
Drive running feedback for the secondary controller.
Secondary controller Set-point operation selection - Local/Remote
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 33
Configuration & Calibration - UNIPROIV SECTION 6
Analogue Inputs
Moisture
Remote Set-point for the Rate controller
Conveyor Inclination (Tilt) angle.
Auxiliary Inputs 1 and 2
External PV input for the Secondary controller
Remote Set-point for the Secondary controller
Analogue Outputs
Belt load
Belt speed
Feed rate
Rate PID controller Control output
Auxiliary Outputs 1 and 2
Secondary PID controller Control output
Notes: 1) The Digital Inputs marked with an asterisk (*) require a momentary
(approx. 500 mS) application of the input. All other Digital Inputs are
effective during their application only.
2) The Digital Outputs are all de-energised on power-up, when they are not
allocated to a variable. If allocated to a variable they will assume an
appropriate condition to match the status of the variable.
3) The Analogue Outputs are set to zero (mA) on power-up, when they are
not allocated to a variable. If allocated to a variable they will assume an
appropriate condition to match the status of the variable.
Selecting “6” from the SETUP menu (msg. 8) brings up msg. 157. This screen allows the
configuration of the I/O used on the I2C expansion bus.
In addition to being able to select the I/O, this screen provides diagnostic information of
absolute slot position (T. IND) and internal card identification type (TYPE). The T.IND is used
to identify where a particular card is situated in the possible address range of 0 through 31.
Depressing F1 scrolls through all the locations 0 through 31. The card type is an internal
representation of the type of card used where 0=None, 1=Analogue Output, 2= Digital
Output, 4=Digital Input and 8 =Analogue Input. The number returned for any particular
location (0-31) will indicate the card type located in this position.
Depressing 1, 2, 3 or 4 will select messages 158, 159, 160 or 161 to configure the Digital
Input, Digital Output, Analogue Input or Analogue Output respectively. The first card of the
type selected will be displayed initially - showing the carrier # and slot # on that carrier.
Depressing F1 in these screens will scroll through all the cards (position) where these cards
have been found by the system. Keys F2/F3 allows the scrolling through all the possible
variables which may be associated with this type of card. A variable (other than none) will
be automatically assigned when depressing the F1 key to move to the following card
location.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 34
Configuration & Calibration - UNIPROIV SECTION 6
The state of Digital I/O is also displayed (ON or OFF). The digitised values of Analogue Inputs
are displayed and should be in the range 200 to 4000 for input signals in the range 4-20 mA.
Analogue Output values are displayed in per cent (%).
(7=) Set-Up OF 4 - 20 mA – Expanded I/O
Selection of “7” from the SET-UP MENU displays msg. 232.
Msg. 232 Allows the selection of (1=) Input(s) and (2=) Output(s). Selecting
“1” takes one sequentially through msgs. 233 through 236.
Selecting “2” takes one to msg. 162.
Msg. 233 Displays the first Analogue Input found by the system, indicating
the carrier number and slot number as well as the (arbitrarily
numbered) input number. The variable assigned to this Input is
also displayed. F1 scrolls between the different Inputs to be
spanned. Depressing ENT takes one to the calibrations.
Msg. 234 The external mA source must now be connected to the Input and
forced to the value equal to the particular variables’ “Bottom of
Scale” condition viz: if the variable associated to the Input is
Moisture covering the range 5% - 20%, then the mA proportional
to 5% must be input at this time. Depressing the ENT key starts
the process of averaging on this input.
Msg. 236 Again displays the carrier number, slot number as well as the
(arbitrarily numbered) input number. The variable assigned to this
Input is also displayed. Now, additionally, the (instantaneous)
count value proportional to the input is also displayed. Depressing
the ENT key stops the process of averaging on this input, and the
gauge automatically assigns this average value to the Bottom of
Scale number e.g. in msg. 217 for Moisture %.
Msg. 235 The external mA source (connected to the Input) must be forced
to the value equal to the particular variables’ “Top of Scale”
condition. viz. If the variable associated to the Input is Moisture
covering the range 5% - 20%, then the mA proportional to 20%
must be input at this time. Depressing the ENT key starts the
process of averaging on this input.
Msg. 236 Again displays the carrier number, slot number as well as the
(arbitrarily numbered) input number. The variable assigned to this
Input is also displayed. Now, additionally, the (instantaneous)
count value proportional to the input is also displayed. Depressing
the ENT key stops the process of averaging on this input, and the
gauge automatically assigns this average value to the Top of Scale
number e.g. as in msg. 217 for Moisture %.
Selection of “2” from the EXPANDED mA SETUP - msg. 232 - brings up msg. 162 which allows
the spanning of the 4 - 20 mA signals being output via the I2C expansion bus.
Msg 162 Displays the first Analogue Output found by the system, indicating
the carrier number and slot number. F4 toggles between the 4mA
and 20 mA condition, while the F2/F3 keys increase/decrease the
output. Connect a milli-Ampere meter across the output terminals
for the appropriate card (ensuring no cables are connected
thereto) and select (F4) the condition for 4 mA output. Use the
F2/F3 keys to adjust this value to as close to 4 mA as is possible.
Depress F4 to select the 20 mA condition and depress the F2/F3
keys to adjust this value to as close to 20 mA as is possible.
Depress F1 to go to the next analogue output and repeat this
procedure until all the outputs have been spanned.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 35
Configuration & Calibration - UNIPROIV SECTION 6
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 36
Configuration & Calibration - UNIPROIV SECTION 6
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 37
Special Functions SECTION 7
Special Functions
All the functions are self-explanatory and partially duplicated from the Initial Set-up. They
are provided separately since quick access to these adjustments is often desirable. Certain
functions also assist in diagnostics.
Hereunder is a list of the current functions with a brief explanation of their operation/uses.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 39
Special Functions SECTION 7
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 40
Special Functions SECTION 7
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 41
Special Functions SECTION 7
(iv) The digital output on the 3000 Load Cell Board (a sampler output
may be configured additionally, via the expansion I/O, should this
output on the 3000 LC board be used for totaliser pulses) may be
configured to provide a sampler control output in place of the
normal totaliser output, should it be required. Depressing “4” from
this menu will toggle allocation of this output between Totaliser and
Sampler. The flashing identifier will indicate the current allocation. In
the event this digital output on the load cell board is used to control
a sampler it must still be turned “ON” under Special Functions #8.
Notes:
1)While the expanded I/O may be used to drive a sampler, only the
digital output on the Load Cell Board may be used to drive a remote
totaliser (due to the pulse width control requirements).
2)It is required that the Mass Sample Interval be greater than the
amount of material between the scale and the sampler else no
output will ever occur. This limits the downstream distance that the
sampler can be located from the scale. i.e. distance between scale &
sampler [m] < Mass Sample Interval [kg] / maximum belt loading
[kg/m].
3)The sampler must always be located after (downstream of) the
scale.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 42
Special Functions SECTION 7
Note: The UNIPROIV remote totaliser is an intelligent totaliser in that it will only pass on to
any remote totaliser, positive tonnage. i.e. should the belt tare be that the scale counts
down (subtracts) tonnage from the local totaliser under empty belt conditions, UNIPROIV
will only pass on a count to the remote totaliser once UNIPROIV has made up for the
amount previously subtracted. For example, assume the belt runs empty for some time and
results in the local tonnage counter removing 2 tonnes. When material next passes over the
scale, it will require that 2 tonnes pass over the scale before it is in a position to pass on
pulses for remote totalisation. This feature allows the use of inexpensive "add only" type
remote counters and saves on cable cost - only 2 conductors are required, and will
additionally ensure the local and remote totalisers remain synchronised.
Likewise, if the remote totaliser pulse width is set too long for the count rate (tonnes per
hour and totaliser resolution), the remote totaliser will appear to lag behind the local
totaliser. The remote totaliser will continue counting (pulsing) even with little or no material
flow) until the remote totaliser equals the local totaliser. It is important to take care when
configuring the totaliser resolution and the remote totaliser pulse width, to avoid the
remote totaliser lagging, since theoretically the buffer for the remote totaliser could
overflow resulting in a mismatch between the local and remote totalisers.
It should also be noted that the max count rate available from the UNIPROIV is 25 pulses per
second when the minimum pulse width of 20ms is set (a maximum of 90000 pulses per
hour). A 50% duty cycle is enforced on these pulses - the minimum OFF time is the same as
the ON time. For count rates > 1000 per hour, the optional Solid State Relay output should
be used in place of the dry contact relay. Care should be taken when selecting the tonnes
per pulse so as to be a practical compromise between relay lifetime and required resolution.
The pulse width setting is also important when connecting to other devices such as PLC etc.
since at high count rates the PLC may miss pulses due to restrictions on the scan
time/window of the PLC. For highest integrity, totalisers are best read using a field bus
connection.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 43
Special Functions SECTION 7
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 44
Special Functions SECTION 7
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 45
Special Functions SECTION 7
Notes: 1) The use of Sampler Control will require the use of a digital output. In the event the
digital output on the 3000 Load Cell Board is used for a remote totaliser, it will
necessitate the use of the expansion I/O and the allocation of this variable to a
particular Digital Output card – see option #6 in the SET-UP MENU. When using the
Expanded I/O the duration of the output pulse is fixed at 500 milliseconds.
2) It is required that the Mass Sample Interval be greater than the amount of material
between the scale and the sampler else no output will ever occur. This limits the
downstream distance that the sampler can be located from the scale. i.e. distance
between scale & sampler [m] < Mass Sample Interval [kg] / maximum belt loading
[kg/m].
3) The sampler must always be located after (downstream of) the scale.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 46
Low Temperature Applications SECTION 8
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 47
Calibration Checks SECTION 9
Calibration Checks
There is a correlation between frequency of calibration, number of idlers on the scale weigh
bridge and the accuracy required. While it may be possible to achieve a reasonably high
degree of accuracy for a short time with a single idler scale in a good installation, it is
obvious that the single idler scale is more susceptible to belt effects and changes in
alignment and therefore the calibration should be checked more frequently. Likewise, a
multiple idler weighbridge will not have to be checked frequently if 1% is all that is required.
A common mistake made in the selection of belt scales is choosing a single idler scale or
short weigh bridge because the accuracy required is only 1%. It may be possible to achieve
1% but it might require calibrating once a day!
The frequency of calibration is best determined from actual operating experience. Initially
the tests should be performed more frequently (once a week) to establish a confidence level
based on accuracy required and the number of times an adjustment had to be made. The
tests should be based on the As-Found-Error (AFE).
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 48
Calibration Checks SECTION 9
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 49
Calibration Checks SECTION 9
Note: If the average belt loading is not close to zero first check for a mechanical
cause i.e. stones, damaged idlers etc. before making any electronic
corrections (tare).
The test weight may now be located in the "SPAN" position, and the average
belt loading noted. Should this be close to the figure on the calibration data
sheet then all is well with the scale and nothing further needs to be done.
Note: If the average belt loading indicated is more than 2% different from the
figure entered on the calibration sheet, do not proceed until the cause for
this has been established i.e. check for stones, mechanical obstruction,
damaged idler sets, mechanical failure etc.
Replace the test weight in the "ZERO" position.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 50
Calibration Checks SECTION 9
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 51
Calibration Checks SECTION 9
A facility (5=) is provided as an aid in the performance of a bulk material test on the belt
scale. Function (6=) allows for the correction to the calibration of the belt scale weight
sensor, if found to be in error.
Note: The automatic calibration correction is an approximation that is correct providing the
belt scale As Found Tare Error is zero. If this tare error is not zero, an error in the
application of the calibration correction will occur. This is due to the nature of the
mathematics involved and may be ignored for zero offsets (tare errors) of less than 5
%. Should there be doubt as to the conditions at hand this may be resolved by
performing a weight sensor zero calibration. This will effectively reset the tare error
to zero, thereby removing any possible problems in this regard.
This selection will result in the transmitter performing a Tare (mass zero) check of the
weight sensor over the defined (calibration) belt length.
Ensure any Test Weight is in the "TARE" (or normal) position. Run the belt conveyor empty
of all material prior to initiating this function.
This feature is very powerful since it will display the value of the average zero condition as
measured (in "signed" kg/m and Per cent Error - referenced to the Full Scale calibration of
the belt loading). It is possible to automatically correct the weight sensor zero calibration
(auto tare) if this is desired. Any correction thus entered will be stored as a "zero (weight)
offset" and may be read from the list of calibration constants. For mathematical reasons,
this zero offset should not be allowed to exceed the value of 5%. Should this figure of 5% or
greater be attained a mass sensor zero calibration should be effected (which will reset this
figure to zero).
The gauge may be instructed to correct any discrepancies in the indication if desired (Auto
Correction).
Should the Tare Error be found to exceed the pre-set allowable value (as defined in
Recalibration Set-up msg. 263), no correction will be permitted (msgs. 264 & 265). The
problem must be corrected and the check repeated. Note: It is possible that the Test Weight
may have been left in the Span position!
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 52
Calibration Checks SECTION 9
This selection will result in the transmitter performing a weight Sensor Span Check using the
Test weight (as previously set-up under RE-CALIBRATION SET-UP). Measurement is made
over the Calibration Belt Length and totalisation will not occur.
Operation will depend on whether the scale has been configured for Automated Test
Weight operation or not. If AUTOMATED TESTING is invoked it will result in msg. 66 being
displayed.
Msg. 66 Will prompt for the selection of automated or manual operation
of the Test Weight. Selection “2” will result in msgs. 45 & 46 being
displayed – as if Automated operation had not been selected.
Selecting “1” from this screen will bring up msg. 84.
Msg. 84 This prompts for the inspection to verify that the conveyor is
empty and clean. Depressing ENT will initiate the test and bring up
msg. 85.
Msg. 85 This screen displays the status of the test, with a continuous
indication of the instantaneous weight measurement (count) and
the belt length remaining for the test. After the measurement has
been completed msg. 46 will be displayed.
Msg. 46 The average loading represented by the Test Weight will be
displayed together with the error % from the stored value and the
gauge will now prompt whether automatic correction, to the
gauge weight calibration, is required or not.
Should the Span Error be found to exceed the pre-set allowable value (as defined in
Recalibration Set-up msg. 263), no correction will be permitted (msgs. 266 & 267). The
problem must be corrected and the check repeated.
If AUTOMATED TESTING is not invoked it will result in msg. 45 being displayed.
Msg. 45 Ensure the belt is running empty end clean. Place the Test Weight
in the Span position. Allow sufficient time for the transmitter
indication to stabilise and depress ENT to start the measurement.
The instantaneous weight measurement (count) and the belt
length remaining for the test will be displayed. After the
measurement has been completed msg. 46 will be displayed.
Msg. 46 The average loading represented by the Test Weight will be
displayed together with the error % from the stored value. This is
a powerful feature since it will display the average value of the
Test Weight (in "signed" kg/m and Per cent Error - referenced to
the Full Scale calibration of the belt load). The transmitter may be
instructed to automatically correct the scale, weight calibration,
for any discrepancies in the indication, if desired (Auto
Correction). Hereafter the display will revert to msg. 42 the
CALIBRATION CHECKS Menu.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 53
Calibration Checks SECTION 9
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 54
Calibration Checks SECTION 9
This extremely useful facility is available to assist with the weight sensor span calibration,
using the bulk (material) test method.
The gauge will prompt for a weight sensor Tare Check/adjustment and belt Speed (zero and
span) check/adjustment prior to any bulk test, since any error found from the bulk test will
be treated as a weight Sensor span calibration error only.
It is possible to select whether the material used for the test, is to be added to the normal
(operating) totaliser or not.
It is also possible to configure UNIPROIV to always make measurements over a complete
revolution of the belt. Since the Tare is determined over an exact number of belt
revolutions, theoretically the Span should be checked over this same length. When enabled,
UNIPROIV will indicate the belt length still to be travelled before the completion of a bulk
test (the totaliser will continue to operate until the belt has completed a whole number of
revolutions).
After the bulk test is completed it is possible to correct the Weight Sensor Span calibration
using option 6 (= Calibration Corrections).
Note: - In the event that the scale is operated with Moisture Correction invoked, this
Moisture Correction function will be automatically disabled during the performance of
a bulk test i.e. moisture correction will be turned off!
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 55
Calibration Checks SECTION 9
Note: Should the door interlock be installed (Assize Applications) no Weight Span
Calibrations are permitted.
UNIPROIV will prompt for two figures viz. Those "indicated" and/or measured by UNIPROIV
and the "actual" mass as determined from a weigh bridge or similar high accuracy check-
weigh system (which has, in turn, had its calibration verified no later than 2 months
previously).
If the belt scale was tested by a bulk material test (menu 5) immediately prior to this
correction, the figure as totalled during the test, will be recalled for convenience (it will not
be necessary to key these figures in again). Should this not be the case, simply key in the
applicable figures.
All that remains is to key in the Actual Mass and to request the gauge to apply the necessary
calibration corrections to remove any error. Alternatively production and figures from the
scale may be keyed in to provide a calibration correction.
The transmitter will also automatically adjust the Test (Reference) Weight, as used in the
weight sensor RE-CALIBRATION SET-UP, for any Weight Sensor Span correction applied.
Note: The automatic calibration correction is an approximation that is correct providing the
belt scale "As Found Tare Error" is zero. If this tare error is not zero an error in the
application of the calibration correction will occur. This is due to the nature of the
mathematics involved but may be ignored for zero offsets (tare errors) less than 5 %.
Should there be doubt as to the situation on hand this may be resolved by performing
a Weight Sensor Zero Calibration under the "Scale Set up Menu". This will effectively
reset the tare error to zero, thereby removing any possible problems in this regard.
As-Found-Error (AFE)
The As-Found-Error calibration check provides a means to determine the error in the scale
prior to making any adjustment. UNIPROIV not only provides for the automatic
determination of the overall “As-Found-Error”, but also allows the further determination of
what the components of this “As-Found-Error” are viz; Mass Zero, Mass Span, Speed Zero
and Speed Span.
.
NOTE: FOR ISO 9000/14000 AND ON HIGH ACCURACY INSTALLATIONS, A LOG BOOK IS
MANDATORY.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 56
Scale Log Book SECTION 10
Date Allowable As Speed Speed Speed Speed Load Load Zero Load Load Span Remarks
Error (%) Found Zero Zero Span Span Zero correction Span correction
Error Error correction Error correction Error performe Error performed
(%) (%) performe (%) performe (%) d (Y/N or (%) (Y/N or
d (Y/N or d (Y/N or value) value)
value) value)
_______________________________________________________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 57
Calibration & Application Data SECTION 11
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 58
Calibration & Application Data SECTION 11
Calibration Constants
The calibration data print-out provides a hard copy of all the parameters required to
properly set-up the gauge. This data may be filed with the Scale Log for use as an aid for
maintenance or to comply with the requirements of ISO 9000/2, which requires certificates
calibration.
Application Data
The technical specification and Serial Numbers of all the equipment as originally shipped,
together with the Application Details.
Should there be a need to order spare parts or should there be a need to alter the
application/ calibration, a copy of this print out should be faxed to the factory. This will
ensure compatibility of spares and make the verification of range/application changes
greatly simplified.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 59
Calibration & Application Data SECTION 11
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 60
Belt Load Statistics SECTION 12
General
Selecting option (6=) STATISTICS from the main menu (access code not required) will result
in the Statistics menu (msg. 280) being presented. The STATISTICS menu offers the
following options:-
1 = Belt Loading.
2 = Print Statistics.
Note: msg. 280 is subject to a time-out of 30 seconds. If an option is not selected during
this period the display will automatically revert to the previous display.
Note: Function keys "F3" and "F4" have the following meaning throughout the Statistics
programme: F3 will take one to the message previous to the one currently being
displayed; F4 will take one to the next message to the one being displayed.
_______________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 61
Belt Load Statistics SECTION 12
_______________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 62
Belt Load Statistics SECTION 12
TAG:CONVEYOR K1-3
PRINTED:12/04/08 11:10 The date and time that this printout was made. The
format is DD/MM/YY HH:MM.
STARTED:11/04/08 11:03 The date and time that the statistical counters were
restarted.
ELAPSED: 24:06:40 The time in HHHHH: MM:SS format that has elapsed
since the statistical counters were restarted (power
to UNIPRO).
BELT RUNNING: The time that the belt speed was greater than 0% of
the speed range. Also expressed as a percentage of
11:53:12 49.3%
the elapsed time.
BELT RUNNING EMPTY: The time that the belt was running with the loading
less than the Totalisation Cut-off level. Expressed as
2:39:02 33.3%
a percentage of the belt running time.
BELT RUNNING LOADED: The time that the belt was running with the loading
more than the Totalisation Cut-off level. Expressed
9:14:10 77.7%
as a percentage of the belt running time.
_______________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 63
Belt Load Statistics SECTION 12
80-90 : 82.1% This example indicates that the loading was in the
range 80% to 90% for 82.1 percent of the time that
the belt was running loaded.
70-80 : 0.3% Equal to or greater than 70% but less than 80%.
60-70 : 5.0% Equal to or greater than 60% but less than 70%.
50-60 : 0.6% Equal to or greater than 50% but less than 60%.
40-50 : 0.6% Equal to or greater than 40% but less than 50%.
30-40 : 0.3% Equal to or greater than 30% but less than 40%.
20-30 : 0.8% Equal to or greater than 20% but less than 30%.
10-20 : 0.3% Equal to or greater than 10% but less than 20%.
<10 : 9.2% Indicates the percentage of the Belt Loaded Time that
the Belt Loading was in the range less than 10%.
_______________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 64
Belt Load Statistics SECTION 12
_______________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 65
Production Reports SECTION 13
Production Reports
Provided the scale is not supplied as “Certified” or “Legal for Trade”, it is possible to obtain a
print out of data as stored within UNIPROIV. This feature is inhibited on certified
applications, via the door interlock switch.
To obtain a printed production report it is necessary to have the PTUII
Note: The printing of Calibration data is provided standard with the scale but if production
reports are required these will have to be purchased additionally.
Reports
The production report printout option allows for the print out of accumulated totals. It is
possible to select either the Current Total or Stored Reports (msg. 125 & 126).
All the printed Reports provide the Instrument Tag, Time and Date as headers.
In addition to the Totaliser values being printed, a further feature has been added to include
the following Performance / Statistical Data:
Run time – This is the time (in hours, minutes & seconds) that the scale has
been in operation (belt speed > zero).
Over Range – this is the percentage of the Run Time that the measured Belt
Loading is in excess of the Over Range Threshold value as programmed in item
9 of the Special Functions menu.
Under Range – this is the percentage of the Run Time that the measured Belt
Loading is less than the Under Range threshold value as programmed in item 9
of the Special Functions menu.
This data is often an invaluable tool for the optimum operation of plant & equipment.
Current Printout
Requesting a current Printout will result in the printout of a Monthly Report plus Statistical
Data for the corresponding period. Additionally the Master Total will be printed.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 66
Production Reports SECTION 13
Weekly printouts
Message 123 allows the selection, and the setting, of a weekly report. The start time and
day of the week is freely programmable. A maximum of four (4) weekly reports will be
stored in memory. In the event of no weekly report printout being made the transmitter will
remEMBer the last four (4) weekly reports only viz.; the oldest weekly report will be
overwritten by the latest. The Weekly Report will record the current master total plus
Performance / Statistical Data for the corresponding period, in addition to the header
details.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 67
Production Reports SECTION 13
Monthly printouts
Message 124 allows the selection, and the setting, of a monthly report. The start time and
day is freely programmable. A maximum of two (2) monthly reports will be stored in
memory. In the event of no monthly report printout being made the transmitter will
remEMBer the last two (2) monthly reports only viz.; the oldest monthly report will be
overwritten by the latest. The Monthly Report will record the current master total plus
Performance / Statistical Data for the corresponding period, in addition to the header
details.
After any printout has been successfully completed the transmitter will prompt for whether
the report may be removed from memory. The report memory in the transmitter is limited
and printed reports should be cleared from it as soon as any report has been printed. The
facility to store the report(s) has been provided only as a security measure to ensure that a
report is not lost due the printer running out of paper, low battery etc.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 68
Routine Maintenance SECTION 14
Routine Maintenance
Due to the unique design of the Process Automation conveyor Scale System, maintenance
requirements are of a minor nature.
The maintenance required by the system is:-
Physical maintenance
Electronic maintenance
Physical Maintenance
Physical maintenance involves the good housekeeping of the system (weigh frame, load cell
assEMBly, weigh idlers, tachometer, tachometer drive etc.). The sensors should be kept
clean and free of spillage.
Special care should be taken to ensure that no build-up of dirt occurs between the weigh
frame and any supporting structure which could interfere with the measurement.
As the weigh idlers, tachometer/drive assEMBly are the only parts making physical contact
with the conveyor belt, the condition of these should be checked frequently.
When supplied with the optional Tachometer Drive AssEMBly (TDA), the drive wheel should
be checked for roundness. Any out of round could be the result of mishandling (dropping/
bashing of the wheel). Depending on the drive model, the drive bearings (where applicable)
should be checked and greased (with Lithium based General Purpose grease) at least once,
preferably twice, per annum for normal applications - more often for applications in harsh
and/or wet environments.
The Test Weight (manual operated, lever operated or electrically actuated) must be secured
in the operating position. The counter balance weights must be securely bolted in position.
The idler alignment on the weigh frame should be checked initially three months after
installation, and at least annually thereafter, but preferably every six months if optimum
performance is required.
Load Cell
The scale force transducer (load cell) requires little maintenance.
Care should be exercised in the handling of low capacity load cells which may be destroyed
by overloads.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 69
Routine Maintenance SECTION 14
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 70
Routine Maintenance SECTION 14
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 71
Routine Maintenance SECTION 14
Lubrication Requirements
The only equipment requiring lubrication is on certain models of the optional Tachometer
Drive Assembly (TDA) used to couple the conveyor to the speed sensor (tachometer).
The two drive bearings on the Tachometer Drive Assembly should be checked and greased
(with a Lithium based General Purpose Grease) at least once, preferably twice, per annum
for normal applications - more often for applications in harsh and/or wet environments.
The internal bearings of tachometer are greased for life and normally require no greasing.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 72
Routine Maintenance SECTION 14
Electronic Maintenance
Electronic System Repair and Adjustment
The electronic instrumentation has been packaged for "Black Box" maintenance. If the
system stops functioning, and a quick visual inspection of the scale and instrument wiring
reveals no cause for concern, the printed circuit boards should be replaced, one by one
until the trouble is corrected. Keep the instrument doors closed tightly to prevent
entrance of moisture or conductive dust.
CAUTION: DO NOT ATTEMPT REPLACEMENT OF PRINTED CIRCUIT BOARDS WHEN THE AC
POWER IS ON.
Electronic maintenance involves the "Standardisation" of the gauge to known calibration
references to compensate for any changes in wear, electronic changes etc. - see
Calibration Checks.
Troubleshooting
When the system fails to operate properly, the recommended troubleshooting procedure
duplicates the procedures outlined in the section entitled "INITIAL CALIBRATION".
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 73
Software/Hardware Compatibility SECTION 15
Notes:
1. Expansion I/O is not supported on Revision 1 3000 CPU Boards.
2. A small number of Revision 5 3000 CPU Boards were produced and have since
been withdrawn from distribution.
3. 3000 CPU Board Revision 4 must be changed according to ECO202 for
compatibility with Software Version EMBC-220.40.25 and higher.
4. For correct operation of Software Version EMBC-220.40.25 and higher, a
32kb EEPROM (28C256) must be fitted in 3000 CPU Board position IC9.
EEPROM Size Jumper JP2 must be set to position 1-2.
5. 3000 CPU Board Revision 7 and higher.
________________________________________________________________________________________
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 74
Software List SECTION 16
EMBC 220.40.00 New Release (with Mk II Load Cell board): This software
runs on the new 3000 Mk2 (containing the PIC processor)
load cell board only– this software will not run with the
Mk I Load Cell board. The load cell input is now sampled
57 times per second and averaged over 0,5 seconds
before being passed to the main (80188) processor for
use in calculations.
This Mk2 load cell board incorporates a digital input (&
software) for speed sensor (tachometer) replacement.
U-NETII on RS-485 support has been added.
The weight, Speed and Rate filtering has been altered to
that as in UNIPROI software.
Serial number and Application data is no longer stored in
RAM. Note: This information must be saved (manually) to
EEPROM before the power is removed / power outage.
Also to view this data it should first be recalled from
EEPROM.
Bugs removed (from EMBC 220.30.01/4): Totaliser now
counts up or down when totaliser cut-off is activated.
The Tag can no longer be edited in the view Calibration
constants screen.
After selecting the mA output from the LC board to be
Control, this would return to Rate whenever any mA set-
up was accessed - even though the configuration would
indicate Control to be allocated.
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 75
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 76
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 77
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 78
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 79
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 80
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 81
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 82
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 83
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 84
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 85
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 86
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 87
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 88
Software List SECTION 16
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 89
Drawings SECTION 17
Drawings
03 May 2018 Operator Manual - Series 220 Conveyor Scales - Rev 13 Page 90