A comparison of International (IEC) and


American (NFPA) power supply systems for
healthcare-facilities

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 1
Standards for electrical installations in medical locations
World map - dated 1994

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 2
Standards for electrical installations in medical locations
World map - dated 2007

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 3
IEC vs. NFPA – The background

IEC NFPA

IEC stands for: NFPA stands for:

International Electro technical Commission National Fire Protection Association


Geneva, CH Quincy, MA, USA

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 4
IEC vs. NFPA – The standards

IEC NFPA
Relevant Standards for medical applications: Relevant Standards for medical applications:

 IEC 60364-7-710 Electrical Safety for  NFPA 70 National Electrical Code


Hospitals and …  NFPA 70B Practice for Electrical
 DIN VDE 0100-710 Electrical Power Supply  Equipment Maintenance
in medical Applications  NFPA 70E Electrical Safety in the
 IEC 60364-4-41 Protection for Safety and Workplace
against Electrical Shock  NFPA 99 Health Care Facilities
 IEC 60364-5-54 EarthingArrangements  UL 1022 Standard for Safety Line
Isolation Monitors
 IEC 60479-1 Effect of current on
 CSA Z32.2 Electrical Safety and
human beings / livestock Essential Electrical
Systems in Health Care
Facilities
 NEC Article 516 Hazardous location
 UL 1047 Isolated Power Systems
Equipment
 AS 3003 Hospital wiring rules

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 5
IEC vs. NFPA – The measuring principle

IEC NFPA

The IEC-Standard requests the The NFPA-standard requests the


measurement of insulation-resistance measurement of resistive, capacitive or hybrid
between an unearthed system and earth: fault currents to ground. This information will
provide an estimation when:
a) At all times (continuously) in kOhm
a) At all times (continuously) in mA
b) Earth fault location is optional

The A-Isometer (absolute) monitors the The LIM (Line Insulation Monitor) detects the
insulation-resistance (in Ohm) and triggers a fault, calculates the Total Hazard Current
visual and audible alarm when the (THC), which could possibly flow under
resistance, the total load-amount or the extreme circumstances and initiates a visual
isolation-transformer-temperature are critical. and audible alarm.
In addition fault location can be provided.

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 6
IEC vs. NFPA – The system
IEC NFPA

The A-Isometer is continuously measuring the The Line Insulation Monitor (LIM) is continuously
insulation-resistance of an unearthed AC- or DC- monitoring the leakage impedance in an
Network and will trigger a visible and audible alarm ungrounded AC-Network. Based on this
at a preset threshold value. It will display the information the maximum Total Hazard Current
amount of insulation-resistance in Ohm. It also (THC) is calculated and displayed. The LIM will
monitors the load- current and the transformer- trigger a visible and audible alarm at a preset
temperature. Earth fault location is optional to threshold (2 or 5 mA).
identify a faulty branch circuit.

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 7
IEC vs. NFPA – The monitors

IEC NFPA

The A-Isometer is connected to an unearthed The THC is displayed on a seven-


system 230VAC. When the resistance-level segment-display and a bar graph of the
falls below 50 kOhm (in industrial LIM. The THC-level will increase as
applications adjustable between 1 kOhm and additional loads are connected to the
10 MOhm) the resistive fault current to be a system as they contribute to the system
maximum of 4,6mA. The normal level of leakage capacitance. The LIM will trigger
insulation-resistance of an IT-system is > 10 an alarm when the THC exceeds 2 mA
MOhm. A current of 5 mA (NFPA) and 30 mA (Canada) or 5 mA (USA) which was
has been defined as not critical for human determined as the critical mark for fire-
health, even under critical conditions. security. The THC is also reduced due to
The normal kVA-rating of an IEC-IT-System reduced kVA-ratings of the IPS (Isolated
is between 3,15 and 8 kVA but IEC has also Power Systems).
allowed 10 kVA since 2007. Smaller systems
with 1,6, 2,0 and 2,5 kVA are compulsory
required in e.g. the Netherlands.

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 8
IEC vs. NFPA – The isolation transformers

IEC NFPA

The IEC-transformers pay attention to a The transformers as per NFPA require a


relatively low inrush current (max of 12 very low capacitive leakage current,
times IN peak), thus allowing a longer resulting in a very low maximal distance
cable distance (for example to install the between transformer and LIM (60 Hz-
transformer in the false ceiling) as well as Noise !!)
a much better short-circuit discrimination.

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 9
IEC vs. NFPA – The remotes

IEC NFPA

The IEC-remotes offer (in the MK2430- and The NFPA-remotes are housed in stainless
the MK800-version) text-displays which can steel, offer (in the new DRI) digital digits for
be programmed according to the e.g. room- the channel and the THC and also the
number and the number of the IT-system in minimum required LED’s and buttons.
addition to the minimum required LED’s and
buttons. The MK7-version offers the
minimum required LED’s and buttons.

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 10
NFPA-Solutions

Compact Isolated Power Systems (IPS) and accessories

Line Isolation Monitor (LIM)

Power and grounding


modules

Isolated power panel (IPP)


Surgical Facility Center
(SFC)
Multiple Isolated Power Panel

© Bender GmbH & Co. KG · 35305 Grünberg · Germany · www.bender-de.com · F xxxx / 200x · rights reserved to modification · Folie Nr. 11
IEC Solutions

Compact Isolated Power Systems (IPS) and accessories

Isolated Power System


(IPS) – wall mounted

Transfer switch
remote indicator

Isolation transformer

Surgical Facility Panels (SFP)

IPS remote indicator


Isolated Power System
(IPS) floor standing Touch screen display