1.0 Introduction
This guide outlines requirements and considerations when planning the site for installation of the
Selenia System.
There are many considerations when selecting a location for the Selenia System. Included in these
considerations are the following:
• location for patient throughput
• space for patient and medical personnel
• input power source and sufficient wall outlets
• environmental concerns
• shielding
• state and local requirements.
• Seismic requirements (see “Seismic Requirements” on page 17.)
P/N 9-600-0100 1
Site Planning and Pre-Installation Guide
P/N 9-600-0100 2
Site Planning and Pre-Installation Guide
P/N 9-600-0100 3
Site Planning and Pre-Installation Guide
3.0 Specifications
3.1 Tubestand (Gantry with C-arm)
Height: 190 cm (75”) ± 1 cm; 213 cm (84”) at highest C-arm travel
Width: 65 cm (26“) ± 1 cm (C-arm at 0° position)
Depth: 110 cm (43”)± 1 cm
Horizontal Rotation: 138 cm (54”)
Weight: 300 kg (660 lbs) ± 50 kg (110 lbs)
P/N 9-600-0100 4
Site Planning and Pre-Installation Guide
P/N 9-600-0100 5
Site Planning and Pre-Installation Guide
P/N 9-600-0100 6
Site Planning and Pre-Installation Guide
Softcopy Workstation
Power Requirements UPS, Domestic:
88 to 148 VAC (120 VAC nominal)
50/60 Hz ± 3 Hz auto sensing
NEMA 5 - 15P input connection type
UPS, ROW:
175 to 295 (230 VAC nominal)
50/60 Hz + 3 Hz auto sensing
IEC-320-C14 input connection type
Computer:
90 to 135 VAC at 50/60 Hz
180 to 265 VAC at 50/60 Hz
Monitors:
90 to 264 VAC, 50/60 Hz
P/N 9-600-0100 7
Site Planning and Pre-Installation Guide
P/N 9-600-0100 8
Site Planning and Pre-Installation Guide
Caution: Provide strain relief for the Mains cable: shall not flex, move, etc. The system is considered
permanently connected according to its safety classification.
P/N 9-600-0100 9
Site Planning and Pre-Installation Guide
P/N 9-600-0100 10
Site Planning and Pre-Installation Guide
P/N 9-600-0100 11
Site Planning and Pre-Installation Guide
Appendix A
Selenia System
Operating Required Actual Comments
Environment
Avg. Temp. 20° C to 30°C(68° F to 86° F)
Avg. Humidity 10% to 80%
P/N 9-600-0100 12
Site Planning and Pre-Installation Guide
Operator
Patient
Room
Acquisition
Required Actual Comments
Workstation
Input Line Voltage 100-240 VAC
3.5 Amps max. @ 100/120 VAC
Input Current
2.0 Amps max @ 220-240 VAC
Hardcopy Film
Required Actual Comments
Printing Devices
Power 90-250 VAC, 47-63 MHz
2.0 Amps @ 115 VAC
Operating Current
1.5 Amps @ 220 VAC
P/N 9-600-0100 13
Site Planning and Pre-Installation Guide
P/N 9-600-0100 14
Site Planning and Pre-Installation Guide
P/N 9-600-0100 15
Site Planning and Pre-Installation Guide
P/N 9-600-0100 16
Site Planning and Pre-Installation Guide
P/N 9-600-0100 17
Site Planning and Pre-Installation Guide
P/N 9-600-0100 18