General Model ZR22G, ZR402G, and ZR202G

Specifications Direct In Situ Zirconia Oxygen Analyzers and


High Temperature Humidity Analyzers
GS 11M12A01-01E

Overview
This analyzer consists basically of a probe and a
converter that are used as both a Zirconia Oxygen
Analyzer and High Temperature Humidity Analyzer.
The probe is of direct insertion type, and the converter
uses a digital display.
Two types of analyzers are available: separate
type and integrated type. As its name implies, the
integrated type combines probe and converter.
Separate and integrated type Zirconia oxygen
analyzers need not use a sampling device, and allow
direct installation of the probe in the wall of a flue
or furnace to measure the concentration of oxygen
in the stack gas. The converter displays the cell
temperature and cell emf in addition to the oxygen
concentration.
This analyzer is most suitable for monitoring the
oxygen concentration of combustion gases in large
or small boilers, various industrial furnace and
combustion devices, or for the control of low-oxygen
combustion.
Separate type and integrated type Zirconia High
Temperature Humidity Analyzers are used to measure
the humidity of hot gases continuously in driers which
use an electrical heater or hot gas as the heat source.
They can also be used in a variety of manufacturing
applications in humidifiers, as well as in driers, for
humidity measurement and control. They can help
improve productivity in these application fields.

Features:
• The built-in heater assembly of the probe can be • This converter can be used as an oxygen analyzer as
replaced on site, reducing maintenance costs. well as a high temperature humidity analyzer.
• The probe uses a long-life, high-reliability Zirconia • The integrated type integrates both probe and
sensor. converter, to reduce wiring, piping, and installation
• The probe uses three-reference gas supply methods costs. This type of unit uses an optical switch for
(natural air convection, instrument air, and pressure ease of operation at the site.
compensated) in its applications. • Remote maintenance using digital communications
• The separate type converter incorporates a LCD (HART) reduces maintenance costs. *1
touchscreen for ease of operation.
*1: HART is a registered trademark of HART
Communication Foundation
Control room
ZR202G 4 to 20 mA DC signal line with
superimposed HART AC signal
Relaying Terminal board
Ao+
+ terminals


Ao-

Receiving instrument
load resistance:
EXA ZR402G

250 Ω to 550 Ω

HART HART
Ao1+ HART
ZR402G Communicator Communicator
Ao1- Communicator
Model 275 Model 275
Model 275 F34.ai

Yokogawa Electric Corporation GS 11M12A01-01E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright June 2000
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 14th Edition Oct. 2019
2
Basic System Configuration
System configuration - Separate type

System configuration Example 1 of Separate type Analyzer


Automatic calibration system uses instrument air for reference gas.
For the calibration gas, a standard gas cylinder may be used for more accurate calibration.
Applications: Oxygen concentration monitoring and control in large boilers
(for private power generation and for business use) and in heating furnaces, and the like.
Humidity monitoring and control in drying furnaces and humidifiers.

ZR22G Separate type Zirconia Oxygen/High ZR402G Converter


Temperature Humidity Analyzer, Detector EXA ZR402G

*2

*1
Signal
(6-core shield cable)
Check valve ~ 100 to 240 V AC
Contact input
Heater (2-core cable)
Analog output, contact output
Flowmeter Needle Digital output (HART)
valve
Reference gas Air Set

Calibration gas
Instrument air

Pressure
ZR40H Automatic Calibration unit regulator
*3 Zero gas cylinder

Calibration gas
unit case F03.ai

System configuration - Integrated type

System configuration Example 1 of Integrated type Analyzer

Automatic calibration system uses instrument air for reference gas.


For the calibration gas, a standard gas cylinder may be used for more accurate calibration.
Applications: Oxygen concentration monitoring and control in large boilers
(for private power generation and for business use) and in heating furnaces, and the like.
Humidity monitoring and control in drying furnaces and humidifiers.

*2 ZR202G Integrated type Zirconia Oxygen/High


Temperature Humidity Analyzer
with automatic calibration (ZR202G-□□□-□-□-A-□-□-□-□-A)

Automatic Calibration Unit ~ 100 to 240 V AC


Contact input
ZR20H
Analog output, contact output
Digital output (HART)
Air Set
Reference gas
Instrument air
Span gas
Calibration gas (Zero)

Pressure
regulator

*3 Zero gas cylinder

Calibration gas
unit case
Note:
The installation temperature limits range for integrated type analyzer is -20 to 55°C.
F05.ai

*1 Shield cable:
Use shielded signal cables, and connect the shields to the FG terminal of the converter.
*2 Select the desired probe from the Probe Configuration table on page 4.
*3 When a zirconia oxygen analyzer is used, 100% N2 gas cannot be used as the zero gas. Use approx.
1 vol% O2 gas (N2-balanced).

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
3
Basic System Configuration
System configuration — Separate type System configuration — Integrated type
System configuration Example 2 of Separate type Analyzer System configuration Example 2 of Integrated type Analyzer
Instrument air is used as the reference gas. A standard gas Instrument air is used as the reference gas. A standard gas
cylinder can be used for the calibration gas for
cylinder can be used for the calibration gas for more accurate
more accurate calibration.
calibration.
Application example: Oxygen concentration monitoring and Application example: Oxygen concentration monitoring and
control in large boilers (for private power generation and
control in large boilers (for private power generation and for
for business use) and in heating furnaces.
Humidity monitoring and control in drying furnaces business use) and in heating furnaces.
and humidifiers. Humidity monitoring and control in drying furnaces and
ZR402G Converter
humidifiers.
ZR22G Separate type Zirconia Oxygen/High
Temperature Humidity Analyzer, Detector
EXA ZR402G

ZR202G Intgrated type Zirconia Oxygen/High


Temperature Humidity Analyzer

Stop valve
or
*1
~ 100 to 240 V AC
Stop valve
or
~ 100 to 240 V AC
Contact input
Check valve ZA8F Flow Setting Unit
Contact input Check valve Analog output, contact output
Analog output, contact output
Needle Digital output (HART) Digital output (HART)
Reference Flowmeter valve
gas Air Set
Calibration
ZA8F Flow Setting Unit
gas Instrument air Needle
Flowmeter valve
Span gas (*4) Reference gas Air Set

Calibration gas
Pressure
Instrument air
regulator Zero gas Span gas (*4)
cylinder

Calibration gas
unit case
F04.ai
*1 Shield cable: Pressure
Use shielded signal cables, and connect the shields to the FG terminal of the converter.
regulator
Zero gas
*4 Calibration gas unit same as for zero gas. cylinder

Calibration gas unit case


F04_01.ai

System configuration Example 3 of Separate type Analyzer


System configuration Example 3 of Integrated type Analyzer
Ambient air is used as the reference gas. A portable
standard gas unit (ZO21S) is used for the calibration. Ambient air is used as the reference gas. A portable
This unit is connected only when the calibration is made. standard gas unit (ZO21S) is used for the calibration.
Application example: This unit is connected only when the calibration is made.
Oxygen concentration monitoring and control in
packaged boilers
Application example:
Humidity monitoring and control in drying furnaces or Oxygen concentraion monitoring and control in
a humidifiers packaged boilers
ZR402G Converter
ZR22G Separate type Zirconia Oxygen/High
Temperature Humidity Analyzer, Detector
Humidity monitoring and control in drying furnaces or
a humidifiers
EXA ZR402G

Stop valve *1
100 to 240 V AC ZR202G Intgrated type Zirconia Oxygen/High
~ Temperature Humidity Analyzer
Contact input
Analog output, contact output
Digital output (HART)

Calibration gas
Stop valve
~ 100 to 240 V AC
Contact input
100/110/115 Analog output, contact output
~ 200/220/240 V AC Digital output (HART)
ZO21S Standard gas unit
*1 Shield cable:
Use shielded signal cables, and connect the shields to the FG terminal of the converter.
F02_01.ai
Calibration gas

~ 100/110/115
200/220/240 V AC
ZO21S Standard gas unit
F02_02.ai

Note: The installation temperature limits range for integrated type


analyzer is -20 to 55°C.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
4
System Components

Separate type Integrated type


System config. System config.
System Components
Ex.1 Ex.2 Ex.3 Ex.1 Ex.2 Ex.3
1 Model ZR22G Separate type Zirconia Oxygen / High Temperature Humidity Analyzers,Detector
2 Model ZR402G Separate type Zirconia Oxygen / High Temperature Humidity Analyzer, Converter(*1)
3 Model ZR202G Integrated type Zirconia Oxygen / High Temperature Humidity Analyzers
4 Model ZO21P High Temperature Probe Adapter for separate type Zirconia Oxygen Analyzer
5 E7046EC, E7046EN Auxiliary Ejector Assembly for High Temperature Probe of separate type Oxygen Analyzer
6 Model ZO21R Probe Protector for Zirconia Oxygen Analyzers
7 K9471UA Dust Filter for Oxygen Analyzer
8 K9471UC Dust Guard Protector
9 Model ZH21B Dust Protector for High Temperature Humidity Analyzers
10 Model ZO21S Standard Gas Unit
11 Model ZA8F Flow Setting Unit for manual calibration
12 Model ZR40H Automatic Calibration Unit for Separate type Analyzers
13 Model ZR20H Automatic Calibration Unit for Integrated type Analyzers (*2)
14 L9852CB, G7016XH Stop Valve for Calibration gas line (*3) ( ) ( )
15 K9292DN, K9292DS Check Valve for Calibration gas line (*3) ( ) ( )
16 G7003XF/K9473XK, G7004XF/K9473XG Air Set
17 G7001ZC Zero gas Cylinder
18 G7013XF, G7014XF Pressure Regulator for Gas Cylinder
19 E7044KF Case Assembly for Calibration gas Cylinder
20 ZR22A, ZR202A Heater Assembly for Spare Parts
: Items required for the above system example
: To be selected depending on each application. For details, refer to Chapter of Options. T01.ai
( ) : Select either
(*1): When used as a high temperature humidity analyzer, specify /HS options.
(*2): When Automatic Calibration of (-A) or (-B) code is speified , ZR20H is installed in ZR202G.
(*3): When ZR22G or ZR202G specifies Stop valve (/SV) or Check valve (/CV) as an option code, they are correspondingly
installed in the equipment.

Detector Components

Sample gas temperature 0 to 700°C Sample gas temperature 700 to 1400˚C


Insertion General-use Probe Application High temperature detector Application
Mounting length

Boiler Sample Heating


Horizontal 0.4 outlet
Detector Heating Absorption furnace
to to (ZR22G or furnace structure

vertical 2m ZR202G) High temperature detector


Probe adapter
for high
temperature use
2.5 m ZO21P-H
Vertical
or more
Sample inlet

Detector Temperature:
Probe Protector (ZR22G or For pulverized Probe material; SUS310S 800˚C
Gas Flow (ZO21R) ZR202G) coal boiler Probe material; SiC 1400˚C
Horizontal with gas flow Mounting: Vertical downwards
3m velocity Insertion length: 1.0 m, 1.5 m
to
or less 10 m/sec or more When duct pressure is atmospheric
vertical Sample inlet
or negative, attach air ejector.
High temperature auxiliary
ejector assembly (E7046EC, E7046EN)

Pressure gauge
Dust filter for Black liquid
Horizontal 0.4 Oxygen Analyzer recovery boiler
to Needle
to (K9471UA)
Cement Kiln valve
vertical 2m Detector(ZR22G or
or ZR202G) Inlet Ejector
Dust guard
protector +
2.5 m
Vertical (K9471UC)
or more Blow

F06.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
5

■STANDARD SPECIFICATIONS Response Time: Response of 90% within 5 seconds.


(Measured after gas is introduced from
(Oxygen Analyzer) calibration gas inlet and analog output
Example of Application starts changing.)
Separate and integrated type Zirconia Oxygen Safety, EMC and RoHS conformity standards for the
Analyzers ZR22G, ZR402G and ZR202G
Installation altitude based on IEC 61010: 2000 m or less
• Large, medium and small boilers (boilers for power Category based on IEC 61010: II (Note)
generation: heavy oil, gas or coal) Pollution degree based on IEC 61010: 2 (Note)
• Various industrial furnaces (refinery process/iron Note: Installation category, called over-voltage category,
manufacture heating furnace, coal kiln, and black liquid specifies impulse withstand voltage.
recovery boilers) For other applications, contact Category II is for electrical equipment.
Yokogawa Electric Corporation. Pollution degree indicates the degree of existence
• May not be applicable corrosive gas such as of solid, liquid, gas or other inclusions which may
ammonia, chlorine is present-check wit reduce dielectric strength. Degree 2 is the normal
YOKOGAWA. indoor environment.
Safety: EN 61010-1, EN 61010-2-030,
General Specifications CAN/CSA-C22.2 No. 61010-1,
UL Std. No. 61010-1
Oxygen Analyzer EMC: EN 61326-1 Class A, Table 2,
Measurement Object: Oxygen concentration in EN 61326-2-3, EN 61000-3-2
combustion exhaust gas and mixed gas E
 MC Regulatory Arrangement in
(excluding inflammable gases may not Australia and New Zealand (RCM)
be applicable corrosive gas such as EN61326-1 Class A
ammonia, chlorine is present-check with  orea Electromagnetic Conformity Standard
K
YOKOGAWA. Note: This instrument is a Class A product, and it is
Measurement System: Zirconia system designed for use in the industrial environment.
Measurement Range: 0.01 to 100 vol% O2 Please use this instrument in the industrial
Output Signal: 4 to 20 mA DC (maximum load environment only.
resistance 550 Ω) RoHS: EN 50581
Setting Range: Any setting in the range of 0 to 5 Information of the WEEE Directive
through 0 to 100 vol% O2 (in 1 vol% O2), This product is purposely designed to be used in a
or partial range large scale fixed installations only and, therefore,
Digital Communication (HART): 250 to 550 Ω, is out of scope of the WEEE Directive. The WEEE
depending on number of field devices Directive does not apply.
connected to the loop (multi-drop mode). The WEEE Directive is only valid in the EU.
Display Range: 0 to 100 vol% O2
Warm-up Time: Approx. 20 min. 1. ZR22G Separate type Zirconia Oxygen
Repeatability: (Excluding the case where the Analyzer, Detector
reference gas is by natural convection) Oxygen Analyzer
± 0.5% Maximum value of set range; Sample Gas Temperature: 0 to 700°C (Probe only)
range from 0 to 5 vol% O2 or more and It is necessary to mount the cell using
less than 0 to 25 vol% O2 range Inconel cell-bolts when the temperature is
± 1% Maximum value of set range; greater than 600°C.
range from 0 to 25 vol% O2 or more and For high temperature sample gas (700
up to 0 to 100 vol% O2 range to 1400°C), apply 0.15 m length probe
Linearity: (Excluding standard gas tolerance) and High Temperature Probe Adapter
(Excluding the case where the reference ZO21P-H.
gas is by natural convection) Sample Gas Pressure: - 5 to + 250 kPa (When the
(Use oxygen of known concentration (with pressure in the furnace exceeds 3 kPa,
in the measuring range) as the zero and it is recommended to use pressure
span calibration gases.) compensated type. When the pressure
± 1% Maximum value of set range; 0 to in the furnace exceeds 5 kPa, pressure
5 vol% O2 or more and less than 0 to 25 compensated type is required.)
vol% O2 range For 0.15 m probe, - 0.5 to + 5 kPa. No pressure
(Sample gas pressure: within ± 4.9 kPa) fluctuation in the furnace should be allowed.
± 3% Maximum value of set range; 0 to Note: When the detector is used in conjunction with a
25 vol% O2 or more and less than 0 to 50 check valve and the ZA8F Flow Setting Unit, the
vol% O2 range maximum pressure of sample gas is 150 kPa.
(Sample gas pressure: within ± 0.49 kPa) When with a check valve and the ZR40H
± 5% Maximum value of set range; 0 to Automatic Calibration Unit, it is 200 kPa. If the
50 vol% O2 or more and up to 0 to 100 pressure of your sample gas exceeds these limits,
vol% O2 range consult with Yokogawa.
(Sample gas pressure: within ± 0.49 kPa) Probe Length: 0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0,
Drift: (Excluding the first two weeks in use) 3.6, 4.2, 4.8, 5.4 m
Probe Material: SUS316 (JIS)
(Excluding the case where the reference Ambient Temperature: -20 to +150°C
gas is by natural convection.) Reference Gas System: Natural Convection,
Both zero and span ± 2% Maximum value Instrument Air, Pressure compensated
of set range/month (other than for probe length 0.15 m)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
6
Instrument Air System (excluding Natural Convection): Storage Temperature: -30 to +70°C
Pressure; 200 kPa + the pressure inside Ambient Humidity: 0 to 95% RH (non-condensing)
the furnace. (It is recommended to use air Power Supply Voltage: Ratings; 100 to 240 V AC
which has been dehumidified by cooling Acceptable range; 85 to 264 V AC
to dew point -20°C or less, and dust Power Supply Frequency: Ratings; 50/60 Hz
or oil mist are removed.) Acceptable range; 45 to 66 Hz
Consumption; Approx. 1 Nl/min Power Consumption: Max. 300 W, approx. 100 W for
Wetted Material: SUS316 (JIS), Zirconia, SUS304 ordinary use.
(JIS) or ASTM grade 304 (flange), Maximum Distance between Detector and Converter:
Hastelloy B, (Inconel 600, 601) Conductor two-way resistance must be
Construction: Heater and thermocouple replaceable
10 Ω or less (when a 1.25 mm2 cable or
construction. Non explosion-proof JIS
C0920 / equivalent to IP44D. Equivalent equivalent is used, 300 m or less.)
to NEMA 4X/IP66 (Achieved when the Construction: Equivalent to NEMA 4X/IP66 (with
cable entry is completely sealed with a conduit holes completely sealed with a
cable gland in the recirculation pressure cable gland)
compensated version.) Wiring Connection: G1/2, Pg 13.5, M20 × 1.5, 1/2
Terminal Box Case: Material; Aluminum alloy NPT (with plug), eight holes
Terminal Box Paint Color: Case; Mint green (Munsell Installation: Panel, wall or 2-inch pipe mounting
5.6BG3.3/2.9) Case: Aluminum alloy
Cover; Mint green (Munsell 5.6BG3.3/2.9) Paint Color: Door: Silver gray (Munsell 3.2PB7.4/1.2)
Finish: Polyurethane corrosion-resistance coating Case: Silver gray (Munsell 3.2PB7.4/1.2)
Gas Connection: Rc1/4 or 1/4 NPT (Female) Finish: Polyurethane corrosion-resistance coating
Wiring Connection: G1/2, Pg 13.5, M20 × 1.5, Weight: Approx. 6 kg
1/2 NPT
Installation: Flange mounting
Probe Mounting Angle: Horizontal to vertically Functions
downward. Display Functions:
When the probe insertion length is Value Display; Displays values of the measured
2 m or less, installing at angles from oxygen concentration, etc
horizontal to vertically downward is Graph Display; Displays trends of measured
possible. oxygen concentration
When the probe insertion length is 2.5 Data Display; Displays various useful data for
m or more, mount vertically downward maintenance, such as cell temperature,
(within ±5°) and use a probe protector. reference junction temperature,
Weight: maximum/minimum oxygen
Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65) concentration, or the like
/ approx. 11 kg (ANSI 150 4) Status Message; Indicates an alarm or error
Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65) occurrence by flashing of the
/ approx. 13 kg (ANSI 150 4) corresponding icon. Indicates status
Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65) such as warming-up, calibrating, or the
/ approx. 15 kg (ANSI 150 4) like by the marks.
Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65) Alarm, Error Display; Displays alarms such as
/ approx. 17 kg (ANSI 150 4) “Abnormal oxygen concentration” or
Insertion length of 3.0 m: approx. 15 kg (JIS 5K 65) errors such as “Abnormal cell e.m.f.”
/ approx. 20 kg (ANSI 150 4) when any such status occurs.
Insertion length of 3.6 m: approx. 17 kg (JIS 5K 65) Calibration Functions:
/ approx. 22 kg (ANSI 150 4) Automatic Calibration; Requires the ZR40H
Insertion length of 4.2 m: approx. 19 kg (JIS 5K 65) Automatic Calibration Unit. It calibrates
/ approx. 24 kg (ANSI 150 4) automatically at specified intervals.
Insertion length of 4.8 m: approx. 21 kg (JIS 5K 65) Semi-automatic Calibration; Requires the
/ approx. 26 kg (ANSI 150 4) ZR40H Automatic Calibration Unit.
Insertion length of 5.4 m: approx. 23 kg (JIS 5K 65) Input calibration direction on the
/ approx. 28 kg (ANSI 150 4) touchscreen or contact, then it calibrates
automatically afterwards.
2. ZR402G Separate type Zirconia Oxygen Manual Calibration; Calibration with opening/closing
Analyzer, Converter the valve of calibration gas in operation
interactively with an LCD touchscreen.
Oxygen Analyzer Blowback Function:
Operated using an LCD touchscreen on the converter. Output through the contact in the set
Display: LCD display of size 320 by 240 dot with period and time. Auto/Semi_Auto
touchscreen. selectable.
Output Signal: 4 to 20 mA DC, two points (maximum Maintenance Functions:
load resistance 550 Ω) Can operate updated data settings in
Contact Output Signal: Four points (one is fail-safe, daily operation and checking. Display
normally open) data settings, calibration data settings,
Contact Input: Two points blowback data settings, current output
Automatic Calibration Output: Two points (for loop check, input/output contact check.
dedicated automatic calibration unit)
Ambient Temperature: -20 to +55°C

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
7
Setup Functions: Range: Any setting between 0 to 5 through 0
Initial settings suit for the plant conditions to 100 vol% O2 in 1 vol% O2, or partial
when installing the converter. Equipment range is available (Maximum range
settings, current output data settings, value/minimum range value 1.3 or more)
alarm data settings, contact data settings, For the log output, the minimum range
other settings. value is fi xed at 0.1 vol% O2.
Self-diagnosis: 4 to 20 mA DC linear or log can be
This function diagnoses conditions of the selected. Input/output isolation.
converter or the detector and indicates
when any abnormal condition occurs.
Password Functions: Output damping: 0 to 255 seconds. Hold/non-hold
Enter your password to operate the selection, preset value setting possible
analyzer excepting data display. Individual with hold.
passwords can be set for maintenance Contact Output: Four points, contact capacity 30 V
and setup. DC 3 A, 250 V AC 3 A (resistive load).
Display and setting content: Three of the output points can be selected
Measuring Related Items: Oxygen concentration to either normally energized or normally
(vol% O2), output current value (mA), air deenergized status.
ratio, moisture quantity (in hot gases) Delayed functions (0 to 255 seconds)
(vol% H2O) and hysteresis function (0 to 9.9 vol%O2)
Display Items: Cell temperature (°C), thermocouple can be added to high/low alarms.
reference junction temperature (°C), The following functions are programmable
maximum/minimum/average oxygen for contact outputs.
concentration (vol% O2), cell e.m.f. (mV), (1) Abnormal, (2) High-high alarm, (3)
cell internal resistance (Ω), cell condition High alarm, (4) Low-low alarm, (5) Low
(in four grades), heater on-time rate alarm, (6) Maintenance, (7) Calibration,
(%), calibration record (ten times), time (8) Range switching answer-back, (9)
(year/month/day, hour/minute) Warm-up, (10) Calibration gas pressure
Calibration Setting Items: Span gas concentration decrease (answer-back of contact
(vol% O2), zero gas concentration (vol% input), (11) Temperature high alarm,
O2), calibration mode (automatic, semi- (12) Blowback start, (13) Flameout gas
automatic, manual), calibration type detection (answer-back of contact input),
and method (zero-span calibration, zero (14) Calibration coefficient alarm, (15)
calibration only, span calibration only), Startup power stabilization timeout alarm
stabilization time (min. sec), calibration Contact output 4 is set to normally
time (min. sec), calibration interval (day/ operated, and fixed error status.
hour), starting time (year/month/day, Contact Input: Two points, voltage-free contacts.
hour/minute) The following functions are programmable
Equipment Related Items: Measuring gas selection for contact inputs:
Output Related Items: Analog output/output mode (1) Calibration gas pressure decrease
selection, output conditions when alarm, (2) Range switching, (3) External
warming-up/maintenance/calibrating calibration start, (4) Process alarm (if this
(during blowback)/abnormal, oxygen signal is received, the heater power turns
concentration at 4 mA/20 mA (vol% off), (5) Blowback start
O2), time constant. Contact capacity: Off-state leakage current; 3 mA or
Alarm Related Items: Oxygen concentration high less
alarm/ high-high alarm limit values (vol% Self-diagnosis: Abnormal cell, abnormal cell
O2), oxygen concentration low alarm/ temperature (low/high), abnormal
low-low alarm limit values (vol% O2), calibration, defective A/D converter,
oxygen concentration alarm hysteresis defective digital circuit
(vol% O2), oxygen concentration alarm Calibration: Method; Zero/span calibration
detection, alarm delay (seconds) Calibration mode; automatic, semi-automatic and
Contact Related Items: Selection of contact input 1 manual (All are operated interactively
and 2, selection of contact output 1 to 4 with an LCD touchscreen). Either zero
(abnormal, high-high alarm, high alarm, or span can be skipped.
low alarm, low-low alarm, maintenance, Zero calibration gas concentration setting range;
calibrating, range switching, 0.3 to 100 vol% O2 (minimum setting:
warming-up, 0.01 vol% O2).
calibration gas pressure decrease, Span calibration gas concentration setting range;
temperature high alarm, blowback, 4.5 to 100 vol% O2 (minimum setting :
fl ameout gas detection, calibration 0.01 vol% O2).
coefficient alarm, stabilization timeout.) Use N2-balanced mixed gas containing
Converter Output: Two points mA analog output (4 0 to 10% scale of oxygen, and 80 to 100
to20 mA DC (maximum load resistance % scale of oxygen for standard zero gas
of 550Ω)) and one of two mA outputs and standard span gas respectively.
is with digital output (HART) (minimum Calibration interval; date/time setting:
load resistance of 250 Ω). maximum 255 days

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
8
3. ZR202G Integrated type Zirconia Oxygen Gas Connection: Rc1/4 or 1/4 NPT(Female)
Analyzer Wiring Connection: G1/2, Pg 13.5, M20 × 1.5,
1/2 NPT select one type (4 pieces)
Oxygen Analyzer Installation: Flange mounting
Can be operated in the field without opening the cover Probe Mounting Angle: Horizontal to vertically
using optical switches. downward. When the probe insertion
Display: 6-digit LCD length is 2 m or less, installing at angles
Switch: Three optical switches from horizontal to vertical downward is
Output Signal: 4 to 20 mA DC, one point (maximum available.
load resistance 550 Ω) When the probe insertion length is 2.5 m or
Digital Communication (HART): 250 to 550 Ω, more, mount vertically downward (within
depending on number of field devices ± 5°) and use a probe protector.
connected to the loop (multi-drop mode). Case: Aluminum alloy
Contact Output Signal: Two points (one is fail-safe, Paint Color:Cover; Mint green (Munsell
normally open) 5.6BG3.3/2.9)
Contact Input Signal: Two points Case; Mint green (Munsell 5.6BG3.3/2.9)
Sample Gas Temperature: 0 to 700°C Finish: Polyurethane corrosion-resistance coating
It is necessary to mount the cell using Weight:
inconel cell-bolts when the temperature is Insertion length of 0.4 m: approx. 8 kg (JIS 5K 65) /
greater than 600°C. approx. 13 kg (ANSI 150 4)
High temperature service - greater than Insertion length of 1.0 m: approx. 10 kg (JIS 5K 65)
700°C - is not available. / approx. 15 kg (ANSI 150 4)
Sample Gas Pressure: - 5 to + 250 kPa (When the Insertion length of 1.5 m: approx. 12 kg (JIS 5K 65)
pressure in the furnace exceeds 3 kPa, / approx. 17 kg (ANSI 150 4)
it is recommended to use pressure Insertion length of 2.0 m: approx. 14 kg (JIS 5K 65)
compensated type. When the pressure / approx. 19 kg (ANSI 150 4)
in the furnace exceeds 5 kPa, pressure Insertion length of 3.0 m: approx. 17 kg (JIS 5K 65)
compensated type is required.) / approx. 22 kg (ANSI 150 4)
No pressure fluctuation in the furnace
should be allowed. Functions
Note: When the detector is used in conjunction with
Display Function: Displays values of the measured
a check valve and the ZA8F Flow Setting Unit,
the maximum pressure of sample gas is 150 oxygen concentration, etc.
kPa. When with a check valve and the ZR20H Alarm, Error Display: Displays alarms such as “AL-
Automatic Calibration Unit, it is 200 kPa. If the 06” or errors such as “Err -01” when any
pressure of your sample gas exceeds these limits, such status occurs.
consult with Yokogawa. Calibration Functions:
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m Automatic Calibration; Requires the ZR20H
Probe Material: JIS SUS316 stainless steel Automatic Calibration Unit. It calibrates
Ambient Temperature: -20 to +55°C (- 5 to +70°C on automatically at specified intervals.
the case surface) Semi-automatic Calibration; Requires the ZR20H
Storage Temperature: -30 to +70°C Automatic Calibration Unit. Input
Ambient Humidity: 0 to 95 %RH (non-condensing) calibration start signal by optical
Power Supply Voltage: Ratings; 100 to 240 V AC switch or contact, then it calibrates
Acceptable range; 85 to 264 V AC automatically afterwards.
Power Supply Frequency: Ratings; 50/60 Hz Manual Calibration; Calibration with opening/closing
Acceptable range; 45 to 66 Hz the valve of calibration gas in operation
Power Consumption: Max. 300 W, approx. 100 W for interactively with the optical switch.
ordinary use. Maintenance Functions:
Reference Gas System: Natural Convection, Can operate updated data settings in
Instrument Air, or Pressure Compensated daily operation and checking. Display data
Instrument Air System (excluding Natural Convection): settings, calibration data settings, test
Pressure; 200 kPa plus the pressure settings (current output loop check, input/
inside the furnace (It is recommended to output contact check).
use air which is dehumidified by cooling Setup Functions:
to dew point -20°C or less, and dust or oil Initial settings suit for the plant conditions
mist are removed.) when installing the converter. Current
Consumption; Approx. 1 Nl/min output data settings, alarm data settings,
Wetted Material: SUS316 (JIS), Zirconia, SUS304 contact data settings, other settings.
(JIS) or ASTM grade 304 (flange),
Hastelloy B, (Inconel 600, 601)
Construction: Heater and thermocouple replaceable
construction. Non explosion proof JIS
C0920 / equivalent to IP44D. Equivalent
to NEMA 4X/IP66 (Achieved when the
cable entry is completely sealed with a
cable gland in the recirculation pressure
compensated version.)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
9
Display and setting content: Contact Output: Two points, contact capacity 30 V
Display Related Items: Oxygen concentration DC 3 A, 250 V AC 3 A (resistive load)
(vol% O2), output current value (mA), One of the output points can be selected
air ratio, moisture quantity (in hot to ether normally energized or normally
gases) (vol% H2O), Cell temperature de-energized status.
(°C ), thermocouple reference junction Delayed functions (0 to 255 seconds)
temperature (°C ), maximum/minimum/ and hysteresis function (0 to 9.9 vol% O2)
average oxygen concentration (vol% can be added to high/low alarms.
O2), cell e.m.f. (mV), cell internal The following functions are programmable
resistance (Ω), cell condition (in four for contact outputs.
grades), heater on-time rate (%), (1) Abnormal, (2) High-high alarm,
calibration record (ten times), time (year/ (3) High alarm, (4) Low-low alarm,
month/day/hour/minute) (5) Low alarm, (6) Maintenance,
Calibration Setting Items: Span gas concentration (7) Calibration, (8) Range switching
(vol% O2), zero gas concentration (vol% answer-back, (9) Warmup,
O2), calibration mode (automatic, semi- (10) Calibration-gas pressure decrease
automatic, manual), calibration type (answer-back of contact input),
and method (zero-span calibration, zero (11) Flameout gas detection (answer-back
calibration only, span calibration only), of contact input).
stabilization time (min.sec), calibration Contact Input: Two points, voltage-free contacts
time (min.sec), calibration interval (day/ The following functions are programmable
hour), starting time (year/month/day/ for contact inputs.
hour/minute) (1) Calibration gas pressure decrease
Output Related Items: Analog output/output mode alarm, (2) Range switching (switched
selection, output conditions when range is fixed), (3) External calibration
warming- up/maintenance/calibrating/ start, (4) Process alarm (if this signal is
abnormal, oxygen concentration at 4 received, the heater power turns off)
mA/20 mA (vol% O2), time constant, Self-diagnosis: Abnormal cell, abnormal cell
preset values when warming-up/ temperature (low/high), abnormal
maintenance/calibrating/abnormal, calibration, defective A/D converter,
output preset values on abnormal defective digital circuit
Alarm Related Items: Oxygen concentration high Calibration: Method; Zero/span calibration
alarm/ high-high alarm limit values (vol% Calibration mode; Automatic, semi-automatic and
O2), oxygen concentration low alarm/ manual (All are operated using optical
low-low alarm limit values (vol% O2), switches). Either zero or span can be
oxygen concentration alarm hysteresis skipped.
(vol% O2), oxygen concentration alarm Zero calibration gas concentration setting range;
detection, alarm delay (seconds) 0.3 to 100 vol% O2 (minimum setting:
Contact Related Items: Selection of contact input 0.01 vol% O2).
1 and 2, selection of contact output Span calibration gas concentration setting range;
1 and 2 (abnormal, high-high alarm, 4.5 to 100 vol% O2 (minimum setting:
high alarm, low alarm, low-low alarm, 0.01 vol% O2).
maintenance, calibrating, range Use N2-balanced mixed gas containing 0
switching, warming-up, calibration to10% scale of oxygen for standard zero
gas pressure decrease, flameout gas gas, and 80 to 100% scale of oxygen for
detection (answer-back of contact input) standard span gas.
Converter Output: One mA analog output point (4 Calibration interval; date/time setting:
to 20 mA DC (maximum load resistance maximum 255 days
of 550 Ω)) with mA digital output point
(HART) (minimum load resistance of 250
Ω).
Range: Any setting between 0 to 5 through 0 to
100 vol% O2 in 1 vol% O2, and partial
range is available (Maximum range value/
minimum range value 1.3 or more)
For the log output, the minimum range
value is fixed at 0.1 vol% O2.
4 to 20 mA DC linear or log can be
selected. Input/output isolation.
Output damping; 0 to 255 seconds.
Hold/non-hold selection, preset value
setting possible with hold.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
10

■STANDARD SPECIFICATIONS (High Response Time: Response of 90% within 5 seconds.


(Measured after gas is introduced from
Temperature Humidity Analyzer) calibration gas inlet and analog output
starts changing.)
Examples of Application (Note 1) These tolerances do not apply to the pressure
Separate/Integrated type Zirconia High Temperature compensated version, or where natural convection
Humidity Analyzer is used for the reference gas.
Safety, EMC and RoHS conformity standards for the
• Coloring processes in the textile industry
ZR402G and ZR202G
• Steam curing processes for concrete products
• Manufacturing processes in the cigarette, food, paper Installation altitude based on IEC 61010: 2000 m or less
or pulp industries Category based on IEC 61010: II (Note)
• Drying processes in various manufacturing of building Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category,
materials, lumber, plasterboard, food or the like specifies impulse withstand voltage.
• Humidifying processes in various manufacturing of Category II is for electrical equipment.
food or the like Pollution degree indicates the degree of existence
Please contact us for other applications. of solid, liquid, gas or other inclusions which may
reduce dielectric strength. Degree 2 is the normal
General Specifications indoor environment.
Safety: EN 61010-1, EN 61010-2-030,
High Temperature Humidity Analyzer CAN/CSA-C22.2 No. 61010-1,
Oxygen concentration in mixed gas which consists of UL Std. No. 61010-1
water vapor and air is proportional to the volumetric EMC: EN 61326-1 Class A, Table 2 (For use in
ratio of oxygen in the air, so the volumetric ratio industrial locations)
of water vapor can be calculated from the oxygen EN 61326-2-3, EN 61000-3-2
concentration. EMC Regulatory Arrangement in
Measurement Object: Water vapor (in vol%) in mixed Australia and New Zealand
gases (air and water vapor) Korea Electromagnetic Conformity
Measurement System: Zirconia system Standard
Measurement Range: 0.01 to 100 vol% O2 ,0 to 100 Note: This instrument is a Class A product, and it is
vol% H2O or 0 to 1.000 kg/kg designed for use in the industrial environment.
Output Signal: 4 to 20 mA DC (maximum load Please use this instrument in the industrial
resistance 550 Ω) environment only.
Setting Range: Any setting in the range of 0 to 5 RoHS: EN 50581
through 0 to 100 vol% O2 (in 1 vol% O2), Information of the WEEE Directive
or partial range. This product is purposely designed to be used in a
Moisture quantity; 0 to 25 through 0 to large scale fixed installations only and, therefore,
100 vol% H2O (in 1 vol% H2O), or partial is out of scope of the WEEE Directive. The WEEE
range. Directive does not apply.
Mixture ratio; 0 to 0.2 through 0 to 1.000 The WEEE Directive is only valid in the EU.
kg/kg (in 0.001 kg/kg), or partial range.
Digital Communication (HART): 250 to 550 Ω, 1. ZR22G Separate type Zirconia High
depending on number of field devices Temperature Humidity Analyzer,Detector
connected to the loop (multi-drop mode).
Display Range: Oxygen concentration; 0 to 100 vol% High Temperature Humidity Analyzer
O2 , Sample Gas Temperature: 0 to 700°C (Probe only)
Moisture quantity; 0 to 100 vol% H2O It is recommended to mount the cell using
Mixture ratio; 0 to 1 kg/kg inconel cell-bolts when the temperature is
Relative humidity; 0 to 100% RH (Note) greater than 600°C.
Dew point; -40 to 370°C (Note) Sample Gas Pressure: - 5 to + 20 kPa (When the
Note: These values are affected by temperature and pressure in the furnace exceeds 3 kPa,
absolute pressure, So accurate temperature and it is recommended to use pressure
pressure values must be inputted to the converter. compensated type. When the pressure
Warm-up Time: Approx. 20 min. in the furnace exceeds 5 kPa, pressure
These characteristics are calculated by oxygen compensated type is required.)
concentration measured in air which include water No pressure fluctuation in the process
vapor. should be allowed.
Repeatability: (see Note 1) Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
± 1 vol% H2O; (Sample gas pressure 2 Probe Material: JIS SUS316 stainless steel
kPa or less) Ambient Temperature: -20 to +150°C
Linearity: (Excluding standard gas tolerance) Reference Gas System: Natural Convection,
(see Note 1), (Use oxygen of known Instrument Air,or Pressure compensated
concentration (in the measuring range) as Instrument Air System (excluding Natural Convection):
the zero and span calibration gas.) Pressure; 200 kPa plus the pressure
± 2 vol% H2O; (Sample gas pressure: inside the furnace. (It is recommended
within ± 0.49 kPa) to use air which has been dehumidified
± 3 vol% H2O; (Sample gas pressure: by cooling to dew point - 20°C or less,
2 kPa or less) and dust or oil mist are removed.)
Drift: (Excluding the first two weeks in use) Consumption; Approx. 1 Nl/min
(see Note 1) Wetted Material: SUS316 (JIS), Zirconia, SUS304
Both zero and span ± 3 vol% H2O/month (JIS) or ASTM grade 304 (flange),
Hastelloy B, (Inconel 600, 601)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
11
Construction: Heater and thermocouple replaceable Wiring Connection: G1/2, Pg 13.5, M20 × 1.5, 1/2
construction. Non explosion proof JIS NPT (with plug), eight holes
C0920 / equivalent to IP44D. Equivalent Installation: Panel, wall or pipe mounting
to NEMA 4X/IP66 (Achieved when the Case: Aluminum alloy
cable entry is completely sealed with a Paint Color: Door; Silver gray (Munsell 3.2PB7.4/1.2)
cable gland in the recirculation pressure Case; Silver gray (Munsell 3.2PB7.4/1.2)
compensated version.) Finish: Polyurethane corrosion-resistance coating
Terminal Box Case: Material; Aluminum alloy Weight: Approx. 6 kg
Terminal Box Paint Color: Case; Mint green (Munsell
5.6BG3.3/2.9) Functions
Cover; Mint green (Munsell 5.6BG3.3/2.9) Display Functions:
Gas Connection: Rc1/4 or 1/4 NPT (Female) Value Display; Displays values of the measured
Wiring Connection: G1/2, Pg 13.5, M20 × 1.5, oxygen concentration, moisture
1/2 NPT quantity,mixture ratio, etc
Installation: Flange mounting Graph Display; Displays trends of measured
Probe Mounting Angle: Horizontal to vertically oxygen concentration, moisture quantity,
downward. When the probe insertion mixture ratio, etc
length is 2 m or less, installing at angles Data Display; Displays various useful data for
from horizontal to vertically downward is maintenance, such as cell temperature,
available. reference junction temperature,
When the probe insertion length exceeds maximum/ minimum moisture quantity,
2.5 m, mount vertically downward (within or the like
± 5°) and use a probe protector. Status Message; Indicates an alarm or error
Weight: occurrence by flashing of the
Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65) / corresponding icon. Indicates status
approx. 11 kg (ANSI 150 4) such as warming-up, calibrating, or the
Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65) / like by the marks.
approx. 13 kg (ANSI 150 4) Alarm, Error Display: Displays alarms such as
Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65) / “Abnormal moisture quantity” or errors
approx. 15 kg (ANSI 150 4) such as “Abnormal cell e.m.f.” when any
Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65) / such status occurs.
approx. 17 kg (ANSI 150 4) Calibration Functions:
Insertion length of 3.0 m: approx. 15 kg (JIS 5K 65) / Automatic Calibration; Requires the ZR40H
approx. 20 kg (ANSI 150 4) Automatic Calibration Unit. It calibrates
automatically at specified intervals.
2. ZR402G Separate type Zirconia High Semi-automatic Calibration; Requires the
Temperature Humidity Analyzer, Converter ZR40H Automatic Calibration Unit.
Input calibration direction on the
High Temperature Humidity Analyzer touchscreen or contact, then it calibrates
Operated using an LCD touchscreen on the converter. automatically afterwards.
Display: LCD display of size 320 by 240 dot with Manual Calibration; Calibration with opening/closing
touchscreen. the valve of calibration gas in operation
Output Signal: 4 to 20 mA DC, two points (maximum interactively with an LCD touchscreen.
load resistance 550 Ω ) Blowback Function:
Contact Output Signal: Four points (one is fail-safe, Output through the contact in the set
normally open) period and time. Auto/Semi_Auto
Contact Input: Two points selectable.
Analog Input: One point (thermal input 4-20 mA) Maintenance Functions:
Automatic Calibration Output: Two points (for Can operate updated data settings in
dedicated automatic calibration unit) daily operation and checking. Display
Ambient Temperature: -20 to +55°C data settings, calibration data settings,
Storage Temperature: -30 to +70°C blowback data settings, current output
Ambient Humidity: 0 to 95 %RH (non-condensing) loop check, input/output contact check.
Power Supply Voltage: Ratings; 100 to 240 V AC Setup Functions:
Acceptable range; 85 to 264 V AC Initial settings suit for the plant conditions
Power Supply Frequency: Ratings; 50/60 Hz when installing the converter. Equipment
Acceptable range; 45 to 66 Hz settings, current output data settings,
Power Consumption: Max. 300 W, approx. 100 W for alarm data settings, contact data settings,
ordinary use. other settings.
Maximum Distance between Detector and Converter: Self-diagnosis:
Conductor two-way resistance must be This function diagnoses conditions of the
10 Ω or less (when a 1.25 mm2 cable or converter or the detector and indicates
equivalent is used, 300 m or less.) when any abnormal condition occurs.
Construction: Equivalent to NEMA 4X/IP66 (with Password Functions:
conduit holes completely sealed with a Enter your password to operate the
cable gland) analyzer excepting data display. Individual
passwords can be set for maintenance
and setup.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
12
Display and setting content: Converter Output: Two points mA analog output (4 to
Measuring Related Items: Oxygen concentration 20 mA DC (maximum load resistance of
(vol% O2), moisture quantity (vol% H2O), 550 Ω ) ) and one of two mA outputs is
mixture ratio (kg/kg), relative humidity with digital output (HART) (minimum load
(%RH) and dew point (°C) resistance of 250 Ω ).
Display Items: Oxygen concentration (vol% O2), Range: Any setting between 0 to 5 through 0 to
moisture quantity (vol% H2O), mixture 100 vol% O2, 0 to 25 through 0 to 100
ratio (kg/kg), relative humidity (%RH), vol% H2O, 0 to 0.200 through 0 to 1.000
dew point (°C), cell temperature (°C), kg/ kg or partial range is available.
thermocouple reference junction For the log output, the minimum range
temperature (°C), maximum/minimum/ values are fixed to 0.1 vol% O2 for the
average oxygen concentration (vol% oxygen concentration, 0.1 vol% H2O for
O2), maximum/ minimum/average the moisture quantity, and 0.01 kg/kg for
moisture quantity (vol% H2O), maximum/ the mixture ratio.
minimum/average mixture ratio (kg/ 4 to 20 mA DC linear or log can be
kg), cell e.m.f. (mV), output 1, 2 current selected. Input/output isolation.
(mA), cell response time (seconds), cell Output damping: 0 to 255 seconds. Hold/non-hold
internal resistance (Ω), cell condition (in selection, preset value setting possible
four grades), heater on-time rate (%), with hold.
calibration record (ten times), time (year/ Contact Output: Four points, contact capacity 30 V
month/day, hour/minute) DC 3 A, 250 V AC 3 A (resistive load).
Calibration Setting Items: Span gas concentration Three of the output points can be selected
(vol% O2), zero gas concentration (vol% to either normally energized or normally
O2), calibration mode (automatic, semi- deenergized status.
automatic, manual), calibration type Delayed functions (0 to 255 seconds) and
and method (zero-span calibration, zero hysteresis function (0 to 9.9 vol%O2) can
calibration only, span calibration only), be added to high/low alarms.
stabilization time (min.sec), calibration The following functions are programmable
time (min.sec), calibration interval (day/ for contact outputs.
hour), starting time (year/month/day, (1) Abnormal, (2) High-high alarm, (3)
hour/minute) Highalarm, (4) Low low alarm, (5) Low
Output Related Items: Analog output/output mode alarm, (6) Maintenance, (7) Calibration,
selection, output conditions when (8) Range switching answer-back, (9)
warmingup/maintenance/calibrating/ Warm-up, (10) Calibration gas pressure
abnormal, oxygen concentration at 4 decrease (answer-back of contact
mA/20 mA (vol% O2), moisture quantity input), (11) Temperature high-alarm,
at 4 mA/20 mA (vol% H2O), mixture ratio (12) Blowback start, (13) Flameout gas
at 4 mA/20 mA (kg/ kg), time constant. detection (answer-back of contact input),
Alarm Related Items: Oxygen concentration high (14) Calibration coefficient alarm, (15)
alarm/ high-high alarm limit values Startup power stabilization timeout alarm
(vol% O2), oxygen concentration low Contact output 4 is set to normally
alarm/low-low alarm limit values (vol% operated, and fixed error status.
O2), moisture quantity high alarm/high- Converter Input: Thermal input one point (4 to 20 mA
high alarm limit values (vol% H2O), DC)
moisture quantity lowalarm/ low-low Contact Input: Two points, voltage-free contacts
alarm limit values (vol% H2O), mixture The following functions are programmable
ratio high alarm/high-high alarm limit for contact inputs:
value (kg/ kg), mixture ratio low alarm/ (1) Calibration gas pressure decrease
low-low alarm limit values (kg/ kg), alarm, (2) Range switching - fixed range
oxygen concentration alarm hysteresis if use range switching (3) External
(vol% O2), moisture quantity alarm calibration start, (4) Process alarm (if this
hysteresis (vol% H2O), mixture ratio signal is received, the heater power turns
alarm hysteresis (kg/ kg), oxygen off), (5) Blowback start
concentration/moisture quantity/mixture Contact capacity: Off-state leakage current; 3 mA or
ratio alarm detection, alarm delay less
(seconds). Self-diagnosis: Abnormal cell, abnormal cell
Contact Related Items: Selection of contact input 1 temperature (low/high), abnormal
and 2, selection of contact output 1 to 4 calibration, defective A/D converter,
(abnormal, high-high alarm, high alarm, defective digital circuit
low alarm, low-low alarm, maintenance, Calibration: Method; Zero/span calibration
calibrating, range switching, warming- Calibration mode; automatic, semi-automatic and
up, calibration gas pressure decrease, manual (All are operated interactively
temperature high alarm blowback, with an LCD touchscreen). Either zero
flameout gas detector calibration or span can be skipped.
coefficient alarm, stabilization timeout) Zero calibration-gas concentration setting range;
0.3 to 100 vol% O2 (minimum setting:
0.01 vol% O2).

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
13
Span calibration-gas concentration setting range; Gas Connection: Rc1/4 or 1/4 NPT (Female)
4.5 to 100 vol% O2 (minimum setting: Wiring Connection: G1/2, Pg 13.5, M20 × 1.5,
0.01 vol% O2). 1/2 NPT select one type (4 pieces)
Use N2-balanced mixed gas containing Installation: Flange mounting
0 to 10% scale of oxygen for standard Probe Mounting Angle: Horizontal to vertically
zero gas, and 80 to 100% scale of downward. When the probe insertion
oxygen for standard span gas. length is 2 m or less, installing at angles
Calibration interval; date/time setting: from horizontal to vertically downward is
maximum 255 days available.
When the probe insertion length is 2.5 m or
3. ZR202G Integrated type Zirconia High more, mount vertically downward (within ±
Temperature Humidity Analyzer 5°) and use a probe protector.
Case: Aluminum alloy
High Temperature Humidity Analyzer Paint Color:Cover; Mint green (Munsell 5.6BG3.3/2.9)
Can be operated in the field without opening the cover Case; Mint green (Munsell 5.6BG3.3/2.9)
using optical switches. Finish: Polyurethane corrosion-resistance coating
Display: 6-digit LCD Weight:
Switch: Three optical switches Insertion length of 0.4 m: approx. 8 kg (JIS 5K 65) /
Output Signal: 4 to 20 mA DC, one point (maximum approx. 13 kg (ANSI 150 4)
load resistance 550 Ω) Insertion length of 1.0 m: approx. 10 kg (JIS 5K 65)
Digital Communication (HART): 250 to 550 Ω, / approx. 15 kg (ANSI 150 4)
depending on number of field devices Insertion length of 1.5 m: approx. 12 kg (JIS 5K 65)
connected to the loop (multi-drop mode). / approx. 17 kg (ANSI 150 4)
Contact Output Signal: Two points (one is fail-safe, Insertion length of 2.0 m: approx. 14 kg (JIS 5K 65)
normally open) / approx. 19 kg (ANSI 150 4)
Contact Input Signal: Two points Insertion length of 3.0 m: approx. 17 kg (JIS 5K 65)
Sample Gas Temperature: 0 to 700°C / approx. 22 kg (ANSI 150 4)
It is necessary to mount the cell using
inconel cell-bolts when the temperature Functions
than 600°C. Display Function: Displays values of the measured
Sample Gas Pressure: - 5 to + 20 kPa (When the oxygen concentration, moisture quantity,
pressure in the furnace exceeds 3 kPa, mixture ratio etc
it is recommended to use pressure Alarm, Error Display: Displays alarms such as “AL-
compensated type. When the pressure
06” or errors such as “Err-01” when any
in the furnace exceeds 5 kPa, pressure
such status occurs.
compensated type is required.)
Calibration Functions:
No pressure fluctuation in the process
Automatic Calibration; Requires the ZR20H
should be allowed.
Automatic Calibration Unit. It calibrates
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
automatically at specified intervals.
Probe Material: JIS SUS316 stainless steel
Semi-automatic Calibration; Requires the ZR20H
Ambient Temperature: -20 to +55°C (- 5 to +70°C on
Automatic Calibration Unit. Input
the case surface)
calibration start signal by optical
Storage Temperature: -30 to +70°C
switch or contact, then it calibrates
Ambient Humidity: 0 to 95%RH (non - condensing)
automatically afterwards.
Power Supply Voltage: Ratings; 100 to 240 V AC
Manual Calibration; Calibration with opening/closing
Acceptable range; 85 to 264 V AC
the valve of calibration gas in operation
Power Supply Frequency: Ratings; 50/60 Hz
interactively with the optical switch.
Acceptable range; 45 to 66 Hz
Maintenance Functions:
Power Consumption: Max. 300 W, approx. 100 W for
Can operate updated data settings in
ordinary use.
daily operation and checking. Display data
Reference Gas System: Natural Convection,
settings, calibration data settings, test
Instrument Air, or Pressure Compensated
settings (current output loop check, input/
Instrument Air System (excluding Natural Convection):
output contact check).
Pressure; 200 kPa plus the pressure
Setup Functions:
inside the process (It is recommended to
Initial settings suit for the plant conditions
use air which is dehumidified by cooling
when installing the converter. Current
to dew point -20°C or less, and dust or oil
output data settings, alarm data settings,
mist are removed.)
contact data settings, other settings.
Consumption; Approx. 1 Nl/min
Wetted Material SUS316 (JIS), Zirconia, SUS304
(JIS) or ASTM grade 304 (flange),
Hastelloy B, (Inconel 600, 601)
Construction: Heater and thermocouple replaceable
construction. Non explosion-proof JIS
C0920 / equivalent to IP44D. Equivalent
to NEMA 4X/IP66 (Achieved when the
cable entry is completely sealed with a
cable gland in the recirculation pressure
compensated version.)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
14
Display and setting content: Converter Output: One mA analog output point (4
Display Related Items: Oxygen concentration to 20 mA DC (maximum load resistance
(vol% O2), moisture quantity (vol% of 550 Ω )) with mA digital output point
H2O), mixture ratio(kg/kg), relative (HART) (minimum load resistance of 250
humidity(%RH), dew point (°C), cell Ω ).
temperature (°C), thermocouple Range; Any setting between 0 to 25 through
reference junction temperature (°C), 0 to 100 vol% H2O, and partial range
maximum/minimum/average oxygen is available (Maximum range value/
concentration (vol% O2), maximum/ minimum range value 1.3 or more) For
minimum/average moisture quantity the log output, the minimum range values
(vol% H2O), maximum/minimum/average are fixed to 0.1 vol% O2 for the oxygen
mixture ratio (kg/kg), cell e.m.f. (mV), concentration, 0.1 vol% H2O for the
output 1, 2 current (mA), cell response moisture quantity, and 0.01 kg/kg for the
time (seconds), cell internal resistance mixture ratio.
(Ω), cell condition (in four grades), 4 to 20 mA DC linear or log can be
heater on-time rate (%), calibration selected. Input/output isolation
record (ten times), time (year/month/ Output damping; 0 to 255 seconds. Hold/non-hold
day/hour/minute) selection, preset value setting possible
with hold.
Calibration Setting Items: Span gas concentration Contact Output: Two points, contact capacity 30 V
(vol% O2), zero gas concentration (vol% DC 3 A, 250 V AC 3 A (resistive load)
O2), calibration mode (automatic, semi- Normally energized or normally de-
automatic, manual), calibration type energized can be selected.
and method (zero-span calibration, zero Delayed functions (0 to 255 seconds) and
calibration only, span calibration only), hysteresis function (0 to 9.9 vol% O2) can
stabilization time (min.sec), calibration be added to high/low alarms.
time (min.sec), calibration interval (day/ The following functions are programmable
hour), starting time (year/month/day/ for contact outputs.
hour/minute) (1) Abnormal, (2) High-high alarm,
Output Related Items: Analog output/output mode (3) Highalarm, (4) Low-low alarm,
selection, output conditions when (5) Low alarm, (6) Maintenance,
warming-up/ maintenance/calibrating/ (7) Calibration, (8) Range switching
abnormal, oxygen concentration at 4 answer-back, (9) Warm-up,
mA/ 20 mA (vol% O2), moisture quantity (10) Calibration gas pressure decrease
at 4 mA/ 20 mA (vol% H2O), mixture (answer-back of contact input),
ratio at 4 mA/ 20 mA (kg/kg), time (11) Flameout gas detection (answer-back
constant, preset values when warming- of contact input).
up/maintenance/calibrating/abnormal, Contact Input: Two points, voltage-free contacts
output preset values on abnormal The following functions are programmable
Alarm Related Items: Oxygen concentration high for contact inputs.
alarm/ high-high alarm limit values (1) Calibration-gas pressure decrease
(vol% O2), oxygen concentration low alarm, (2) Range switching (switched
alarm/low-low alarm limit values (vol% range is fixed), (3) External calibration
O2), moisture quantity high alarm/ start, (4) Process alarm (if this signal is
high-high alarm limit values (vol% received, the heater power turns off)
H2O), moisture quantity low alarm/ Contact capacity: Off-leakage current; 3 mA or less.
low-low alarm limit values (vol% H2O), Self-diagnosis: Abnormal cell, abnormal cell
mixture ratio high alarm/high-high temperature (low/high), abnormal
alarm limit values (kg/kg), mixture ratio calibration, defective A/D converter,
low alarm/low-low alarm limit values defective digital circuit
(kg/kg), oxygen concentration alarm Calibration: Method; Zero/span calibration
hysteresis (vol% O2), moisture quantity Calibration mode; automatic, semi-automatic and
alarm hysteresis (vol% H2O), mixture manual (All are operated using optical
ratio alarm hysteresis (kg/kg), oxygen switches). Either zero or span can be
concentration/moisture quantity/ mixture skipped.
ratio detection, alarm delay (seconds) Zero calibration gas concentration setting range;
Contact Related Items: Selection of contact input 0.3 to 100 vol% O2 (minimum setting:
1 and 2, selection of contact output 0.01 vol% O2).
1 and 2 (abnormal, high-high alarm, Span calibration gas concentration setting range;
high alarm, low alarm, low-low alarm, 4.5 to 100 vol% O2 (minimum setting:
maintenance, calibrating, range 0.01 vol% O2).
switching, warming-up, calibration gas Use N2-balanced mixed gas containing 0
pressure decrease, flameout gas to10% scale of oxygen for standard zero
detection gas, and 80 to 100% scale of oxygen for
standard span gas.
Calibration interval; date/time setting:
maximum 255 days

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
15
■ OPTIONS Insertion Length: 1.05, 1.55, 2.05 m.
Flange: JIS 5K 65A FF equivalent. ANSI Class
4. ZO21P-H High Temperature Probe Adapter 150 4 FF (without serration) equivalent.
for separate type Oxygen Analyzer However, flange thickness is different.
Material: SUS316 (JIS), SUS304 (JIS) or ASTM
Measuring O2 in the high temperature gases (exceeds grade 304 (Flange)
700°C) requires a general-use probe ZR22G of 0.15 m
Weight: 1.05 m; Approx. 6/10/8.5 kg (JIS/ANSI), 1.55
length and a high temperature probe adapter.
m; Approx. 9/13/11.5 kg (JIS/ANSI), 2.05 m;
Sample gas temperature: 0 to 1400°C (when using Approx. 12/16/14.5 kg (JIS/ANSI)
SiC probe) Installation: Bolts, nuts, and washers are provided for
0 to 800°C (when using SUS310S probe detector, probe adapter and process-side
adapter) flange.
Sample gas pressure: -0.5 to + 5 kPa. When using
in the range of 0 to 25 vol% O2 or more, 7. K9471UA Dust Filter for Oxygen Analyzer
the sample gas pressure should be in
the range of -0.5 to +0.5 kPa. (Where This filter is used to protect the cell from corrosive dust
the sample gas pressure for the high- components or high velocity dust in recovery boilers
temperature probe is negative, an ejector and cement kiln. Sample gas flow rate is needed to be
assembly is necessary.) 1m/sec or more to replace gas inside zirconia sensor.
Insertion length: 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.5 m Mesh: 30 microns
Material in Contact with Gas: SUS316 (JIS), SiC or Material: SiC (Filter), SUS316 (JIS)
SUS310S, SUS304 (JIS) or ASTM grade Weight: Approx. 0.2 kg
304 (flange)
Probe Material: SiC, SUS310S (JIS) 8. K9471UC Dust Guard Protector
Installation: Flange mounting (FF type or RF type)
Probe Mounting Angle: Vertically downward within ± Recommended to be used when sample gas is likely
5°. Where the probe material is SUS310S, to flow directly into the cell due to its flow direction in
horizontal mounting is available. the stack or the like, flammable dust may go into the
Construction: Non explosion-proof. Rainproof cell, or water drops are likely to fall and remain in the
construction cell during downtime or the like due to the installation
Weight(example): Insertion length of 1.0 m: approx. position.
5.3 kg (JIS) / approx. 11.3 kg (ANSI) Material: SUS316 (JIS)
Insertion length of 1.5 m: approx. 5.8 kg Weight: Approx. 0.3 kg
(JIS) / approx. 11.8 kg (ANSI)
9. ZH21B Dust Protector for High
5. E7046EC/E7046EN Ejector Assembly for
High Temperature Detector of separate temperature Humidity Analyzer
type Oxygen Analyzer This protector is designed to protect the probe output
For use in cases where pressure of sample gas for high from dust agitation (i.e., to prevent combustible
temperature detector is negative. materials from entering the probe cell) where humidity
measurements are made under dusty environments.
5.1 Needle Valve Insertion length: 0.440 m
Connection: Rc1/4 or 1/4 NPT (Female) Flange: JIS 5K 80 FF equivalent or ANSI
Material: SUS316 (JIS) Class150 4 FF equivalent. (However,
(Note) Pipes and connectors are not provided.
flange thickness is different.)
5.2 Pressure Gauge Assembly Material: SiC, SUS316 (JIS), SUS304 (JIS) or
Material in Contact with Gas: SUS316 (JIS) ASTM grade 304 (flange)
Case Material: Aluminum alloy (Paint color; black) Weight: Approx. 6 kg (JIS), approx. 8.5 kg (ANSI)
Scale: 0 to 100 kPa G Mounting: Mounted on the probe or process flange
Connection: R1/4 or 1/4 NPT, SUS304 (JIS) withbolts and the associated nuts and
(with Bushing G3/8 x R1/4 or 1/4 NPT (Female)) washers.
5.3 Ejector
Ejector Inlet Air Pressure: 29 to 68 kPa G 10. ZO21S Standard Gas Unit
Air Consumption: Approx. 30 to 40 l/min The ZO21S does not conform to CE marking.
Suction gas flow rate: 3 to 7 l/min Function: Portable unit for calibration gas supply
Connection: Rc1/4, SUS304 (JIS) consisting of span gas (air) pump, zero
Tube Connection: (ø6/ø4 mm or 1/4 inch copper tube gas cylinder with sealed inlet, flow rate
or stainless tube) checker and flow rate needle valve.
Sealed Zero Gas Cylinders (6 provided): E7050BA
6. ZO21R Probe Protector for Zirconia Capacity: 1 l
Oxygen Analyzer Filled pressure: Approx. 686 kPa G (at 35°C)
Used when sample gas flow velocity is approx. 10m/ Composition: 0.95 to 1.0 vol% O2 (N2-balanced)
sec or more and dust particles wears the detector in Power Supply: 100, 110, 115, 200, 220, 240V AC
cases such as pulverized coal boiler of fluidized bed ± 10%, 50/60 Hz
furnace (or burner) to protect the detector from wearing Power Consumption: Max. 5 VA
by dust particles. When probe insertion length is 2.5 m Paint Color:
or more and horizontal installation, specify the ZO21R- Mainframe;Munsell 2.0 GY3.1/0.5 equivalent
L-200-□*B to reinforce the probe. Cover; Munsell 2.8 GY6.4/0.9 equivalent
Weight: Approx. 3 kg

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
16
11. ZA8F Flow Setting Unit Materials: Body: Aluminum alloy, Piping: SUS316
Used when instrument air is provided. (JIS), SUS304 (JIS), Flowmeter: MA
(Methacrylate resin)
This unit consists of flowmeter and flow control valve to Finish: Polyurethane corrosion-resistance coating
controls flow rates of calibration gas and reference gas. Case: Mint green (Munsell 5.6BG3.3/2.9),
FIowmeter Scale: Calibration gas; 0.1 to 1.0 l/min. Cover: Mint green (Munsell 5.6BG3.3/2.9)
Reference gas; 0.1 to 1.0 l/min. Piping Connection: Rc1/4 or 1/4 NPT(Female)
Construction: Dust-proof and rainproof construction Power Supply: 24V DC (from ZR202G), Power
Case Material: SPCC (Cold rolled steel sheet) consumption: 1.3 W
Painting: Baked epoxy resin, Dark-green (Munsell 2.0 Reference Gas Pressure: Sample gas pressure
GY 3.1/0.5 or equivalent) plus Approx. 150 kPa (690 kPa max.),
Tube Connections: Rc1/4 or 1/4 NPT (Female) (Pressure at inlet of automatic calibration
Reference Gas Pressure: Clean air supply of sample unit)
gas pressure plus approx. 50 kPa G (or Air Consumption: Approx. 1.5 l/min
sample gas pressure plus approx.150 kPa Weight: Approx. 2 kg
when a check valve is used.) Pressure Ambient Temperature: -20 to +55°C, no condensing
at inlet of the flow setting unit. (Max. 300 and freezing
kPa G) Ambient Humidity: 0 to 95%RH
Air Consumption: Approx. 1.5 l/min Storage Temperature: -30 to +65°C
Weight: Approx. 2.3 kg
14. L9852CB/G7016XH Stop Valve
12. ZR40H Automatic Calibration Unit (for The stop valve is mounted on the calibration gas line.
Separate type) To include Nipple Stop Valve with the product, select
Used when automatic calibration is required for the the suffix code (/SV) for ZR22G Separate type Zirconia
separate type and instrument air is provided. The Oxygen/High Temperature Humidity Analyzer, Detector
solenoid valves are provided as standard. or ZR202G Integrated type Zirconia Oxygen/High/High
Construction: Dust-proof and rainproof construction: Temperature Humidity Analyzer.
NEMA 4X/IP67 - only for case coating Connection: Rc1/4 or 1/4 NPT (Female)
solenoid valve, not flowmeter (excluding Material: SUS316 (JIS)
flowmeter) Weight: Approx. 150 g
Mounting: 2-inch pipe or wall mounting, no vibration
Materials: Body: Aluminum alloy, Piping: SUS316 15. K9292DN/K9292DS Check Valve
(JIS), SUS304 (JIS), Flowmeter: MA This is used to prevent entry of sample gas into
(Methacrylate resin) Bracket : SUS304 calibration gas line. Purpose is the same as stop valve,
(JIS) but is convenient, as it does not need to be opened or
Finish: Polyurethane corrosion-resistance closed for calibration.
coating, Mint green (Munsell Mount directly on calibration gas inlet of detector in
5.6BG3.3/2.9) place of stop valve. However as source pressure of 150
Piping Connection: Rc1/4 or 1/4 NPT (Female) kPa G or more is needed, standard gas unit cannot be
Power Supply: 24V DC (from ZR402G), Power used.
consumption: Approx. 1.3 W
Reference Gas Pressure: Sample gas pressure When option code “/CV” of the ZR22G or the ZR202G
plus Approx. 150 kPa (690 kPa max.), is specified, check valve is provided.
(Pressure at inlet of automatic calibration Connection: Rc1/4 or 1/4 NPT (Female)
unit) Material: SUS304 (JIS)
Air Consumption: Approx. 1.5 l/min Pressure: Between 70 kPa G or more 350 kPa G or
Weight: Approx. 3.5 kg less
Ambient Temperature: -20 to +55°C, no condensing Weight: Approx. 90 g
and freezing
Ambient Humidity: 0 to 95%RH 16. Air Set
Storage Temperature: -30 to +65°C
G7003XF/K9473XK
13. ZR20H Automatic Calibration Unit (for Primary Pressure: Max. 1 MPa G
Secondary Pressure: 0.02 to 0.2 MPa G
Integrated type) Connection: Rc1/4 or 1/4 NPT (F) with joint adapter
Used when automatic calibration is specified for the
G7004XF/K9473XG
integrated type and instrument air is provided.
Primary Pressure: Max. 1 MPa G
Equipped with the analyzer when automatic calibration Secondary Pressure: 0.02 to 0.5 MPa G
is specified in the suffix code of the ZR202G Integrated Connection: Rc1/4 or 1/4 NPT (F) with joint adapter
type by selecting either “-A (Horizontal mounting)”
or “-B (Vertical mounting)”. The ZR20H should be 17. G7001ZC Zero Gas Cylinder
arranged when automatic calibration is to be required Capacity: 3.4 l
after the ZR202H has been installed. Ask Yokogawa Filled pressure: 9.8 to 12 MPa G
service station for its mounting. Composition: 0.95 to 1.0 vol% O2 (N2-balanced)
Construction: Dust-proof and rainproof construction; (Note) Export of such high pressure filled gas cylinders
NEMA 4X/IP67 (excluding flowmeter) to most countries is prohibited or restricted.
Mounting: Mounted on ZR202G, no vibration

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
17
18. G7013XF/G7014XF Pressure Regulator for Instruction -J - - - - - - Japanese
Gas Cylinder manual -E - - - - - - English
Primary Pressure: Max.14.8 MPa G, -C - - - - - - Chinese
Secondary Pressure; 0 to 0.4 MPa G — -A - - - - - - Always -A
Connection: Inlet W22 14 threads, right hand screw
Option /HS Set for Humidity Analyzer (*1)
Outlet Rc1/4 or 1/4 NPT (Female)
Material: Brass body /H Hood (*3)
Tag plate /SCT Stainless steel tag plate (*2)
19. E7044KF Case Assembly of Calibration /PT Printed tag plate (*2)
Gas Cylinder
NAMUR NE43 /C2 Failure alarm down-scale:
Case Paint: Baked epoxy resin, complian Output status at CPU failure
Jade green (Munsell 7.5 BG 4/1.5) and hardware error is 3.6 mA
Installation: 2B pipe mounting or less (*4)
Weight: Approx. 10 kg
(Note) Export of such high pressure filled gas cylinders /C3 Failure alarm up-scale:
to most countries is prohibited or restricted. Output status at CPU failure
and hardware error is 21.0
20. ZR22A, ZR202A Heater Assembly mA or more (*4)
ZR22A: Spare Parts for ZR22G Standard /EQ EAC with PA (*5)
ZR202A: Spare Parts for ZR202G /ER EAC (*5)
(Note) Yokogawa shall not guarantee the heater *1 For humidity measurements, be sure to specify /HS
assembly after its replacement. options.
*2 Specify either /SCT or /PT option code.
STANDARD ACCESSARIES *3 Sun shield hood is still effective even if scratched.
ZR402G *4 Output signal limits: 3.8 to 20.5 mA. Specify either /C2
or /C3 option code.
Item Parts. No. Q'ty Description *5 “/EQ” is EAC with Pattern Approval for Russia.
“/ER” is EAC for Kazakhstan and Belarus.
Fuse A1113EF 1 3.15 A (Note) If AC line voltage is 125 V AC or greater, or in the
Bracket F9554AL 1 For pipe, panel,or EEC, the ZO21D cannot be used with the ZR402G.
wall mounting
Language Japanese English German French
Screws for Bracket F9123GF 1
Model
ZR22G ZA8C K9290LF K9290KF K9290MF K9290MG
HA400 (kg) K9293HT K9293HU K9293HW K9293HV
Item Parts. No. Q'ty Description
HA400 (%) K9293HP K9293HQ K9293HS K9293HR
Allen wrench L9827AB 1 For lock screw
AV8V K9296CN K9296CN K9296CN K9296CN
ZR202G Note for ZR22G combination use with existing older
model converters
Item Parts. No. Q'ty Description When the ZR22G is used with existing older model
converters, ZA8C, AV8C and HA400, ROM
Fuse A1113EF 1 3.15 A
replacement and addition of a cold junction
Allen wrench L9827AB 1 For lock screw temperature compensation board are required.
The part numbers of each language version of ROM
refer to table below.
The part numbers of cold junction temperature
 Model and Code compensation boards are K9471JA for the ZA8C
For replacing the ROM by using ROM extraction tool
1. Separate type General Parpose Zirconia (Part No. K9471JT) and mounting the cold
Oxygen / High Temperature Humidity junction temperature compensation board, it is
Analyzer, Converter recommended that you ask Yokogawa service
station.

Model Suffix Option Description


code code
ZR402G - - - - - - - - - - - - - - Separate type Zirconia
Oxygen Analyzer, Converter
Converter -P ------ G1/2
thread -G ------ Pg13.5
-M ------ M20x1.5
-T ------ 1/2NPT
Display -J ------ Japanese
-E ------ English
-G ------ German
-F ----- - French
-C ------ Chinese

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
18
2. Separate type Zirconia Oxygen / High Temperature Humidity Analyzer, Detectors
Style : S2
Model Suffix code Option code Description
ZR22G ---------------------------------- ---------- Separate type Zirconia Oxygen/ High Temperature Humidity Analyzer, Detector
Length -015 ---------- 0.15 m (for high temperature use) (*1)
-040 ---------- 0.4 m
-070 ---------- 0.7 m
-100 ---------- 1.0 m
-150 ---------- 1.5 m
-200 ---------- 2.0 m
-250 ---------- 2.5 m (*2)
-300 ---------- 3.0 m (*2)
-360 ---------- 3.6 m (*2)
-420 ---------- 4.2 m (*2)
-480 ---------- 4.8 m (*2)
-540 ---------- 5.4 m (*2)
Wetted material -S ---------- Stainless steel
-C ---------- Stainless steel with Inconel calibration gas tube (*10)
Flange -A ---------- ANSI Class 150 2 RF
(*3) -B ---------- ANSI Class 150 3 RF
-C ---------- ANSI Class 150 4 RF
-E ---------- DIN PN10 DN50 A
-F ---------- DIN PN10 DN80 A
-G ---------- DIN PN10 DN100 A
-K ---------- JIS 5K 65 FF
-L ---------- JIS 10K 65 FF
-M ---------- JIS 10K 80 FF
-P ---------- JIS 10K 100 FF
-Q ---------- JIS 5K 32 FF (for high temperature use) (*4)
-R ---------- JPI Class 150 4 RF
-S ---------- JPI Class 150 3 RF
-W ---------- Westinghouse

Reference gas -C ---------- Natural convection


-E ---------- External connection (Instrument air) (*11)
-P ---------- Pressure compensated (*11)
Gas Thread -R ---------- Rc1/4
-T ---------- 1/4NPT(Female)
Connection box thread -P ---------- G1/2
-G ---------- Pg13.5
-M ---------- M20 x1.5
-T ---------- 1/2 NPT
-Q ---------- Quick connect (*9)
Instruction manual -J ---------- Japanese
-E ---------- English
-C ---------- Chinese
-- -A ---------- Always -A
Options /C Inconel bolt (*5)
Valves /CV Check valve (*6)
/SV Stop valve (*6)
Filter /F1 Dust Filter (*7)
/F2 Dust Guard Protector (*7)
Tag plates /SCT Stainless steel tag plate (*8)
/PT Printed tag plate (*8)
Standard /EQ EAC with PA (*12)
/ER EAC (*12)
*1 Used with the ZO21P High Temperature Probe Adapter. Select flange (-Q).
*2 When installing horizontally the probe whose insertion length is 2.5 m or more, use the Probe Protector. Be sure to

specifyZO21R-L-200- . Specify the flange suffix code either -C or -K.
*3 The thickness of the flange depends on its dimensions.
*4 Not used in conjunction with —P (pressure compensation) for reference gas. The flange thickness does not conform to JIS
specification
*5 Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700 °C).
*6 Specify either /CV or /SV option code.
*7 Not used with the high temperature humidity analyzer.
*8 Specify either /SCT or /PT option code.
*9 Not waterproof, avoid rain. Operating maximum temperature is 80°C. Available only in the U.S.
*10 Recommended if sample gas contains corrosive gas like chlorine.
*11 Piping for reference gas must be installed to supply reference gas constantly at a specified flow rate.
*12 “/EQ” is EAC with Pattern Approval for Russia. “/ER” is EAC for Kazakhstan and Belarus.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
19
3. Integrated type Zirconia Oxygen / High temperature Humidity Analyzer

Model Suffix code Option code Description


ZR202G -------------------------------------- ---------- Integrated type Zirconia Oxygen/ High Temperature Humidity
Analyzer
Length -040 ---------- 0.4 m
-070 ---------- 0.7 m
-100 ---------- 1.0 m
-150 ---------- 1.5 m
-200 ---------- 2.0 m
-250 ---------- 2.5 m (*1)
-300 ---------- 3.0 m (*1)
Wetted material -S ---------- Stainless steel
-C ---------- Stainless steel with Inconel calibration gas tube (*10)
Flange -A ---------- ANSI Class 150 2 RF
(*2) -B ---------- ANSI Class 150 3 RF
-C ---------- ANSI Class 150 4 RF
-E ---------- DIN PN10 DN50 A
-F ---------- DIN PN10 DN80 A
-G ---------- DIN PN10 DN100 A
-K ---------- JIS 5K 65 FF
-L ---------- JIS 10K 65 FF
-M ---------- JIS 10K 80 FF
-P ---------- JIS 10K 100 FF
-R ---------- JPI Class 150 4 RF
-S ---------- JPI Class 150 3 RF
-W ---------- Westinghouse

Auto Calibration -N ---------- Not required


-A ---------- Horizontal mounting (*8)
-B ---------- Vertical mounting (*8)
Reference gas -C ---------- Natural convection
-E ---------- External connection (Instrument air) (*11)
-P ---------- Pressure compensated (*11) (*14)
Gas Thread -R ---------- Rc1/4
-T ---------- 1/4 NPT (Female)
Connection box thread -P ---------- G1/2
-G ---------- Pg 13.5
-M ---------- M20x1.5
-T ---------- 1/2 NPT
Instruction manual -J ---------- Japanese
-E ---------- English
-C ---------- Chinese
-- -A ---------- Always -A
Options /C Inconel bolt (*3)
/HS Set for Humidity Analyzer (*4)
Valves /CV Check valve (*5)
/SV Stop valve (*5)
/H Hood (*9)
/F1 Dust Filter (*6)
/F2 Dust Guard Protector (*6)
Tag plates /SCT Stainless steel tag plate (*7)
/PT Printed tag plate (*7)
NAMUR NE43 compliant /C2 Failure alarm down-scale:
Output status at CPU failure and hardware error is 3.6 mA or less
(*12)
/C3 Failure alarm up-scale:
Output status at CPU failure and hardware error is 21.0 mA or more
(*12)
Standard /EQ EAC with PA (*13)
/ER EAC (*13)

*1 For the horizontally installed probe whose insertion length is 2.5 m or more, use the Probe Protector. Be sure to specify

ZO21R-L-200- . Specify the flange suffix code either -C or -K.
*2 The thickness of the flange depends on its dimensions.
*3 Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700°C).
*4 For humidity measurements, be sure to specify /HS options. Pressure compensation of reference gas can not be selected.
*5 Specify either /CV or /SV option code.
*6 Not used with the high temperature humidity analyzer.
*7 Specify either /SCT or /PT option code.
*8 No need to specify the option codes, /CV and /SV, since the check valves are provided with the Automatic Calibration Unit.
Automatic calibration cannot be used when natural convection is selected as reference air.
*9 Sun shield hood is still effective even if scratched. Hood is necessary for outdoor installation out of sun shield roof.
*10 Recommended if sample gas contains corrosive gas like chlorine.
*11 Piping for reference gas must be installed to supply reference gas constantly at a specified flow rate.
*12 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
*13 “/EQ” is EAC with Pattern Approval for Russia. “/ER” is EAC for Kazakhstan and Belarus.
*14 When Pressure Compensated is specified with “-A” or “-B” in Auto Calibration, consult with Yokogawa.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Nov. 07, 2019-01
20
4. High Temperature Probe Adapter for 6. Probe Protector for Zirconia Oxygen
Separate type Oxygen Analyzer Analyzers

Option Suffix Option


Model Suffix code Description Model Description
code code code
High Temperature Probe ZO21R -L ------ Probe Protector (0 to 700°C)
ZO21P -H -------
Adapter
Insertion -100 ------ 1.05 m
Material -A ------- SiC length -150 ------ 1.55 m
-B ------- SUS 310S (JIS) -200 ------ 2.05 m
Insertion -050 ------ 0.5 m Flange ( *1) -J ------ JIS 5K 65 FF
length -060 ------ 0.6 m -A ------ ANSI Class 150 4 FF
-070 ------ 0.7 m
-080 ------ 0.8 m Style code *B ------ Style B
-090 ------ 0.9 m
*1 Thickness of flange depends on dimensions of flange.
-100 ------ 1.0 m
-150 ------ 1.5 m
7. Dust Filter for Zirconia Oxygen Analyzers
Flange -J ------- JIS 5K 50 FF
-N ------- JIS 10K 65 FF Part No. Description
-M ------- JIS 10K 80 FF K9471UA Filter
-L ------- JIS 10K 100 FF
-A ------- ANSI Class 150 4 RF K9471UX Tool
-R ------- ANSI Class 150 2 1/2 RF
-Q ------- ANSI Class 150 3 RF 8. Dust Guard Protector
-T ------- JPI Class 150 3 RF
-S ------- JPI Class 150 4 RF Part No. Description
-E ------- DIN PN10 DN50 A
K9471UC Dust guard protector
Style code *B ------- Style B
Option Ejector /EJ1 Ejector Assy with E7046EC
/EJ2 Ejector Assy with E7046EN
Tag plate /SCT Stainless steel tag plate 9. Dust Protector for High Temperature
Humidity Analyzers
Note: For this high-temperature use probe adapter, be
sure to specify the ZR22G probe of its insertion length
Suffix Option
0.15 meters. Model Description
code code
ZH21B ----- ------ Dust Protector (0 to 600°C)
High Temperature Probes (Spare Parts)
Insertion
Part No. Description -040 ------ 0.440 m
length
K9292TP SiC, insertion length 0.5 m Flange -J ------ JIS 5K 80 FF *(1)
E7046CF SiC, insertion length 0.6 m -A ------ ANSI Class 150 4B FF *(2)
K9292TQ SiC, insertion length 0.7 m Style code *B ------ Style B
E7046CG SiC, insertion length 0.8 m * The flange thickness varies.
E7046CH SiC, insertion length 0.9 m □ □
Specify the probe ZR22G-040- -K or ZR202G-040- -K in case of (1)
□ □
ZR22G-040- -C or ZR202G-040- -C in case of (2)
E7046AL SiC, insertion length 1.0 m
E7046BB SiC, insertion length 1.5 m 10. Standard Gas Unit
K9292TV SUS310S (JIS), insertion length 0.5 m
Suffix Option
E7046CR SUS310S (JIS), insertion length 0.6 m Model Description
code code
K9292TW SUS310S (JIS), insertion length 0.7 m ZO21S ----- ------ Standard gas unit
E7046CS SUS310S (JIS), insertion length 0.8 m
-2 ------ 200 V AC 50/60 Hz
E7046CT SUS310S (JIS), insertion length 0.9 m -3 ------ 220 V AC 50/60 Hz
Power -4 ------ 240 V AC 50/60 Hz
E7046AP SUS310S (JIS), insertion length 1.0 m
supply -5 ------ 100 V AC 50/60 Hz
E7046AQ SUS310S (JIS), insertion length 1.5 m -7 ------ 110 V AC 50/60 Hz
-8 ------ 115 V AC 50/60 Hz
5. Ejector Assembly or High Temperature -J ------ Japanese version
Panel
Use of separate type Oxygen Analyzer -E ------ English version

Part No. Description Style code *A ------ Style A

E7046EC Needle valve; Rc1/4, Pressure gauge; R1/4,


Ejector; ∅6/∅4 mm TUBE joint: SUS304 (JIS)
E7046EN Needle valve; 1/4 NPT(F), Pressure gauge;1/4 NPT(M),
Ejector; 1/4 TUBE joint: SUS304 (JIS)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
21
11. Flow Setting Unit for manual calibration 15. Check Valve for Calibration-gas line
(Needs instrument air.)
Part No. Description
Suffix Option K9292DN Joint: Rc1/4, Material: SUS304 (JIS)
Model Description
code code
K9292DS Joint: 1/4 NPT (F), Material: SUS304 (JIS)
ZA8F -------- -------- Flow setting unit
-J -------- Rc1/4
Joint
-A -------- With 1/4 NPT (F) adapter
16. Air Set
Style code *C -------- Style C Part No. Description
G7003XF Joint: Rc1/4, Material: Zinc alloy
12. Automatic Calibration Unit for Separate
K9473XK Joint: 1/4 NPT (F), Material: Zinc alloy with adapter
type Analyzer (Needs instrument air.)
G7004XF Joint: Rc1/4, Material: Zinc alloy
Suffix Option K9473XG Joint: 1/4 NPT (F), Material: Zinc alloy with adapter
Model Description
code code
ZR40H --------- -------- Automatic calibration 17. Zero gas Cylinder
unit for ZR402G
Gas piping -R -------- Rc1/4 Part No. Description
connection -T -------- 1/4 NPT (F) G7001ZC 3.4 l container, 0.95 to 1.0 vol % O2, N2-balanced.
Wiring -P -------- G1/2 (Note) Export of such high pressure filled gas cylinders
connection -G -------- Pg 13.5 to most countries is prohibited or restricted.
-M -------- M20 x 1.5
-T -------- 1/2 NPT
18. Pressure Regulator for Gas Cylinder
-- -A -------- Always -A
Part No. Description
13. Automatic Calibration Unit for Integrated G7013XF Inlet: W22 14 threads, Outlet: Rc1/4
type Analyzer (Needs instrument air.) G7014XF Inlet: W22 14 threads, Outlet: 1/4 NPT (F)

Suffix Option
Model
code code
Description 19. Case Assembly for Calibration-gas
Cylinder
ZR20H --------- -------- Automatic calibration
unit for ZR202G *1
Part No. Description
Gas piping -R -------- Rc1/4 E7044KF Calibration gas unit case
connection -T -------- 1/4 NPT (F)
(Note) Export of such high pressure filled gas cylinders
Reference air -E -------- Instrument air to most countries is prohibited or restricted.
*2 -P -------- Pressure compensated
Mounting -A -------- Horizontal mounting
-B -------- Vertical mounting
– -A -------- Always -A

*1 Ask Yokogawa service station for additional mounting


of ZR20H to the preinstalled ZR202G.
*2 Select the appropriate reference gas of ZR20H
according to the one of ZR202G.

14. Stop Valve for Calibration gas line

Part No. Description


L9852CB Joint: Rc1/4, Material: SUS316 (JIS)
G7016XH Joint: 1/4 NPT (F), Material: SUS316 (JIS)

Part No. Description


G7209XA Nipple: R1/4, Material: SUS304 (JIS)
K9470ZN Nipple: 1/4 NPT, Material: SUS304 (JIS)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
22
20. Heater Assembly

Style: S2

Model Option code Description

ZR22A ---------- ----------- Heater Assembly for ZR22G


Length -015 ----------- 0.15 m
(*1) -040 ----------- 0.4 m
-070 ----------- 0.7 m
-100 ----------- 1m
-150 ----------- 1.5 m
-200 ----------- 2m
-250 ----------- 2.5 m
-300 ----------- 3m
Jig for -A ---------- with Jig (*2)
change -N ---------- None
Reference gas (*3) -A ----------- Natural convention,
External connection (Instrument air)
-B ----------- Pressure compensated (for ZR22G S2)
-C ----------- Pressure compensated (for ZR22G S1)

*1 Suffix code of length should be selected as same as ZR22G installed.


*2 Jig part no. is K9470BX to order as a parts after purchase.
*3 Select appropriately among "-A", "-B", "-C" according
to the reference gas supply method and style.
(Note) The heater is made of ceramic, do not drop or subject it to pressure stress.

Model Suffix code Option code Description

ZR202A --------- ----------- Heater Assembly for ZR202G


Length (*1) -040 ----------- 0.4 m
-070 ----------- 0.7 m
-100 ----------- 1m
-150 ----------- 1.5 m
-200 ----------- 2m
-250 ----------- 2.5 m
-300 ----------- 3m
Jig for change -A ----------- with Jig (*2)
-N ----------- None
― -A ----------- Always -A

*1 Suffix code of length should be selected as same as ZR202G installed.


*2 Jig part no. is K9470BX to order as a parts after purchase.
(Note) The heater is made of ceramic, do not drop or subject it to pressure stress.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
23

■EXTERNAL DIMENSIONS
1. Model ZR22G Separate type Zirconia Oxygen / High Temperature Humidity Analyzer,
Detectors

Unit : mm

L 283 to 292
t

Ø124±3
Ø50.8

85
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0 Rc1/4 or 1/4 NPT (F)
Reference gas inlet
3.6, 4.2, 4.8, 5.4 (m)
155 to 163 69 C
2-G1/2, 2-1/2 NPT etc.
Cable connection port
(Female)

ØA
25

48

ØB

Rc1/4 or 1/4 NPT (F)


Flange Calibration gas inlet Flange

C
Flange A B C t
ANSI Class 150 2 RF 152.4 120.6 4 - Ø19 19
ANSI Class 150 3 RF 190.5 152.4 4 - Ø19 24
ANSI Class 150 4 RF 228.6 190.5 8 - Ø19 24 ØA
DIN PN10 DN50 A 165 125 4 - Ø18 18
DIN PN10 DN80 A 200 160 8 - Ø18 20
DIN PN10 DN100 A 220 180 8 - Ø18 20
JIS 5K 65 FF 155 130 4 - Ø15 14 ØB
JIS 10K 65 FF 175 140 4 - Ø19 18
JIS 10K 80 FF 185 150 8 - Ø19 18
Flange
JIS 10K 100 FF 210 175 8 - Ø19 18
JIS 5K 32 FF 115 90 4 - Ø15 5
JPI Class 150 4 RF 229 190.5 8 - Ø19 24
JPI Class 150 3 RF 190 152.4 4 - Ø19 24 F07_01.ai
Westinghouse 155 127 4 - Ø11.5 14

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
24
Model ZR22G...-P (with pressure compensated) Separate type Zirconia Oxygen / High
Temperature Humidity Analyzer, Detectors
Unit : mm

L 303
t

ø124±3
Ø50.8

85
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0
3.6, 4.2, 4.8, 5.4 (m) Rc1/4 or 1/4 NPT (F)
Reference gas inlet
C
156 87
2-G1/2, 2-1/2 NPT etc.
t
Cable connection port
(Female)
ØA
25

48
ØB

Reference gas outlet PIPING


:B Rc1/4 or 1/4 NPT (F)
Flange Flange
Calibration gas inlet
Stop Valve
PIPING : A
C
Flange A B C t PIPING
ANSI Class 150 2 RF 152.4 120.6 4 - Ø19 19 A
ANSI Class 150 3 RF 190.5 152.4 4 - Ø19 24 B ØA
ANSI Class 150 4 RF 228.6 190.5 8 - Ø19 24 B
DIN PN10 DN50 A 165 125 4 - Ø18 18 A
DIN PN10 DN80 A 200 160 8 - Ø18 20 B
DIN PN10 DN100 A 220 180 8 - Ø18 20 B ØB
JIS 5K 65 FF 155 130 4 - Ø15 14 A
JIS 10K 65 FF 175 140 4 - Ø19 18 A Flange
JIS 10K 80 FF 185 150 8 - Ø19 18 B
JIS 10K 100 FF 210 175 8 - Ø19 18 B
JPI Class 150 4 RF 229 190.5 8 - Ø19 24 B
JPI Class 150 3 RF 190 152.4 4 - Ø19 24 B
Westinghouse 155 127 4 - Ø11.5 14 A F07_02.ai

●Check Valve (option code /CV), Stop valve (option code /SV) -specified Calibration gas inlet
Detector case Detector case

Approx. 44

Approx. 100

40 Ø48

19
Full open height
Calibration gas inlet 58
Rc1/4 or 1/4NPT
Calibration gas inlet Rc1/4 or 1/4NPT

with Stop Valve (option: /SV) with Check Valve (option: /CV)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
25
2. Model ZR402G Separate type Zirconia Oxygen/ High Temperature Humidity Analyzer,
Converter
Unit: mm
1 to 6 (Panel Thickness)
2B mounting pipe

EXA ZR402G
4 - Ø6 holes
for Wall mounting
280

126.5
100
(*1/2 NPT)
20.2

8 10 57.3 36
228 136.3 54.7
40 40 40
111

8-G1/2, *8-1/2 NPT etc


(Wiring connection) (Female)
36 23

*: 1/2 NPT with plug

38 24 14 38
( for wall mounting)
4 - Ø6 holes
+2
0274
126.5

4-R8 to R10
or
4-C5 to C8

+2
190 183 0

Wall mounting Panel Cut-out F08_01.ai

●With sun shield hood (option code /H)


Unit : mm
243 ±4 205.5 ±4

55 ±2 123 ±3 155.5 ±3
39±2

ZR402G
94.5 ±3
251.5 ±4

F10.ai

Material of HOOD : Aluminum

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
26
3. Model ZR202G Integrated type Zirconia Oxygen/ High Temperature Humidity Analyzers
Unit: mm
L 338 to 351 Ø123
t Display side
L= 0.4, 0.7, Rc1/4 or 1/4 NPT (F)
1.0, 1.5, 2.0, Reference gas inlet
2.5, 3.0 (m)

122
Ø50.8

25

48.5

125
Rc1/4 or 1/4 NPT (F)
Calibration gas inlet

153 to 164
4-G1/2,2-1/2 NPT etc. Terminal side
252 to 265 49 Cable connection port 170
(Female)

ØA

ØB

Flange
Flange
C

Flange A B C t
ANSI Class 150 2 RF 152.4 120.6 4 - Ø19 19
ØA
ANSI Class 150 3 RF 190.5 152.4 4 - Ø19 24
ANSI Class 150 4 RF 228.6 190.5 8 - Ø19 24
DIN PN10 DN50 A 165 125 4 - Ø18 18
DIN PN10 DN80 A 200 160 8 - Ø18 20
DIN PN10 DN100 A 220 180 8 - Ø18 20 ØB
JIS 5K 65 FF 155 130 4 - Ø15 14
JIS 10K 65 FF 175 140 4 - Ø19 18
Flange
JIS 10K 80 FF 185 150 8 - Ø19 18
JIS 10K 100 FF 210 175 8 - Ø19 18
JPI Class 150 4 RF 229 190.5 8 - Ø19 24
JPI Class 150 3 RF 190 152.4 4 - Ø19 24 F11_01.ai
Westinghouse 155 127 4 - Ø11.5 14

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
27
Model ZR202G...-P (with pressure compensated) Integrated type Zirconia Oxygen / High
Temperature Humidity Analyzers
Unit: mm

L 342 ± 4 Ø123
t Display side
Rc1/4 or 1/4 NPT (F)
L= 0.4, 0.7, Reference gas inlet
1.0, 1.5, 2.0,
2.5, 3.0 (m)

122
Ø50.8

25

48.5

125
Reference gas outlet PIPING
:B

PIPING:A 4-G1/2,2-1/2 NPT etc. Terminal side


156 ± 3 Cable connection port 170
Stop (Female)
256 ± 4 valve 49
Rc1/4 or 1/4 NPT (F) C
Calibration gas inlet

ØA

ØB

Flange
Flange

Flange A B C t PIPING
ANSI Class 150 2 RF 152.4 120.6 4 - Ø19 19 A
B ØA
ANSI Class 150 3 RF 190.5 152.4 4 - Ø19 24
ANSI Class 150 4 RF 228.6 190.5 8 - Ø19 24 B
DIN PN10 DN50 A 165 125 4 - Ø18 18 A
DIN PN10 DN80 A 200 160 8 - Ø18 20 B
DIN PN10 DN100 A 220 180 8 - Ø18 20 B ØB
JIS 5K 65 FF 155 130 4 - Ø15 14 A
JIS 10K 65 FF 175 140 4 - Ø19 18 A
Flange
JIS 10K 80 FF 185 150 8 - Ø19 18 B
JIS 10K 100 FF 210 175 8 - Ø19 18 B
JPI Class 150 4 RF 229 190.5 8 - Ø19 24 B
JPI Class 150 3 RF 190 152.4 4 - Ø19 24 B F11_02.EPS
Westinghouse 155 127 4 - Ø11.5 14 A

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
28

● With sun shield hood (option code /H)


Unit : mm
274 ± 4 150 ± 3

150 ± 3
F13.ai

Material of HOOD : Aluminum


Check Valve (option code /CV), Stop valve (option code /SV) -specified Calibration gas inlet
Unit: mm

Detector case Detector case

Approx. 44

Approx. 100

40 Ø48

19
Full open height
Calibration gas inlet 58
Rc1/4 or 1/4NPT
Calibration gas inlet Rc1/4 or 1/4NPT

with Stop Valve (option: /SV) with Check Valve (option: /CV)

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
29
4. Model ZO21P Adapter for High Temperature Probe of separate type Oxygen Analyzer

Approx. 351 Unit: mm


Sample gas outlet 180 Flange (Thickness 5)
Rc1/2(Note2) JIS 5K 32 FF equivalent
Gasket (Thickness 1.5)

Ø60.5
Approx. 215

Ø127
Ø115
170

Ø60.5

85
ØA

110
t
Flange <1> Detector(ZR22G)
Approx. 100
L (Insertion length) (Note1)

Flange provided
Approx. 48 by customer
Ø52 over
Reference gas inlet Rc1/4 or 1/4NPT
69
High temperature
Probe SiC pipe
25

Ø30 42
Pipe hole (2- G1/2, 2-1/2 NPT, etc)
Calibration gas inlet Rc1/4 or 1/4 NPT
C
ØA
(Note 1) L= 0.5、0.6、0.7、0.8、0.9、1.0、1.5(m)
ØB (Note 2) Sample gas outlet
(if the sample gas pressure is negative, connect the auxiliary ejector.)

<1> Flange A B C t
JIS 5K 50 FF 130 105 4 - Ø15 14
JIS 10K 65 FF 175 140 4 - Ø18 18
JIS 10K 80 FF 185 150 8 - Ø19 18
JIS 10K 100 FF 210 175 8 - Ø19 18
ANSI Class 150 4 RF 228.6 190.5 8 - Ø19 24
ANSI Class 150 3 RF 190.5 152.4 4 - Ø19 24
ANSI Class 150 2 1/2 RF 177.8 139.7 4 - Ø19 24
JPI Class 150 3 RF 229 190.5 8 - Ø19 24
JPI Class 150 4 RF 190 152.4 4 - Ø19 24
DIN PN10 DN50 A 165 125 4 - Ø18 18
F12.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
30
5. E7046EC, E7046EN Auxiliary Ejector Assembly for High Temperature Detector of separate
type Oxygen Analyzer
Unit : mm

Pressure gauge assembly

Pressure gauge

39 Ejector
Needle valve
<2>
Nozzle (Note1)
Approx. 88

Ø43
Detector
Full open height
Approx. 67

Approx. 70
<3> R1/2
Instrument air inlet
<1>

38 20
40 Tee
Blow Rc1/4

(Note1) The connecter of ejector assembly is a


<1> Rc1/4 or 1/4NPT (F)
dedicated connecter with nozzle function.
<2> Ø6/Ø4 mm or 1/4 inch copper tube (stainless)
with ejector to connect
<3> R1/4 or 1/4 NPT (F)

F19.ai

6. Model ZO21R Probe Protector for Zirconia Oxygen Analyzers


Unit: mm
Flange <1>
(with bolts, nuts and washer)
Washer (M12) gasket (t3.0)
Gas flow Mounting nut (M12)
Ø60.5

ØB
ØA

SUS316 (JIS)
D
t ØB
l (Insert length) C
l=1050,1550,2050
Dimensions of holes
on opposing surface

Flange<1> A B C t D
JIS 5K 65 FF 155 130 4 - Ø15 5 40
ANSI Class 150 4 FF 228.6 190.5 8 - Ø19 12 50
F17.ai

7. K9471UA Dust Filter for Oxygen Analyzer


Unit : mm
Carborundum filter (SiC)
32 Detector
Ø51

Screw
10
Increasing of insertion length
F31.EPS

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
31
8. K9471UC Dust Guard Protector
Unit : mm
Increasing of insertion length
100

Ø50.8
4-Ø6

122
F11-1.ai

9. Model ZH21B Dust Protector for High Temperature Humidity Analyzers


440 Unit: mm
t
Ø76.3

ØA
Ø72

in case of JIS 5K 80A FF ØB


D
Install facing upwards.
in case of ANSI CLASS 150 4B FF

ØB
ØB

C
C
Hole dimensions on the opposite side

Flange A B C t D
JIS 5K 80 FF 180 145 4-Ø19 14 40
ANSI Class 150 4B FF 228.5 190.5 8-Ø19 12 50
F09.ai

10. Model ZO21S Standard Gas Unit


Unit : mm
92

Flow checker
253
228

Span gas valve


Zero gas valve
Gas outlet
1600

354

Zero gas cylinder (6 cylinder): E7050BA F24.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
32
11. Model ZA8F Flow setting unit for manual calibration
180 Unit : mm
ø6 Hole
140
7

REFERENCE CHECK

REFERENCE SPAN ZERO


235.8 2B mounting pipe

222.8

Calibration gas outlet


Reference gas outlet Span gas inlet
Zero gas inlet
32 Piping connection port A
OUT OUT IN IN IN
REF CHECK AIR SPAN ZERO

Model Piping connection port A


70
35 20 35 35 35 35 20 ZA8F-J*C 5 - Rc1/4

8 ZA8F-A*C 5 - 1/4 NPT (F)


Instrument air inlet

Weight : Approx. 2.3 kg


PIPNG INSIDE THE FLOW SETTING UNIT

CHECK REF
OUT OUT

Flowmeter Flowmeter

AIR IN

ZERO SPAN Instrument air


GAS IN GAS IN Approx. 1.5 l/min.

Air Set
Air pressure ;
without check valve ; sample gas pressure + approx. 50 kPaG
with check valve ; sample gas pressure + approx. 150 kPaG F2.6E.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
33
12. Model ZR40H Automatic Calibration Unit for Separate type Analyzer

2B pipe mounting example


Unit: mm

Wiring inlet ; 2-G1/2,Pg 13.5, M20 X 1.5 or 1/2 NPT(F)


(wiring inlet is at same position on rear)
*1 with four ISO M6 screws can wall-mount
90 116.5
26 54 71.5
*1
41.2 41.2 4 - Ø6.5 140

12
Connection port

49.5
Flowmeter

Needle valve OCK

Terminal box

223
250

Setting Valve for


reference gas

Setting Valve for


calibration gas

2B mounting pipe
Calibration gas outlet Zero gas inlet
Rc1/4 or 1/4 NPT(F) Rc1/4 or 1/4 NPT(F)
58 MAX
102

16
46
42
26

40 30 47.5 25

Reference gas outlet Reference gas inlet


Rc1/4 or 1/4 NPT(F) Rc1/4 or 1/4 NPT(F)

F35.ai

CHECK REF
ZR402G Converter ZR40H Automatic OUT
OUT
Calibration unit
AC-Z Zero Flowmeter Flowmeter
EV1
AC-S Solenoid valve
*2 *2 EV1, 2
AC-C
Span
EV2
EV1 EV2

F35_00.ai

AIR IN

ZERO GAS IN
Instrument air Approx.
1.5 l/min.
*2
Needle valve is supplied as accessory with flowmeter F35_01.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
34
13. Model ZR20H Automatic Calibration Unit for Integrated type Analyzer

Horizontal mounting on the ZR202G (-A)


Unit: mm

display side terminal box side

258

244
zero gas inlet
214 44 MAX 40 40 66.5 Rc1/4 or 1/4 NPT(F)

166.5
reference gas inlet
span gas inlet
Rc1/4 or 1/4 NPT(F)
Rc1/4 or 1/4 NPT(F)

Vertical mounting on the ZR202G (-B)


166.5

Span gas inlet


Rc1/4 or 1/4 NPT(F)
160

60
45

reference gas inlet zero gas inlet


180 44 MAX Rc1/4 or 1/ NPT(F) 40 40 66.5 Rc1/4 or 1/4 NPT(F)

F36.ai

14. L9852CB /G7016XH Stop Valve for Calibration gas line


Unit : mm
Ø48
(Full open length)
58

Rc1/4 or 1/4 NPT(F)


40
F15.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
35
15. K9292DN /K9292DS Check Valve for Calibration-gas line
Unit: mm
K9292DN : Rc1/4(A), R1/4(B)
K9292DS : 1/4 NPT(A), 1/4 NPT(Male)(B)

A
B

Approx. 19 Approx. 54

F30.EPS

16. G7003XF/K9473XK, G7004XF/K9473XG Air Set


Unit : mm Unit : mm
Panel cut dimensions View A
Horizontal Vertical
mounting mounting
22 ø15

+0.5 40
2-ø2.2 -0
2-ø6.5 40

2-M6 screw depth 8


max. 55

Panel (Horizontal mounting)

Secondary
pressure
Ø74 gauge Panel (Vertical mounting)
Max. 210

Primary Secondary
A
88

Approx. 122

G7003XF, G7004XF: Rc 1/4


K9473XK, K9473XG: 1/4NPT connector

17. G7001ZC Zero gas Cylinder

Unit : mm
485
325

ø140

(Note) Export of such high pressure filled gas cylinders to


most countries is prohibited or restricted.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
36
18. G7013XF, G7014XF Pressure Regulator for Gas Cylinder

Unit: mm
Secondary Primary
pressure gauge pressure gauge
Approx.112

Regulator handle
W22 (Right-
Stop valve handed screw)

ACH IN
TAK

IH
O

* Outlet Secondary
safety valve
Primary
safety valve
Approx. 59
Approx. 82

Approx. 163 Approx. 174

Part No. * Outlet


G7013XF Rc1/4
G7014XF 1/4 NPT (F)

19. E7044KF Case Assembly for Calibration gas Cylinder

Unit : mm

Pressure regulator
200
324 G7013XF/
180
G7014XF
Zero gas cylinder
(G7001ZC)
496

(158.3) 2B mounting pipe


( Ø60.5 )
(160)

The oblique line is an opening portion.

(Note)The zero gas cylinder and the regulator valve


are not included in the E7044KF (case assembly)

F23.ai

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
37
WIRING CONNECTIONS
Model ZR402G Separate type Zirconia Oxygen / Model ZR22G Separate type Zirconia Oxygen /
High Temperature Humidity Analyzer, Converter High Temperature Humidity Analyzer, Detector
Analog
output 1
4-20 mA DC Analog
Digital output output 2
(HART) 4-20 mA DC

1 2 3 4 5 6 7 8 9 10 11
FG AO1 AO1 AO2 AO2 CJ CJ TC TC CELL CELL CELL CELL TC TC CJ CJ
(+) (-) (+) (-) (+) (-) (+) (-) (+) (-) (+) (-) (+) (-) (+) (-)
12 13 14 15 16 17 18 19 20 21 22
FG DI-1 DI-2 DI-C AI AI AC- AC- AC- FG FG
(+) (-) ZERO SPAN COM

Contact
input 2
Span gas Solenoid valve for
Contact input 1 automatic calibration

*2 Solenoid valve for zero gas


for automatic calibration
Temperature Zero gas
transmitter
input 4-20 mA DC ZR40H Automatic Calibration Unit
23 24 25 26 27 28 29 30 31 32 33 34 35 36
DO-1 DO-1 DO-2 DO-2 DO-3 DO-3 DO-4 DO-4 HTR HTR L N G FG HTR HTR

*1
*1 100-240V AC
50/60 Hz *1
Contact Contact Contact Contact
output 1 output 2 output 3 output 4
F27.EPS

*1 Grand resistance is 100 ohm or less.


*2 Option (Temperature transmitter provide by user) for humidity measurement.

Model ZR202G Integrated type Zirconia Oxygen Analyzer

Contact output 1 Contact output 2


Contact input 1

Contact input 2

1 2 3 4 5 6 7
DI-1 DI-2 DI-C DO-1 DO-1 DO-2 DO-2

8 9 10 11 12 13 14
FG AO AO L N G FG ○
(+) (-)

Analog output 100 to 240 V AC, *1


4-20 mA DC 50 or 60 Hz
Digital output
(HART)
F28.EPS

The protective grounding for the analyzer shall be connected either the protective ground terminal
in the equipment or the ground terminal on the case.
Standard regarding grounding: Ground to earth, ground resistance: 100Ω or less.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
38

Inquiry Sheet for Models ZR22G, ZR402G, and ZR202G Direct In Situ Zirconia Oxygen
Analyzers and High Temperature Humidity Analyzers
Please place checkmarks in the appropriate boxes and fill in the necessary information in
the blanks.
1. General information
Customer Type of analyzer : □ Oxygen Analyzer □ High Temperature Humidity Analyzer
Destination of delivery □ Separate type □ Integrated type
Plant name Object : □ indication □ record □ control □ alarm
Measurement points Fuel : □ gas □ oil □ coal □
Power requirements V AC Hz
2. Process conditions
2.1 Measurement gas components
2.2 Oxygen concentration Nor. Min. Max. □ vol% O , 2 □
Moisture contents Nor. Min. Max. □ kg/kg, □ Vol% H O 2

2.3 Temperature Nor. Min. Max. □ °C, □


2.4 Pressure Nor. Min. Max. □ kPa, □
2.5 Gas flow Nor. Min. Max. □ m/sec, □
2.6 Dust type, Size Nor. Min. mm quantity
□ g/Nm , 3 □
2.7 Corrosive gas □ No gas □ Gas , quantity □ ppm, □
, quantity □ ppm, □
2.8 Combustible gas □ No gas □ Gas , quantity □ ppm, □
, quantity □ ppm, □
2.9 Others

3. Installation site conditions


3.1 Ambient temperature 1. Around Probe temp. from to °C, 2. Around Converter temp. from to °C
3.2 Vibration □ No vibration □ Vibration
3.3 1 Probe installation location □ Furnace □ Stack □ Others
2 Probe position □ Horizontal □ Vertical □ Others
□ Indoor □ Outdoor □ Covered
3 Probe insertion length (m) (Note) □ 0.4, □ 0.7, □ 1.0, □ 1.5, □ 2.0, □ 2.5, □ 3.0, □ 3.6, □ 4.2, □ 4.8, □ 5.4
4 Flange □ DIN □ ANSI □ Others
3.4 Instrument air supply □ Cannot be used. □ Can be used. kPa
3.5 Converter location □ Indoor □ Outdoor □ Covered (under roof)
3.6 Cable length between probe and converter meters

3.7 Calibration method □ Manual □ Automatic


(Note) 3.6m or more is available only in the U.S.
4. Quotation data

Quotation Qantity Description

Probe Model ZR22G General-use Probe Refer to the Probe Configuration


Model ZO21P-H High Temperature Probe Adapter for probe selection.

E7046EC /E7046EN Ejector Assembly for high temperature use.


Model ZH21B Dust Protector for High Temperature Humidity Analyzer.
Options
(for general Model ZO21R Probe Protector for Oxgen Analyzer
use)
K9471UA Filter for Oxgen Analyzer
Model ZR402G Separate type Analyzer, Converter
Model ZR202G Integrated type Zirconia Oxygen /High Temperature Humidity Analyzer
Model ZO21S Standard Gas Unit Select any one of Model ZO21S,
ZA8F, ZR40H, ZR20H.
Model ZA8F Flow Setting Unit
Model ZR40H, ZR20H Automatic Calibration Unit
L9852CB /G7016XH Stop Valve Not required if probe options are
specified.
K9292DN /K9292DS Check Valve (*1)
G7003XF/K9473XK, G7004XF/K9473XG Air Set
G7001ZC Zero Gas Cylinder (*2)
G7013XF /G7014XF Pressure Regulator
ZR22A, ZR202A Heater Assembly (Spare Parts)
T20.ai
*1 When Automatic Calibration of "-A" or "-B" code is specified, ZR20H is installed in ZR202G.
*2 Export of such high pressure filled gas cylinders to most countries is prohibited or restricted.

All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 11M12A01-01E 14th Edition Oct. 25, 2019-00
Subject to change without notice.