General/Product Category ICP MS

Instruments Varian810/820, Bruker Aurora

Notification of change: Replacement Kit for Obsolete Vent Valve

Dear colleagues,

This instruction is only applicable for the Replacement Kit 10-5000-500-11 of the obsolete Aurora Vent
Valve 10-5-3610008300.
Kit Contents:

NR PART NUMBER DESCRIPTION

1 10-5000-500-12 Valve Adapter Aurora

2 10-5000-500-91 Adaptor Cable Vent Valve

3 10-405-648 Magnetic Valve

4 10-406-346 Silencer

5 10-406-347 O-Ring 8,73x1,78

Analytik Jena AG Telefon +49 3641 77 70


Konrad-Zuse-Straße 1 Telefax +49 3641 77 9279
07745 Jena  Deutschland info@analytik-jena.de
www.analytik-jena.de 1/3
Service Information

1. Remove the damaged Vent Valve


2. Remove the Valve Adaptor 10-5000-500-12 from the Magnetic Valve 10-405-648 and install it to
the interface. Make sure the O-Ring 10-406-347 is mounted perfectly.

3. Mount the Magnetic Valve 10-405-648 now to the adaptor again.

Analytik Jena AG Telefon +49 3641 77 70


Konrad-Zuse-Straße 1 Telefax +49 3641 77 9279
07745 Jena  Deutschland info@analytik-jena.de
www.analytik-jena.de 2/3
Service Information

4. Before connecting the flat blade terminal to DIN valve type C connector remove the long screw from
the center of the plug. If you do not remove this screw it will cause an electrical short with the two
blade terminals.

5. Connect the Adaptor Cable 10-5000-500-91 to the Magnetic Valve and the blade terminals to the
DIN valve type C connector.

The orientation of the cable contacts (wire colors) Use the left and right contact of the DIN valve plug
does not matter

If you have further questions, I will gladly assist.

Best regards

Ismail Sel

Lead Service Support ICP-MS