Raw Data Module (RDCM)


RDCM Installation Manual
Software Version v202
Installation Manual for installation on Image Vault 5.0.1.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

ADVANCED SERVICE DOCUMENTATION.


COPYRIGHT GENERAL ELECTRIC COMPANY.
GE CONFIDENTIAL. UNLICENSED USE BY NON-GE
PARTIES IS STRICTLY PROHIBITED.

Part Number: IV294004


Revision: 1
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

TRANSLATION POLICY
Important Precautions

-
GE CONFIDENTIAL

i
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

ii
GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

iii
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

iv
GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)

v
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

vi
GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

vii
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

viii
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

ix
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

x
GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

xi
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

xii
GE
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL
RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE personnel. In
performing all electrical work on these products, GE will use its own specially trained field engineers. All
of GE’s electrical work on these products will comply with the requirements of the applicable electrical
codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.

OMISSIONS & ERRORS


If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Ultrasound Global Documentation Group with specific information listing the system type, manual
title, part number or direction number, revision number, page number and suggestion details.

Mail the information to:

GE Vingmed Ultrasound AS
Service Documentation
P.O.Box 141
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

3191 HORTEN
NORWAY

GE employees should use Post-Market Quality Management (PQM) to report service documentation
issues.

SERVICE SAFETY CONSIDERATIONS


See the GEHC Myworkshop System to determine the status of this document.

DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN


THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.

WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.

- xiii
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

LEGAL NOTES

The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE.

GE makes no representations or warranties with respect to the information herein. In addition, the
information is subject to change without notice. Every precaution has been taken in the preparation of
this document. Nevertheless, GE assumes no responsibility for errors or omissions or any damages,
including special or consequential, resulting from the use of this information. GE will issue updates to
this information periodically, as needed. If there are any questions regarding the information contained
in this manual, please contact your GE Representative.

PROPRIETARY TO GE
Permission to use this Advanced Service Software and related documentation (herein called the
material) by persons other than GE employees is provided only under an Advanced Service Package
License relating specifically to this Proprietary Material. This is a different agreement from the one under
which operating and basic service software is licensed. A license to use operating or basic service
software does not extend to or cover this software or related documentation.

If you are a GE employee or a customer who has entered into such a license agreement with GE to use
this proprietary software, you are authorized to use this Material according to the conditions stated in
your license agreement.

However, you do not have the permission of GE to alter, decompose or reverse-assemble the software,
and unless you are a GE employee, you MAY NOT COPY the Material. The Material is protected by
Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal
prosecution.

If you are not party to such a license agreement or a GE Employee, you must exit this Material now.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

TRADEMARKS
All products and their name brands are trademarks of their respective holders.

COPYRIGHTS
© 2016 by General Electric Company. All Rights Reserved.
See the GEHC Myworkshop System to determine the status of this document.

xiv -
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Revision History

Date
Revision (YYYY-MM-DD) Reason for change
1 2016-11-25 Initial release of manual.

List of Effected Pages (LOEP)

Pages Revision Pages Revision Pages Revision


Title Page 1 1-1 to 1-4 1 Appendix A 2

i to -xvi 1 2-1 to 2-2 1 Back Cover N/A

TOC 1 3-1 to 3-10 1

4-1 to 4-6 1

5-1 to 5-2 1

Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on ePDM (GE
Healthcare electronic Product Data Management). If you need to know the latest revision, contact your distributor, local GE Sales Representative
or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

- xv
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

xvi
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

This page was intentionally left blank.


RAW DATA MODULE (RDCM)
RDCM INSTALLATION MANUAL
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Purpose of this installation manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Contents in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Tools in kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Other Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2

Effectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3

Prerequisite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3

Obsolete Service Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3

CHAPTER 2
Platform requirement
Platform requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

CHAPTER 3
Software installation procedure
Overview of the RDCM installation DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1

Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
See the GEHC Myworkshop System to determine the status of this document.

Checking the prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3

Install the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5


Resource File Reconciliation dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8

CHAPTER 4
Verification of successful installation
Verify that the RDCM is running correctly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1

Verify that the RDCM user account has sufficient permissions . . . . . . . . . . . . . . . . 4 - 3

Check link between RDCM software and Image Vault 5 software. . . . . . . . . . . . . . 4 - 4


Table of Contents xvii
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Dispose of any excess material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5

CHAPTER 5
Known limitations
Content in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1

Issue after RDCM update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1

How to change the IP address of RDCM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2

APPENDIX A
Enabling Server Feature ‘Desktop Experience’
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

xviii -
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Chapter 1
Introduction

Section 1-1
Overview
1-1-1 Purpose of this installation manual
NOTE: These instructions apply to installation of RDCM on Windows Server 2008 64 bit (Windows
Server 2008 R2) running Image Vault 5.0.1.

This installation manual describes how to install the Vivid Raw Data Module (RDCM) software, available
on the DVD Part Number EZ100532, on the Image Vault 5.0.1.

1-1-2 Contents in this chapter


1-1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-3 Effectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-4 Prerequisite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-5 Obsolete Service Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

1-1-3 Contents in this manual


• Chapter 1 - Introduction
• Chapter 2 - Platform requirement
• Chapter 3 - Software installation procedure
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

• Chapter 4 - Verification of successful installation


• Chapter 5 - Known limitations
• Appendix A - Enabling Server Feature ‘Desktop Experience’
See the GEHC Myworkshop System to determine the status of this document.

Chapter 1 - Introduction 1-1


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 1-2
Tools
1-2-1 Tools in kit

Table 1-1 Content needed for the Vivid Raw Data Module installation

Item Part Number Description


1 EZ100532 Vivid Raw Data Module DVD, software version v202

Image Vault: Vivid Raw Data Module (RDCM) Installation Manual


2 IV294004
(This manual)

1-2-2 Other Tools


• Image Vault Installation Manual
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

1-2 Section 1-2 - Tools


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 1-3
Effectivity
Please refer to: Chapter 2 - Platform requirement.

Section 1-4
Prerequisite
Please refer to: Chapter 2 - Platform requirement.

Section 1-5
Obsolete Service Notes
• SN76380 RDCM Activity logging to file, RDCM action with 3rd party data.
• SN76398 RDCM integration with customer domain security.
Both above mentioned Service Notes are obsolete with the installation of this version of RDCM. The
issues discussed in the Service Notes have been incorporated into the software package and/or
covered in sections of this manual.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 1 - Introduction 1-3


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

1-4
DIRECTION IV294004, REVISION 1
GE CONFIDENTIAL

Section 1-3 - Effectivity


This page was intentionally left blank.
RDCM INSTALLATION MANUAL
RAW DATA MODULE (RDCM)
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Chapter 2
Platform requirement

Section 2-1
Platform requirement
• The RDCM software contained on this DVD is intended for installation on Image Vault 5.0.1 servers
running Windows Server 2008 64 bit (Windows Server 2008 R2).
• Image Vault installation DVDs may contain an older version of RDCM. Any existing version of
RDCM will be properly updated by any newer RDCM installation package.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 2 - Platform requirement 2-1


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

2-2
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

This page was intentionally left blank.


RDCM INSTALLATION MANUAL
RAW DATA MODULE (RDCM)
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Chapter 3
Software installation procedure

Section 3-1
Overview of the RDCM installation DVD
The installation DVD contains a zipped copy of the DVD content.

• File name: RDCM_202_x_x.zip


(the ‘x_x’ in the file name indicates that the file name may be slightly different than stated here.)
• Location: The zip file is located on root on the DVD.
The intended use is for electronic distribution through standard service connectivity methods.

The installation DVD also includes an unzipped version of the installation files on the root of the DVD.
These files may be used for installation if you want to install direct from the DVD.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 3 - Software installation procedure 3-1


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 3-2
Prerequisites
Before starting the installation of the RDCM software, it is important to verify if any prerequisites are
missing. The prerequisites required by RDCM are summarized in Table 3-1.

NOTE: Server restarts are required by some of the prerequisites.

Table 3-1 RDCM prerequisites

Number Prerequisite Comment


Operating system (OS): Win 7 64-bit SP1 or Win
1 Must be provided by customer
Server 2008 R2 SP1

If OS is Win Server 2008 R2: Activate the Server


2 Manual action needed, see: Appendix A
Feature 'Desktop Experience'

DirectX version 11.0 or greater. This is part of the OS,


3 No action needed
i.e. always present.

Windows platform update KB2670838.


4 Download and install if missing
Alternatively, install Internet Explorer 11.

5 .NET Framework v3.5.30729 or greater Taken care of by RDCM setup

6 .NET Framework v4.5.50709 or greater Taken care of by RDCM setup

Visual C++ Redistributable for Visual Studio 2015


7 Taken care of by RDCM setup
Update 3 (v14.0.24212 x64)

Visual C++ Redistributable for Visual Studio 2015


8 Taken care of by RDCM setup
Update 3 (v14.0.24212 x86)

The 'TEMP' environment variable must exist and point


9 Manual action needed
to an existing directory.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

Read and write access for user “Network Service”


(RDCM runs as this user) to the following directories:
1.) The directory pointed to by the TEMP environment
10 Manual action needed
variable
2.) The staging area, i.e. the “Archive” share
3.) The STS area, i.e. the “IV_Archive” share
See the GEHC Myworkshop System to determine the status of this document.

3-2 Section 3-2 - Prerequisites


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

3-2-1 Checking the prerequisites


The presence of prerequisites can be checked by doing the following steps:

1.) Insert the installation disk into the DVD drive or download the zipped file to the ImageVault.
2.) Open a Windows Explorer window and navigate to the location where you have the installation files
(typically to the root of the DVD, or alternatively to the media or folder where the files from the
installation DVD have been copied).
The ‘RDCMCheckPrerequisites.exe’ program is in the installation root and not part of Installation
CAB files, so it can be run at any time without the need to run ‘setup.exe’ first.

Figure 3-1 RDCM installation files

3.) If not already opened, open a Command Prompt window.


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Figure 3-2 Open Command Prompt window

Chapter 3 - Software installation procedure 3-3


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

3-2-1 Checking the prerequisites (cont’d)


4.) Start RDCMCheckPrerequisites.exe from the Command Prompt window.
Use the command: RDCMCheckPrerequisites.exe -v
Press Enter to run the program.

Before pressing Enter After pressing Enter

Figure 3-3 Run ‘RDCMCheckPrerequisites.exe’ (before and after pressing Enter)

Any missing prerequisites will be written to the window.

NOTE: The ‘setup.exe’ portion of the installation will attempt to install missing prerequisites, but it is the
customer responsibility to verify the latest installed windows components are the recommended
versions by the vendor (Microsoft).
Installation of Desktop Experience windows feature is the most challenging due to possible
missing Windows' prerequisites of the missing RDCM prerequisite.

Reference:
http://technet.microsoft.com/en-us/library/cc742809.aspx
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

3-4 Section 3-2 - Prerequisites


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 3-3
Install the software
When preparing the system, you should run RDCMCheckPrerequisites.exe prior to running
setup.exe. Exit all running Windows and/or Anti Virus programs.

See: Section 3-2 "Prerequisites" on page 3-2 for instructions on what prerequisites are required and the
valid URL links for the prerequisites at the time of this release.

Follow these steps to install the new RDCM software:

1.) If not already done, insert the installation disk into the DVD drive or download the zipped file to the
ImageVault.
2.) Open a Windows Explorer window and navigate to the location where you have the installation files
(typically to the root of the DVD, or alternatively to the media or folder where the files from the
installation DVD have been copied).

Figure 3-4 Folder with installation files

3.) Double click on the file setup.exe to start RDCM Setup.


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

- If all prerequisites are satisfied, the setup will skip the prerequisite installation graphic shown
below.
- If not, after you run (double click) on setup.exe, the prerequisite process will begin with the
dialog shown below.
See the GEHC Myworkshop System to determine the status of this document.

Figure 3-5 Missing requirements

Chapter 3 - Software installation procedure 3-5


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 3-3 Install the software (cont’d)


If you suppress a forced reboot, ‘Setup.exe’ may close after one or more prerequisites have been
installed. Restart ‘Setup.exe’ until complete.

NOTE: The ‘setup.exe’ portion of the installation will attempt to install missing prerequisites, but it is the
customer responsibility to verify the latest installed windows components are the recommended
versions by the vendor (Microsoft).
Installation of Desktop Experience windows feature is the most challenging due to possible
missing Windows' prerequisites of the missing RDCM prerequisite.

Reference:
http://technet.microsoft.com/en-us/library/cc742809.aspx

4.) After the missing requirements have been installed, the Preparing Setup screen is displayed.
- If you are installing this on Windows Server 2008 with User Access Controls (UAC)
enabled, it is recommended that you Right Click on Setup and select Run As
Administrator.
- If you have previously un-installed RDCM (not required or recommended), the installation
dialogs may be slightly different. Follow the installation prompts and dialogs and this manual
to achieve a complete and successful installation or upgrade of existing RDCM.
See Figure 3-6 ‘Preparing Setup” on page 3-6.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Figure 3-6 Preparing Setup

It is possible to stop the preparations by clicking Cancel as long as this dialog is displayed.

Next, the Vivid RDCM Installation dialog is shown on the screen (see next page).

3-6 Section 3-3 - Install the software


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 3-3 Install the software (cont’d)

Figure 3-7 Version information

The installation dialog displays the existing and installing version numbers of RDCM. Be certain the
version you are installing is newer than the existing one.

5.) Read the text in the dialog.


- If needed, select Cancel, then exit any Windows programs and/or anti virus programs before
you continue.
- If no Windows programs and/or anti virus programs are running on the server, select Next >.

NOTE: Installation while anti virus programs are active may increase installation issues or errors. If anti
virus software can not be paused, extra verification should be done after installation to verify
the system is completely operational.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 3 - Software installation procedure 3-7


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 3-3 Install the software (cont’d)


The wizard starts to install the program files.

Figure 3-8 Installing

3-3-1 Resource File Reconciliation dialog


The installation will record a “Resource File Reconciliation log.” This indicates which resource files
appear to have been changed. This is informational only. If the Image Vault installation of RDCM
included customization of RDCM resource files, the installer will need to repeat those customizations
again. A backup copy of the previous resource files has been placed on the system in a folder similar
in name to the example below.

C:\Program Files (x86)\RDCM\RESOURCES.112.00.004


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

Such changes are not typical and repeating the resource file customization is usually not required.
See the GEHC Myworkshop System to determine the status of this document.

3-8 Section 3-3 - Install the software


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

3-3-1 Resource File Reconciliation dialog (cont’d)


After all the new files has been installed, the installation is complete, and the InstallShield Wizard
Complete dialog is displayed.

Figure 3-9 Installation is complete

6.) Select Finish to exit the wizard.


RDCM installation does not require a system reboot. Rebooting the Image Vault system is suggested
to ensure all services restart automatically.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 3 - Software installation procedure 3-9


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

3 - 10
DIRECTION IV294004, REVISION 1
GE CONFIDENTIAL

Section 3-3 - Install the software


This page was intentionally left blank.
RDCM INSTALLATION MANUAL
RAW DATA MODULE (RDCM)
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Chapter 4
Verification of successful installation

Section 4-1
Verify that the RDCM is running correctly
After you have installed the software following the procedure above, you should verify that the system
is running correctly.

First we will check that the RDCM itself is running:

1.) Navigate with Windows Explorer to the RDCM program folder, by default
C:\Program Files\RDCM\bin
and double click on the file RDCMLogWindow.exe.

1
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

3
See the GEHC Myworkshop System to determine the status of this document.

Figure 4-1 RDCM Log Window

It is possible that RDCMLogWindow will scroll past the version information before you can view it.
If that happens, an alternative method to verify the installation is to look at the RDCM Logfile.txt and
read from the bottom up to verify the latest RDCM version.

Chapter 4 - Verification of successful installation 4-1


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 4-1 Verify that the RDCM is running correctly (cont’d)


The Log file is located on this path: %Test_Root%\Log\RDCM\logfile.txt

A typical location for this path is: D:\GEHC\ImageVault\User_data\Log\RDCM\logfile.txt

2.) Verify that the RDCM version number is correct, i.e. that the log window contains a message as
indicated below, refer ‘1’ in Figure 4-1.
RDCM(632); RDCMService version=x.x.x.x

3.) Verify that you get only “Info” messages in the log, i.e. no unexpected “Error” or “Warning”
messages.
Refer to Figure 4-1:

- ‘2’ shows the path to the TEMP folder.


RDCM(632); TEMP: S:\TEMP

- ‘3’ RDCM server ready:


RDCM(5040); Waiting for a connection at port 13900…
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

4-2 -
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 4-2
Verify that the RDCM user account has sufficient permissions
If the RDCM does not have sufficient permissions to a directory, it will manifest it in the RDCM log as
follows (see the red line):

Failure when closing file Access to the path ‘c:\Any folder containing file for RDCM\Image001’

Figure 4-2 Access issue

In the example below, RDCM is failing while trying to close the file.

In this case, the problem was resolved by adding the “Write” permission for account “Network Service”
to the failing directory. Individual system security including domain membership may require advanced
folder permission access:
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Figure 4-3 Adding the ‘Write’ permissions

Chapter 4 - Verification of successful installation 4-3


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 4-3
Check link between RDCM software and Image Vault 5 software
Next, check that the link between the RDCM software and Image Vault 5 software is operative:

1.) Open the Image Vault 5 user interface and log on as Administrator.
2.) Select the top-level tab page Info, then the Service Manager tab in the SERVER INFORMATION
window.
3.) Press the button Restart All.
You will get the following window:
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

4.) Verify that all the services are running as appropriate for the individual system and licensed options.
5.) With every patch and change, you should verify the system is working by performing a system
checkout of services and functions. This checkout is covered in the Image Vault installation manual.

4-4 Section 4-3 - Check link between RDCM software and Image Vault 5 software
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 4-4
Dispose of any excess material

WARNING THE WASTE OF ELECTRICAL AND ELECTRONIC EQUIPMENT MUST NOT BE


DISPOSED AS UNSORTED MUNICIPAL WASTE AND MUST BE COLLECTED
SEPARATELY.
PLEASE CONTACT THE MANUFACTURER OR OTHER AUTHORIZED DISPOSAL
COMPANY TO DECOMMISSION YOUR EQUIPMENT.

• Properly dispose of any excess material in accordance with current GE’s policy, according to local
policy.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 4 - Verification of successful installation 4-5


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

This page was intentionally left blank.


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

4-6 Section 4-3 - Check link between RDCM software and Image Vault 5 software
GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Chapter 5
Known limitations

Section 5-1
Content in this chapter
• Section 5-2 "Issue after RDCM update" on page 5-1
• Section 5-3 "How to change the IP address of RDCM" on page 5-2

Section 5-2
Issue after RDCM update
Previous version of RDCM may have a different product ID in registry which is not overwritten. The
remaining entry is cosmetic and does not affect operation of the product RDCM or Image Vault.

If it is desired to remove the duplicate entry from Control Panel Add Remove Programs follow the
commands below:

1.) To manually clean this, Reg Delete:


“HKLM\SOFTWARE\Microsoft\Windows\CurrentVersion\ Uninstall\{302D7C9A-C09F-4E7D-9AE4-
608C1418C52B}”
2.) Reg Delete:
“HKLM\SOFTWARE \Classes\Installer\Products\A9C7D203F90CD7E4A94E06C841815CB2”
3.) Restart Windows.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

Chapter 5 - Known limitations 5-1


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Section 5-3
How to change the IP address of RDCM
NOTE: The step below shall be performed after RDCM has been installed.

During RDCM Startup, the following line is displayed in the RDCM log:

Starting server at nnn.nnn.nnn.nnn:13900

Depending on the network configuration, RDCM may bind to a different type of address than the one
we want. A typical scenario is a mixed IP v4 and IP v6 network configuration.

By default, RDCM binds to the first available IP v4 address. This behavior can be modified so that
RDCM binds to the second IP v4 address in the list, or to a IP v6 address.

The file to modify is RDCM.res, which by default is located in the C:\Program Files\RDCM\Resources
directory.

Backup the file(s) before any changes are made. An incorrectly formatted line may prevent RDCM from
launching.

It is not important to force RDCM to bind to any one specific IPv4 address. It is only important to modify
this if you want RDCM to bind to a IPv6 address.

1.) Open the file RDCM.res in Notepad.

2.) Locate the [NetworkAdapter] section

3.) Change the line in bold below:

AddressFamily = IPv6 # Supported values: IPv4, IPv6. Default IPv4.

#AddressIndex = 0 # 0 is the first in the list. Default value is 0.

#Port = 13900 #Default port is 13900


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.

4.) Save the file.


5.) Restart the RDCM service.
RDCM will now bind to the first available IPv6 address.

The first address in the list corresponds to AddressIndex=0. The leading ‘#’ character means that
the setting is not active. If a setting is not active, like the AddressIndex or Port in this case, the
See the GEHC Myworkshop System to determine the status of this document.

default values (as indicated at the end of the lines) will be used.

6.) Check the RDCM log and repeat the step above by incrementing the value of AddressIndex until
RDCM binds to the correct address.

5-2 Section 5-3 - How to change the IP address of RDCM


GE GE CONFIDENTIAL RAW DATA MODULE (RDCM)
DIRECTION IV294004, REVISION 1 RDCM INSTALLATION MANUAL

Appendix A
Enabling Server Feature ‘Desktop Experience’
This appendix applies only to the Windows Server 2008 R2 operating system. The Windows 7 already
has this functionality embedded.

The Server Feature “Desktop Experience” is by default not enabled. The feature is enabled by executing
the following manual procedure:

1.) Log on to the machine where RDCM shall be installed.


2.) Click the Start button, click Control Panel, click Programs, and then
click Turn Windows features on or off.
3.) In the Server Manager windows, expand the Features node, then click on Add Features.
To turn on the Desktop Experience feature, tick off the check box next to that feature.
Click OK.
4.) Restart the machine.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

- Enabling Server Feature ‘Desktop Experience’ A-1


State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.

GE

A-2
DIRECTION IV294004, REVISION 1

-
GE CONFIDENTIAL

This page was intentionally left blank.


RDCM INSTALLATION MANUAL
RAW DATA MODULE (RDCM)
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.
State: COMPLETE - Production parts may not be defined by this document; it is an uneditable finished draft.
See the GEHC Myworkshop System to determine the status of this document.