Sei sulla pagina 1di 7

ANKE - The Reliable Medical Equipment Manufacturer

OPENMARK 4000 Magnetic Resonance Imaging System

Specifications
OPENMARK 4000 Magnetic Resonance Imaging System Specifications

No. Technical and performance names Specific parameters Note

1 Magnet system
Permanent magnet (Automatic
1.01 Magnet type constant temperature
system)

1.02 Magnetic field strength 0.38 ± 5%T

1.03 Magnet appearance C-shape

1.04 Shimming method Active / passive

1.05 Uniformity of magnetic field ≤ 2.0 ppm @ 40cm , DSV , VRMS


Horizontal front opening angle of
1.06 320°
magnet
1.07 5 Gaussian line range X , Y , Z directions ≤ 2.5m

2 Patient table and interaction environment

2.01 Patient table electric / manual dual mode

2.02 Max. weight load 200kg

2.03 Cross cursor positioning Laser cross cursor positioning

2.04 positioning accuracy 1mm

2.05 Patient table control Touch screen control

2.06 Emergency call button Equipped

2.07 Two-way intercom system Equipped

3 Gradient system

3.01 Gradient field strength (single axis) 28mT/m

3.02 Gradient switching rate (single axis) 93mT/m/ms

3.03 Climb time 0.3ms

3.04 Gradient cooling Air cooling

4 Radio frequency system

4.01 System type Full digital transmission and reception

4.02 RF power 6kW

2 Ver: OPENMARK4000 V18.06


4.03 Number of transmission channels 4 channels

4.04 Receiving channel number 4 channels

4.05 Receiving bandwidth 1MHz

4.06 Phase array head coil Equipped

4.07 Phase array neck coil Equipped

4.08 Phase array 17” body coil Equipped

4.09 Phase array 20” body coils Equipped

4.10 Phase array knee coil Equipped

4.11 Phase array shoulder coil Equipped

4.12 Phase array ankle coil Optional

4.13 Phase array 14” body coil Optional

4.14 Phase array wrist coil Optional

4.15 Phase array breast coil Optional

5 Console
Professional workstation
5.01 Host computer
for MRI
5.02 Operation interface APEX operating software

5.03 operating system Windows 7

5.04 CPU frequency 3.6 GHz

5.05 RAM 8GB

5.06 Memory 2.0GB

5.07 monitor 24 -inch LCD monitor

5.08 Keyboard and mouse Standard


Image reconstruction speed
5.09 3300 frames / sec
( 256 × 256 matrix) (FFT)
5.10 Hard drive capacity 1TB × 2
The number of images stored on the
5.11 ≥ 4 million
hard disk ( 256 × 256 matrix)
5.12 Storage media DVD burning

5.13 DICOM 3.0 interface provided

5.14 Ethernet connection provided

3 Ver: OPENMARK4000 V18.06


5.15 Console provided

5.16 Operation chair provided

6 Scan parameters

6.01 Maximum FOV 400mm

6.02 Minimum FOV 30mm

6.03 Shortest TE (spin echo) 5ms

6.40 Shortest TR (spin echo) 11ms

6.05 Shortest TE (gradient echo) 1ms

6.06 Shortest TR (gradient echo) 3ms

6.07 Minimum 2D layer thickness 1.0mm

6.08 Minimum 3D layer thickness 0.1mm

6.09 Maximum acquisition matrix 1024 × 1024

6.10 DWI maximum b value 1500s/mm2

7 Imaging Technology

7.01 Spin echo ( SE 2D/3D ) provided

7.02 Dual echo ( DE ) provided

7.03 Spin dual echo ( DSE ) provided

7.04 Multi-slice multi-echo ( MSME ) provided

7.05 Fast spin echo ( FSE 2D/3D ) provided


Fast recovery of fast spin
7.06 provided
echo ( FRFSE )
Single shot fast spin echo ( SSFSE
7.07 provided
2D/3D )
Multiple shot of fast spin
7.08 provided
echo ( MSFSE )
7.09 Inversion recovery sequence ( IR ) provided

7.10 Inversion recovery spin echo ( IRSE ) provided


Inverse recovery fast spin
7.11 provided
echo ( IRFSE )
Short tau time inversion recovery
7.12 provided
sequence ( STIR )
7.13 Water-fat separation imaging ( DIXON ) provided

4 Ver: OPENMARK4000 V18.06


Fast inversion recovery
7.14 provided
sequence ( FIR )
Fluid attenuated inversion
7.15 provided
recovery ( FLAIR )
T2 weight fluid attenuated inversion
7.16 provided
recovery ( T2-FLAIR )
T1 Weight fluid attenuated Inversion
7.17 provided
recovery ( T1-FLAIR )
7.18 Gradient recall echo ( GRE 2D/3D ) provided

7.19 Dual gradient echo ( DGRE ) provided

7.20 Turbo gradient echo ( TFE 2D/3D ) provided


Steady state free process gradient
7.21 provided
echo ( SSFP )
Balanced steady state free process
7.22 provided
gradient echo ( SSFP )
Magnetic resonance
7.23 provided
angiography ( TOF 2D/3D )
7.24 Linear scan diffusion imaging ( LSDW ) provided

7.25 Diffusion weighted imaging ( DWI ) provided

7.26 Planar echo imaging ( EPI ) provided


Propeller artifact suppression Available, apply to both
7.27
technology head and abdomen
8 Imaging technology

8.01 Automatic RF correction provided

8.02 Automatic frequency locking provided

8.03 Automatic increasment adjustment provided

8.04 Automatic coil identification provided

8.05 Automatic phase correction provided

8.06 Automatic shiming linear compensation provided

Scanning scheme presetting and


8.07 provided
selection techniques
GRSCOUT three-dimensional
8.08 provided
positioning scan
Three-dimensional three-slice rapid
8.09 provided
scanning technology

5 Ver: OPENMARK4000 V18.06


Three-dimensional nine-slice rapid
8.10 provided
positioning scanning technology
Multiple groups of arbitrary positioning
8.11 provided
technology
Batch imaging technology ( Scan
8.12 provided
Sequence Queued Waiting )
8.13 Interval scanning technology provided

8.14 Off-center Imaging provided

8.15 Rectangular FOV imaging technology provided

8.16 Arbitrary slice scanning technology provided

8.17 Partial acquisition technology provided

8.18 Partial echo technique provided

Optimizition bandwidth acquisition


8.19 provided
technology
Phase encode gradient optional
8.20 provided
technology
Ellipse-encoded acquisition imaging
8.21 provided
technology
8.22 Parallel acquisition technology provided

8.23 Flow compensation technology provided

Multi-slice & multi-angle


8.24 provided
pre-saturation technology
Pre-saturated automatic tracking
8.25 provided
technology
Magnetization transfer contrast
8.26 provided
technology ( MTC )
Tilted optimized non-saturating
8.27 provided
excitation technology ( TONE )
8.28 Breath-holding imaging technology provided

8.29 Respiratory gating technology provided

8.30 Motion artifact suppression provided

8.31 Metal artifact suppression technology provided

8.32 Over sampling technology provided

Phase encoding direction anti-wrap


8.33 provided
technology

6 Ver: OPENMARK4000 V18.06


Slice encoding direction anti-wrap
8.34 provided
technology
Magnetic resonance
8.35 cholangiopancreatography MRCP ( 2D/ provided
3D )
Magnetic resonance urography
8.36 provided
MRU ( 2D/3D )
Magnetic resonance myelography
8.37 provided
MRM ( 2D/3D )
8.38 Ultra-fast water imaging technology provided

8.39 Maximum intensity projection provided

8.40 Minimum intensity projection provided

8.41 Multiplanar reconstruction provided

8.42 Online image filtering technology provided

8.43 Online post-processing technology provided

7 Ver: OPENMARK4000 V18.06