Sei sulla pagina 1di 7

MAX-302 Xenon Light Source 300W

Technical Information
MAX-302 Technical Information

MAX-302
UV Light Guide
(250-385nm) (Quartz) *Please regard the following data as a reference.

Measuring Condition
Quartz Light Guide
Sensitivity of Detector Irradiation Distance φ5mmx1m
100 S254
Relative Spectral

80 S365
Sensitivity

60
40
Detector
20 Xenon Light Source MAX-302
A.Accumulated UV Power Meter UIT-150 S254(Ushio)
0 B.Accumulated UV Power Meter UIT-150 S365(Ushio) ・UV Lamp
200 300 400 C.Power Meter MODEL2835C (Newport) ・UV Mirror Module
Wavelength(nm)

Relati

Spectral Distribution
1
This graph shows the spectral distribution of
0.8 MAX-302 with UV Lamp and UV Mirror
Module measured by fiber spectrometer.
Spectral Irradiance
(Relative Value)

0.6

0.4

0.2

0
250 300 350 400 450 500
Wavelength(nm)

Illuminated Area & Center Illuminance at Different Distance

Irradiance(mW/㎝2) Irradiance(mW/㎝2)
Detector A B C
Filter
Distance --- --- XHQA254 XHQA320 XHQA380
Illuminated area
10mm φ9mm 664.00 1699.00 --- --- ---
20mm φ13mm 309.00 700.00 --- --- ---
30mm φ18mm 160.00 342.00 9.10 33.90 36.70
40mm φ24mm 95.50 196.00 6.74 24.90 27.10
50mm φ30mm 63.10 128.00 4.91 18.20 20.00
60mm φ35mm 44.60 90.10 3.75 13.60 15.10
70mm φ40mm 32.60 65.40 2.89 10.50 11.70
80mm φ45mm 25.10 50.10 2.30 8.37 9.34
90mm φ50mm 19.70 39.60 1.87 6.82 7.64
100mm φ53mm 16.00 32.10 1.54 5.60 6.29

Reference for XHQA (Bandpass Filter)


http://www.asahi-spectra.com/opticalfilters/bandpass_filter.html

1
MAX-302 Technical Information

MAX-302
UV Light Guide Collimator Lens
(250-385nm) (Quartz) (x 1.0) *Please regard the following data as a reference.

Measuring Condition
Collimator Lens Quartz Light Guide
Sensitivity of Detector x 1.0 φ5mmx1m
100 S254 Irradiation Distance
Relative Spectral

80 S365
Sensitivity

60
40
Detector
20 A.Accumulated UV Power Meter UIT-150 S254(Ushio) Xenon Light Source MAX-302
0 B.Accumulated UV Power Meter UIT-150 S365(Ushio) ・UV Lamp
200 300 400 C.Power Meter MODEL2835C (Newport) ・UV Mirror Module
Wavelength(nm)

Spectral Distribution
1
This graph shows the spectral distribution of
0.8 MAX-302 with UV Lamp and UV Mirror Module
Spectral Irradiance
(Relative Value)

measured by fiber spectrometer.


0.6

0.4

0.2

0
250 300 350 400 450 500
Wavelength (nm)

Illuminated Area & Center Illuminance at Different Distance

Irradiance(mW/㎝2) Irradiance(mW/㎝2)
Detector A B C C C
Filter
Distance --- --- XHQA254 XHQA320 XHQA380
Illuminated area
80mm 21 x 21mm 35.10 73.00 2.13 7.92 8.99
100mm 26 x 26mm 24.00 49.50 1.49 5.44 6.21
200mm 48 x 48mm 6.83 13.80 0.48 1.65 1.88
300mm 71 x 71mm 3.16 6.42 0.23 0.78 0.90
400mm 93 x 93mm 1.82 3.69 0.13 0.46 0.53
500mm 116 x 116mm 1.19 2.42 0.09 0.30 0.35
600mm 140 x 140mm 0.83 1.69 0.06 0.21 0.24
1000mm 230 x 230mm 0.29 0.59 0.02 0.08 0.09

Reference for XHQA (Bandpass Filter)


http://www.asahi-spectra.com/opticalfilters/bandpass_filter.html

2
MAX-302 Technical Information

MAX-302
VIS Light Guide
(385-740nm) (Quartz) *Please regard the following data as a reference.

Measuring Condition
Sensitivity of Detector QuartzQuartz
Collimator 石英ファイバー
Lens Light Guide
Lightguide
Sensitivity Characteristics of
ディテクタの感度特性 1.0倍 φ5×1m φ5×1m
100 T-10M irradiation Distance
Irradiation
照射距離 Distance φ5mmx1m
Detector
100
Relative Spectral

100
80 S254
S254
Spectral irradiance
相対分光感度 Value)

80
80 S365
Sensitivity

S365
60
60
60 T-1M
T-1M
ディテクタ
Detector
(Relative

40
40
40
20
Detector
A.高速分光ユニット
A.High-resolution Spectrometer
HSU-100S (朝日分光製)
HSU-100S (Asahi Spectra)
20
20
B.紫外線積算光量計
B.Accumulated UV actinometer
UIT-150 S254/S365(ウシオ電機製)
UIT-150 S254/S365(Ushio Inc) Xenon LightLight
Xenon Source MAX-302
Source MAX-
キセノン光源 MAX-302
0 A.Illuminance Meter T-10M (Konica Minolta)
0
0200200 300
300 400
400 500
500 600
600 700
700 800
800
C.Light
C.照度計
B.Power Meter
T-1M
MeterT-1M (Konica Minolta)
(ミノルタ製)
MODEL2835C (Newport) ・UV 302
・UVランプ
Lamp
300 400 Wavelength(nm)
500 600 700 800 900
波長(nm) D.Power
D.パワーメーターMeter MODEL2835C
MODEL2835C (New Port)
(ニューポート製) ・VIS・UV
・VISミラーモジュール
Lamp
Mirror Module
Wavelength(nm) ・UV Mirror Module

Spectral Distribution
1
This graph shows the spectral distribution of
0.8 MAX-302 with UV Lamp and VIS Mirror
Spectral Irradiance
(Relative Value)

0.6 Module measured by fiber spectrometer.


0.4

0.2

0
300 400 500 600 700 800 900
Wavelength (nm)

Illuminated Area & Center Illuminance at Different Distance

Center Illuminance(lx) Irradiance(mW/㎝2)


Detector A B
Filter ---
Distance XBPA400 XBPA550 XBPA700
Illuminated area
10mm φ9mm 10,253,000 --- --- ---
20mm φ13mm 6,897,000 --- --- ---
30mm φ18mm 4,474,000 40.00 51.80 41.00
40mm φ24mm 2,983,000 27.80 36.30 28.90
50mm φ30mm 2,051,000 19.50 25.50 20.30
60mm φ35mm 1,491,000 14.40 18.90 15.10
70mm φ40mm 1,137,000 10.90 14.30 11.40
80mm φ45mm 895,000 8.55 11.30 9.02
90mm φ50mm 713,000 6.93 9.12 7.28
100mm φ53mm 583,000 5.79 7.63 6.11

Reference for XBPA (Bandpass Filter)


http://www.asahi-spectra.com/opticalfilters/bandpass_filter.html

Uniformity of Irradiation
Distance: 20mm Distance: 50mm Distance: 100mm
90-100 90-100 90-100
80-90 80-90 80-90
-10 70-80 70-80 70-80
-24 -38
60-70 60-70 60-70
-6 -14 -22
50-60 50-60 50-60
-2 -4 -6
40-50 40-50 40-50
2 6 10
30-40 30-40 30-40
6 20-30 16 20-30 26 20-30
42
10

26

30
19
7

18
12
4

10-20
6
5

10-20 10-20 2
1

-6

10
-2

26
-2

-18
-9
-5

-30
-16
-8

-424
-23

0-10 0-10 0-10

3
MAX-302 Technical Information

MAX-302
VIS Light Guide Collimator Lens
(385-740nm) (Quartz) (x 1.0) *Please regard the following data as a reference.

Measuring Condition
Collimator Lens Quartz Light Guide
ロッドレンズ 石英ファイバー
石英ファイバー
Sensitivity of Detector
ディテクタの感度特性 x 1.0 φ5mmx1m
ディテクタの感度特性 Irradiation
照射距離 Distance
照射距離 1.0倍 φ5×1m φ5×1m
100 T-10M
100
100
Relative Spectral

S254
S254
80
相対分光感度

80
80 S365
S365
Sensitivity

60
60
60 T-1M
T-1M
40
40
40
ディテクタ
Detector
A.高速分光ユニット HSU-100S (朝日分光製)
20
20
20 A.Illuminance
B.紫外線積算光量計 Meter T-10M (Konica
UIT-150 Minolta)
S254/S365(ウシオ電機製) Xenon Light Source MAX-302
キセノン光源 MAX-302
000 B.Power
200
200 300 400
300 400 500
500 600
600 700
700 800
800 C.照度計MeterT-1M MODEL2835C
(ミノルタ製) (Newport) ・UV ・UVランプ
Lamp
300 400 500 600 700 800 900
波長(nm)
波長(nm) D.パワーメーター MODEL2835C (ニューポート製) ・VIS ・VISミラーモジュール
Mirror Module
Wavelength(nm)

Spectral Distribution
1
This graph shows the spectral distribution
0.8 of MAX-302 with UV Lamp and VIS Mirror
Spectral Irradiance
(Relative Value)

0.6 Module measured by fiber spectrometer.


0.4

0.2

0
300 400 500 600 700 800 900
Wavelength (nm)

Illuminated Area & Center Illuminance at Different Distance

Center Illuminance(lx) Irradiance(mW/㎝2)


Detector A B
Filter ---
Distance XBPA400 XBPA550 XBPA700
Illuminated area
80mm 21 x 21mm 1,318,000 12.70 16.40 13.70
100mm 26 x 26mm 898,000 8.79 11.30 9.56
200mm 48 x 48mm 265,000 2.56 3.30 2.77
300mm 71 x 71mm 122,400 1.20 1.55 1.29
400mm 93 x 93mm 70,800 0.69 0.89 0.74
500mm 116 x 116mm 46,200 0.45 0.58 0.48
600mm 140 x 140mm 32,100 0.31 0.40 0.33
1000mm 230 x 230mm 11,700 0.11 0.15 0.12
Reference for XBPA (Bandpass Filter)
http://www.asahi-spectra.com/opticalfilters/bandpass_filter.html

Uniformity of Irradiation
Distance: 100mm Distance: 200m Distance: 500mm
90-100
80-90 90-100 90-100
80-90 80-90
70-80
-11 -24 70-80 70-80
60-70
-7 -16 60-70 -44 60-70
50-60
-3 -8 50-60 -24 50-60
40-50 0 -4
1 40-50 40-50
5 30-40 8 20
30-40 30-40
9 20-30 16 20-30 40 20-30
13

24

60
9

16

40
5

20

10-20
1

4
0

13 60
-4

10-20 10-20
-3

-8

-24
-7

-16

-44
-11

-242

0-10 0-10 0-10

4
MAX-302 Technical Information

Comparison of Spectrum

7
6 UV Mirror Module

Irradiance[μW/(cm2・nm)]
VIS Mirror Module
5
UV-VIS Mirror Module
4 IR Mirror Module
3
2
1
0 Note:
200 300 400 500 600 700 800 900 1000 1100 1200 Irradiated Area: 400 x 400mm
Wavelength(nm)

Lamp Start-Up Characteristics

120
Relative Intensity(%)

110

100

90

80

70
0 1 2 3 4 5 6 7 8 9 10
Time(min)

Lamp Fluctuation

103
Relative Intensity(%)

102

101

100

99

98

97
0 30 60 90 120 150 180
Time(sec)

Lamp Life

100

80
Relative Intensity(%)

60

40

20

0
0 100 200 300 400 500 600 700 800 900 1000
Continuous Lighting Time(hrs)

5
*We accept no responsibility for any errors or omissions, and we reserve the right to modify design, characteristics and products at any time without obligation.