USER MANUAL

Model: DRSHD® 1080p


U5010-12, 6.0 2017-03-28
USER MANUAL
Model: DRSHD® 1080p
U5010-12, 6.0 2017-03-28
TABLE OF CONTENTS

INTRODUCTION4 Media Tab32


About the DRSHD 1080p4 Location Page33
Video and Still Image Formats and Quality4 Printing Tab35
Capacity Chart4 Work List Tab36
QUICK INSTALL 5 DICOM Storage Tab37
Contents5 DICOM Storage38
DRSHD 1080p Software Overview7 Advanced DICOM Storage Settings38
Peripheral Input Devices7 Patient Fields Tab39
Automatic Functions8 Physicians Tab40
Auto Delete Old Cases8 Physician Profile Sub Tabs41
Archiving8 Procedures Tab41
Auto Archive to USB9 Security Tab42
Auto Segmenting of Video Files9 TECHNICAL43
Auto Printing10 DRSHD 1080p Product Specifications43
Auto DICOM Store10 Symbol Glossary44
WORK LIST 11 SAFETY / ELECTRICAL / MAINTENANCE 45
Work List Page11 Technical45
Patient Study Information11 DRSHD 1080p Computer45
Creating a Case12 Connecting Computer to Power45
Starting a Case from the Work List Page13 Classifications45
Deleting a Case from the Work List Page in Local Notes46
Work List configuration13 Warranty46
Filter Work List14 SYSTEM COMPLIANCE STATEMENTS 47
Work List Page Controls15 System Compliance Statements & Tables47
RECORD16 IEC 60601-1-2:2007 Table 1 Requirements47
Record Page16 IEC 60601-1-2:2007 Table 2 Requirements48
Video Input Source Selection16 IEC 60601-1-2:2007 Table 6 Requirements49
No Video Input17 Additional50
Media Options17 CONTACT INFORMATION 51
Standard and Dual Recording18 Technical Assistance or Injuries and/or
Speech to Text Annotations18 Malfunction, Complaints51
Record Page Controls19 Product Return for Service and/or Investigation51
Capturing Video20 Open Service Request Status51
Capturing Still Images20 Authorized Representative Europe51
REVIEW22
Review Page22
Filter by Patient Study Field Criteria22
Review Page Controls23
Print Page24
Print Page Controls25
Copy Page26
Copy Page Controls26
Location Page27
Location Page Controls27
Burn Page28
Burn Page Controls28
SETUP29
Setup Page29
General Tab29
System Sub Tab29
Camera Sub Tab30
Network Sub Tab31
Regional Sub Tab31
Manual Sub Tab32

PG003 U5010-12
INTRODUCTION

ABOUT THE DRSHD 1080P VIDEO AND IMAGE FORMATS AND QUALITY
The Med X Change DRSHD 1080p is a complete medical The Med X Change DRSHD 1080p utilizes the latest in
recording and documentation system that can capture H.264 encoding and compression, supporting 720p,
High Definition & Standard Definition video and images 1080i, 1080p at 50, 59.94, or 60hz HD resolutions and
from multiple sources, capture digital images without NTSC (480i) or PAL (576i) SD video resolutions. Videos
interruption of video capture, and store the captured are stored as AVCHD @Main Level 4.2 H.264 encoded
digital f iles with patient information. The DRSHD MP4 files.
1080p can duplicate captured videos and images to
a portable drive, optical storage, smart devices (using Images can be stored as uncompressed bitmap files
the MedXStream app), and PACS (Picture Archiving and (.BMP), compressed JPEG, or lossless compressed PNG
Communication Systems) for documentation, storage, files.
and presentation.
All digital f ile formats are O SX , iO S , A ndroid, and
The DRSHD 1080p video capture software is built upon Windows 7 compatible.
H.264 technology; utilizing the best compression and
the highest quality video capture available. Videos are CAPACITY CHART
stored as HD or SD H.264 MP4 files. Images are stored The following charts detail the maximum capacities of
as JPEG, BMP, or PNG files. To keep pace with the the DRSHD 1080p storage drive.
rapid advancements in medical and portable storage
Video Type Number of
te chnology, the DRS HD 10 8 0 p sof t ware is easily
Hours
updateable as new features and advancements become
available. HD Videos (High Quality) 141
HD Videos (Medium Quality) 206
The DRSHD 1080p is a Class I medical device intended
HD Videos (Low Quality) 413
to be used as an accessory to various surgical cameras
to record video and capture images during surgical SD NTSC Videos (High Quality) 397
procedures and operations. The DRSHD 1080p is not SD NTSC Videos (Medium Quality) 689
intended for use with radiological imaging or diagnostic
SD NTSC Videos (Low Quality) 1000
systems or in any function other than what is outlined
in this manual. SD PAL Videos (High Quality) 492
SD PAL Videos (Medium Quality) 816
The DRSHD 1080p system has no patient applied parts.
SD PAL Videos (Low Quality) 1193

BACK UP VIDEOS AND IMAGES TO EXTERNAL Image Type Number of


STORAGE PERIODICALLY. Files
THE DRSHD 1080p IS NOT A LONG TERM
HD BMP 676979
STORAGE OR ARCHIVING DEVICE.
HD PNG 744677
HD JPEG 6808477
SD NTSC BMP 2118193
SD NTSC PNG 2723391
SD NTSC JPEG 22427926
SD PAL BMP 2118193
SD PAL PNG 3466134
SD PAL JPEG 38127475

U5010-12 PG004
QUICK INSTALL

CONTENTS
Upon receiving the DRSHD 1080p, the following contents
should be in the box:

• 1 - DRSHD 1080p Video Recording System • 2 - BNC to RCA adapters


• 1 - User Manual Download Card • 1 - Medical Grade Power cable
• 1 - Keyboard with built in touchpad • 1 - GPIO cable
• 1 - DVI to DVI cable • 2 - Remote trigger cables
• 1 - S-Video cable • 5 - Blank DVD-R discs
• 1 - S-Video cable Installation Guide

DRSHD 1080P FRONT

Front Microphone In (enabled by default)


Front Speaker Out
USB Port Disc Drive Activity Light Eject Disc

DVD/Blu-ray Drive Hard Drive Activity Light Power

PG005 U5010-12
QUICK INSTALL (CONTINUED)

DRSHD 1080P BACK

System Integration Por ts Camera Remote Control AC Source


(CMIS) Main Video Output HD DVI Input 100-240 VAC, 50/60 HZ

REF DRSHD 1080p Rev7

COM 3
SN

HD IN
Med X Change, Inc. This device complies with Part 15 of
525 8th Street West the FCC Rules. Operation is subject
Bradenton, FL 34205 USA to the following two conditions: (1)
this device may not cause harmful
interference, and (2) this device must
AC Input accept any interference received,
COM1

100-240V~ Attention: Consult including interference that may


60-50Hz accompanying cause undesired operation.
5-3A documents Assembled in USA

DISTRIBUTED BY: GPIO


CAPTURE RECORD

S-VIDEO
N/A
C O R P O R A T I O N
CONMED CORPORATION
525 FRENCH ROAD

DVI OUT
UTICA, NY 13502-5994 USA
SPEAKER OUT

1
N/A
LINE IN
NETWORK
FIREWIRE N/A

COM 4

2
N/A
S-VIDEO OUT
N/A
COM2 VGA OUT

N/A
N/A
ON/OFF
USB USB MIC

Recording Indicator Main Audio Output Foot Pedal Port


S-Video (Secondary Video Output) SD Video Inputs
Potential Equalization Terminal S-Video Master Power Switch
Composite 1
Composite 2

Accessories should only be connected as shown.


• To use an ELO 1528L (DRS02) Touch Screen Monitor
• A HD or SD video source that is to be used for or ELO 15.6” LCD Accutouch Monitor, Widescreen
capture, may be connected to the four video capture (DRS34); connect the cables accordingly: connect
inputs; DVI, S-Video, Composite 1, and Composite the DVI cable to the Main Video Output (DVI) on the
2. The DRSHD 1080p is capable of simultaneous DRSHD® 1080p, the USB cable to any USB port on
video capture from the HD DVI Input and one of the the back of the DRSHD 1080p, and the audio cable
three SD Inputs. to the Main Audio Output on DRSHD 1080p back
panel. Connect the monitor’s AC Power Brick to the
• Unplugging the power cable will completely isolate power port and power on the ELO Monitor.
the DRSHD 1080p from the electrical energy source.
• If a microphone is to be used, it may be connected
• An appropriate UPS battery back-up/surge protector to the microphone input jack on the front or back
is recommended if the system is placed in an of the system.
environment where electrical dropout, spike, or
brownout can occur. • The DRSHD 1080p is loaded with Sony UP-DR80MD,
Sony UP-DR55, Sony UPD-25 and HP Officejet 8000
• The DRSHD 10 8 0p has not been approved for printer drivers for plug and play ability. Simply
domain authentication. Archiving and copying to connecting any of these printers to DRSHD 1080p
shared folders within a domain is possible with the will allow an automatic install of the printer. To install
proper permissions to the shared folder. any other printer, Med X Change recommends using
the manufacturer’s “basic driver” if available at their
support site.

U5010-12 PG006
QUICK INSTALL (CONTINUED)

DRSHD 1080P SOFTWARE OVERVIEW PERIPHERAL INPUT DEVICES


The DRSHD 1080p Software is designed to capture digital The DRSHD 1080p is equipped with several methods of
files with ease. It can be operated using a mouse and controlling the software with peripheral input devices.
keyboard, touch screen monitor, surgical camera remote Camera Remote Controls, mouse click, touch screen,
controls, foot pedal, or integrated system controls. There foot pedal, and serial command with integrated systems
is also an option to use the MedXStream software on can control the trigger of stills and videos. The DRSHD
a smart device in conjunction with the aforementioned 1080p capture software provides an On Screen Keyboard
devices. feature that can be used with a touch screen monitor
or disabled if a standard keyboard is preferred to enter
Work List Page – used to organize and display patient Patient Study Information and Annotations.
information pertaining to a case that is created either on
the DRSHD 1080p or sent from a work list server. The Camera Remote Control - camera control unit (CCU) that
user has the option to create and save cases to the work have two 3.5mm stereo (2 conductor ring) connections
list in Local Work List configuration, create scheduled for remote control capture can be connected to the
cases in MWL and HL7 Configurations, or to select and DRSHD 1080p via the Camera Remote Control Capture
begin a case that was sent from a work list server. The and Camera Remote Control Record connections using
Local Work List may also be populated from a smart two 3.5mm stereo (2 conductor ring) Camera Control
device using the MedXStream software. cables. This connection facilitates the DRSHD 1080p
capture software’s ability to respond to camera button
Record Page – used to capture digital HD and SD video pushes with video and still image capture on command.
and images during a surgical case. The user has the
ability to select one of four video sources for capture, Touch Screen Monitor Control - the DRSHD 1080p is
optionally select a video source for PIP, and add verbal easy to pair with a touch screen monitor connected via
and typed annotations to digital files captured. The USB. The DRSHD 1080p capture software features large
Record Page appears after a case is started from the graphic controls that are easy to touch, an On Screen
Work List Page. Keyboard, and integrated touch screen calibration. The
DRSHD 1080p capture software can complete a surgical
Review Page – used to view, copy, delete, burn to disc, case using a touch screen monitor without requiring any
and print digital files. The user can also append (add additional peripheral devices.
digital files) to a case, alter patient information from
a previous case, and send previous cases to DICOM Foot Pedal - the DRSHD 1080p capture software will
Storage (when active). The cases on the Review Page respond to a foot pedal command to star t and stop
may also be viewed and managed from a smart device recordings and to capture still images. The foot pedal
using the MedXStream software. connection is the 9 pin serial port on the back of the
DRSHD 1080p labeled COM 4.
Setup Page – used to configure software preferences,
set working configurations, select digital file types, Keyboard and Mouse - a USB keyboard and mouse can
adjust camera settings, select printing options, configure be used to enter Patient Study Information and trigger
server communications, adjust per user settings, and digital file capture alone or in combination with other
set software security options. control methods.

PG007 U5010-12
QUICK INSTALL (CONTINUED)

AUTOMATIC FUNCTIONS ARCHIVING


The DRSHD 1080p capture software can be configured The DRSHD 1080p Archive feature creates and sends
to perform automatic functions that allow the user to a duplicate copy of a surgical case to a sele cted
minimize repetitive workflow tasks and complete surgical secondary storage location as digital files are created.
cases with more efficient usage of time. The secondary storage location can be a network folder,
a smart device connected via USB, or a USB hard drive.
AUTO DELETE OLD CASES The Archiving feature overrides the Auto Archive To USB
The DRSHD 1080p capture software can be configured to feature, when the checkbox labeled Archiving is active;
automatically delete previously captured cases that have the Auto Archive To USB feature is turned off. Archiving
reached a predetermined age to save storage space and can be enabled on the Setup Page > Media Tab (Figure
to automate machine storage maintenance. Auto Delete 2-1) or set per user on the Setup Page > Physicians Tab
Cases can be enabled on the Setup Page > General Tab > Media Sub Tab (Figure 2-2, Media Sub Tab).
> System Sub Tab (Figure 1, System Sub Tab). FIGURE 2-1

FIGURE 1

FIGURE 2-2

U5010-12 PG008
QUICK INSTALL (CONTINUED)

AUTO ARCHIVE TO USB AUTO SEGMENTING OF VIDEO FILES


T he DRS HD 10 8 0 p c apture sof t ware provides an The DRSHD 1080p capture software automatically divides
automatic archive feature that creates and sends a captured video files into segments of a predetermined
duplicate copy of a surgical case to a connected USB maximum time length. This feature controls the size
hard drive or smart device connected via USB without of the captured videos and prevents incompatibility
user inter vention and as soon as the USB device is problems when copying the video segment files to a
available. This feature is activated by default. The Auto CD/DVD or a USB flash drive. The default maximum
Archive to USB feature will send a copy of each surgical segment time is 5 minutes. The maximum Segment
case to the largest connected external hard drive or Length can be adjusted or disabled on the Setup Page
smart device depending on the Auto Archive Priority > Media Tab (Figure 4-1) or set per user on the Setup
setting. Auto Archiving can be enabled on the Setup Page > Physicians Tab > Media Sub Tab (Figure 4-2,
Page > Media Tab (Figure 3-1) or set per user on the Media Sub Tab).
Setup Page > Physicians Tab > Media Sub Tab (Figure FIGURE 4-1
3-2, Media Sub Tab). To Archive surgical cases to a user
specified location, see Archiving (Archiving).

N o te: D igi tal f il es that are c apture d p r i o r to th e


connection of a USB device will not be copied.
FIGURE 3-1

FIGURE 4-2

FIGURE 3-2

PG009 U5010-12
QUICK INSTALL (CONTINUED)

AUTO PRINTING AUTO DICOM STORE


T he DRS HD 10 8 0 p c apture sof t ware provides an The DRSHD 1080p capture software can automatically
automatic printing feature that sends print jobs to a send surgical cases to a preset DICOM Storage Server
preset printer without user intervention and as soon as when the user clicks the Close Case Button on the
a print job is available. This feature expedites printing Record Page. Auto Store can be enabled on the Setup
tasks to make prints available as soon as possible after Page > DICOM Storage Tab (Figure 6, DICOM Storage
digital files are captured and will begin the printing Tab).
process while a case is in progress. Auto Print can be FIGURE 6
enabled on the Setup Page > Printing Tab (Figure 5-1)
or set per user on the Setup Page > Physicians Tab >
Printing Sub Tab (Figure 5-2, Printing Sub Tab).
FIGURE 5-1

FIGURE 5-2

U5010-12 PG010
WORK LIST

WORK LIST PAGE PATIENT STUDY INFORMATION


The DRSHD 1080p Work List Page is the starting point Patient demographic information is entered here to
to which a user will go to begin a surgical case. The create a case. Patient Study Fields may be resorted,
user may enter Patient Study Information to create and set as required to start a case, or disabled and hidden
start a case, schedule future cases, organize the work depending on user preference and the present Work
list using various filter buttons, or receive and display List configuration (Figure 7-1). When DICOM Storage
pending cases from a work list server. is enabled, cer tain Patient Information Fields are
required to begin and store a case. Required Fields will
The DRSHD 1080p Work List can work in one of three be highlighted in red when DICOM Storage is enabled
Work List configurations. These Configurations can (Figure 7-2).
be chosen on the Setup Page > Work List and DICOM FIGURE 7-1
Storage Tabs (DICOM Storage Tab). Depending on the
Work List and DICOM Storage, a user may chose to
enter no Patient Study Information and begin a case,
enter required Patient S tudy Information to store a
case using DICOM Storage, or to enter as much Patient
Study Information as desired including Custom Patient
Information Fields.

Local Work List (default) – all cases are created and


maintained on the DRSHD 1080p locally. Patient Study
Information is not required to create and start a case
or to schedule a future case.

DICOM Modality Work List (MWL) – the Work List Server


delivers cases to the DRSHD 1080p. The DRSHD 1080p FIGURE 7-2
works as a client to the work list server. Scheduled cases
can also be created and completed locally. Patient Study
Information is not required to create and start a case or
to schedule a future case.

HL7 – the Work List Server delivers cases to the DRSHD


1080p. Unscheduled cases can also be created and
completed locally. Patient Study Information is not
required to create and star t a case or to schedule a
future case.

DICOM Storage – DICOM Storage is used to send digital


files to DICOM Storage. It can be used in conjunction
with the above three Work List configurations. When
active, a case created MUST have specific Patient Study
Fields populated before the user is permitted to begin
a case. The required fields (highlighted red) are Last
Name, Patient ID, Sex, Birth Date, and Modality. With
information entered into these Patient Study Fields, the
Start Case Button will become available.

PG011 U5010-12
WORK LIST (CONTINUED)

CREATING A CASE
The DRSHD 1080p user may create a case on the Work FIGURE 8-1
List Page at any time. A scheduled or unscheduled case
may be created and started in any of the three Work List
configurations.

To create a case:
1. Click the New Button on the lower left side of the
page (Figure 8-1).
2. Enter patient information into the appropriate Patient
Information Fields. Pressing the Enter key on the
keyboard moves the curser down to the next Patient
Information Field (Figure 8-2).
3. After entering all desired Patient Study Information,
click the Start Case Button on the lower left side
of the page to begin the case or optionally the user
may click the Save Button to add the case to the FIGURE 8-2
work list for future use when using Local Work List
configuration (Figure 8-2).
4. The Record page will display and the DRSHD 1080p
will be ready to capture videos and images for the
case created (Figure 8-3).

When in Local Work List configuration the user may


optionally click the Save Button to add the case to the
Work List for future use. There is no limit to how many
cases can be created and saved to the Local Work
List; this allows cases to be created and scheduled in
advance.

The DRSHD 1080p pre-populates the Study Date field FIGURE 8-3
with the current Study Date and Time each time a new
patient entry is started. The Study Date can be edited
and changed to a future date if desired. If the Study Date
and Time does not match the System Date and Time,
verify that the DRSHD 1080p’s Date and Time is correctly
setup. This can be adjusted on the Setup Page > General
> Regional > Date/Time Text Box (Regional Sub Tab).

U5010-12 PG012
WORK LIST (CONTINUED)

STARTING A CASE FROM THE WORK LIST PAGE DELETING A CASE FROM THE WORK LIST PAGE
Selecting and starting a case from the DRSHD 1080p IN LOCAL WORK LIST CONFIGURATION
Work List Page is the same process in each of the Work When in Local Work List configuration, the user has the
List configurations (Figure 9-1). ability to delete cases from the work list. Deleting a case
from the DRSHD 1080p Work List Page is only possible
1. Browse or search the cases listed in the middle of in Local Work List configuration (Figure 10).
the Work List Page and click on the desired case to
select. The case will highlight to show that it has 1. Browse the cases listed in the middle of the Work
been selected. List Page and click on the desired case to select. The
2. The Patient Study Information section will populate case will highlight to show that it has been selected.
with the information from the selected case. 2. The Patient Study Information section will populate
3. Click the Start Case Button to begin the case. with the information from the selected case.
3. Click the Delete From Work List Button to delete the
When a case is selected and star ted, the case will selected case from the local work list.
be removed from the DRSHD 1080p Work List Page. FIGURE 10
After a case has been started and/or completed, it will
appear on the Review Page. From the DRSHD 1080p
Review Page, the user has the ability to review, copy, and
print digital files. If the user wishes to alter the Patient
Study Information or add digital files to a case that was
previously created with the DRSHD 1080p software, the
Review Page offers this function (Review Page Controls).
FIGURE 9

PG013 U5010-12
WORK LIST (CONTINUED)

FILTER WORK LIST


The DRSHD 1080p Work List Page provides the user the FIGURE 11-1
ability to sort the cases in the work list by any default
Patient Study Information Field, by date range, and by
cases scheduled for the present date (default). Click the
Today’s Work List Button at any time to revert back to
viewing only the cases scheduled for the present date
(default).

To filter the work list by a keyword in the Patient Study


Information, add text to the Filter Box above each of the
fields and click the magnifying glass icon on the right to
search. Sex and Performing Physician use a drop down
selection to filter the list. Study Date can be filtered by
cases scheduled for Today, This Week, or This Month.
Click on the X that appears to the right of the Filter Box
to remove the specific filter, or optionally click the Clear FIGURE 11-2
Filters Button at the top of the Work List Page to remove
all sorting filters (Figure 11-1).

Click the Patient Study Field at the top of the work list to
alphabetize the list by that field. Click the Patient Study
Field at the top of the work list a second time to reverse
alphabetize the list by that field. Click the Patient Study
Field at the top of the work list a third time to remove
the filter. The “alphabetize by field” sorting will remain
constant until the Patient Study Field is clicked three
times to remove the filter (Figure 11-2).

Note: In DICOM Modality Work List (MWL) Configuration,


the incoming work list from the work list server is filtered FIGURE 11-3
by the AE Title assigned to the DRSHD 1080p by default.
This prevents any cases that are not explicitly assigned
to the AE Title of the DRSHD 1080p from displaying on
the DRSHD 1080p Work List Page. This filter can be
removed on the DRSHD 1080p Setup Page if desired
(Figure 11-3).

U5010-12 PG014
WORK LIST (CONTINUED)

WORK LIST PAGE CONTROLS


Clear Filters Button – removes all filters that have been Scroll Bars – move the orientation of the Patient Study
applied to the work list on the DRSHD 1080p Work List Information Fields to enable the user to view all of the
Page. cases and data in the work list and in the Patient Study
Information sections of the DRSHD 1080p Work List
Today’s Work List Button – displays only cases on the Page.
Work List that are scheduled by the Study Date for the
present day. Scroll Bar Arrow Buttons – move the Scroll Bars in
small increments.
Delete From Work List Button – delete the case that is
presently selected in the work list (only visible in Local On Screen Keyboard (OSK ) – enters Patient Study
Work List configuration). Information.

Refresh Button – requests the present work list from Patient S tudy Information Tex t Boxes – create a
the work list server. collection of patient demographics that are associated
with a case to be created.
Start Case Button – begins the case that is presently
entered into the Patient Study Information or selected A rc hiving Indicator - displays the present status
in the work list. of digital files to be automatically copied to external
storage.
Save Button - saves the case that is presently entered
into the Patient Study Information to the Local Work List Auto Printing Indicator – displays the present status
(only visible in Local Work List configuration). of automatic print jobs.

New Button – clears all information entered into the On Screen Keyboard Button - displays the on screen
Patient Study Information, restores all preset default keyboard when clicked. Click the “enter” button on the
values to the Patient Study Information. to close the On Screen Keyboard.

Work List Server Indicator – informs the user of the


present connection status of the selected work list
server. Green indicates a connection. Red and flashing
indicates no connection.

Storage Server Indicator – informs the user of the


present connection status of the selected DICOM storage
server. Green indicates a connection. Red and flashing
indicates no connection.

Filter Box – enter text and numbers to filter the work


list by a keyword.

Activate Filter Button – (magnifying glass shaped icon)


sorts the work list by the filter text entered.

Deactivate Filter Button – (x shaped icon)removes the


filter from the work list sorting.

PG015 U5010-12
RECORD

RECORD PAGE
The DRSHD 1080p Record Page appears after a case FIGURE 12-1
is started from the Work List Page. The Record Page
is the starting point to which a user will capture videos
and images for the surgical case in progress. The
Record page allows users to record videos and capture
still images from live video, delete captured images,
and optionally alter Patient Study Information for the
case in progress when in Local Work List and HL7
Configurations. The user has the ability to select one
of four video sources for capture, select a video source
for PIP, add verbal and typed annotations to digital files
captured, view live video in full screen mode, and to
signal the finish of a case to initiate automatic printing,
automatic storage to a DICOM storage server.

FIGURE 12-2
VIDEO INPUT SOURCE SELECTION
The DRSHD 1080p capture sof tware provides four
choices of video input source for capture; one HD source
(HD DVI) and three SD sources (S-Video, Comp 1, and
Comp 2). The default primary video source for capture
is the HD DVI Input (Figure 12-1). The default secondary
video source for capture is S-Video (Figure 12-2). The
default SD video source changes based upon the last
SD video input source selection that was made by the
user. To change the default SD video source, click on
any SD Video Input Source Button. The title of the SD
video source will change to the new default selection
on the Status Tab and the new default SD video source
is saved. The stream from the primary video source
displays in the window on the center left of the page. The FIGURE 12-3
stream from the secondary video source can optionally
be displayed as a picture in picture (PIP) window in the
upper left corner of the Primary Video Source Window
if the PIP feature is enabled (Figure 12-3) and the PIP
button for the secondary video source is selected. Still
Images will only be captured from the Primary Video
Source Window.

U5010-12 PG016
RECORD (CONTINUED)

NO VIDEO INPUT MEDIA OPTIONS


The primary video source and PIP windows will display By default the DRSHD 1080p capture sof tware will
an image of color bars and a “No Video Input” warning limit captured video segments to 5 minutes maximum
message when the DRSHD 1080p does not detect an Segment Length and will capture images in the JPEG
incoming video stream on a selected HD or SD video Image File Type. The user may change the maximum
input (Figure 13). When this warning is displayed the Segment Length and Image File Type on the Setup Page
DRSHD 1080p capture software will not allow the user > Media Tab (Figure 14, Media Tab). The Image File Type
to begin video or still image capture; check the camera, can be BMP, JPEG, or PNG. Image File Types can be BMP,
signal cabling, and video input connection for trouble JPEG, or PNG. Video Segments can be from 15 seconds
or switch to a video input source that has an incoming to 10 minutes and 45 seconds; Video Segments can also
video stream from a connected camera. be limited to the maximum 2 hour segment length. For
FIGURE 13 synchronized starts and stops of Dual Recording, check
the Dual Recording checkbox.
FIGURE 14

PG017 U5010-12
RECORD (CONTINUED)

STANDARD AND DUAL RECORDING SPEECH TO TEXT ANNOTATIONS


The DRSHD 1080p capture software is able to capture The DRSHD 1080p Record Page has a speech to text
video from one or two video sources simultaneously, the annotation feature that will take a verbal annotation from
primary video source and the secondary video source. a connected microphone, convert it to editable text, and
The primary video source defaults to the HD DVI input then attach it to the last still image captured. The list
when a case is begun. The default selection for the of possible annotations is maintained per user on the
primary video source is adjustable via the Setup Page Setup Page > Physicians Tab (Figure 16-1, Physicians
> General Tab > Camera Sub Tab (Figure 15-1, Camera Tab). The DRSHD 1080p will listen for approximately 5
Sub Tab). The secondary video source defaults to the seconds after a still image is captured for the user to
last SD source that was selected using the Video Input speak into the microphone. When the user verbalizes
Source Buttons (Figure 15-2). a preset annotation into the microphone, the DRSHD
FIGURE 15-1 1080p capture software will select the proper annotation
and save it to the captured image (Figure 16-2). Erasing
Dr. Sample will remove all default annotations from the
DRSHD 1080p.
FIGURE 16-1

FIGURE 15-2

FIGURE 16-2

U5010-12 PG018
RECORD (CONTINUED)

RECORD PAGE CONTROLS Annotation Button - used to type an annotation to the


Status Tab - displays the video segment time of the last still image captured at any time. When clicked, the
HD and SD video capture sources while video capture On Screen Keyboard will appear, allowing the user to
is in progress, the storage space that has been filled, enter an annotation for the image. The user may also
and the total number of video segments that have been click on the microphone icon and verbalize a preset
captured in the present surgical case. annotation into the microphone, the DRSHD 1080p
capture software will select the proper annotation and
Archiving Tab - displays status of digital files to be save it to the captured image. When video capture is
sent to the Archiving Location set on the Setup Page not in progress, the DRSHD 1080p Review Page can be
> Media Tab (Media Options). A green dot next to a accessed to add annotations to any of the images that
digital file indicates the digital file has successfully been have been captured in the present surgical case.
delivered to the designated Archive Location. A yellow
dot indicates the digital file is waiting to be delivered to Delete Button - delete the captured image shown on
the designated Archive Location. A red dot indicates the the upper right of the DRSHD 1080p Record Page. When
digital file has failed to be delivered to the designated clicked, a warning pop up window will appear asking
Archive Location. Clicking the X to the right side of a for confirmation that the user wishes to delete the
digital file that has been copied to the Archive Location image. When the warning pop up window is present,
will remove item from the Archive Status List, but will the software graphic controls will not respond until the
not delete the digital file from the designated Archive user clicks the OK or the Cancel Button on the pop up
Location. Clicking the X to the right side of a digital file window, however the DRSHD 1080p capture software
that has not been copied to the Archive Location will will continue to capture video and images in response
cancel the Archive job for that item and the digital file to Camera Remote Controls and foot pedal commands.
will not be copied to the Archive Location. Digital files
will remain in the DRSHD 1080p main storage location Close Case Button - used to signal the completion of
without regard to Archiving activities. If a digital file a surgical case and initiates all automated background
fails to copy to the designated Archive Location, the functions like automatic printing (Auto Printing),
DRSHD 1080p capture software will attempt to copy automatic send to DICOM (Auto DICOM Store), and
the digital file until the Archive job is completed, until automatic copy to Archive Location (Archiving).
the user clicks the X to the right side of a digital file to
cancel the Archive job, or until a new case is started. Mute Button - disable microphone input.

Information Tab - displays the Patient Study Information Capture Button - capture a still image from the primary
for the case in progress. When the DRSHD 1080p capture video source.
software is in Local Work List or HL7 configuration, the
Edit Button will be available. Click the Edit Button to alter Record Button - capture video from the selected primary
the Patient Study Information for the case in progress. video source or the primary and secondary video sources
when Dual Recording is active (S tandard and Dual
Levels Tab - features sliders that adjust the Brightness, Recording).
Contrast, Hue, and Saturation of the SD video stream.
The adjustments will show in the video files that are Pause - stop capture of video and complete a video
captured. The Levels Tab is only visible when a SD video segment file.
source is selected as the primary input. There are no
level adjustments for the HD video stream. Video Input Source and PIP Buttons - select the primary
and secondary video source to be captured.

PG019 U5010-12
RECORD (CONTINUED)

CAPTURING VIDEO CAPTURING STILL IMAGES


The DRSHD 1080p Record Page captures HD or SD video The DRSHD 1080p Record Page captures HD or SD
from the selected primary video input and optionally from still images from the selected primar y video input.
the primary and secondary video inputs simultaneously. Images are captured in JPEG format by default, and
Video segments are limited to 5 minutes maximum can optionally be changed to capture BMP or PNG image
Segment Length by default. The DRSHD 1080p Record files (Media Options). The DRSHD 1080p Record Page
Page will display an image of color bars and a “No Video will display an image of color bars and a “No Video
Input” warning message if an incoming video stream Input” warning message if an incoming video stream
on a selected HD or SD video input is not detected on on a selected HD or SD video input is not detected on
the Primary Video Source Window or the PIP Window the Primary Video Source Window or the PIP Window
(No Video Input). Video and still image capture can be (No Video Input). Still image capture can be initiated
initiated using peripheral input devices or using the using peripheral input devices or using the software
sof tware graphic controls displayed on the DRSHD graphic controls displayed on the DRSHD 1080p Record
1080p Record Page (Figure 17). Page. Still image capture only occurs on the video input
FIGURE 17 source that is selected as primary and is displayed on
the Primary Video Source Window.

To capture a video or a still image:


1. Start a case to navigate to the DRSHD 1080p Record
Page (Figure 18-1).
2. Select the desired Video Input Source Button to
focus control on this input (Figure 18-1).
3. Observe the incoming video stream displayed in the
Primary Video Source Window (Figure 18-1).
4. Begin video capture when desired via peripheral
input device or the Record Button (Figure 18-1).
5. Stop video capture via peripheral input device or
the Pause Button (Figure 18-1).
6. The user may capture a still image at any time via
peripheral input device or the Capture Button (Figure
18-1).
7. Optionally the user may speak into a connected
microphone immediately following the capture of
a still image to annotate (Figure 18-2).

U5010-12 PG020
RECORD (CONTINUED)

FIGURE 18-1

FIGURE 18-2

PG021 U5010-12
REVIEW

REVIEW PAGE FILTER BY PATIENT STUDY FIELD CRITERIA


The DRSHD 1080p Review Page is the digital file control When in List View, DRSHD 1080p Review Page can be
center of the capture software. The user is able to use the filtered by any of the pertinent patient fields by typing a
Review Page to review and expedite digital file workflow value into the search box above the field or clicking on
for cases that have been captured on the Record Page. a Patient Study Field at the top of the list. These filters
Surgical case digital files can be viewed, annotated, operate in the same manner as the filters on the Work
deleted, copied to a portable drive or network folders, List Page (Filter Work List) and persist over reboots of
burned to optical disc, printed, and sent to a DICOM the system (Figure 19).
Storage server while a surgical case is in progress or FIGURE 19
after a surgical case is completed.

The DRSHD 1080p Review Page has two default views


that will display based upon the presence of a surgical
case in progress or no surgical case in progress. When
a case is open and in progress on the Record Page, the
Review Page default to Case View and will only show
digital files for the case in progress. When a case is not
in progress, the Review Page will default to List View and
display all of the surgical cases stored on the DRSHD
1080p that are ready for review.

REVIEW PAGE VIEWS


List View - displays all of the surgical cases stored on
the DRSHD 1080p that are ready for review. To select
a case click the Checkbox to the left side of the page
next to the desired case.

Case View - displays a list of Thumbnails for each still


image and video captured. To select an image or video
click in the middle of the Thumbnail and a check icon
will appear over the Thumbnail.

Playback View - displays a large screen view of the


digital file selected. Videos will play back when selected.
To select an image or vide o click the Media Item
Checkbox on the upper right of the page and a check
will appear in the box.

U5010-12 PG022
REVIEW (CONTINUED)

REVIEW PAGE CONTROLS


List View Controls:
Clear Filters Button – removes all filters that have been Delete Item Button - (trash pail icon) erases an individual
applied to the case list on the DRSHD 1080p Review digital file. A warning dialog pop-up window will appear to
Page. confirm that the user wishes to delete the specified item.

Filter Box – enter text and numbers to filter the work DICOM Button - sends a selected case the preset DICOM
list by a keyword. Storage server when clicked. This button is only visible
when DICOM storage is enabled.
Activate Filter Button – (magnifying glass shaped icon)
sorts the case list by the filter text entered. Case View Controls:
Back Button – navigates to List View. This button is not
Deactivate Filter Button – (x shaped icon) removes the available when a surgical case is in progress.
filter from the case list sorting.
Images/Videos Buttons - show only the Thumbnails
Scroll Bars – move the orientation of the Patient Study for the images or videos for the selected surgical case.
Information Fields to enable the user to view all of the
cases and data in the case list. Thumbnails - small images that represent videos and
still captures for the selected surgical case. Click in the
Scroll Bar Arrow Buttons – move the Scroll Bars in center of the Thumbnail to select the digital file for print,
small increments. copy, burn, or delete.

Top Checkbox - selects and deselects all cases. Annotate Button - allows a user to add an annotation
to the digital file.
Checkbox - selects and deselects the individual case
or multiple cases to copy, burn, or delete. Playback View Button - navigates to Playback View
and displays the digital file. Videos will begin playback
View Button - changes the page to Case View and for immediately.
the corresponding surgical case.
Delete Item Button - (trash pail icon) allows a user to
Edit Button - opens pop-up window that allows the user delete the displayed digital file. A warning dialog pop-up
to edit Patient Study Information of the corresponding window will appear to confirm that the user wishes to
surgical case. delete.

Append Button - gives user the option to reopen the Select All Button - selects all case digital files for copy,
corresponding surgical case to add additional digital burn, or delete.
files. When clicked, a warning dialog pop-up window will
appear to confirm that the user wishes to navigate to Select Annotated Button - selects all annotated case
the Record Page with the corresponding surgical case digital files for copy, burn, or delete.
loaded to append digital files to the case.
Select None Button - deselects all selected digital files.
Copy Button - navigates to the Copy Page to duplicate
case digital files to a portable drive or network folders. Copy Button - navigates to the Copy Page to duplicate
case digital files to a portable drive or network folders.

Burn Button - navigates to the Burn Page to duplicate Burn Button - navigates to the Burn Page to duplicate
case digital files to optical disc. case digital files to an optical disc.

PG023 U5010-12
REVIEW (CONTINUED)

PRINT PAGE
Delete Button - allows a user to delete the selected The DRSHD 1080p Print Page is accessed from the
digital files. A warning dialog pop-up window will appear Review Page and offers full featured printing options.
to confirm that the user wishes to delete. The user can send the print job to a list of printers or
create a PDF file to print later, select a page layout,
DICOM Button - sends a selected digital file to the preset crop HD images to a 4:3 ratio to save black ink with
DICOM Storage server when clicked. endoscopic images, add a personalized heading and logo
to the top of the printed page, make adjustments to the
Playback View Controls: images prior to printing. The user can also include dates,
Back Button – navigates to Case View. annotations, and a number of Patient Study Fields to the
printed page. The printing options can be preset to the
Media Item Checkbox - selects the displayed digital desired default settings on the Setup Page > Printing
file for copy, burn, or delete when the user navigates Tab (Figure 20), set per Physician on the Setup Page >
back to Case View. Physicians Tab (Auto Printing) or adjusted for each print
job individually on the Print Page. Any adjustments made
Annotate Button - allows a user to add an annotation to the print layout on the Print Page will not be saved
to the digital file. after the page is closed, it is recommended that the
user set the preferred default print settings on the Setup
Delete Button - allows a user to delete the displayed Page > Printing Tab (Printing Tab) during the DRSHD
digital file. A warning dialog pop-up window will appear 1080p initial setup.
to confirm that the user wishes to delete. FIGURE 20

Case View Button – navigates to Case View.

Controls specific to video playback:


Mute Button 'Speaker Icon' - activates and deactivates
audio playback of the video displayed.

Volume Slider - adjusts volume of the audio playback


of the video displayed.

Play/Pause Button - starts and pauses playback of the


video displayed.

.25x Button - slows video playback to 25% of full speed.

.50x Button - slows video playback to 50% of full speed.

Normal Button - returns video playback to full speed.

Capture Button - captures a still image from the video


displayed and adds it to the case digital files.

Loop Button - when activated the selected video repeats


playback until stopped. Looped playback is activated by
default as the Playback View Page is displayed.

Seek Bar - click and drag the dot on the Seek Bar to
navigate to a desired place in the video displayed.
U5010-12 PG024
REVIEW (CONTINUED)

PRINT PAGE CONTROLS


Print Preview Navigation Buttons - used to display each Enhancements Checkboxes - when activated the
page of a multiple page print job. Enhancement Sliders will become available and a print
job will include preset image color, brightness, contrast,
Print Preview Window - displays the print job with all of and sharpen adjustments.
the set options as it will look when printed.
Enhancement Sliders - used to adjust the selected
Selected Printer - used to select an installed printer to Enhancement. The result of the adjustment will be seen
send the present print job. in the Print Preview Window as it is made.

Page Layout Dropdown - used to select 1, 2, 4, 6, 8, or Copies - adjusts the number of copies for the present
15 images per page. print job.

4:3 Crop Checkbox - activate this checkbox to crop a Print Button - sends the print job to the specified printer
16:9 image to a 4:3 image. This is useful with scope with all of the preset printing options.
images.
Save as PDF Button - saves the print job to a PDF file
Include Annotations Checkbox - when activated the all of the preset printing options.
print job will include annotations under the images.
Exit Button - closes the Print Page without sending the
Header Text Box - add text here to add a header or a present print job.
title for all of the pages in the present print job.

Logo Checkbox - when activated the print job will include


a preselected logo at the top of the pages

Logo Dropdown - used to select a custom logo JPEG


file. This file will be saved into the software and able to
be reused at any time.

Footer Fields Dropdown - used to select Patient Study


Fields to be added to the present print job.

Footer 1-6 - used to set the specified Patient Study Field


to respective position of the numbered footer.

Include Image Date Checkbox - when activated the


print job will include the date of image capture under
the images.

E n ha n ce me nts E nable d - when activated the


Enhancements Dropdown will be available.

Enhancements Dropdown - used to access digital


Gamma, Brightness, Saturation, Sharpen, and Contrast
adjustments for the images in the present print job.

PG025 U5010-12
REVIEW (CONTINUED)

COPY PAGE COPY PAGE CONTROLS


The Copy Page displays the size of the digital files to be Include Patient Data Checkbox - when activated the
copied, and the percentage of completion of the present Patient Study Information will be included in a text file for
copy job. To copy a case to an external device or network each case that is copied to a portable drive or network
folder, select the desired digital files then click the Copy location.
Button on the Review Page to display the Copy Page.
The user can optionally copy the case into a password Secure Zip Checkbox - when activated the software will
protected zip folder (Figure 21). If a copy job that is copy the case to the specified location in a ZIP file that
too large for the selected Copy Location is created and can optionally be password protected. The password
begun, the software will copy all of the digital files that can be set on the Setup Page > Media Tab (Media Tab).
can be sent, then prompt the user with a warning that
there was not enough space on the disk. Copy Location Button - when clicked, the Location Page
FIGURE 21 will become available.

Eject Button - used to safely eject a USB device after a


copy job has completed.

Abort - used to cancel a copy job in progress.

Exit - closes the Copy Page.

U5010-12 PG026
REVIEW (CONTINUED)

LOCATION PAGE LOCATION PAGE CONTROLS


The DRSHD 1080p Location Page is accessed from the Removable Drives Dropdown - an arrow will appear
Copy Page when the Copy Location Button is clicked to the left of the dropdown when a removable drive
and is used to select a location to copy surgical cases is connected and available. When an arrow is present
(Figure 22). click the dropdown to display the available devices and
FIGURE 22 locations.

MedXStream (iOS) Dropdown - an arrow will appear to


the left of the dropdown when an iOS device is connected
via USB cable and available. When an arrow is present
click the dropdown to display the available devices.

Remote Dropdown - an arrow will appear to the left of


the dropdown when a network location is connected and
available. When an arrow is present click the dropdown to
display the available locations. To connect to a network
folder click the Connect to UNC Button.

Connect to UNC Button - when clicked the user will be


able to enter a network path, username and password
to access a shared folder or network location. Leave the
username and password blank if they are not required.
After a UNC path is entered, it must be selected on the
Remote Dropdown to be the chosen Copy Path. For more
help with UNC locations contact the facility network
administrator.

PG027 U5010-12
REVIEW (CONTINUED)

BURN PAGE BURN PAGE CONTROLS


The DRSHD 1080p Burn Page is accessed from the Include Patient Data Checkbox - when activated the
Review Page and is used to burn a surgical case or Patient Study Information will be included in a text file
multiple cases to optical disc. The Burn Page displays for each case that is burned to optical disc.
the number of discs needed for a burn job, the count of
how many discs have been burned in the present job, the Burn Button - begins burning the optical digital files. If
disc type that is in use, the capacity of the disc type, the more than one optical disc is required to complete the
size of the digital files to be burned, the speed at which burn job, the burn page will display the number of discs
the optical digital files are burned, and the percentage required and the user will be prompted to insert another
of completion of the present burn job. To burn a case disc as the burn job finishes with the disc in the tray.
to optical disc, select the desired digital files then click
the Burn Button on the Review Page to display the Burn Erase Button - erases rewritable disc digital files.
Page (Figure 23).
FIGURE 23 Eject Button - opens the tray of the optical drive.

Abort - cancels a burn job and makes the Exit Button


available. If the Abort button is clicked during a disc
burn; the disc will become unusable. Discard or erase
any optical digital files that has been aborted during a
burning process.

Exit Button - closes the Burn Page.

U5010-12 PG028
SETUP

SETUP PAGE SYSTEM SUB TAB


The DRSHD 10 8 0 p S etup Page allows the user to The System Sub Tab contains controls and information
conf igure sof t ware preferences, sele ct Work List that relate to system maintenance and global system
configurations, select digital file types, set default controls. The System Sub Tab is displayed by default
printing options, configure work list and DICOM server when the user navigates to the Setup Page (Figure 25).
communications, adjust per user settings, and set
software security options. The Setup Page is divided Version Numbers - displays the sof tware GUI and
into Tabs on the left side of the page and Sub Tabs near Firmware version numbers.
the top of the page. Each Tab and Sub Tab contains a
number of controls that relate to its title. The Tabs and Calibrate Touch Screen Button - launches the ELO touch
Sub Tabs on the Setup Page are arranged in a logical screen monitor calibration utility.
order to assist the user in finding the correct setting
quickly. Reset System Settings Button - resets all global settings
to factory default. This button does not erase work list
GENERAL TAB entries, cases, Physician settings, or Security settings.
The General Tab contains Sub Tabs for System, Camera,
Network, and Regional settings. It also contains a Sub Software Update Button - used in conjunction with a
Tab that displays the DRSHD 1080p User Manual. The USB flash drive to update the DRSHD 1080p software.
General Tab is displayed by default when the user
navigates to the Setup Page. Time, date, time zone, Wireless Configuration But ton - used to create a
network, and language settings are just some of the dedicated wireless network for use with the MedXStream
settings that can be found in the General Tab and its app.
Sub Tabs (Figure 24).
FIGURE 24 Delete Cases, Work List Entries, Physicians and Users
- deletes all cases, Work List entries, Security Users, and
Physicians with one click.

Delete Physicians - deletes all Physicians and their


settings.

Delete Cases - deletes all stored cases and Work List


entries.

Delete Users - deletes all Security Users.

Delete Storage History - deletes the internal list of all


cases that have been sent to DICOM. This control can
be used when a case has been successfully sent to
DICOM storage and the user wishes to send the case
a second time.

Auto Delete Cases - set the system to automatically


delete cases older than 60, 90 or 120 days star ting
when the control is set and continuing until the control
is deactivated.

PG029 U5010-12
SETUP (CONTINUED)

CAMERA SUB TAB


Delete Cases - manually delete cases older than the 30, The Camera Sub Tab contains controls for the Video
60, 90, 120 days one time. Use the Older Than dropdown Input Source and PIP Buttons. Inputs may be disabled,
to select the number of days. Click the Delete button to renamed, or set as the default (Figure 26).
delete the cases older than the number of days selected.
Video Input Source Label Text Boxes - add a custom
Mute System Audio Checkbox - deactivates the DRSHD label to each Video Input Source Button on the Record
1080p Audible Notification feature that gives the user Page. There is a 9 character limit for the custom label.
audible feedback when a still image is captured and
when a video is started or stopped. Video Input Source Enabled Checkboxes - enables or
disables the corresponding Video Input Source Button
E nable O S K C he c k box - activates the On S cre en on the Record Page.
Keyboard feature. This control is activated by default.
Default Camera - selects the default Video Input Source
Enable Voice Annotation - activates the Speech to Text Button that will always be selected as a surgical case
feature for Annotations. star ts. If a Video Input Source Button is disabled, it
FIGURE 25 cannot be set as the Default Camera.
FIGURE 26

U5010-12 PG030
SETUP (CONTINUED)

NETWORK SUB TAB REGIONAL SUB TAB


The Network Sub Tab contains settings that allow the The Regional Sub Tab contains settings that allow the
user to view or adjust Internet Protocol version 4 (IPv4) user to adjust the system time and date, time zone,
settings for the DRSHD 1080p. IPv6 is not supported language and time/date format (Figure 28).
(Figure 27).
Date/Time Text Box - used to adjust the system time and
DHCP/Manual Dropdown - activates or disables Dynamic date. Enter a date and time then click the Save Button
Host Configuration Protocol for the DRSHD 1080p to adjust the time and date. The present time and date
network connection to the Ethernet Port. Select DHCP are displayed by default.
to allow the facility’s network infrastructure to assign
network values to the DRSHD 1080p automatically. Time Zone Dropdown - sets the time zone for the DRSHD
Choose Manual to set network values to the DRSHD 1080p capture software. (UTC -5:00) Eastern Time (US
1080p that do not change. & Canada) is the default setting. Changing the time zone
will automatically adjust the current time.
IP Address - displays the present IP address of the
DRSHD 1080p. Language Dropdown - sets the language displayed
for the DRSHD 10 8 0p capture sof tware. Available
Subnet - displays the present IP subnet mask of the languages are Arabic, Chinese (Traditional), Deutch,
DRSHD 1080p. English, Espanol, Francias, Italiano, Japanese, Korean,
Portuguese, Russian, Svenska, and Turkish.
Gateway - displays the present default gateway of the
DRSHD 1080p. Culture Dropdown - sets user interface to use the date
FIGURE 27 and time format common to the specified region.
FIGURE 28

PG031 U5010-12
SETUP (CONTINUED)

MANUAL SUB TAB MEDIA TAB


The Manual Sub Tab displays the DRSHD 1080p User The Media Tab contains settings that relate to surgical
Manual (Figure 29). case digital files and how they are stored. Default
case folder names, archive controls, image file types,
Export Manual Button - This allows the user manual to video compression settings, camera color spacing,
be copied to an external location in PDF format. PIP activation, video segment length, and microphone
muting are found in the Media Tab (Figure 30). The Media
Page Down Button - navigates to the next page. Tab Settings can be set Per Physician on the Setup Page
> Physicians Tab (Physicians Tab).
Page Up Button - navigates to the previous page.
Media Tab Controls:
Search Button - (magnifying glass shaped icon) search Case Storage Folder Name Dropdown - selects the
the manual for keywords. folder name convention which cases are stored on a
FIGURE 29 portable drive.

Available options are:


Timestamp - includes a date and time stamp with a
randomly generated number.

Patient Full Name - includes the first and last name. If


no name is entered, the folder name will be “Unknown(#)”.

Patient ID - folder name is taken from the Patient ID


field. If no Patient ID is entered, the folder name will be
“Unknown(#)”.

Secure Zip Password Button - displays the Secure


Zip Password window used to assign a password to
secure ZIP files. Users are required to enter the previous
password if changing passwords. If no password is set,
all ZIP files will open without a password. Secure .ZIP
files are selectable in the copy dialog.

HD Video Channel Sub Tab - contains adjustments for


the HD video quality and color space.

(HD) Quality Dropdown - Used to adjust the amount of


compression (bit rate) used when creating HD video files.
Changing the Quality setting adjusts both video quality
and file size. The user can select High at 30Mbit/s,
Medium at 20Mbit/s, or Low at 8Mbit/s.

Color Space Dropdown - adjusts the HD capture color


space to RGB or Y U V to adapt to multiple surgical
cameras. Consult camera documentation to identify
the color space required.

SD Video Channel Sub Tab - contains adjustments for


the SD video quality.
U5010-12 PG032
SETUP (CONTINUED)

LOCATION PAGE
(SD) Quality Dropdown - Used to adjust the amount of The DRSHD 1080p Location Page is accessed from the
compression (bit rate) used when creating SD video files. Media Tab when the Archiving Location Button is clicked
Changing the Quality setting adjusts both video quality and is used to select a location to archive surgical cases
and file size. The user can select High at 10Mbit/s, to external devices and network folders (Figure 31-1).
Medium at 8Mbit /s, or Low at 5Mbit /s. The quality
setting for SD video recordings adjusts the quality for Location Page Controls:
all three of the SD inputs. Removable Drives Dropdown - an arrow will appear
to the left of the dropdown when a removable drive
Archiving Location Button - displays the Location is connected and available. When an arrow is present
window that allows the user to select a Removable Drive, click the dropdown to display the available devices and
a smart device with the MedXStream app, or a Remote locations.
network folder.
FIGURE 30 MedXStream (iOS) Dropdown - an arrow will appear
to the lef t of the dropdown when an iOS device is
connected and available. When an arrow is present click
the dropdown to display the available devices.

Remote Dropdown - an arrow will appear to the left of


the dropdown when a network location is connected and
available. When an arrow is present click the dropdown to
display the available locations. To connect to a network
folder click the Connect to UNC Button.

Connect to UNC Button - when clicked the user will be


able to enter a network path, username and password
to access a shared folder or network location. Leave the
username and password blank if they are not required.
After a UNC path is entered, it must be selected on the
Remote Dropdown to be the chosen Archive Path.

To specify a Removable Drive as the Archive Location:


1. Connect an external drive to the DRSHD 1080p. A
new USB drive may take up to 30 seconds to install
and become usable.
2. Click the Archive Location Button. If no archive has
been set, this box will show “None” (Figure 31-1).
3. Click the Removable Drives Dropdown after the
arrow to the left of the dropdown appears (Figure
31-2).
4. Select the drive letter and path desired (Figure 31-2).
5. Click OK (Figure 31-2).

Image File Type Dropdown - used to select the file type


to be used for still image capture. The choices are BMP,
JPEG, or PNG.

PG033 U5010-12
SETUP (CONTINUED)

Segment Length Checkbox - activates video segment FIGURE 31-2


time limitation. The segment time limitation is active by
default. When deactivated, the captured video files are
unlimited in length or size.

Segment Length Minutes and Seconds Dropdown -


used to set the amount of time of continuous recording
that is possible before a video segment file is completed
and a new one is created. The default segment time is 5
minutes. A 5 minute HD video segment at High Quality
setting is approximately 1 GB in size. It is recommended
that a segment length limitation be set when the user
expects to copy digital files to a USB flash drive or to
burn digital files to a CD/DVD.

Enable PIP Checkbox - activates the Video Input Source FIGURE 31-3
PIP Buttons on the Record Page.

D u al Re c o rdin g C he c k b ox - when activated a


synchronized start and stop of recordings of both the
primary and secondary video sources will occur with the
Record and Pause Button on the Record Page.

Mute Microphone Checkbox - when activated the


microphone input will be muted by default. This control
can be overridden on the Record Page by clicking the
Mute Button.
FIGURE 31-1

U5010-12 PG034
SETUP (CONTINUED)

PRINTING TAB
The Printing Tab contains controls to select the default Include Image Date Checkbox - when activated a print
printing settings that are displayed as the Print Button job will include the image date of capture under the
is clicked on the Review Page (Print Page) and used with images.
the Auto Print feature. The Printing Tab Settings can be
selected Per Physician on the Setup Page > Physicians Number of Copies - adjusts the default number of copies
Tab (Figure 32). for a print job.

Printing Tab Controls: E n ha n ce me nts E nable d - when activated the


Print Preview Window - displays the print job with all of Enhancements Dropdown will be available.
the set defaults as it will look when printed.
Enhancements Dropdown - used to access digital
Printer Dropdown - used to select the default printer Gamma, Brightness, Saturation, Sharpen, and Contrast
for print jobs. adjustments for the images in the present print job.

Page Layout Dropdown - used to select 1, 2, 4, 6, 8, or Enhancements Checkboxes - when activated the
15 images per page. Enhancement Sliders will become available and a print
job will include preset image color, brightness, contrast,
4:3 Crop Checkbox - activate this checkbox to crop a and sharpen adjustments.
16:9 image to a 4:3 image. This is useful with scope
images. Enhancement Sliders - used to adjust the selected
Enhancement. The result of the adjustment will be seen
Auto Print Checkbox - activates the Auto Print feature. in the Print Preview Window as it is made.
Auto Print is a expedited printing feature that sends a
print job to the printer without user intervention as a Print Test - sends a sample print job to the specified
surgical case is performed. When Auto Print is enabled, printer with all of the preset printing options. The sample
still image captures are printed using the default printing print job as it will be sent to the printer is displayed in
settings immediately as a print job becomes available. the Print Preview Window.
When automatic printing is in progress, a printer icon FIGURE 32
will appear on the title bar of the recording software.

Include Annotations Checkbox - when activated a print


job will include annotations under the images.

Header Text Box - add text here to add a header or a


title for all of the pages in a print job.

Logo Checkbox - when activated the print job will include


a preselected logo at the top of the pages

Logo Dropdown - used to select a custom logo JPEG


file. This file will be saved into the software and able to
be reused at any time.

Footer Fields Dropdown - used to select Patient Study


Fields to be added to the present print job.

Footer 1-6 - used to set the specified Patient Study Field


to respective position of the numbered footer.
PG035 U5010-12
SETUP (CONTINUED)

WORK LIST TAB


The Work List Tab contains Sub Tabs titled Local, Server 1-4 Sub Tabs - used to select connection settings
MWL, and HL7 used to select one of the three Work and select from one of four DICOM servers to be the
List configurations (Figure 33). In addition to being a default DICOM Storage Server. Each Server Sub Tab
work list client the DRSHD 1080p can combine work list contains Text Box controls for each of the server settings
functionality with DICOM storage. This combination will that are required to connect to a MWL Server.
help facilitate the import of patient demographics within
MWL and HL7 Work List configurations as well as the Filter Work List Dropdown - used to hide cases that
export of patient demographics and digital files within are older that the selected value from appearing on the
surgical cases. This makes the DRSHD 1080p a client Work List Page when in MWL Work List configuration.
machine that can be assigned cases to be completed by
a work list administrator; The DRSHD 1080p can then Server Sub Tab Controls:
optionally send the captured digital files and Patient Host IP Address Text Box - enter the IP address of the
Study Information to be stored on a digital file storage work list server.
ser ver. A PACS ser ver can be programmed to send
cases via MWL to many DRSHD 1080p machines in one Port - enter the port number used to communicate with
facility and then to store the captured digital files from the work list server.
the surgical cases via DICOM Storage. This creates a
master work list and case reconciliation system that Calling AE Title - enter the Application Entity Title
can be used to help manage multiple OR workflow. For selected for the DRSHD 1080p.
more information on automated digital file storage see
Archiving (Archiving), and DICOM Storage Tab (DICOM Called AE Title - enter the Application Entity Title of
Storage Tab). the work list server.

Local Work List Sub Tab (default) – all cases are Test Connection Button - when clicked the DRSHD
created and maintained on the DRSHD 1080p locally 1080p capture software will attempt to communicate
when this Sub Tab is selected and the Save Button is with the work list server using the selected settings.
clicked. There is no required Patient Study Information The software will reply with a pop up window that tells
to create and begin or to schedule a future case in this the user that the communication succeeded or failed. If
Work List configuration. the communication fails check the Server settings then
check the network connection and settings. Contact the
Modality Work List Sub Tab (MWL) – the Work List facility DICOM and Network Administrators for more
Server delivers cases to the DRSHD 1080p. The DRSHD help.
1080p works as a client to the work list server. Scheduled
cases can also be created and completed locally. Patient HL7 Sub Tab - The DRSHD 1080p is capable of receiving
Study Information is not required to create and start a HL7 Scheduling Information Unsolicited messages
case or to schedule a future case. (SIU^S12, SIU^S13, SIU^S14, SIU^S15) for impor ting
patient demographic information from an Electronic
MWL Sub Tab Controls: Medical Record system or Electronic Health Record
Filter By AE Title - allows only cases assigned to the system (EMR/EHR). The Patient Information Fields are
specified Application Entity Title to be imported from imported onto the Work List Page. If DICOM Storage is
the work list server. This filter defaults to the Calling AE active, the Patient Information Fields are translated into
Title when it is entered and set. The filer can be cleared DICOM tags automatically and then stored with the case
manually if the user wishes to import the complete work digital files on the DICOM Storage Server. Unscheduled
list without this filter. cases can also be created and completed locally. There
is no required Patient Study Information to create and
begin a case.

HL7 Version - displays the compatible HL7 version


U5010-12 PG036 number
SETUP (CONTINUED)

Receiver Port - enter the port for incoming messages. Host IP Address Text Box - enter the IP address of the
work list server.
Receiver Timeout - enter the number of seconds before
timeout. Port - enter the port number used to communicate with
FIGURE 33 the work list server.

Calling AE Title - enter the Application Entity Title


selected for the DRSHD 1080p.

Called AE Title - enter the Application Entity Title of


the work list server.

Test Connection Button - when clicked the DRSHD


1080p capture software will attempt to communicate
with the work list server using the selected settings.
The software will reply with a pop up window that tells
the user that the communication succeeded or failed. If
the communication fails check the Server settings then
check the network connection and settings. Contact the
facility DICOM and Network Administrators for more
DICOM STORAGE TAB
help.
The DICOM Storage Tab contains all of the controls and
settings used to enable copy and storage of surgical
With these settings entered, the DICOM Button on the
cases to a DICOM Storage Server (Figure 34).
Review Page will become visible enabling the user to
send cases to DICOM Storage. When a case is selected
The default Single Frame Compression Configuration,
on the Review Page List View or a digital file is selected
Single Frame SOP Class, Multi Frame SOP Class Settings
on the Review Page Case View, the user clicks the DICOM
are set to the most common settings that are used
Button to send this item to the preset DICOM Storage
in DICOM Storage protocol by default. The Commit
server.
Checkbox deactivated by default is also the most
common correct setting. The typical set up only requires FIGURE 34
the user to activate the Storage Enabled Checkbox, enter
a Host IP Address, Port, Calling AE Title and a Called AE
Title to activate the DICOM Storage.

Setting Default Value

Single Frame Implicit VR Little Endian


Compression
Configuration
Single Frame SOP Class Secondary Capture Image
Storage
Multi Frame SOP Class Multi Frame Secondar y
Capture Image Storage
Commit Deactivated

PG037 U5010-12
SETUP (CONTINUED)

DICOM STORAGE ADVANCED DICOM STORAGE SETTINGS


A Storage Server can be set up to upload digital files S in gle Fra me C o m p ressio n C o nf ig u ratio n - T he
created on the DRSHD 1080p, to a PACS via DICOM DRSHD 1080p offers four compression configurations
protocol. While Storage Server is enabled, the DRSHD or transfer syntaxes to transmit still capture images to
1080p can automatically transmit digital files to the DICOM Storage. The compression configuration should
selected server upon close of the active case. Conversely, be selected according to the supported configuration(s)
users may choose to manually upload case digital files of the DICOM Storage server. The table below describes
to the Selected Server on the Review Page after the case the available Single Frame Compression Configurations
has been completed. and their Unique Identifiers.

Before implementing DICOM Storage on the DRSHD Transfer Syntax DICOM UID
10 8 0 p , t h e M e d X C h a n g e D I C O M C o nfo r m a n c e
S tatement should be read to determine basic Implicit VR Little Endian 1.2.840.10008.1.2
interoperability with the storage server. The DRSHD Explicit VR Little Endian 1.2.840.10008.1.2.1
1080p DICOM Conformance Statement has been created JPEG Baseline (Process 1) 1.2.840.10008.1.2.4.50
to provide a DICOM Administrator with the information
JPEG Lossless, 1.2.840.10008.1.2.4.70
required to take the first step in determining whether
Nonhierarchical, First-
and to what extent DICOM implementations are able to
Order Prediction
inter-operate. In addition to review of the DRSHD 1080p
(Processes 14 [Selection
DICOM Conformance Statement, it is suggested that
Value 1])
a test case be performed and a complete exchange
of communications be completed to ensure that all
of the settings, digital file types, and SOP classes are Single Frame SOP Class - The DRSHD 1080p offers two
compatible between the DRSHD 1080p and the DICOM Service-Object Pair (SOP) Classes or abstract syntaxes
Storage server. to transmit still images to DICOM Storage. The table
below describes the available Single Frame SOP Classes
The DRSHD 1080p capture software requires the fields and their Unique Identifiers.
displayed in the table below to be populated when
creating a surgical case to be sent to the DICOM Storage Abstract Syntax DICOM UID
server.
Secondary Capture 1.2.840.10008.5.1.4.1.1.7
Image Storage
Patient Study Information Field DICOM Tag VL Endoscopic Image 1.2.840.10008.5.1.4.1.1.77.1.1
Patient Last Name 0010,0010 Storage

Patient ID 0010,0020
Multi-Frame SOP Class - The DRSHD 1080p offers two
Patient Date of Birth 0010,0030 Service-Object Pair (SOP) Classes or abstract syntaxes
Patient Sex 0010,0040 to transmit videos to DICOM Storage. The table below
describes the available Multi- Frame SOP Classes and
Study Date 0008,0020
their Unique Identifiers.
Modality 0008,0060

Abstract Syntax DICOM UID

Multi-Frame 1.2.840.10008.5.1.4.1.1.7.4
Secondary Capture
Image Storage
Video Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1.1
Image Storage

U5010-12 PG038
SETUP (CONTINUED)

Mutli-Frame Compression Configuration - The DRSHD Arrow Up and Down Buttons - used to reorder the Patient
1080p proposes the MPEG2 1.2.840.1000 8.1.2.4.100 Information Fields on the Work List Page. Select a Field
compression configuration or transfer syntax to transmit and then click the Arrow Buttons to move the field up
video to DICOM Storage. or down the list. The list order from top to bottom on
the Patient Fields Tab will order from left to right side
Commit - Users may enable the storage commit feature in of the page on the Work List Page and Review Page in
the event the PACS requires storage commitment. There Case View.
is no notification to the user regarding commitment.
Commit is disabled by default for the most compatibility Display Name - click on the names of the Patient
with DICOM Storage Servers. Information Fields to select them and move them with
the Arrow Buttons or click the Manage button on the
Retries - Set the number of times the commit right side of the Display name to alter it.
service attempts to commit.
Re quire d a n d E na ble d C he c k b oxes - display the
Time Interval - Set the number of minutes between Required and Enabled status of the corresponding
commit retries. Patient Information Field. These settings can be altered
by clicking the Manage Button.
AE Title - Specify the Application Entity Title of the
Commit server. Scroll Bars – move the orientation of the Patient Study
Information Fields to enable the user to view all of the
SCP Port - Specify the port number which is used Display Names listed.
to commit.
Scroll Bar Arrow Buttons – move the Scroll Bars in
Separate Association Checkbox - Commit files and small increments.
wait for result on a separate association. FIGURE 35

PATIENT FIELDS TAB


The Patient Fields Tab controls what Patient Information
Fields are visible on the Work List Page, are set to be
required to star t a surgical case, the default values
for Patient Information Fields, and Custom Patient
Information Fields (Figure 35). The default Patient
Infor mation Fields are displayed on the Tab with
checkboxes that tell the user if the Patient Information
Field is required or not required, and visible or not visible.
The user can click the Manage Button on any of the
Patient Information Fields to Update the Patient Field
the Display Name, Data Type, Default Value and to set
the Patient Information Field as required or not required,
and visible or not visible.

Patient Fields Tab Controls:


New Patient Field Button - used to create Custom
Patient Fields

PG039 U5010-12
SETUP (CONTINUED)

PHYSICIANS TAB
The Physicians Tab contains the user profiles that are Manage Button - displays the Sub Tabs for each of the
saved for the Performing Physicians that are assigned Physician Profile Settings.
to a surgical case on the Work List Page (Figure 36). FIGURE 36
Physician profiles can be created prior to a surgical case
or as a case is created. Preset Physician profiles will be
selectable from the Performing Physician Dropdown on
the Work List Page in the Patient Study Information. The
user profile for the Physician can be set to the default
settings or set uniquely for the user’s preferences. The
Physician user profile will be loaded automatically as
the Physician is selected as the Performing Physician
in a case on the Work List Page. A single Physician may
also create multiple profiles to be used with each type
of surgical case.

Each Physician can select unique Patient Fields, Printer


Settings and Media Settings. They can also create unique
Annotations and Procedures to be selectable when they
are assigned to the case in progress. With a preset profile
the user has no settings to adjust as the different profiles
are loaded automatically as a case is started.

Performing Physicians contained in the DICOM Modality


Work List and HL7 work lists are automatically added
to the DRSHD 1080p capture software Physicians list.

Physicians Tab Controls:


Add Physician Button - activates the Add Physician
Window used to create a new Physician profile.

Sync Settings To Global Checkbox - when active the


Physician Profile will always sync to the default Printing,
Media, and Patient Fields settings. When not active the
user may select unique Printing, Media, and Patient
Fields settings that will be loaded when the Physician
is assigned to a case as the Performing Physician. This
checkbox can be activated when a new Physician is
created or when the user clicks the Edit Button.

Edit Button - activates the Update Physician window


used to alter the Full Name of the Physician and to edit
the Sync Settings To Global Checkbox.

Delete Button - erases the Physician profile. Erasing


Dr. Sample will remove all default annotations from the
DRSHD 1080p.

U5010-12 PG040
SETUP (CONTINUED)

PHYSICIAN PROFILE SUB TABS PROCEDURES TAB


The Physician Profile Sub Tabs contain controls for The Procedures Tab contains the list of Procedure
unique user set tings for Annotations, Procedures, D e s c r ipti o ns a n d t h e ir as s o c i ate d P ro c e d u re ID
Printing, Media, and Patient Fields. The settings created (Figure 37). There are a number of preset Procedure
in these tabs are unique to the Physician; if the user Descriptions that are displayed on the Procedures Tab
wishes to set the default global settings, they can by default. Users may add Procedure IDs to the default
be found on the Media Tab (Media Tab), Printing Tab Procedure list or make a custom list of Procedure IDs and
(Printing Tab), and the Patient Fields Tab (Patient Fields associated Procedure Descriptions commonly used in
Tab) on the Setup Page. their practice. When entering Patient Study Information,
stored Procedures may be selected from a the Procedure
Annotations Sub Tab - use the Add Annotation Button Description and the Procedure ID dropdowns in the
to create a new Annotation and optionally share it with Patient Study Information on the Work List Page.
the rest of the users. Sharing an Annotation will make
it appear in all user Annotation lists. Add Procedure Button - activates the Add Procedure
window used to create a new Procedure Description and
Annotations Sub Tab Controls: associated Procedure ID.
Add Annotation Button - use the Add Annotation Button
to create a new Annotation. Edit Button - activates the Update Procedure window
used to alter the Procedure Description and associated
Edit Button - use to edit the Annotation and Shared Procedure ID.
status.
Delete Button - erases the Procedure Description and
Delete Button - deletes the Annotation. associated Procedure ID.
FIGURE 37

Procedures Sub Tab - used to create unique Physician


prof ile set tings. S e e Procedures Tab for controls
(Procedures Tab).

Printing Sub Tab - used to create unique Physician


profile settings. See Printing Tab for controls (Printing
Tab).

Media Sub Tab - used to create unique Physician profile


settings. See Media Tab for controls (Media Tab).

Patient Fields Sub Tab - used to create unique Physician


profile settings. See Patient Fields Tab for controls
(Patient Fields Tab).

PG041 U5010-12
SETUP (CONTINUED)

SECURITY TAB
The Security Tab is used to administrate the DRSHD
1080p capture software Security feature (Figure 38).
When the optional security feature is activated users
must provide a preset Username and Password to gain
access to software functions. This feature helps prevent
unauthorized users from accessing the DRSHD 1080p
recording software. The Usernames and Passwords
stored on the Security Tab also allow users to log into the
DRSHD 1080p capture software using the MedXStream
smart device software.

Security Tab Controls:


Securit y Enabled Checkbox - used to activate the
Security feature enforcement. Users must have a profile
in the Security Username list and provide a preset
Username and Password to gain access to software
functions when the Security Feature is activated. When
a user activates the Security Enabled Checkbox, the
Security feature will be active on the next launch of the
DRSHD 1080p capture software.

Add User Button - displays the Add User window used


to create a new Security User. The user created will
have the ability to sign in to the DRSHD 1080p capture
software.

Edit Button - used to alter Username, Password, and


Notes of a user. The user’s current password is required
to edit a Security Username, Password, or Notes.
FIGURE 38

U5010-12 PG042
TECHNICAL

DRSHD 1080P PRODUCT SPECIFICATIONS


Still Capture: Overall Dimensions:
HD, SD - .BMP, .JPEG, .PNG CPU: 6 in H x 14.0 in W x 14.0 in D
14.5 cm H x 35.5 cm W x 35.5 cm D
Motion Video Capture: Touch Screen (Panel and base): 10.9” (276.52mm) H x
HD 720p 50/6 0fps, 1080p 25/30fps, SD 4 80i/576i 15.1” (383mm) W x 8.4” (213mm) D
25/30fps - AVCHD H.264 Encoding .MP4
Weights:
Hard Drive Capacity: CPU: 9.3 kg / 20.5 lb.
2TB Total capacity, 1.77TB for storage Touch Screen: 5.2 kg / 11.5 lb.

Inputs:
DVI-I x 1: 720p, 1080i, 1080p, 50/59.94/60hz
S-Video x 1, PAL/NTSC
Composite x 2, PAL/NTSC

Outputs:
Video output: DVI-D x 1
S-Video x 1: (Outputs last image(s))

Data I/O:
USB 2.0: Front: Type 1 x 1, Rear: Type 1 x 4
Ethernet: Gigabit RJ45 x 1
Keyboard/Mouse I/O: USB
Operating System: Windows 7 Embedded
GPIO: RJ-11

Power Input:
100 - 240 VAC, 50/60 Hz

Environmental (Operating):
Temperature: 0 to 40º C
Relative Humidity: 0 to 85%
Atmospheric: 700 to 1060hPa

Environmental (Storage):
Temperature: -40 to +70º C
Relative Humidity: 10 to 100%
Atmospheric: 500 to 1060hPa

PG043 U5010-12
TECHNICAL (CONTINUED)

SYMBOL GLOSSARY
Symbol Reference(s) Title Symbol Reference(s) Title
IEC 60417:5009 Stand by IEC 60417:5031 Direct current

IEC 60417:5021 Equipotentiality EN 980:4.11 Manufacturer

47 CFR 15 FCC compliance D i r e c t i v e 9 3 /4 2 / CE compliance


EEC

EN 50419:2005 “New” waste ISO 7000:0434 Caution, consult


Directive 2002/96/ ISO 15223:3.4 accompanying
EC (WEEE) EN 980:4.9 documents

ISO 7010:M002 Refer to instruction IEC 60601-1:2005 TÜV certification


manual/booklet IEC 60601-1-2:2014

ISO 15223:3.16 Serial number ISO 15223:3.15 Catalog number


EN 980:4.4 EN 980:4.8

ISO 7000:2497 Date of EN 980:4.12 Authorized


ISO 15223:3.13 manufacture representative
EN 980:4.5 of the European
community
ISO 7000:1641 Consult
ISO 15223:3.3 instructions for use
EN 980:4.17

U5010-12 PG044
SAFTEY/ELECTRICAL/MAINTENANCE

TECHNICAL CONNECTING COMPUTER TO POWER


The DRSHD 1080p computer has no internal serviceable The DRSHD 1080p system comes with 1 HOSPITAL
parts. If for any reason the DRSHD 1080p computer fails GR ADE power cord which is indicated by the Green
please contact customer service at (888) 982-4469 or dot on the male end of the cord for connecting the
Email support@medxchange.com computer to the electrical outlet. In order to achieve
proper grounding of the DRSHD 1080p system, use the
DRSHD 1080P COMPUTER provided HOSPITAL GRADE power cord or equivalent.
Warning: Disconnect power supply before servicing.
The computer system and its internal components meet Warning: A multiple portable socket-outlet or extension
the requirements for applicable ETL, FCC, CE, CB, CIB cord should not be connected to the system.
and other required certifications.
Warning: Do not connect items that have not been
The plastic material used for the front of the DRSHD specified as par t of the DRSHD 1080p computer or
10 8 0 p P C e n c l o s u r e h a s a U L 9 4 f l a m m a b i l i t y specified as being compatible with the DRSHD 1080p.
classification of 94H-2.
Note: A multiple port power strip should not be used to
A warm damp cloth can be used to wipe the outside of connect the system because it can increase the over all
the unit. leakage for the system.

Warning: Do not touch the DRSHD 1080p computer and CLASSIFICATIONS


patient simultaneously. The DRSHD 1080p is classified as a Class I Device:
The product uses a grounded plug.
The DRSHD 1080p is not to be used within 1½ meters There are no patient applied parts.
of patient’s chair or table in the patient environment. Does not protect against ingress of liquids.
Not to be installed near an MRI or within any other
CMOS battery: magnetic field.
BR 1225 *Note for California: Perchlorate This system can be continuously used.
Material - special handling may apply, See www.dtsc.
ca.gov/hazadouswaste/perchlorate

PG045 U5010-12
SAFTEY/ELECTRICAL/MAINTENANCE (CONTINUED)

NOTES
Warning: There are no known potential electromagnetic Med X Change, Inc will not be held responsible for any
or other interference between the DRSHD 1080p system personal/internet use. The DRSHD 1080p is to be used
and other devices. If the system does cause harmful for Digital Recording only. The warranty will be null and
interference, the user is encouraged to correct the void, in response to any repairs that may be needed, if
interference by one of the following methods: personal/internet use is apparent.

1. Relocate the DRSHD 1080p or the device that is WARRANTY


receiving the interference. Med X Change, Inc. will provide a warranty against
2. Increase the separation between the DRSHD 1080p manufacturer’s defect of all DRSHD 1080p hardware,
and the device that is receiving the interference. which will ex tend for a period of 12 months from
3. Connect the DRSHD 1080p into an outlet on a circuit manufacturer date; covering parts and labor only. During
different from that to which the device is connected. the respective warranty periods Med X Change, Inc. will
repair or replace defective parts and maintain software
There are preventative measures that can be taken applications to original specifications. Claims under
to ensure proper functionality of the DRSHD 1080p warranty are null and void if the defect has been caused
system. The system should be shutdown and restarted by improper handling/use, or if the DRSHD 1080p at
periodically (at least every other day). A proper shutdown issue has been serviced by unauthorized persons or
sequence would be to press the power button on the product/service providers. Med X Change, Inc. warrants
front of the DRSHD 1080p one time and release. Holding that the DRSHD 1080p is free of defects in material and
the power button down until power is off should only be manufacture at the time of delivery and that the DRSHD
done when directed by Med X Change Technical Support. 1080p meets the specifications.

T he fuses use d in the DR S HD 10 8 0 p syste m are


nonreplaceable by the user. If there is a problem with a
fuse, please contact customer service at (888) 982-4469
or Email support@medxchange.com to set up repair.

There are no repairable component parts in the DRSHD


1080p system. Any system that needs repair should be
returned to the manufacturer for such repair.

This equipment is not suitable for use in the presence


of flammable anesthetic mixture or with nitrous oxide.
Med X Change, Inc cannot be held responsible for any
data loss in accordance to the DRSHD 1080p.

Med X Change, Inc. strongly recommends the client burn


all videos immediately to a USB drive or use the archiving
feature in accordance to the DRSHD 1080p User Manual.

Med X Change, Inc. strongly recommends you purchase


and hook up a battery back up system as well as a surge
protector (ex. 1100 VA surge protector), in accordance
to the DRSHD 1080p User Manual.

U5010-12 PG046
SYSTEM COMPLIANCE STATEMENTS

SYSTEM COMPLIANCE STATEMENTS & TABLES


Electrostatic Discharge Precaution If the system is installed near or on other electronic
When using USB devices do not touch the USB ports. devices, please observe to verify normal operation. Other
Contacting the ports may create electrostatic discharge devices may emit harmful radiation or interference that
(ESD) which may cause the system to freeze and prevent could affect system performance.
the USB peripherals from functioning properly. Before
restarting the computer, attempt to restore functionality Use of accessories and cables other than those provided
by unplugging the USB device and re-plugging it back into with the system could result in increased emissions or
the port. Wait approximately 10 seconds to determine decreased immunity of the equipment and may cause
if functionality has been restored. If not, restar t the the system to be non-compliant with the requirements
computer. of IEC 60601-1-2:2007.

Precautions This may be used with an optional Wi-Fi ZyXEL router/


This equipment/system is intended for use by healthcare IEEE 802.11B/G/N 2.412 GHz to 2.462GHz transmitter
professionals only. T his equipment /system may operating at up to 250 mW. Fur ther details can be
cause radio interference or may disrupt the operation accessed through the FCC using the following FCC ID:
of nearby equipment. It may be necessar y to take ZV Y-FW152A. The received signal may be interfered/
mitigation measures, such as reorienting or relocating disrupted by other equipment operating in this frequency
the equipment or shielding the location. band (i.e. Bluetooth, other non associated access points,
microwave ovens, ISM devices, etc) even if the other
The system must be installed per the requirements of equipment complies with its applicable regulations such
this manual. Failure to properly install the system may as CISPR 11.
result in degraded performance or loss of functionality.
Mobile and portable RF communications equipment
may affect the system.

IEC 60601-1-2:2007 TABLE 1 REQUIREMENTS


The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 Except for the intentional data transmission at 2.412 GHz to 2.462 GHz,
this equipment uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment. This device may cause interference,
or be disrupted by other devices operating in the 2.412 GHz to 2.462
GHz spectrum.
RF emissions CISPR 11 Class A The equipment is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those
directly connected to the public low-voltage power supply network that
Harmonic emissions Class A supplies buildings used for domestic purposes, provided the following
IEC 61000-3-2 warning is heeded:
Warning: This equipment/system is intended for use by healthcare
Voltage fluctuations/ Complies
professionals only. This equipment/system may cause radio interference
flicker emissions
or may disrupt the operation of nearby equipment. It may be necessary
IEC 61000-3-3
to take mitigation measures, such as re -orienting or relocating the
equipment or shielding the location.

PG047 U5010-12
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)

IEC 60601-1-2:2007 TABLE 2 REQUIREMENTS


The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
Equipment should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment
– guidance
Electrostatic discharge Level three ±6 kV contact Level three ±6 kV contact Floors should be wood, concrete or ceramic
(ESD) IEC 61000-4-2 discharge ±8 kV air discharge ±8 kV air tile. If floors are covered with synthetic
discharge discharge material, the relative humidity should be at
least 30%.
Electrical fast transient/ ± 2 kV for power supply ± 2 kV for power supply Mains power quality should be that of a
burst IEC 61000-4-4 l i n e s ±1 k V f o r i n p u t / l i n e s ±1 k V f o r i n p u t / typical commercial or hospital environment.
output lines output lines
Surge IEC 61000-4-5 ±1 kV line(s) to line(s) ±2 ±1 kV line(s) to line(s) ±2 Mains power quality should be that of a
kV line(s) to earth kV line(s) to earth typical commercial or hospital environment.
Voltage dips, shor t <5% UT (>95% dip in UT) < 5 % U T ( > 9 5 % d i p i n M a i n s p o w e r q u a l i t y s h o u l d b e t h a t
interruptions and voltage for 0,5 cycle 40% UT (60% U T ) fo r 0 , 5 c y c l e 4 0 % of a t ypical commercial or hospital
v a r i a t i o n s o n p o w e r dip in U T ) for 5 cycles UT (60% dip in UT) for 5 environment. If the user of the equipment
su p p l y i n p u t li n e s I E C 70% UT (30% dip in UT ) cycles 70% UT (30% dip requires continued operation during power
61000-4-11 UT = 230 Vac for 25 cycles <5% U T in UT ) for 25 cycles No mains interruptions, it is recommended
(>95% dip in UT) for 5 sec anomalies 95% dip meets that the equipment be powered from an
requirements. uninterruptible power supply or a battery.
Power frequency (50 3 A/m 3 A/m Power frequency magnetic fields should
Hz) magnetic f ield IEC b e at levels character istic of a t y pic al
61000-4-8 location in a typical commercial or hospital
environment.

U5010-12 PG048
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)

IEC 60601-1-2:2007 TABLE 6 REQUIREMENTS


The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
Equipment should assure that it is used in such an environment.
Immunity IEC 60601 Compliance Electromagnetic environment – guidance
test test level level
Conducted 3 Vrms 150 3V Portable and mobile RF communications equipment should be used no closer to
RF IEC kHz to 80 any part of the equipment including cables, than the recommended separation
61000-4-6 MHz 3 V/m distance calculated from the equation applicable to the frequency of the transmitter.

Radiated 3 V/m 80 Recommended separation distance


RF IEC MHz to 2,5 3,5
61000-4-3 GHz V1
3,5
80 MHz to 800 MHz
E1
7
800 MHz to 2,5 GHz
E1
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF transmitters, as determined
by an electromagnetic site surveys should be less than the compliance level in
each frequency range.

Interference may occur in the vicinity of known RF transmitting devices and


equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and T V broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above,
the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the equipment
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

PG049 U5010-12
SYSTEM COMPLIANCE STATEMENTS (CONTINUED)

IEC 60601-1-2:2007 TABLE 6 REQUIREMENTS


Recommended separation distances between portable and mobile RF communications equipment and the DRSHD 1080p.
The Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Equipment can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Equipment as recommended below,
according to the maximum output power of the communications equipment.
3,5
Rated maximum output power Separation distance according to frequency of transmitter M V1
of transmitter W
150 kHz to 80 MHz 80 MHz3,5
to 800 MHz 3,5to 2.5 GHz
800 MHz
80 MHz to 800 MHz
V1 E1
3,5 3,5 7
80 MHz to 800 MHz 800 MHz to 2,5 GHz
V1 E1 E1
0.01 0.12 0.12 0.24
3,5 7
0.1 0.37 80 MHz to 800 MHz
0.37 800 MHz to 2,50.74
GHz
E1 E1
1 1.17 1.17 2.34
7
10 3.69 800 MHz to 2,5 3.69
GHz 7.38
E1

ADDITIONAL
Special skills, training, or knowledge: Operator should In the United States, some states have
be familiar with computer based equipment. laws regarding the disposal of electronic
w a s t e ( s e e h t t p : // w w w . e l e c t r o n i c s r e c y c l i n g .
Start up procedure: org/public/ContentPage.aspx?pageid=14).
Power up the DRSHD 1080p. Upon booting up, the
DRSHD 1080p will be ready to operate. Personal patient infor mation protected by HIPA A
i s s to r e d o n t h e D R S H D 10 8 0 p h a r d d r i v e . I t i s
The DRSHD 1080p utilizes the Microsoft® Windows 7 highly recommended to clean the hard drive
Embedded operating system software. In the event of b e f o r e d i s p o s i n g ( h t t p : // o n g u a r d o n l i n e . g o v /
an operating system error, Windows system error codes articles/0010-disposing-oldcomputers).
can be referenced at: http://msdn.microsoft.com/en-us/
library/windows/desktop/ms681381(v=vs.85).aspx Warning: No modification of this equipment is allowed.
Failure to comply could result in electric shock or death.
Shutdown Procedure:
Press the power button on the front of the DRSHD 1080p Keep combustible materials away from DRSHD 1080p
in and release. while in use.

Proper disposal of the DRSHD 1080p at the end of its


service life:
The DRSHD 1080p is an electronic device that should be
disposed of in accordance the local laws of the region
where it is being disposed. Even when local laws do
not regulate the disposal of electronic equipment, Med
X Change, Inc. recommends contacting an electronic
waste recycler to dispose of the DRSHD 1080p.

U5010-12 PG050
CONTACT INFORMATION

TECHNICAL ASSISTANCE OR INJURIES AND/OR


MALFUNCTION, COMPLAINTS
If you need technical assistance regarding the use C onMed C orporation will issue a Return Material
or application of this product, or to report any events Authorization (RMA). All units are to be shipped to
involving injuries, malfunctions or a complaint, contact ConMed Corporation with the RMA number noted on
ConMed Corporation Customer Service or your ConMed the shipping memo.
Corporation S ales Representative and you will be
transferred to Med X Change, Inc. for handling. If Product Return Address: Manufacturer’s Address:
Outside U.S., contact your local ConMed Corporation ConMed Corporation Med X Change, Inc.
Representative via Email. 11155 Concept Boulevard 401 8th Street West
Largo, FL 33773-4908, USA Bradenton, FL 34205, USA
PRODUCT RETURN FOR SERVICE AND/OR
INVESTIGATION OPEN SERVICE REQUEST STATUS
If you encounter a problem that requires servicing or If you need to contact ConMed Corporation regarding an
repair, contact ConMed Corporation Customer Service open Service Request, contact servicesupport@conmed.
or your ConMed Corporation Sales Representative. com or number/fax listed below.
Outside the U.S. contact your local ConMed Corporation
Representative.
PRIOR TO SENDING THE DRSHD 1080P IN FOR
SERVICE; PLEASE BACKUP VIDEO AND STILL
Returning products for any reason requires an authorized
IMAGES TO EXTERNAL STORAGE.
Service Request (S.R.) number prominently displayed
on the box and included on all paperwork. Refer to this
MED X CHANGE, INC. AND CONMED
number if making inquiries about the repair status.
CORPORATION ARE NOT LIABLE FOR DATA
Please call ConMed Corporation Customer Service (1-
LOSS ON THE DRSHD 1080P.
800-237-0169) and provide the following information
to obtain an S.R. number prior to returning any product
for repair: AUTHORIZED REPRESENTATIVE EUROPE

1. Product Number RMS - UK, Limited


2. Original Invoice Number 28 Trinity Road
3. Date of Purchase Nailsea, Somerset
4. Serial/Lot Number BS48 4NU England
5. Detailed description of the problem
6. Reason for Return

Contact Within US (Phone) Within US (Fax) Email

Primary Customer Service, (800) 237-0169 (727) 399-5256 Inside US:


Technical Assistance or customer_service@conmed.com
Injuries and/or Malfunction,
Outside US:
Complaints, Product Returns internationalrepairs@conmed.com
for Service and/or Investigation
Open Service Request Status (800) 325-5900 (727) 399-5269 servicesupport@conmed.com
ext 6031
Secondary Med X Change, Inc. (941) 794-9977 (941) 794-8432 support@medxchange.com
Technical Support
(888) 982-4469
(24/7 Toll Free)

PG051 U5010-12