INTRODUCTION
ABOUT THE DRSHD 1080P VIDEO AND IMAGE FORMATS AND QUALITY
The Med X Change DRSHD 1080p is a complete medical The Med X Change DRSHD 1080p utilizes the latest in
recording and documentation system that can capture H.264 encoding and compression, supporting 720p,
High Definition & Standard Definition video and images 1080i, 1080p at 50, 59.94, or 60hz HD resolutions and
from multiple sources, capture digital images without NTSC (480i) or PAL (576i) SD video resolutions. Videos
interruption of video capture, and store the captured are stored as AVCHD @Main Level 4.2 H.264 encoded
digital f iles with patient information. The DRSHD MP4 files.
1080p can duplicate captured videos and images to
a portable drive, optical storage, smart devices (using Images can be stored as uncompressed bitmap files
the MedXStream app), and PACS (Picture Archiving and (.BMP), compressed JPEG, or lossless compressed PNG
Communication Systems) for documentation, storage, files.
and presentation.
All digital f ile formats are O SX , iO S , A ndroid, and
The DRSHD 1080p video capture software is built upon Windows 7 compatible.
H.264 technology; utilizing the best compression and
the highest quality video capture available. Videos are CAPACITY CHART
stored as HD or SD H.264 MP4 files. Images are stored The following charts detail the maximum capacities of
as JPEG, BMP, or PNG files. To keep pace with the the DRSHD 1080p storage drive.
rapid advancements in medical and portable storage
Video Type Number of
te chnology, the DRS HD 10 8 0 p sof t ware is easily
Hours
updateable as new features and advancements become
available. HD Videos (High Quality) 141
HD Videos (Medium Quality) 206
The DRSHD 1080p is a Class I medical device intended
HD Videos (Low Quality) 413
to be used as an accessory to various surgical cameras
to record video and capture images during surgical SD NTSC Videos (High Quality) 397
procedures and operations. The DRSHD 1080p is not SD NTSC Videos (Medium Quality) 689
intended for use with radiological imaging or diagnostic
SD NTSC Videos (Low Quality) 1000
systems or in any function other than what is outlined
in this manual. SD PAL Videos (High Quality) 492
SD PAL Videos (Medium Quality) 816
The DRSHD 1080p system has no patient applied parts.
SD PAL Videos (Low Quality) 1193
U5010-12 PG004
QUICK INSTALL
CONTENTS
Upon receiving the DRSHD 1080p, the following contents
should be in the box:
PG005 U5010-12
QUICK INSTALL (CONTINUED)
COM 3
SN
HD IN
Med X Change, Inc. This device complies with Part 15 of
525 8th Street West the FCC Rules. Operation is subject
Bradenton, FL 34205 USA to the following two conditions: (1)
this device may not cause harmful
interference, and (2) this device must
AC Input accept any interference received,
COM1
S-VIDEO
N/A
C O R P O R A T I O N
CONMED CORPORATION
525 FRENCH ROAD
DVI OUT
UTICA, NY 13502-5994 USA
SPEAKER OUT
1
N/A
LINE IN
NETWORK
FIREWIRE N/A
COM 4
2
N/A
S-VIDEO OUT
N/A
COM2 VGA OUT
N/A
N/A
ON/OFF
USB USB MIC
U5010-12 PG006
QUICK INSTALL (CONTINUED)
PG007 U5010-12
QUICK INSTALL (CONTINUED)
FIGURE 1
FIGURE 2-2
U5010-12 PG008
QUICK INSTALL (CONTINUED)
FIGURE 4-2
FIGURE 3-2
PG009 U5010-12
QUICK INSTALL (CONTINUED)
FIGURE 5-2
U5010-12 PG010
WORK LIST
PG011 U5010-12
WORK LIST (CONTINUED)
CREATING A CASE
The DRSHD 1080p user may create a case on the Work FIGURE 8-1
List Page at any time. A scheduled or unscheduled case
may be created and started in any of the three Work List
configurations.
To create a case:
1. Click the New Button on the lower left side of the
page (Figure 8-1).
2. Enter patient information into the appropriate Patient
Information Fields. Pressing the Enter key on the
keyboard moves the curser down to the next Patient
Information Field (Figure 8-2).
3. After entering all desired Patient Study Information,
click the Start Case Button on the lower left side
of the page to begin the case or optionally the user
may click the Save Button to add the case to the FIGURE 8-2
work list for future use when using Local Work List
configuration (Figure 8-2).
4. The Record page will display and the DRSHD 1080p
will be ready to capture videos and images for the
case created (Figure 8-3).
The DRSHD 1080p pre-populates the Study Date field FIGURE 8-3
with the current Study Date and Time each time a new
patient entry is started. The Study Date can be edited
and changed to a future date if desired. If the Study Date
and Time does not match the System Date and Time,
verify that the DRSHD 1080p’s Date and Time is correctly
setup. This can be adjusted on the Setup Page > General
> Regional > Date/Time Text Box (Regional Sub Tab).
U5010-12 PG012
WORK LIST (CONTINUED)
STARTING A CASE FROM THE WORK LIST PAGE DELETING A CASE FROM THE WORK LIST PAGE
Selecting and starting a case from the DRSHD 1080p IN LOCAL WORK LIST CONFIGURATION
Work List Page is the same process in each of the Work When in Local Work List configuration, the user has the
List configurations (Figure 9-1). ability to delete cases from the work list. Deleting a case
from the DRSHD 1080p Work List Page is only possible
1. Browse or search the cases listed in the middle of in Local Work List configuration (Figure 10).
the Work List Page and click on the desired case to
select. The case will highlight to show that it has 1. Browse the cases listed in the middle of the Work
been selected. List Page and click on the desired case to select. The
2. The Patient Study Information section will populate case will highlight to show that it has been selected.
with the information from the selected case. 2. The Patient Study Information section will populate
3. Click the Start Case Button to begin the case. with the information from the selected case.
3. Click the Delete From Work List Button to delete the
When a case is selected and star ted, the case will selected case from the local work list.
be removed from the DRSHD 1080p Work List Page. FIGURE 10
After a case has been started and/or completed, it will
appear on the Review Page. From the DRSHD 1080p
Review Page, the user has the ability to review, copy, and
print digital files. If the user wishes to alter the Patient
Study Information or add digital files to a case that was
previously created with the DRSHD 1080p software, the
Review Page offers this function (Review Page Controls).
FIGURE 9
PG013 U5010-12
WORK LIST (CONTINUED)
Click the Patient Study Field at the top of the work list to
alphabetize the list by that field. Click the Patient Study
Field at the top of the work list a second time to reverse
alphabetize the list by that field. Click the Patient Study
Field at the top of the work list a third time to remove
the filter. The “alphabetize by field” sorting will remain
constant until the Patient Study Field is clicked three
times to remove the filter (Figure 11-2).
U5010-12 PG014
WORK LIST (CONTINUED)
Refresh Button – requests the present work list from Patient S tudy Information Tex t Boxes – create a
the work list server. collection of patient demographics that are associated
with a case to be created.
Start Case Button – begins the case that is presently
entered into the Patient Study Information or selected A rc hiving Indicator - displays the present status
in the work list. of digital files to be automatically copied to external
storage.
Save Button - saves the case that is presently entered
into the Patient Study Information to the Local Work List Auto Printing Indicator – displays the present status
(only visible in Local Work List configuration). of automatic print jobs.
New Button – clears all information entered into the On Screen Keyboard Button - displays the on screen
Patient Study Information, restores all preset default keyboard when clicked. Click the “enter” button on the
values to the Patient Study Information. to close the On Screen Keyboard.
PG015 U5010-12
RECORD
RECORD PAGE
The DRSHD 1080p Record Page appears after a case FIGURE 12-1
is started from the Work List Page. The Record Page
is the starting point to which a user will capture videos
and images for the surgical case in progress. The
Record page allows users to record videos and capture
still images from live video, delete captured images,
and optionally alter Patient Study Information for the
case in progress when in Local Work List and HL7
Configurations. The user has the ability to select one
of four video sources for capture, select a video source
for PIP, add verbal and typed annotations to digital files
captured, view live video in full screen mode, and to
signal the finish of a case to initiate automatic printing,
automatic storage to a DICOM storage server.
FIGURE 12-2
VIDEO INPUT SOURCE SELECTION
The DRSHD 1080p capture sof tware provides four
choices of video input source for capture; one HD source
(HD DVI) and three SD sources (S-Video, Comp 1, and
Comp 2). The default primary video source for capture
is the HD DVI Input (Figure 12-1). The default secondary
video source for capture is S-Video (Figure 12-2). The
default SD video source changes based upon the last
SD video input source selection that was made by the
user. To change the default SD video source, click on
any SD Video Input Source Button. The title of the SD
video source will change to the new default selection
on the Status Tab and the new default SD video source
is saved. The stream from the primary video source
displays in the window on the center left of the page. The FIGURE 12-3
stream from the secondary video source can optionally
be displayed as a picture in picture (PIP) window in the
upper left corner of the Primary Video Source Window
if the PIP feature is enabled (Figure 12-3) and the PIP
button for the secondary video source is selected. Still
Images will only be captured from the Primary Video
Source Window.
U5010-12 PG016
RECORD (CONTINUED)
PG017 U5010-12
RECORD (CONTINUED)
FIGURE 15-2
FIGURE 16-2
U5010-12 PG018
RECORD (CONTINUED)
Information Tab - displays the Patient Study Information Capture Button - capture a still image from the primary
for the case in progress. When the DRSHD 1080p capture video source.
software is in Local Work List or HL7 configuration, the
Edit Button will be available. Click the Edit Button to alter Record Button - capture video from the selected primary
the Patient Study Information for the case in progress. video source or the primary and secondary video sources
when Dual Recording is active (S tandard and Dual
Levels Tab - features sliders that adjust the Brightness, Recording).
Contrast, Hue, and Saturation of the SD video stream.
The adjustments will show in the video files that are Pause - stop capture of video and complete a video
captured. The Levels Tab is only visible when a SD video segment file.
source is selected as the primary input. There are no
level adjustments for the HD video stream. Video Input Source and PIP Buttons - select the primary
and secondary video source to be captured.
PG019 U5010-12
RECORD (CONTINUED)
U5010-12 PG020
RECORD (CONTINUED)
FIGURE 18-1
FIGURE 18-2
PG021 U5010-12
REVIEW
U5010-12 PG022
REVIEW (CONTINUED)
Filter Box – enter text and numbers to filter the work DICOM Button - sends a selected case the preset DICOM
list by a keyword. Storage server when clicked. This button is only visible
when DICOM storage is enabled.
Activate Filter Button – (magnifying glass shaped icon)
sorts the case list by the filter text entered. Case View Controls:
Back Button – navigates to List View. This button is not
Deactivate Filter Button – (x shaped icon) removes the available when a surgical case is in progress.
filter from the case list sorting.
Images/Videos Buttons - show only the Thumbnails
Scroll Bars – move the orientation of the Patient Study for the images or videos for the selected surgical case.
Information Fields to enable the user to view all of the
cases and data in the case list. Thumbnails - small images that represent videos and
still captures for the selected surgical case. Click in the
Scroll Bar Arrow Buttons – move the Scroll Bars in center of the Thumbnail to select the digital file for print,
small increments. copy, burn, or delete.
Top Checkbox - selects and deselects all cases. Annotate Button - allows a user to add an annotation
to the digital file.
Checkbox - selects and deselects the individual case
or multiple cases to copy, burn, or delete. Playback View Button - navigates to Playback View
and displays the digital file. Videos will begin playback
View Button - changes the page to Case View and for immediately.
the corresponding surgical case.
Delete Item Button - (trash pail icon) allows a user to
Edit Button - opens pop-up window that allows the user delete the displayed digital file. A warning dialog pop-up
to edit Patient Study Information of the corresponding window will appear to confirm that the user wishes to
surgical case. delete.
Append Button - gives user the option to reopen the Select All Button - selects all case digital files for copy,
corresponding surgical case to add additional digital burn, or delete.
files. When clicked, a warning dialog pop-up window will
appear to confirm that the user wishes to navigate to Select Annotated Button - selects all annotated case
the Record Page with the corresponding surgical case digital files for copy, burn, or delete.
loaded to append digital files to the case.
Select None Button - deselects all selected digital files.
Copy Button - navigates to the Copy Page to duplicate
case digital files to a portable drive or network folders. Copy Button - navigates to the Copy Page to duplicate
case digital files to a portable drive or network folders.
Burn Button - navigates to the Burn Page to duplicate Burn Button - navigates to the Burn Page to duplicate
case digital files to optical disc. case digital files to an optical disc.
PG023 U5010-12
REVIEW (CONTINUED)
PRINT PAGE
Delete Button - allows a user to delete the selected The DRSHD 1080p Print Page is accessed from the
digital files. A warning dialog pop-up window will appear Review Page and offers full featured printing options.
to confirm that the user wishes to delete. The user can send the print job to a list of printers or
create a PDF file to print later, select a page layout,
DICOM Button - sends a selected digital file to the preset crop HD images to a 4:3 ratio to save black ink with
DICOM Storage server when clicked. endoscopic images, add a personalized heading and logo
to the top of the printed page, make adjustments to the
Playback View Controls: images prior to printing. The user can also include dates,
Back Button – navigates to Case View. annotations, and a number of Patient Study Fields to the
printed page. The printing options can be preset to the
Media Item Checkbox - selects the displayed digital desired default settings on the Setup Page > Printing
file for copy, burn, or delete when the user navigates Tab (Figure 20), set per Physician on the Setup Page >
back to Case View. Physicians Tab (Auto Printing) or adjusted for each print
job individually on the Print Page. Any adjustments made
Annotate Button - allows a user to add an annotation to the print layout on the Print Page will not be saved
to the digital file. after the page is closed, it is recommended that the
user set the preferred default print settings on the Setup
Delete Button - allows a user to delete the displayed Page > Printing Tab (Printing Tab) during the DRSHD
digital file. A warning dialog pop-up window will appear 1080p initial setup.
to confirm that the user wishes to delete. FIGURE 20
Seek Bar - click and drag the dot on the Seek Bar to
navigate to a desired place in the video displayed.
U5010-12 PG024
REVIEW (CONTINUED)
Page Layout Dropdown - used to select 1, 2, 4, 6, 8, or Copies - adjusts the number of copies for the present
15 images per page. print job.
4:3 Crop Checkbox - activate this checkbox to crop a Print Button - sends the print job to the specified printer
16:9 image to a 4:3 image. This is useful with scope with all of the preset printing options.
images.
Save as PDF Button - saves the print job to a PDF file
Include Annotations Checkbox - when activated the all of the preset printing options.
print job will include annotations under the images.
Exit Button - closes the Print Page without sending the
Header Text Box - add text here to add a header or a present print job.
title for all of the pages in the present print job.
PG025 U5010-12
REVIEW (CONTINUED)
U5010-12 PG026
REVIEW (CONTINUED)
PG027 U5010-12
REVIEW (CONTINUED)
U5010-12 PG028
SETUP
PG029 U5010-12
SETUP (CONTINUED)
U5010-12 PG030
SETUP (CONTINUED)
PG031 U5010-12
SETUP (CONTINUED)
LOCATION PAGE
(SD) Quality Dropdown - Used to adjust the amount of The DRSHD 1080p Location Page is accessed from the
compression (bit rate) used when creating SD video files. Media Tab when the Archiving Location Button is clicked
Changing the Quality setting adjusts both video quality and is used to select a location to archive surgical cases
and file size. The user can select High at 10Mbit/s, to external devices and network folders (Figure 31-1).
Medium at 8Mbit /s, or Low at 5Mbit /s. The quality
setting for SD video recordings adjusts the quality for Location Page Controls:
all three of the SD inputs. Removable Drives Dropdown - an arrow will appear
to the left of the dropdown when a removable drive
Archiving Location Button - displays the Location is connected and available. When an arrow is present
window that allows the user to select a Removable Drive, click the dropdown to display the available devices and
a smart device with the MedXStream app, or a Remote locations.
network folder.
FIGURE 30 MedXStream (iOS) Dropdown - an arrow will appear
to the lef t of the dropdown when an iOS device is
connected and available. When an arrow is present click
the dropdown to display the available devices.
PG033 U5010-12
SETUP (CONTINUED)
Enable PIP Checkbox - activates the Video Input Source FIGURE 31-3
PIP Buttons on the Record Page.
U5010-12 PG034
SETUP (CONTINUED)
PRINTING TAB
The Printing Tab contains controls to select the default Include Image Date Checkbox - when activated a print
printing settings that are displayed as the Print Button job will include the image date of capture under the
is clicked on the Review Page (Print Page) and used with images.
the Auto Print feature. The Printing Tab Settings can be
selected Per Physician on the Setup Page > Physicians Number of Copies - adjusts the default number of copies
Tab (Figure 32). for a print job.
Page Layout Dropdown - used to select 1, 2, 4, 6, 8, or Enhancements Checkboxes - when activated the
15 images per page. Enhancement Sliders will become available and a print
job will include preset image color, brightness, contrast,
4:3 Crop Checkbox - activate this checkbox to crop a and sharpen adjustments.
16:9 image to a 4:3 image. This is useful with scope
images. Enhancement Sliders - used to adjust the selected
Enhancement. The result of the adjustment will be seen
Auto Print Checkbox - activates the Auto Print feature. in the Print Preview Window as it is made.
Auto Print is a expedited printing feature that sends a
print job to the printer without user intervention as a Print Test - sends a sample print job to the specified
surgical case is performed. When Auto Print is enabled, printer with all of the preset printing options. The sample
still image captures are printed using the default printing print job as it will be sent to the printer is displayed in
settings immediately as a print job becomes available. the Print Preview Window.
When automatic printing is in progress, a printer icon FIGURE 32
will appear on the title bar of the recording software.
Local Work List Sub Tab (default) – all cases are Test Connection Button - when clicked the DRSHD
created and maintained on the DRSHD 1080p locally 1080p capture software will attempt to communicate
when this Sub Tab is selected and the Save Button is with the work list server using the selected settings.
clicked. There is no required Patient Study Information The software will reply with a pop up window that tells
to create and begin or to schedule a future case in this the user that the communication succeeded or failed. If
Work List configuration. the communication fails check the Server settings then
check the network connection and settings. Contact the
Modality Work List Sub Tab (MWL) – the Work List facility DICOM and Network Administrators for more
Server delivers cases to the DRSHD 1080p. The DRSHD help.
1080p works as a client to the work list server. Scheduled
cases can also be created and completed locally. Patient HL7 Sub Tab - The DRSHD 1080p is capable of receiving
Study Information is not required to create and start a HL7 Scheduling Information Unsolicited messages
case or to schedule a future case. (SIU^S12, SIU^S13, SIU^S14, SIU^S15) for impor ting
patient demographic information from an Electronic
MWL Sub Tab Controls: Medical Record system or Electronic Health Record
Filter By AE Title - allows only cases assigned to the system (EMR/EHR). The Patient Information Fields are
specified Application Entity Title to be imported from imported onto the Work List Page. If DICOM Storage is
the work list server. This filter defaults to the Calling AE active, the Patient Information Fields are translated into
Title when it is entered and set. The filer can be cleared DICOM tags automatically and then stored with the case
manually if the user wishes to import the complete work digital files on the DICOM Storage Server. Unscheduled
list without this filter. cases can also be created and completed locally. There
is no required Patient Study Information to create and
begin a case.
Receiver Port - enter the port for incoming messages. Host IP Address Text Box - enter the IP address of the
work list server.
Receiver Timeout - enter the number of seconds before
timeout. Port - enter the port number used to communicate with
FIGURE 33 the work list server.
PG037 U5010-12
SETUP (CONTINUED)
Before implementing DICOM Storage on the DRSHD Transfer Syntax DICOM UID
10 8 0 p , t h e M e d X C h a n g e D I C O M C o nfo r m a n c e
S tatement should be read to determine basic Implicit VR Little Endian 1.2.840.10008.1.2
interoperability with the storage server. The DRSHD Explicit VR Little Endian 1.2.840.10008.1.2.1
1080p DICOM Conformance Statement has been created JPEG Baseline (Process 1) 1.2.840.10008.1.2.4.50
to provide a DICOM Administrator with the information
JPEG Lossless, 1.2.840.10008.1.2.4.70
required to take the first step in determining whether
Nonhierarchical, First-
and to what extent DICOM implementations are able to
Order Prediction
inter-operate. In addition to review of the DRSHD 1080p
(Processes 14 [Selection
DICOM Conformance Statement, it is suggested that
Value 1])
a test case be performed and a complete exchange
of communications be completed to ensure that all
of the settings, digital file types, and SOP classes are Single Frame SOP Class - The DRSHD 1080p offers two
compatible between the DRSHD 1080p and the DICOM Service-Object Pair (SOP) Classes or abstract syntaxes
Storage server. to transmit still images to DICOM Storage. The table
below describes the available Single Frame SOP Classes
The DRSHD 1080p capture software requires the fields and their Unique Identifiers.
displayed in the table below to be populated when
creating a surgical case to be sent to the DICOM Storage Abstract Syntax DICOM UID
server.
Secondary Capture 1.2.840.10008.5.1.4.1.1.7
Image Storage
Patient Study Information Field DICOM Tag VL Endoscopic Image 1.2.840.10008.5.1.4.1.1.77.1.1
Patient Last Name 0010,0010 Storage
Patient ID 0010,0020
Multi-Frame SOP Class - The DRSHD 1080p offers two
Patient Date of Birth 0010,0030 Service-Object Pair (SOP) Classes or abstract syntaxes
Patient Sex 0010,0040 to transmit videos to DICOM Storage. The table below
describes the available Multi- Frame SOP Classes and
Study Date 0008,0020
their Unique Identifiers.
Modality 0008,0060
Multi-Frame 1.2.840.10008.5.1.4.1.1.7.4
Secondary Capture
Image Storage
Video Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1.1
Image Storage
U5010-12 PG038
SETUP (CONTINUED)
Mutli-Frame Compression Configuration - The DRSHD Arrow Up and Down Buttons - used to reorder the Patient
1080p proposes the MPEG2 1.2.840.1000 8.1.2.4.100 Information Fields on the Work List Page. Select a Field
compression configuration or transfer syntax to transmit and then click the Arrow Buttons to move the field up
video to DICOM Storage. or down the list. The list order from top to bottom on
the Patient Fields Tab will order from left to right side
Commit - Users may enable the storage commit feature in of the page on the Work List Page and Review Page in
the event the PACS requires storage commitment. There Case View.
is no notification to the user regarding commitment.
Commit is disabled by default for the most compatibility Display Name - click on the names of the Patient
with DICOM Storage Servers. Information Fields to select them and move them with
the Arrow Buttons or click the Manage button on the
Retries - Set the number of times the commit right side of the Display name to alter it.
service attempts to commit.
Re quire d a n d E na ble d C he c k b oxes - display the
Time Interval - Set the number of minutes between Required and Enabled status of the corresponding
commit retries. Patient Information Field. These settings can be altered
by clicking the Manage Button.
AE Title - Specify the Application Entity Title of the
Commit server. Scroll Bars – move the orientation of the Patient Study
Information Fields to enable the user to view all of the
SCP Port - Specify the port number which is used Display Names listed.
to commit.
Scroll Bar Arrow Buttons – move the Scroll Bars in
Separate Association Checkbox - Commit files and small increments.
wait for result on a separate association. FIGURE 35
PG039 U5010-12
SETUP (CONTINUED)
PHYSICIANS TAB
The Physicians Tab contains the user profiles that are Manage Button - displays the Sub Tabs for each of the
saved for the Performing Physicians that are assigned Physician Profile Settings.
to a surgical case on the Work List Page (Figure 36). FIGURE 36
Physician profiles can be created prior to a surgical case
or as a case is created. Preset Physician profiles will be
selectable from the Performing Physician Dropdown on
the Work List Page in the Patient Study Information. The
user profile for the Physician can be set to the default
settings or set uniquely for the user’s preferences. The
Physician user profile will be loaded automatically as
the Physician is selected as the Performing Physician
in a case on the Work List Page. A single Physician may
also create multiple profiles to be used with each type
of surgical case.
U5010-12 PG040
SETUP (CONTINUED)
PG041 U5010-12
SETUP (CONTINUED)
SECURITY TAB
The Security Tab is used to administrate the DRSHD
1080p capture software Security feature (Figure 38).
When the optional security feature is activated users
must provide a preset Username and Password to gain
access to software functions. This feature helps prevent
unauthorized users from accessing the DRSHD 1080p
recording software. The Usernames and Passwords
stored on the Security Tab also allow users to log into the
DRSHD 1080p capture software using the MedXStream
smart device software.
U5010-12 PG042
TECHNICAL
Inputs:
DVI-I x 1: 720p, 1080i, 1080p, 50/59.94/60hz
S-Video x 1, PAL/NTSC
Composite x 2, PAL/NTSC
Outputs:
Video output: DVI-D x 1
S-Video x 1: (Outputs last image(s))
Data I/O:
USB 2.0: Front: Type 1 x 1, Rear: Type 1 x 4
Ethernet: Gigabit RJ45 x 1
Keyboard/Mouse I/O: USB
Operating System: Windows 7 Embedded
GPIO: RJ-11
Power Input:
100 - 240 VAC, 50/60 Hz
Environmental (Operating):
Temperature: 0 to 40º C
Relative Humidity: 0 to 85%
Atmospheric: 700 to 1060hPa
Environmental (Storage):
Temperature: -40 to +70º C
Relative Humidity: 10 to 100%
Atmospheric: 500 to 1060hPa
PG043 U5010-12
TECHNICAL (CONTINUED)
SYMBOL GLOSSARY
Symbol Reference(s) Title Symbol Reference(s) Title
IEC 60417:5009 Stand by IEC 60417:5031 Direct current
U5010-12 PG044
SAFTEY/ELECTRICAL/MAINTENANCE
PG045 U5010-12
SAFTEY/ELECTRICAL/MAINTENANCE (CONTINUED)
NOTES
Warning: There are no known potential electromagnetic Med X Change, Inc will not be held responsible for any
or other interference between the DRSHD 1080p system personal/internet use. The DRSHD 1080p is to be used
and other devices. If the system does cause harmful for Digital Recording only. The warranty will be null and
interference, the user is encouraged to correct the void, in response to any repairs that may be needed, if
interference by one of the following methods: personal/internet use is apparent.
U5010-12 PG046
SYSTEM COMPLIANCE STATEMENTS
PG047 U5010-12
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)
U5010-12 PG048
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and T V broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above,
the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the equipment
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
PG049 U5010-12
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)
ADDITIONAL
Special skills, training, or knowledge: Operator should In the United States, some states have
be familiar with computer based equipment. laws regarding the disposal of electronic
w a s t e ( s e e h t t p : // w w w . e l e c t r o n i c s r e c y c l i n g .
Start up procedure: org/public/ContentPage.aspx?pageid=14).
Power up the DRSHD 1080p. Upon booting up, the
DRSHD 1080p will be ready to operate. Personal patient infor mation protected by HIPA A
i s s to r e d o n t h e D R S H D 10 8 0 p h a r d d r i v e . I t i s
The DRSHD 1080p utilizes the Microsoft® Windows 7 highly recommended to clean the hard drive
Embedded operating system software. In the event of b e f o r e d i s p o s i n g ( h t t p : // o n g u a r d o n l i n e . g o v /
an operating system error, Windows system error codes articles/0010-disposing-oldcomputers).
can be referenced at: http://msdn.microsoft.com/en-us/
library/windows/desktop/ms681381(v=vs.85).aspx Warning: No modification of this equipment is allowed.
Failure to comply could result in electric shock or death.
Shutdown Procedure:
Press the power button on the front of the DRSHD 1080p Keep combustible materials away from DRSHD 1080p
in and release. while in use.
U5010-12 PG050