EDX 3600
Version:2008.06.02
1 www.skyray-instrument.com
This service manual describes the structure and
maintenance of the EDX3600 model.Users are
recommened to refer to this manual for the
troubleshooting,repair,maintenance and inspection
procedure of the instrument.This manual does not
contain operating instructions.The specifications of
the unit are subject to change due to equipment
improvement.Therefore,the contents and diagrams
of this manual may be slightly inconsistent with the
equipment delivered to the user.The contents of this
manual may be also revised without notice.
2 www.skyray-instrument.com
CONTENTS
3 www.skyray-instrument.com
Chapter1 Environment
4 www.skyray-instrument.com
Chapter 2
Structure Of The
EDX3600
5 www.skyray-instrument.com
EDX3600
6 www.skyray-instrument.com
Display Pannel
7 www.skyray-instrument.com
Measurement Chamber
8 www.skyray-instrument.com
Faucet
9 www.skyray-instrument.com
Side View
10 www.skyray-instrument.com
Sideview Opposite
11 www.skyray-instrument.com
Rearview
12 www.skyray-instrument.com
Inside view
13 www.skyray-instrument.com
14 www.skyray-instrument.com
Chapter3 Parts list
15 www.skyray-instrument.com
High Voltage Unit
16 www.skyray-instrument.com
HV Parallel Interface
Pin 1:Grounding
Pin 2:Voltage display
Pin 3:Current display
Pin 4:INTLK
Pin 7:Voltage adjustment(DB-3 pin)
Pin 10:Current adjustment(DB-6 pin)
Pin 15:HV grounding(DB-9 pin)
17 www.skyray-instrument.com
HV Parallel Interface
18 www.skyray-instrument.com
MCA Board
19 www.skyray-instrument.com
What does every interface stand for?
20 www.skyray-instrument.com
MCA board
21 www.skyray-instrument.com
Interface 1:DB faucet
DB-3 pin-High voltage adjustment
DB-6 pin-Current adjustment
DN-9 pin-grounding
Interface 2:Powersupply for MCA board
Interface 3:Signal wire of detector
Potentiometer 4:Upper channel limit adjust
Potentiometer 5:Lower channel limit adjust
Interface 6:The 14 pin chip (The 10th pin from top to bottom
controls the amplification).
Light 8:Indicator light(3lights shine;1light flickering )
22 www.skyray-instrument.com
Detector
23 www.skyray-instrument.com
Faucet 1:Signal wire connection
Faucet 2:SNE wire connection
Faucet 3:Power supply for detector(5 pin)
The correct voltage for the number 3 are
+12DCV ,-12DCV,+5DCV, 0,grounding
respectively.
24 www.skyray-instrument.com
Main Power Box
25 www.skyray-instrument.com
Power box for Spellman
26 www.skyray-instrument.com
Main Power box
Input :220ACV
Output: +5DCV, -5DCV,+12DCV,
-12DCV,2AMP
Power box for spellman
Input:220ACV
Output: 24DCV,5AMP
27 www.skyray-instrument.com
Chapter 4
Maintenance and
Inspection
28 www.skyray-instrument.com
Maintenance precautions
29 www.skyray-instrument.com
Make sure you complete the process of preheat in
advance of a measurement.
Make sure that you input the correct collimater with
right filters and confirm they are placed in the best
position before a test.
For safety,do not put heavy articles on top of the
instrument.
Do not run a test with the side cover open.
Add in pump oil before you run the vacuum pump
30 www.skyray-instrument.com
Keep the measurement chamber clean and
sealed.
In the case of fuse replacement,always use
parts with the specified capacity.use of
improper parts may lead to accidents.
Always check to assure that all the connector
are adequately inserted.
31 www.skyray-instrument.com
Warning
32 www.skyray-instrument.com
Never ever poke sharp stick or anything
else into the detector window which is
significantly fragile.Any man-made
damage to the detector will be out of
guarantee for a free replacment.
Loosening of a joint or screw while they are
hot may cause scorching.Make sure the
temperature is sufficiently low before
loosening them.
33 www.skyray-instrument.com
When test liquid or powder,make sure that
the sample will not leak or penetrate into the
hardware.
Make sure there is no intense magnetic or
electric interference around the instrument.
Never pull out the power plug when the
instrument is running.
34 www.skyray-instrument.com
Chapter 5
Troubleshooting
35 www.skyray-instrument.com
Malfunction
Classification & Troubleshooting
36 www.skyray-instrument.com
The HV lamp is off while testing 45
The time does not run 46
The time runs very slowly and seems
stopped but the HV lamp is on 47
The instrument can not start 48
Remarks 49
37 www.skyray-instrument.com
No spectrum or count rate display
The possible reasons for failure are as follows:
(1) High voltage is faulty.
(2) X - ray tube is faulty.
(3) MCA is faulty.
Check whether the Be window is broken or not
1st. check the connection with instrument and pc including the BIOS in pc mother
board.
Lpt or USB cable
BIOS parameter
2nd. Power supply box
Check the 12V, -12V, +5V , -5V and 24V
3rd. MCA
Check TB-9 terminal 3pin and 6pin. For example ,in the software voltage
40 ,current 600 ;the value in the multimeter should be 8DCV(3rdpin) and
6DCV(6thpin)respectively.
38 www.skyray-instrument.com
4th. High voltage box
Check the 2nd、3rd pins which indicates the current and
voltage .check the value when testing ,the voltage value should
be 1/5 of the value in the software , and current value should be
1/100 of that in the software .For example in the software
voltage 40 ,current 600 ;the value in the multimeter should be
8DCV and 6DCV
39 www.skyray-instrument.com
Peak drift
(1)Detector is faulty
(2)MCA is faulty
1st . Take a look at the indication light of MCA
2nd.Test the DB-9 value of the MCA.For example,when the volage ,
current value is 40,600 in software, the correct value in the multimeter
is 8DCV ,6DCV correspondingly. In addition, try to adjust the value in
the software,and test if there is corresponding change in the interface.
3rd. Check the detector . First of all, make sure the powersupply wire and
the signal cable is correctly installed,then test the value of the 5pin
powersupply faucet.Finally,you can judge from the spectrum . after
continuous testing several times , and open a reference spectrum ,
then check whether the spectrum drift compared to others
40 www.skyray-instrument.com
Count rate decreases while testing.
41 www.skyray-instrument.com
The HV lamp is flickering during testing
42 www.skyray-instrument.com
Initialization can not complete
43 www.skyray-instrument.com
The shape of spectrum is abnormal
44 www.skyray-instrument.com
The HV lamp is off while testing
45 www.skyray-instrument.com
The time does not run
46 www.skyray-instrument.com
The time runs very slowly and seems
stopped but the HV lamp is on
47 www.skyray-instrument.com
The instrument can not start
48 www.skyray-instrument.com
Remarks
All the problems above will not come out alone ,so
once you meet a problem listed above, don't take it
as a single problem , you should take into account all
the possibility , for example , once the initialization
not completed ,according to the methods
above ,maybe it is because of MCA , but in a
meanwhile ,the countrate is very low ,nearly close to
0, so the reason why initialization failed is because
of low countrate , no count rate no spectrum ,hereby
you should refer to the question 1 and follow the
steps to check.
49 www.skyray-instrument.com
Once it is low count rate (or no count rate and no spectrum)
maybe you think it is because of the high voltage and x-ray tube
according to question 1 .but in fact (take EDX2800 for example )
it is because of the wrong collimator , the collimator is chosen
automatically for EDX2800 ,but sometimes it is not chosen
rightly ,maybe the smallest one is chosen while we need the
largest one for testing, so the count rate is very low and the
hardware is no problem .
So once you meet a problem above ,just follow the steps
above ,but if the problem you met is not listed above ,don't
worry ,take it easy ,just analyze the possibility step by step ,and
there will be no problem for you.
50 www.skyray-instrument.com
Chapter6
COPYRIGHT
51 www.skyray-instrument.com
The copyright of this manual is owned by Skyray
Instrument Inc. Without the prior written permission
obtained from Skyray Instrument Inc, no part of this
documentation shall in any form or by any means be
reproduced, excerpted, stored in a retrieval system,
modified, distributed, translated into other languages,
in whole or in part applied for a commercial purpose
unless be admitted by copyright law.
52 www.skyray-instrument.com
Disclaimer
53 www.skyray-instrument.com
Copyright@ Skyray corporation 1992.All rights are
reserved,including all those to reproduce this publication or
parts theory in any form without permission in writing from
skyray corporation.
Any errors which may occur in this publication despite upmost
care will be corrected as soon as possible.
China office address: Skyray building, Tsinghua science park
1666,south Weicheng Rd.,Kunshan, jiangsu Province.
Tel:86-0512-57017000
FAX:86-0512-57017060
Service Hotline:86-800-999-3800
54 www.skyray-instrument.com