Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Calibration protocol
Biomed equipment: UV-meter
Performed by: Technical (or qualified) staff
Calibration set-up
The UV-meter needs to be calibrated every year, to confirm radiation is around 0.2 uW/cm2 and
not too low or too high. Calibration can be done by the manufacturer but this is expensive and the
transport can cause problems on the calibration1.
The alternative is to calibrate the meter at the location by a designated user/supervisor once a year.
This should be done according the use of a standard set-up situation (fig.1) and comparing the
outcome every year.
Position 1
Document [uW/cm2] of the 3 positions, every year
Unused bulb
UV light Position 2
Determine fixed
distances + angles (3x)
Position 3
1
The costs for calibration are:
ECAL/1400 ILT1400 Radiometer Electrical $183 Electrical accuracy and linearity check, and recalibration to NIST
standards. Includes I/O check.
PIR(IPIR) Peak Irradiance Response $272 Irradiance response, (amps)(cm²)(watt-1), at peak sensitivity of an IL
detector.
1
Written by biomed referent FSU, MSF-OCA
Protocol calibration UV-meter January 2013
Calibration procedure
2
Written by biomed referent FSU, MSF-OCA