Sei sulla pagina 1di 18

C18 Column

Care & Use Sheet


HALO Description • Water and all common organic solvents are
HALO C18 is a high-speed, high-performance liquid
compatible with HALO C18 columns.
chromatography column based on a new Fused-CoreTM
particle design. The Fused-Core particle provides a thin • Halo C18 columns are best used at temperatures
porous shell of high-purity silica surrounding a solid below 60 ºC for maximum column life.
silica core. This particle design exhibits very high • Mobile phase pH for HALO C18 columns is best
column efficiency due to the shallow diffusion paths in maintained in the range of pH = 2 to 9 for maximum
the 0.5-micron thick porous shell and the small overall column stability.
particle size of 2.7-microns. The densely bonded, • HALO C18 columns are stable to operating
extensively endcapped dimethyloctadecyl stationary pressures up to 600 bar (9000 psi).
phase of HALO C18 provides a stable, reversed-phase
packing that can be used for basic, acidic, or neutral Column Care
compounds. To maximize column life, ensure that samples and
mobile phases are particle-free. The use of guard
Column Characteristics columns or an in-line filter with 0.5-micron porosity
Figure 1 shows a sample Quality Control test between the sample injector and the column is highly
chromatogram for a 4.6 mm x 75 mm column. A printed recommended. The 2-micron porosity frits on HALO
report including the actual test chromatogram and C18 columns are less subject to pluggage than are the
performance results is enclosed with every column. 0.5-micron frits typically used with other small-particle
The Fused-Core particle has a surface area of ~ 150 m2/g columns. Should the operating pressure of the column
and an average pore size of 90Å. The Fused-Core suddenly increase beyond normal levels, reversing the
particles are 30% to 50% heavier than commercially flow direction of the column may be attempted to
available totally porous particles due to the density of the remove debris on the inlet frit.
solid cores. Therefore, the effective surface area per To remove strongly retained materials from the column,
column is similar to columns packed with totally porous flush the column in the reverse direction with very
particles having surface areas in the 225-300 m2/g range. strong solvents such as 100% of the organic component
of the mobile phase in use. A mixture (95/5 v/v) of
dichloromethane and methanol is often effective at this
Figure 1 - HALO C18 QC Chromatogram
TEST CONDITIONS
task. Extreme cases may require the use of very strong
70 Column: 4.6 x 75 mm HALO C18 (2.7 µm)
Mobile Phase: 60% ACN/ 40% Water
solvents such as dimethylformamide (DMF) or
60
2
Flow: 1.8 ml/min
Temperature : Ambient (24ºC)
dimethylsulfoxide (DMSO).
Detector: UV (254 nm)
50 Sample Volume: 2 µl

40
1
Peak Identities:
1. Uracil 5 µg/ml Column Storage
2. Phenol 200 µg/ml
3. 4-Chloro-1-nitrobenzene 20 µg/ml Long-term storage of silica-based, reversed-phase
mAU

30
3 4. Naphthalene 20 µg/ml
columns is best in 100% acetonitrile. Columns may be
20
4
safely stored for short periods (up to 3 or 4 days) in most
10 common mobile phases. However, when using buffers,
0
it is best to protect both the column and the HPLC
equipment and remove the salts by flushing the column
-10
0 0.5 1 1.5 2 2.5 with the same mobile phase without the buffer (e.g.,
Time, minutes
when using 60/40 ACN/buffer, flush the column with
60/40 ACN/H2O) to eliminate any danger from
Operation Guidelines corrosion from the salts while providing rapid re-
equilibration of the column with the original mobile
• The direction of flow is marked on the column label.
phase.
• Reversed flow may be used to attempt removal of
Before storing the column, the end-fittings should be
inlet pluggage or contamination.
tightly sealed with the end-plugs that came with the
• A new column contains a mixture of acetonitrile and column to prevent the packing from drying.
water. Initial care should be taken to avoid mobile
phases that are immiscible with this mixture or could
cause a precipitate.
Safety • Detector – Flow cells should be of low-volume
• HPLC columns are for laboratory use only. Not design (preferably < 2µl).
for drug, household, or other use. • Detector – To properly sense and integrate the often
• Users of HPLC columns should be aware of the very fast peaks that elute from low-volume columns,
toxicity or flammability of the mobile phases chosen the detector response time should be set to the
for use with the columns. Precautions should be taken fastest level (~ 0.1 second) and the integration
to avoid contact and leaks. software should sample the detector signal at least
• HPLC columns should be used in well-ventilated 20 points per second.
environments to minimize concentration of solvent • Injector – The injection system should be of a low-
fumes. volume design (e.g., Rheodyne Model 8125). Auto-
samplers will often cause band spreading with low-
Applications volume columns but may be used for convenience
The HALO C18 bonded phase is nonpolar in nature. It is with the expectation of some loss in column
best utilized with mobile phases that are mixtures of efficiency.
methanol and water or acetonitrile and water. Higher • Connecting Tubing – The shortest possible lengths
levels of the organic solvent component will typically of connecting tubing with narrow internal diameters
reduce the retention of the sample compounds. Using (at most 0.005-inch, 0.12 mm ID) should be used to
elevated temperatures (e.g., 40 – 60 ºC) will reduce the connect the column to the injector and the detector
viscosity of the mobile phase and allow the use of faster cell. The tubing must have flat ends and should
flow rates and lower column pressure for high sample bottom out inside all fittings. Zero-dead-volume
throughput. Gradient elution techniques using 5 -10% fittings should always be used where required.
organic component as the initial mobile phase and • Peak Retention – As retention is increased, the
increasing to 100% organic component as the final volume of a peak increases, decreasing the effects
mobile phase often can effect separations of complex on band spreading caused by components of the
sample mixtures in minimal time. instrument.
HALO C18 columns are highly suited for the reversed- • Sample Solvent – For isocratic separations, the
phase separation of basic, neutral, or acidic compounds. sample should be dissolved in the mobile phase or in
Ionizable compounds, such as acids and bases, are a solvent that is weaker (less polar) than the mobile
generally best separated with mobile phases buffered at phase. For gradient separations, the sample should
pH of 2 to 3. The use of 20-50 mM buffers is always be dissolved in the initial mobile phase or in a
recommended for optimum results and long-term stability solvent substantially weaker than the final mobile
when separating ionizable compounds. Additional phase.
information on solvent selection and separation • Injection Volume – For isocratic separations, the
techniques can be found in Chapters Six, Seven, and volume of sample injected should be kept as small
Eight, Practical HPLC Method Development, Second as possible (typically 2 µl or less). Sample volumes
Edition, L.R. Snyder, J.L. Glajch, and J.J. Kirkland, (John are less critical for gradient separations, especially
Wiley & Sons, 1997). if the sample is dissolved in a weak solvent.

Guidelines for Low-Volume Columns Ordering Information


High performance columns with small internal volumes For ordering information or for technical support on this
(shorter lengths, internal diameters < 3 mm) are being product, in the U.S. call MAC-MOD Analytical at 800-
increasingly used for high speed separations, especially 441-7508 For information in other countries go to the
with specialty detection systems such as mass website:
spectrometers. These low-volume columns generate www.advanced-materials-tech.com
peaks having considerably less volume than those eluting
from columns of larger dimensions (e.g., 4.6 mm x 150 TM
mm). The efficiency of separations performed in low- HALO and Fused-Core are trademarks of Advanced
volume columns is highly dependent on the HPLC Materials Technology, Inc.
system having components designed to minimize band
spreading. All low-volume columns perform best when
used with proper attention to the following factors:
C8 Column
Care & Use Sheet
HALO-5 Description Water and all common organic solvents are
HALO-5 C8 is a high-speed, high-performance liquid
compatible with HALO-5 C8 columns.
chromatography column based on the highly efficient
HALO-5 C8 columns are best used at temperatures
Fused-Core® particle design. The Fused-Core particle
below 60 ºC for maximum column life.
provides a thin porous shell of high-purity silica
Mobile phase pH for HALO-5 C8 columns is best
surrounding a solid silica core. This particle design
maintained in the range of pH = 2 to 9 for maximum
exhibits very high column efficiency due to the shallow
column stability.
diffusion paths in the 0.6-micron thick porous shell and
the highly uniform overall particle size of 5-microns. HALO-5 C8 columns are stable to operating
The densely bonded, extensively endcapped pressures up to 600 bar (9000 psi).
dimethyloctyl stationary phase of HALO-5 C8 provides a
stable, reversed-phase packing that can be used for basic, Column Care
acidic, or neutral compounds. To maximize column life, ensure that samples and
mobile phases are particle-free. The use of guard
Column Characteristics columns or an in-line filter with 0.5-micron porosity
Figure 1 shows a sample Quality Control test between the sample injector and the column is highly
chromatogram for a 4.6 mm x 150 mm column. A printed recommended. The 2-micron porosity frits on HALO-5
report including the actual test chromatogram and C8 columns are less subject to pluggage than are the 0.5-
performance results is enclosed with every column. micron frits typically used with 3-micron particle
The 5-micron Fused-Core particle has a surface area of columns. Should the operating pressure of the column
~ 100 m2/g and an average pore size of 90Å. The Fused- suddenly increase beyond normal levels, reversing the
Core particles are 30% to 50% heavier than commercially flow direction of the column may be attempted to
available totally porous particles due to the density of the remove debris on the inlet frit.
solid cores. Therefore, the effective surface area per To remove strongly retained materials from the column,
column is similar to columns packed with totally porous flush the column in the reverse direction with very
particles having surface areas in the 160-200 m2/g range. strong solvents such as 100% of the organic component
of the mobile phase in use. A mixture (95/5 v/v) of
dichloromethane and methanol is often effective at this
task. Extreme cases may require the use of very strong
solvents such as dimethylformamide (DMF) or
dimethylsulfoxide (DMSO).

Column Storage
Long-term storage of silica-based, reversed-phase
columns is best in 100% acetonitrile. Columns may be
safely stored for short periods (up to 3 or 4 days) in most
common mobile phases. However, when using buffers,
it is best to protect both the column and the HPLC
equipment and remove the salts by flushing the column
with the same mobile phase without the buffer (e.g.,
Operation Guidelines when using 60/40 ACN/buffer, flush the column with
The direction of flow is marked on the column label. 60/40 ACN/H2O) to eliminate any danger from
Reversed flow may be used to attempt removal of corrosion from the salts while providing rapid re-
inlet pluggage or contamination. equilibration of the column with the original mobile
A new column contains a mixture of acetonitrile and phase.
water. Initial care should be taken to avoid mobile Before storing the column, the end-fittings should be
phases that are immiscible with this mixture or could tightly sealed with the end-plugs that came with the
cause a precipitate. column to prevent the packing from drying.
Safety Detector – Flow cells should be of low-volume
HPLC columns are for laboratory use only. Not
design (preferably < 2µl).
for drug, household, or other use.
Detector – To properly sense and integrate the often
Users of HPLC columns should be aware of the
very fast peaks that elute from low-volume columns,
toxicity or flammability of the mobile phases chosen
the detector response time should be set to the
for use with the columns. Precautions should be taken
fastest level (~ 0.1 second) and the integration
to avoid contact and leaks.
software should sample the detector signal at least
HPLC columns should be used in well-ventilated 20 points per second.
environments to minimize concentration of solvent
Injector – The injection system should be of a low-
fumes.
volume design (e.g., Rheodyne Model 8125). Auto-
samplers will often cause band spreading with low-
Applications volume columns but may be used for convenience
The HALO-5 C8 bonded phase is nonpolar in nature. It with the expectation of some loss in column
is best utilized with mobile phases that are mixtures of efficiency.
methanol and water or acetonitrile and water. Higher Connecting Tubing – The shortest possible lengths
levels of the organic solvent component will typically of connecting tubing with narrow internal diameters
reduce the retention of the sample compounds. Using (at most 0.005-inch, 0.12 mm ID) should be used to
elevated temperatures (e.g., 40 – 60 ºC) will reduce the connect the column to the injector and the detector
viscosity of the mobile phase and allow the use of faster cell. The tubing must have flat ends and should
flow rates and lower column pressure for high sample bottom out inside all fittings. Zero-dead-volume
throughput. Gradient elution techniques using 5 -10% fittings should always be used where required.
organic component as the initial mobile phase and Peak Retention – As retention is increased, the
increasing to 100% organic component as the final volume of a peak increases, decreasing the effects
mobile phase often can effect separations of complex on band spreading caused by components of the
sample mixtures in minimal time. instrument.
HALO-5 C8 columns are highly suited for the reversed- Sample Solvent – For isocratic separations, the
phase separation of basic, neutral, or acidic compounds. sample should be dissolved in the mobile phase or in
Ionizable compounds, such as acids and bases, are a solvent that is weaker (more polar) than the mobile
generally best separated with mobile phases buffered at phase. For gradient separations, the sample should
pH of 2 to 3. The use of 20-50 mM buffers is always be dissolved in the initial mobile phase or in a
recommended for optimum results and long-term stability solvent substantially weaker than the final mobile
when separating ionizable compounds. Additional phase.
information on solvent selection and separation Injection Volume – For isocratic separations, the
techniques can be found in Chapters Six, Seven, and volume of sample injected should be kept as small
Eight, Practical HPLC Method Development, Second
as possible (typically 2 µl or less). Sample volumes
Edition, L.R. Snyder, J.L. Glajch, and J.J. Kirkland, (John are less critical for gradient separations, especially
Wiley & Sons, 1997).
if the sample is dissolved in a weak solvent.

Guidelines for Low-Volume Columns Ordering Information


High performance columns with small internal volumes For ordering information or for technical support on this
(shorter lengths, internal diameters < 3 mm) are being product, in the U.S. call MAC-MOD Analytical at 1-
increasingly used for high speed separations, especially 800-441-7508. For information in other countries go to
with specialty detection systems such as mass the website: www.advanced-materials-tech.com
spectrometers. These low-volume columns generate
peaks having considerably less volume than those eluting
from columns of larger dimensions (e.g., 4.6 mm x 150
mm). The efficiency of separations performed in low- HALO and Fused-Core are registered trademarks of
volume columns is highly dependent on the HPLC Advanced Materials Technology, Inc.
system having components designed to minimize band
spreading. All low-volume columns perform best when
used with proper attention to the following factors:
ES-CN Column
Care & Use Sheet
HALO-5 Description Water and all common organic solvents are
HALO-5 ES-CN is a high-speed, high-performance
compatible with HALO-5 ES-CN columns.
liquid chromatography column based on the highly
HALO-5 ES-CN columns are best used at
efficient Fused-Core® particle design. The Fused-Core
temperatures below 60 ºC for maximum column life.
particle provides a thin porous shell of high-purity silica
Mobile phase pH for HALO-5 ES-CN columns is
surrounding a solid silica core. This particle design
best maintained in the range of pH = 1 to 8 for
exhibits very high column efficiency due to the shallow
maximum column stability.
diffusion paths in the 0.6-micron thick porous shell and
the highly uniform overall particle size of 5-microns. HALO-5 ES-CN columns are stable to operating
The sterically protected, extensively endcapped pressures up to 600 bar (9000 psi).
diisopropyl-cyanopropylsilane stationary phase of
HALO-5 ES-CN provides a stable, reversed-phase Column Care
packing that can be used for basic, acidic, or neutral To maximize column life, ensure that samples and
compounds. mobile phases are particle-free. The use of guard
columns or an in-line filter with 0.5-micron porosity
Column Characteristics between the sample injector and the column is highly
Figure 1 shows a sample Quality Control test recommended. The 2-micron porosity frits on HALO-5
chromatogram for a 4.6 mm x 50 mm column. A printed ES-CN columns are less subject to pluggage than are the
report including the actual test chromatogram and 0.5-micron frits typically used with 3-micron particle
performance results is enclosed with every column. columns. Should the operating pressure of the column
The 5-micron Fused-Core particle has a surface area of suddenly increase beyond normal levels, reversing the
~ 100 m2/g and an average pore size of 90Å. The Fused- flow direction of the column may be attempted to
Core particles are 30% to 50% heavier than commercially remove debris on the inlet frit.
available totally porous particles due to the density of the To remove strongly retained materials from the column,
solid cores. Therefore, the effective surface area per flush the column in the reverse direction with very
column is similar to columns packed with totally porous strong solvents such as 100% of the organic component
particles having surface areas in the 160-200 m2/g range. of the mobile phase in use. A mixture (95/5 v/v) of
dichloromethane and methanol is often effective at this
task. Extreme cases may require the use of very strong
solvents such as dimethylformamide (DMF) or
dimethylsulfoxide (DMSO).

Column Storage
Long-term storage of silica-based, reversed-phase
columns is best in 100% acetonitrile. Columns may be
safely stored for short periods (up to 3 or 4 days) in most
common mobile phases. However, when using buffers,
it is best to protect both the column and the HPLC
equipment and remove the salts by flushing the column
with the same mobile phase without the buffer (e.g.,
when using 60/40 ACN/buffer, flush the column with
60/40 ACN/H2O) to eliminate any danger from
Operation Guidelines corrosion from the salts while providing rapid re-
The direction of flow is marked on the column label. equilibration of the column with the original mobile
Reversed flow may be used to attempt removal of phase.
inlet pluggage or contamination. Before storing the column, the end-fittings should be
A new column contains a mixture of acetonitrile and tightly sealed with the end-plugs that came with the
water. Initial care should be taken to avoid mobile column to prevent the packing from drying.
phases that are immiscible with this mixture or could
cause a precipitate.
Safety
HPLC columns are for laboratory use only. Not Detector – Flow cells should be of low-volume
for drug, household, or other use. design (preferably < 2µl).
Users of HPLC columns should be aware of the Detector – To properly sense and integrate the often
toxicity or flammability of the mobile phases chosen very fast peaks that elute from low-volume columns,
for use with the columns. Precautions should be taken the detector response time should be set to the
to avoid contact and leaks. fastest level (~ 0.1 second) and the integration
HPLC columns should be used in well-ventilated software should sample the detector signal at least
environments to minimize concentration of solvent 20 points per second.
fumes. Injector – The injection system should be of a low-
volume design (e.g., Rheodyne Model 8125). Auto-
samplers will often cause band spreading with low-
Applications volume columns but may be used for convenience
The HALO-5 ES-CN bonded phase is moderately polar
with the expectation of some loss in column
in nature. It is best utilized with mobile phases that are
efficiency.
mixtures of methanol and water or acetonitrile and water.
Connecting Tubing – The shortest possible lengths
Higher levels of the organic solvent component will
of connecting tubing with narrow internal diameters
typically reduce the retention of the sample compounds.
(at most 0.005-inch, 0.12 mm ID) should be used to
Using elevated temperatures (e.g., 40 – 60 ºC) will reduce
connect the column to the injector and the detector
the viscosity of the mobile phase and allow the use of
cell. The tubing must have flat ends and should
faster flow rates and lower column pressure for high
bottom out inside all fittings. Zero-dead-volume
sample throughput. Gradient elution techniques using 5 -
fittings should always be used where required.
10% organic component as the initial mobile phase and
increasing to 100% organic component as the final Peak Retention – As retention is increased, the
mobile phase often can effect separations of complex volume of a peak increases, decreasing the effects
sample mixtures in minimal time. on band spreading caused by components of the
HALO-5 ES-CN columns are highly suited for the instrument.
reversed-phase separation of basic, neutral, or acidic Sample Solvent – For isocratic separations, the
compounds. Ionizable compounds, such as acids and sample should be dissolved in the mobile phase or in
bases, are generally best separated with mobile phases a solvent that is weaker (more polar) than the mobile
buffered at pH of 2 to 3. The use of 20-50 mM buffers is phase. For gradient separations, the sample should
always recommended for optimum results and long-term be dissolved in the initial mobile phase or in a
stability when separating ionizable compounds. solvent substantially weaker than the final mobile
Additional information on solvent selection and phase.
separation techniques can be found in Chapters Six, Injection Volume – For isocratic separations, the
Seven, and Eight, Practical HPLC Method Development, volume of sample injected should be kept as small
Second Edition, L.R. Snyder, J.L. Glajch, and J.J. as possible (typically 2 µl or less). Sample volumes
Kirkland, (John Wiley & Sons, 1997). are less critical for gradient separations, especially
if the sample is dissolved in a weak solvent.

Guidelines for Low-Volume Columns Ordering Information


High performance columns with small internal volumes For ordering information or for technical support on this
(shorter lengths, internal diameters < 3 mm) are being product, in the U.S. call MAC-MOD Analytical at 1-
increasingly used for high speed separations, especially 800-441-7508. For information in other countries go to
with specialty detection systems such as mass the website: www.advanced-materials-tech.com
spectrometers. These low-volume columns generate
peaks having considerably less volume than those eluting
from columns of larger dimensions (e.g., 4.6 mm x 150
mm). The efficiency of separations performed in low-
HALO and Fused-Core are registered trademarks of
volume columns is highly dependent on the HPLC Advanced Materials Technology, Inc.
system having components designed to minimize band
spreading. All low-volume columns perform best when
used with proper attention to the following factors:
HILIC Column
Care & Use Sheet
HALO Description
HALO-5 HILIC is a high-speed, high-performance liquid HALO-5 HILIC columns are best used at temperatures
chromatography column based on a 5-micron Fused-CoreTM below 60 ºC for maximum column life.
particle design. The Fused-Core particle provides a thin porous Mobile phase pH for HALO-5 HILIC columns is best
shell of high-purity silica surrounding a solid silica core. This maintained in the range of pH = 1 to 8 for maximum
particle design exhibits very high column efficiency due to the column stability.
shallow diffusion paths in the 0.6-micron thick porous shell and HALO-5 HILIC columns are stable to operating pressures
an overall particle size of ~5-microns. The bare silica up to 600 bar (9000 psi).
stationary phase of HALO-5 HILIC provides a column that can
be used for traditional normal phase separations using non- Column Care
polar, totally organic mobile phases (not discussed in this To maximize column life, ensure that samples and mobile
document) or for aqueous normal phase chromatography with phases are particle-free. The use of guard columns or an in-
the typical reversed-phase mobile phases for hydrophilic line filter with 0.5-micron porosity between the sample
interactive liquid chromatography (HILIC) of basic, acidic, or injector and the column is highly recommended. The 2-
neutral compounds. micron porosity frits on HALO-5 HILIC columns are less
subject to pluggage than are the 0.5-micron frits typically used
Column Characteristics with other small-particle columns. Should the operating
Figure 1 shows a sample Quality Control test chromatogram pressure of the column suddenly increase beyond normal
for a 4.6 mm x 50 mm column. A printed test report including levels, reversing the flow direction of the column may be
the actual test chromatogram and performance results for a attempted to remove debris on the inlet frit.
specific column is enclosed with the column. To remove strongly retained materials from the column, flush
The Fused-Core particle has a surface area of ~ 100 m2/g and the column in the reverse direction with very strong solvents
an average pore size of 90Å. The Fused-Core particles are such as 50/50 methanol and deionized water. Extreme cases
30% to 50% heavier than commercially available totally porous may require the use of very strong solvents such as 100% of
particles due to the density of the solid cores. Therefore, the the most polar component of the mobile phase in use.
effective surface area per column is similar to columns packed Alternatively, a mixture (95/5 v/v) of dichloromethane and
with totally porous particles having surface areas in the 160- methanol is often effective at this task.
200 m2/g range.
Column Storage
Figure 1 - HALO-5 HILIC QC Chromatogram Long-term storage of silica-based columns is best in 100%
3
acetonitrile. Columns may be safely stored for short periods
Test Conditions:
Column: 4.6 x 50 mm HALO-5 HILIC (up to 3 or 4 days) in most common mobile phases. However,
2 Mobile Phase: 88/12 ACN/0.1 M Ammonium
Formate, pH = 3.0
when using buffers, it is best to remove the salts to protect
Flow rate: 1.0 mL/min
o
both the column and the HPLC equipment by first flushing the
1 Temperature: ambient (~ 24 C)
Detection: UV at 254 nm column with the same mobile phase without the buffer (e.g.,
Injection: 1.0 µL
Sample components:
when using 90/10 ACN/buffer, flush the column with 90/10
1. acenaphthene 80 µg/mL ACN/H2O) to eliminate any concern about salt precipitation or
2. adenosine 35 µg/mL
3. cytosine 75 µg/mL corrosion from the salts then flush the column with 100%
acetonitrile for storage.
Before storing the column, the end-fittings should be tightly
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 sealed with the end-plugs that came with the column to
Time, minutes prevent the packing from drying.

Safety
Operation Guidelines HPLC columns are for laboratory use only. Not for
The direction of flow is marked on the column label. drug, household, or other use.
Reversed flow may be used to attempt removal of inlet Users of HPLC columns should be aware of the toxicity
pluggage or contamination. or flammability of the mobile phases chosen for use with
the columns. Precautions should be taken to avoid contact
A new column contains 100% acetonitrile.
and leaks.
Water and all common organic solvents are compatible HPLC columns should be used in well-ventilated
with HALO-5 HILIC columns. environments to minimize concentration of solvent
fumes.
Applications
HILIC is a useful and complimentary method to reversed-phase
chromatography (RPC) and is especially attractive in situations Guidelines for Low-Volume Columns
where compound retention is poor in RPC and very high levels High performance columns with small internal volumes
of water are required in the mobile phase for adequate (shorter lengths, internal diameters < 3 mm) are being
retention. Retention in HILIC is not well-understood but increasingly used for high speed separations, especially with
appears to be a combination of hydrophilic interaction, ion- specialty detection systems such as mass spectrometers.
exchange and some reversed-phase retention. The aqueous These low-volume columns generate peaks having
layer which forms on the surface of HILIC particles promotes considerably less volume than those eluting from columns of
interaction with polar solutes. Retention in HILIC as a larger dimensions (e.g., 4.6 mm x 150 mm). The efficiency of
function of the mobile phase is just opposite from that in RPC. separations performed in low-volume columns is highly
The strongest mobile phase has a high concentration of water dependent on the HPLC system having components designed
and the weakest has a high concentration of organic solvent. to minimize band spreading. All low-volume columns
Therefore, for gradient separations, the initial mobile phase has perform best when used with proper attention to the following
a high concentration of organic solvent and the gradient is factors:
formed by increasing the aqueous concentration. Greatest Detector – Flow cells should be of low-volume design
retention for basic and acidic analytes is found when using (preferably < 2µl).
more than about 70% organic (e.g., acetonitrile) in acidic Detector – To properly sense and integrate the often very
mobile phases. High organic concentrations are used in the fast peaks that elute from low-volume columns, the
mobile phases, therefore, HILIC is especially favorable for detector response time should be set to the fastest level (~
separations using mass spectrometry (MS) detection. 0.1 second) and the integration software should sample
the detector signal at least 20 points per second.
Acetonitrile is typically used as the weak organic solvent in the Injector – The injection system should be of a low-
mobile phase. With this solvent, 95% is typically the upper volume design (e.g., Rheodyne Model 8125). Auto-
limit and 60 - 65% the lower limit for adequate retention. At samplers will often cause band spreading with low-
least 5% of the mobile phase should be the highly polar solvent volume columns but may be used for convenience with
such as water or methanol. Water should be the polar solvent if the expectation of some loss in column efficiency.
a buffer is included because of solubility limitations. The
Connecting Tubing – The shortest possible lengths of
organic solvent type can be varied to change retention and
connecting tubing with narrow internal diameters (at most
separation selectivity, much as in RPC. Solvent strength (from
0.005-inch, 0.12 mm ID) should be used to connect the
weakest to strongest) for HILIC generally is tetrahydrofuran <
column to the injector and the detector cell. The tubing
acetone < acetonitrile < isopropanol < ethanol < methanol <
must have flat ends and should bottom out inside all
water, where water is the strongest elution solvent. To further
fittings. Zero-dead-volume fittings should always be
increase retention in HILIC, replacing some of the water in the
used where required.
mobile phase with another polar solvent such as methanol or
Peak Retention – As retention is increased, the volume of
isopropanol sometimes is effective.
a peak increases, decreasing the effects on band spreading
caused by components of the instrument.
For optimum column efficiency and reproducibility, buffers in
Sample Solvent – For isocratic separations, the sample
the range of 10 - 20 mM concentration or additives in the 0.5%
range are used the mobile phase. Phosphate buffers are not should be dissolved in the mobile phase or in a solvent
recommended because of their poor solubility in high organic that is weaker (less water) than the mobile phase. For
mobile phases and incompatibility with MS detection. gradient separations, the sample should be dissolved in
Additives such as formic acid, trifluoroacetic acid and the initial mobile phase or in a solvent substantially
phosphoric acid at concentrations up to about 1% can be a part weaker than the final mobile phase.
of the mobile phase. Volatile ammonium formate/formic acid Injection Volume – For isocratic separations, the volume
buffers up to a final concentration of about 20 mM and pH 3 of sample injected should be kept as small as possible
are especially effective for separating both basic and acidic (typically 2 µl or less). Sample volumes are less critical
compounds when using MS detection. (Acetonitrile/formate for gradient separations, especially if the sample is
mobile phases seem to be a good starting point for many dissolved in a weak solvent.
separations of both basic and acidic compounds.) Ammonium
acetate at pH ~5 also have been used at concentrations of 5 - 20 Ordering Information
mM, but are generally less effective for separating stronger For ordering information or for technical support on this
basic and acidic compounds. Buffers or additives above pH 6 product, in the U.S. call MAC-MOD Analytical at 800-
usually are not recommended because of slow dissolution of 441-7508 For information in other countries go to the
the silica support. website: www.advanced-materials-tech.com
HALO and Fused-Core are trademarks of Advanced Materials Technology, Inc.
TM
HILIC Column
General Information

HALO Description
HALO HILIC is a high-speed, high-performance liquid HALO HILIC columns are best used at temperatures
chromatography column based on a new Fused-CoreTM particle below 60 ºC for maximum column life.
design. The Fused-Core particle provides a thin porous shell of Mobile phase pH for HALO HILIC columns is best
high-purity silica surrounding a solid silica core. This particle maintained in the range of pH = 2 to 8 for maximum
design exhibits very high column efficiency due to the shallow column stability.
diffusion paths in the 0.5-micron thick porous shell and the HALO HILIC columns are stable to operating pressures
small overall particle size of 2.7-microns. The bare silica up to 600 bar (9000 psi).
stationary phase of HALO HILIC provides a column that can
be used for traditional normal phase separations using non- Column Care
polar, totally organic mobile phases (not discussed in this To maximize column life, ensure that samples and mobile
document) or for aqueous normal phase chromatography with phases are particle-free. The use of guard columns or an in-
the typical reversed-phase mobile phases for hydrophilic line filter with 0.5-micron porosity between the sample
interactive liquid chromatography (HILIC) of basic, acidic, or injector and the column is highly recommended. The 2-
neutral compounds. micron porosity frits on HALO HILIC columns are less
subject to pluggage than are the 0.5-micron frits typically used
Column Characteristics with other small-particle columns. Should the operating
Figure 1 shows a sample Quality Control test chromatogram pressure of the column suddenly increase beyond normal
for a 2.1 mm x 50 mm column. A printed test report including levels, reversing the flow direction of the column may be
the actual test chromatogram and performance results for a attempted to remove debris on the inlet frit.
specific column is enclosed with the column. To remove strongly retained materials from the column, flush
The Fused-Core particle has a surface area of ~ 150 m2/g and the column in the reverse direction with very strong solvents
an average pore size of 90Å. The Fused-Core particles are such as 50/50 methanol and deionized water. Extreme cases
30% to 50% heavier than commercially available totally porous may require the use of very strong solvents such as 100% of
particles due to the density of the solid cores. Therefore, the the most polar component of the mobile phase in use.
effective surface area per column is similar to columns packed Alternatively, a mixture (95/5 v/v) of dichloromethane and
with totally porous particles having surface areas in the 225- methanol is often effective at this task.
300 m2/g range.
Column Storage
Figure 1 - HALO HILIC QC Chromatogram Long-term storage of silica-based columns is best in 100%
180
acetonitrile. Columns may be safely stored for short periods
160
(up to 3 or 4 days) in most common mobile phases. However,
140
3 Test Conditions when using buffers, it is best to remove the salts to protect
Column: 2.1 x 50 mm Halo HILIC

120 2
Mobile Phase: 90/10 ACN/0.1 M Ammonium
Formate, pH=3.0
both the column and the HPLC equipment by first flushing the
100
Flow: 0.4 ml/min
Temperature: Ambient (24 ºC)
column with the same mobile phase without the buffer (e.g.,
1 Detection: UV (254 nm)
Sample Volume: 0.5µl when using 90/10 ACN/buffer, flush the column with 90/10
80 Sample Components:
1. acenaphthene
2. adenosine
80 µg/ml
35 µg/ml
ACN/H2O) to eliminate any concern about salt precipitation or
60
3. cytosine 75 µg/ml
corrosion from the salts then flush the column with 100%
40
acetonitrile for storage.
20
Before storing the column, the end-fittings should be tightly
0
sealed with the end-plugs that came with the column to
-20
prevent the packing from drying.
0 0.5 1 1.5 2
Time, minutes
Safety
HPLC columns are for laboratory use only. Not for
drug, household, or other use.
Operation Guidelines Users of HPLC columns should be aware of the toxicity
or flammability of the mobile phases chosen for use with
The direction of flow is marked on the column label. the columns. Precautions should be taken to avoid contact
Reversed flow may be used to attempt removal of inlet and leaks.
pluggage or contamination. HPLC columns should be used in well-ventilated
A new column contains 100% acetonitrile. environments to minimize concentration of solvent
Water and all common organic solvents are compatible fumes.
with HALO HILIC columns.
Applications Guidelines for Low-Volume Columns
HILIC is a useful and complimentary method to reversed-phase High performance columns with small internal volumes
chromatography (RPC) and is especially attractive in situations (shorter lengths, internal diameters < 3 mm) are being
where compound retention is poor in RPC and very high levels increasingly used for high speed separations, especially with
of water are required in the mobile phase for adequate specialty detection systems such as mass spectrometers.
retention. Retention in HILIC is not well-understood but These low-volume columns generate peaks having
appears to be a combination of hydrophilic interaction, ion- considerably less volume than those eluting from columns of
exchange and some reversed-phase retention. The aqueous larger dimensions (e.g., 4.6 mm x 150 mm). The efficiency of
layer which forms on the surface of HILIC particles promotes separations performed in low-volume columns is highly
interaction with polar solutes. Retention in HILIC as a dependent on the HPLC system having components designed
function of the mobile phase is just opposite from that in RPC. to minimize band spreading. All low-volume columns
The strongest mobile phase has a high concentration of water perform best when used with proper attention to the following
and the weakest has a high concentration of organic solvent. factors:
Therefore, for gradient separations, the initial mobile phase has Detector Flow cells should be of low-volume design
a high concentration of organic solvent and the gradient is (preferably < 2µl).
formed by increasing the aqueous concentration. Greatest Detector To properly sense and integrate the often very
retention for basic and acidic analytes is found when using fast peaks that elute from low-volume columns, the
more than about 70% organic (e.g., acetonitrile) in acidic detector response time should be set to the fastest level (~
mobile phases. High organic concentrations are used in the 0.1 second) and the integration software should sample
mobile phases, therefore, HILIC is especially favorable for the detector signal at least 20 points per second.
separations using mass spectrometry (MS) detection. Injector The injection system should be of a low-
volume design (e.g., Rheodyne Model 8125). Auto-
Acetonitrile is typically used as the weak organic solvent in the samplers will often cause band spreading with low-
mobile phase. With this solvent, 95% is typically the upper volume columns but may be used for convenience with
limit and 60 - 65% the lower limit for adequate retention. At the expectation of some loss in column efficiency.
least 5% of the mobile phase should be the highly polar solvent Connecting Tubing The shortest possible lengths of
such as water or methanol. Water should be the polar solvent if connecting tubing with narrow internal diameters (at most
a buffer is included because of solubility limitations. The 0.005-inch, 0.12 mm ID) should be used to connect the
organic solvent type can be varied to change retention and column to the injector and the detector cell. The tubing
separation selectivity, much as in RPC. Solvent strength (from must have flat ends and should bottom out inside all
weakest to strongest) for HILIC generally is tetrahydrofuran < fittings. Zero-dead-volume fittings should always be
acetone < acetonitrile < isopropanol < ethanol < methanol < used where required.
water, where water is the strongest elution solvent. To further Peak Retention As retention is increased, the volume of
increase retention in HILIC, replacing some of the water in the a peak increases, decreasing the effects on band spreading
mobile phase with another polar solvent such as methanol or caused by components of the instrument.
isopropanol sometimes is effective. Sample Solvent For isocratic separations, the sample
should be dissolved in the mobile phase or in a solvent
For optimum column efficiency and reproducibility, buffers in that is weaker (less polar) than the mobile phase. For
the range of 10 - 20 mM concentration or additives in the 0.5% gradient separations, the sample should be dissolved in
range are used the mobile phase. Phosphate buffers are not the initial mobile phase or in a solvent substantially
recommended because of their poor solubility in high organic weaker than the final mobile phase.
mobile phases and incompatibility with MS detection.
Injection Volume For isocratic separations, the volume
Additives such as formic acid, trifluoroacetic acid and
of sample injected should be kept as small as possible
phosphoric acid at concentrations up to about 1% can be a part
(typically 2 µl or less). Sample volumes are less critical
of the mobile phase. Volatile ammonium formate/formic acid
for gradient separations, especially if the sample is
buffers up to a final concentration of about 20 mM and pH 3
dissolved in a weak solvent.
are especially effective for separating both basic and acidic
compounds when using MS detection. (Acetonitrile/formate
mobile phases seem to be a good starting point for many Ordering Information
separations of both basic and acidic compounds.) Ammonium For ordering information or for technical support on this
acetate at pH ~5 also have been used at concentrations of 5 - 20 product contact Hichrom Limited on 0118 930 3660.
mM, but are generally less effective for separating stronger TM
HALO and Fused-Core are trademarks of Advanced Materials Technology, Inc.
basic and acidic compounds. Buffers or additives above pH 6
usually are not recommended because of slow dissolution of
the silica support.

,
Penta-HILIC Column
Care & Use Sheet
HALO-5 Penta-HILIC columns are best used at
Description temperatures below 60 ºC for maximum column life.
HALO-5 Penta-HILIC is a high-speed, high-performance
Mobile phase pH for HALO-5 Penta-HILIC columns is
liquid chromatography column based on a 5-micron Fused-
best maintained in the range of pH = 2 to 9 for maximum
CoreTM particle design. The Fused-Core particle provides a thin
column stability.
porous shell of high-purity silica surrounding a solid silica
core. This particle design exhibits very high column efficiency HALO-5 Penta-HILIC columns are stable to operating
due to the shallow diffusion paths in the 0.6-micron thick pressures up to at least 600 bar (9000 psi).
porous shell and an overall particle size of ~5-microns. The
HALO-5 Penta-HILIC stationary phase is a highly polar ligand Column Care
that possesses 5 hydroxyl groups tethered to the silica via novel To maximize column life, ensure that samples and mobile
proprietary linkage chemistry. This high performance material phases are particle-free. The use of guard columns or an in-
provides a column that can be used for traditional normal phase line filter with 0.5-micron porosity between the sample
separations using non-polar, totally organic mobile phases (not injector and the column is highly recommended. The 2-
discussed in this document) or for aqueous normal phase micron porosity frits on HALO-5 Penta-HILIC columns are
chromatography with the typical mobile phases for hydrophilic less subject to pluggage than are the 0.5-micron frits typically
interactive liquid chromatography (HILIC) of basic, acidic, used with other small-particle columns. Should the operating
zwitterionic, or neutral compounds. pressure of the column suddenly increase beyond normal
levels, reversing the flow direction of the column may be
attempted to remove debris on the inlet frit.
Column Characteristics To remove strongly retained materials from the column, flush
Figure 1 shows a sample Quality Control test chromatogram
the column in the reverse direction with very strong solvents
for a 4.6 mm x 50 mm column. A printed test report including
such as 10/90 methanol and deionized water. Extreme cases
the actual test chromatogram and performance results for a
may require the use of very strong solvents such as 100% of
specific column is enclosed with the column.
the most polar component of the mobile phase in use, which is
The Fused-Core particle has a surface area of ~ 100 m2/g and
typically water.
an average pore size of 90Å. The Fused-Core particles are
30% to 50% heavier than commercially available totally porous Column Storage
particles due to the density of the solid cores. Therefore, the
Long-term storage of silica-based columns is best in 100%
effective surface area per column is similar to columns packed
acetonitrile. Columns may be safely stored for short periods
with totally porous particles having surface areas in the 160-
(up to 3 or 4 days) in most common mobile phases. However,
200 m2/g range.
when using buffers, it is best to remove the salts to protect
both the column and the HPLC equipment by first flushing the
Figure 1 - HALO-5 Penta-HILIC QC Chromatogram column with the same mobile phase without the buffer (e.g.,
when using 90/10 ACN/buffer, flush the column with 90/10
Test Conditions:
Column: 4.6 x 50 mm HALO-5 Penta-HILIC ACN/H2O) to eliminate any concern about salt precipitation or
3 Mobile Phase: 88/12 ACN/0.1 M Ammonium
Formate, pH = 3.0 corrosion from the salts then flush the column with 100%
1 2
Flow rate: 1.0 mL/min
Temperature: ambient (~ 24 oC)
acetonitrile for storage.
Detection: UV at 254 nm Before storing the column, the end-fittings should be tightly
Injection: 1.0 µL
Sample components: sealed with the end-plugs that came with the column to
1. acenaphthene
2. adenosine
80 µg/mL
35 µg/mL
prevent the packing from drying.
3. cytosine 75 µg/mL After storing the column in 100% acetonitrile for a prolonged
period of time, the packing material may become dehydrated,
necessitating a rehydration treatment. We recommend
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 flushing the column at reduced flow rate at room temperature
Time, minutes
overnight in 50% acetonitrile/water mixture.

Safety
Operation Guidelines HPLC columns are for laboratory use only. Not for
The direction of flow is marked on the column label. drug, household, or other use.
Reversed flow may be used to attempt removal of inlet Users of HPLC columns should be aware of the toxicity
pluggage or contamination. or flammability of the mobile phases chosen for use with
the columns. Precautions should be taken to avoid contact
A new column contains 90% acetonitrile/10% water.
and leaks.
Water and all common organic solvents are compatible HPLC columns should be used in well-ventilated
with HALO-5 Penta-HILIC columns. environments to minimize concentration of solvent
fumes.
Applications
HILIC is a useful and complimentary method to reversed-phase Guidelines for Low-Volume Columns
chromatography (RPC) and is especially attractive in situations High performance columns with small internal volumes
where compound retention is poor in RPC. Retention in HILIC (shorter lengths, internal diameters < 3 mm) are being
is not fully understood but appears to be a combination of increasingly used for high speed separations, especially with
hydrophilic interaction, ion-exchange and hydrophobic specialty detection systems such as mass spectrometers.
retention. The aqueous layer which forms on the surface of These low-volume columns generate peaks having
HILIC particles promotes interaction with polar solutes. As a considerably less volume than those eluting from columns of
first approximation, retention in HILIC as a function of the larger dimensions (e.g., 4.6 mm x 150 mm). The efficiency of
mobile phase is opposite that in RPC. The strong mobile phase separations performed in low-volume columns is highly
has high concentration of water and the weakest has a high dependent on the HPLC system having components designed
concentration of organic solvent. For gradient separations the to minimize band spreading. All low-volume columns
initial mobile phase has a high concentration of organic solvent perform best when used with proper attention to the following
and the gradient is formed by increasing the aqueous factors:
concentration. Greatest retention for many analytes is found Detector – Flow cells should be of low-volume design
when using more than about 70% organic (e.g., acetonitrile) in (preferably < 2µl).
acidic mobile phases. High organic concentrations are used in Detector – To properly sense and integrate the often very
the mobile phases, therefore, HILIC is especially favorable for fast peaks that elute from low-volume columns, the
separations using mass spectrometry (MS) detection. detector response time should be set to the fastest level (~
0.1 second) and the integration software should sample
Acetonitrile is typically used as the weak organic solvent in the the detector signal at least 20 points per second.
mobile phase. With this solvent, 95% is typically the upper Injector – The injection system should be of a low-
limit and 50-60% the lower limit for adequate retention. At volume design (e.g., Rheodyne Model 8125). Auto-
least 5% of the mobile phase should be the highly polar solvent samplers will often cause band spreading with low-
such as water or methanol. Water should be the polar solvent if volume columns but may be used for convenience with
a buffer is included because of solubility limitations. The the expectation of some loss in column efficiency.
organic solvent can be varied to alter retention and selectivity. Connecting Tubing – The shortest possible lengths of
Solvent strength (from weakest to strongest) for HILIC connecting tubing with narrow internal diameters (at most
generally is tetrahydrofuran < acetone < acetonitrile < 0.005-inch, 0.12 mm ID) should be used to connect the
isopropanol < ethanol < methanol < water, where water is the column to the injector and the detector cell. The tubing
strongest elution solvent. To further alter retention, replacing must have flat ends and should bottom out inside all
water in the mobile phase with another polar solvent may be fittings. Zero-dead-volume fittings should always be
effective. used where required.
Peak Retention – As retention is increased, the volume of
For optimum column efficiency and reproducibility, buffers in a peak increases, decreasing the effects on band spreading
the range of 5 - 50 mM concentration or additives in the 0.1%- caused by components of the instrument.
0.5% range are used the mobile phase. Phosphate buffers are
Sample Solvent – For isocratic separations, the sample
not generally recommended because of their poor solubility in should be dissolved in the mobile phase or in a solvent
high organic mobile phases and incompatibility with MS that is weaker (less water) than the mobile phase. For
detection, although phosphoric acid at low concentrations can gradient separations, the sample should be dissolved in
be employed, as long as sufficient water is present in the the initial mobile phase or in a solvent substantially
mobile phase. Additives such as formic acid, trifluoroacetic weaker than the final mobile phase.
acid and acetic acid at concentrations up to about 1% can be a
Injection Volume – For isocratic separations, the volume
part of the mobile phase. HILIC separations can exhibit high
of sample injected should be kept as small as possible
load tolerance and efficiency, but with dependence on
(typically 2 µl or less). Sample volumes are less critical
buffering capacity of the mobile phase. Injection from a
for gradient separations, especially if the sample is
suitable organic/water mixture is a must, when injection
dissolved in a weak solvent.
volume becomes significant. Ammonium formate/formic acid
buffers up to a final concentration of about 50 mM and pH 3-5
are especially effective for analysis of basic and acidic Ordering Information
compounds using MS detection with either ESI or APCI For ordering information or for technical support on this
interfaces. (This volatile mobile phase buffer seems to be a product, in the U.S. call MAC-MOD Analytical at 1-
good starting point for many separations of both basic and 800-441-7508 For information in other countries go to
acidic compounds.) Ammonium acetate at pH 5-8 also have the website: www.advanced-materials-tech.com
been used at concentrations of 1-20 mM, but are generally less
favored for separating stronger basic and acidic compounds. HALO and Fused-Core are trademarks of Advanced Materials Technology, Inc.
Buffers or additives above pH 9-10 usually are not
recommended, particularly at elevated temperature because of
slow dissolution of the underlying silica support.
PFP Column
Care & Use Sheet
HALO Description Operation Guidelines
HALO PFP is a high-speed, high-performance liquid The direction of flow is marked on the column label.
chromatography column based on a new fused core Reversed flow may be used to attempt removal of
particle design. The fused core particle provides a thin inlet pluggage or contamination.
porous shell of high-purity silica surrounding a solid A new column contains a mixture of acetonitrile and
silica core. This particle design exhibits very high water. Initial care should be taken to avoid mobile
column efficiency due to the shallow diffusion paths in phases that are immiscible with this mixture.
the 0.5-micron thick porous shell and the small overall Water and all common organic solvents are
particle size of 2.7-microns. The bonded, endcapped, compatible with HALO PFP columns.
dimethylpentafluorophenylpropyl stationary phase of Halo PFP columns are best used at temperatures
HALO PFP provides a stable, reversed phase packing below 60ºC for maximum column life.
with electron-deficient phenyl rings due to the presence Mobile phase pH for HALO PFP columns is best
of electronegative fluorines. In addition to forming • -• maintained in the range of pH = 2 to 8 for maximum
and mildly steric interactions, PFP phases also retain column stability.
compounds by polar interactions. As a result of having HALO PFP columns are stable to operating
both polar and non-polar character, PFP phases can show pressures up to 600 bar (9000 psi).
dual-mode retention behavior, sometimes producing a
“U-shaped” retention as a function of acetonitrile content
of the mobile phase, with retention increasing at both low Column Care
and high concentrations of ACN (reversed-phase and To maximize column life, ensure that samples and
HILIC retention modes). HALO PFP can be used for mobile phases are particle-free. The use of guard
basic, acidic, or neutral compounds with alternate columns or an in-line filter with 0.5-micron porosity
selectivity from C18. between the sample injector and the column is highly
recommended. The 2-micron porosity frits on HALO
PFP columns are less subject to pluggage than are the
Column Characteristics 0.5-micron frits typically used with other small-particle
Figure 1 shows a sample Quality Control test columns. Should the operating pressure of the column
chromatogram for a 4.6 mm x 50 mm column. A printed suddenly increase beyond normal levels, reversing the
report including the actual test chromatogram and flow direction of the column may be attempted to
performance results is enclosed with every column.
remove debris on the inlet frit.
The fused core particle has a surface area of ~ 150 m2/g To remove strongly retained materials from the column,
and an average pore size of 90Å. The fused core particles
flush the column in the reverse direction with very
are 30% to 50% heavier than commercially available
strong solvents such as 100% of the organic component
totally porous particles due to the density of the solid of the mobile phase in use. A mixture (95/5 v/v) of
cores. Therefore, the effective surface area per column is dichloromethane and methanol is often effective at this
similar to columns packed with totally porous particles task. Extreme cases may require the use of very strong
having surface areas in the 225-300 m2/g range. solvents such as dimethylformamide (DMF) or
dimethylsulfoxide (DMSO).
FIGURE 1 - PFP QC Chromatogram
3
Test Conditions:
Column Storage
Column: 4.6 x 50 mm HALO PFP
Mobile Phase: 60/40 ACN/H2O
Flow: 1.8 ml/min
Long-term storage of silica-based, reversed-phase
1 2
4
Temperature: Ambient (~24ºC)
Detection: UV at 254 nm
columns is best in 100% acetonitrile. Columns may be
Sample Volume: 1 µL
safely stored for short periods (up to 3 or 4 days) in most
Peaks:
common mobile phases. However, when using buffers,
1) uracil 30 µg/ml
2) phenol 500 µg/ml it is best to protect both the column and the HPLC
3) 4-chloro-nitrobenzene
120 µg/ml
4) naphthalene 200 µg/ml
equipment to remove the salts by flushing the column
with the same mobile phase without the buffer (e.g.,
when using 60/40 ACN/buffer, flush the column with
60/40 ACN/H2O) to eliminate potential corrosion from
the salts while providing rapid re-equilibration of the
column with the original mobile phase.
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Time, minutes
Before storing the column, the end-fittings should be system having components designed to minimize band
tightly sealed with the end-plugs that came with the spreading. All low-volume columns perform best when
column to prevent the packing from drying. used with proper attention to the following factors:
Detector – Flow cells should be of low-volume
design (preferably < 2µl).
Safety Detector – To properly sense and integrate the often
HPLC columns are for laboratory use only. Not very fast peaks that elute from low-volume columns,
for drug, household, or other use. the detector response time should be set to the
Users of HPLC columns should be aware of the fastest level (~ 0.1 second) and the integration
toxicity or flammability of the mobile phases chosen software should sample the detector signal at least
for use with the columns. Precautions should be taken 20 points per second.
to avoid contact and leaks. Injector – The injection system should be of a low-
HPLC columns should be used in well-ventilated volume design (e.g., Rheodyne Model 8125). Auto-
environments to minimize concentration of solvent samplers will often cause band-spreading with low-
fumes. volume columns but may be used for convenience
with the expectation of some loss in column
Applications efficiency.
The HALO PFP columns are best utilized with mobile Connecting Tubing – The shortest possible lengths
phases that are mixtures of methanol and water or of connecting tubing with narrow internal diameters
acetonitrile and water. Higher levels of the organic (at most 0.005-inch, 0.12 mm ID) should be used to
solvent component will typically reduce the retention of connect the column to the injector and the detector
the sample compounds. Using elevated temperatures cell. The tubing must have flat ends and should
(e.g., 40 – 60 ºC) will reduce the viscosity of the mobile bottom out inside all fittings. Zero-dead-volume
phase and allow the use of faster flowrates and lower fittings should always be used where required.
column pressure for high sample throughput. Gradient- Peak Retention – As retention is increased, the
elution techniques using 5 -10% organic component as volume of a peak increases, decreasing the effects
the initial mobile phase and increasing to 100% organic on band spreading caused by components of the
component as the final mobile phase often can effect instrument.
separations of complex sample mixtures in minimal time. Sample Solvent – For isocratic separations, the
Ionizable compounds, such as acids and bases, are sample should be dissolved in the mobile phase or in
generally best separated with mobile phases buffered at a solvent that is weaker (less polar) than the mobile
pH of 2 to 3. The use of 20-50 mM buffers is always phase. For gradient separations, the sample should
recommended for optimum results and long-term stability be dissolved in the initial mobile phase or in a
when separating ionizable compounds. Additional solvent substantially weaker than the final mobile
information on solvent selection and separation phase.
techniques can be found in Chapters Six, Seven, and Injection Volume – For isocratic separations, the
Eight, Practical HPLC Method Development, Second volume of sample injected should be kept as small
Edition, L.R. Snyder, J.L. Glajch, and J.J. Kirkland, (John as possible (typically 2 µl or less). Sample volumes
Wiley & Sons, 1997). are less critical for gradient separations, , especially
if the sample is dissolved in a weak solvent.
Guidelines for Low-Volume Columns
High performance columns with small internal volumes
(shorter lengths, internal diameters < 3 mm) are being
increasingly used for high speed separations, especially
with specialty detection systems such as mass
spectrometers. These low-volume columns generate
peaks having considerably less volume than those eluting
from columns of larger dimensions (e.g., 4.6 mm x 150
mm). The efficiency of separations performed in low-
volume columns is highly dependent on the HPLC
Phenyl-Hexyl Column
Care & Use Sheet
HALO-5 Description Water and all common organic solvents are
HALO-5 Phenyl-Hexyl is a high-speed, high-
compatible with HALO-5 Phenyl-Hexyl columns.
performance liquid chromatography column based on the
HALO-5 Phenyl-Hexyl columns are best used at
highly efficient Fused-Core® particle design. The Fused-
temperatures below 60 ºC for maximum column life.
Core particle provides a thin porous shell of high-purity
Mobile phase pH for HALO-5 Phenyl-Hexyl
silica surrounding a solid silica core. This particle design
columns is best maintained in the range of pH = 2 to
exhibits very high column efficiency due to the shallow
9 for maximum column stability.
diffusion paths in the 0.6-micron thick porous shell and
the highly uniform overall particle size of 5-microns. HALO-5 Phenyl-Hexyl columns are stable to
The bonded, endcapped, dimethylphenylhexyl stationary operating pressures up to 600 bar (9000 psi).
phase of HALO-5 Phenyl-Hexyl provides a stable,
reversed phase packing with enhanced • -• and mild steric Column Care
interactions due to the relatively long-chain hexyl linkage To maximize column life, ensure that samples and
to the base silica. HALO-5 Phenyl-Hexyl can be used for mobile phases are particle-free. The use of guard
basic, acidic, or neutral compounds with alternate columns or an in-line filter with 0.5-micron porosity
selectivity from C18. between the sample injector and the column is highly
recommended. The 2-micron porosity frits on HALO-5
Column Characteristics Phenyl-Hexyl columns are less subject to pluggage than
Figure 1 shows a sample Quality Control test are the 0.5-micron frits typically used with 3-micron
chromatogram for a 4.6 mm x 50 mm column. A printed particle columns. Should the operating pressure of the
report including the actual test chromatogram and column suddenly increase beyond normal levels,
performance results is enclosed with every column. reversing the flow direction of the column may be
The 5-micron Fused-Core particle has a surface area of attempted to remove debris on the inlet frit.
~ 100 m2/g and an average pore size of 90Å. The Fused- To remove strongly retained materials from the column,
Core particles are 30% to 50% heavier than commercially flush the column in the reverse direction with very
available totally porous particles due to the density of the strong solvents such as 100% of the organic component
solid cores. Therefore, the effective surface area per of the mobile phase in use. A mixture (95/5 v/v) of
column is similar to columns packed with totally porous dichloromethane and methanol is often effective at this
particles having surface areas in the 160-200 m2/g range. task. Extreme cases may require the use of very strong
solvents such as dimethylformamide (DMF) or
dimethylsulfoxide (DMSO).

Column Storage
Long-term storage of silica-based, reversed-phase
columns is best in 100% acetonitrile. Columns may be
safely stored for short periods (up to 3 or 4 days) in most
common mobile phases. However, when using buffers,
it is best to protect both the column and the HPLC
equipment and remove the salts by flushing the column
with the same mobile phase without the buffer (e.g.,
when using 60/40 ACN/buffer, flush the column with
60/40 ACN/H2O) to eliminate any danger from
corrosion from the salts while providing rapid re-
Operation Guidelines equilibration of the column with the original mobile
The direction of flow is marked on the column label. phase.
Reversed flow may be used to attempt removal of Before storing the column, the end-fittings should be
inlet pluggage or contamination. tightly sealed with the end-plugs that came with the
A new column contains a mixture of acetonitrile and column to prevent the packing from drying.
water. Initial care should be taken to avoid mobile
phases that are immiscible with this mixture or could
cause a precipitate.
spreading. All low-volume columns perform best when
Safety used with proper attention to the following factors:
HPLC columns are for laboratory use only. Not Detector – Flow cells should be of low-volume
for drug, household, or other use. design (preferably < 2µl).
Users of HPLC columns should be aware of the
Detector – To properly sense and integrate the often
toxicity or flammability of the mobile phases chosen
very fast peaks that elute from low-volume columns,
for use with the columns. Precautions should be taken
the detector response time should be set to the
to avoid contact and leaks.
fastest level (~ 0.1 second) and the integration
HPLC columns should be used in well-ventilated software should sample the detector signal at least
environments to minimize concentration of solvent 20 points per second.
fumes.
Injector – The injection system should be of a low-
volume design (e.g., Rheodyne Model 8125). Auto-
Applications samplers will often cause band spreading with low-
The HALO-5 Phenyl-Hexyl bonded phase is nonpolar in volume columns but may be used for convenience
nature. It is best utilized with mobile phases that are with the expectation of some loss in column
mixtures of methanol and water or acetonitrile and water. efficiency.
Higher levels of the organic solvent component will Connecting Tubing – The shortest possible lengths
typically reduce the retention of the sample compounds. of connecting tubing with narrow internal diameters
Using elevated temperatures (e.g., 40 – 60 ºC) will reduce (at most 0.005-inch, 0.12 mm ID) should be used to
the viscosity of the mobile phase and allow the use of connect the column to the injector and the detector
faster flow rates and lower column pressure for high cell. The tubing must have flat ends and should
sample throughput. Gradient elution techniques using 5 - bottom out inside all fittings. Zero-dead-volume
10% organic component as the initial mobile phase and fittings should always be used where required.
increasing to 100% organic component as the final Peak Retention – As retention is increased, the
mobile phase often can effect separations of complex volume of a peak increases, decreasing the effects
sample mixtures in minimal time. on band spreading caused by components of the
HALO-5 Phenyl-Hexyl columns are highly suited for the instrument.
reversed-phase separation of basic, neutral, or acidic Sample Solvent – For isocratic separations, the
compounds. HALO-5 Phenyl-Hexyl columns will have sample should be dissolved in the mobile phase or in
similar retention to C8 phases but different selectivity for a solvent that is weaker (more polar) than the mobile
some compounds compared to alkane phases. Many phase. For gradient separations, the sample should
aromatic compounds interact more strongly with the be dissolved in the initial mobile phase or in a
phenyl phase compared to C18/C8. Ionizable compounds, solvent substantially weaker than the final mobile
such as acids and bases, are generally best separated with phase.
mobile phases buffered at pH of 2 to 3. The use of 20-50 Injection Volume – For isocratic separations, the
mM buffers is always recommended for optimum results volume of sample injected should be kept as small
and long-term stability when separating ionizable
as possible (typically 2 µl or less). Sample volumes
compounds. Additional information on solvent selection are less critical for gradient separations, especially
and separation techniques can be found in Chapters Six, if the sample is dissolved in a weak solvent.
Seven, and Eight, Practical HPLC Method Development,
Second Edition, L.R. Snyder, J.L. Glajch, and J.J.
Kirkland, (John Wiley & Sons, 1997). Ordering Information
For ordering information or for technical support on this
product, in the U.S. call MAC-MOD Analytical at 1-
Guidelines for Low-Volume Columns 800-441-7508. For information in other countries go to
High performance columns with small internal volumes the website: www.advanced-materials-tech.com
(shorter lengths, internal diameters < 3 mm) are being
increasingly used for high speed separations, especially
with specialty detection systems such as mass
spectrometers. These low-volume columns generate HALO and Fused-Core are registered trademarks of
peaks having considerably less volume than those eluting Advanced Materials Technology, Inc.
from columns of larger dimensions (e.g., 4.6 mm x 150
mm). The efficiency of separations performed in low-
volume columns is highly dependent on the HPLC
system having components designed to minimize band
RP-Amide Column
Care & Use Sheet
HALO-5 RP-Amide Description
HALO-5 RP-Amide is a high-speed, high-performance Water and all common organic solvents are
liquid chromatography column based on a new Fused- compatible with HALO-5 RP-Amide columns.
Core® particle design. The Fused-Core particle provides HALO-5 RP-Amide columns are best used at
a thin porous shell of high-purity silica surrounding a temperatures below 60ºC for maximum column life.
solid silica core. This particle design exhibits very high Mobile phase pH for HALO-5 RP-Amide columns is
column efficiency due to the shallow diffusion paths in best maintained in the range of pH = 2 to 9 for
the 0.6-micron thick porous shell and the highly uniform maximum column stability.
overall particle size of 5-microns. The one-step bonded, HALO-5 RP-Amide columns are stable to operating
endcapped, amide-based, polar-embedded stationary pressures up to 600 bar (9000 psi).
phase of HALO-5 RP-Amide provides a stable, reversed The polar-embedded HALO-5 RP-Amide columns
phase packing with decreased hydrophobic character. are compatible with 100% aqueous mobile phases
HALO-5 RP-Amide can be used for basic, acidic, or without suffering from “phase collapse”.
neutral compounds with orthogonal selectivity from C18
or C8. Column Care
To maximize column life, ensure that samples and
Column Characteristics mobile phases are particle-free. The use of guard
Figure 1 shows a sample Quality Control test columns or an in-line filter with 0.5-micron porosity
chromatogram for a 4.6 mm x 50 mm column. A printed between the sample injector and the column is highly
report including the actual test chromatogram and recommended. The 2-micron porosity frits on HALO-5
performance results is enclosed with every column. RP-Amide columns are less subject to pluggage than are
The Fused-Core particle has a surface area of ~ 100 m2/g the 0.5-micron frits typically used with 3-micron particle
and an average pore size of 90Å. The Fused-Core columns. Should the operating pressure of the column
particles are 30% to 50% heavier than commercially suddenly increase beyond normal levels, reversing the
available totally porous particles due to the density of the flow direction of the column may be attempted to
solid cores. Therefore, the effective surface area per remove debris on the inlet frit.
column is similar to columns packed with totally porous To remove strongly retained materials from the column,
particles having surface areas in the 160-200 m2/g range. flush the column in the reverse direction with very
strong solvents such as 100% of the organic component
Figure 1 - HALO-5 RP-Amide QC Chromatogram of the mobile phase in use. A mixture (95/5 v/v) of
120
dichloromethane and methanol is often effective at this
task. Extreme cases may require the use of very strong
TEST CONDITIONS
Column: 4.6 x 50 mm HALO-5 RP-Amide
Mobile Phas e: 60% ACN
100
3 Flow: 1.0 mL/min
Temperature: 30 C
40% Water
o
solvents such as dimethylformamide (DMF) or
80
1
2
Detec tor: UV (254 nm)
Sample Volume: 1 µL
Peak Identities :
dimethylsulfoxide (DMSO).
1. Urac il 30 µg/mL
2. Phenol 500 µg/mL
mAU

3. 4-Chloro-1-nitrobenz ene 120 µg/mL


60 4. Naphthalene 200 µg/mL
Column Storage
40
4
Long-term storage of silica-based, reversed-phase
columns is best in 100% acetonitrile. Columns may be
20
safely stored for short periods (up to 3 or 4 days) in most
0 common mobile phases. However, when using buffers,
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Time, minutes it is best to protect both the column and the HPLC
equipment to remove the salts by flushing the column
with the same mobile phase without the buffer (e.g.,
Operation Guidelines when using 60/40 ACN/buffer, flush the column with
The direction of flow is marked on the column label. 60/40 ACN/H2O) to eliminate any danger from
Reversed flow may be used to attempt removal of corrosion from the salts while providing rapid re-
inlet pluggage or contamination. equilibration of the column with the original mobile
phase.
A new column contains a mixture of acetonitrile and
Before storing the column, the end-fittings should be
water. Initial care should be taken to avoid mobile
tightly sealed with the end-plugs that came with the
phases that are immiscible with this mixture or could
column to prevent the packing from drying.
cause a precipitate.
Safety system having components designed to minimize band
HPLC columns are for laboratory use only. Not spreading. All low-volume columns perform best when
for drug, household, or other use. used with proper attention to the following factors:
Users of HPLC columns should be aware of the Detector – Flow cells should be of low-volume
toxicity or flammability of the mobile phases chosen design (preferably < 2µl).
for use with the columns. Precautions should be taken Detector – To properly sense and integrate the often
to avoid contact and leaks. very fast peaks that elute from low-volume columns,
HPLC columns should be used in well-ventilated the detector response time should be set to the
environments to minimize concentration of solvent fastest level (~ 0.1 second) and the integration
fumes. software should sample the detector signal at least
20 points per second.
Injector – The injection system should be of a low-
Applications volume design (e.g., Rheodyne Model 8125). Auto-
The HALO-5 RP-Amide columns are best utilized with
samplers will often cause band-spreading with low-
mobile phases that are mixtures of methanol and water or
volume columns but may be used for convenience
acetonitrile and water. Higher levels of the organic
with the expectation of some loss in column
solvent component will typically reduce the retention of
efficiency.
the sample compounds. Using elevated temperatures
Connecting Tubing – The shortest possible lengths
(e.g., 40 – 60 ºC) will reduce the viscosity of the mobile
of connecting tubing with narrow internal diameters
phase and allow the use of faster flowrates and lower
(at most 0.005-inch, 0.12 mm ID) should be used to
column pressure for high sample throughput. Gradient-
connect the column to the injector and the detector
elution techniques using 5 -10% organic component as
cell. The tubing must have flat ends and should
the initial mobile phase and increasing to 100% organic
bottom out inside all fittings. Zero-dead-volume
component as the final mobile phase often can effect
fittings should always be used where required.
separations of complex sample mixtures in minimal time.
Ionizable compounds, such as acids and bases, are Peak Retention – As retention is increased, the
generally best separated with mobile phases buffered at volume of a peak increases, decreasing the effects
pH of 2 to 3. The use of 20-50 mM buffers is always on band spreading caused by components of the
recommended for optimum results and long-term stability instrument.
when separating ionizable compounds. Additional Sample Solvent – For isocratic separations, the
information on solvent selection and separation sample should be dissolved in the mobile phase or in
techniques can be found in Chapters Six, Seven, and a solvent that is weaker (more polar) than the mobile
Eight, Practical HPLC Method Development, Second phase. For gradient separations, the sample should
Edition, L.R. Snyder, J.L. Glajch, and J.J. Kirkland, (John be dissolved in the initial mobile phase or in a
Wiley & Sons, 1997). solvent substantially weaker than the final mobile
phase.
HALO-5 RP-Amide columns will have similar retention Injection Volume – For isocratic separations, the
but different selectivity for some compounds compared to volume of sample injected should be kept as small
alkane phases. Phenols and anilines interact more as possible (typically 2 µl or less). Sample volumes
strongly with the amide phase compared to C18/C8. are less critical for gradient separations, , especially
Typically, bases move to shorter retentions compared to if the sample is dissolved in a weak solvent.
C18 columns. Non-polar solutes show little differences.
Ordering Information
For ordering information or for technical support on this
Guidelines for Low-Volume Columns product, in the U.S. call 1-800-441-7508 For
High performance columns with small internal volumes information in other countries go to the website:
(shorter lengths, internal diameters < 3 mm) are being www.advanced-materials-tech.com
increasingly used for high speed separations, especially
with specialty detection systems such as mass HALO and Fused-Core are registered trademarks of
spectrometers. These low-volume columns generate Advanced Materials Technology, Inc.
peaks having considerably less volume than those eluting
from columns of larger dimensions (e.g., 4.6 mm x 150
mm). The efficiency of separations performed in low-
volume columns is highly dependent on the HPLC

Potrebbero piacerti anche