Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Abstract
On the basis of a six-degree-of-freedom adjustable fracture reduction hexapod external fixator, a system which can be used for
measuring axial and shear forces as well as torsion and bending moments in the fixator in vivo was developed. In a pilot study on 9
patients (7 fresh fractures and 2 osteotomies of the tibia), the load in the fixator during the healing process was measured after 2, 4, 8
and 12 weeks and at fixator removal. The measured values enabled both the type of fracture to be determined as well as the
monitoring of the healing process. In well-reduced type A3 fractures small axial (direction of the bone axis) forces were found in the
fixator. A2, B2 and C3 fractures showed distinct axial forces, which decreased during the healing process, according to an increasing
load transfer over the bone. Bending moments in the fixator showed good correspondence with the clinical healing process, except in
the case of a C3 fracture. A combination of bending moment and axial force proved to be particularly suitable to assess fracture
healing. In transverse fractures, the well-known resorption phenomenon of bone in the fracture gap at approximately 4 weeks was
detected by the system. Compared with other external fixator load measurements in vivo, the hexapod offers the advantage of being
able to measure all forces and moments in the fixator separately and with a relatively simple mechanical arrangement. In our
opinion, it will be possible to control fracture healing using this system, thereby minimizing radiation exposure from radiographs.
Furthermore, the measurement system is a step towards the development of external fixator systems that enable automatic
adjustments of the callus mechanical situation (‘‘automatic dynamization’’) and inform the patients about the optimal weight
bearing of their extremity (‘‘intelligent fixator’’).
r 2004 Elsevier Ltd. All rights reserved.
0021-9290/$ - see front matter r 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.jbiomech.2003.12.025
ARTICLE IN PRESS
1362 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369
Table 1
Patients
Diagnosis (classification after AO) Time to fixator Clinical decision after Further course of healing, remarks
removal (days) 12 weeks
1 Osteotomy proximal lower leg (equals A3) 112 Leave fixator in place Later discussion of X-rays concluded that
consolidation could have been assumed already
after 8 weeks
2 Osteotomy proximal lower leg (equals A3) 169 Leave fixator in place Consolidation in fixator, measurement at time of
fixator removal was not possible
3 1 Open shaft fracture middle third (A3) 153 Leave fixator in place Consolidation in fixator with partial bone defect,
temporary brace with 10 kg weight bearing after
fixator removal
4 1 Open shaft fracture middle third (A3) 93 Fixator removal Fracture not consolidated at fixator removal,
because of a pin consolidation in brace
infection and
application of a brace
5 1 Open fracture proximal lower leg (A2) 105 Fixator removal, full Immediate mobilization without crunches or
weight bearing other safety measures
6 Oblique shaft fracture middle third (A2) 144 Leave fixator in place Consolidation in fixator, then immediate
mobilization without crunches or other safety
measures after fixator removal
7 Oblique shaft fracture middle third (A2) 108 Fixator removal and 4 weeks plaster cast with full weight bearing for
full weight bearing precautionary reasons
8 2 Open oblique shaft fracture middle/ 130 Leave fixator in place Consolidation in fixator with partial bone defect,
distal third (B2) temporary brace with 10 kg weight bearing after
fixator removal
9 2 Open shaft fracture middle/distal third 270 Leave fixator in place Cancellous bone graft necessary, consolidation in
with long comminution zone (C3) fixator, after fixator removal cast for
precautionary reasons
have superposed the fracture zone when taking radio- that no pain appeared. Maximal loads applied were
graphs in many cases. Mounting parameters of the individually different and ranged between 100 and
hexapod were noted and the hexapod was re-mounted 400 N, found in fresh and healed fractures, respectively.
for each session in an identical manner. For the application of the bending load, the patient was
The hexapod elements were mounted force free placed supine with a free-hanging fixator and the heel
(controlled by measurement, hanging leg). The standard supported by the ground sensor. At the fracture gap
fixator rods were removed afterwards. The measuring location, a ventro-dorsal force was applied manually.
hexapod took over the fixator load completely during The load was limited in such a way that no pain in
the measurements, including inherent stresses of the the fracture arose (up to 60 N). The effective lever arm,
bone-soft-tissue-fixator mounting. i.e., the distance between the point of force application
The measurements were performed 2, 4, 8 and 12 and the heel was noted. The measurements under
weeks after the trauma or the osteotomy. If the fixator bending loads were not performed in the first 4 patients,
remained longer than 12 weeks, a further measurement since this measurement procedure was later included in
was made before removal of the fixator. An influence of the protocol. The measurement at 2 weeks was missing
the measurements on the healing process was excluded. for two patients because the clinical condition of the
Neither were fragment position and fixator assembly patients after a polytrauma did not allow for a
modified during the measurements, nor were the results measurement in the laboratory. In one patient the
of the measurements used for clinical decisions. measurement before the fixator removal could not be
Measurements were performed under axial and completed due to organizational reasons.
bending loads of the leg. The external axial force was The patients were informed in detail (‘‘informed
measured with a one-degree-of-freedom vertical ground consent’’) and participated in very motivated manner
sensor positioned under the heel. For the application of in the sessions, which lasted about 2 h.
axial loads (direction of the bone axis) the patient was The evaluations were performed with the Excel
instructed to stand relaxed with a stretched leg and to spread-sheet program (Microsoft). For the measure-
put weight on the heel, slowly increasing and decreasing ments under axial load, linear regression lines were
the ground force. Three repetitions were performed. The computed with respect to the ground force. The
total measuring period was 20 s. The scanning rate was regression coefficient (upward slope gradient) obtained
0.2 s. The patients were instructed to limit the load so describes the stiffness of the bone–fixator system. Each
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1365
Fz/Fground
lated regression coefficient values were charted against
0.4
the healing time. The fixator bending moment under
external bending load was charted with respect to the
external bending load so that the evaluation was 0.2
performed in the mode of a relative bending stiffness.
The y-axis-intersections of the regression lines are to 0
be interpreted as offset stresses. These stresses, which 2 4 8 12 fixator
are, e.g., due to primary pressure or tension on the (a) weeks removal
fracture site, were excluded by evaluating the slope
0.8
values of the regression lines in this study.
5 A2
6 A2
0.6 7 A2
Fz/Fground
3. Results 8 B2
9 C3
0.4
3.1. Forces and moments under axial load of the
extremity
0.2
30 1 A3
2 A3
25 3 A3
Mxy/Fground (Nmm/N)
4 A3
20
15
10
0
2 4 8 12 fixator
(a) weeks removal
30 5 A2
6 A2
25 7 A2
Mxy/Fground (Nmm/N)
8 B2
20 9 C3
15
10
0
2 4 8 12 fixator
(b) weeks removal
Fig. 5. Anteroposterior and lateral radiographs of a C3 fracture, 12
Fig. 6. Bending load in the fixator under axial load of the leg during
weeks after trauma (patient 9). The Fracture is not consolidated, the
the course of fracture healing. Each point represents the regression
load measurements show a decreasing, but still significant load transfer
coefficient between fixator bending moment and ground force. Values
in the fixator. Note that the radiographs do not show the hexapod
decrease in A3 (a), A2 and B2 fractures (b). In contrast, consistently
elements, since these were mounted before and replaced by standard
small bending moments are seen in C3 fractures (b). Error bars
rods after the measurement sessions in the lab (see text).
represent the standard error of the regression coefficient.
measuring curves for patient 1, it had to be assumed that between 12 weeks and fixator removal. Both these
a fixator removal would have been possible after 8 decreases correlate to the curves found in the bending
weeks. A renewed discussion of the radiographs with the moments.
treating surgeon confirmed this assumption.
In contrast, in two cases (patients 3 and 4) a flattening 3.2. Bending moment under bending load of the extremity
of the curve was observed. In case 4 the highest
measurement value of all patients at 12 weeks In 5 cases a bending load case was examined in
(9 Nmm/N) correlated with a pin infection, which then addition to the axial load case (Fig. 9). In all patients a
led to the removal of the fixator. Likewise, in case 3 a decrease of the bending moment transferred in the
relatively high bending moment was present after 12 fixator was found during fracture healing. This corre-
weeks. The fixator was left in place. sponds to a shift in the bending stiffness in the overall
A characteristic pattern showed the comminuted system from the fixator to the bone. The inter-individual
fracture (C3, patient 9). Here the fixator bending comparison indicated that the courses of the values were
moment was in a low range over the entire healing similar to the corresponding curves under axial load.
process. In this case only the combination with the However, the temporary increase in instability in cases 3
measuring curve of the axial force Fz showed that the and 4 at 4 weeks is not apparent. In the case of patient 4,
fracture was slowly consolidating. in whom the fixator had to be removed because of a pin
Measurements with the hexapod allow the determina- infection, an increase of the fixator bending load was
tion of shear forces in the fixator. Values of the order of found parallel to pin loosening. As under axial load, the
magnitude 1/20 of the axial load (Fig. 8) resulted. C3-fracture (patient 9) showed a spacial behavior with
Fracture consolidation was found to reduce shear forces almost constant values up to 8 weeks, but followed by a
similar to the other parameters. For example, patient 7 slight decrease. Patient 8, who showed distinctly more
showed a definite decrease at 8 weeks and patient 8 fixator bending moment than the other patients at 12
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1367
1 1 A3
3 A3
0.8 6 A2
4 A3
My/Mload
0.6 8 B2
9 C3
0.4
0.2
0
2 4 8 2 fixator
weeks removal
Fig. 9. Bending moment in the fixator under bending load of the leg
during the course of fracture healing. Each point represents the
regression coefficient between bending moment measured in the fixator
and applied bending moment, with respect to the transverse axis. The
results correspond to the measurements of the bending moment under
axial load. However, the temporary instability in A3 fractures after 4
weeks is not apparent. Error bars represent the standard error of the
regression coefficient.
4. Discussion
9 C3
0.06 ‘‘parallel robot mechanism’’) is based on the fact that
exactly one degree of freedom is carried and measured
0.04 by each measurement unit. According to the theory of
the hexapod kinematics no interrelation between the
0.02
elements is given. So, there are six degrees of freedom in
0 space on the one hand and six independent degrees of
2 4 8 12 fixator freedom load carrying/measuring elements on the other.
(b) weeks removal
A mathematical relationship is given between the linear
Fig. 8. Shear forces in the fixator under axial load of the leg for A3 (a), force measurements in the hexapod fixator and the six
A2, B2, and C3 fractures (b) during the course of fracture healing.
load degrees of freedom (axial, shear, bending, tor-
Each point represents the regression coefficient between fixator shear
force and ground force. Values are of the order of magnitude 1/20 of sional) in the Cartesian coordinate system.
the axial load. Error bars represent the standard error of the regression The fixator load situation is different from other
coefficient. fixators, where, e.g., in a ring fixator 3 rods are fixed on
ARTICLE IN PRESS
1368 K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369
the rings in all degrees of freedom, summing up to 18 the interfragmentary movement was found. A correlat-
degrees of freedom and thus resulting in prestresses ing bone absorption was determined histologically
between the rods. Common clinical experience is that (Claes et al., 1989).
these stresses can be substantial. Bending measurements under external bending loads
In the performed measurements a distribution be- showed the course of callus stiffness, comparable to
tween the applied limb load and the forces and moments those under axial load. However, the former were less
in the fixator was determined. As a measure for the sensitive, e.g., the temporary instability in A3 fractures
callus formation, it would be most interesting, to after 4 weeks was not apparent. Thus a conclusion of the
compute an ‘‘absolute fracture stiffness’’ (Evans et al., study is that the latter is to be recommended for clinical
1988). For this a mechanical model would have to be practice. However, it is also possible that the mode of
applied. Theoretically, measurement values depend on bending load application in the study was not optimal
dimensions and material properties of callus, Kirschner and a special load frame might improve the accuracy as
wires and/or Schanz’ screws as well as the carrying well as the sensitivity of these measurements.
elements of the fixator. However, as opposed to in vitro A difference between earlier investigations with
or animal experiments, many simplifying assumptions external fixators and the investigations with the hexapod
would have to be made in such calculations, when one fixator is given. Cases examined with the ‘‘Fraktometer’’
deals with real clinical situations, e.g., fracture gaps showed maximum gap widths of 5–6 mm in type A and
found in a clinical situation are considerably irregular type B fractures and average gap widths of 10 mm in
and linearity for axial force in assemblies with tensioned cases with type C fractures or delayed healing fractures
wires cannot be presumed. Also, the authors are aware (Wentzensen, 1995). Due to the exact reduction, which is
of the fact that soft tissues play an additional mainly typically performed when the hexapod is used (Seide
unknown role in the load distribution. Therefore, only a et al., 1999), a fragment contact has to be assumed.
summarizing parameter rather than a biomechanically Therefore, in A3 fractures, i.e., with an almost
precisely defined callus stiffness can be measured. So, transverse gap, only a small axial load was measured
the main question to be answered by this study was, in the fixator, which means that the axial load was
whether, despite these theoretical limitations, the transferred by the bone from the beginning. However,
hexapod fixator load measurements could be useful the fixator was substantially loaded on bending under
as—relative—indicators of the fracture healing progress. axial load of the extremity, which is explainable by
In our opinion—with reservation due to the relatively ‘‘buckling’’ in the fracture site. This is increasingly
small number of measured patients—this can be stabilized during the healing process through endostal
assumed. and periostal callus. The measured fixator bending loads
The optimal measurement parameter was dependent decreased accordingly. The hypothesis of an axial
on the type of the fracture. In transverse fractures (A3) deformation in the fracture gap under load, which
the fixator bending moment was most predictive, decreases during fracture healing, can be supported with
whereas the fixator axial load gave nearly no relevant type B and C fractures by the hexapod measurements. In
information. In contrast, in a comminuted fracture A2 (simple oblique) fractures, a mechanism of the
fixator axial load was found to be the only meaningful interfragmentary movement in the sense of sliding and
predictor. The evaluation of the healing process by tilting is conceivable, which explains the measured
means of a combination of the parameters bending simultaneous changes of the axial force and the bending
moment and axial force in the fixator under axial load of moment in the fixator during the healing process.
the extremity was possible in all examined patients. So, In vivo interfragmentary movements by patients with
the measurements suggest that two well-selected degrees ring and hexapod fixators have been documented and
of freedom are the minimum criteria necessary for the reported by Duda et al. (Duda et al., 2002, 2003). The
mechanical monitoring of fracture consolidation. The interfragmentary movement rather than the load in
measurements agreed with the radiologically diagnosed the fracture site has been shown to be important for the
stability and the treatment measures taken after 12 healing process (Claes et al., 1997). For this reason it has
weeks (e.g., removal of the fixator, continuing treatment to be discussed, whether the deformation or the load
in the fixator, unprotected or protected weight bearing), needs to be measured, in order to utilize the measure-
which were accomplished without knowledge of the ment values concerning the influence and optimization
measurement results. of the callus’ mechanical situation. In contrast to
It is interesting that the measured reduction of fracture healing measurements in transverse osteotomies
stability after 4 weeks in transverse osteotomies and under laboratory conditions in animals (Claes et al.,
transverse fractures can be explained by the results of 1997) callus deformation values are more difficult to
experimental investigations of fracture gap movement in access in humans, because their measurement requires
sheep. For fracture gap widths between 0.5 and 1 mm invasive techniques, e.g., the insertion of additional
and relative movements of 18%, a temporary increase of Schanz’ screws. On the other hand, there is a mechanical
ARTICLE IN PRESS
K. Seide et al. / Journal of Biomechanics 37 (2004) 1361–1369 1369