Sei sulla pagina 1di 2

pressure instrumentation

1.1.3.

Badotherm model BDT6 series is only available as 100 mm dia and is fillable.
These gauges are suitable for gaseous and liquid, non-aggressive media that
will not obstruct the pressure system or attack copper alloy parts and has a
ABS casing.

ABS 100 mm pressure gauge

mounting variations
type A bottom connection, direct mounting
type B panel mounting, clamp fixing, back connection
BDT6 type C surface mounting, bottom connection
type D back connection, direct mounting
type E panel mounting, back connection
type F panel mounting, bottom connection

technical specification model BDT6


diameter connection window ranges
100 mm 1/2” BSP/NPT glass from -1/1600 bar
casing tube movement dial
ABS copper alloy brass aluminium
bezel socket pointer accuracy
ABS threaded ring brass fixed 1.0% FSD

Note: only available with bottom connection

options and special executions at extra cost


window safety glass
dials customized – i.e., non standard ranges or logos
certificates calibration -; conformity -; material
connection METRIC; BSPT; acc JIS in various sizes
restrictor brass; AISI316
AISI316 internals type BDT6-18
Monel internals type BDT6-M
filling glycerine
red pointer on dial
connection position top or side
receiver gauge 3-15 psi; 0.2-1 bar; 0-100% lin; 0-10 sq.rt.
ranges other on request
non fillable type BDT 5

56 For technical data sheets and dimensional drawings please refer to our website www.badotherm.com
01/01/2006 rev 2
BDT6 - technical data sheet

pressure details
working pressure

Steady 3
/4 x full scale value
Fluctuating 2
/3 x full scale value
Short time see over-pressure table

pressure element over-pressure


<100 bar 100 bar max. scale in bar over-pressure at 20 °C
BDT6, BDT5 Copper alloy AISI316 -1 to 100 1,25 x full scale value
BDT6-18, BDT5-18 AISI316 AISI316 101 to 600 1,15 x full scale value
601 to 1600 1,10 x full scale value

Note : only withstand for a short period of time

temperature details
temperature limitations
ambient process The variation of indication caused by the effects of tem-
perature shall not exceed :
glycerine filling -25 ºC to 60 ºC 60 ºC
±0,04 x (ambient temperature-reference temperature)%
brass internal, no filling -40 ºC to 60 ºC 65 ºC
of the span
st.st. internal, no filling -40 ºC to 60 ºC 200 ºC

design information
ingress protection design
IP 65 per EN 60 529 / IEC 529 EN 837-1
IP 54 per EN 60 529 / IEC 529 (BDT5, BDT5-18 only)

drawings

type A type A

dimensions
ns d d1 L H h weight
100 114 100 50 145 - 0,60 kg
Note : dimensions are in mm