Sei sulla pagina 1di 9

CHAPTER 10

10 USE OF NAVIGATOR ECHOES IN CARDIOVASCULAR MAGNETIC RESONANCE AND FACTORS AFFECTING THEIR IMPLEMENTATION
Use of Navigator Echoes
in Cardiovascular Magnetic
Resonance and Factors Affecting
Their Implementation
David Firmin and Jennifer Keegan

Respiration has been shown to be an important factor influ In CMR, there have been a number of developments in
encing the quality of cardiovascular magnetic resonance the use of navigator measurement to reduce the problems
(CMR) images. In addition to cardiac motion, which can of respiratory motion. This chapter discusses these devel
be addressed reasonably well by electrocardiographic opments, considers the various choices that have been
(ECG) triggering, respiratory motion moves the position studied in the implementation of navigators, and discusses
and distorts the shape of the heart by several millimeters their importance. There are many variables in the applica
between inspiration and expiration. In 1991, Atkinson tion and use of navigator echoes, and although there have
and Edelman1 showed the detrimental effects of breathing been some attempts to study these, it is unlikely that we
on the quality of cardiac studies by showing improved are close to optimizing their application.
detail (fast low angle shot) in breath hold segmented fast
gradient echo images compared with conventional non
breath hold images. Although breath holding produces
images that are free of respiratory motion artifact, it is
USE OF NAVIGATOR
not without problems. The breath hold position may vary INFORMATION
from one breath hold scan to the next, giving rise to mis
registration effects, and it may also vary during the breath There are two distinct ways of using navigator echoes to
hold period itself,2 resulting in image blurring and arti reduce the problems of respiratory motion in CMR, which
facts. In addition, the scan parameters are limited by the are multiple breath holding with feedback and free breath
need to perform imaging within the duration of a comfort ing methods. The first of these uses the navigator informa
able breath hold period, and for a number of patients, this tion to provide visual feedback on the diaphragm position
period may be very short. to subjects to allow them to hold their breath at the same
An alternative to breath holding is to monitor respira point repeatedly.5 The second uses the navigator echo mea
tory motion throughout the data acquisition period and surement as an input to some form of respiratory gating
to correct the data for that motion, either in real time or algorithm while the patient breathes normally. Figure 10 1
through postprocessing, with the efficacy of both techni shows actual respiratory trace data in a subject when
ques being strongly dependent on the accuracy of the performing multiple breath holds and when free breathing.
method of motion assessment. During normal tidal respira In both cases, a navigator acceptance window, typically
tion, the superior inferior (SI) motion of the diaphragm is 5 mm wide, is defined, and all data acquired when the nav
approximately four to five times the anterior posterior igator is outside of this window are ignored. The resulting
motion of the chest wall,3 and consequently, diaphragm image therefore consists of data acquired over a narrow
motion is the most sensitive measure of respiratory motion. range of respiratory positions. The respiratory or scan effi
In 1989, Ehman and Felmlee4 were the first to introduce ciency is defined as the percentage of ECG triggers that fall
navigator echoes for measuring the displacement of a within the navigator acceptance window and is a measure
moving structure and to demonstrate their use in determin of the data rejection rate, which in turn determines the
ing diaphragm motion during respiration. The navigator overall scan duration. As the navigator acceptance window
echo is the signal from a column of material oriented per is reduced, the rejection rate increases and the scan
pendicular to the direction of the motion to be monitored. efficiency decreases. Figure 10 2 shows the residual dia
On Fourier transformation, this signal results in a well phragm displacements that occurred during data acquisi
defined edge of the moving structure. The navigator echoes tions performed during conventional breath holding,
may be interleaved with the imaging sequence and conse breath holding with navigator feedback, and navigator free
quently enable the motion to be determined throughout breathing in normal subjects.6 Both navigator techniques
the data acquisition period. result in images acquired over a reduced range of

Cardiovascular Magnetic Resonance 129


BASIC PRINCIPLES OF CARDIOVASCULAR MAGNETIC RESONANCE
5mm NE acceptance window Standard deviation
±2.4 mm ±0.7 mm ±0.7mm
Diaphragm displacement (mm) 0 30

20
20

40

Diaphragm displacement (mm)


10

60

0
A 0 Time (s) 60
Diaphragm displacement (mm)

−10
0
BH

LED
20 −20
+ FR

40
−30

B Figure 10 2 Mean diaphragm displacement in 17 normal subjects


with conventional breath holding (open circles), breath holding
Figure 10 1 Navigator echo respiratory trace data during multiple with navigator feedback (closed circles), and free breathing (plus
breath holding with navigator feedback (A) and free breathing marks). The navigator controlled studies used a 5 mm navigator
(B). In each case, the shaded region shows the position of a 5 mm acceptance window. (Adapted from Taylor AM, Keegan J, Jhooti P,
navigator acceptance window outside of which data is rejected. Gatehouse PD, Firmin DN, Pennell DJ. Differences between
(Adapted from Taylor AM, Keegan J, Jhooti P, Gatehouse PD, normal subjects and patients with coronary artery disease for
Firmin DN, Pennell DJ. Differences between normal subjects and three different MR coronary angiography respiratory suppression
patients with coronary artery disease for three different MR coronary techniques. J Magn Reson Imaging. 1999; 9:786 793, with permission.)
angiography respiratory suppression techniques. J Magn Reson
Imaging. 1999;9:786 793, with permission.)
it might be expected that breath holding with respiratory
feedback would enable the completion of a cardiac study
much more quickly than when using the free breathing
diaphragm displacement compared with those acquired methods described later, because of the time required for
using repeated conventional breath holding. In addition, training and the required rest periods between breath holds,
they allow a longer overall scan time. This allows for aver the overall examination times are longer than anticipated. In
aging of data to improve the signal to noise ratio, increas fact, in a group of patients with coronary artery disease,
ing the k space coverage for improved spatial resolution there was no significant difference between the overall
and increasing the temporal resolution by reducing the examination times with the two techniques,6 although the
number of individual image views acquired per cardiac same study showed that, in a group of normal healthy sub
cycle. jects, multiple breath holding resulted in a time reduction of
20%.

Multiple Breath Hold Methods


Wang and colleagues7 were the first to show the use of a
Free Breathing Methods
respiratory feedback monitor to reduce misregistration arti Free breathing methods require very little cooperation from
facts in consecutive breath hold segmented fast gradient the patient. The main disadvantage is the potential for
echo coronary artery images and to show improved image respiratory drift, which can cause considerably reduced
quality from averaging scans acquired over multiple breath scan efficiency.9 Recently, therefore, most effort has gone
holds. When used in informed healthy volunteers, this tech toward improving scan efficiency with this approach.
nique has been shown to produce good results with reason Much of the early work used retrospective respiratory
able scan efficiency.8 However, a period of training is gating.10 With this method, data acquisition was over
required, and the process can be problematic, particularly sampled, typically by a factor of five, and then sorted retro
with patients who have difficulty holding their breath spectively so that the final image was constructed from data
because of a combination of illness and anxiety.6 Although acquired over the narrowest possible range of respiratory

130 Cardiovascular Magnetic Resonance


positions. In 1995, Hofman and associates11 showed that, much greater scan efficiency than other methods, while

10 USE OF NAVIGATOR ECHOES IN CARDIOVASCULAR MAGNETIC RESONANCE AND FACTORS AFFECTING THEIR IMPLEMENTATION
using this approach, the image quality of three dimensional retaining scan quality (Table 10 1 and Figure 10 3). An
(3D) coronary acquisitions was improved over those alternative method, initially developed by Sinkus and
acquired with multiple averages. However, scan efficiency Bornert to address general respiratory motion16 and more
was poor (20% for an oversampling ratio of five), and recently applied to imaging of the coronary arteries, used
although the final image was constructed from the narrow a tailored acceptance window through k space as opposed
est respiratory window possible, the range was highly to phase encode ordering to obtain a very similar result.17
dependent on the subject’s breathing pattern during the Both of these phase ordering or windowing techniques
long acquisition period and was often still unacceptably use a predefined navigator acceptance window, and scan
high. efficiency is reduced when the respiratory pattern changes
After the introduction of prospective control techni during study acquisition. This has been more recently
ques, navigators have most commonly been used with a addressed by Jhooti and colleagues, who developed a tech
simple accept reject algorithm where data are acquired nique that combines the benefits of phase ordering with an
or not, depending on whether the navigator measurement automatic window selection that enables the highly effi
is within a predefined acceptance window. Oshinski cient acquisition of high quality coronary artery images
and coworkers12 were the first investigators to show without the need for a predefined acceptance window.18
high quality coronary artery images with such an Three dimensional motion adapted gating19 is a similar
approach. The problem with this method, however, is that technique that yields images comparable to standard pro
for reasonably high scan quality, a narrow acceptance win spective navigator gating, with significantly improved scan
dow of 5 mm or less is required, and this generally results efficiency.20
in relatively poor scan efficiency. In addition, as noted ear
lier, many subjects and patients undergo a drift in dia
phragm position over time,9 such that the predefined NAVIGATOR ECHO
acceptance window becomes less and less suitable as the
study progresses. The diminishing variance algorithm IMPLEMENTATION
overcomes this problem because it does not use a prede
fined acceptance window.13 With this method, one com
plete scan is acquired and the navigator positions are
Method of Column Selection
saved for each line of data. At the end of the initial scan, Two methods have been used for the generation of a navi
the most frequent diaphragm position during that scan gator echo.
is determined, and a process of reacquiring lines of data With the spin echo technique, a spin echo signal is gen
that were acquired with diaphragm positions furthest off erated from the column of material formed by the intersec
set from this position begins. As time progresses, the tion of two planes, one excited by a 90 radiofrequency
range of diaphragm positions for the data making up the (RF) pulse and the other by a 180 RF pulse. The column
final set is considerably reduced. In addition to the lack cross section may be either rectangular or rhombic,
of requirement of an acceptance window, this method depending on the orientation of the two planes. This
has the advantage that an image can be reconstructed at approach is very robust and produces an extremely well
any time after the initial dataset is complete. defined column. However, it cannot be repeated rapidly,
Another alternative to the simple accept reject algorithm and care must be taken to ensure that the column selection
that can improve both image quality and scan efficiency is planes do not impinge on the region of interest.
to use a k space ordering that depends on diaphragm posi The alternative approach is to use a selective two
tion. Two similar approaches have been suggested, based dimensional (2D) RF pulse to excite a column of approxi
on the finding that the center of k space appears to be more mately circular cross section.21 Although this technique is
sensitive to motion than the edges.14 Jhooti and colleagues much more sensitive to factors such as shimming errors,
developed a phase encode ordered method that used a dual which can potentially cause blurring and distortion of the
acceptance window of 5 mm for the center of k space and column, with a reduced flip angle, it can be repeated more
10 mm for the outer regions.15 This approach allowed rapidly and the navigator artifact is less extensive.

Table 10-1 Image Quality Scores and Scan Efficiencies{ for Three-Dimensional Magnetic Resonance
Angiography*
Phase Ordered ARA DVA{ RRG
Image Quality Mean Score 4.4 4.7 6.6 6.8
Scan Efficiency 72 ( 11.6) 48 ( 11.5) 72 ( 11.6) 20

*Mean image quality scores (1 ¼ excellent, 10 ¼ very poor) and scan efficiencies{ ( SD) for data acquired using phase ordering, an accept/reject algorithm
(ARA), the diminishing variance algorithm (DVA), and retrospective respiratory gating (RRG) in 15 subjects.
{
Scan time for the DVA technique is set to that of the phase ordered technique.
(Adapted from Jhooti P, Keegan J, Gatehouse PD, et al. 3D coronary artery imaging with phase reordering for improved scan efficiency. Magn Reson Med.
1999; 41: 555.)

Cardiovascular Magnetic Resonance 131


Figure 10 3 A single slice from a
BASIC PRINCIPLES OF CARDIOVASCULAR MAGNETIC RESONANCE

three dimensional dataset showing


a long section of the right coronary
artery. The phase ordered images
(A) are of comparable quality to
those acquired with the accept
reject algorithm (B) and better than
those acquired with both the
diminishing variance algorithm
(C) and retrospective respiratory
gating (D). Scan efficiency is also
significantly higher for phase
ordering than for both the accept
reject algorithm and retrospective
respiratory gating techniques.
(Adapted from Jhooti P, Keegan J,
A C Gatehouse PD, Collins S, Rowe A,
Taylor AM, Firmin DN. 3D coronary
Phase reordered (scan eff. = 73%) DVA (scan eff. = 73%) artery imaging with phase
reordering for improved scan
efficiency. Magn Reson Med 1999.
41:555 562, with permission.)

B D
ARA (scan eff. = 49%) RRG (scan eff. = 20%)

Both methods are used routinely for research studies on or slice tracking, is now used routinely for both 2D and
coronary imaging, without any reported problems. 3D methods of acquisition. Of note, however, is the rela
tively high standard deviation of the correction factor noted
earlier, which reflects considerable intersubject variation in
Correction Factors the degree of cardiac motion with respiration. This was also
observed by Danias and coworkers, who used real time 2D
In CMR, navigator echoes are most frequently used to mea echo planar imaging to study the SI motion of the heart as a
sure the position of the diaphragm. However, the motion of function of navigator position.24 The accuracy of slice fol
the heart is not straightforward, and only the inferior bor lowing techniques will obviously depend on the accuracy
der that sits on the diaphragm will move to the same of the correction factor implemented. In 1997, Taylor and
extent, whereas superiorly, the relative motion will be colleagues25 showed how a subject specific factor could
reduced. This was first studied by Wang and associates,3 be measured rapidly with end inspiratory and end expira
who measured the displacement of the right coronary tory breath hold scans before the coronary imaging proto
artery root, the origin of the left anterior descending artery, col. Figure 10 4 shows the relationship between the
and the superior and inferior margins of the heart relative motion of the right hemi diaphragm and the coronary ostia
to the diaphragm in 10 healthy subjects. For the right cor measured in one subject, with the slope of the graph giving
onary artery origin, the mean ( SD) relative displacement the correction factor. Figure 10 5 shows two examples of
(or correction factor) was 0.57 ( 0.26). McConnell and subjects with very different correction factors, showing
coworkers22 first used this correction factor to track the how a wider acceptance window can be used, thus improv
position of the imaging slice during breath holding and ing scan efficiency. The need for a subject specific correc
showed improved image registration relative to untracked tion factor has further been confirmed in 3D coronary
acquisitions. In free breathing studies, the correction factor angiography, where its use was found to yield optimal
was first applied by Danias and colleagues,23 who showed image quality.26 In 2002, Keegan and associates further
that tracked image quality was maintained as the navigator developed this area of work by studying the variability of
acceptance window increased from 3 mm to 7 mm, correction factors in the SI, anterior posterior, and right left
whereas in untracked images, it decreased significantly. directions for both breath holding and free breathing.27
This technique, called real time prospective slice following, The study concluded that subject variability in correction

132 Cardiovascular Magnetic Resonance


supports the use of a subject specific factor, as described

10 USE OF NAVIGATOR ECHOES IN CARDIOVASCULAR MAGNETIC RESONANCE AND FACTORS AFFECTING THEIR IMPLEMENTATION
0
in the previous section.
McConnell and colleagues32 studied the effects of vary
Downward coronary displacement (mm) y = –0.08 – (0.45 x) ing the navigator location on the image quality of coronary
r = 0.99 artery studies. Navigators were positioned through the
4
dome of the right hemi diaphragm, through the posterior
portion of the left hemi diaphragm, through the anterior
and posterior left ventricular walls, and through the ante
rior left ventricular wall, as shown in Figure 10 6. The
8 advantage of the latter navigator position is that it would
eliminate the need for a correction factor, as described
in the previous section, relating the navigator echo
measured displacement to the coronary artery motion.
12 The results are summarized in Table 10 2 and show no
significant differences in the image quality scores obtained
with varying navigator location. There was a tendency for
the anterior left ventricular wall navigator scans to be lon
16 ger in duration, but the difference did not reach statistical
0 10 20 30 significance. One of the problems with monitoring the
heart itself is the complex anatomy, making it more diffi
Downward diaphragm displacement (mm)
cult to find a suitable position for the navigator column.
Figure 10 4 Plot of superior inferior right coronary artery More sophisticated methods of positioning the column
displacement against superior inferior diaphragm displacement for a may further improve this method of monitoring cardiac
single subject. The gradient of the linear regression line is the subject motion.
specific correction factor. (Modified from Taylor AM, Keegan J, Jhooti P,
Firmin DN, Pennell DJ. Calculation of a subject specific adaptive
motion correction factor for improved real time navigator echo gated
magnetic resonance coronary angiography. J Cardiovasc Magn Reson.
1999; 1:131 138, with permission.)
MULTIPLE COLUMN
ORIENTATIONS
factors, together with within subject differences between
breath holding and free breathing, is such that slice follow There is a linear relationship between the SI and anterior
ing should be performed with subject specific factors deter posterior motions of the heart, with the SI motion being
mined from free breathing acquisitions. approximately five times that of the anterior posterior
An additional or alternative approach to the real time slice motion.3 For this reason, the real time slice following
following described earlier is to use a postprocessing adap methods first used by McConnell and colleagues22 and
tive motion correction technique4 to correct an image retro by Danias and associates23 included a correction for
spectively for movement occurring during data acquisition. anterior posterior motion of the heart, assuming it to be
This technique, which can be used to correct a 2D acquisi equal to 20% of the SI motion. Unfortunately, there is
tion for in plane displacement or a 3D acquisition for in not always such a strong relationship between the direc
plane and through plane displacement, may not appear to tions of motion of the heart with respiration. Sachs and
be an attractive option initially, but it has the advantages of colleagues showed this by using three navigators to
allowing the correction factor to be optimized for each indi measure the SI, anterior posterior, and right left motions
vidual patient and provides an alternative approach to those of the heart.33 Figure 10 7 shows an example from this
centers with scanners that do not have a real time decision study illustrating the scatter of SI, right left, and anterior
making capability. This approach has been implemented posterior measurements, made over a period of approxi
with both segmented gradient echo28 and interleaved spi mately 10 minutes. The group went on to compare the
ral29 coronary artery acquisitions, with promising results. use of one, two, and three navigators for imaging the
right coronary artery and showed an improvement when
multiple directions of motion were considered. This
Column Positioning improvement in image quality, however, must be offset
against the main disadvantage, which is that scan effi
The degree of diaphragm motion detected by the navigator ciency is reduced, potentially introducing more problems
echo is dependent on the positioning of the navigator col associated with long term drift in the breathing pattern. A
umn. The dome of the right hemi diaphragm is higher than more recent study by Jahnke and coworkers used a new
that of the left, and the two move coherently with respira cross correlation based approach and showed the poten
tion, but to differing degrees.30 Motion of the diaphragm is tial advantage of combining three orthogonal navigators.34
also greater posteriorly than anteriorly (anterior and dome
excursions are 56% and 79%, respectively, of posterior
excursions), and at the level of the dome, it is greater later
ally than medially.31 The correction factor implemented in
Navigator Timing
real time slice following or postprocessing adaptive motion Navigator timing is one of the more important parameters;
correction, as described earlier, is strongly dependent on however, flexibility to alter this is often limited by the com
the positioning of the navigator column and further puting architecture of the scanner being used (discussed

Cardiovascular Magnetic Resonance 133


BASIC PRINCIPLES OF CARDIOVASCULAR MAGNETIC RESONANCE

6 mm
16 mm

A
CF 0.70 0.00 1.00
6 mm
16 mm

B
CF 0.25 0.00 1.00

Figure 10 5 Right coronary artery origin images acquired with navigator acceptance windows of 6 mm and 16 mm in subjects with subject
specific correction factors (CFs) of 0.7 (A) and 0.25 (B). For both subjects, images were also acquired with CFs of 0 and 1. In the absence
of slice following (CF 0), image quality is reduced as the navigator acceptance window increases from 6 mm to 16 mm. When slice
following with a subject specific CF is used, however, image quality is maintained. (Modified from Taylor AM, Keegan J, Jhooti P, Firmin DN,
Pennell DJ. Calculation of a subject specific adaptive motion correction factor for improved real time navigator echo gated magnetic
resonance coronary angiography. J Cardiovasc Magn Reson. 1999; 1:131 138, with permission.)

134 Cardiovascular Magnetic Resonance


10 USE OF NAVIGATOR ECHOES IN CARDIOVASCULAR MAGNETIC RESONANCE AND FACTORS AFFECTING THEIR IMPLEMENTATION
A B C D

Figure 10 6 Navigator column locations positioned on tranverse, coronal and sagittal pilot images: (A) through the dome of the right hemi
diaphragm, (B) through the posterior left hemi diaphragm, (C) through both anterior and posterior left ventricular walls and (D) through the
anterior left ventricular wall.

Table 10-2 Image Quality Scores, Registration Errors, and Total Scan Times*
Right Diaphragm Left Diaphragm Left Ventricle Anterior LV Wall
Parameter Navigator Navigator Navigator Navigator
Image Quality Score (0 4) 2.3  0.1 2.3  0.1 2.4  0.1 2.2  0.2
Registration error (mm)
Craniocaudal 0.5  0.1 0.4  0.1 0.6  0.1 0.4  0.1
Anteroposterior 0.3  0.1 0.3  0.1 0.3  0.1 0.4  0.1
Total Scan Time{ (sec) 294  28 314  30 342  62 427  111

*Image quality scores (0 ¼ very poor, 4 ¼ excellent) registration errors, and total scan times for different navigator column positions during free breathing
MR coronary angiography. There were no significant differences between the navigator column locations. Data are presented as mean  standard error of
the mean (SEM); LV ¼ left ventricle
{
Total scan time is the time from start to finish for 6 scans.
Adapted from McConnell MV, Khasgiwala VC, Savord BJ, et al. Comparison of respiratory suppression methods and navigator locations for MR coronary
angiography. Am J Roentgenol. 1997;168:1369.

Figure 10 7 Anterior posterior


(A/P; A) and right left (R/L; B) 80
navigator echo measurements as a 60
function of superior inferior
navigator echo measurements in a
healthy subject. (Data provided by
Todd Sachs, Stanford University.)

R/L A/P

40

A 0 S/I 150 B 0 S/I 150

later). Figure 10 8 shows the three main alternatives: (1) postnavigators overcome this problem, but of course, they
pre , (2) pre and post , and (3) navigators repeated regu also reduce scan efficiency. In our experience, the use of
larly throughout the cardiac cycle. A simple prenavigator prenavigators only produces acceptable results for free
provides the highest scan efficiency when a navigator breathing studies, whereas multiple breath hold acquisi
acceptance window is used, but may not be reliable if there tions certainly require both pre and postnavigators. An
is a sudden change in breathing between the navigator important factor that depends on the computer hardware
measurement and image data acquisition. Pre and and architecture is the time required after the navigator

Cardiovascular Magnetic Resonance 135


Finally, the diaphragm edge may be detected by edge detec
BASIC PRINCIPLES OF CARDIOVASCULAR MAGNETIC RESONANCE

tion, correlation, or least squares fit algorithms. For rapid


tracking (repetition time < 100 msec) or narrower col
umns, the signal to noise ratio in the diaphragm trace
could be too poor for simple edge detection methods to
succeed. Of the remaining two techniques, the least squares
D. acq. fit method has been shown to be more resistant to the
effects of noise and to the diaphragm profile deformation
Pre- that occurs during respiration than the correlation method
Pre- and post- and therefore would be the technique of choice.36 How
D. acq. D. acq. D. acq. D. acq. ever, most navigator techniques acquire only one or two
navigators per cardiac cycle, and in such cases, the signal
Rep.
to noise ratios are usually relatively high and edge detec
tion algorithms are generally adequate.
Figure 10 8 Timing of the navigators for pre , pre and post ,
and repeated (Rep.) navigator echo controlled data acquisition
(D. acq.). MORE RECENT APPROACHES
acquisition before the start of the imaging sequence. Partic Other Forms of Navigators
ularly if prenavigators only are being used, the longer this
interval, the greater the potential for errors caused by respi As has been mentioned, there are problems with the con
ratory motion. Also, for ECG R wave triggered scans, this ventional navigators that have been described earlier
may also have implications for the minimum gating delay because they generally do not give a direct measure of
that can be obtained and for short gating delay or cine the respiratory related motion of the heart and they can
scans, post only navigators can be used as an alternative. not be implemented simply and efficiently to give a mea
This approach was recently implemented in left ventricular sure of this motion in 3D.37 A number of ingenious
function studies, where it was found that image quality in a alternative approaches have been described. In 2003,
group of patients with heart failure was significantly improved Nguyen and colleagues38 described a method that selec
over conventional breath hold scans.35 tively excited the epicardial fat, followed by a rapid read
Repeated navigators allow for improved cine or multi out scheme that instantaneously gave three 1D images of
slice imaging and also provide some potential for estimat its position in the x, y, and z directions. Tested on six sub
ing internavigator respiratory motion. The potential prob jects, the method showed a slight improvement over con
lem with this is that the navigator signal to noise ratio ventional navigators; however, the authors noted a
could be reduced and this may affect the accuracy of navi number of problems that would need to be resolved
gator edge detection. In addition, as the time for navigator before it could be in routine use. In the same year, another
output increases, the time for imaging decreases and the method of rapidly localizing heart signals for measure
number of phases or slices that can be acquired is reduced. ment of its position was suggested by Pai and Wen, who
used a phase contrast angiographic approach to selectively
image the flowing blood in the heart chambers.39 These
blood signals were used to define the heart position in
Precision of Navigator the SI and anterior posterior directions. Despite the
Measurement potential advantages of truly tracking the heart position,
this method does not appear to have been developed fur
Commonly, a spatial resolution of 1 mm is used along the ther or used.
navigator echo column, for example, having a field of view Subsequently, Stehning and associates used radial
of 512 mm and sampling 512 points on the navigator read imaging for “self navigation.” They developed an inter
out. However, the precision of the measurement is depen leaved 3D radial acquisition modified in such a way that
dent to a large extent on the signal to noise ratio of the the first readout was always in the SI direction.40 This
navigator measurement. The most important factor affect readout could then be reconstructed every cardiac cycle
ing the signal to noise ratio is the coil arrangement. If, for to give an SI projection for motion extraction. In this
example, a single coil is used for imaging and navigator work, the authors showed improved definition compared
detection, it must be large enough to cover both the imag with conventional navigators when imaging a moving
ing area of interest and the region of navigator detection. phantom and similar image quality on initial in vivo cor
On the other hand, if phased array coils are used, it is pos onary scans. Hardy and colleagues41 used cross correla
sible to position one coil specifically for navigator detec tion of low resolution real time 2D spiral coronary
tion, possibly over the region of the right diaphragm. artery images to accept or reject images for averaging.
Another important factor in the precision of the measure This adaptive averaging technique was extended to high
ment is the quality of the edge on the navigator trace. To resolution segmented acquisitions by cross correlating subi
obtain a well defined edge of the diaphragm, for example, mages reconstructed from individual data segments. Breath
it is important to have a reasonably small column cross sec ing autocorrection with spiral interleaves (BACSPIN) is a
tion and to position it through the dome of the diaphragm, similar technique42 that involves the acquisition of a multi
so that the column is perpendicular to the diaphragm edge, slice spiral dataset during breath holding, followed by
rather than more posteriorly, where motion is greatest. repeated acquisition of the same slices during free breathing.

136 Cardiovascular Magnetic Resonance


Each heavily undersampled free breathing spiral interleaf is and image quality was also considerably better than for

10 USE OF NAVIGATOR ECHOES IN CARDIOVASCULAR MAGNETIC RESONANCE AND FACTORS AFFECTING THEIR IMPLEMENTATION
compared with the same interleaf acquired during breath those acquired simply with a 20 mm acceptance window.
holding, and those that match closely are incorporated into A more recent study showed that the residual coronary
a multi slice, multi average dataset. This technique has been artery motion observed using affine navigators with a 10
shown in a group of six healthy volunteers where increased mm acceptance window is similar to that observed with
signal to noise ratio has been achieved with minimal motion conventional navigator gating with a 5 mm window and
blurring. An alternate approach to respiratory motion cor that observed using a single breath hold.48
rection that has been applied to spiral imaging is to acquire
a low resolution 3D dataset on the fat resonance immedi
ately before a high resolution interleaf on the water reso
nance. Cross correlation of a selected region of interest of
Computer Architecture
the low resolution fat images from beat to beat is then used The computer architecture of modern CMR scanners can
to determine the x, y, and z translations of that region and be very complex, generally incorporating three main com
can be used to correct the next high resolution interleaf puters. The host computer runs the user interface and
retrospectively. This approach has been shown in dark allows connection to the image database, the reconstruction
blood coronary vessel wall imaging, where high quality computer is a dedicated rapid processor for reconstruction
images have been obtained without the need for a gating of the CMR image data, and the scan computer allows con
window.43 trol and adjustment of parameters associated with the scan
ning sequence. The architecture of these computers can
significantly affect the potential and usefulness of navigator
Motion Models echoes. On many systems, for example, the navigator signal
must be reconstructed and processed on the reconstruction
Because of the complexity of cardiac motion and the diffi computer, but the measurement made must be passed
culty in extracting measures to correct for it, there has been through the host to control the parameters on the scan
an interest in developing modeling methods to assist with computer. This arrangement inevitably adds a variable
this process. Manke and coworkers compared one and unknown delay that is dependent on other tasks being
dimensional (1D) translation (SI direction), 3D translation, performed by the host operating system. To overcome this,
and 3D affine transformation motion models in a group of either a direct and rapid link is required, allowing transfer
healthy subjects.44 By using an elastic image registration of data from the reconstruction computer to the scan com
algorithm on 3D coronary images acquired at different puter, or the scan computer itself must be capable of
breath hold positions, the superiority of the 3D transla acquiring and reconstructing the navigator data, so that
tional model over the 1D translational model was shown. no data transfer is required. Newer scanners are generally
The authors also used a fast model based image registration being designed with rapid acquisition, processing, and con
to extract motion information from a time series of low trol in mind.
resolution 3D images. This was used in conjunction with
conventional navigators to calibrate a respiratory motion
model that allowed the prediction of affine transformation CONCLUSION
parameters, including 3D translation, rotation, scale, and
shear motion from the navigator measurements, and was Navigator echo has been shown to be an important method
shown with coronary artery imaging.45 McLeish and col for monitoring respiration that has been used for defining
leagues46 acquired images at a number of breath hold posi the position of the heart, enabling improved coronary and
tions and studied the accuracy of both rigid and nonrigid other cardiac imaging. The limited number of studies and
registration methods in registering the other breath hold the many parameters and variables involved in their use
images with those acquired at end expiration. They used suggest that an optimal method of application may not
principal component analysis to produce patient specific yet have been developed. With future system development,
statistical motion models and suggest how this could be there will be minimal cost, in imaging time or other factors,
used to assist motion correction in CMR. In 2005, Nehrke involved in obtaining this positional information. There
and Bornert described their study in which they used fore, it would seem worthwhile to collect and use it where
a patient specific model to control the acquisition.47 appropriate. The methods are not robust, probably because
Initially, multiple low resolution 3D images were preac of their relative lack of sophistication.
quired during free breathing. An affine model of the respi One of the major advantages of this technique is that
ratory related cardiac motion was then extracted from these navigators allow images to be acquired during free respira
images, and this was steered by real time navigators to con tion, eliminating the need for patient cooperation. They
trol the high resolution acquisition. The method was also allow longer acquisition times, enabling higher spatial
demonstrated in both phantoms and volunteers when a and temporal resolution and increasing the potential for
20 mm navigator acceptance window was used. The results more sophisticated techniques, such as detailed flow imag
in the volunteers showed slightly inferior image quality to ing.49,50 A balance must be maintained, however, and
images acquired simply with a 5 mm navigator acceptance imaging time should not be increased so much that
window. However, scan efficiency was considerably higher increased respiratory drift cancels any potential benefit.

Cardiovascular Magnetic Resonance 137

Potrebbero piacerti anche