Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User's Manual
Publication Reference : 105/12192-01SA
Issue 7
Warranty
NDC Infrared Engineering guarantees its equipment against faulty components for a period of twelve
months from delivery. Replacement components will be free of charge. NDC Infrared Engineering shall not
in any event be liable for consequential damage or loss. NDC Infrared Engineering operates a customer care
facility and all requests for repair and replacement should be directed to the Customer Care Manager. The
serial number of the equipment should be quoted in any communications. The right to change specification
and price is reserved by NDC Infrared Engineering.
Copyright
The equipment described in this publication contains a computer program or programs which is or are the
confidential and copyrighted property of NDC Infrared Engineering. NDC Infrared Engineering grants to the
purchaser of the equipment the right and licence to use the program or programs in the normal use of the
equipment only, but grants no further licence. The copying, disassembly, reverse compilation or
transcription or transmission of the said program or programs in any form for the use of the purchaser or
others is strictly prohibited. The contents of this publication must not be reproduced in part or in whole in any
form without the written permission of the copyright owner.
NDC Infrared Engineering makes no guarantees of any kind with regard to this manual. NDC Infrared
Engineering shall not be liable for errors contained herein or for consequential or incidental damages
incurred as a result of acting on information contained in the manual.
INTERNATIONAL OFFICES
NDC Infrared Engineering NDC Infrared Engineering NDC Infrared Engineering
5314 North Irwindale Avenue, Robert-Bosch-Str. 11B, Suite 1210, Parkview Square,
California 91706, 63225 Langen, 960 Jie Fang Bei Road,
USA Germany Guangzhou 510040, China
Tel +1 626 960 3300 Tel +49 61 0397 7112 Tel +86 (0)20 8 666 2790
Fax +1 626 939 3870 Fax +49 61 0397 7116 Fax +86 (0)20 8 666 5963
Page (ii)
TM710
Prelims
TM710
User's Manual
CONTENTS
Preliminary pages
Chapter
1 DESCRIPTION
Gauge
GaugePort
Operator Interface Unit
Sample Display Unit
Handheld Unit
2 INSTALLATION
Gauge
GaugePort
Operator Interface Unit
Sample Display Unit
Handheld Unit
3 OPERATION
Operator Interface Unit
Handheld Unit
Sample Display Unit
4 CALIBRATION
6 REFERENCE
continued overleaf
Page (iii)
TM710
User's Manual
Annex
A Temperature Measurement
Page (iv)
TM710
Prelims
WARNINGS PAGE
LETHAL VOLTAGE
!
IF THE GAUGE IS NOT INSTALLED AND USED IN THE MANNER
PRESCRIBED IN THIS MANUAL, THE SAFETY PROTECTION
AFFORDED BY THE EQUIPMENT MAY BE IMPAIRED.
ELECTRICAL SAFETY
!
FOR COMPLETE ELECTRICAL SAFETY IN OPERATION THE
EQUIPMENT HAS BEEN SUPPLIED WITH DOUBLE-POLE NEUTRAL
FUSING INCORPORATED. DO NOT MODIFY IN ANY WAY.
Page (v)
TM710
User's Manual
Page (vi)
TM710
Prelims
CAUTIONS PAGE
!
OPTICAL COMPONENTS
Page (vii)
TM710
User's Manual
Page (viii)
TM710
Prelims
READER’S NOTES
GENERAL
CUSTOMER CARE
For enquiries relating to the operation and use of the TM710 equipment
described in this Manual please contact
ASSOCIATED PUBLICATIONS.
AMENDMENTS
Page (ix)
TM710
User's Manual
Page (x)
TM710
Prelims
ABBREVIATIONS
A Amp(s)
a.c. Alternating current
d.c. direct current
dia diameter
E Mains Earth
EMC Electromagenetic compatibility
EMI Electromagnetic interference
Hz Hertz (cycles per second
l.e.d. Light emitting diode
L Mains Live
mA milliamps
mm millimetre
N Mains Neutral
NI Not illustrated
p.s.i. pounds per square inch
PCB Printed Circuit Board
rpm revs per minute
SDU Sample Display Unit
SC Supplied by customer
sync synchronisation
UNC Unified National Coarse (thread)
V Volt(s)
º degrees of arc
ºC degrees Celsius
Page (xi)
TM710
User's Manual
Page (xii)
TM710
Prelims
FACTORY CONFIGURATION
MEASUREMENTS
Measurement Application
Channel Span Trim
name code
Channel 1
Channel 2
Channel 3
Channel 4
SET PASSWORD
The enable/disable feature associated with the BCD INPUT function, and
the DIAG (diagnostics) function, are each protected by the factory set
password “1234”. This code may be modified as described in Chapter 3
(Operation/Configuration/Passwords) of this manual.
%WEIGHT
Page (xiii)
TM710
User's Manual
Page (xiv)
TM710
Prelims
RECORD OF AMENDMENTS
10
Page (xv)
TM710
User's Manual
Page (xvi)
Description
Chapter 1
CHAPTER 1
DESCRIPTION
CONTENTS
Page
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
The Gauge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
GaugePort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Operator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Sample Display Unit (SDU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Handheld Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
PC Interface Software (GaugeView) . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
PRINCIPLE OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
OPTICAL layout of THE TM 710 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
The Dual Detector Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
THE GAUGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
MECHANICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
ELECTRICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
GAUGEPORT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
MECHANICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
ELECTRICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
RS232 communications port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
Motherboard connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
710 Databus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Gated input signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Plug-In Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Network Termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Free Topology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
Bus Topology. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
OPERATOR INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
MECHANICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
ELECTRICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Analogue Output Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Alarm Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Gated input signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Plug-In Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
RS232 communications port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
710 Databus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Network Termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
SAMPLE DISPLAY UNIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Page 1-1
TM710
User's Manual
CONTENTS (continued)
Page
TABLES
Table Page
1 Details of the signals and power supplies on the Gauge connector . . . . . . 1-14
2 Recommendedcable entry locations . . . . . . . . . . . . . . . . . . . . . . 1-17
3 Pin out details for RS232 Comms channel on GaugePort, PL1 . . . . . . . . 1-18
4 GaugePort motherboard user connection details . . . . . . . . . . . . . . . 1-19
5 The Operator Interface motherboard connection details . . . . . . . . . . . 1-25
6 Connection details of the SDU PCB connectors . . . . . . . . . . . . . . . . 1-28
6 Connection details of the Hand Held Interface external plug, PL1. . . . . . 1-30
ILLUSTRATIONS
Figure Page
1 (top) The gauge supported on tube mounts with and without air purge window
fitted, (middle) the GaugePort and the Operator Interface (bottom) the Sample
Display Unit and the Handheld Interface . . . . . . . . . . . . . . . . . . . 1-4
2 Connected elements of the Series 710 system . . . . . . . . . . . . . . . . . 1-7
3 The optical path of the 710 system . . . . . . . . . . . . . . . . . . . . . . . 1-8
4 The dual detecor arrangement in the TM710 gauge . . . . . . . . . . . . . . 1-9
5 (sheet 1) - Typical TM710 system configurations . . . . . . . . . . . . . . . 1-11
5 (sheet 2) - Typical TM710 system configurations . . . . . . . . . . . . . . . 1-12
6 Pin locations for the socket on the Gauge . . . . . . . . . . . . . . . . . . . 1-14
7 Recommended connection of the screen to a cable gland . . . . . . . . . . . 1-16
8 Cable entry positions in the base section of the GaugePort enclosure . . . . 1-16
9 Pin out details for SK1 on the Gaugeport . . . . . . . . . . . . . . . . . . . 1-18
10 GaugePort and Operator Interface “daisy chain” connection
of the 710 Databus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
11 Pin details for connector PL1 on the Hand Held Interface Unit . . . . . . . 1-30
Page 1-2
Description
Chapter 1
OVERVIEW
GENERAL
(2) GaugePort
Page 1-3
TM710
User's Manual
Figure 1 - (top) The gauge supported on tube mounts with and without air purge window
fitted, (middle) the GaugePort and the Operator Interface (bottom) the Sample Display Unit
and the Handheld Interface
Page 1-4
Description
Chapter 1
The Gauge
GaugePort
Operator Interface
The SDU is normally located close to the gauge and displays the gauge
measurements (up to 4). Its sampling facility allows accurate cross
checking of the gauge against the laboratory reference method.
Handheld Interface
Page 1-5
TM710
User's Manual
PRINCIPLE OF OPERATION
The Gauge contains a quartz halogen source lamp which projects light
through an optical system to provide light in the infrared spectrum at
wavelengths selected to suit the particular constituents being measured.
Infrared light is radiated from the Gauge on to the product where some
is reflected back to the Gauge.
Page 1-6
Description
Chapter 1
HANDHELD
INTERFACE UNIT
SAMPLE
710 GAUGE
DISPLAY UNIT GAUGEPORT
SAMPLE
710 GAUGE
DISPLAY UNIT GAUGEPORT
SAMPLE
710 GAUGE
DISPLAY UNIT GAUGEPORT
ETHERNET OR FIELDBUS
RS232
PERSONAL COMPUTER
OPERATOR
INTERFACE UNIT
710 DATABUS
CD-0702
Page 1-7
TM710
User's Manual
All elements of the optics have been designed to provide modules, which
can be fitted together during manufacturing without the need for any
optical alignment. The optical schematic of the gauge is shown in
Figure 3.
SEGMENTED
MIRROR
PRIMARY
SOURCE DETECTOR
LAMP
SECONDARY
DETECTOR
BEAM
SPLITTER
OFF AXIS
COLLECTING
-
MIRROR
FILTER WHEEL
The light then passes onto the beam splitter. This device projects a
small part of the outgoing beam, uniformly and without abberation, to
the optical system behind it. This is used for internal referencing
purposes. The benefit of this device is that it ensures much more faithful
tracking of the two detectors which is critical for the Dual Detector
Referencing concept (explained below) to work accurately and confer
high stability on the gauge.
The light that passes through the beam splitter is then collected by
another mirror, returned and focussed onto the secondary detector.
Page 1-8
Description
Chapter 1
The majority of the light passes out of the gauge onto the product. The
product absorbs some of it. A small amount of the backscattered
(diffusely reflected) light is collected by the segmented, collecting mirror.
The dual detector principle ensures very high long term stability of the
gauge. A detailed view of the arrangement of the detectors in the optical
layout illustration is shown in Figure 4.
The secondary detector only sees signals that can have been influenced
by the Gauge, for example, the effects of lamp ageing, whereas the
primary detector sees a composite signal made up of the “Gauge and
Product” characteristics.
SECONDARY
DETECTOR
INTERNAL
REFERENCE
LIGHT
>
>
OUTGOING
LIGHT
BEAM SPLITTER
MIRROR
Page 1-9
TM710
User's Manual
The two detectors are always spectrally matched to have the same
infrared properties and are both mounted on a heat sink to ensure they
operate at the same temperature.
Configurations
Page 1-10
Description
Chapter 1
GAUGEPORT
TM710
MAINS SUPPLY
(A)
GAUGEPORT
710 Sample Display Unit
MAINS SUPPLY
(B)
SAMPLE
DISPLAY UNIT
TM710 NETWORK
MAINS SUPPLY
(C)
CD-0509
Page 1-11
TM710
User's Manual
TM710
MAINS SUPPLY
(D)
HAND HELD
GAUGEPORT INTERFACE UNIT
OPERATOR
INTERFACE UNIT 710 Hand Held Interface
NETWORK
710 Operator Interface
MAINS SUPPLY
SAMPLE
DISPLAY UNIT #1
SAMPLE
RS232 LINK (PC) DISPLAY UNIT #2
710 Sample Display Unit
NETWORK
710 Sample Display Unit
MAINS SUPPLY
710 GAUGE #1
710 GAUGE #2
TM710
TM710
CD-0519
(E)
Page 1-12
Description
Chapter 1
THE GAUGE
GENERAL DESCRIPTION
The Gauge is known as a backscatter type due to the fact that it collects
a portion of the infrared light that is scattered back towards the Gauge
after it re-emerges from the product.
MECHANICAL DESCRIPTION
The cast aluminium case comprises two sections, the lid (upper part) and
the box (lower part) that are screwed together using six screws. These
screws are captive in the lower section. When fully screwed in to the
threaded holes in the upper section the screw heads sit in recessed holes
in the lower section. There is a gasket sandwiched between the two case
sections to prevent the ingress of dust and water and other liquids rated
at IP65. The gasket is glued to the lower section.
Page 1-13
TM710
User's Manual
The top aspect of the lid contains four M8 threaded holes. These holes
are provided to permit the Gauge to be mounted at its location in the
plant.
An air purge window may be fitted to the bezel on the Gauge. When the
air purge window is supplied with compressed air, it minimizes the
contamination of the window of the gauge by dust, condensation or other
condensates.
ELECTRICAL DESCRIPTION
The TM710 Gauge has a single connector fitted to the rear face of the
unit. The outline of the 12-way connector is shown in Figure 6. The pin
connection details are given in Table 1.
4 12
5
1 11
6 3
2 10
7 9
8
CD-0453
Table 1 - Details of the signals and power supplies on the Gauge connector
5 0 V Power 11 Gate LO
Page 1-14
Description
Chapter 1
(4) Flashing RED and GREEN. Error detected in Gauge, but the
gauge continues to work.
GAUGEPORT
(1) To supply d.c. power to the Gauge via an internal mains power
supply unit
(7) To provide a connection point and power supply for the Sample
Display Unit (SDU)
Page 1-15
TM710
User's Manual
MECHANICAL DESCRIPTION
The two sections are fabricated from sheet metal. The base section
contains a motherboard and an a.c. mains input power supply unit. The
lid is empty.
The base section has a single externally mounted 9-pin D-type plug for
connection to a remote RS232 device ( a PC for example) or the network
device, such as the Hand Held Interface Unit. This connector has a metal
protective cover fitted when supplied.
1 2 3 4 5
6 7 8 9 10
CD-0455
Figure 8 - Cable entry positions in the base section of the GaugePort enclosure
Page 1-16
Description
Chapter 1
1 - Spare
8 - Spare
10 - Spare
The metal glands provide 360 degree screening when used with a
suitably prepared screened cable.
Page 1-17
TM710
User's Manual
ELECTRICAL DESCRIPTION
+ +
9 8 7 6
Table 3 - Pin out details for RS232 Comms channel on GaugePort, PL1
Pin 5 0V Orange *
* Internal wiring
Page 1-18
Description
Chapter 1
Motherboard connectors
PL12, pin 1 +24 V d.c. Sampling Display Unit (SDU) J1 pin 1 (Red)
PL12, pin 2 0V Sampling Display Unit (SDU) J1 pin 2 (Green)
PL12, pin 3 710 Databus B Sampling Display Unit (SDU) J2 pin 1 (Blue)
PL12, pin 4 710 Databus A Sampling Display Unit (SDU) J2 pin 2 (Yellow)
Page 1-19
TM710
User's Manual
PL9 PL9
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
DATABUS B DATABUS B
CD-0671
DATABUS A DATABUS A
GAUGEPORT OR GAUGEPORT OR
OPERATINTERFACE UNIT OPERATINTERFACE UNIT
Figure 10 - GaugePort and Operator Interface “daisy chain” connection of the 710 Databus
710 Databus
There is a current sense input available to the user that can be used to
control the acquisition of data by the Gauge. This is the Gated Input
signal, named Gate HI and Gate LO on the User Interface connector PL9
pin 3 and pin 4. The input current required is 5 mA (compliant with a
voltage between 5 and 36 volts). Calculations for output are carried out
using the last collected data. The recommended minimum gate pulse
width is 1 ms.
When the input level is zero (no current flowing) data is enabled and the
data is used by the Gauge in its calculations for output. When the gated
input level is high (current flowing) data is disabled and no new data is
collected. If the input is left on indefinitely then the calculated output
from the gauge will be out of data.
Page 1-20
Description
Chapter 1
Plug-In Modules
(2) Alarm Output Node PCB. This module provides two channels of
HI and LO alarm outputs. There are two HI and two LO fused
no-volt relay contacts that close when preset alarm conditions
are attained.
Network Termination
NOTE
A networked 710 system supplied by NDC IE would normally be pre-configured
with the termination circuitry set to the optimum positions. However if a system is
extended, the network termination may need to be re-configured. This is an
important consideration and should be observed.
There are two network topologies catered for in the 710 system :
Page 1-21
TM710
User's Manual
Free Topology
Bus Topology
The bus topology is an electrically superior system where the units are
daisy-chained in a serial fashion, termed a bus. Here the network
termination switch SW1 in the first and last GaugePort units (or
Operator Interface) are set so that the switch marked BUS is in the ON
position and the switch marked FREE is in the OFF postion. All other
GaugePort (and Operator Interfaces) must have their network
termination switch (SW1) set with BUS to OFF and FREE to OFF.
Using this topology, total cable runs of 2700 metres can be achieved,
again subject to the amount of electrical interference that the network is
exposed to.
Page 1-22
Description
Chapter 1
OPERATOR INTERFACE
GENERAL
It can replace the GaugePort and be used with a Gauge forming the
other sytem building block or be a fully floating network device. It
displays up to four measurements simultaneously from one gauge or a
selection of gauges (up to four) from all the gauges connected to the
network and has comprehensive calibration and diagnostic facilities.
Page 1-23
TM710
User's Manual
(5) Can talk to up to any ten of the gauges connected to the network
and display data in any combination from any of them
(7) Sampling facility for when used as a device close to the gauge
MECHANICAL DESCRIPTION
NOTE
The mechanical and electrical details of the Operator Interface are very similar to
the GaugePort. All features described in the GaugePort apply also to the Operator
Interface. The Operator Interface has the additional Mechanical and Electrical
features described below.
The lid is a fully removeable cast aluminum lid. When in normal use the
lid is screwed down using the four captive screws. The lid has a central
aperture (126 x 70 mm) which is fitted with an l.e.d. backlit liquid
crystal display (240 x 128 pixels).
The circuit boards attached to the lid connect to the motherboard in the
lower section via a cable. These cables have connectors at either end
that may be unplugged at the lid end and at the motherboard end.
All metal surfaces of the case are painted and have a textured finish.
There are gaskets between the display panel and the lid, and between
the two case sections to make the enclosure watertight (IP65).
Page 1-24
Description
Chapter 1
ELECTRICAL DESCRIPTION
For details of connectors PL9, PL10, PL11 and PL12, refer to Table 4.
PL2, pin 1,2 RL1 contacts Customer's equipment for Alarm output 1 LO
PL2, pin 3,4 RL2 contacts Customer's equipment for Alarm output 1 HI
PL2, pin 5,6 RL3 contacts Customer's equipment for Alarm output 2 LO
PL2, pin 7,8 RL4 contacts Customer's equipment for Alarm output 2 HI
Alarm Outputs
There are four independent sets of fused no-volt relay contacts for use by
the operator (one contact pair for each measurement channel). These
contacts are activated when calulated measurements exceed preset
Page 1-25
TM710
User's Manual
Plug-In Modules
710 Databus
Network Termination
Page 1-26
Description
Chapter 1
GENERAL DESCRIPTION
This unit will normally be located close to the gauge and allows accurate
cross checking of the gauge against the laboratory reference method.
There is an output available that switches “on” during the time that the
SDU is sampling. This may be used, for example, to light an external
illumination device when the unit is sampling.
MECHANICAL DESCRIPTION
The enclosure is available in two versions (1) painted steel and (2)
stainless steel. The size and shape of the versions are identical.
The lower side has two holes, one fitted with a cable gland and one fitted
with a blanking plug. These are used for cables to enter the enclosure to
reach the internal connectors. The blanking plug can be removed a fitted
with a second gland when the switched output is required.
Page 1-27
TM710
User's Manual
ELECTRICAL DESCRIPTION
Connection to the SDU is via a single cable through a gland fitted to the
lower side of the enclosure. This cable carries both 710 Databus signals
and the power supply. Inside the SDU these signals are terminated with
two screw terminal sockets that mate with plugs on the SDU PCB.
The SDU is normally supplied with the cable already wired and fitted,
and the other end of the cable terminated with bootlace ferrules for
connection to the connectors on the GaugePort or Operator Interface.
J1, pin 1 LonWorks B (IN) PL12 pin 3 GaugePort (or Operator |Interface)
motherboard
J1, pin 2 LonWorks A (IN) PL12 pin 3 GaugePort (or Operator |Interface)
motherboard
J1, pin 3 LonWorks B (OUT) Next network connected unit
J1, pin 4 LonWorks A OUT) Next network connected unit
Page 1-28
Description
Chapter 1
GENERAL
The Hand Held Interface is a portable unit that connects to the internal
710 Databus network via the GaugePort or Operator Interface external
connector and provides the following features:
Page 1-29
TM710
User's Manual
MECHANICAL
The lid has a central aperture which is fitted with an l.e.d. backlit liquid
crystal display.
The unit can be fitted with a rubber boot to protect it from ambient
substances. There is a apearture in the boot to provide access to the
touch sensitive display panel.
ELECTRICAL
The lower side panel has a single externally mounted 9-pin D-type plug
for connection to the GaugePort or Operator Interface, see Figure 11.
The interconnecting cable carries network signals, 710 Databus A and B
and the power supply for the Hand Held Interface.
+ +
6 7 8 9
Figure 11 - Pin details for connector PL1 on the Hand Held Interface Unit
Table 7 - Connection details of the Hand Held Interface external plug, PL1
Page 1-30
Description
Chapter 1
The GaugeView 710 Software allows the user to communicate with the
gauges from a terminal connected to the network. The terminal screen
displays results and trends and can carry out calibration analysis and
adjust calibration settings of a gauge.
Page 1-31
TM710
User's Manual
Page 1-32
Installation
Chapter 2
CHAPTER 2
INSTALLATION
CONTENTS
Page
Page 2-1
TM710
User's Manual
ILLUSTRATIONS
Figure Page
Page 2-2
Installation
Chapter 2
TABLES
Table Page
Page 2-3
TM710
User's Manual
Page 2-4
Installation
Chapter 2
EMC PRECAUTIONS
(2) Lay signal cables in a low-power signal conduit, and mains cables
in low power mains supply conduit.
(3) Ensure that no part of the gauge/ system is placed near sources of
strong ‘EMI’ (Electromagnetic Interference). Examples of EMI:
(a) Large electric motors
(b) Welding equipment
(c) Large static discharges
(d) Infrared ovens
(e) Microwave ovens
(f) Large transformers
(g) Transmitters
(h) Power control circuits
NOTE
Do not alter any cables or enclosures without prior permission from the
Engineering Department at NDC Infrared Engineering Ltd.
For advice or help on installation concerning ‘EMC’, please contact the
Customer Care Helpdesk at NDC Infrared Engineering.
Page 2-5
TM710
User's Manual
INTRODUCTION
EQUIPMENT SUPPLIED
The TM710 system can comprise the main items that are listed in
Table 1.
Page 2-6
Installation
Chapter 2
UNPACKING PROCEDURE
Open the packaging carefully and remove each item. Take the items to a
clean area for inspection. Check that the items are free from external
damage, connectors are free from dirt and cables are not chaffed or
kinked.
Check against the delivery note that all parts listed have been supplied,
and that no obvious external damage has occurred in transit. In the
event of loss and/or damage notify the carrier, NDC Infrared
Engineering or the agent of NDC Infrared Engineering as appropriate.
Check that the details on the serial number plate of each unit is as
specified on the Advice Note accompanying the supplied equipment.
STORAGE
INSTALLATION CONSIDERATIONS
Gauge
The gauge mounting equipment or any objects in contact with the gauge
mounting equipment must not be subjected to excessive vibration as
this may cause premature failure of the source lamp in the gauge.
Page 2-7
TM710
User's Manual
The gauge should be mounted with due regard for accessibility for
maintenance.
The TM710 gauge will give accurate measurements only when correctly
mounted relative to the product and located in a suitable environment.
The gauge must be situated to see a consistent representative sample of
the product under measurement. Refer to Figure 3 for the distance from
the glass window to the product and the inclination of the gauge to the
product.
Ambient Light
The gauge should be shielded from strong sources of natural light as this
will reduce the sensitivity of the gauge. Artificial light sources powered
from the mains supply have virtually no effect on the measurement
being made by the gauge and therefore permit the gauge to be situated
in normally lit factory settings, for example light from mains lamps and
flourescent tubes. However, the light source, or its direct reflection must
not be directed at the glass window of the gauge.
Temperature
The enclosure is equipped with 0.25 inch tube inlet and outlet fittings for
easy connection to a water supply.
The air supplied to the air purge window must be clean and dry. If it is
not it will contaminate the glass window and affect the measurements
being made. The recommended flow rate is 75 litres per minute (+/- 10
litres per minute).
Page 2-8
Installation
Chapter 2
The cables running into the GaugePort and Operator Interface should be
placed away from heavy duty power cables and cables carrying high
frequencies to other equipment.
The cable from the SDU should be placed away from heavy duty power
cables and cables carrying high frequencies to other equipment.
Page 2-9
TM710
User's Manual
Page 2-10
Installation
Chapter 2
INSTALLATION PROCEDURES
GAUGE
General
Do not subject the Gauge mounting, nor any objects in contact with the
head to excessive vibration as this may cause premature failure of the
source lamp.
Mounting
NOTE
For some installations imperial fixings are required to mount the gauge to the
production line. This is achieved by the use of thread adaptors. These are studs
with a metric thread at one end (for screwing into the gauge) and an imperial
thread at the other for fixing to the customers fixing structure.
The threaded Stud Adaptors need to be fitted to the gauge. They have a
10 mm long M8 thread at one end and a one inch long 3/8 inch UNC
thread at the other. Screw the four Stud Adapters supplied into each
gauge. Use threadlock (such as Loctite 638) on the thread of the M8
threads when screwing this end into the gauge fixing locations. When
they have been fitted leave aside to set for 12 hours before further use.
Mount the Gauge to its selected mounting bracket using the four 3/8 inch
UNC studs (ref. 105/12061-01).
Page 2-11
TM710
User's Manual
75 47.5
WATER COOLING GS
0.25 INCH OUTSIDE
DIAMETER (2 OFF)
75
MOUNTING HOLES
M8 x 10 DEEP
ALL DIMENSIONS IN MM
CD-0451-TM
29
310 18.5 177
TM710 CONNECTOR
REMOVAL
172
CLEARANCE
140
230
101 (DIA)
Page 2-12
Installation
Chapter 2
Tube mounting
SADDLE CLAMP
TUBE
MOUNTING PLATE
GAUGE
SADDLE CLAMP
ALL DIMENSIONS IN MM
CD-0766
33
TUBE
205
TM710
244 85
177
Page 2-13
TM710
User's Manual
Place the Gauge at the correct distance from, and at the correct angle to
the product that is to be measured. Refer to Figure 3. At this distance
the patch size will be 50 mm diameter. The gauge is inclined at 20
degrees to the product to avoid specular reflections. The permissible
pass height fluctuation of the product is +/- 100 mm.
Cooling
10
TM7
20 DEGREES
PRODUCT
Page 2-14
Installation
Chapter 2
Caution
Ensure that the air supply to be fed to the Gauge is clean, dry and free
from oil.
Normal atmospheric dust will not affect the gauge measurements. If the
atmosphere is likely to contain significant quantities of contaminants in
dust or droplet form, an Air Purge Window assembly should be fitted
Gauge windows must be kept free from water droplets and condensation.
NOTE
Window contamination up to a preset threshold can be tolerated and is
compensated for by the internal window contamination monitor.
The air purge window is fitted to the Gauge window bezel. A groove is
provided around the rim of the bezel for attaching the air purge window.
There are three set screws fitted to the air purge window which when
screwed into this groove hold the air purge window secure.
(1) Unscrew the three set screws sufficiently so they do not protrude
beyond the inner wall of the air purge window
(2) Check that the cylindrical insert and gaskets are located
correctly in the air purge housing
(4) Align the set screws with the groove in the bezel and rotate the
air purge window so that the air connector points in the right
direction to receive the air supply
(5) Evenly tighten the three set screws to secure the air purge
window in its required position
The air supply connects to the air purge window using a 6 mm outside
diameter plumbing fitting. Fit this compression fitting as follows:
Page 2-15
TM710
User's Manual
(1) Remove the nut and olive from the air purge window and slide
the nut and olive in this order on to the end of the nylon or metal
pipe that carries the air to the Gauge.
(2) Push the pipe into the fitting on the side of the air purge window
and tighten the nut sufficiently so that the connection is made
airtight.
(3) Connect the unit to a supply of clean, dry and oil-free air and set
the delivery pressure to approximately 30 p.s.i at a flow rate of
around 5 litres per minute.
(4) Support the connecting pipe and tighten the nut further, if
required, to make the connection airtight.
Page 2-16
Installation
Chapter 2
GAUGEPORT
208
230 F.C.
710 GaugePort
ALL DIMENSIONS IN MM
6 7 8 9 10 SK1
CD-0449A
Page 2-17
TM710
User's Manual
Mounting
The rear panel of the base section has four threaded bushes fitted.
These are provided to allow the unit to be mounted to a structure or
fitted to a suitable mounting bracket such as those described below.
(1) Mark, drill and prepare the four fixing holes that hold the
bracket to the surface. Do not fit the bracket at this stage.
1.6
265
248
10
20
Page 2-18
Installation
Chapter 2
(2) Secure the bracket to the underside of the unit using the four
M6 x 10 mm socket head screws provided.
Tube mounting
ADAPTOR PLATE
MOUNTING
PLATE TUBE
SADDLE CLAMP
217 600
182 270
115
TUBE
230
TUBE
141
SADDLE
CLAMP
26
ALL DIMENSIONS IN MM
Figure 6 - Tube mounting arrangement for the GaugePort and Operator Interface
Page 2-19
TM710
User's Manual
Modules
The GaugePort unit can accept up to three plug-in modules which are
fitted inside the unit. These are inserted into connectors PL1 and PL6
and PL7 on the Motherboard PCB. The analogue and alarm out node
modules are shown in Figures 25 and 26. They may be plugged into any
module location connector. The motherboard of the GaugePort and
Operator Interface showing the locations of PL1, PL6 and PL7 is shown
in Figure 13. The modules may be supplied already fitted or as separate
items.
Caution
(1) Undo the four screws that secure the lid to the base of the unit
and remove the lid.
(2) From the kit of parts supplied with the module, select the pillar
and refer to Figure 13.
(3) Place this pillar into the slot at position C alongside the existing
shorter pillar at location B and screw the pillar into the stud
fitted to the enclosure wall under the slot. Tighten the pillar on
the stud.
(4) Take the module and with reference to Figures 13, and 24 or 25
align the connector SK1 on this PCB with PL1, PL6 or PL7 on the
motherboard. Ensure that the PCB is between the two pillars
and locate SK1 with the motherboard plug. Push SK1 fully on to
the motherboard plug until the connector is fully inserted and
the connector parts lock.
Page 2-20
Installation
Chapter 2
(5) From the module kit of parts select the module retaining clip and
its pan head screw and fit the clip to the top of the long pillar so
that when the fixing screw is tightened, the return on the clip
holds the module secure.
(6) Repeat stages (4) to (6) for each other module to be fitted.
(8) Fit a cable gland from the module kit of parts supplied, to each of
the locations required.
Page 2-21
TM710
User's Manual
OPERATOR INTERFACE
The lid is fitted to the case by means of four M6 captive socket head
screws in the lid that screw into threaded holes in the top of the base
STATUS INDICATOR
270
CONDUCTIVE GASKET
248 FIXING CENTRES
LID
208
230 F.C.
ALL DIMENSIONS IN MM
6 7 8 9 10 SK1
CD-0450
Page 2-22
Installation
Chapter 2
Modules
Page 2-23
TM710
User's Manual
There are two holes in the lower panel of the base section for cables to
pass through. These are normally supplied fitted with a gland (with
cable) and a blanking plug. The latter must be replaced by a gland if the
user requires to connect to the the internal switched output connector on
the SDU PCB. The glands and blanking plugs are of the same type used
and described in the GaugePort and Operator Interface sections.
The rear face of the base section of the enclosure has threaded inserts
fitted. These are accessible from the outside and are intended to be used
to mount the SDU in its chosen location.
195 54
32.5
50
FIXING
115
CENTRES
CABLE GLAND
CD-0459 50 72.5
FIXING CENTRES
Page 2-24
Installation
Chapter 2
NOTE
The maximum cable run from SDU to GaugePort or Operator Interface is 20
metres.
Mounting
The rear panel of the base section has four threaded bushes fitted.
These are provided to allow the unit to be mounted to a structure or
fitted to a suitable mounting bracket such as those described below.
10
190
170
20
A A
1.6
145 X 95 CENTRED B
B
A A
50 X 50 115
155 139 CENTRED
A A
B B
CD-0458
A A
Figure 9 - Fixing bracket for the Sample Display Unit : outline drawing
Page 2-25
TM710
User's Manual
Drill the SDU bracket fixing holes in the wall for example at the SDU
fixing site according to the dimensions given in Figure 9.
Fit the bracket to the SDU using 4 x M5 screws (plus washers) into the
threaded inserts in the SDU rear panel. Secure the bracket to the wall
and route the cable to reach the GaugePort or Operator Interface.
Tube mounting
MOUNTING
PLATE
TUBE
SADDLE CLAMP
87 195
57 TUBE
115 38
78
ALL DIMENSIONS IN MM
Page 2-26
Installation
Chapter 2
Connecting
Remove the lid from the GaugePort of Operator Interface taking care to
unplug the lid from the base section in the Operator Interface.
Fit a gland in place of the blanking plug at gland location #5 and pass
the cable through the gland assembly and into the base section of the
enclosure. Fit SK12 to the cable according to the wiring schedule and
connect it to PL12 on the motherboard. Adjust the cable path in the
enclosure and tighten the gland.
Page 2-27
TM710
User's Manual
The outline drawing of the Hand Held Interface is shown in Figure 11. It
is supplied with a protective rubber boot fitted. It is not intended that
this unit is permanently fitted. It can be connected to the GaugePort
and Operator Interface by means of a cable that plugs into the 9-way
socket on the latter and the plug on the lower panel of the enclosure.
The rubber boot is fitted on the Hand Held Interface by pushing the unit
through the aperture in the front face of the boot.
+ +
CONNECTOR ACCESSIBLE
THROUGH APERATURE IN
RUBBER BOOT
232 57
151
Page 2-28
Installation
Chapter 2
INTERCONNECTION
Page 2-29
TM710
User's Manual
5
710 NETWORK
TO SAMPLING
INDICATOR #1 710 Operator Interface
MAINS SUPPLY
SAMPLE 1
DISPLAY UNIT #1
6 SAMPLE
RS232 LINK (PC) DISPLAY UNIT #2
11 TO SAMPLING
710 Sample Display Unit
INDICATOR #2
710 NETWORK
4 710 Sample Display Unit
MAINS SUPPLY 11
3 1
710 GAUGE #1
3
710 GAUGE #2
2 2
TM710
TM710
CD-0522TM
Page 2-30
Installation
Chapter 2
Page 2-31
TM710
User's Manual
C
A A
TR1
R1
PL2 R2
1 R3
FS1 8
TR2
2 7
3 PL3
6
FS2
4
5
5
4
6 FS3
3
D1
7
D2
PL5 2
8 FS4 D3
1
D4
PL4
PL6
B
MODULE LOCATION #2
C
A A
CLEAR
PL7
POLYCARBONATE
PROTECTIVE COVER
B
MODULE LOCATION #3
PL13 8
C D5
D6 7
PL8 R8
L1
R12 6
R9 C1
R11 SK1
L2 5
D9
LP1 C6 D7
FS7 D10 4
D8
C7
IC2 R5 3
IC1 C5
OFF ON 2
R10 R6 FREE
R4 FS6
C8 R13 R7 BUS 1
SW1
L N E 1 2 3 4 5 6 7 8 9 10 11 12 1 2 3 4
PL9
A A
Page 2-32
Installation
Chapter 2
AMERICAN SUPPLY
SK10
EYELET FOR CONNECTION L
L TO ENCLOSURE STUD 1
N
N 2
E
E 80 MM LONG 3
THIS CABLE IS SUPPLIED FITTED MATES WITH PL10
WITH A MAINS PLUG TO SUIT THE ON THE GAUGEPORT
CONNECTOR AT THE CUSTOMER'S SITE. OR OIU MOTHERBOARD
Page 2-33
TM710
User's Manual
WIRING SCHEDULE #2
GAUGE CABLE
20 METRES (MAX)
RED
1 1
BLUE
2 2
GREEN
3 3
YELLOW
4 4
WHITE
5 5
BLACK
1
6 6
BROWN
7 7
VIOLET
8 8
SCREEN ORANGE
PL1, MATES WITH SK1 9 9
ON THE GAUGE PINK
1 0 10
CD-0510 TURQUOISE
1 1 11
GREY
1 2 12
SK11
(MATES WITH PL11 ON THE
GAUGEPORT MOTHERBOARD
Page 2-34
Installation
Chapter 2
WIRING SCHEDULE #3
SAMPLE DISPLAY UNIT TO OPERATOR INTERFACE or GAUGEPORT
BOTTOM
PCB
J1 J2 RIBBON
1 234 1 2
J3
2 0V
1 +24 V
CD-0460
4 LONWORKS A (OUT)
3 LONWORKS B (OUT)
2 LONWORKS A (IN)
1 LONWORKS B (IN)
CD-0512
SK12
1
MAXIMUM LENGTH = 20 M
2
3
4
REF SIGNAL CABLE COLOUR
1 +24 V RED
2 0V GREEN
3 710 NETWORK B BLUE SK12 MATES WITH PL12 ON THE
4 710 NETWORK A YELLOW GAUGEPORT OR OIU MOTHERBOARD
Figure 17 - Wiring Schedule #3 (part) : The Sample Display Unit with integral connection
cable. SK12 connects with PL12 in the GaugePort unit or the Operator Interface Unit
Page 2-35
TM710
User's Manual
WIRING SCHEDULE #3
SAMPLE DISPLAY UNIT TO OPERATOR INTERFACE or GAUGEPORT
Page 2-36
Installation
Chapter 2
WIRING SCHEDULE #4
GAUGEPORT (or OPERATOR INTERFACE) to
GAUGEPORT (or OPERATOR INTERFACE)
PL9 ON MOTHERBOARD
IN SECOND GAUGEPORT
PL9 ON MOTHERBOARD OR SECOND OIU
IN FIRST GAUGEPORT
OR FIRST OPERATOR
INTERFACE
8
7
710 DATABUS OUT
6
5
710 DATABUS IN
CD-0513
SK9 SK9
MATES WITH PL9 MATES WITH PL9
ON FIRST ON SECOND
MOTHERBOARD MOTHERBOARD
Page 2-37
TM710
User's Manual
WIRING SCHEDULE #5
HAND HELD INTERFACE to GAUGEPORT UNIT (or OPERATOR INTERFACE)
SK1 PL1
Pin 1 Pin 1 LonWorks A
Pin 5 Pin 5 0V
Pin 6 Pin 6 LonWorks B
Pin 9 Pin 9 + 24 V d.c.
Pins 2,3,7 Note that these pins are wired for RS232 on the
GaugePort and Operator Interface connector.
Pins 4,8 Note that these pins are not connected at the
GaugePort and Operator Interface connector..
YELLOW YELLOW
HEATSHRINK HEATSHRINK
SLEEVING SLEEVING
Figure 19 - Wiring Schedule #5 : The cable for connecting the Hand Held Interface to the
GaugePort and Operator Interface
Page 2-38
Installation
Chapter 2
WIRING SCHEDULE #6
GAUGEPORT or OPERATOR INTERFACE to PC SERIAL PORT
Pins 1,5,6,9 Note that these pins are wired for use by the Hand
Held Interface when connected to SK1.
Pins 4,8 Note that these pins are not connected at the
GaugePort and Operator Interface connector.
Figure 20 - Wiring Schedule #6 : The cable for connecting the GaugePort and Operator
Interface to the serial port of a PC
Page 2-39
TM710
User's Manual
WIRING SCHEDULE #7
INBUILT ANALOGUE OUTPUT CHANNELS IN OPERATOR INTERFACE
SK4
TO THE USER'S 4-20 mA 1
MEASURING EQUIPMENT.
REFER TO THE CONFIGURATION 2
DIAGRAMS
3
4
5
CD-0517
6
7
8
Figure 21 - Wiring Schedule #7 : The cable for connecting to the in-built Analogue Outputs
Page 2-40
Operation
Chapter 2
WIRING SCHEDULE #8
INBUILT ALARM OUTPUTS IN OPERATOR INTERFACE UNIT
NOTE
The relays are normally configured “Normally Open”
SK2
1
2
TO CUSTOMERS EQUIPMENT 3
CABLE SCREEN 4
5
CD-0681A
6
7
8
Figure 22 - Wiring Schedule #8: The cable for connecting to the in-built Alarm Outputs
Page 2-41
TM710
User's Manual
WIRING SCHEDULE #9
ANALOGUE OUTPUT NODE (4-20 mA), FITTED MODULE
TO CUSTOMER'S SK1
4 - 20 mA INPUT EQUIPMENT SCREEN 1
CABLE
* 2
3
TO CUSTOMER'S 4
4 - 20 mA INPUT EQUIPMENT CABLE
*
5
SCREEN 6
CD-0514
* REFER TO POSSIBLE
CONFIGURATIONS DIAGRAMS
SK1 (MATES WITH PL1 ON THE
ANALOGUE OUTPUT NODE PCB
Figure 23 - Wiring Schedule #9 : the cable for connection to an Analogue Output Node PCB
Module when fitted in the GaugePort and Operator Interface
Page 2-42
Installation
Chapter 2
NOTE
The relays are normally configured “Normally Open”
SK2
1
2
TO CUSTOMERS EQUIPMENT 3
CABLE SCREEN 4
5
CD-0681
6
7
8
Figure 24 - Wiring schedule #10 : the cable for connection to an Alarm Output Node PCB
Module when fitted in the GaugePort and Operator Interface
Page 2-43
TM710
User's Manual
J3, Pin 1 Customer’s equipment Relay contact pair Clean (no-volt) contacts
closed when sampling
J3, Pin 2 Customer’s equipment
BOTTOM
PCB
J1 J2 RIBBON
1234 12
J3
1 2
CD-0460A
} SAMPLING
INDICATOR
TO J3 SCREW TO CUSTOMER’S
CONNECTOR EQUIPMENT
1 ON SDU PCB CONNECTOR TO SUIT 1
Page 2-44
Installation
Chapter 2
CD-0512A
TO CUSTOMER’S
LENGTH TO SUIT EQUIPMENT
1
Page 2-45
TM710
User's Manual
PL2 XL1
IC2
IC5
1
SK1
IC3
Figure 28 - The Alarms Output Node PCB for use in the TM710 system
3
FS2
4 IC3
5
IC5
6
SK1
5 4 3 2 1
IC1
IC10
Figure 29 - The Analogue Output Node PCB for use in the TM710 system
Page 2-46
Installation
Chapter 2
PL1
}
1 +15V
V+ LOAD 1 2 CURRENT SINK #1
0V 3 0V
}
4 +15V
V+ LOAD 2 5 CURRENT SINK #2
0V 6 0V
ALTERNATIVE LOAD ANALOGUE OUTPUT NODE PCB #1
CONNECTION IS SHOWN
AS DOTTED BOX
CD-0433AB
PL1
}
LOAD 1
1 +15V
2 CURRENT SINK #1
3 0V
}
LOAD 2
4 +15V
5 CURRENT SINK #2
6 0V
CD-0433CD
Page 2-47
TM710
User's Manual
Page 2-48
Operation
Chapter 3
CHAPTER 3
OPERATION
CONTENTS
Page
INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
OPERATOR INTERFACE (and hand held interface) . . . . . . . . . . . . . . . . . . 3-5
Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
RUN PAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Product selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Gauge selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Data entry page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
MENU PAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Sample page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Contrast page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Diag (Diagnostics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
CONFIGURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Channel and Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
%Deadband and Trend FRQ - Explained . . . . . . . . . . . . . . . . . . . . . . 3-20
%Weight explained . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Setting the passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Analogue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
BCD OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
CALIBRATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
In use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Trim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Autotrim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
RECIPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Editing an existing recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Creating a new recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Deleting a recipe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
PRODUCT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Edit an existing product. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Creating a new product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
WORKED EXAMPLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
CONDITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
METHOD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
Page 3-1
TM710
User's Manual
CONTENTS (continued)
Page
ILLUSTRATIONS
Figure Page
Page 3-2
Operation
Chapter 3
ILLUSTRATIONS (continued)
Figure Page
Page 3-3
TM710
User's Manual
Page 3-4
Operation
Chapter 3
INTRODUCTION
This chapter describes the operation of the following items of the TM710
system
The operation of the Operator Interface and the Hand Held Interface are
almost identical so the following description applies to both.
NOTE
The following descriptions assume that the system has been correctly installed
and interconnected as described in Chapter 2 of this manual.
The two Operator Interface and the Hand Held Interface each have a
touch sensitive display panel, refer to Figures 1 and 2. This means that,
besides functioning as a display panel for the displayed outputs, it also
functions as a key pad for data entry and selection where the keys are
redefined for the particular page being displayed at that time.
The display panels are LCD types with l.e.d. backlighting for easy
readability. Functions and their values are contained in a hierarchical
set of pages accessed through the touch screen keys. A flow diagram
showing this structure appears in Figure 3.
Touch Screen
The touch screen layer on the display screen comprises 36 cells arranged
on a 6 cell x 6 cell matrix having a 2:1 aspect ratio. The active cells in this
layer are re-defined according to the image appearing on the display
where the active cells are those sited directly above the image of a key.
Page 3-5
TM710
User's Manual
FIRSTMIX BLEND29
CD-0644
MOISTURE MENU
12.52
MOISTURE MENU
CD-0644
CD-0646
CD-0645
Figure 1 - Shows the Operator Interface Figure 2 - Shows the Hand Held Interface
Initialization
NOTE
If a fault is detected on any network device then when the RUN page is entered the
fault will be reported on a pop-up fault page and show the nature of the fault and
the name of the device reporting the fault. This may be cleared by the operator to
return to the RUN page, or the operator can go to the Diagnostic facility for
further details.
At this time the Operator Interface display panel shows the Welcome
page. After about 15 seconds when the internal checks have been
completed by the gauge and no errors detected, the Operator Interface
display panel shows the RUN page.
Page 3-6
Operation
Chapter 3
DIAG
SAMPLE CONTRAST CONFIGURE CALIBRATE RUN
SELF TEST
TO SET/SELECT TO RETURN TO LAST BOOT
TO START TO INCREASE OR
1. GAUGE NAME THE RUN PAGE MOTOR DATA
AND STOP DECREASE THE
2. MEASUREMENT TEMP
TAKING A CONTRAST OF THE
PARAMETERS WINDOW
"SAMPLE" DISPLAY PANEL
3. DISPLAY LAMP DATA
MEASUREMENT
4. PASSWORDS
5. LANGUAGE
6. ANALOGUE
7. BCD ENABLE/DISABLE
Figure 3 - Flow diagram showing the page structure used in the Operator Interface
Page 3-7
TM710
User's Manual
RUN PAGE
-32.14 4
6 CONSTITUENT MENU 5
CD-0600
The key to the elements of the RUN page are listed below:
(2) GAUGE NAME key. This is the gauge name for the measurement
being displayed, which also acts as a touch screen key to select
other gauges for display. If there are only two gauges in the
network, this will toggle between the gauges. If there are more
than two, then a pick list of gauges is displayed, allowing any
gauge in the network to be selected. Touch the gauge name to
select that gauge. When selected, the first measurement channel
for that gauge will be displayed as in '1' as well as its
corresponding measurement name as in '6' and current product
recipe name as in '3'.
NOTE
The gauge name displayed will be the gauge serial number that is embedded in
the gauge. The displayed name may be changed by the operator as required - refer
to Configure later in this chapter.
Page 3-8
Operation
Chapter 3
(3) PRODUCT RECIPE NAME key. This is the Product Recipe name for
the current measurement being displayed, which also acts as a
touch screen key to select other product recipes that have been
set up for the gauge being displayed.
Touch this key to present the Product selection page, showing all
the names of all the Product recipes that have been This will
enable the User to change the Product recipe to another.
Touch this key to display a pick list of the products loaded in the
Operator Interface pre-configured for that gauge. Touch the key
displaying the name of the product you wish to download to the
gauge. The screen beeps and displays DOWNLOAD and each recipe
name as it is being downloaded. When the download has been
successfully completed the screen reverts to the previously
displayed RUN page
(5) MENU key. This is a touch key that selects the main MENU page
which gives access to the functions of the system that is fully
described in the following pages.
The four measurement display has the same functionality as the single
measurement display except that the Gauge name, Product name and
GAUGE NAME
GAUGE NAME
Page 3-9
TM710
User's Manual
GAUGE NAME
PRODUCT NAME
MENU
ENABLE
RUN
CD-0706
Figure 7 - The Sub-menu page activated with the MENU key on the 4 display screen
Main Menu are now viewed and accessed through a sub-menu. The
sub-menu is displayed by touching the MENU touch key (in Figure 6)
which displays the screen in Figure 7. The functions are as follows:
(1) The operation and actions of the GAUGE NAME and PRODUCT NAME
keys are identical to the functions for the single measurement
display mode described above.
(2) Touch the MENU key to switch to the primary MENU page. Touch
RUN to return to the four measurement display mode.
The ENABLE key has a toggle action, that is to say, the next time
that the ENTER key is touched the measurement is displayed as
the screen display returns to the RUN PAGE.
(4) RUN key. Touch the RUN key to return to the current RUN DISPLAY
page.
Product selection
To view the Product recipes loaded touch the PRODUCT RECIPE NAME key
on the RUN page. This will cause the Product Selection page to be
displayed as is shown in Figure 8.
Page 3-10
Operation
Chapter 3
The Product Selection page shows the names of the Product recipes for
the named gauge, which can be selected by the operator as a touch key.
The up/down arrow touch keys are used to page the names up and down.
PRODUCT #1
PRODUCT #2
PRODUCT #3
PRODUCT #4
PRODUCT #5
PRODUCT #6 RUN
CD-0603
Gauge selection
To view the list of all the gauges connected touch the GAUGE NAME key on
the RUN page. This will cause the Gauge Selection page to be displayed
as is shown in Figure 9.
This page will show the names of all the gauges on the network, which
can be selected by the operator to show its measurements on the RUN
page. The up/down arrow touch keys are used to scroll the page up and
down.
When a Gauge is selected, the display will then revert to run mode (RUN)
page displayed), displaying the first measurement channel for the newly
selected Gauge.
Page 3-11
TM710
User's Manual
GAUGE NAME #1
GAUGE NAME #2
GAUGE NAME #3
GAUGE NAME #4
GAUGE NAME #5
CD-0661
This data entry page is used to enter both text and numeric data for
configuration and recipe set up. It operates in a similar fashion to a
mobile phone:
(1) This is the data entry line. It can be up to 20 characters long, and
shows the current entry by the operator from the touch keys
below the horizontal line. The data line will be initially shown
blank and will be filled in from the left.
1 MOISTURE_
1% 2 abc 3 def
2
C 4 ghi 5 jkl 6 mno
3 7 pqrs
wx
8 tuv 9 yz
4 RUN ENTER - * 0. #
CD-0604 5 6 7
Figure 10 - The Data Entry Page (DEP) for entering text and values
Page 3-12
Operation
Chapter 3
touch the right arrow to advance the cursor to the right. Enter
the next character and so on. Press ENTER to accept the data.
(2) The Left/Right touch key arrows are used to move the cursor over
a character to edit a correction. Edit operates in insert/delete
and not overwrite mode.
(3) The 'C' (cancel) touch key performs the delete function on the
character prior to the cursor position. A long touch results in the
data entry line being completely cleared.
(4) The RUN touch key will take the operator back to the RUN page
(abort operation).
(5) The ENTER touch key will load the alphanumeric entry and take
the operator back to the previous level.
(6) The S key allows the special characters found in foreign language
alphabets, ( for example 8 q Ç ) to be included in the descriptive
names and passwords entered via DEP keypad. Each press of
the S key scrolls through the total list of special characters and
displays the character at the cursor position. To accept the
displayed character press the right-arrow key to advance the
cursor to the next position (or ENTER if it is the last character).
(7) For the key pointed to in Figure 10, press once to enter the
number 4. Press twice for alpha G, three times for H and so on.
Press n times for the nth character, which will include special
signed alpha characters, as found on a typical mobile phone. This
action is cyclic.
(8) This is the space character for word separation. Note that the
Right-Arrow (to the right of “C”) key does not introduce a space.
MENU PAGE
As can be seen from the flow diagram in Figure 1, the MENU page has five
options plus the return to RUN page. Touch the MENU key to view the MENU
page as shown in Figure 11.
The six touch keys from top to bottom have the following functions:
(1) The SAMPLE feature enables the operator to collect a sample from
the production line and to get corresponding average
measurement values and standard deviations (qualifier) from
the Operator Interface.
Page 3-13
TM710
User's Manual
SAMPLE
CALIBRATE
CONFIGURE
CONTRAST
DIAG
RUN
CD-0605
(2) The CALIBRATE touch key gives access to the TRIM, AUTOTRIM,
PRODUCT and RECIPE features.
(5) The DIAG key calls the diagnostic function which displays the
results of a range of internal checks that have been carried out
(6) Touch the RUN key to return directly to the RUN page.
The following pages describe the functions of the keys on the MENU page
in the following order: (1) Sample Page (2) Contrast Page,
(3) Diagnostics Page, (4) Configure Page (5) Calibrate Page
Sample page
The SAMPLE start page is shown in Figure 12. The SAMPLE function
enables the operator to collect a sample from the production line and
obtain corresponding average measurement values and standard
deviations (qualifier) from the Operator Interface display.
To take a sample reading with a particular gauge carry out the following
procedure.
Page 3-14
Operation
Chapter 3
(2) Touch the key having the name of the gauge you wish to use.
(3) Touch the START key to start the sampling in the selected gauge
for all its available measurements.
When the STOP key is pressed, the gauge will stop sampling and
return all the sample averages and standard deviations for the
available measurements to the Operator Interface where they
are displayed, see Figure 14.
START
RUN
CD-0606
AVERAGE S.D.
CHANNEL 2 CONSTITUENT
CHANNEL 3 CONSTITUENT
CHANNEL 4 CONSTITUENT
RUN
CD-0659 CD-0660
Page 3-15
TM710
User's Manual
At any time touch the RUN key to abort the SAMPLING procedure and
return to the RUN PAGE.
Contrast page
Touch the INCREASE key to increase the contrast of the display by a fixed
increment. Continue to touch the key to further increase the contrast of
the display.
Touch the DECREASE key to decrease the contrast of the display by a fixed
increment. Continue to touch the key to further decrease the contrast of
the display.
INCREASE
DECREASE
RUN
CD-0607
Diag (Diagnostics)
Touch the DIAG key on the menu page to display the diagnostics page.
This is shown in Figure 16.
Touch the UPDATE key to re-run the internal tests and update the
information on the screen.
Page 3-16
Operation
Chapter 3
NAME WO 123456
CX-0093
NOTE
The AUTO-REF function should only be used with an ARS fitted to the Gauge
otherwise the function can corrupt the measurement.
CONFIGURE
The CONFIGURE PAGE display is shown in Figure 17. The functions of the
six touch keys shown are described on the next page in the order they
appear on screen.
Gauge
The GAUGE key allows the operator to select a gauge from a number of
gauges on the network and name or change the name of the connected
gauge.
CHANNEL MOISTURE
DISPLAY 1
LANGUAGE ENGLISH
ANALOGUE RUN
CD-0608
Page 3-17
TM710
User's Manual
To change the name of the gauge from either the pre-configured name
(which is the Serial Number of the gauge) or a previously entered name
carry out the following procedure.
Touch the GAUGE key. The GAUGE EDIT page is displayed as shown in
Figure 18 .
GAUGE WO-000000/0
BACK
CD-0711
(1) If more than one gauge is connected to the network, touch the
GAUGE key to list the connected gauges with their serial number
(or name) displayed so that they can be identified.
(2) Touch the key that contains the name of the required gauge.
(3) The screen returns to the GAUGE EDIT page with the selected
name alongside the GAUGE name key
(6) Press ENTER to accept and return to the GAUGE EDIT page (or RUN
to cancel the entry.)
Page 3-18
Operation
Chapter 3
There are four entities in the gauge called 'channels', These software
channels are designated CHANNEL #1 to CHANNEL #4. The pre-delivery
configuration of the gauge assigns each constituent in the product that is
to be measured (there can be up to four) to a CHANNEL number. For
convenience in the illustrations that follow, these have been abbreviated
to CH.#1 to CH.#4.
The CHANNEL key allows the operator to view and amend the
measurement (the screen output) configuration details.
When the CHANNEL key is touched the CHANNEL page is displayed. This
contains a list of up to four constituents as shown in Figure 19. Touch the
key of the constituent of interest to display the second level CHANNEL
page, Figure 20. This page enables the operator to re-name (and
automatically store in the Operator Interface) the renamed key-legend;
it also allows the number of decimal places (DECIMALS) in the displayed
measurement to be set, the % DEADBAND and TREND FRQ parameters be
set and the dry-weight conversion function (%WEIGHT) to be enabled or
disabled if the gauge has been supplied with this option installed.
The second level CHANNEL page allows the following settings to be carried
out. Note that when each of the NAME key, %DEADBAND key and the TREND
FRQ key are touched the DATA ENTRY page is presented.
%WEIGHT WET
BACK BACK
CD-0609 CX-0081
Figure 19 - The 1st level CHANNEL page Figure 20 - The 2nd level CHANNEL page
Page 3-19
TM710
User's Manual
The arrow that appears on the RUN PAGE alongside the measurement
value is an indication of any long term trend in the process variable
being measured, ignoring the affects due to short term variation in
process. If the indicator is shown for a prolonged period of time in the one
direction, this would be a good indication that the process is out of
control.
For example if the trend frequency (TREND FRQ) is set to 5 seconds and
the %DEADBAND is set to 100%. The average of all the measurements
made over two consecutive periods would be calculated and compared. If
the second average deviates by more than 100% from the first average,
the indicator will be shown in the direction of the trend.
(3) Press the %DEADBAND key to set this parameter (100% max.).
Type in the number on the DATA ENTRY PANEL presented.
(4) Press the TREND FRQ key to set the time interval (60 seconds
maximum) used for the measurement trend indicator.
Page 3-20
Operation
Chapter 3
%Weight explained
IMPORTANT
The gauge must be calibrated for the “wet” product over the advised range of
moisture content. The %Dry Weight is derived from the measured %Wet Weight
through a known relationship in this range.
(5) Press the %WEIGHT key to change the status of the dry-weight
conversion function.
(7) The BACK key takes the operator back to the previous page.
The default settings are %DEADBAND - 100%, and TREND FRQ - 5 seconds.
Display
For the display configuration shown in Figure 4, the run values setting
is 1 so the single measurement display mode will be entered when the
RUN PAGE is next displayed.
Page 3-21
TM710
User's Manual
Passwords
The password feature has been included to maintain the security of the
gauge settings where only personnel who have knowledge of the set
passwords may access the features to change existing settings.
The process engineers only have access to the settings under PRODUCT,
CONFIGURATION and CALIBRATION, as they only have password
level 1, whereas the operators only have access to the settings under
PRODUCT and CALIBRATION since they only have password level 2.
Note that both groups of engineers have access to the PRODUCT and
CALIBRATE settings.
Each time the OPTION key is pressed it steps through the functions shown
in the inset list. Touch the OPTION key the number of times required
until the function you wish to protect appears alongside the OPTION key.
Page 3-22
Operation
Chapter 3
CONFIGURE
OPTION FUNCTION CALIBRATE
TRIM
AUTO-TRIM
1 PRODUCT
CFG.PRDCT
RECIPE
CHANNEL
PASSWORD NO / YES GAUGE
BCD INPUT
DIAG.
PASSWORDS
2
PASSWORD NO / YES
BACK
CD-0704
There are two password levels, level 1 and level 2. The upper key labeled
PASSWORD is associated with the level 1 password and the lower key
labeled PASSWORD is associated with the level 2 password. These keys
set YES or NO alongside the password key that applies to the currently
selected function displayed alongside the OPTION key.
The square keys numbered 1 and 2 when touched begin the password
entry procedure by presenting the DATA ENTRY PANEL.
(1) Use the OPTION key and step through the list of functions.
(2) At each function, touch the level 1 and/or level 2 PASSWORD keys
to set either YES or NO as required for each level.
(3) Touch the number 1 key and set the password in the DEP.
(5) Repeat from step (3) using number 2 key instead of number 1
for the level 2 password.
Page 3-23
TM710
User's Manual
Language
The LANGUAGE key leads to the LANGUAGE PAGE where the operator is able
to select the language that will be used to display all text appearing on
the Operator Interface and SDU displays. The choice is between either
US English and one other language. The other languages are:
Touch the LANGUAGE key to toggle between English and the other loaded
language.
Analogue
The ANALOGUE key on the CONFIGURE page provides the operator with the
means to scale the 4-to-20 mA outputs available. There are two analogue
pages as shown in Figure 22 and Figure 23.
BACK BACK
CD-0708 CD-0709
Figure 22 - The 1st level ANALOGUE Figure 23 - The 2nd level ANALOGUE
page page
For each constituent in the named gauge shown in the Analogue page in
Figure 22 there are two columns for the analogue output signals. The
first column displays the value assigned to the 4 mA output level and the
second column the value assigned to the 20 mA output level. Figure 22
represents the Analogue page for a gauge with four constituents loaded
where only channel #1 constituent has so far been set.
Note that the name of the constituent being edited is shown at the top of
the 2nd level ANALOGUE page.
Page 3-24
Operation
Chapter 3
NOTE
Only the gauge channels configured in the gauge will be displayed on screen.
To set the maximum and minimum levels carry out the following
procedure:
BCD OFF
The BCD OFF button informs the user that the BCD Input function has
been disabled. If the legend on this button reads BCD ON then the BCD
Input function is enabled. For full details of the BCD Input function
please refer to Annex B in this manual.
Caution
Run
Page 3-25
TM710
User's Manual
CALIBRATE
Overview
In use
The CALIBRATE key is the gateway to the gauge parameters that are being
used to determine the calibration curve of the constituents in the
product being measured. A set of such parameters are grouped and
stored as a recipe. Each recipe is given a name by which it can be called
up.
The CALIBRATE PAGE is entered when the CALIBRATE key on the MENU PAGE
is touched. The CALIBRATE PAGE is shown in Figure 24.
TRIM
AUTOTRIM
PRODUCT
RECIPE
RUN
CD-0612
Page 3-26
Operation
Chapter 3
Trim
Touch the TRIM key to display a pick list of gauges. Select (touch) the
gauge to be edited to display the TRIM EDIT page which shows the
calibration trim values currently being used by the gauge for each of the
available measurement channels in the gauge, see Figure 25.
GAUGE BAY 1
CHANNEL MOISTURE
ENTER
RUN
CD-0614
Figure 25 - The TRIM EDIT page for the gauge named BAY 1.
The operator is able to set a new trim value for any of those
measurements by touching the TRIM which causes the DATA ENTRY PAGE to
be displayed. The display shows both old and new values (left and right
columns respectively), so that the change can be checked before it is
entered.
Touch the ENTER touch key to save the new TRIM value in the selected
gauge and in the stored recipe within the Operator Interface. The screen
beeps and displays DOWNLOAD as the TRIM value is being downloaded to
each of the selected gauges.
Page 3-27
TM710
User's Manual
Autotrim
Touch the AUTOTRIM key to display a pick list of gauges. Select (touch)
the gauge to be edited to display the AUTOTRIM EDIT page which shows the
calibration AUTOTRIM values currently being used by the gauge for each of
the available measurement channels in the gauge, see Figure 26.
GAUGE BAY 2
CHANNEL NICOTINE
PERIOD 10
DESIRED 20.00
ENTER
RUN
CD-0615-TM
Figure 26 - The AUTOTRIM EDIT page for the gauge named BAY 2
Touch the PERIOD key. The DATA ENTRY PAGE is displayed for the operator
to type in the required averaging period for determining AUTOTRIM. Press
ENTER on the DEP to return to the above AUTOTRIM EDIT PAGE.
Touch the DESIRED key to display the DATA ENTRY PAGE and type in the
desired value to appear on the measurement output. Press ENTER on the
DATA ENTRY PAGE to return to the above AUTOTRIM EDIT PAGE.
Touch the ENTER key to start the averaging over the set time interval
(PERIOD). On completion, the average value will be compared with the
desired value and the difference corrected for by automatically adjusting
Page 3-28
Operation
Chapter 3
the Trim in the gauge and in the stored recipe within the Operator
Interface.
Touch the ENTER touch key to save the AUTOTRIM value in the selected
gauges and in the stored recipe in the Operator Interface. The screen
beeps and displays DOWNLOAD as the AUTOTRIM value is being
downloaded to each of the selected gauges.
RECIPE
Overview
Each constituent has its own set of pre-configured recipes of which one
will need to be loaded into the gauge channel before a valid
measurement can be made for a particular product.
All measurements that are on the one production line have their recipes
grouped together under the one Product recipe name, so that when a
Product recipe is selected all the measurements will be updated to reflect
the new product being manufactured.
The operator selects a gauge a pick list and a constituent from a scroll
list displayed alongside CHANNEL the key. On selection, the available
recipes for that constituent are displayed as a pick list. On selection of a
recipe the recipe parameters are presented. Any or all of the parameters
can be modified and saved to the same recipe name or to another named
recipe (up to a maximum of 20 alphanumeric characters) by the
operator.
The effect of modifying the recipe parameters Span, Trim, Alarm Hi and
Lo, Response Time and Algorithm are described in Chapter 4
Calibration.
The RECIPE PAGE, see Figure 27, is entered by touching the RECIPE key on
the CALIBRATE page.
Page 3-29
TM710
User's Manual
GAUGE LOC A
CHANNEL SUGAR
EDIT
NEW
DELETE
RUN
CD-0613-TM
(1) Touch the GAUGE key to select a gauge from the pick list that
appears if more than one gauge is connected to the network. Else
this key beeps but has no other associated action.
(2) Touch the CHANNEL key a number of times until the required
constituent is displayed alongside the CHANNEL key.
To edit an existing recipe carry out the following procedure, from (2)(a)
above.
(1) Touch the EDIT key to list the recipes associated with the
currently selected constituent.
The list of parameters that comprise the recipe are listed in two
pages (press the ARROW key to reach the second page), see
Figure 28 and Figure 29. Touch the BACK key on the 2nd
parameters list to return to the first parameters list.
Page 3-30
Operation
Chapter 3
SPAN 1.00
TRIM 0.00
ALARM HI 20.00
ALARM LO 10.00
CD-0617
Figure 28 - Page 1 of 2 of the list of the RECIPE PARAMETERS page presented when a
recipe is chosen (touched) from the list of recipes presented for editing
(3) Touch in turn each of the keys shown in Figure 28 and Figure 29
that are to be changed. Each key causes the DATA ENTRY PAGE to
be presented for input of the new parameter values in the normal
way.
NOTE The RESPONSE TIME units when entering at the DATA ENTRY
PANEL are tenths of a second ( 0.1 sec). To set a response time of
1 second, enter 10 at the data entry panel, [10 x 0.1 sec = 1 sec].
(4) To store the modified recipe in place of the old recipe touch the
SAVE key.
NOTE The old recipe settings will be lost, even if the new recipe
has a different name. This is understood when it is known that it
is the recipe storage location in the Operator Interface that has
RESP. TIME 10
ALGORITHM 0
BACK
CD-0618
Page 3-31
TM710
User's Manual
NOTE Take care to rename the recipe (using the NAME function)
else the old and modified recipes will have the same name. (This
is quite different to the way a PC carries out a SAVE AS
command, for example).
To create a new recipe carry out the following procedure, from (2)(b)
above.
(2) Touch in turn each of the keys shown in Figure 28 and Figure 29
that are to be changed. Each key causes the DATA ENTRY PAGE to
be presented for input of the new parameter values in the normal
way.
NOTE The RESPONSE TIME units when entering at the DATA ENTRY
PANEL are tenths of a second ( 0.1 sec). To set a response time of
1 second, enter 10 at the data entry panel, [10 x 0.1 sec = 1 sec].
(3) To store the new recipe touch the SAVE or SAVE AS key.
Deleting a recipe
(2) Use the up and down ARROW keys to view long lists of recipes.
Page 3-32
Operation
Chapter 3
PRODUCT
EDIT
NEW
DELETE
RUN
CD-0616
(2) Use the up and down ARROW keys to view long lists of products.
(3) Touch the key containing the name of the PRODUCT to be edited.
Page 3-33
TM710
User's Manual
PRODUCT #1
PRODUCT #2
PRODUCT #3
PRODUCT #4
PRODUCT #5
PRODUCT #6 RUN
CD-0619
RECIPE
CD-0620
(4) If required, touch the product name key and alter the name of the
product using the DATA ENTRY PANEL when presented.
(5) Touch the RECIPE key to display the list of the constituents
(channels) and assigned recipes in the gauge named on the
screen. See Figure 33.
(6) In multi-gauge systems touch the gauge key to display a pick list
of gauges on the network. Choose (touch) the required gauge
name.
Page 3-34
Operation
Chapter 3
BACK
CD-0622
(9) Repeat the procedure from step (6) for each gauge associated
with the product.
(10) Touch the BACK key to return to the Product Recipe definition
page, Figure 32.
RECIPE #1
RECIPE #2
RECIPE #3
RECIPE #4
RECIPE #5
RECIPE #6 RUN
CD-0622B
Page 3-35
TM710
User's Manual
CONSTITUENT IN CH.#3
CONSTITUENT IN CH.#4
BACK
CD-0621A
(11) To store the modified (son) product in place of the old (father)
product touch the SAVE key.
NOTE The father product settings will be lost, even if the son
product has a different name. This is understood when it is
known that it is the product storage location in the Operator
Interface that has been overwritten with the modified settings.
(This is quite different to the way a PC carries out a SAVE
command for example).
(12) To store the modified (son) product in addition to the old (father)
product touch the SAVE AS key.
(1) Touch the NEW key (refer to the PRODUCT page Figure 30).
(2) Touch the NAME key to present the DATA ENTRY PAGE and type in
the chosen name of the Product ( say BLEND-01 ). Touch ENTER and
view the Product page as shown in Figure 37.
Page 3-36
Operation
Chapter 3
NAME
RECIPE
CD-0620B
NAME BLEND-01
RECIPE
CD-0623
(3) Touch the RECIPE key to display the list of the constituents in the
currently selected gauge, see Figure 38.
(4) In multi-gauge systems touch the gauge key to display a pick list
of gauges on the network. Choose (touch) the required gauge
name.
Page 3-37
TM710
User's Manual
CONSTITUENT IN CH.#1
CONSTITUENT IN CH.#2
CONSTITUENT IN CH.#3
CONSTITUENT IN CH.#4
BACK
CD-0621B
RECIPE #1
RECIPE #2
RECIPE #3
RECIPE #4
RECIPE #5
RECIPE #6 RUN
CD-0622B-2
CONSTITUENT IN CH.#3
CONSTITUENT IN CH.#4
BACK
CD-0621C
Page 3-38
Operation
Chapter 3
(7) Repeat the procedure from step (6) for each gauge associated
with the product.
(8) Touch the BACK key to return to the completed New Product page
as previously shown in Figure 37.
To download the product to the gauge touch the PRODUCT key on the RUN
PAGE and touch the PRODUCT NAME key of the product you wish to
download from the pick list of products displayed. This is fully described
in the RUN PAGE description earlier in this chapter.
WORKED EXAMPLE
CONDITIONS
The task is to set the gauge to measure a new product called Blend029
that contains the three constituents listed below. The gauge is to be
called Exit Dryer.
The Alarm Levels are to be set at +/-10 units about the target values
which are 10, 20 and 30 units for Moisture, Sugar and Nicotine
respectively.
The %DEADBAND value required is 30% and the TREND FRQ period is to be
15 seconds.
The gauge has been supplied with the following hardware set-up.
Page 3-39
TM710
User's Manual
(2) The names of the constituents have been loaded into the gauge
as: (a) M-01, (b) S-01 and (c) N-01.
METHOD
(1) From the RUN PAGE touch the MENU key to display the MENU PAGE.
DISPLAY 1
LANGUAGE ENGLISH
CX-0079 CD-0711A
Figure 43 is displayed.
CX-0080
Page 3-40
Operation
Chapter 3
The before and after screens for the Moisture channel are shown
in Figure 45 and 46.
When all data for the first channel has been input, touch BACK to
return to the list of constituents and then select the next
constituent and repeat the above procedure for Sugar and
Nicotine. The before and after screens for the Sugar and Nicotine
DECIMALS 1 DECIMALS 2
%DEADBAND 20 %DEADBAND 30
BACK BACK
CX-0082 CX-0083
DECIMALS 1 DECIMALS 2
%DEADBAND 20 %DEADBAND 30
BACK BACK
CX-0084 CX-0085
Page 3-41
TM710
User's Manual
DECIMALS 3 DECIMALS 1
%DEADBAND 20 %DEADBAND 30
BACK BACK
CX-0086 CX-0087
(3) From the CALIBRATE PAGE touch the RECIPE key to display the
RECIPE PAGE, see Figure 51
EDIT EDIT
NEW NEW
DELETE DELETE
RUN RUN
CD-0633 CD-0634
(4) Touch the CHANNEL key a number of times to read the required
constituent name alongside this key.
(5) Touch the NEW key to enter the NEW RECIPE PAGE. This is shown in
Figure 53 ( and Figure 54. The latter is reached by touching the
ARROW key).
Page 3-42
Operation
Chapter 3
TRIM 0.00
ALARM HI 0.00
ALARM LO 0.00
CD-0635 CD-0636
Figure 53 - Parameters for CAL-M1 recipe Figure 54 - More parameters for CAL-M1
(6) Enter the values as shown where they are different from those
shown in Figures 53 and 54 by touching the appropriate key and
typing in the value/name using the DATA ENTRY PAGE when it is
presented. Touch BACK to return to the previous page.
(7) Repeat the procedure for the other two constituents using the
name CAL-S1 and CAL-N1 for the Sugar and Nicotine
constituents respectively.
(3) From the CALIBRATE PAGE touch the PRODUCT key to display the
PRODUCT PAGE, see Figure 55.
(4) Touch the NEW key to display the NEW PRODUCT PAGE, as shown in
Figure 56.
EDIT NAME
NEW RECIPE
DELETE
CD-0639A CD-0620B
Page 3-43
TM710
User's Manual
SUGAR
NICOTINE
CD-0620C CD-0642-TM
(5) Touch the NAME key to display the DATA ENTRY PAGE and type in
the name of the Product followed by the ENTER key, to display
Figure 57. In this example the product name is BLEND29.
(6) Touch the RECIPE key to view the assignment page, Figure 58.
SUGAR CAL-S1
A list of recipes for this
constituent is displayed (not NICOTINE CAL-N1
illustrated).
BACK
(8) Touch the required recipe, in
this case it is CAL-M1.
CD-0643-TM
12.52
The display should now be
as in Figure 58.
Page 3-44
Operation
Chapter 3
(13) Touch the BLEND29 key in the list of products. The screen displays
DOWNLOAD as the data is sent to the gauge and the Operator
Interface beeps as it is doing so. At the end of the download the
display returns to the RUN PAGE and displays the measurements
of the new product as shown in Figure 60.
The Sample Display Unit (SDU) has an l.e.d. backlit liquid crystal
display (LCD) displaying two lines of 20 characters per line. It is capable
of displaying up to four measurements simultaneously. There are three
membrane input keys below the display area that are used for
controlling the unit. The unit is shown in Figure 61.
The default display at power up is shown in Figure 61, where the SDU
displays the measurements received from the network from four
channels. These channels are
labeled CH1, CH2, CH3 and CH4. To
the right of each label there are six
> C H 1 2 6 . 5 C H 2 9 . 2 character spaces to contain the
measurement output.
C H3 1 3 2 . 7 C H 4 8 2 . 5
The SDU has two main modes of operation : the RUN DISPLAY mode and
the MENU mode. The latter mode provides access to the following
Page 3-45
TM710
User's Manual
RUN DISPLAY
> C H 1 - 1 2 . 5 C H 2 3 5 . 2
C H3 2 . 4 C H 4 - 1 3 . 4
CD-0648
Figure 62 - A typical RUN display with the cursor shown in top left position
exception of values less than 1, where one leading zero will be shown (i.e.
0.23). The display layout is shown in Figure 62. The character locations
in the display are indicated by dotted boxes in this illustration.
Press the UP key to move the cursor forward. At each press of the UP key
the cursor jumps to the next display value ( next will be in the space
immediately in front of the label CH2, then the space before label CH3 and
then the space before label CH4 and next CH1.) When the DOWN key is
pressed the direction of cursor movement is reversed.
MENU
To display the MENU page press the SELECT key to show the display in
> S A M P L E C ON F I G
E N A B L E N A ME Q U I T
CD-0649
Figure 63. The menu option is selected by pressing the UP or DOWN key
until the cursor is situated alongside the chosen option, and to then
press the SELECT key to enter the function. A description of the elements
of this menu is given below.
Page 3-46
Operation
Chapter 3
Sample
The SAMPLE feature enables the operator to collect a sample from the
production line and get a corresponding average measurement and
standard deviation (qualifier) data from the SDU.
(1) Use the UP and DOWN arrows to set the cursor alongside the
SAMPLE function on the menu
At the end of the DELAY period the SDU commands the gauge
(over the network link) to start sampling. This is seen by the
SDU display changing to the SAMPLING display as shown in
Figure 65.
The length of time assigned for taking the sample can be set by
the operator in the CONFIG options using the function called
PERIOD (see CONFIG). The sampling period may be prematurely
terminated by pressing the SELECT button a second time before
PERIOD times out.
P R E P A R E 0 . 1 0 M I N
CD-0651
Figure 64 - The time decrements over the period set by the value in DELAY
S A M P L I NG . . 0 . 1 0 M I N
CD-0653
Figure 65 - The display on the SDU during Sampling - the time decrements to zero
Page 3-47
TM710
User's Manual
C H 1 2 . 4 - 0 . 4 s
C H2 5 . 0 1 . 2 s
CD-0654
The results from the sampling period will remain on the display
until the SELECT key is once again pressed whereupon the unit
returns to the RUN DISPLAY.
Config
To display the CONFIG menu press the UP or DOWN key to position the
cursor alongside CONFIG. Press the SELECT key to display the sub-menu
shown in Figure 67.
> D E LA Y P E R I O D
GA U G E Q U I T
CD-0650
(1) The DELAY function. This allows the operator to set the time that
elapses (in minutes and seconds.) from the time when the SELECT
key is pressed when sampling is selected in the MENU display,
until the time that sampling begins. This time should be set to
allow the operator time to reach the sample location after the
SELECT key is pressed to start sampling.
To set the DELAY time, use the UP or DOWN arrow keys to move the
cursor to the DELAY option and then press SELECT.
Page 3-48
Operation
Chapter 3
S A MP L E D E LA Y 0 . 1 0
CD-0655
Press the UP (or DOWN) arrow key to increment (or decrement) the
current delay time. The time is incremented (or decremented) in
steps of 1 second per key press.
(2) Press SELECT to return to CONFIG sub-menu with the set time
stored.
(3) The PERIOD function allows the operator to set the duration of the
sampling period in minutes and seconds.
To set the PERIOD time, press the UP or DOWN arrow key to move
the cursor to the PERIOD option and then press SELECT.
S A MP L E P E R I O D 0 . 1 5
CD-0656
Press the UP (or DOWN) arrow key to increment (or decrement) the
current PERIOD time. The time is incremented (or decremented)
in steps of 1 second per key press. The display in Figure 70
results after one DOWN key press (decrement).
S A MP L E P E R I O D 0 . 1 4
CD-0656A
Page 3-49
TM710
User's Manual
(4) The GAUGE function. Place the cursor at the GAUGE option on the
CONFIG display page and press SELECT.
(5) The QUIT function. Place the cursor at the QUIT option on the
CONFIG display page to return to the MENU page.
Enable
(1) On the RUN DISPLAY page set the cursor at CH2 measurement and
press the SELECT key.
(2) Set the cursor at the ENABLE option on the MENU page.
C H 1 - 2 . 5
C H3 1 3 . 4 C H 4 2 2 . 5
CD-0658
Figure 71 - The RUN DISPLAY page with CH2 measurement display inhibited
To re-instate a CH2 measurement that has been "disabled" from the RUN
DISPLAY page carry out the following procedure.
(1) On the RUN DISPLAY page set the cursor at the CH2 measurement
previously displayed position and press the SELECT key.
(2) Set the cursor at the ENABLE option on the MENU page.
Page 3-50
Operation
Chapter 3
Name
To modify the NAME of the measurement label carry out the following
procedure.
(1) On the RUN DISPLAY page set the cursor at the measurement
whose label is to be modified and press the SELECT key.
(2) Set the cursor at the NAME option on the MENU page.
(4) Press the UP (or DOWN) arrow key to step forward (or reverse)
through the character set until the required character is reached,
see Figure 72.
(5) Press the SELECT key to advance to the second character and
M H 3
CD-0657
Figure 72 - The NAME CHANGING page with the third measurement label being changed
repeat using the UP and DOWN arrow keys until the required new
character is reached.
(7) Press SELECT to exit to the RUN DISPLAY page with the new label
for the previously chosen measurement.
Page 3-51
TM710
User's Manual
FAULT MESSAGES
When a fault occurs in the system and is detected by the gauge, the
gauge broadcasts an updated fault code along the 710 databus. This
results in an error or warning message being displayed on the screens of
all devices (Operator Interfaces, Sample Display Units and Handheld
Interfaces) connected to the system. The exception to this is the LINK
ERROR fault message which is generated by the device unable to
communicate with the gauge, and in this case is displayed only on this
device.
When the condition causing the fault is removed the gauge issues an
updated fault code that removes the previously displayed fault message
from the screen.
Multiple faults are handled using a priority system where the fault
message of highest priority is displayed. When its fault condition is
removed the next highest priority fault message is displayed and so on.
On removal of the last fault message the screen reverts to the page of
information displayed prior to the display of the fault message.
Page 3-52
Calibration
Chapter 4
CHAPTER 4
CALIBRATION
CONTENTS
Page
INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
CALIBRATION FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Trim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Autotrim. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
CALIBRATION OVERVIEW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
SpeedCal (Autotrim) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Fine tuning at target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Full range calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
CALIBRATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Speedcal procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Fine tuning at the target procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Full range calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Additional Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
ILLUSTRATIONS
Figure Page
TABLES
Table Page
Page 4-1
TM710
User’s Manual
Page 4-2
Calibration
Chapter 4
INTRODUCTION
The factors affecting the calibration of the gauge are named : SPAN, TRIM,
AUTOTRIM and ALGORITHM. Should the operator wish to edit (re-calibrate)
the gauge for any reason, or create a new calibration for a product having
some or all of the same constituents, the values of SPAN and/or TRIM
and/or ALGORITHM may need to be changed. In addition the calibration
related function SAMPLE is required to be run to determine the
average-measurements made by the gauge.
CALIBRATION FUNCTIONS
Span
20
INCREASE IN
The SPAN function enables the SPAN VALUE
Page 4-3
TM710
User’s Manual
Trim 20
710 OUTPUT
applies a positive or negative shift
to the final output. The default 10
CD-0164A
value can be seen in Figure 2.
0 5 10 15 20
LABORATORY VALUE
Sample
NOTE
A useful tool to determine the calibration parameters is the NDC IE product
called GaugeView. This is a software package that runs on a PC linked to the
gauge using the RS232 comms link. The software calculates, recommends or
rejects the calculations, and when commanded downloads the calculated values
for SPAN, TRIM and ALGORITHM directly to the gauges requiring the new
values.
Page 4-4
Calibration
Chapter 4
CALIBRATION OVERVIEW
(1) SpeedCal
(a) Autotrim
(b) Fine tuning at target
SpeedCal
Autotrim
The result can be put into service as soon as the gauge is installed and
will take less than a few minutes to implement.
NOTE
The expected value is supplied by an operator who knows the value of the
constituent that is being measured for the product currently passing the gauge.
Page 4-5
TM710
User’s Manual
For some applications, both TRIM and SPAN values may need adjustment
to obtain correct measurement sensitivity and achieve direct
correspondence with the laboratory reference over the full range of
interest. This is likely to be necessary where agreement with an unusual
laboratory reference method is required.
The optimum calibration parameters are calculated from the gauge and
corresponding laboratory values. The calibrated output of the gauge is
calculated as follows:
where G is the calibrated gauge output and BASE is the base calibration
output ( i.e. with SPAN = 1.000 and TRIM= 0).
The Operator Interface, Hand Held Interface allow new TRIM and new
SPAN values to be downloaded to the gauge using the TRIM and SPAN
functions.
Page 4-6
Calibration
Chapter 4
CALIBRATION PROCEDURES
Initial Conditions
The gauge is installed and operating on-line and the Operator Interface is set at
the RUN DISPLAY page.
To run the AUTOTRIM function, carry out the following procedure at the
Operator Interface Unit.
The PERIOD is the length of time (in minute and seconds) that is
used to take and average the measurement of the constituent in
the product.
(7) Enter the value of the PERIOD. This should be set to 10 seconds, or
up to 20 seconds for products with high variation. Next touch the
ENTER key to return to the AUTOTRIM page.
(9) Enter the measurement expected from the gauge for the product
on line and then touch the ENTER key to return to the AUTOTRIM
page.
(10) Touch ENTER to start the measurement over the set time.
Page 4-7
TM710
User’s Manual
Touch RUN to return to the RUN DISPLAY page and read the measurement
value. This should be close to the DESIRED value previously entered.
The specific method of collecting the samples for laboratory analysis will
depend upon the product and measurement being made.
Initial Conditions
The gauge is installed and operating on-line and the Operator Interface is set at
the RUN DISPLAY page.
Page 4-8
Calibration
Chapter 4
(6) Throughout the sample period collect the product from the
process from a position just after, and in line with the
measurement point. Place this in a sealed container.
(8) Repeat the sample procedure at least four times. More samples
will improve the accuracy to which the gauge is calibrated.
T1 = T0 + ( labM - GM ) … (1)
where
Example
The results shown in Table 1 were recorded from samples taken for the
production process using a gauge with initial TRIM = 4.0.
Page 4-9
TM710
User’s Manual
2 3.75 4.50
3 4.40 5.30
4 4.05 4.80
5 4.10 4.90
6 3.60 4.90
Substituting these values in the equation (1) the new TRIM, T1 is given
by:
Initial Conditions
The gauge is installed and operating on-line with the product to be re-calibrated
selected and the Operator Interface is set at the RUN DISPLAY page.
(3) Touch TRIM. A pick list of gauges (if more than one gauge is
connected) is displayed. Touch the name of the gauge to be edited
to display the TRIM EDIT page.
(4) Touch the CHANNEL key a number of times until the required
measurement is displayed alongside the CHANNEL key. Note that
the set TRIM value changes for each measurement displayed. This
will be the measurement in the gauge that is modified when the
new TRIM value is entered and ENTER is touched.
Page 4-10
Calibration
Chapter 4
(5) Touch the TRIM key. The DATA ENTRY PAGE is displayed.
(6) Type in the new TRIM value that was calculated earlier and touch
ENTER to complete the data entry and return to the TRIM EDIT
page.
(7) Touch ENTER to send the new TRIM value to the gauge and save the
new TRIM value in the stored recipe in the Operator Interface.
NOTES
In normal circumstances the SpeedCal calibration will be all that is required.
However if the operator feels that a full calibration is required then the longhand
procedure described in this section be carried out.
By far the most convenient and easiest way to carry out a full range calibration is
to use the ‘calibrate feature’ in the GaugeView software as described in the
GaugeView User’s Manual.
Initial Conditions
The gauge is installed and operating on-line with the product to be re-calibrated
being measured by the gauge and the Operator Interface is set at the RUN
DISPLAY page.
Page 4-11
TM710
User’s Manual
(4) Record the collected data. Refer to the foregoing procedures for
suitable sample integration period, (t) and the method of
collecting samples, for each measurement type.
Y = mX + c
X X
710 OUTPUT
X X
X Y
X
X SLOPE (= m) = Y
X
X
INTERCEPT (= c)
CD-0167
LABORATORY VALUE
S 1 = S0 / m
Page 4-12
Calibration
Chapter 4
T1 = (T0 - c) / m
(11) Touch the GAUGE key to select a gauge from the gauge pick list
that is displayed if there is more than one gauge in the system,
else touching this key has no action.
(12) Touch the CHANNEL key a number of times until the required
constituent is displayed alongside the CHANNEL key.
(13) Options (a) and (b) result in the RECIPE PARAMETERS PAGE 1 being
displayed : (a) has current values of measurement displayed and
(b) has default settings displayed.
(14) For option (b) only, touch the NAME key and type in at the DATA
ENTRY PAGE the name to assign to the recipe.
(15) For either option, touch the SPAN key. The DATA ENTRY PAGE is
displayed.
(16) Type in the new SPAN value and then touch the ENTER key.
The display returns to the previous page with the new SPAN value
shown alongside the SPAN key.
(17) Touch the TRIM key. The DATA ENTRY PAGE is displayed.
Page 4-13
TM710
User’s Manual
(18) Type in the new TRIM value and then touch the ENTER key.
The display returns to the previous page with the new TRIM value
shown alongside the TRIM key.
(20) At RECIPE PARAMETERS PAGE 1 touch the SAVE key to save the
recipe under the name of the recipe.
(21) To save under a different name, touch NAME and type in the new
name in the DATA ENTRY PAGE and then touch the ENTER key.
(22) Touch the SAVE key to store this recipe in the Operator Interface.
Touch the EDIT key to view the recipe list and see that the new
recipe is included in this list.
Repeat the above procedure from stage (8) for each constituent recipe
requiring modification of SPAN and TRIM.
Existing products that have previously been assigned the recipes just
modified will download the new parameters to the gauge when the
product is selected to be measured.
Existing products may be modified and new products created using the
new recipes.
Additional Comments
Table 1 is provided as a template for the operator to copy for the purpose
of recording results from the gauge and laboratory that will be used to
determine the new TRIM and SPAN values according to the above
procedures.
Page 4-14
Calibration
Chapter 4
TM710
Sample number Laboratory result Gauge measurement
10
Averages
Page 4-15
TM710
User’s Manual
Page 4-16
Maintenance
Chapter 5
CHAPTER 5
CONTENTS
Page
INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
SAFETY PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
PREVENTATIVE MAINTENANCE. . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
CLEANING AND INSPECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Cleaning solvents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
GaugePort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Gauge Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Removal of the air-purge window housing . . . . . . . . . . . . . . . . . . . . . 5-7
Replacement of the air purge window housing. . . . . . . . . . . . . . . . . . . . 5-7
Cables and leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
DIAGNOSTIC MAINTENANCE. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
SELF DIAGNOSTICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
THE DIAGNOSTICS PAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Motor data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Filter wheel target speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Actual filter wheel target speed . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Mains frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Window contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Lamp data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
AUTO REFERENCE STANDARD . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Auto Reference (Auto-Ref) procedure. . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
CORRECTIVE MAINTENANCE. . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
SCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
GAUGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Source Lamp Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Removal of the Source Lamp Assembly . . . . . . . . . . . . . . . . . . . . . . . 5-17
Replacement of the Source Lamp Assembly . . . . . . . . . . . . . . . . . . . . 5-18
Filter Wheel Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Removal of the Filter Wheel Motor . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Replacement of the Filter Wheel Motor . . . . . . . . . . . . . . . . . . . . . . . 5-21
SPARES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
Page 5-1
TM710
User's Manual
ILLUSTRATIONS
Figure Page
TABLES
Table Page
Page 5-2
Maintenance
Chapter 5
INTRODUCTION
Certain tools and materials are required for effective and safe
maintenance of the TM710. The tools and materials required are listed
below.
5 Allen key driver for M3 cap head 2.5 mm A/F GaugePort and Operator Interface
power supply mounting bracket
Page 5-3
TM710
User's Manual
WARNINGS
Cautions
Page 5-4
Maintenance
Chapter 5
SAFETY PROCEDURE
(1) Turn off and isolate the mains supply to the GaugePort.
(2) Disconnect the cable that connects the GaugePort to the Gauge.
(3) Turn off and disconnect the air and water supplies to the Gauge.
PREVENTATIVE MAINTENANCE
At first, the items listed below should be checked often to determine the
periodicity of cleaning the equipment. The periodicity will vary
according to the level of ambient contaminants.
(4) GaugePort
Cleaning solvents
In all cases, check with the Customer Care Department of NDC Infrared
Engineering or their agent as to the suitability of the solvent and give
precise details of the solvent before using on the equipment. Obtain and
obey the safety procedures relevant to the substance/solvent selected.
Page 5-5
TM710
User's Manual
GaugePort
In most cases the GaugePort unit will be remote from the Gauge and
should not become excessively contaminated and can be cleaned as
follows.
(1) Clean all sides of the GaugePort unit using a damp cloth and
soapy water or a weak detergent.
(2) Wipe with a rinsed damp cloth and then dry off with a clean dry
cloth.
Gauge
Gauge Window
When the measured level reaches a preset upper threshold a fault signal
is generated by the gauge and the operator is notified of the nature of the
fault by a message automatically appearing on the Operator Interface
and Hand Held Interface screen. For a window contamination fault the
window should be cleaned to continue to receive accurate output from
the gauge.
Note also that a generated fault signal causes the status indicator on the
gauge (stainless steel version only) and on the GaugePort, to flash off
and on. For systems without a message screen the nature of the fault
must be determined, via the customer’s equipment connected to the 710
network.
NOTE
Page 5-6
Maintenance
Chapter 5
Caution
Do not remove the air purge window from the gauge head while the
sensing head is fitted on the production line. Take care when removing
the air purge window from the sensing head casing as it will separate
into three parts.
(1) Turn off the air and water supplies to the Gauge.
(2) Remove the air and water supply pipes from the Gauge.
(5) Take the Gauge to a clean room and place it on the work top with
the air-purge window uppermost.
(6) Use a 2 mm Allen key to evenly loosen the three grub screws
which secure the air purge window to the rim of the Gauge
window.
(7) The air purge window housing has a cylindrical insert and
rubber gaskets which come loose when the grub screws are
unscrewed.
(8) Firmly grip the inner cylinder and outer housing and lift off the
air purge window using two hands. Note how the component
parts of the housing fit together.
(9) Clean the Gauge window. Take care when cleaning the Gauge
window. Use only a soft clean lint-free cloth to clean the glass
window. (Refer to solvent paragraph above)
Page 5-7
TM710
User's Manual
(1) Place the Gauge on the clean workbench with the window
uppermost.
(2) Take the assembled air purge window and carefully seat it in
position on the bezel of the Gauge window.
(3) Using a 2 mm Allen key evenly tighten the three grub screws
around the perimeter of the air purge window to secure it to the
Gauge.
(4) Return the Gauge to its mounted position and fix it in place using
the previously removed fixings.
(5) Re-attach the cable to the Gauge and any other local cables that
have been disconnected.
Regularly inspect all cables in the system for signs of chaffing, kinking
and their general condition. Rectify if chaffing or kinking are seen. Clear
any debris or residue from the cables that could cause mechanical or
chemical damage to the cables.
Page 5-8
Maintenance
Chapter 5
DIAGNOSTIC MAINTENANCE
SELF DIAGNOSTICS
NOTE
Some messages have a “warning” fault type (indicated by a W suffix to the screen
message) and inform the operator of an impending error.
For example, TEMP. HI W warns that the temperature has reached a pre-set level
(below the maximum) whereas TEMP. HI E is an error and reports that the
temperature has exceeded the maximum allowed.
Table 2 - Fault screen messages and descriptions
FAULT: DUP GAUGE Two gauges on the network with the same
address
SELF TEST Warning Self Test warning
Page 5-9
TM710
User's Manual
The layout of the diagnostics page is shown in Figure 1. The values of the
data fields of the display are explained in the Table 2.
Touch the UPDATE key to re-run the internal tests and update the
information on the screen.
1 2 3
NAME WO 123456 4
CX-0092
Page 5-10
Maintenance
Chapter 5
3 SELF TEST The result of the internal self-test carried out by the gauge (PASS)
4 LAST BOOT The time since the gauge was last powered up in hours and minutes
Motor data
The target speed is set at the factory to be a set factor of the supplied
mains frequency. At 50.000 Hz mains frequency, this would result in a
target speed of 7385 rpm (revs per minute) and at 60.000 Hz a speed of
7680 rpm.
The actual speed is controlled to the target speed, within +/- 5 rpm of
that specified above. However, if the mains frequency is not sampled by
the gauge (due to a fault condition) the control speed will be set at the
default speed of 6857 rpm.
Mains frequency
Temperature
Page 5-11
TM710
User's Manual
displayed. When the temperature rises above 80°C (176°F) the lamp and
motor (main sources for internally generated heat) will be shut down.
They will automatically restart when the internal temperature falls
below 70°C (158°F).
Window contamination
Lamp data
The lamp data value is proportional to the current flowing through the
lamp and is typically at a level of 26963. If the lamp ruptures, this level
will fall to zero. The Lamp ON/OFF indicator shows the control status of
the lamp. The indicator shows OFF when the high temperature
threshold has been exceeded.
Page 5-12
Maintenance
Chapter 5
NOTE
If an Air Purge assembly is fitted to the gauge, It is NOT
necessary to remove the air purge assembly in order to fit
the ARS. When the Air Purge assembly is fitted, the
index markers on the bezel are partially obscured.
Figure 2 - General view
Attach the ARS to the Gauge as follows: of the ARS
(2) Carefully position the ARS window end over the gauge window
(3) Align the yellow dot on the ARS with the yellow line on the bezel
and feel that the two locating pins on the window bezel are
squarely located in the slots on the ARS.
NOTE
If this is not easily visible, align the yellow dot on the ARS centrally and
pointing towards the front face (opposite to the face fitted with the cable
connector).
(4) Twist clockwise (when viewed from knurled end of ARS) to lock
into position.
The value of the standard sample appears on the User Interface display
(normally an Operator Interface unit). Select the appropriate channel
and note the reading.
Page 5-13
TM710
User's Manual
Caution
If you require to run the AUTO REF function carry out the following
procedure.
(1) Ensure that the correct gauge is named alongside the gauge
name key on the DIAGNOSTICS page.
(3) Touch the AUTO-REF key. The question “Is there an ARS fitted ? “
is displayed.
(4) Answer as appropriate using the answer keys YES and NO at the
foot of the display. Do not answer yes if an ARS is not fitted.
CORRECTIVE MAINTENANCE
SCOPE
Corrective maintenance procedures are included here for the items that
have been designated Customer Replaceable Items. The following pages
describe the removal and replacement of the following items:
Page 5-14
Maintenance
Chapter 5
GAUGE
General
For each of the two versions of the TM710 Gauge, the electronic and
optical assemblies, are contained within a two-part enclosure (1) the
upper section and (2) the lower section.
The two sections are screwed together by means of six machine screws. A
gasket is fitted to the lower section by adhesive to provide a sealed
enclosure when the two sections are screwed together.
The upper section of the case contains the Gauge Main PCB assembly
(electronics and optics). The lower section contains only the viewing
window. There are no user replaceable items in the lower section.
WARNING
NOTE
The following procedures assume that the equipment has already been
dismantled from its plant location and has been brought to a clean room, and that
anti-static precautions have been implemented.
The source lamp assembly is mounted on the Mirror, Motor and Lamp
Block. Refer to Figure 3 for its location and fixings
Cautions
It is important that the glass of the source lamp assembly is not touched
or contaminated in any way. When handling the source lamp assembly,
hold the assembly lamp leads between the forefinger and thumb
immediately behind the backing plate.
Page 5-15
TM710
User's Manual
A REPLACEMENT MOTOR
IS SUPPLIED ALREADY
FITTED TO THE MOTOR MOUNT
BLOCK. THE EXPLODED VIEW
OF THIS ASSEMBLY IS SHOWN
FOR INFORMATION ONLY
MOTOR MOUNT BLOCK
MOTOR
MOTOR FIXING SCREWS
AND WASHERS (M2 X 4MM)
LAMP ASSEMBLY
CD-0447
LAMP ASSEMBLY HOLES FOR LAMP
MOUNTING POWER SUPPLY WIRES
SURFACE
Page 5-16
Maintenance
Chapter 5
Figure 4 - The TM710 Gauge. The six fixing Figure 5 - The case section containing the PCB
screws to be removed to separate the case assemblies. The upper PCB must be hinged open
sections can be seen to gain full access to the lamp and motor
Figure 6 - The source lamp assembly being Figure 7 - Shows the Motor Mount block fixing
withdrawn from its housing after the fixing screws screws being undone. When undone, unplug
have been removed and the leads unplugged the ribbon cable before removing the block
(1) Place the Gauge on the bench with the Gauge window
uppermost.
(2) Use a 4mm Allen key driver to undo the six machine screws that
hold the enclosure sections together (Figure 4).
Page 5-17
TM710
User's Manual
(4) Place the section without the PCB assembly to one side while the
other section is worked on.
(5) Refer to Figure 3, 5 and 6 and locate the source lamp assembly,
the lamp assembly fixings (2 off) and the lamp supply leads that
plug into the main PCB.
NOTE
The small PCB piggy-back fitted to the larger PCB must be hinged back as
described below to gain full access to the lamp assembly fixing screws.
(6) Support the case so that the source lamp assembly is positioned
for easy access.
(8) Use an Allen key to loosen the screws from the threaded spacers.
(9) Lift the board and hinge it fully open to gain access to the lamp
assembly fixing screws.
(10) Unplug the two source lamp assembly leads from the sensing
head PCB
(11) Use a 2 mm hexagonal screw driver and loosen the two captive
fixings that secure the lamp assembly to the mounting surface.
(12) Carefully withdraw the source lamp assembly from its mounting
taking care not to damage any surrounding circuitry.
To replace the source lamp assembly carry out the following procedure.
(2) Carefully insert the new source lamp assembly into the lamp
mounting bracket on the circuit board. Avoid touching the glass
envelope if not wearing lint-free gloves.
Page 5-18
Maintenance
Chapter 5
(3) Fasten the assembly to the bracket by means of the two captive
M3 screws and crinkle washers in the lamp assembly.
(4) Connect the two lamp leads to the sockets on the PCB.
(5) Return the hinged PCB to its previous position and carefully
screw the two captive fixing screws into the hexagonal pillars
below and tighten.
(6) Place the two sections of the case together and evenly tighten the
six machine screws using a 4 mm Allen key driver.
WARNING
Cautions
The filter wheel motor assembly is fitted on the Gauge main PCB
assembly. Removal of the motor requires the removal of the filter
motor/filter wheel assembly and for the Signal Processor PCB to be
hinged fully open.
Page 5-19
TM710
User's Manual
To remove the filter wheel motor, refer to Figure 3 and 7 and carry out
the following procedure.
(1) Place the Gauge on the bench with the Gauge window
uppermost.
(2) Use a 4 mm Allen key driver to undo the six machine screws that
hold the enclosure sections together.
(4) Place the section without the PCB assembly to one side while the
other section is worked on.
(6) Lift the board and hinge it fully open to gain access to the motor
fixing screws (2 off).
(7) Unplug the motor ribbon cable from the PCB at PL3.
(8) Use a 4 mm Allen key driver to loosen and remove the two M4 x
45 mm socket head screws, crinkle and plain washers, that
secure the motor/filter wheel assembly in its location. Retain the
fixings.
Page 5-20
Maintenance
Chapter 5
To replace the filter wheel motor carry out the following procedure.
(1) Take the replacement Motor Assembly and slide the filter wheel
fully onto the motor shaft with the bush containing the grub
screw furthest from the motor plate. Tighten the grub screw.
(2) Place the motor/filter wheel assembly in its final position and
secure in place using the retained M4 x 45 mm socket head
screws, crinkle washers and plain washers.
(3) Plug the motor ribbon cable socket into PL3 on the main PCB
assembly.
(4) Return the hinged PCB to its previous position and carefully
screw the two captive fixing screws into the hexagonal pillars
below and tighten.
(5) Place the two sections of the case together and evenly tighten the
six machine screws using a 4 mm Allen key driver.
Page 5-21
TM710
User's Manual
SPARES
The parts may be obtained by ordering the item by the part number
given below from:
Gauge
Lamp Assembly 64/7158-02SA
Motor Assembly 105/12151-01SA
Page 5-22
Reference
Chapter 6
CHAPTER 6
REFERENCE
CONTENTS
Page
COMPLIANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
GaugePort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Operator Interface Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Sample Display Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Hand Held Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Page 6-1
TM710
User's Manual
Page 6-2
Reference
Chapter 6
COMPLIANCE
EMC
EN50081-1 (1992) (Generic emission standard - Residential, commercial, and light industry), and
EN50082-2 (1994) (Generic immunity standard - industrial environment).
Low Voltage Directive
EN 61010-1 (1993) (Safety requirements for electrical equipment for measurement, control, and
laboratory use - general requirements).
UL Standard
UL 3111-1: (1994) 'Electrical measuring and test equipment part 1 - General Requirements'
Declaration of Conformity
Page 6-3
TM710
User's Manual
SPECIFICATION
GAUGE
Dimensions
Width 177 mm
Height 172 mm
Length 328.5 mm
Weight 7 kg
Connector removal clearance : 140 mm
Environmental sealing IP65 NEMA 4
Ambient Temperature Range
Storage 0º C - 70 ºC
Operating 0º C - 50 ºC
Cable length 20 metres maximum
Services
Air For the Air Purge Window A clean air supply at a flow rate of 75 litres per minute
+/- 10 litres per minute
Water For cooling the cast case A clean water supply at normal mains pressure
version when water adjusted to give a minimum flow rate of 2 litres per
cooling is required) minute.
Measurement Scope Four measurements simultaneously
Serial Communications RS232 Full duplex
Network LonWorks
Power Supply +24 V d.c. @ 1.3 A
Working Distance 250 mm +/- 100 mm
Sampling area 50 mm diameter
Ambient Light Rejection The mean Gauge output does not deviate for the
conditions given below :
Page 6-4
Reference
Chapter 6
GAUGEPORT
Dimensions
Length 270 mm
Width 230 mm + 25 mm (length of gland)
Height 133 mm
Note : Above values do not include cable clearance and connector removal
clearance
Weight 4.7 kg excluding connected cables
Environmental sealing IP65
NEMA 4
Enclosure Fabricated steel sheet with painted textured finish
Ambient Temperature Range
Storage 0º C - 70 ºC
Operating 0º C - 50 ºC
Relative Humidity 5% to 95% (non-condensing) over the full operating
temperature range
Power Supply input voltage range 80 V a.c. to 265 V a.c @ 50/60 Hz
Power consumed 50 watts maximum
Inputs / Outputs EXTERNAL CONNECTOR
(1) 9-way D-type socket on enclosure wired for both
(a) 710 Databus and
(b) RS232 connection
MOTHERBOARD CONNECTORS
(2) Mains input, PL10 on motherboard
(3) SDU, PL12 on motherboard
(4) Gauge, PL11 on motherboard
(5) User Interface, PL9 on motherboard
PANEL INDICATOR
(6) Gauge STATUS
Page 6-5
TM710
User's Manual
MOTHERBOARD CONNECTORS
(2) Mains input, PL10 on motherboard
(3) SDU, PL12 on motherboard
(4) Gauge, PL11 on motherboard
(5) User Interface, PL9 on motherboard
(6) Two Analogue Output channels, PL4 on motherboard
(7) Four independent Alarm Output channels, PL2 on
motherboard
PANEL INDICATOR
(8) Gauge STATUS
Page 6-6
Reference
Chapter 6
(3) +24 V d.c. Supply via cable through a cable gland from
the GaugePort or Operator Interface (usually a shared
cable with Input (2)
Page 6-7
TM710
User's Manual
Page 6-8
Reference
Chapter 6
Notes
Page 6-9
TM710
User's Manual
Notes
Page 6-10
DECLARATION OF CONFORMITY
Declaration
The above products comply with the requirements of the Low voltage Directive
73/23/EEC as amended and the EMC Directive 89/336/EEC as amended, by the
application of the following Transposed Harmonised European Standards.
For EMC
EN55022 1994 class B. EN61000-3-2, and EN61000-3-3 Generic emission
standard - Residential, Commercial and Light Industry
For LVD
EN61010-1 (1993) Safety requirements for electrical equipment for
measurement, control, and laboratory use - General requirements.
ATTENTION
The attention of the specifier, installer, or user is drawn to special measures and
limitations to use which must be observed when the product is taken into service to
maintain compliance with the above directives. Details of these special measures and
limitations to use are contained in the product manuals and are also available on request.
TM710
Temperature Measurement
Annex A
TM710 Temperature Measurement
Annex A
TM710
Temperature Measurement
CONTENTS
Page
SCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
SET UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Air purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Temperature Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fig Page
SCOPE
INTRODUCTION
Page 3
TM710 Temperature Measurement
Annex A
The received infrared signals are digitized and then processed by the 710
software and made available as one of the measurement channels from
the gauge.
The sensor is fitted in the base of the TM710 casing adjacent to the main
gauge-viewing window. When the gauge is mounted at the
recommended distance from the product (250 mm) the temperature
sensor receives infrared emissions from a 25 mm diameter area of the
product at a known location with respect to the main measurement
patch. This area is not illuminated by the gauge as is the main
measurement patch.
SET UP
The gauge is factory set and does not require further adjustment on site.
INSTALLATION
Page 4
TM710 Temperature Measurement
Annex A
75 47.5
TOP VIEW
WATER COOLING GS
0.25 INCH OUTSIDE
DIAMETER (2 OFF)
75
MOUNTING HOLES
M8 x 10 DEEP
SIDE VIEW
29
310 18.5 END VIEW
177
TM710 CONNECTOR
REMOVAL 172.5
CLEARANCE
230 140
101 (DIA)
ALL DIMENSIONS IN MM
34.5
AIR PURGE
76.5 TEMPERATURE
SENSOR
Page 5
TM710 Temperature Measurement
Annex A
Air Purge
To clean the viewing face of the temperature sensor, first remove the air
purge body by unscrewing the retaining nut whereupon the viewing face
will be exposed. Use a lint-free cloth to carefully wipe the surface clean
and then refit the air purge body.
10
TM7
20 DEGREES
TEMPERATURE
MEASUREMENT
CX-0046
PATCH SIZE = 25 MM
(NOT VISIBLE)
Figure 2 - Gauge viewing distance and angle for the TM710 fitted with Temperature Sensor
76.5 MM
25 MM DIAMETER
TEMPERATURE
BEAM PATCH
34.5 MM
CX-0045
50 MM DIAMETER
MAIN BEAM PATCH
Figure 3 - Relative positions of the main measuring beam patch and the temperature sensing
beam patch (not visible to the naked eye) viewed from above
Page 6
TM710 Temperature Measurement
Annex A
The air supplied to the air purge window must be clean and dry. If it is
not it will contaminate the viewing face of the sensor and affect the
measurement being made.
It is suggested that the air supply to the temperature sensor air purge is
obtained from a tapping on the air supply to the air-purge window fitted
to the gauge viewing window using a T-junction fitment.
OPERATION
Temperature Calibration
Set the gauge to measure the temperature of the product and follow the
CALIBRATE /Trim section in Chapter 3 - Operation / Calibration of the
TM710 User’s Manual, and adjust the TRIM value so that the gauge
displays the temperature value observed on the reference temperature
measurement equipment.
MAINTENANCE
Page 7
TM710 Temperature Measurement
Annex A
SPECIFICATION
Temperature Sensor
Beam patch size 25 mm diameter
Working distance 250 mm +/- 100 mm
Measurement Range 0 °C to 120 °C
Accuracy +/- 1 °C
Repeatability +/- 0.5% of reading or +/- 0.5 °C whichever is the
greater over the full temperature range
Air supply required 10 litres per minute minimum, clean, dry, instrument
quality air.
Measurement output One of the four available measurement channels
Page 8
TM710
Remote Recipe Selection
Annex B
TM710 Remote Recipe Selection
Annex B
TM710
Remote Recipe Selection
CONTENTS
Page
SCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Recipe address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
PRODUCT RECIPES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Selecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Deleting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Adding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
ILLUSTRATIONS
Fig Page
SCOPE
Page 3
TM710 Remote Recipe Selection
Annex B
The annex provides a brief description of the BCD Input equipment and
the additional information required to install and operate the gauge
where it differs from that described in the TM710 User’s Manual.
DESCRIPTION
Introduction
The BCD input is an 8-bit wide parallel digital input to the Operator
Interface that enables recipes to be automatically and remotely selected
based on a Binary Coded Decimal code. The code is generated by a plant
control system such as a Programmable Logic Controller (PLC ). This
configuration allows any one-of-eighty customer-configured recipes to be
selected.
Hardware
Page 4
TM710 Remote Recipe Selection
Annex B
The BCD signals are fed to the connector on the Touch Screen Board
Assembly via a cable which passes through one of the glands in the
Operator Interface enclosure.
Software
The BCD Input feature is introduced with version 1.05 of the software
installed in the Operator Interface. The BCD Input function software is
present regardless of whether the BCD Input hardware is installed or
not. When the latter is not installed the equipment is supplied with the
BCD Input function disabled by a password protected software setting.
Caution
Recipe address
Example
Say for example there are a number of product recipes in the list of
Product Recipes (01 to 80). If the value read at the BCD coded input
changes from a valid number to a new and different number (let’s say
“10”), then if product recipe “10” is in the list, then this recipe is selected.
If product recipe “10” is not in the list then the read value is ignored.
Page 5
TM710 Remote Recipe Selection
Annex B
INSTALLATION
Connector Function
Pin Number
Multiplier Weighted
value
1 1
2 2
Units
3 (x 1) 4
4 8
5 1
6 2
Tens
7 (x 10) 4
8 8
9 0 V (spare)
10 0V
To connect the BCD Input cable from the plant to the 710, carry out the
following procedure:
(1) Remove the lid from the Operator Interface to access the BCD
Input connector. The connector can be reached through the
cut-out in one end of the protective cover fitted over the Touch
Screen Board Assembly and Combi Board, see Figure 2.
(2) If necessary fit a gland in the chosen entry position for the BCD
Input cable.
(3) Prepare the cable by cutting back the outer sleeving sufficiently
to expose adequate lengths of the inner wires to reach the
connector.
(5) Pass the cable through the gland and tighten the gland with
sufficient length of cable to reach the BCD input connector when
the lid is re-fitted.
Page 6
TM710 Remote Recipe Selection
Annex B
(6) Identify the individual cores of the cable and fit each to the
connector according to Table 1 (and Figure 1), including the 0V
reference to either pin 9 or pin 10 on the connector.
(7) Replace the lid and tighten the recessed captive fixing screws.
(8) Route the cable to reach the plant equipment that will provide
the BCD coded signals (Figure 3).
PERFORATED COVER
CX 0048
Figure 2 - Location of the BCD input connector and pin numbering details
Page 7
TM710 Remote Recipe Selection
Annex B
CX 0054
Figure 3 - The cable from the plant control system to the BCD Input connector
OPERATION
In order to use the BCD Input function, the Operator Interface must be
fitted with v1.05 software or later, and have the necessary hardware
fitted, see page 4 of this annex. The version number appears on the
screen as the system powers up. From v1.05, the software features the
BCD STATUS button on the main CONFIGURATION PAGE as shown in Figure 4.
The BCD STATUS button displays as either BCD ON or BCD OFF.
When the BCD STATUS button is set to BCD OFF, the product recipe can be
selected in the usual way. However, when it is set to BCD ON, the product
recipe can only be changed using the BCD Input function.
CHANNEL MOISTURE
DISPLAY 1
PASSWORDS BCD ON
LANGUAGE ENGLISH
ANALOGUE RUN
CX 0049
Figure 4 - The CONFIGURATION PAGE when the BCD INPUT feature is installed
Page 8
TM710 Remote Recipe Selection
Annex B
CX 0055
PRODUCT RECIPES
Selecting
The PRODUCT RECIPE selection page that is displayed when the BCD
Input is enabled (from software v1.05) is shown in Figure 6. It contains
the product recipe names, each with an associated number to the left of
the name. These numbers (01 to 80) represent the address of each recipe
in the list.
To select the correct recipe, the control system providing the BCD input
must have knowledge of this list and the associated address. For
example, for the control system to select a recipe from this list, the
address of the required recipe must be supplied to the BCD Input. For
the screen displayed in Figure 6, to select PRODUCT RECIPE #E the control
system must send “05” to the BCD Input.
Page 9
TM710 Remote Recipe Selection
Annex B
ADDRESS IN LIST
1 PRODUCT RECIPE #A
2
3 PRODUCT RECIPE #B
#C
3
4 PRODUCT RECIPE #C
#D
4
6 PRODUCT RECIPE #D
#F
5 PRODUCT RECIPE #E
CX-0050
Figure 6 - The PRODUCT RECIPE selection page displayed for software from v1.05
Deleting
When a recipe is deleted from the list, the list closes up. If for example
PRODUCT RECIPE #B, and PRODUCT RECIPE #E was deleted from the list, the
display will change to that shown in Figure 7 (assuming only the recipes
displayed in Figure 6 are loaded).
ADDRESS IN LIST
1 PRODUCT RECIPE #A
3 PRODUCT RECIPE #C
4 PRODUCT RECIPE #D
6 PRODUCT RECIPE #F
RUN
CX-0056
Page 10
TM710 Remote Recipe Selection
Annex B
Adding
When recipes are added to a list of previously loaded recipes, the added
recipes will occupy the highest available address in the list and the list
will expand. This is illustrated in the following example.
Let’s assume that the following recipes are to be added to the list shown
in Figure 7 in the following order: PRODUCT RECIPE X, PRODUCT RECIPE Y,
PRODUCT RECIPE Z. When this has been carried out and the recipe list is
next displayed, the display of the first six recipes will then be as shown
in Figure 8. PRODUCT RECIPE #Z will have address” 7” and can be selected
by presenting a value of “7” at the BCD Input. This entry in the list can
be viewed on screen by pressing the down-arrow button.
ADDRESS IN LIST
1 PRODUCT RECIPE #A
2
3 PRODUCT RECIPE #X
#C
3
4 PRODUCT RECIPE #C
#D
4
6 PRODUCT RECIPE #D
#F
5 PRODUCT RECIPE #Y
CX-0057
MAINTENANCE
SPECIFICATION
As well as the inputs listed in Chapter 6 of this manual, when the BCD
Input is installed the following additional input specification applies.
Page 11
TM710 Remote Recipe Selection
Annex B
Page 12