Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
SD0000012
IMAGING ARTIFACTS
Comparative Sludy in
X- Ray Ct & Medical Ultrasound
Kha\\cVMoh&VY\ecl
MayVAohamed W\edan\
W\ahas\n Fad\ KWa
NU2H\\ Saved
Supervised by.
* Sad\Q KA. Tam
July 1998-Khartoum
I
2. CONTENTS.
1.2 INTRODUCTION.
Contents
A emem
To pur imehersrm the DiplmM leva! how.
in Sudan,
My B9S Khartoum
Chapter One
Research Proposal:
Imaging Artifacts: A Comparative Study
In X-Ray CT & Medical Ultrasound
Introduction:
This study draws attention forwards the quality of imaging process.
Concerning the factors that may observe the diagnosis out come which
represents the aim of the whole process of imaging technology. Imaging
artifacts is one of the major limitation factor that after and decrease the
value of diagnosis and out come in conventional radiography we can
estimate the factors that cause artifacts which could be easily evaluated
because most of the parameters in imaging process are fixed to some
extend and the accessing of the imaging approximately in direct mode.
But the errors which leads to artifacts skill arise in digital imaging
"Concerning CTV U/S", the accessing process taken over several steps
involve conversion from analogue format to digital and vice versa; and
application of computer technology program needs careful awareness.
Tiny errors many arise artifact, issue, errors leads to artifact, on digital
imaging can't predicted unless listed and discussed on a cord of practices
according to the operational protocol.
Statement of problem:
The digital imaging nowadays cover most of investigations done in
radiography department ever the conventional process started earlier and
cover along of time and the technologist gain experience skill errors lead
to artifacts seem to be more while digital imaging replace the conventional
imaging in a fast steps it carry also the same errors with the new errors,
that may warser the new practice as a whole.
Reasons for Choice This Project:
There were a lot of errors in different department that deals with
digital imaging can be observed by any general observer this issue
reflected on the fetal report at the level of diagnosis and this will affect the
quality of patient cane and may be the morbidity rate as well as mortality
rate.
Objectives of the Project:
The main objective of this study is to high light the artifacts and the
source of errors that lead to artifact formation and we can summarize the
main objectives on the following points:
(1) Consider and identify the artifacts on each modality.
(2) Show the cause of artifact.
(3) Resolve and eliminate the reasons that lead to artifacts
formation in imaging.
(4) Establish proper quality control techniques issue set tha base
line limits of artifact causation.
Hypothesis:
1- The presence of artifact creates problems in diagnosis.
2-The cause errors leading to imaging artifact that affect the
diagnosis more in digital U/S than CT.
3- Lack of quality assurance programs maximizes the presence of
faults leading to imaging artifact.
4- Imaging artifacts affects the quality of patient care and delay
medical diagnosis and treatment, and hence increase patient
waiting time.
Previous Studies:
From what we read in the previous research studies (1,2,3) we did
notice that there was just listing of faults without indicating any reason (s)
or causes except in medical ultrasound. However, we planned to research
it from our locate experts in CT and medical ultrasound in the teaching
hospitals and medical centres according to the values of the code of
practice.
Research Methodology:
In this study of have taken the scientific method in the field of X-ray
CT and medical ultrasound imaging.
Data Collection:
1-Questionnaire.
2- Interviews.
Using both in the following variable:
a. Equipment factors.
b. Patient factors.
c. Technique.
Place & Period of Study:
Sudan- Khartoum March/July 1998 (Teaching Hospitals and Medical
Imaging Centers).
Introduction
Diagnostic imaging is an important element in the practice of the
modern medicine without it medical treatment would have been
impossible. Radiation medicine had seen many changes since 1895, where
only early tuals. A radiation therapy and radoidiagnosis. Today imaging
has many source of electromagnetic radiation "EMRs" and each source
has its advantages over the other. However, medical radiation education
had also seen more advances in the medical sciences and medical
education, which lead to the discoveries of new methods, and techniques
in these fields.
There were also new approaches to refine and perfect the practice of
radiologic techniques and to reduce radiation exposure, patient waiting
time and cost. We have seen lots of new imaging modalities, quality
assurance and quantity control procedures in all modalities to produce
images of high quality and more diagnostic information.
Here, we have taken the lead to research and study carefully how we
could participate in the area of reducing image retakes and study the
reasons of image artifacts in diagnostic imaging.
In this regaid we are researching on artifacts in X-ray computed
tomography and medical ultrasound and considering how these artifacts
may be eliminated.
PART 10W
2.1.1 TECHNIQUE
2.1.2 EQUIPMENT
2.1.3 PATIENT
'ffifflf '* V'f$"*'/lbfff&$*S* $ **f * .•'*'">• y V f '
8
Artifacts
This section deals with experience gained with the original model
with EMI CT 1010 scanner and other scanning systems, is expected. This
section should be of considerable value in the understanding of the basic
causes of artifact production and sources of error despite such
modifications.
Artifacts:
Motion and high-differential attenuation value of adjacent tissues are
principal cause of artifacts. Technical errors are also responsible.
Motion Artifacts:
Motion of a point places that point different computed positions
during the scanning cycle. This false representation usually produces a
linear artifact called a "streak". Point A scanned at 0 degree is again
scanned at 180 degrees with an interval time delay. A change in point A
position will produce computer error in the form of a streak. Computer
error due to movement will, therefore, be more marked along the vertical
(Odegree/180degrees) line, because of motion during the time that elapses
between the two readings. The head can rotate in all directions, producing
a variety of streaking outside or inside the skull vault. Assuming that the
scanning time is constant, head support, patient cooperation and sedation
are factors to consider in order to diminish motion artifact. The faster
scanning cycle of the newer CT systems reduces the degree of motion
artifact. The EMI CT 1010 scanner has a 240-degree scanning angle
option designed to decrease streaking that result from patient motion.
10
Source of Error:
Technical source of error includes improper selection of window
level and window width, improper Polaroid printing and improper patient
positioning. Several other situation are also potential sources of error,
including "partial-volume" phenomenon and incorrect measurement of
surface area. The partial volume phenomenon is the effect of matrix
averaging of heterogeneous tissues. Partial volume phenomenon leads to
misinterpretation, such as spurious "hydrocephalus" Postpneumeonce-
phalography. It also leads to Vat and volumetric measurement error,
particularly of small highly contrasting regions. Use of the higher detail
matrix systems, combined with thin-section technique, diminishes this
averaging problem. Magnification factors for measurement must be
accurately calculated in order to avoid misrepresentation of size. This
calculation should based on data obtained by scanning a phantom of
known dimension, or by measurement of structure of known size that
appears on the CT scan.
2.12
TECHNIQUE
Technical Artifacts:
Figure (2)
llasilnr Impression, Prominent Dens, and
Jugular Tubercles
Figure(3)
IP-
PART TOW
1.2
is
Figure (1!)
Figure ( 1 2 )
Figure ( 1 3 )
16
Figure (14)
Artifacts
Figure ( 1 9 )
i9
PART TOW
2.1.3
Figure(5)
Figure(6)
Motion Artifact-streak
Motion Artifact-streak
Figure(8)
Figure(9)
3.1
3.1.1
EQUIPMENT
Sonogram Abbreviations:
BI: Bladder
D: Diaphragm
E: Echoes
GBI: Gallbladder
K: Kidney
L: Liver
P: Pleural effusion
T: Tornado effect
Ut: Uterus
FIGURE X ° Interference from nearby equipment causes artifacts on the CRT (arrow).
Figure (20): Interference from nearby equipment causes artifacts on the CRT
(arrow).
Calibration Problems-Incorrect
Distance Markers
riOUBE 400 Main bang artifact With older units this is caused by a strong interface between
skin and the transducer. Too much near gain (arrow) can also be a reason
Figure (22): Main bang artifact. With older units this is caused by a strong interface
between the skin and the transducer. Too much near gain (arrow) can also be a reason
Veiling:
Bands of increased echogenicity can be seen at certain depth if all
focal zones are used simultaneously, producing the veiling artifact (figure
23).
DIAGNOSTIC CONFUSION The impression of a mass may be created
within the area of veiling. Masses may be overlooked at the interface of
the different focal zones.
FIGURE J^TVeiling. Focusing zones are well delineated transverse echo areas (arrows). Utilize
only the focusing zone in the area of interest to eliminate the focal banding.
Figure (23): Veiling Focusing zones are well-delineated trs isverse echo areas
(arrows) Utilize only the focusing zone in the area of interest to eliminate the focal
banding
J27
A bsence of Focusing:
Electronic focusing and the use of acoustic lenses have increased the
number of focal zones available with a single transducer and greatly
increased the resolution of the image. If the focal zone option is not used
with newer electronic systems much blurring of echo interfaces is seen
(Figure 24).
Figure (24): Absence focusing. There is blurring of the echoes when focusing is not
utilized (arrow)
28
Focusing and Persistence versus Festal Heart Motion:
"Focusing" and "persistence" improve the quality of the image at the
expense of frame rate; a wavy motion across the image is commonplace if
these controls are used.
Figure (26): A grating artifact (arrow) may sometimes occur above or below a strong
linear interface ( e g the diaphragm) when using an array system, particularly a linear
array
Photographic Artifacts:
Figure (27): A
31
PART THREE
%Mm l a MM
3-2
Noise:
Side Lobes:
Side lobes are secondary echoes outside the main beam, that exist
with all transducers.
Figure (29): Dropout lines (arrows) are created when the scanning speed is too rapid
Operator Pressure:
Applying too much or uneven pressure while scanning can distort the
image.
Figure (30): Pressure artifacts. A. Too much pressures over the fetal trunk produces a
flattened ovoid shape. B. A. Lighter scanning pressure creates a round mink and
correct measurement
PART THREE
3.1.3
Patient
Artifacts from Strongly Reflective Structures (Shadowing):
Gas, bone, and, to a much lesser extent, muscle do not conduct sound
well. When sound strikes a strong interface such as gas or bone, one of
two responses may be produced. Either there is no sound conduction
through the area (shadowing), or numerous secondary reverberations are
produced, causing a series of echogenic lines extending into the tissues
(ring down).
Figure (32): Large acoustic interfaces due to gallstone are associated with shadowing
(arrow). Shadowing is accentuated with higher frequency.
Reverberation Artifacts:
Whenever sound passes out of a structure with an acoustic impedance
that is markedly different from its neighbor, a large amount of sound is
returned to the transducer. The amount of sound returning may be so great
that it is sent from the transducer back into tissues, causing s duplication
of the original structure. The second wave has traveled twice as far as the
first one, the third echo three times as far, and so fourth. The distance
between each successive echo will equal the distance between the original
two interfaces. The second echo and each successive echo parallel the
original interface.
Figure (34): Echoes due to reverberations are parallel to the anterior body wall
(arrow) of the bladder
Figure (35): Reverberations (r) from the body wall may be seen to
extend down into a renal cyst. B. when the scan angle is changed, the
reverberations are no longer seen within the cyst. (c).
Figure (36): Mirror anifact behind the posterior wall of the bladder with creation of
an apparent cystic lesion posterior to the bladder (arrow)
4-f
Transducer
Diaphragm
Figure (37): Mirror artifacts. A. In the normal patient there is a mirror image of the
liver tissue above the diaphragm at the site of the lung (arrow). B Diagram of how the
artifact is created. C. When there is a pleural effusion, an echo-free area is seen above
the diaphragm.
4-2
Enhancement Effect:
As the sound beam passes through fluid-filled structures or structure
containing many cysts, it is not attenuated and there is an increase in the
amplitude (brightness) of the echoes distal to the fluid (figure 38).
Figure (38): Enhancement effect. A. Increased echoes obscure the structures behind
the bladder owing to enhancement of the sound passing through the bladder (arrow). E
Decreasing the gain allows.the uterus to be seen clearly.
Frezonal Zone Artifacts:
The near field of the transducer contains artifactiual echoes.
RECOGNITION. Echoes are seen at the posterior aspect of the cyst and
develop as the transducer moves from the center of the cyst.
Slice thickness
Artifact echoes
Figure (41): Slice-thickness artifact. A. Echoes in the posterior part of the gallbladder
relate to the slice thickness artifact. The diagram shows the beam intersecting an
oblique segment of the cyst wall B. Sonogram demonstrates low level echoes
apparently in the posterior part of the gallbladder where the gallbladder angle is steep.
4-6
Comet Effect:
A very strong acoustical interface, such as an air bubble, or a metallic
structure, such as a suture, creates a dense echogenic line extending
through the image known as the comet effect (figure 42).
BENEFIT. The presence of the line indicates a very strong interface and
may allow recognition of metallic structures such as clips.
Figure (42): The comet effect is demonstrated on this longitudinal view of the liver.
At the diaphragm echogenic lines can be seen extending towards the lung (arrow).
Static Scanning Artifacts:
Figure (43): Misregistration. A. Transverse view of the right kidney scanning in one
direction. Note gas-filled bowel to the right of the kidney (arrow) and echoes from the
main bang artifact at the skin. B. Scanning the right kidney but angling from the other
direction. Note the distortion of the kidney borders due to misregistration (arrows). C.
There is no distortion of the right kidney. Registration is now in alignment. Note
irregular shadowing due to gas in bowel adjacent to the kidney.
Beam Depth Problem:
Artifacts are present beyond the focal zone of the transducer in the
far part of the field when there are beam depth problems (figure 44). The
echoes in this region are much coarser, and major lesions may be missed if
a long-focus transducer is not used. Because considerable lateral beam
spread occurs, small pinpoint structures appear as transverse 1ine,s.
DIAGNOSTIC CONFUSION: Subtle small lesions may be missed
because of the coarse echogenic structure at depth (e.g. small metastases
in the liver).
Figure (44): The near segment of this liver contains artifactiual information owing to
beam distortion in the frezonal zone (small arrow). Lateral beam spread beyond the
focal zone (large arrow) causes wide echoes with little information.
Compounding:
Often the best way to complete a B-scan is to form numerous small
sector scans to create one overall image. At the junction of the small
sector scans artifact is created because the transducer can be accurately
aligned only rarely (figure 45).
DIAGNOSTIC CONFUSION: The intersection of two sector scans can
be thought to represent a pathologic process.
Figure (45): compound scanning causes this artifact; the image is not aligned (left
arrow).
" ^"; > ^v; A : >;>-*. ^ \ - r
! ! - i .
Real-Time and Static Scanning Artifacts
Artifacts Caused By Equipment
Calibration Problems-Incorrect
Distance Markers Reduction:
Calibration checks should be performed frequently (once a month).
See chapter 49. Measurements should be performed in the center of the
image where calibration is most correct and not at the edge of the video
monitor.
Veiling Reduction:
When the veiling cannot be corrected by adjusting the time gain
compensation controls, use only one focal zone.
3.1.2
Redaction Of Noise (Technique):
Decrease gain without losing structural information.
Redaction Of Banding:
Use a transducer with a different frequency and focus and alter the
TGC setting.
PATIENT
'Wit*'
Redaction Of Interactions
The sonographer should attempt to scan around gas or bone,
obtaining scans of the areas below these structures from an oblique angle.
Redaction Of Compounding:
Repeat the scan using a smoother technique, preferable using a single
pass. Recognition of the artifact is possible if one observes where the
transducer skin lines join.
^ ;,; y . 74/ % ,
& ?>****•& ** **' ** ** ********** f • **** *-* * **s*** ** **fr£*** ***y*****
CT U/S
1. X-Raytube 1- Noise
artifact overshoot.
artifact.
6o
4- Prominent dens and jugular 6- Operator scanning speed.
artifact streak.
6- Patient motion
(A) Pt movement
4.1
DATA COLLECTION
AND
DISCUSSION OF RESULTS
62
Questionnaire onem & ,-
Artifacts in CT Imaging
if ^_JU (V)
A. CT Equipment Artifacts
1-X-Raytube
2- Detectors
3- Operator errors
1-Technical errors.
droplets.
C. CT patient Artifact
phenomenon.
6- Patient motion
(A) Pt movement
£-Uai
1- Noise
2- Calibration problem
4- Veiling artifact
5- Absence of focusing
6- Focusing persistence
7- Pixel mismatch.
1-Noise Artifact
3- Transducer artifact
4- Contract problem
5- Movement artifact
7- Operator pressure
1- Shadow.
2- Reverberation.
3 - Enhancement effect
4- Frezonal zone.
5- Split-image.
6- Comet effect.
7- Static scanning
9- Compounding.
66
Interview No. one
CT scan artifacts
Table (1)
Participant Number %
sonographer 3 30%
Sonologist 4 40%
Engineers 3 30%
100%
50%
40%
30% 30%
A B C
A sonologists
B sonolgraphers
C engineers
from the table " 1 " above we find that the staff that answered the
questioner consist of sonologists, represent 40% and sonolgraphers
represent 30% and medical engineers represent 30%.
Table (2)
50%
20%
A B
A Japanese
B American
C Others
from the table "2" above we find that most of the participant use
Japanese machines which represent 80% and the type of the machines
which are American types and others which represents 20%, this means
the result is mostly dependence on Japanese machines.
Table (3)
type of U.S Equipment artifact.
Type of calibration dust of poloroid noise veiling absence of focusin-g grating pixel
artifact the artifact focusing presistence lopes mismach
camera
number 6 4 1 6 2 1 1
% 60% 40% 10% 60% 30% 20% 10% 10% 10%
Fig (A)
100%
60%
50%
40%
30%
2Q% 10%
A B C D
From Table (3) above we find that there are tow peaks can concerning
calibration problem and noise artifact caused by equipment represent 60%
and the lowest cause contributed derived from focusing presistence,
grating lobes and pixel mismach, and ploroid artifact which represent 10%.
For all general view look Fig (A).
Table (4)
Type of U/S Technique artifact
50%
40%
30%
20% 10%
A B D
From Table (4) we find that there is one peak offered by causes of U/S
technique artifact due to transducer selection problem which represent
80%, and the lowest artifact caused by technique is concerning operator
scanning speed which represent 10%.
Table (5)
Type of U/S Patient artifact.
Type of shadow reverbrat- enhancem- frezonal split comet static beam compounding
artifact ion ent effect zone image effect effect depth
problem
number 6 7 3 2 1 5 1
% 60% 70% 30% 30% 20% 10% 30% 50% 10%
n
Fig (A)
100%
70%
60%
50%
50°A
30%
20%
A B C D F
A reverberation
B shadow
C beam depth problem
D enhancement effect + frezonal zone + static effect
E there is no patient artifact
F compound + comet effect ^^^
From Table (4) we find that there is one peak concerning patient
artifact (reverberation) which represent 70% and the lowest caused by the
patient is comet effect and compounding which represent 10%.
Summary of information collected by the personal contact (interview)
which shown on tables with conclusion summary.
Table (1)
100%
60%
50%
40% 40%
2H%
A B C D
A noise
B no preparation
C probes problem
From the table above we recognize that the most predominant artifacts
is due to noise and gases artifact which represent 40% and the lowers
cause is due to probes problem which represent 20%.
Table (2)
Effect of artifact in the imaging result
u
100%
80%
50%
211%
A B
A Yes
B No
From Table (2) above we find that 80% of the participant agreed that
artifact affect the final imaging result, 20% of them agreed that artifact do
not have any effect on the imaging.
Table (3)
Expectation of U/S artifact
100%
80%
50%
20%
A B
75
A wrong diagnosis
B No affect in the diagnosis
From Table (3) above we have concluded that 80% of the participant
agreed that artifact lead to wrong diagnosis in U/S imaging while 20% said
that artifact does not affect the final result.
Table (4)
Types of U/S due to cause (by %)
100%
80%
50%
40%
A B C D
A noise
B unknown
C calibration
D skills
From Table (4) above we can conclude that the most causes of U/S
artifact in the department is due to noise from the power supply and bad
earthing which represent 40% and the other causes by miscalibration of
U/S unit and skills of the operator, which represent 20% of the total.
Table (5)
Suggestion for the reduction of the artifact in the LJ/S imaging
100%
80%
50%
40%
A B C D
A no preparation of patient
B good earthing
C check calibration
D operator skill
100%
80%
50%
40%
20%
10% 10% 10%
A B C D E
A preparation of pt.
B un co-operative pt.
C movement of the pt.
D obese pt.
E uncontact probe
From Table (6) we conclude that 40% agreed that artifact are due to pt.
Preparation. 20% suggested that those artifact are due to co-operative pt.
10% suggested that artifact were due to movement of the patient, 10%
have also suggested the cause due to patient obesity and finally 10% have
suggested the cause to be due to uncontact probe.
For details of data collection of questioner look table A,B, and C.
Table (A)
U/S Equipment artifact
No. Department Participant job Type of equipment Type of artifact
1. government sonographer Japanese 1- calibration problem
2- dust of camera
2. government engineer Japanese 1- calibration problem
2- poloroid artifact
3. private sonologist Japanese 1- noise
4. government sonologist Japanese 1- calibration problem
2- veiling artifact
5. government sonologist American 1-noise
2-calibration problem
3- main bang artifact
4- dust of camera
5- absence of focusing
6. private engineer Japanese 1-noise
2- absence of focusing
3- dust of camera
7. government engineer Japanese American others 1- noise
2- calibration
3- veiling artifact
4- absence of focusing
5- focusing presistance
6- grating lobes artifact
7- pixel mismach
8- dust of camera
9- noise
8. private sonographer Japanese 1 - noise
9. government sonographer others 1- calibration
10. government sonologist Japanese 2- veiling artifact
8o
Table (C)
U/S Patient Artifact
No. Department Participant job Type of equipment Type of artifact
1. government sonographer Japanese 1 - not found
engineer Japanese 1 - no found
2. government sonologist Japanese 1- reverberation
2- frezonal zone
3. private sonologist Japanese 1- shadow
4. government 2- reverberation
3- beam depth problem
sonologist American 1- enhancement effect
5. government 2- split image
3- static scanning
engineer Japanese 1- shadow
2- reverberation
6. private 3- enhancement effect
4- frezonal zone
5- static scanning
7. government 6- beam depth problem
7- compounding
engineer Japanese American others 1- shadow
2- reverberation
3- enhancement effect
4- frezonal zone
5- split image
6- comet effect
7- static scanning
8- depth problem
8. private sonographer Japanese 1- shadow
2- reverberation
9. government sonographer others 3- beam depth problem
10. government sonologist Japanese 1- shadow
2- reverberation
1- shadow
2- reverberation
3- beam depth problem
82
Table
U/S Technique Artifact
No. Department Participant job Type of equipment Type of artifact
1. government sonographer Japanese 1- not found
2. government engineer Japanese 1- transducer selection problem
3. private sonologist Japanese 1- noise artifact
4. government sonologist Japanese 1- transducer selection problem
2- movement artifact
3- operator pressure
5. government sonologist American 1 - noise artifact
2- transducer selection problem
3- transducer artifact
4- contact problem
6. private engineer Japanese 1- transducer selection problem
2- transducer artifact
3- operator pressure
7. government engineer Japanese, American, others 1-noise artifact
2- transducer selection problem
3- transducer artifact
4- contact problem
5- movement artifact
6- operator scanning speed
7- operator pressure
8. private sonographer Japanese 1-noise artifact
2- transducer selection problem
9. government sonographer others 1- transducer selection problem
10. government sonologist Japanese 1- transducer selection problem
2- contact problem
3- operator pressure
Table (1)
CT Equipment Artifact
90%
80%
70%
60% 1—1
50%
40%
30%
20%
10%
n n
6 7 8
Key:
1- X-Ray tube
detectors
operator errors
high differential metal-brain artifacts streak.
High differential air brain artifact over shoot
Overshoot artifact and partial volume phenomena cephalogram
high differential brain-vat artifact overshoot
Indexing error artifact
X-Ray crystal. Alignment artifact
From the table above we have recognized that the most predominant
artifact is due to " high differential metal brain artifact streak "represents
8 $ •
80% the second peak is detectors artifact and it represents 60% the lowest
contribution came from " high differential brain vat artifact overshoot "
and indexing error artifact represent 10% for both of them.
Table (2)
Technique Artifact
30%
20%
10%
1
Key:
1- Patient motion streak.
High differential air-skull artifact streak.
High differential air-brain artifact streak. Partial volume phenomena.
High differential bore-brain artifact streak .
5- patient motion.
From the table above we can recognize that the most predominant
artifact is technical errors and it represent 70%, the second peak is
technical errors artifact troticollosed pt. pantopaque droplets and represent
20%, the lowest contribution come from Basil or impression prominent
deus and jugular tubercles represent 10%.
Table (3)
CT Patient Artifact
80%
70%
60%
50%
40%
30%
20%
10%
Key:
1- Patient Motion streak.
2- High differential air-skull artifact streak.
3- High differential air-brain artifact streak. Partial volume phenomena.
4- High differential bore-brain artifact streak .
5- patient motion.
From the table above we can recognized that the most predominant
artifact pt. Motion and represent 80% the second peak is high differential
air-skull artifact streak represent 60% the lowest contribution come from
high differential bone-brain artifact streak. Partial volume phenomena and
high differential bone-brain artifact streak represent 40% for both of them.
89-
Summary of information collected by personal contact which is shown
on tables with conclusion summary
Most probable artifacts in CT image cause by table one
Types No. %
1- Blurring of the image " streaks " 3 70%
2- Calibration problem 1 20%
3- High atomic number material in teeth 1 10%
100%
70 %
70%
40% II
40%
10%
III 10%
lit
HI
A B C
Key:
A Blurring of the image
B calibration
C High atomic number material in teeth
From Table we can conclude that the most predominant artifacts are
streaks which represent 70%, the second peaks formed by patient motion
which represent 20%, the lowest artifact source were due to high atomic
number material which represent 10%.
Table (2)
Change in image information
Types No. %
1- Yes 4 80%
2- Yes
3- Yes
4- Yes
5- No. 1 20%
100%
80%
80%
20%
20%
A B
From the table above we can conclude that most artifacts due to the
level of diagnosis which represent 80% and the lowest cause which is
about 20% considered not having any change in diagnosis.
Table (3)
Expectation of Artifact in image result
Types No. %
1- Unknown 1 20%
2- Wrong diagnosis 3 60%
3-
4- Ci
20% 20%
20%
A B C
Key:
A Unknown
B Wrong diagnosis
C Not affect to the diagnosis
From the table above we can conclude that 60% of the faults lead to
wrong diagnosis, 20% were due to unknown causes and only 20% which
did not have any affect on image information or diagnosis.
Table (4)
Reasons of faults in CT image
Types No. %
1- Due calibration of CT 1 20%
2- Use of CT control 2 40%
3- Due motion of pt. 2 50%
4- Due 3 rd generation 1 10%
60%
50% 50%
A B C D
Key: of table (4)
A Due contribution of the CT unit
B Use of CT control
Due motion of pt.
rd
D Due 3 generation of CT
From the table above we can conclude that the most predominant
artifacts due to motion due to pt. Which represent 50%, the second peaks
is due to use of apparatus which represent 20%, the third peaks is due to
wrong calibration which represent 20% and finally the lowest contribution
were caused by third generation which represent 10%.
Table (5)
Suggestion for reduction of artifacts in CT image
Types No. %
1- Instruction to pt. To put away all F,B 2 40%
2- Equipment maintenance. 2 40%
3- Sedation. 2 20%
20%
A B C
Key:
A Instruction to the pt. To put away all F.B
B Equipment maintenance
C Sedation
From the table above we have concluded that the most predominant
artifacts are due to equipment calibration (40%), 40% is also found due to
lack of communication with patient and 20% due sedation.
9/
Table (6)
Artifacts due to patient
Types No. %
1- Streak 5 100%
2- «
3-
4-
5-
100% 100%
From the above table (6) we have concluded that streaks took 100% of
the image artifacts.
5.1
There are varieties of tests that can be carried out using the above
phantoms, which include:
This test should be carried out daily using the same phantom and a
consistent technique. For example, 120 kVp, 200mA3 seconds with 10mm
collimation on a standard algorithm using a 256x256 matrix.
The patient alignment lights should be checked weekly at the time as (1).
Electronic Circuit:
Noise Artifacts:
A common source of spurious echoes in any ultrasound equipment
for medical diagnostics is the inherent noise generated in the electronic
circuitry. It is generally the noise level of the input amplifier alone. Which
limits equipment performance with respect to low-level signals. With
doppler equipment, however, distortion and spurious modulation will also
have a considerable limiting effect with respect to low-level performance.
Detection Thresholds:
It is quite possible that the minimum signal, which the system can
process, is several dB above the noise level. If this is the case, the useful
dynamic range will be less than the stated signal-to-noise level. If the
minimum signal which could be processed is 8 dB above the noise level,
and the signal-to-noise ratio is 10 dB with a particular echo signal, this
signal can only decrease a further 2 dB before it is no longer registered. A
common reason for this might be deliberate rejection effect at an echo
envelope detector.
Storage:
Conversion circuits for gray-scale displays should not added all the
amplitudes registered for any given point on the display, because this will
not allow several scans of the same structure in one single store operation
without blurring the picture excessively with analogue and causing
overflow with digital storage. Thus last-value-write or speak-hold types of
circuits should be used avoid artifacts.
Compensation Setting:
The compensation should be set to minimize the number of artifacts,
but at the same time giving the maximum detail in the representation of
the examined tissue. A reasonable initial setting will normally be given in
the instruction manual for the particular equipment. With digital systems
automatic resetting to this position will often occur at switch-on. This
setting can then be modified at will.
Care should be taken not to get artifacts from transducer ringing (too
high an initial gain) and amplifier noise or spurious echoes (excessive
final gain). Too low gain settings should also be avoided, as this will
aggravate possible threshold and noise problems by not utilizing the
maximum useful-signal handling capacity of the equipment.
\QQ
Gray-Scale Adjustment:
Further adjustments will cause the gray-scale range to fall outside the
dynamic range of the processing circuit output, making the picture either
too bright or too dark. In either case, the usable gray-scale part of the
display is reduced as a result of either too small a signal or overload of the
conversion circuit, and the gray-scale information obtainable is reduced.
Scanning Rate:
When scanning moving objects in the body with a static display, the
picture will not be very sharp, and may be almost useless as far as the
moving objects are concerned except for such moderately moving
structures as the aorta. This is due to the very long scanning time, i.e.
effectively an extremely low scanning rate compared with the rate of the
movement of the objects. When using a dynamic scanner, the problem no
longer exists if the scanning rate is sufficiently high. A scanning rate of
more than 16-20 picture updates per second will, dependent on scanning
and display system generally be sufficient to avoid artifacts.
lot
Dependent on the particular structure studied, the effect of too low a
scanning rate with a dynamic scanner may be far worse than the blurring
caused by moving objects with the static display systems. Besides the
unavoidable reduction of sharpness, the shape of the moving structure and
its rate of movement can be heavily distorted due to a stroboscopic
phenomenon known as aliasing. What actually happens is that the system
is unable to differentiate between successive movement cycle, and is
displaying the information more or less at random so that both the shapes
and the movement of the moving hinge is heavily distorted. It should be
emphasized, that this is an extreme case unlikely to occur with normal
scanners, except when scanning rapidly moving heart valves.
Resolution:
The effective diameter of the beam is dependent on the diameter of
the piezoelectric crystal. The gain level of receiver will also affect the
apparent effectively beamwidth since with a low gain level, only the
highest-amplitude portions of the beam will give detectable echoes.
1oZ
normally be reduced as compared to a steady-state situation, as many of
the interference causing the steady-state side-lobes will not have time to
arise.
Internal Echoes:
Spurious Reflections:
Re-reflections may arise from the strong echoes from the ribs (bone),
when not scanning in the intercostal spaces. They will usually be
accompanied by some sort of shadow effect, when trying to find a
compromise between a high gain to avoid the effect of the attenuation and
strong reflections of the ribs, and the low gain preferable to dimmish the
re-reflections.
The so-called "rain" that can occur close to the anterior bladder wall
when scanning the uterus through the full bladder is also most likely to be
re-reflections.
Granulation:
Air-Field Structures:
The effect of air in the body structure is that the ultrasound is almost
totally reflected. Thus no structures behind a lung or any other internal
cavity filled with air can be seen. However, it is to some degree possible
to overcome such problems by scanning from different positions.
Due to the spurious echoes from the radiated spherical waves, it will
be difficult to discriminate between cystic and solid structures. Generally,
the picture will look blurred.
Fat:
Fat gives unwanted diffraction and high attenuation of the ultrasound,
making it difficult to obtain echoes underneath the far layer, if this is
abnormally thick. There is also a tendency to get scattering in fat layers.
Fat has largely the same effect as calcification and dehydration, and
makes it difficult to discriminate between cystic and solid structures.
Generally, the picture will look blurred.
p6
PARTins
5.2
1. Equipment artifacts.
2. Technique artifacts.
3. Patient artifacts.
Each of the above fault contribute by same way or another to the
detraction of image quality and hence would lead to retake of image.
However, from the analysis of the data collected for this research,
which was conducted by questioner and personal interviews we did find
that most of the artifact in ultrasound is due equipment calibration
problems and noise interference ( from radio and improper earthing 60%).
While artifacts from computed tomography is due to high differential metal
brain artifacts etc. (80%).
From the other hand, the occurrence of technique artifact in the U/S
is due transducer selection problem (80%) where as in CT technical errors
artifact is about (30%) patient artifacts in ultrasound is only by
reverbration which is about 70% while in CT is patient motion which is
about 80%, other patient problem artifacts due to patient preparation is
40% while patient movement is 70%.
Find to reduce the pressure of artifacts in image the operator should
have good command of the control console and he/she should the meaning
of all keys in addition to the better patient preparation during the exam as
well effective patient communication and use of immobilizing devices.
RECOMMENDATION
/of
References
Avruch, L , and cooperberg, P.L. The ring-down artifact. J
Ultrasound Med. 4:21-28, 1985.
Barturm, R., and Crow, H.C. (Eds.). A Manual for Physician and
Technical Personnel: Gray-Scale Ultrasound, Real-time in Ultrasound.
Philadelphia: Saunders, 1983.
U/S: Ultrasound
ill