Sei sulla pagina 1di 9

ARTICLE IN PRESS

Journal of Biomechanics 37 (2004) 1361–1369

Three-dimensional load measurements in an external fixator


.
K. Seide*, N. Weinrich, M.E. Wenzl, D. Wolter, C. Jurgens
Berufsgenossenschaftliches Unfallkrankenhaus (Trauma Hospital) Hamburg, Bergedorfer StraX e 10, 21033 Hamburg, Germany
Accepted 16 December 2003

Abstract

On the basis of a six-degree-of-freedom adjustable fracture reduction hexapod external fixator, a system which can be used for
measuring axial and shear forces as well as torsion and bending moments in the fixator in vivo was developed. In a pilot study on 9
patients (7 fresh fractures and 2 osteotomies of the tibia), the load in the fixator during the healing process was measured after 2, 4, 8
and 12 weeks and at fixator removal. The measured values enabled both the type of fracture to be determined as well as the
monitoring of the healing process. In well-reduced type A3 fractures small axial (direction of the bone axis) forces were found in the
fixator. A2, B2 and C3 fractures showed distinct axial forces, which decreased during the healing process, according to an increasing
load transfer over the bone. Bending moments in the fixator showed good correspondence with the clinical healing process, except in
the case of a C3 fracture. A combination of bending moment and axial force proved to be particularly suitable to assess fracture
healing. In transverse fractures, the well-known resorption phenomenon of bone in the fracture gap at approximately 4 weeks was
detected by the system. Compared with other external fixator load measurements in vivo, the hexapod offers the advantage of being
able to measure all forces and moments in the fixator separately and with a relatively simple mechanical arrangement. In our
opinion, it will be possible to control fracture healing using this system, thereby minimizing radiation exposure from radiographs.
Furthermore, the measurement system is a step towards the development of external fixator systems that enable automatic
adjustments of the callus mechanical situation (‘‘automatic dynamization’’) and inform the patients about the optimal weight
bearing of their extremity (‘‘intelligent fixator’’).
r 2004 Elsevier Ltd. All rights reserved.

Keywords: Load measurements; Fracture healing; External fixator; Hexapod

1. Introduction investigated in robotics and has been applied in position


control applications, for example, in flight simulators for
Monitoring of the fracture healing process is routinely the three-dimensional positioning of the cockpit (Stewart,
done by radiographic examinations which permit form 1965). The hexapod mechanism in an external fixator
and intensity of callus calcification to be determined. allows a computer-assisted universal three-dimensional
However, after primary demonstration of an exact fracture reduction or deformity correction (Seide and
reduction, the mechanical properties of the osteosynth- Wolter, 1996; Seide et al., 1999). By adding 6 uniaxial
esis are the main determinants for therapeutic decisions. force-measuring elements, the hexapod was extended into
Estimation of mechanical callus strength from radi- a six-spacial-degree-of-freedom measuring fixator.
ological images depends largely on the expertise of the The goal was to construct a system for monitoring
surgeon, but in some cases the identification of healing fracture healing by mechanical measurement and thus
problems by radiological means is very difficult. minimize radiation exposure. Therefore parameters
A new system for in vivo three-dimensional load which optimally describe an emerging callus had to be
measurements during fracture healing in an external identified. Furthermore, it had to be shown that
fixator is introduced in this study. It is based on a hexapod different fracture types in patients can be distinguished
mechanism also known as parallel robot manipulator using the system. The hypothesis of the study was that
or Stewart platform. This mechanism has been well conclusions drawn from the mechanical parameters
*Corresponding author. Tel.: +49-40-7306-2701; fax: +49-40-7306-
measured are consistent with the decisions of the
2703. treating surgeons made on the basis of the correspond-
E-mail address: k.seide@buk-hamburg.de (K. Seide). ing routine radiographs.

0021-9290/$ - see front matter r 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.jbiomech.2003.12.025
ARTICLE IN PRESS
1362 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369

2. Material and methods

The system is based on a modified ring fixator.


Instead of parallel interconnecting longitudinal bars,
three pairs of ball joints are attached at the proximal
and distal ring. The joints are connected by six linear
actuator elements (‘‘distractors’’) in the form of
circularly arranged triangles (Figs. 1 and 2). Bone
fixation was performed either by Kirschner wires
(Ilizarov ring fixator configuration) or with Schanz’
screws. The mechanism was also mounted secondarily to
AO, Hoffman or Orthofix fixators.
Force sensors (Burster, measuring range 1000 N with
an accuracy of 1% of the final value and a maximum
load of 3000 N) were inserted in line with the distractor
elements of the hexapod fixator. This insertion resulted
in measuring distractors, which were mechanically
compatible with the original non-measuring ones. The
sensors were connected to a PC-based measurement
system (National Instruments, DAQ AT-MIO-64F-5).
A measuring software was developed using Microsoft
Visual Basics.
The peculiarity of the hexapod is based on the fact
that the ball joints that connect the rings to the linear
actuators or force sensor elements are not blocked. So
bending, torsional or shear load transfer is theoretically
not possible along these elements and there is exactly
one load component (axial) carried and measured by
each unit. Additionally, according to the theory of the
hexapod kinematics, there is no interrelationship be- Fig. 2. Patient during measurement of fixator forces and moments
under axial load of the leg.
tween the units and thus exactly six independent degrees
of freedom are allowed.
From the force values FD1 yFD6 measured in the six
distractor elements, the components of the load L in the as well as the moments Mx ; My ; Mz with respect to the
fixator, i.e., the axial force Fz and the shear forces Fx ; Fy three spacial axes can be calculated. The calculation
provides transformation of six-degree-of-freedom linear
force values to 6-degree-of-freedom spatial force and
moment values. The transformation was mathematically
realized with a matrix multiplication of the distractor
force vector with a 6  6 transformation matrix:
0 1
Fx
B C
B Fy C
B C
BF C
B z C
L ¼B C
B Mx C
B C
BM C
@ yA
Mz
0 10 1
f1x f2x f3x f4x f5x f6x FD1
B CB C
B f1y f2y f3y f4y f5y f6y CB FD2 C
B CB C
B f f2z f3z f4z f5z f6z C B C
B 1z CB FD3 C
¼B CB C: ð1Þ
B m1x m2x m3x m4x m5x m6x CB FD4 C
B CB C
Bm CB C
@ 1y m2y m3y m4y m5y m6y A@ FD5 A
Fig. 1. Measuring hexapod external fixator.
m1z m2z m3z m4z m5z m6z FD6
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1363

z In order to facilitate the evaluation of the quantities


Mz
(i.e., by eliminating the direction), the shear forces Fx ; Fy
and the bending moments Mx ; My were summarized as
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
Fxy ¼ Fx2 þ Fy2 ; ð5Þ
fulcrum K oi qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
Fz Mxy ¼ Mx2 þ My2 : ð6Þ
ri

Mxy Fi The accuracy of the overall system, i.e., mechanical


arrangement and software, was investigated. A typical
Di Fxy hexapod configuration for the lower leg (symmetrical
callus assembly, 150 mm ring diameter, 130 mm ring distance)
bone
was exposed to defined forces and moments in a
ui material testing machine (Zwick1455). Fixator forces
y and moments were calculated from the distractor values.
This procedure showed maximum errors of 11%. An
x explanation for the relatively high error value is that the
Fground sensors available in the clinic’s laboratory had a range of
Fig. 3. Sketch of the experimental model and illustration of the 1000 N (measuring error 1%) each, resulting in an
parameters used in the mathematical computations of the measure- unfavorable approx. 5000 N measuring range for the
ment software. whole system. Also, the involvement of 6 sensors and
errors from determination of the mounting parameters
of the hexapod have to be considered. The accuracy was
The elements of the matrix depend on the geometrical considered appropriate, comparing well with deviations
arrangement of the hexapod in the individual case. caused by biological influences normally encountered in
fix ; fiy ; fiz in Eq. (1) correspond to the components clinical investigation. Additionally, the force-measuring
of a unit vector f i in the direction of the measured force, distractors were routinely calibrated every month to
i.e., the direction of the respective distractor. For the avoid errors caused for example by mechanical damage.
calculation of f i ; first the forward kinematics is solved, Clinical measurements were performed on 9 patients
which means the computation of the ring position of the (Table 1), in which fresh fractures (o2 weeks) or
hexapod from ring sizes, distractor lengths and ball joint osteotomies with acute correction (i.e., without callus
positions relative to the rings (Seide et al., 2002). Then distraction osteoneogenesis) were treated solely with an
the positions of the joints oi and ui in the global external fixator. Typically, non-complicated crural
coordinate system are computed (Fig. 3).1 From this f i is fractures are stabilized internally when possible or by
given as a secondary exchange of the external fixator by an
f i ¼ oiDui i with Di ¼ joi  ui j: ð2Þ internal implant. So, the investigated cases represent a
selected sample (e.g., polytrauma, open fractures, severe
For the computation of the moments the definition of a soft-tissue damage) and the number of patients available
fulcrum is necessary. The software is laid out in such a for the measurement was therefore limited. In 7 cases
way that moments can be computed both with respect to fractures and in 2 cases osteotomies of the tibia were
the ring center point and with respect to a clinically given. An assembly with Kirschner wires in Ilizarov’s
determined point (e.g., fracture location, assuming an technique was used in 6 cases, and an assembly with
idealized solid connection between bone and ring). The Schanz’ screws in 3 cases. For Ilizarov’s assembly, full
position vector of this point is designated as K. Thus weight bearing was allowed; for Schanz’ screws assem-
the lever arm ri for the forces working on a distractor i is bly, 10 kg partial load bearing was permitted, according
the spacial difference between K and the upper joint to our usual clinical practice.
position oi : In all cases, a fracture reduction had been performed
with the hexapod mechanism (Seide and Wolter, 1996;
ri ¼ oi  K: ð3Þ
Seide et al., 1999). The hexapod elements were ex-
Due to the definition of the moment this leads to changed by lighter standard fixator elements (long-
itudinal rods, joints) after the primary reduction as well
m i ¼ ð oi  K Þ  f i ð4Þ
as after each measurement session and remounted
with the components mix ; miy ; miz ; to be used in Eq. (1). before the next. This was performed for better patient
1
o corresponds to German ‘‘oberer Ring’’, which means upper ring, comfort, as the measurement elements with its
u corresponds to German ‘‘unterer Ring’’, which means lower ring cables were not suitable for outpatient activities.
(Seide et al., 2002). Additionally, the remaining measurement sensors would
ARTICLE IN PRESS
1364 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369

Table 1
Patients

Diagnosis (classification after AO) Time to fixator Clinical decision after Further course of healing, remarks
removal (days) 12 weeks

1 Osteotomy proximal lower leg (equals A3) 112 Leave fixator in place Later discussion of X-rays concluded that
consolidation could have been assumed already
after 8 weeks
2 Osteotomy proximal lower leg (equals A3) 169 Leave fixator in place Consolidation in fixator, measurement at time of
fixator removal was not possible
3 1 Open shaft fracture middle third (A3) 153 Leave fixator in place Consolidation in fixator with partial bone defect,
temporary brace with 10 kg weight bearing after
fixator removal
4 1 Open shaft fracture middle third (A3) 93 Fixator removal Fracture not consolidated at fixator removal,
because of a pin consolidation in brace
infection and
application of a brace
5 1 Open fracture proximal lower leg (A2) 105 Fixator removal, full Immediate mobilization without crunches or
weight bearing other safety measures
6 Oblique shaft fracture middle third (A2) 144 Leave fixator in place Consolidation in fixator, then immediate
mobilization without crunches or other safety
measures after fixator removal
7 Oblique shaft fracture middle third (A2) 108 Fixator removal and 4 weeks plaster cast with full weight bearing for
full weight bearing precautionary reasons
8 2 Open oblique shaft fracture middle/ 130 Leave fixator in place Consolidation in fixator with partial bone defect,
distal third (B2) temporary brace with 10 kg weight bearing after
fixator removal
9 2 Open shaft fracture middle/distal third 270 Leave fixator in place Cancellous bone graft necessary, consolidation in
with long comminution zone (C3) fixator, after fixator removal cast for
precautionary reasons

have superposed the fracture zone when taking radio- that no pain appeared. Maximal loads applied were
graphs in many cases. Mounting parameters of the individually different and ranged between 100 and
hexapod were noted and the hexapod was re-mounted 400 N, found in fresh and healed fractures, respectively.
for each session in an identical manner. For the application of the bending load, the patient was
The hexapod elements were mounted force free placed supine with a free-hanging fixator and the heel
(controlled by measurement, hanging leg). The standard supported by the ground sensor. At the fracture gap
fixator rods were removed afterwards. The measuring location, a ventro-dorsal force was applied manually.
hexapod took over the fixator load completely during The load was limited in such a way that no pain in
the measurements, including inherent stresses of the the fracture arose (up to 60 N). The effective lever arm,
bone-soft-tissue-fixator mounting. i.e., the distance between the point of force application
The measurements were performed 2, 4, 8 and 12 and the heel was noted. The measurements under
weeks after the trauma or the osteotomy. If the fixator bending loads were not performed in the first 4 patients,
remained longer than 12 weeks, a further measurement since this measurement procedure was later included in
was made before removal of the fixator. An influence of the protocol. The measurement at 2 weeks was missing
the measurements on the healing process was excluded. for two patients because the clinical condition of the
Neither were fragment position and fixator assembly patients after a polytrauma did not allow for a
modified during the measurements, nor were the results measurement in the laboratory. In one patient the
of the measurements used for clinical decisions. measurement before the fixator removal could not be
Measurements were performed under axial and completed due to organizational reasons.
bending loads of the leg. The external axial force was The patients were informed in detail (‘‘informed
measured with a one-degree-of-freedom vertical ground consent’’) and participated in very motivated manner
sensor positioned under the heel. For the application of in the sessions, which lasted about 2 h.
axial loads (direction of the bone axis) the patient was The evaluations were performed with the Excel
instructed to stand relaxed with a stretched leg and to spread-sheet program (Microsoft). For the measure-
put weight on the heel, slowly increasing and decreasing ments under axial load, linear regression lines were
the ground force. Three repetitions were performed. The computed with respect to the ground force. The
total measuring period was 20 s. The scanning rate was regression coefficient (upward slope gradient) obtained
0.2 s. The patients were instructed to limit the load so describes the stiffness of the bone–fixator system. Each
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1365

regression calculation included three load cycles (100 0.8 1 A3


values, 20 s, as described above). To allow judgement of 2 A3
the repeatability of measurements, the standard error of 3 A3
0.6 4 A3
the regression coefficient was determined. The calcu-

Fz/Fground
lated regression coefficient values were charted against
0.4
the healing time. The fixator bending moment under
external bending load was charted with respect to the
external bending load so that the evaluation was 0.2
performed in the mode of a relative bending stiffness.
The y-axis-intersections of the regression lines are to 0
be interpreted as offset stresses. These stresses, which 2 4 8 12 fixator
are, e.g., due to primary pressure or tension on the (a) weeks removal
fracture site, were excluded by evaluating the slope
0.8
values of the regression lines in this study.
5 A2
6 A2
0.6 7 A2

Fz/Fground
3. Results 8 B2
9 C3
0.4
3.1. Forces and moments under axial load of the
extremity
0.2

Linear relationships between ground force Fground and


0
bending moment Mxy as well as ground force Fground and 2 4 8 12 fixator
axial force Fz in the fixator were found in the range of (b) weeks removal
the applied loads (R2 between 0,7 and 0,99, po0:01).
Fig. 4. Axial force in the fixator under axial load of the leg during the
The regression coefficients were calculated. These were course of fracture healing. Each point represents the regression
found to be particularly suitable for the clinical coefficient between fixator axial force and ground force. In type A3
measurements to monitor fracture healing. Different fractures values increase at 4 weeks and are then small (a). In A2 and
behaviors of these parameters were found for different B2 fractures larger and in a C3 fracture (b) maximum values are found,
fracture types and different healing progression (Figs. 4, decreasing during the healing process. Error bars represent the
standard error of the regression coefficient.
6, 8 and 9). The standard errors of the regression
coefficients were relatively small, due to the high number
of 100 single measurements included in each regression remained after 12 weeks. This correlated with the
coefficient. The applied method of evaluating the clinical decisions: a further fixator treatment and a
regression between external load and fixator seems to cancellous bone graft in the later course of healing were
be a favorable approach to enhance the accuracy in the necessary.
clinical experiment, as compared to single readings. The bending moment in the fixator under axial load of
Axial fixator forces were small in A3 (transverse) the leg also showed a clear dependence on the type of
fractures over the entire healing process (Fig. 4a). In all fracture (Fig. 6a and b). Even more clearly than for the
cases less than 10% of the ground force load was axial force measurements three of the four A3 cases
transferred by the fixator; the axial load transfer was showed an increase of the fixator bending load between
performed by the fractured bone. In A2, B2 and C3 the second and the fourth week and a decrease there-
fractures distinct axial loads at the beginning and a after.
decreasing load transfer during fracture healing ap- In two cases (patients 1 and 7) a high bending stability
peared in the fixator (Fig. 4b). In 5 cases a slight increase of the fracture resulted already after 8 weeks. In cases 5
around the fourth week was noted. and 7 (Fig. 7) maximally 1.5 Nmm/N resulted at 12
In the case of the C3 (comminuted, Fig. 5) fracture, weeks. Correlating with this measured high callus
64% of the external axial load was transferred by the stability were the radiographic findings and the clinical
fixator at the beginning. Since the fracture was probably decisions to immediately mobilize under full load after
not load carrying at this time, for the other 36% a load- fixator removal. The other patients received a temporary
carrying function of the soft tissues (Schmidt et al., brace after fixator removal for 3 months as a safety
1995) has to be assumed. The decrease of the load measure to protect from peak loads. In patient 6, a low
transfer in the fixator, i.e., an increase of the load value at 12 weeks and a slight further decrease under
transfer over the healing bone, was also found in this ongoing fixator treatment was observed. Here, the
case, in particular, between the fourth and the twelfth measurements showed that fixator removal would have
week (Fig. 4b). Contrary to the other cases 15% already been possible after 12 weeks. From the
ARTICLE IN PRESS
1366 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369

30 1 A3
2 A3
25 3 A3

Mxy/Fground (Nmm/N)
4 A3
20

15

10

0
2 4 8 12 fixator
(a) weeks removal

30 5 A2
6 A2
25 7 A2

Mxy/Fground (Nmm/N)
8 B2
20 9 C3
15

10

0
2 4 8 12 fixator
(b) weeks removal
Fig. 5. Anteroposterior and lateral radiographs of a C3 fracture, 12
Fig. 6. Bending load in the fixator under axial load of the leg during
weeks after trauma (patient 9). The Fracture is not consolidated, the
the course of fracture healing. Each point represents the regression
load measurements show a decreasing, but still significant load transfer
coefficient between fixator bending moment and ground force. Values
in the fixator. Note that the radiographs do not show the hexapod
decrease in A3 (a), A2 and B2 fractures (b). In contrast, consistently
elements, since these were mounted before and replaced by standard
small bending moments are seen in C3 fractures (b). Error bars
rods after the measurement sessions in the lab (see text).
represent the standard error of the regression coefficient.

measuring curves for patient 1, it had to be assumed that between 12 weeks and fixator removal. Both these
a fixator removal would have been possible after 8 decreases correlate to the curves found in the bending
weeks. A renewed discussion of the radiographs with the moments.
treating surgeon confirmed this assumption.
In contrast, in two cases (patients 3 and 4) a flattening 3.2. Bending moment under bending load of the extremity
of the curve was observed. In case 4 the highest
measurement value of all patients at 12 weeks In 5 cases a bending load case was examined in
(9 Nmm/N) correlated with a pin infection, which then addition to the axial load case (Fig. 9). In all patients a
led to the removal of the fixator. Likewise, in case 3 a decrease of the bending moment transferred in the
relatively high bending moment was present after 12 fixator was found during fracture healing. This corre-
weeks. The fixator was left in place. sponds to a shift in the bending stiffness in the overall
A characteristic pattern showed the comminuted system from the fixator to the bone. The inter-individual
fracture (C3, patient 9). Here the fixator bending comparison indicated that the courses of the values were
moment was in a low range over the entire healing similar to the corresponding curves under axial load.
process. In this case only the combination with the However, the temporary increase in instability in cases 3
measuring curve of the axial force Fz showed that the and 4 at 4 weeks is not apparent. In the case of patient 4,
fracture was slowly consolidating. in whom the fixator had to be removed because of a pin
Measurements with the hexapod allow the determina- infection, an increase of the fixator bending load was
tion of shear forces in the fixator. Values of the order of found parallel to pin loosening. As under axial load, the
magnitude 1/20 of the axial load (Fig. 8) resulted. C3-fracture (patient 9) showed a spacial behavior with
Fracture consolidation was found to reduce shear forces almost constant values up to 8 weeks, but followed by a
similar to the other parameters. For example, patient 7 slight decrease. Patient 8, who showed distinctly more
showed a definite decrease at 8 weeks and patient 8 fixator bending moment than the other patients at 12
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1367

1 1 A3
3 A3
0.8 6 A2
4 A3

My/Mload
0.6 8 B2
9 C3
0.4

0.2

0
2 4 8 2 fixator
weeks removal

Fig. 9. Bending moment in the fixator under bending load of the leg
during the course of fracture healing. Each point represents the
regression coefficient between bending moment measured in the fixator
and applied bending moment, with respect to the transverse axis. The
results correspond to the measurements of the bending moment under
axial load. However, the temporary instability in A3 fractures after 4
weeks is not apparent. Error bars represent the standard error of the
regression coefficient.

Fig. 7. Anteroposterior and lateral radiograph of an A2 fracture, 12


weeks after trauma (patient 7). The fracture is consolidated, and the
load measurements show small forces and moments in the fixator. weeks under axial load, also showed persisting high
Note that the radiographs do no show the hexapod elements, as these values until 12 weeks under bending load.
were mounted before and replaced by standard rods after the
measurement sessions in the lab (see text).

4. Discussion

Clinical in vivo applications of mechanical measure-


0.1 1 A3 ments in the process of the fracture healing have been
2 A3 .
published. Typically, the techniques used (Jorgensen,
0.08 3 A3
4 A3
1972; Burny et al., 1984; Cunningham et al., 1987; Evans
Fxy/Fground

0.06 et al., 1988) measure bending moments in unilateral


external fixators with sensors on longitudinal rods. With
0.04 the Fraktometer (Claes, 1991) the fracture gap move-
ment is unidirectionally measured. Compared with these
0.02
in vivo load measurements in external fixators, the
0 hexapod offers the advantage that the fixator load can
2 4 8 12 fixator be measured in all six spacial degrees of freedom, which
(a) weeks removal
are 3 forces and 3 moments. Measurements of fracture
healing in six degrees of freedom have been realized so
0.1 5 A2 far in some very special internal implants (nail (Schnei-
6 A2
0.08 7 A2 der et al., 1990), spine fixator (Rohlmann et al., 1995)).
8 B2 The characteristic of the hexapod (also known as
Fxy/Fground

9 C3
0.06 ‘‘parallel robot mechanism’’) is based on the fact that
exactly one degree of freedom is carried and measured
0.04 by each measurement unit. According to the theory of
the hexapod kinematics no interrelation between the
0.02
elements is given. So, there are six degrees of freedom in
0 space on the one hand and six independent degrees of
2 4 8 12 fixator freedom load carrying/measuring elements on the other.
(b) weeks removal
A mathematical relationship is given between the linear
Fig. 8. Shear forces in the fixator under axial load of the leg for A3 (a), force measurements in the hexapod fixator and the six
A2, B2, and C3 fractures (b) during the course of fracture healing.
load degrees of freedom (axial, shear, bending, tor-
Each point represents the regression coefficient between fixator shear
force and ground force. Values are of the order of magnitude 1/20 of sional) in the Cartesian coordinate system.
the axial load. Error bars represent the standard error of the regression The fixator load situation is different from other
coefficient. fixators, where, e.g., in a ring fixator 3 rods are fixed on
ARTICLE IN PRESS
1368 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369

the rings in all degrees of freedom, summing up to 18 the interfragmentary movement was found. A correlat-
degrees of freedom and thus resulting in prestresses ing bone absorption was determined histologically
between the rods. Common clinical experience is that (Claes et al., 1989).
these stresses can be substantial. Bending measurements under external bending loads
In the performed measurements a distribution be- showed the course of callus stiffness, comparable to
tween the applied limb load and the forces and moments those under axial load. However, the former were less
in the fixator was determined. As a measure for the sensitive, e.g., the temporary instability in A3 fractures
callus formation, it would be most interesting, to after 4 weeks was not apparent. Thus a conclusion of the
compute an ‘‘absolute fracture stiffness’’ (Evans et al., study is that the latter is to be recommended for clinical
1988). For this a mechanical model would have to be practice. However, it is also possible that the mode of
applied. Theoretically, measurement values depend on bending load application in the study was not optimal
dimensions and material properties of callus, Kirschner and a special load frame might improve the accuracy as
wires and/or Schanz’ screws as well as the carrying well as the sensitivity of these measurements.
elements of the fixator. However, as opposed to in vitro A difference between earlier investigations with
or animal experiments, many simplifying assumptions external fixators and the investigations with the hexapod
would have to be made in such calculations, when one fixator is given. Cases examined with the ‘‘Fraktometer’’
deals with real clinical situations, e.g., fracture gaps showed maximum gap widths of 5–6 mm in type A and
found in a clinical situation are considerably irregular type B fractures and average gap widths of 10 mm in
and linearity for axial force in assemblies with tensioned cases with type C fractures or delayed healing fractures
wires cannot be presumed. Also, the authors are aware (Wentzensen, 1995). Due to the exact reduction, which is
of the fact that soft tissues play an additional mainly typically performed when the hexapod is used (Seide
unknown role in the load distribution. Therefore, only a et al., 1999), a fragment contact has to be assumed.
summarizing parameter rather than a biomechanically Therefore, in A3 fractures, i.e., with an almost
precisely defined callus stiffness can be measured. So, transverse gap, only a small axial load was measured
the main question to be answered by this study was, in the fixator, which means that the axial load was
whether, despite these theoretical limitations, the transferred by the bone from the beginning. However,
hexapod fixator load measurements could be useful the fixator was substantially loaded on bending under
as—relative—indicators of the fracture healing progress. axial load of the extremity, which is explainable by
In our opinion—with reservation due to the relatively ‘‘buckling’’ in the fracture site. This is increasingly
small number of measured patients—this can be stabilized during the healing process through endostal
assumed. and periostal callus. The measured fixator bending loads
The optimal measurement parameter was dependent decreased accordingly. The hypothesis of an axial
on the type of the fracture. In transverse fractures (A3) deformation in the fracture gap under load, which
the fixator bending moment was most predictive, decreases during fracture healing, can be supported with
whereas the fixator axial load gave nearly no relevant type B and C fractures by the hexapod measurements. In
information. In contrast, in a comminuted fracture A2 (simple oblique) fractures, a mechanism of the
fixator axial load was found to be the only meaningful interfragmentary movement in the sense of sliding and
predictor. The evaluation of the healing process by tilting is conceivable, which explains the measured
means of a combination of the parameters bending simultaneous changes of the axial force and the bending
moment and axial force in the fixator under axial load of moment in the fixator during the healing process.
the extremity was possible in all examined patients. So, In vivo interfragmentary movements by patients with
the measurements suggest that two well-selected degrees ring and hexapod fixators have been documented and
of freedom are the minimum criteria necessary for the reported by Duda et al. (Duda et al., 2002, 2003). The
mechanical monitoring of fracture consolidation. The interfragmentary movement rather than the load in
measurements agreed with the radiologically diagnosed the fracture site has been shown to be important for the
stability and the treatment measures taken after 12 healing process (Claes et al., 1997). For this reason it has
weeks (e.g., removal of the fixator, continuing treatment to be discussed, whether the deformation or the load
in the fixator, unprotected or protected weight bearing), needs to be measured, in order to utilize the measure-
which were accomplished without knowledge of the ment values concerning the influence and optimization
measurement results. of the callus’ mechanical situation. In contrast to
It is interesting that the measured reduction of fracture healing measurements in transverse osteotomies
stability after 4 weeks in transverse osteotomies and under laboratory conditions in animals (Claes et al.,
transverse fractures can be explained by the results of 1997) callus deformation values are more difficult to
experimental investigations of fracture gap movement in access in humans, because their measurement requires
sheep. For fracture gap widths between 0.5 and 1 mm invasive techniques, e.g., the insertion of additional
and relative movements of 18%, a temporary increase of Schanz’ screws. On the other hand, there is a mechanical
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1369

relationship between deformation and loads, depending .


Claes, L., Wilke, H.J., Rubenacker, S., 1989. Interfragmentary strain
on the stiffness parameters of the involved structures and bone healing—an experimental study. 35th annual meeting,
(bone, fixator elements, callus). Thus, either measure- Orthopaedic Research Society, February 6–9, 1989, Las Vegas,
Nevada.
ment should be appropriate. This should be especially Claes, L., Augat, P., Suger, G., Wilke, H.J., 1997. Influence of size and
true in relative measurements during the course of stability of the osteotomy gap on the success of fracture healing.
healing. Journal of Orthopedic Research 15, 577–584.
After developing the measuring external fixator a Cunningham, J.L., Evans, M., Harris, J.D., Kenwright, J., 1987. The
pilot study in nine patients was carried out. The measurement of stiffness of fractures treated with external fixation.
Engineering in Medicine 16 (4), 229–232.
measured parameters resulted in curves in good agree- Duda, G.N., Kirchner, H., Wilke, H.J., Claes, L., 1998. A method to
ment with the healing process. This leads to the determine the 3-D stiffness of fracture fixation devices and its
conclusion that it will be possible to control fracture application to predict inter-fragmentary movement. Journal of
healing using a measurement fixator system, minimizing Biomechanics 31, 247–252.
Duda, G.N., Sollmann, M., Sporrer, S., Hoffmann, J.E., Kassi, J.-P.,
the use of radiographs with their radiation exposure.
Khodadadyan, C., Raschke, M., 2002. Interfragmentary motion in
However, this has to be proven in further investigations. tibial osteotomies stabilized with ring fixators. Clinical Orthopae-
The future goal is the development of intelligent dics and Related Research 396, 163–172.
implants and external fixators, which automatically Duda, G.N., Bartmeyer, B., Sporrer, S., Taylor, W.R., Raschke, M.,
adjust load distribution in the callus according to the Haas, N.P., 2003. Does partial weight bearing unload a healing
results of measurements (‘‘automatic dynamization’’) bone in external ring fixation? Langenbecks Archive 388,
298–304.
and thus optimize fracture healing. In addition, it can Evans, M., Kenwright, J., Cunningham, J.L., 1988. Design and
inform the patients about the optimal load of their performance of a fracture monitoring transducer. Journal of
extremity (‘‘intelligent fixator’’, Wolter, 1991; Wolter Biomedical Engineering 10, 64–68.
and Seide, 1997). A precursor of such systems was an .
Jorgensen, T.E., 1972. Measurements of stability of cruaral fractures
external fixator provided with force sensors for the treated with Hoffmann osteotaxis. Acta Orthopaedica Scandina-
vica 43, 207–218.
application of controlled longitudinal forces in an .
Muller, K.H., Bowe, K.-H., Becker, J., 1982. Kraftbestimmung bei
external compression osteosynthesis (Muller . et al., externer Kompressionsosteosynthese am Frakturspalt des
1982). Duda (Duda et al., 1998) developed a method Menschen. Unfallheilkunde 85, 321–337.
for the description of the spacial rigidity of a fixator Rohlmann, A., Bergmann, G., Graichen, F., Weber, U., 1995. In vivo
system over a rigidity matrix of 36 elements. It is measurement of implant loads in a patient with a fractured
vertebral body. European Spine Journal 4, 347–353.
conceivable to develop a fixator with controllable .
Schmidt, H.G.K., Wolter, D., Sasse, S., Schneider, E., Schumann, U.,
rigidity matrix using motors and flexible elements in a .
Jurgens, C., 1995. Die Lastverteilung im Ringfixateur bei
hexapod and adjust it as a function of fracture type and segmentalem Knochendefekt. In: Wolter, D., Hansis, M., Have-
measured healing status. Further, by applying the mann. D. (Hrsg.) (Eds.), Externe und interne Fixateursysteme.
hexapod system to control the acting forces during a Springer, Berlin, pp. 37–43.
Schneider, E., Michel, M.C., Genge, M., Perren, S.M., 1990. Loads
callus distraction (Aronson and Harp, 1994; Wolfson acting on an intramedullary femoral nail. In: Bergmann, G.,
et al., 1990), the distraction time and quality of the Rohlmann, A., Graichen, F. (Hrsg.) (Eds.), Implantable Telemetry
developing bone could be optimized. in Orthopaedics. Forschungsermittlung der FU, Berlin.
Seide, K., Wolter, D., 1996. Universelle dreidimensionale Korrektur
und Reposition mit dem Ringfixateur unter Anwendung der
Acknowledgements Hexapod-Anordnung. Unfallchirurg 6, 422–424.
Seide, K., Wolter, D., Kortmann, H.R., 1999. Fracture reduction and
deformity correction with the hexapod Ilizarov fixator. Clinical
The authors wish to thank the Biomechanics Section Orthopaedics and Related Research 363, 186–195.
of the Technical University Hamburg-Harburg for help .
Seide, K., Wollnack, J., Weinrich, N., Jurgens, C., 2002. Theorie und
in manufacturing the sensor-elements and Mrs. Sarah Software des Hexapod Fixateurs externe. Biomedizinische Technik
Strametz for correcting the translation. 47, 326–333.
Stewart, D., 1965. A platform with six degrees of freedom. Proceedings
of the Institute of Mechanical Engineers 180, 371–378.
Wentzensen, A., 1995. In-vivo-Messung der Knochenheilung bei
References Fixateur-externe-Osteosynthesen. In: Wolter, D., Hansis, M.,
Havemann, D., (Hrsg.) (Eds.), Externe und interne Fixationssys-
Aronson, J., Harp, J.H., 1994. Mechanical forces as predictors of teme. Springer, Berlin, pp. 107–111.
healing during tibial lengthening by distraction osteogenesis. Wolfson, N., Hearn, T.C., Thomason, J.J., Armstrong, P.F., 1990.
Clinical Orthopaedics and Related Research 301, 73–79. Force and stiffness changes during Ilizarov leg lengthening. Clinical
Burny, F., Donkerwolcke, M., Bourgois, R., Domb, M., Saric, O., Orthopaedics and Related Research 125, 58–60.
1984. Twenty years experience in fracture healing measurement Wolter, D., 1991. Das intelligente Implantat. Abstraktband BIOME-
with strain gauges. Orthopedics 7, 1823–1826. CHANIKA 1991, Hamburg.
Claes, L., 1991. Die Messung der Knochenheilung bei Fixateur- Wolter, D., Seide, K., 1997. Das intelligente Implantat. In: Schneider
externe-Osteosynthesen mit dem Fraktometer FM 100. Chirurg 62, E. (Hrsg.) (Eds.), Biomechanik des menschlichen Bewegungsappar-
354–355. ats. Vol. 261, Hefte zur Zeitschrift Der Unfallchirurg, pp. 236–240.

Potrebbero piacerti anche