1)ER01,ER02 temperature alarming

2)ER03 inverter alarming

3)ER04 lamp filament alarming

4)ER05 rotating anode alarming

5)ER06 ,ER07 overflow time alarming

6)The lamp of Limited beam device does not light

7)LCD displays a black screen

8)Connecting failure

9)Solenoid valve failure

10)Failed to turn on

11)The fuse-blew
ER01,ER02 temperature alarming
Checking and processing steps
1. Check the two temperature probe connectors upper computer,check whether the
temperature switches connectors is correct, make sure the power supply voltage is
normal,its temperature probe resistance about 605,temperature switches resistance
should be 0 .
2. Enterinto debug interface, check whether T1, T2 temperature range is in allowed
scope 37 degrees, if it is beyond the temperature range, check whether the cooling fan
rotation is normal or not.
3. make sure the above all are normal,then check the PC, the machine board, change
the circuit board.
ER03 inverter alarming
1. Enter into the debugging interface,Check whether the parameter is normal,if it isnot
normal,you should restore the parameters first.
2. make a false IGBT load experiment of the two groups, check whether the IGBT are
normal,whether the drive ability is consistent, if it isnot,change the damaged. If the
IGBT are all not working in the two groups, check the driver board,observe whether
the board is functioning properly.check whether the Kvset signal and voltage of
interface board to its backplane is normal or not.
3. If the given signal or voltage stabilizer of interface board are not normal, replace
the corresponding inspection of lower level computer board or manostat board, and
related components.
4. If the interface board KVset signals are normal,and the drive adjustment plate is
normal,but IGBT components of two groups still not working, replace the backplane.
5. If the above examination results are all normal, check whether the tube and the kv
feedback circuit board,and the kv feedback board to the Inverter backplane
wiring,circuit board are normal or not.
6. If the IGBT close of two groups are all normal work, disconnect the adjustment
board,measurement the KV feedback board whether KV + and KV- to GND
resistance are balance,disconnect KV feedback board,in the tube, measuring
whether KV + and KV- to GND resistance are balance, then judge whether KV
feedback board and the tube is normal, if not normal, replace them.
ER04 lamp filament alarming
1. confirm the connection from interface board to filament board,5v,24v direct
voltage,and the connection from the tube filanment and filament board to the tube are
normal.
2. Short sub of GOOD signal,if the early warning device is no longer alarming,the
filament board is failure, replace the filament board.
ER05 rotating anode alarming
1. confirm the related wirings of circuit board is correct,220 V AC voltage, 15 V, 5 V
dc voltage are normal.
2. If 220V AC voltage is not normal,the front power circuit or the filter is breakdown;
if the lines connected with the toroidal transformer, each of the AC input and output
voltages are not normal,pls check the toroidal transformer.
3. If the DC power supply is breakdown,it must be the manostat board and the
realated lines,and the transformer are breakdown.
4. confirm the connection from rotating anode plate output to the tube is normal,and
Winding resistance is correct(PS70Ω,PC20Ω, SC50Ω).
5. The above tests are normal, short sub alarm signal should turn to the OK, if the
failure of the rotating anode plate is not alarming, pls replace the rotating anode plate.
ER06 ,ER07overflow time alarming
1. check the parameters normal or not under the debugging interface,especially Mul
parameters are reset.
2. check the filament plate is normal or not.
3. check the connection from the tube mA feedback circuit to the inverter backplane
and the interface board is normal or not.
4. MA stopwatch test feedback mAs,check the mA normal or not,if not,pls check the
fixed variable of mA normal nor not.
5. If the fixed variable of mA is normal,the kv feedback board or back-plane is
breakdown.If it isnot normal,pls replace the lower PC-PLC component assembly.
6. if the above tests are normal,check IGBT assembly.False load experiment,confirms
the working of the IGBT assembly is normal or not,if not,pls replace it.
The lamp of Limited beam device does not light
Make sure that the mask button of limited beam device and linkline from the button to
the delay board is in good condition and can use normally.
1. If the 24V DC voltage can input the delay board normally, that is the problem of
delay board or the bulb, you should change them.
2. If the 24V voltage is not normal, that is the problem of 250W switching power
supply failure, or 250W switching power supply 220V AC input is not normal.
LCD displays a black screen.
Starting up the machine to check whether the 9V micrococontroller power supply of
the host computer is normal of not. If it is not normal, please check whether the lines
and manostat board output power is normal or not.
1 .If the microcontroller power supply is normal, check the cable connection to see
whether there has 5V voltage to supply LCD screen, If there is no 5V supply, please
replace the PC board.
2 .If above are normal, that is the problem that the LCD screen is damaged, please
replace it.
Connecting failure
Confirm the communication power and the lines of communication is normal and
check whether the line in the towline is disconnecting.
1. If the interface board output communication port voltage is correct, that shows the
PC board (S260-01) is damaged, you should replace the PC board.
2. If the above are normal, please replace the PC-PLC attachment (S260--09).
Solenoid valve failure
1. Bed solenoid valve does not pull, if the 24V switching power supply is working
properly and the line is normal, that shows the power supply electromagnetic valve
plate failure.
2. The solenoid valve is not controlled and it can not loosen, you should check
whether the trip switch or mask switch is working properly. If it working normally, it
is the power supply electromagnetic valve plate failure, replace the power supply
solenoid valve plate.
Failed to turn on
1. If the power light is on, and the circuit for turn on process works, then check
whether 2A fuse wire works,. If the parts mentioned above all worked well, the
problem maybe cause by the Solenoid valve for Power Management Board, then you
replace it will be ok.
2. If the power light is off, please check whether the voltage for equipment is 220V
first , then check the fuse wire and the powe line of Solenoid valve for Power
Management Board whether works, If the parts mentioned above are all works, then
the problem may caused by the Solenoid valve for Power Management Board
The fuse-blew
1. If the 6A fuse is blown while the machine was working. Please check the output
voltage whether steady. If the voltage is steady, then check the rotating anode, or
replace it directly. If the fuse is blown at once when the machine turns on, you need to
check the Solenoid valve for Power Management Board and rotating anode working
condition or works.
2. If the 2A fuse is blown, and the voltage under the condition of normal operating
range, please check filters and isolating transformer, and at last check the two
transformers for voltage-stabilized power source board which are most likely caused
the question
3.It is rarely to meet the problem that 63A fuse is blown, at most case this is caused by
unstable voltage.