Sei sulla pagina 1di 5

Principles of X-Ray Imaging

Already a few weeks after the discovery of X-rays in 1895 3. The X-rays are attenuated differently by the various body
by Wilhelm Conrad R€ ontgen the first medical images with tissues.
photographic plates and fluorescent screens were made. This 4. Scattered radiation, which impairs image contrast, is
was the origin of projection radiography and fluoroscopy. reduced.
The greatest steps forward in X-ray diagnostic radiology 5. The transmitted photons are detected.
since Roentgen’s observations were the development of the 6. The image is processed and – in the case of CT –
image intensifier systems and then above all the announce- reconstructed.
ment of computed tomography (CT) in a clinical environment This makes it possible to discuss the aspects of image
by Hounsfield at the 1972 British Institute of Radiology quality and radiation exposure for both systems together in
annual conference. A further important step was the introduc- the main parts of the book (cf. Chap. 2).
tion of digital image receptors in projection radiography In radiography/fluoroscopy with digital image receptors
during the last years. Compared to conventional film-screen and in computed tomography the digital image consists of a
systems these receptors allow the separate optimisation of (typically square) matrix of picture elements (pixels) which
photon detection and image processing, resulting in signifi- represent the corresponding volume elements (voxels) and –
cant advantages for image quality and dose. after the exposure – carry the local intensity information (gray
Although today projection radiography is still the most scale value). Quality of digital images depends primarily on
frequent examination with X-rays the use of computed tomo- the image matrix size, i.e. the pixel size (cf. Chap. 9). As the
graphy increases rapidly, and – because it involves larger matrix size is increased resolution improves but the number
radiation doses than the conventional imaging procedures of photons in each pixel must be increased in order to maintain
(cf. Table 10.1) – contributes significantly to the annual a certain minimum noise level.
collective dose (see Fig. 1.1). Therefore CT also obtains
growing attention in radiation protection (Brenner and
Hall 2007). 1.1 Projection Radiography
In X-ray diagnostic radiology the image is generated by and Fluoroscopy
the interaction of X-ray photons, which have transmitted the
patient, with a photon detector. These photons can either be In projection radiography and fluoroscopy the image is a
primary photons, which have passed through the tissue with- two-dimensional projection of the attenuating properties of
out interacting, or secondary photons, which result from an all the tissues along the paths of the X-rays. The components
interaction along their path through the patient. The second- of a typical radiographic/fluoroscopic system are shown in
ary photons will in general be deflected from their original Fig. 1.3.
direction and result in scattered radiation. The photons emitted by the X-ray tube are collimated by
The basic principles of projection radiography/fluoroscopy a beam-limiting device. Then they enter the patient, where
and CT are shortly explained in Sects. 1.1 and 1.2 respec- they may be scattered, absorbed or transmitted without
tively. Although totally different in image character, both interaction. The primary photons recorded by the image
imaging systems have in common certain features, which receptor form the image. The secondary photons create a
can be recognised in Fig. 1.2: certain amount of background radiation which degrades
1. X-rays are produced in an X-ray tube. contrast. If necessary, the majority of the scattered photons
2. The energy distribution of the photons is modified by can be removed by placing an anti-scatter device between
inherent and additional filtration. the patient and the image receptor. This device can simply be

H. Aichinger et al., Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology, 3
DOI 10.1007/978-3-642-11241-6_1, # Springer-Verlag Berlin Heidelberg 2012
4 1 Principles of X-Ray Imaging

Remainder Dental
CT 0.7% 0.2%
Angiography and 7%
intervention Remainder 1% 3% Thorax
2%
Mammography 9%
4% Skeleton
GI and urogen
3% Dental and bile tract
GI and urogen 8%
37%
and bile tract
1%
Mammo-
graphy

CT 18%
Skeleton 60% Angiography
33% and
intervention
Thorax
13%

Fig. 1.1 Contribution of various examination types to total frequency (left) and to collective effective dose (right) in 2006 for Germany adapted
from BMU (2009)

X-ray tube
x
Collimation

Production of X-rays

Filtration

Object transmission

Patient
Scatter reduction

Photon detection

Patient support
Anti-scatter device
AEC system
Image reconstruction and processing
Image receptor

Fig. 1.2 Basic principles of radiography/fluoroscopy and CT imaging Fig. 1.3 Typical arrangement of a radiography/fluoroscopy system
1.2 Computed Tomography 5

an air gap or a so-called anti-scatter grid formed from a reconstructed as image signal by computation. In practice
series of parallel metal strips. An automatic exposure control CT numbers or Hounsfield units are used instead of mtissue
system (AEC) provides for the correct exposure of the image where the Hounsfield unit HU is defined by:
receptor. Today digital image receptors predominate in radi-
ography and fluoroscopy, but film-screen systems and image ðmtissue  mwater Þ
HU ¼ 1000  (1.1)
intensifiers are also still in use. mwater

where mwater is the linear attenuation coefficient of water.


1.2 Computed Tomography The experimental set-up of Hounsfield corresponded
largely to the arrangement sketched in Fig. 1.4.
Whereas it is not possible in projection radiography to This set-up was termed the ‘first generation’ of CT
gain any depth information from a single image, computed (Kalender 2006). To speed up scanning and to utilise the
tomography separates the superimposed anatomical details available X-ray power more efficiently the first commercial
and produces sectional or axial slice images with excellent scanners (the ‘second generation’) used some more detectors
soft tissue contrast. Compared to projection radiography and and a small fan beam. The typical scan time for an 80  80
fluoroscopy computed tomography is a rather new imaging image matrix was 5 min (Kalender 2006).
technique. Therefore it seems to be reasonable to present its Continuously rotating CT systems (‘third generation’)
fundamental principles in some more detail. according to Fig. 1.5 with a fan beam covering the total patient
The principle of computed tomography is illustrated in cross-section and a corresponding detector array, consisting
Fig. 1.4. of gas proportional detectors or scintillation detectors (cf.
A well-collimated X-ray pencil beam is attenuated by Sect. 8.2), were introduced in the 1980s. Continuous rotation
the tissues along its path and the transmitted radiation is was made possible by a slip-ring technology for electrical
detected. In order to generate one projection the tube-detector power supply and data acquisition. Scan time was reduced
assembly scans the object in a linear translatory motion. This down to 2 s for a single slice with a 256  256 matrix.
procedure is repeated at many viewing angles (typically at A major step forward in CT technology was the introduc-
least 180 projections are received with a rotational increment tion of spiral or helical CT by Kalender and Vock in 1989
of 1 ). From these projections a two-dimensional discrete (Kalender et al. 1989; Vock et al. 1989): Slice-by-slice
distribution of the linear attenuation coefficients mtissue is imaging was replaced by volume scanning. The principle

X-ray tube

Detector

Fig. 1.4 Principle of data acquisition in CT imaging (Adapted from Fig. 1.5 Continuously rotating CT system with a fan beam and
Bunke 2003) corresponding detector array (Adapted from Bunke 2003)
6 1 Principles of X-Ray Imaging

Fig. 1.6 Principle of spiral CT


imaging (From Bunke 2003)

of this method is illustrated in Fig. 1.6: While the fan beam is [ mm ] 5 2.5 1.5 1 1 1.5 2.5 5
continuously rotating the patient is moved with constant
velocity along his body axis (the z-axis) through the gantry; 2×8
this results in a spiral track of the focal spot around the
patient and accordingly in a spiral data set. 4×5
Direct image reconstruction from these data would give
rise to image artefacts (similar to motion artefacts). This 4 × 2.5
can mostly be avoided by data interpolation. The interpolation
method developed at first was the 360 linear interpolation 4×1
(LI) algorithm, which used data from a full rotation of the
tube-detector assembly. Since for a complete interpolated data 2 × 0.5
set at a definite slice position two successive 360 rotations on
either side of the selected plane were necessary, considerable
Fig. 1.7 Adaptive array detector with detector combinations for
widening of the slice profile resulted, thus reducing image different slice thicknesses (from Bunke 2003), e.g. the uppermost
quality. Therefore the 360 LI was soon replaced by a 180 combination allows slice widths in the longitudinal direction from
LI where interpolation from opposing 180 points reduces the 1 to 5 mm at the isocentre
spiral range used for reconstruction. This is possible since
X-ray beam attenuation at a distinct rotation angle j is equiv-
alent to the X-ray beam attenuation traversing the body from The MSCT detector arrays could be divided into two
the opposite side, at 180 + j. As the distance of the data groups: Those with detector elements of unequal width along
points is now smaller, effective slice width will be less. the z-axis (adaptive array detector) and those with elements of
In 1992 CT scanners were introduced, which used two equal width (linear or matrix detector). Figure 1.7 shows as an
parallel banks of detectors. This was followed by multirow example an adaptive array detector with the possibility of the
detector CT scanning in 1998 using solid detectors and simul- setting of different slice thicknesses.
taneously imaging four slices in each rotation of the X-ray tube At present 64-slice scanning represents the state of the art,
(Kalender 2006). A great advantage of multislice CT (MSCT) allowing the imaging of all body regions with submilli-
scanners over single section spiral CT is the opportunity for metre isotropic spatial resolution and scan times of 5–15 s
longer anatomic coverage during the same scanning time. (Kalender 2006).
References 7

Scan time can be further reduced with recent developments Kalender WA (2006) X-ray computed tomography. Phys Med Biol 51:
of CT such as dual source CT or cone beam CT with C-arm R29–R43
Kalender WA, Seissler W, Vock P (1989) Single-breath-hold spiral
systems (Kalender 2006). This is especially interesting for volumetric CT by continuous patient translation and scanner rota-
cardiac imaging, angiography and interventions. Dual source tion. Radiology 173:414
CT scanners are equipped with an ultrafast dual detector BMU (Bundesministerium f€ ur Umwelt, Naturschutz und Reaktor-
system and two X-ray tube assemblies. Cone beam CT (see sicherheit) (2009) Umweltradioaktiuit€at und Strahlenbelastung im
Jahr 2008: Unterrichtung durch die Bundesregierung http://nbn-
Sect. 8.2) uses a flat-panel-detector with up to 1,920 rows resolving.de/urn:nbn:de:0221-201003311019
and 2,480 columns (Oppelt 2005). This enables enhanced use Oppelt A (ed) (2005) Imaging systems for medical diagnostics.
of X-ray quanta, but also leads to a higher fraction of scattered Publicis, Erlangen
radiation (see Sect. 11.2.4). Vock P, Jung H, Kalender WA (1989) Single breathhold spiral volu-
metric CT of the lung. Radiology 173:400

References
Brenner DJ, Hall EJ (2007) Computed tomography – an increasing
source of radiation exposure. N Engl J Med 357:2277–2284
Bunke J (2003) Computertomographie. In: Schmidt T (ed)
Strahlenphysik Strahlenbiologie Strahlenschutz. Springer, Berlin,
pp 84–98

Potrebbero piacerti anche