Sei sulla pagina 1di 2

IES LM-79-08

geometry). They are not operated; only the auxiliary With knowledge of srel(m) and the relative spectral
lamp is operated. The auxiliary lamp can be a halogen power distribution STEST(m) of the SSL product under
or incandescent lamp or a white LED source. test, the spectral mismatch correction factor F is
given by:
9.2.6 Determination of f1' and Spectral Mismatch
Correction Factor The spectral responsivity of the y m SREF (m)srel(m)dm y m STEST (m)V (m)dm
F= (9)
integrating sphere photometer cannot be perfectly
matched to the V(l) function. An error (called spec-
y m SREF (m)V(m)dm y m STEST (m)srel (m)dm
tral mismatch error) occurs when a test SSL product
where, SREF(m) is the spectral distribution of the refer-
has a different spectral power distribution from that
ence standard source. Spectral mismatch errors can
of the standard source. The f1' value10 is an index that
be corrected by multiplying the factor F to the mea-
indicates the degree of mismatch in spectral respon-
sured lumen value of the SSL product. The accuracy
sivity, and the value (in %) gives a rough indication
of STEST(m) is generally not very critical, and thus, the
of the magnitude of errors that can occur for general
nominal spectral distribution of a product may be used.
white light sources, but errors can be larger than the
f1' value for SSL products consisting of only a few
For further details on f1' and spectral mismatch cor-
narrowband emissions.
rection, see Refs. 10 and 7.
To determine f1' value, the relative spectral respon-
9.2.7 Calibration The integrating sphere photometer
sivity of the total sphere system srel(m) must be
shall be calibrated against total luminous flux stan-
obtained. srel(m) is given as a product of rela-
dards traceable to an NMI.
tive spectral responsivity of the photometer head
sph,rel(m) and the relative spectral throughput of the
9.3 Goniophotometer
sphere Trel(m):
Goniophotometers are normally used for measure-
srel(m)= sph,rel (m) Trel(m) (6)
ment of luminous intensity distribution, from which
total luminous flux can be obtained.
The sph,rel(l) should be measured in hemispheri-
cal illumination geometry. If it is measured only in
9.3.1 Type of Goniometer Goniophotometers shall
normal direction, uncertainty should be determined.
be the type that maintains the burning position
Trel(l) is theoretically given by
unchanged with respect to gravity; therefore, only
ta (m) Type C goniophotometers9 are allowable. Type C
Trel(m)= k • (7)
1- ta (m) goniophotometers include moving detector gonio-
meters and moving mirror goniophotometers. Care
where ta (m) is the average reflectance of the entire
should be exercised to prevent light reflected from
inner sphere surfaces (including t=0 for an opening,
the mechanical structure of the goniophotometer
if there is one) and k is a normalization factor. If ta (m)
or any other surface including secondary reflec-
of the integrating sphere in use is measured accu-
tions from surfaces of the SSL product itself from
rately, Trel(m) may be obtained using this equation.
reaching the photodetector. The speed of rotation
However, integrating spheres in use are more or less
of the positioning equipment should be such as to
contaminated and the data of coating samples tend
minimize the disturbance of the thermal equilibrium
to deviate from the reflectance of the real sphere
of the SSL product.
surface. Therefore, it is recommended that Trel(l) be
directly measured on the integrating sphere using
9.3.2 Principle of Total Luminous Flux Measure-
the procedures given in the Annex B of Ref. 7.
ment By measuring the luminous intensity distribu-
tion I(i,z) of the source, the total luminous flux is
Once srel(l) is determined, f1' is calculated by
obtained by
y | |
m s*rel (m) – V (m) dm
2p p
f1' =  100% with U= y z=0 y i = 0 I(i,z)sini di dz (10)
y
m V (m) dm (8)
If the photometer head is calibrated for measuring
y mSA(m)V (m) d m illuminance E(i,z),
s*rel(m) = • srel(m) 2p p
y mSA(m) srel (m) d m U= r 2 y z=0 y i = 0 E(i,z)sini di dz (11)

where SA(m) is spectral distribution of CIE Illuminant A where r is the rotation radius of the reference plane
and V(m) is the spectral luminous efficiency function. of the photometer head. A sufficiently long photo-


IES LM-79-08

metric distance, r, is required for measurement of ment of such SSL products, goniophotometers that
luminous intensity distribution (See section 9.3.1). do not use a mirror are recommended. Some mirror
type goniophotometers have an option to mount a
The distance requirement is not critical if only total photometer head directly on the rotating arm for
luminous flux is to be measured. As indicated by such purposes.
Eq. (11), as long as the illuminance is measured
accurately, the total luminous flux can be measured 9.3.6 Photometer head The photometer head of
accurately even with a relatively short photometric the goniophotometer should have relative spectral
distance (radius r), thus less space for the goniopho- responsivity matched to the V(l) function. The f'1
tometer is required for a given size of light source to value10 of the spectral responsivity shall be less than
be measured. In this case, the detector must have 3 %. It is further desirable to apply spectral mismatch
cosine corrected angular responsivity within its field corrections for the photometer reading. For determi-
of view for the test SSL product. By definition given nation of f'1 and spectral mismatch correction factor,
in Eq. (11), the location of the light source relative to see eqs. (8) and (9) in section 9.2.6, with srel(l)
the rotation center is theoretically not relevant, and being the relative spectral responsivity of the pho-
therefore, the alignment of light source is not critical tometer head, measured in normal direction.
for measurement of total luminous flux.
For the total luminous flux measurement described
9.3.3 Scanning resolution Scanning resolution fine in section 9.3.2, the photometer head shall have
enough to accurately define the test sample shall good cosine response within the angle range where
be used. For typical wide-angle, smooth intensity light is incident, with the f2(f,z) value (relative devia-
distributions, a 22.5° lateral (horizontal) and 5° longi- tion from the cosine function)10 of less than 2 %
tudinal (vertical) grid may be acceptable. Finer angle within the acceptance angle range. The field-of-view
resolution (smaller test increments) shall be used of the photometer head should be limited (e.g., using
where the luminous intensity from the SSL product aperture screens) in order to shield stray light reflect-
is changing rapidly or is erratic, such as in beam ed from the angles other than from the light source
forming sources. Further guidance on selecting the being measured. To minimize stray light errors within
correct scanning resolution, based on experience the field-of-view of the photometer, it is recommend-
gained over the years testing other luminaire and ed to use a light trap on the other side of the detector
lamp types, are in Refs. 9, 11-17. arm and/or use low reflectance materials (such as
black velvet) for the wall and floor surfaces.
9.3.4 Angle coverage The range of the angular scan
must cover the entire solid angle to which the SSL 9.3.7 Calibration The goniophotometers for measur-
product emits light. A disadvantage of a goniopho- ing luminous intensity distribution shall be calibrated
tometer, when measuring total luminous flux, is that against the illuminance or luminous intensity stan-
a goniophotometer in general has some angular dards traceable to national standards. In addition,
region where emission from a light source under goniophotometers for measuring total luminous flux
test is blocked by its mechanism (e.g., an arm for shall be validated by measurement of total luminous
SSL product holder) so that measurement in that flux standard lamps traceable to national standards.
direction cannot be made (such angle is called dead Such validation measurements should use standard
angle). This is not a problem for SSL products that lamps having similar angular intensity distributions
emit light only in the forward direction similar to many (directional / omni-directional) as the types of SSL
existing fixtures. However, this can be a problem for products to be tested with the goniophotometer.
SSL products that emit light in all directions (e.g.,
integrated LED lamps similar to compact fluorescent
lamps). Goniophotometers with a large dead angle 10.0 Luminous Intensity Distribution
are not suited for total flux measurement of such
SSL products. If the dead angle is small (e.g., ±10°
or less), it is possible to interpolate the missing data The recommendations given in section 9.3 pertain
points with an additional uncertainty. to goniophotometers used to measure luminous
intensity distribution as well as total luminous flux.
9.3.5 Polarization It should be noted that mirror For measurement of luminous intensity distribution,
type goniophotometers have a detection system a sufficient photometric distance should be used –
that is polarization sensitive due to the slightly generally, more than five times of the largest dimen-
polarizing characteristics of mirrors themselves. sion of the test SSL product having broad angular
Sensitivity to polarized light can cause significant distributions. A longer distance may be needed for
errors when measuring the total luminous flux of narrow beam sources (e.g., see Ref. 13).
SSL products that emit polarized light. For measure-

10

Potrebbero piacerti anche