Infrared Temperature Measurement System

Modline 3 - the most advanced micro computer based


non-contact temperature measurement system,
featuring high performance, reliability and versatility.

The Modline 3 is a non-contact, infrared radiation thermo-


meter designed to meet the critical needs of today’s appli-
cation requirements. Modline 3 features a broad line of
infrared thermometers to match specific applications in
many industries. The system offers a choice of sensors
with sophisticated single lens reflex (SLR) focusing that
can view spots as small as 0.012 inches (0.3mm), or
sensors using two-color (ratio) technology to tolerate more
than 95% reduction in radiant intensity, or fiber optic
sensors for difficult to reach or obstructed targets.
Standard Features (cont’d)
• Track and Hold
The system consists of a high performance sensor and
Hold instrument readings/outputs or allow the instrument to
micro-computer based digital Indicator/ Processor. Many
track real events, all via a simple contact closure from a
standard features are included, only a two-point On/Off
remote switch or a relay.
controller/alarm or a PID controller are optional. Addi-
• Analog Outputs
tionally, Ircon offers standard accessories to meet the
Select one from four available analog outputs:
demanding application requirements, such as air purge
0 to 20mA, 4 to 20 mA, 5µA / °F or °C or 0 to 10 Vdc
and water cooling, sight tubes and mounting assemblies.
• Remote Analog Input
Modline 3 utilizes advanced micro-computer technology 4 to 20mA signal input to the Modline 3 permits external
to process temperature signals received from the sensor. adjustment of Emissivity/E-Slope or Controller Setpoint.
This data is accurately displayed on a highly visible LED
• RS-485 Digital Interface
digital display in °F or °C. Programming features and
control functions are viewed on a bright, highly readable Digital communication port provides a complete two-way
Alpha Numeric, Vacuum Fluorescent display. access to Modline 3 features and functions.
• CE Compliance
Standard Features Meets EMC directive 89/366/EEC
• High Performance Sensors Meets Low Voltage directive 73/23/EEC
Select from 9 narrow spectral regions and temperature
Optional Features
ranges to match specific applications
• 2- Point On/Off Control Options
• Micro-computer Technology Two completely independent channels of control and/or
Indicator/Processor is programmed to match specific alarming.
sensor requirements and control functions using front
• 3-Mode Proportional Controller
panel buttons
With a multitude of features; Auto tune, Bumpless Output
• High System Accuracy Transfer, 4 to 20 mA isolated output, Internal/External
± 0.6% of reading ± 1 digit, or ± 0.6% of F.S. ± 1 digit, Setpoint control and Hi-Lo Deviation Alarms with relay
depending on model; ± 0.1% Repeatability F.S. outputs.
• Smart Peak Picker
With Decay rate controlled directly in degrees per
second. Manual and Automatic Reset modes.
Modline 3 Catalog

System Overview

E-SLOPE OR
CONTROLLER
SET POINT

Modline 3 System Configuration for SLR and Fiber Optic Versions

Sensor Selection Chart

Series Temp Span Wavelength Applications


200 900 - 4800°F or 500 - 2600°C 0.7 - 1.0µ Forging, annealing, hardening, foundries,
200 Fiber Optic 1200 - 4800°F or 650 - 2600°C 0.7 - 1.0µ most iron and steel, incandescent processes
including semiconductor processes.

340 75 - 1500°F or 25 - 800°C 3.43µ Thin films of polyethylene, polypropylene,


vinyls and nylons.

600 150 - 1400°F or 80 - 800°C 2.0 - 2.6µ Medium temperature applications for thick
plastics, rubber, textiles and metals.

700 100 - 2500°F or 50 - 2500°C 4.8 - 5.3µ Glass surface temperature in forming, bending.
tempering, annealing and sealing.

800 75 - 800°F or 25 - 400°C 7.92µ Thin films of polyester (PET) and fluorocarbon
plastics.

3G 500 - 2500°F or 250 - 1400°C 1.5 - 1.6µ Medium temperature applications, particularly
3G Fiber Optics 650 - 2500°F or 350 - 1400°C 1.5 - 1.6µ ferrous and non-ferrous metals.

3L 500 - 1800°F or 250 - 1000°C 1.00 - 1.20µ Low temperature non-ferrous metals including
or 1.5 - 1.6µ, aluminum using two-color ratio technique.
and 1.65 - 1.71µ

3R 1300 - 6500°F or 700 - 3500°C 0.7 - 1.08 and 1.08µ For difficult high temperature processes such
3R Fiber Optic 1300 - 6500°F or 700 - 3500°C 0.7 - 1.08 and 1.08µ as molten metal, wire rod, vacuum furnaces
and kilns using two-color ratio technique.

3V 400 - 1200°C 0.91 - 0.97µ Specifically designed for gallium arsenide


3V Fiber Optic 500 - 1500°C 0.91 - 0.97µ wafer temperature in MBE, MOCVD and
MOMBE applications.

2
Modline 3 Catalog

Sensors
The Modline 3 system offers you the widest
choice of non-contact infrared sensors to
match any process. Modline 3 sensors
offer sophisticated single lens reflex (SLR)
focusing, that can view spots as small as
0.012 inches (0.3 mm); or for difficult, high
temperature and harsh environments,
some model series incorporate fiber optics
for added flexibility.
The optical system, detector and electron-
ics are housed in a rugged sealed casting Series 340, 700, 800, 3R & 3L with SLR focusing Series 200, 600, 3V and 3G with SLR focusing
rated NEMA 4 (IP66) to perform in the
harsh industrial environments. Modline 3
sensors are designed to match existing
mounting configurations and offer a choice
of cooling and environmental protection
accessories.

200 Series
Measures temperatures ranging from 900
to 4800°F (500 to 2600°C) and is intended
for high temperature incandescent appli- 3G Series The series also uses Ircon’s two-color
cations. It operates in a narrow spectral Medium temperature thermometer mea- technique to measure difficult high
region from 0.7 to 1.0µm, utilizing an sures temperatures ranging from 500 to temperature applications. It is ideal for
extremely stable silicon detector. 2500°F (250 to 1400°C) and operates at processes such as molten metal, wire
1.5 to 1.6µm. It utilizes an InGaAs detector and rod, vacuum furnaces and kilns.
340 Series for exceptional stability, wider temperature
Measures temperatures ranging from 75 spans and ability to operate at higher ambi- 3V Series
to 1500°F (25 to 800°C) and is ideal for ent temperatures. This series is ideal for Gallium Arsenide thermometer measures
clear films of C-H type plastics or paints, medium temperature applications such as temperatures ranging from 400 to 1200°C
waxes and oils. It operates in a very nar- ferrous and non-ferrous metals. and operates in a narrow spectral region
row wavelength band centered at from 0.91 to 0.97µm. This series is speci-
3.43µm. 3L Series fically tailored for measuring gallium
Low temperature two-color thermometer arsenide (GaAs) wafer temperatures.
600 Series measures temperatures ranging from 500
Measures temperatures ranging from to 1800°F (250 to 1000°C) and operates at Fiber Optic Sensor Option
150 to 1400°F (80 to 800°C) and operates 1.55 and 1.68µm lower ranges and 1.10 Our fiber optic sensors are the ultimate
at 2.0 to 2.6µm. It’s ideal for general pur- and 1.68µm upper ranges. It measures in non-contact temperature measurement
pose medium temperature applications temperatures from the ratio of radiation for difficult-to-read or obstructed targets.
such as thick plastics, rubber, textiles signals of two adjacent wavelengths and The following series are available as a
and metals. not from absolute intensity. It will tolerate a fiber optic version:
more than 95% reduction in radiant intensity
700 Series Series and Temperature Range
with virtually no error. This series is most 200 200 to 4800°F (650 to 2600°C)
Measures temperatures ranging from
appropriate for low temperature non-ferrous 3G 650 to 2500°F (350 to 1400°C)
100 to 2500°F (50 to 2500°C) and oper-
metal applications including aluminum. It 3R 1300 to 6500°F (700 to 3500°C)
ates at 4.8 to 5.3µm. It is ideal for glass
features two Indium Gallium Arsenide 3V 500 to 1500°C
surface temperatures in forming, bending,
(InGaAs) detectors which measure target
tempering, annealing and sealing.
signals simultaneously delivering highly sta-
800 Series ble, highly accurate and fast measurements.
Measures temperatures ranging from
3R Series
75 to 800°F (25 to 400°C) and operates
High temperature two-color thermometer
at 7.92µm. This series measures thin film
measures temperatures ranging from 1300
of all polyester (PET) and fluorocarbon
to 6500°F (700 to 3500°C) and operates at
plastics including thin glass and ceramics.
0.7 to 1.08 µm and 1.08 µm.

3
Modline 3 Catalog

Standard Features Optional Features


The Modline 3 Indicator/Processor is a micro computer On/Off Control
based signal processor designed to operate with all Modline The On/Off controller provides two independent set-
3 sensors, an internal menu allows the user to indicate the points for alarms or control. Each setpoint is adjustable
sensor connected and the micro-computer provides the over the entire span of the instrument. Setpoint 1 can
functions for that specific sensor. No calibration is necessary be externally set via the 4-20 mA linear input.
when changing sensors.
PID Control
Analog Outputs The Modline 3 PID Controller provides a fully automat-
Four selectable isolated and linear outputs are available; ed proportional controller for controlling heaters, lamps
4-20 mA, 0-20 mA, 5µA per degree, or 0 to 10 Vdc. The out- or speed controls. The controller can be operated in a
puts are adjustable to cover any portion of the instruments manual mode using front panel controls or a remote
available temperature span to a minimum of 10 degrees. setpoint via a 4-20 mA input. The PID control features
Auto-tune, Bumpless Transfer and two Deviation
Digital Interface Alarms which can be independently set either side of
RS-485 digital communication interface communicates in the operating setpoint.
half duplex to the computer. Command signals are used to
make temperature request, setting adjustments, system sta- Modline 3 Functions
tus, system configuration, on/off control functions, service All functions, either standard or optional, are easily set
functions and PID control functions. using the four controls on the front panel. Easily under-
stood menus help you to make quick selections for fast
Auxiliary input start up and operation.
A remote 4-20 mA input can be field-selected to set the
emissivity, the E-Slope or controller setpoint.

Smart Peak Picker


The Pick Picker circuitry responds to the highest instanta-
neous value of temperature and holds this value even if the
temperature source is interrupted by smoke, steam, dust or
intermittent targets. A rate of decay adjustment allows the
instrument to follow the true target temperature but not be
affected by the conditions previously mentioned.

The Modline 3 Peak Picker is a smart peak picker. When


viewing products that have a hot leading edge like a piece
of glass, the peaking function can be delayed for as long as
10 seconds to avoid this peak temperature. In addition, it
can be reset manually or a reset setpoint can be selected.
Should the temperature decay below this point, the Peak
Picker will automatically reset. This setpoint is adjustable
anywhere within the range of the instrument.

Track and Hold


Many processes may require that a control output be held
even if the hot target is not in sight. For example, when a
billet exits an induction coil, a temperature is obtained - if no
billet is present, the controller would call for heat because
it senses a cool target.

With a track and hold feature, the instrument receives a Typical Dimension Drawing shows panel mounting for
contact closure and holds the last seen temperature of the Modline 3 Indicator/Processor
billet. All functions are held (the indicator, linear outputs
and controller outputs) until the contact is opened.

4
Modline 3

Operation Features
Front Panel Displays and Controls Rear Panel Connections

4 Digit LED Display indicates temperature calibration in °F


or°C and out of temperature range alarm. Analog Output Connection
0 to 20 mAdc; 4 to 20 mAdc, PID Controller
5µA per/degree or 0 to 10 Vdc Output (optional)

Alarm Output for


On/Off Controller
(optional)

“M” Button
allows you Power Supply
to scroll Connections for
through 90 to 250 Vac
menu
selections
Invalid Alarm
for 3L and 3R
series only
“F” Button
provides
control Up-Down Buttons Sensor Input Remote Reset for Peak
function. allow you to change Picker and Track /Hold
Auxiliary Analog Signal Input
Function Display indicates control values to adjust remotely Emissivity, RS-485 Digital Interface
function settings and menus in a E-Slope or Set Point-1 For digital communication
2 x 20 line character display.

Control Function Display (In English, French or German) Optional Accessories


Menus and functions are easily selected and displayed Ircon provides a variety of optional accessories for the
using two buttons, variable values are changed using up Modline 3 System for the SLR sensors or the Fiber Optic
and down buttons. Version to simplify mounting and to protect the sensors
from hostile environments.

SLR Sensors
• Model AA-3 Air Purge
• Models SB-1 Swivel Mounting Base
• Model WA-3 Water Cooling Accessory
• Model WJ-5 Water Cooling Jacket
Select Emissivity and Value 0.100 to 1.000
• All other Flange Mount Accessories and Sight
Tubes.

Change Value using up/down buttons

Other Standard Function Selections


• E-Slope: 0.850 to 1.150
• Response Time: 0.01 to 60 sec.
• Peak Picker: On/Off, Decay Rate, Reset, Remote
Reset, Auto Reset, Reset Below, Delay Fiber Optic Sensors
• Track and Hold: On/Off • Model AA-5 Air Purge
• Security Access: Panel Locked/Closed/Open • Model AP-5 Adapter Plate
• Calibration: Cycle Cal. • Model MB-5 Angle Mounting Bracket
• System ConfigurationModel Series: Text Language • Model SB-5 Swivel Mounting Base with
(English, French, German, Japanese) MC-5 Mounting Clamp

5
Modline 3 Catalog

System Specifications
Spectral Response Operating Ambient Select. Analog Outputs (cont’d)
CE Directives: Each series is designed to operate (without ambient cooling) 0 to 20 mAdc with a 1000 W max.
in a discrete wavelength band (see Series 200 and 3 G: 4 to 20 mAdc with a 1000 W max
Meets EMC Directive Model Selection Chart.) 32 to 150°F (0 to 60°C) 5 mAdc / degree F or C, (1mV/ °
89/336/EEC for EMI/RFI Series 3L and 3R: 0 to 10 Vdc
and Safety Compliance Calibration / Accuracy 32 to 130°F (0 to 55°C)
tested to: Accuracy @25°C: Series 3V: 50 to 113°F (10 to 45°C) [max 4000 mV] across 200 W
EN 50081-2: Generic 200, 3G, 3R, 3V series within Series 340, 600, 700, and 800: resistor, typical deviation two degrees
Emissions Standard 0.6% of reading (+ 1 digit) or 5°F 50 to 130°F from display)
EN 50082-2: Generic (3°C), whichever is greater. Sensor Cable Silicone Rubber:
Immunity Standard 400°F (200°C) Max. Temp. Digital Communication
EN 61010-1: Safety Standard 340, 600, 700, 800 and 3L series Indicator / Processor: RS-485 Digital Interface
within 0.6% F.S. (+ 1 digit) or 32 to 122°F (0 to 50°C)
5°F (3°C) whichever is greater. Emissivity Range*
Humidity 0.100 to 1.000.
RepeatabilitY @25°C 10 to 90% non-condensing *see pages 7 & 8 for limitations.
Within 0.1% of full-scale temperature
(+1 digit) Enclosure E Slope Range*
allows mounting two Indicator/ 0.850 to 1.150.
Response Time at Analog Output Processors side-by-side in 19 *see pages 7 & 8 for limitations.
Adjustable from 0.01 to 60 seconds inch rack; Size: 7.56 x 5.67” (192
x 144 mm) Power Requirements
90 to 250 Vac, 50/60 Hz, 40VA max.
NEMA 4 (IP66 rated) Selectable Analog Outputs
Sensor Housing and Indicator/ are isolated from ground Weight
processor enclosure Sensor: 7 to 8 lbs (3.2 to 3.6 kg)
Indicator/Processor:
9 lbs (4 kg) - approximate weight

Unique to Sensor Series 3L 3L Spectral Response Signal Reduction Range Invalid Alarm
& 3R Two-Color (Ratio) Adjacent bands: 1.50 to 1.60µ Caused by low emissivity, non- Form C relay outputs rated at
Thermometer and 1.65 to 1.71m or 1.00 to resolved or obliterated target 24 VAC/DC;@ 1 Amp.
1.20µ and 1.65 to 1.71µ. above 575°F (300°C) for 3L Resistive or Inductive.
series and above 1500°F (800°C)
3R Spectral Response for 3R Series. Will tolerate about
Adjacent bands: 0.7 to 1.08µ and 95% reduction in radiation intensity.
a narrow band centered at 1.08µ.

Peak Picker Rise Time Adjustable Decay Rate Peak Picker Reset
Same as selected Response 0.00 to 300°F (0.00 to 300°C) per Internal Reset
Time (0.01 to 60 seconds). second, depending on the instrument two Modes: Manual and Auto
model. External Resetvia contact closure.

On/Off Controller/Alarm with Response Time Remote Set Point 1 Relay Delay Time
Relay Outputs Same as selected Response Time Input Signal of 4 to 20 mA, daisy 1 to 2 milliseconds
(0.01 to 60 seconds). chain compatible.
Hysteresis
Internal Set Point Relay Outputs 2°F (2°C)
Adjustable over full temperature span Two Form C relay outputs rated at
of the instrument model. 24 V AC/DC; @ 1 Amp. Resistive
or Inductive.

PID Controller with Auto Pid Output Proportional Band (P) Manual Adjust
Tune and Deviation Alarms 4 to 20 mAdc isolated. 0.1 to 200.0% 0.0 to 100.0% of output
1000 Ohm maximum.
Reset Rate Deviation Alarm Set Points
Remote Set Point (I) - 0.01 to 99.00 repeats per minute Adjustable for HI and LO set points
Input Signal 4 to 20 mA, daisy with an Off position. over full temperature span of instru-
chain compatible. ment model, each with Off position.
Rate Time
(D) - 0.001 to 9.900 minutes, with Deviation Alarm Output
Internal Set Point Relay Outputs for HI and LO
Off position.
Adjustable over full temperature Alarms. Use two Form C relay
span of the instrument model. Load Demand outputs rated at 24 V AC/DC; @
0 to 100% of output 1Amp. Resistive or Inductive.

6
Modline 3 Catalog

Model Identification Chart


A B C D E F

Model Number Explanation Example: 33-05F05-0-0-0-0

BLOCK A 3 3 = Modline 3 Series 340 BLOCK D 0 = Analog Signal Output - 4 to 20mA,


0 to 20mA, and 5mA per degree
BLOCK B 0 5 F 0 5 = 150 to 500°F Temperature Range
D/50 Resolution BLOCK E 0 = Digital Communication
BLOCK C 0 = Signal Conditioning RS-485 Digital Interface
Peak Picker and Track and Hold BLOCK F 0 = No Controller Output

BLOCK A - Series Designation BLOCK B (cont.) BLOCK B (cont.)


32 = 200 Series
33 = 340 Series 200 Series Fiber Optic Options (cont.) 600 Series SLR Version (cont’d)
36 = 600 Series 10 ft (3 m) fiber optic cable 02C15 = 80 – 200°C D/150*
37 = 700 Series 15CF5 = 650 – 1500°C D/30 03C15 = 120 – 300°C D/150*
38 = 800 Series 15CF8 = 650 – 1500°C Extension Tip 04C15 = 160 – 400°C D/150
3G = 3G Series 20CF6 = 700 – 2000°C D/60 06C15 = 250 – 600°C D/150
3L = 3L Series 20CF7 = 700 – 2000°C D/30 x D/150 08C15 = 350 – 800°C D/150
3R = 3R Series 20CF8 = 700 – 2000°C Extension Tip
3V = 3V Series 26CF6 = 750 – 2600°C D/60 * Limited range of response time adjustment
26CF7 = 750 – 2600°C D/30 x D/150 is 0.10 to 60 seconds.
26CF8 = 750 – 2600°C Extension Tip
BLOCK B - Temperature Range, Optical
Characteristics and Fiber Optic Options ** Emissivity span is restricted to 0.2 to 1.0 for the 700 Series – 4.8 to 5.3µ
(Check One) first 100°F (55°C) for all temperature ranges. 06F05 = 100 – 600°F D/50*
Note: Some sensor models are marked with response 10F05 = 200 – 1000°F D/50**
time restrictions. All other models have an adjustable 15F10 = 500 – 1500°F D/100
range of 0.01 to 60 seconds. 340 Series – 3.43 µ (0.14µ Band) 25F10 = 500 – 2500°F D/100
04F05 = 75 – 400°F D/50*
05F05 = 150 – 500°F D/50** 03C05 = 50 – 300°C D/50*
200 Series* – 0.7 to 1.0 µ 06F05 = 200 – 600°F D/50 06C05 = 100 – 600°C D/50**
20F05 = 900 – 2000°F D/50 10F10 = 300 – 1000°F D/100 08C10 = 300 – 800°C D/100
24F10 = 1000 – 2400°F D/100 15F10 = 500 – 1500°F D/100 13C10 = 300 – 1300°C D/100
30F20 = 1200 – 3000°F D/200 25C10 =1000 – 2500°C D/100
02C05 = 25 – 200°C D/50*
48F30 = 1400 – 4800°F D/300
03C05 = 75 – 300°C D/50** * Limited range of response time adjustment
11C05 = 500 – 1100°C D/50 04C05 = 100 – 400°C D/50 s 1.0 to 60 seconds.
13C10 = 550 – 1300°C D/100 05C10 = 150 – 500°C D/100
08C10 = 200 – 800°C D/100 ** Limited range of response time adjustment
14C10 = 600 – 1400°C D/100 is 0.10 to 60 seconds.
15C20 = 650 – 1500°C D/200 * Limited range of response time adjustment
16C30 = 800 – 1600°C D/300 is 2.00 to 60 seconds.
26C30 = 750 – 2600°C D/300 800 Series* – 7.92µ (0.3µ Band)
** Limited range of response time adjustment 08F02 = 75 – 800°F D/20
* Emissivity span is restricted to 0.3 to 1.0 for the is 0.10 to 60 seconds.
first 100°F (55°C) for all temperature ranges. 04C02 = 25 – 400°C D/20

600 Series SLR Version – 2.0 to 2.6µ * Limited range of response time adjustment
200 Series Fiber Optic Options** 04F15 = 150 – 400°F D/150* is 0.25 to 60 seconds.
10 ft (3 m) fiber optic cable 05F15 = 200 – 500°F D/150*
06F15 = 250 – 600°F D/150 Note: Model Selection Chart continued
25FF5 = 1200 – 2500°F D/30
on following page.
25FF8 = 1200 – 2500°F Extension Tip 08F15 = 350 – 800°F D/150
10F15 = 500 – 1000°F D/150
30FF6 = 1300 – 3000°F D/60
14F15 = 600 – 1400°F D/150
30FF7 = 1300 – 3000°F D/30 x D/150
30FF8 = 1300 – 3000°F Extension Tip
48FF6 = 1800 – 4800°F D/60
48FF7 = 1800 – 4800°F D/30 x D/150
48FF8 = 1800 – 4800°F Extension Tip

7
Modline 3 Catalog
Model Selection Chart (continued)
BLOCK B (cont.) BLOCK B (cont.)
3G Series SLR Version* – 1.5 to 1.6 µ
18F05 = 500 – 1800°F D/50 3R Series SLR Version – 0.7 to 1.08 µ and 1.08 µ BLOCK B (cont.)
20F10 = 600 – 2000°F D/100 25F05 = 1300 – 2500°F D/50 3V Series Fiber Optic Version
25F15 = 700 – 2500°F D/150 32F05 = 1800 – 3200°F D/50 10 ft (3 m) fiber optic cable
40F10 = 2000 – 4000°F D/100 15CF5 = 500 –1500°C D/30 †
10C05 = 250 – 1000°C D/50 65F15 = 2500 – 6500°F D/150 15CF8 = 500 –1500°C Extension Tip †
11C10 = 350 – 1100°C D/100
14C15 = 400 – 1400°C D/150 14C05 = 700 – 1400°C D/50 † These models are restricted to sensing head
16C05 = 900 – 1600°C D/50 ambient temperatures of 50 to 113°F (10 to
* Emissivity span is restricted to 0.3 to 1.0 for 20C10 = 1100 – 2000°C D/100 45°C) and a minimum emissivity setting of
the first 100°F (55°C) for all temperature 24C05 = 900 – 2400°C D/50 0.400.
ranges. 35C15 = 1500 – 3500°C D/150

3G Series Fiber Optic Options* 3R Series Fiber Optic Version BLOCK C Signal Conditioning
10 ft (3 m) fiber optic cable 10 ft (3 m) fiber optic cable 0 = Peak Picker and Track & Hold
22FF5 = 650 – 2200°F D/30 25FF5 = 1300 – 2500°F D/30
22FF8 = 650 – 2200°F Extension Tip 25FF8 = 1300 – 2500°F Extension Tip
25FF6 = 800 – 2500°F D/60 32FF5 = 1800 – 3200°F D/30 BLOCK D Analog Signal Output
25FF7 = 800 – 2500°F D/30 x D/150 32FF8 = 1800 – 3200°F Extension Tip 0 = 4 to 20mA, 0 to 20mA, or 5µA per
25FF8 = 800 – 2500°F Extension Tip 40FF5 = 2000 – 4000°F D/30 Degree
40FF8 = 2000 – 4000°F Extension Tip 1 = 0 to 10 Vdc
12CF5 = 350 – 1200°C D/30 65FF6 = 2500 – 6500°F D/60
12CF8 = 350 – 1200°C Extension Tip 65FF7 = 2500 – 6500°F D/30 x D/150 BLOCK E Digital Communication
14CF6 = 450 – 1400°C D/60 0 = RS 485 Digital Interface
14CF7 = 450 – 1400°C D/30 x D/150
14CF5 = 700 – 1400°C D/30
14CF8 = 450 – 1400°C Extension Tip
14CF8 = 700 – 1400°C Extension Tip
16CF5 = 900 – 1600°C D/30 BLOCK F Controller Outputs
* Emissivity span is restricted to 0.3 to 1.0
0 = None
for the first 100°F (55°C) for all temperature 16CF8 = 900 – 1600°C Extension Tip
ranges. 20CF5 =1100 – 2000°C D/30 1 = PID Controller with deviation alarms
20CF8 = 1100 – 2000°C Extension Tip and 4 to 20mA isolated output
3L Series** – 1.5 to 1.6 µ and 1.65 to 1.71µ 24CF6 = 900 – 2400°C D/60 2 = 2-Point On/Off Controller with relay
10F05 = 500 – 1000F D/50 24CF7 = 900 – 2400°C D/30 x D/150 outputs
14F10 = 700 – 1400F D/100 35CF6 = 1500 – 3500°C D/60
18F10 = 1000 – 1800F D/100* 35CF7 = 1500 – 3500°C D/30 x D/150
3V Series SLR Version * – 0.91 to 0.97 µ
05C05 = 250 – 550°C D/50
10C02 = 400 –1000°C D/20
07C10 = 400 – 750°C D/100
10C10 = 550 – 1000°C D/100* 12C05 = 450 –1200°C D/50

* Indicates spectral response of 1.00 to * Adjustable range of response time 0.10 to 60


1.20 µ and 1.65 to 1.71 µ. seconds
**Response time of 0.02 to 60 seconds for
all models. Note: Specifications subject to change without notice.

The Worldwide Leader in Noncontact Temperature Measurement

IRCON, Inc.
Worldwide Headquarters
Santa Cruz, CA USA
Tel: 1 800 227 8074 (USA and Canada, only)
1 831 458 3900 www.ircon.com
info@ircon.com
European Headquarters
Berlin, Germany United Kingdom
Tel: 49 30 4 78 00 80 Tel: +44 1908 630 800
China Headquarters
Beijing, China
Tel: 8610 6438 4691

To find an IRCON office near you, please visit www.ircon.com

Worldwide Service
IRCON offers services, including repair and calibration.
For more information, contact your local office or e-mail info@ircon.com © 2009 IRCON, Inc. (3309357 Rev J ) 9/2009