Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
CSH/CWH
CAT.NO.E15114
1
Three promises to guarantee performance
meeting ICH guideline for stability testing
CWH-40A
CSH-112
CSH-122
CSH-132
3
Full-view inner glass door Vacuum Insulation
Full-view inner glass door is equipped as standard. (CSH) HG model stability test chamber is the first
It lets you easily check on samples without temperature in the environment test chamber industry
and humidity fluctuation that caused by opening and to adopt vacuum insulation, which reduces
closing the outer door. the effect of ambient temperature.
CWH-40A CSH-122
4
Characteristics
Air blow
Clean humidifier
The growth of bacteria is suppressed with
a st r uct ure that maintains a high
humidifying water temperature. The
humidifier is also easier to clean. (CWH)
Easy maintenance
Maintenance work such as cleaning the
condenser fins and filling the water tank
can be easily operated. (CSH)
CSH chamber front
5
Characteristics
A new developed Virtual Air Jacket system Nine points in the effective space of 4
Air blown from below circulates along the ■ Temperature Set point: +40°C
chamber walls for stable airflow that is not 42
Temperature (°C)
Max-min range
Measurement
affected by specimen volume, etc. 41
(±2°C)
point
Storage conditions stipulated by ICH 40
39
guidelines are maintained, regardless of the
38
position in which specimens are located Run 10 min. 20 min. 30 min. 40 min.
(within effective dimensions). (CSH)
■ Humidity Set point: 75%rh
循環方式
80
● Air flow (Virtual Air Jacket) 79
77
Measurement
76
point
75
74
73
72
71
70
Run 10 min. 20 min. 30 min. 40 min.
5
Frost-free, continuous operationRefrigeration circuit 1 300 W
7 2
6
Evaporator 1
Double refrigerator circuit for ■ 温度 設定値:+40℃
reliable design 42
● 循環方式 Expansion mechanism Solenoid
最大最小幅
温度
valve
Two independent refrigerator circuits are 41
(バーチャルエアージャケット方式)
測定値
77
測定値
湿度
76
Evaporator frosting is prevented to allow
ダブル冷凍回路
75 (CWH)
continuous operation without interrupting (%rh) 74
test. 73 (±5%rh)
72
圧縮機1 6
71 300W
凝縮
70
Characteristics
Saving energy
● Power consumption comparison
The CSH-132, a newly developed large
・CSH Operating conditions: +40°C/75%rh capacity type, can be used with the same
CSH-112 power consumption as the CSH-122, which
CSH-110 (previous series) is expected to hold down running costs
during long-term usage. The maximum
CSH-122 load current has been reduced 36%
CSH-120 (previous series)
compared to previous model.
CSH-132
安定性試験器消費電力量比較
安定性試験器
新製品
従来器
C
消費電力量(W)
運転条件 500 550 600 650 700 750 800 850
200L A A
100mm 1240mm 100mm 省エネ率
+40℃/
75%rh
最大 % 36
400L
B
7
N-Instrumentation
Easy-to-use instrumentation
Unlike the smartphones, the controller
comes with resistive touchscreen, which
allows you to operate without taking off
your gloves.
Various items, including operation settings
and chamber setup, can be selected with
the tabs at the bottom of the screen.
Absolute temperature/
humidity limit alarm
This chamber is equipped with a standard
function to transmit an alarm when a
process value has deviated from the
temperature/humidity set points. The
temperature/humidity allowable range and
temperature/humidity stability time can
also be registered as desired. Registering
the ICH stability testing guideline
standards of ±2°C/±5%rh allows a rapid Allowable range
response when problems occur.
Email alert
When an alarm is triggered, an e-mail is
PV has deviated, alarm issued
sent to the registered PC or mobile address.
A notification can also be sent at the time
Instrumentation
of test completion.
* Requires an intranet environment capable of Name N-instrumentation P-200
sending emails. Operating modes Constant, program, stop
E-mail 通知 8
SPECIFICATIONS
Inner door (reinforced glass), Power cable (about 2m), Drain hose (x2), Temperature detect terminal (Pt100),
Accessories Humidity sensor terminal, External alarm terminal, Ethernet port (LAN port),
Through-hole for sensor (ø25 mm, right side), Quick joint for water circuit drain, Leveling feet casters (x4), Dew tray
Volume 235 L 470 L 794 L
Inside dimensions W600×H700×D560 mm W750×H950×D660 mm W1100×H950×D760 mm
Outside dimensions *3 W740×H1550×D774 mm W890×H1800×D874 mm W1240×H1800×D974 mm
Water supply Pure water (Conductivity 0.1 to 10 µS/cm)
Heat exhaust 1650 kJ/h 2130 kJ/h
Weight 200 kg 240 kg 370 kg
Maximum chamber load
Maximum 100 kg
capacity
Specimen shelf load capacity Maximum 30 kg/shelf (Equally distributed load)
Allowable ambient conditions Ambient temperature +5 to +40°C
Utility requirements
100V AC 1ø 13A
50/60 Hz
Power 220V AC 1ø
5.9A
supply 50/60 Hz
230V AC 1ø 5.7A
50 Hz
*1 The performance values are based on IEC 60068-3-6:2001.
Performance figures are given for a +5°C to +35°C ambient temperature, relative humidity 65±20%rh, rated power supply and no specimens inside the test area.
*2 Temperature and humidity maximum and minimum range means maximum difference after stabilization, at any moment in time in the working space against the
setting values; ambient temperature of +5 to +32°C, no load, no specimen.
*3 Excluding protrusions.
CSH-122/CSH-122HG: W650×D610 mm
70
CSH-132/CSH-132HG: W1000×D710 mm
60
• Cartridge fuse (7 A) ························································ 1 50
• Temperature-detecting terminal connector····················· 1 40
• Humidity-detecting terminal connector ··························· 1 30
20
• Socket adapter ································································ 1
10
• Filter for water ································································· 1 0
10 20 30 40 50 60 70 80 90 100
• Water tank (about 10 L)··················································· 1 35 Temperature °C 75
• Hose with quick joint ······················································· 1
Temperature/humidity control range
• Level gauge····································································· 1
• Silicon rubber plug ·························································· 1
• Operation manual (CD, Installation manual) ············ 1 set
9 ● 温湿度制御可能範囲
100
SPECIFICATIONS
ACCESSORIES DIMENSIONS
• Key (for door) ·································································· 2
Unit: mm
• Rubber plug (for 50ø cable port) ···································· 1
• Cartridge fuse (3 A) ························································ 2 40 A
• Temperature-detecting terminal connector····················· 1
20 A 30 A
• Humidity-detecting terminal connector ··························· 1
Air conditioning
compartment
system
2168
3088
1170
Test chamber
100
90
80 2838
75
70 1938
Relative humidity %rh
60 20 A 40 A
50 30 A
40
30
2248
1934
20
1479
10
0
10 20 30 40 50 60 70 80 90 100
25 Temperature °C
500 446
* Excluding protrusions.
● 温湿度制御可能範囲 10
SAFETY DEVICES Stability Test Chamber CSH SAFETY DEVICES Stability Test Chamber CWH
• Leakage breaker for power supply • Leakage breaker for power supply
• Short circuit protection fuse for control circuit • Short circuit protection fuse for control circuit
• Electrical compartment door switch • Electrical compartment door switch
• Chamber thermal fuse • Chamber thermal fuse
• Humidifier boil-dry protector • Humidifier boil-dry protector
• Temperature switch for air circulator • Temperature switch for air circulator
• Refrigerator overcurrent protection • Refrigerator overcurrent protection
• Overheat protector • Overheat protector
• Temperature burn-out circuit • Temperature burn-out circuit
• Humidity burn-out circuit (with built-in temperature/humidity controller)
Safety precautions
11
OPTIONS Stability Test Chamber CSH
• 220 V AC 1ø 2 W 50/60 Hz Records the temperature and humidity of each In case of power failure, power is supplied
• 230 V AC 1ø 2 W 50 Hz section such as the temperature inside the chamber. to the temperature/humidity recorder and
Rocation: Left side, right side or lower left humidity sensor, and test area temperature/
* External dimensions change when attaching the humidity is recorded.
recorder at thel eft or right side.
Direct water coupling to tap water (Please refer to the recorder location.)
Recharge time: 12 h
Backup time: 40 min.
Data saving cycle: 5 sec.
A water circuit to supply pure water
Internal memory: 4 MB (Nonvolatile flash memory)
continuously to the chamber.
External recording media:
• Pure water coupling with pressure- Thermocouple
CF memory card port (Includes a 256 MB CF card)
reducing valve
USB memory port
• Pure water coupling without pressure- Attached to specimen to measure
< Temperature & humidity type >
reducing valve specimen temperature.
Temperature 1, Humidity 1
Thermocouple with a brass ball tip
(4 more channels can be turned ON)
Thermocouple type T (Copper/Copper-Nickel)
Water purifier (reverse osmosis) •2m
•4m
Use to continuously supply pure water. •6m
• WS-1
Produced water capacity: 12 L/h
Size: W480 × H400 × D280 mm Anchoring fixtures
WS-1
190
1440
Lower left • CD
下部固定
• Booklet
Shelf/shelf bracket
13
http://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500