“Blood coagulation detection Using Atmega328P-PU”


Anuja Marathe1, Hiteshi Turakhia2
1Student,Dept of Electronics & Tele. Engg, KKWIEER, Nashik, India.
2Student,Dept of Electronics & Tele. Engg, KKWIEER, Nashik, India.

---------------------------------------------------------------------***---------------------------------------------------------------------
Abstract – Coagulation disorders include a number of
The rapidly increasing growth of technology in all the disorders that are related to defects in the
fields has made our lives much easier. The arrival of clotting of blood. Deficiencies in any of the
technology has also increased the hazardous dangers in protein factors involved in coagulation can
our lives which can be in every field where the technology result in hemorrhages following minor
growth is rapidly taking place and which causes huge loss injuries.[1]
of life and property because of the poor emergency
facilities. Here, the proposed system detects the blood
clots in the body the proposed detection system will 2. LITERATURE SURVEY
provide an optimum solution to detect blood clots by Coagulation is a complex network of
using ATmega328P-PU. In this system the module is used interaction involving vessels and factors the
to track and find the accident location of the blood clots. A ability to form and to remove a clot is truly a
heartbeat sensor is used to find the pulse rate of the blood system dependant on many synergistic
which will help to detect the blood clots. In this module, forces, haemostasis depends on a system of
heartbeat sensor calculates the pulse rate of the patient checks and balances between thrombosis and
which constantly changes according to the heartbeat. In haemorrhage that includes both
this module we have used the timer to measure a pulse
procoagulantes and anticoagulants. This scale
rate for a specific time period if the pulse rate is changing
for that time period then blood clot is not detected as the needs to be kept in balance.[1]
blood is constantly flowing but for clot detection the timer
will measure the pulse rate and if pulse rate is not Thrombosis is an activation of the
changing during that time period than blood clot is haemostatic system at an inappropriate time
detected. Here, we have used Arduino because it has in vessels. Thrombi formed in this fashion are
inbuilt analog inputs and digital outputs which takes pathological and beyond the normal
analog input using heartbeat sensor and gives digital hemostatic mechanism if physiological
output at LCD display. This project is useful in detecting anticoagulants are decreased in the
the Blood clots precisely by means of heartbeat sensor. As circulation there will be clot if procoagulants
a future implementation a GPS system can be built which
or clotting factors are decreased this scale
can help to give messages to doctors and give appropriate
treatments to the patients. will tip toward bleeding, hemorrhage or
excessive bleeding may be due to blood
vessel disease, platelets abnormalities and
Key Words: ATmega328P-PU, Heartbeat sensor, acquired or congenital abnormalities.[2]
Pulse rate detection, Blood clot detection, GPS
system Hemostasis is comprised of the vascular
system, platelet and series of enzymatic
1. INTRODUCTION reactions of the of coagulation factors. Under
physiological conditions, fluidity is
Coagulation (also known as clotting) is the process
by which blood changes from a liquid to a gel, maintained by the anticoagulants,
profibrinolytic, and antiplateletes properties
forming blood clot. It potentially results
in hemostatis, the cessation of blood loss from a of the normal endothelium. Coagulation is
divided into two major systems, the primary
damaged vessel, followed by repair.[2]
The mechanism of coagulation involves activation, and secondary systems of hemostasis.
(Lozier, 2004) The primary systems comprise
adhesion, and aggregation of platelets along with
deposition and maturation of fibrin. Disorders of platelets function and vasoconstriction. [9]
coagulation are disease states which can result in
bleeding or obstructive clotting.[2]
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 04 | Apr 2019 www.irjet.net p-ISSN: 2395-0072

The secondary system involves coagulation proteins Thus, we are using arduino to display digital
and a series of enzymatic reactions once the output which is taken from the heartbeat
coagulation protein become involved fibrin is formed sensor as analog input. Arduino based board
and this reinforces platelets plug formation until exposes most of the microcontroller’s I/O
healing is complete the product of the coagulation pins for use by other circuits. Arduino
cascade is the conversion of soluble fibrinogen into provides 14 digital I/O pins, six of which can
insoluble fibrin clot. produce pulse-width modulated signals, and
[2] six analog inputs which can also be used as
six digital I/O pins.
3. SYSTEM DESCRIPTION Here we can use liquid crystal display as
Here we are taking inputs from the sensors,we are digital output which is taken from heartbeat
detecting blood flow rate of the patient which will sensor as analog input .LCD are available to
detect whether there are blood clots or not. Thus we display arbitrary images or fixed images with
are using hertbeat sensor which will detect blood low information content which can be
flow rate of the patient. Inputs to the hearbeat pulse displayed or hidden, such as preset words,
sensor is attached to a finger and get analog out digits and seven segment displays as in
from the sensor based on heartbeat pulse.we can digital clock.
read the analog output with microcontroller and plot
it or calculate the readings like hearbeat per 4. FLOWCHART
minute.In this way, we can calculate the hearbeat
pulse which will give blood flow rate.
Blood temperature changes according to
environmental temperature. Temperature sensor is
to be used to measure temperatue through an
electrical signal(analog input taken from heartbeat
sensor). Here we are using temperature sensor
LM35.

Figure - 1 Block Diagram

Here,we can use microcontroller to calculate analog


output from the heartbeat sensor, i.e we get blood
flow which willl be digitally displayed using
microprocessor.

© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 2
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 04 | Apr 2019 www.irjet.net p-ISSN: 2395-0072

[8]. Deng B., Lin Y., Wang C., Li F., Wang Z.,
Zhang H., Li X., and Le X. C., Anal. Chim.
5. CONCLUSIONS Acta 837, 1
Hence, the Blood Clot Detection System for Detecting (2014).10.1016/j.aca.2014.04.055
Blood Clots has been implemented using [9]. Shuman M. A. and Majerus P. W., J. Clin.
ATMega328P microcontroller .This System is Invest. 58, 1249 (1976).10.1172/jci108579
Designed which can detect Blood Clots in less time [10]. Uzlova S. G., Guria K. G., and Guria G.
and precisely at a perfect doubted location in the T., Philos. Trans. A. Math. Phys. Eng. Sci. 366,
body. The System uses Heartbeat Sensor to detect 3649 (2008).10.1098/rsta.2008.0109
the Blood clots. The Heartbeat Sensor determines the [11]. Tennico Y. H., Hutanu D., Koesdjojo M.
pulse rate of the patient which is compared with the T., Bartel C. M., and Remcho V. T., Anal.
pulse rate of the body. Thus, if the pulse rate is less Chem. 82, 5591 (2010).10.1021/ac101269u
than the threshold value then the Blood clot has been [12]. Voros E., Cho M., Ramirez M., Palange A.
detected and if Threshold is greater than the pulse L., De Rosa E., Key J., Garami Z., Lumsden A.
rate then Blood clot is not detected. Thus, in this way B., and Decuzzi P., Adv. Funct. Mater. 25, 1709
system determines the blood clots in the patient’s (2015).10.1002/adfm.201404354
body. [13]. Drozdov A. S., Vinogradov V. V.,
Dudanov I. P., and Vinogradov V. V., Nat. Publ.
6. Future Scope Gr. 1(2016).10.1038/srep28119
1. Currently this project tells us about blood clots in [14]. Starmans L. W. E., Moonen R. P. M.,
the body i.e. in the veins, capillaries, arteries. Aussems-Custers E., Daemen M. J. A. P.,
2. This n is further implemented by adding GSM Strijkers G. J., Nicolay K., and Grüll H., PLoS
Module in the system. One 10, 1
3. GSM Module will help to directly give information (2015).10.1371/journal.pone.0119257
to the doctors which will help doctors to give further [15]. Schmitz S. A., Winterhalter S., Schiffler
treatment required. S., Gust R., Wagner S., Kresse M., Coupland S.
E., Semmler W., and Wolf K. J., Radiology 221,
237 (2001).10.1016/s1076-6332(03)80223
7. REFERENCES
[16]. Wilson R., Cossins A., Nicolau D. V., and
[1]. Jiangtao ji, Feifei Hu, Zhitao He. Application of
Missailidis S., Nucleic Acid Ther. 23, 88
four-rotor micro micro aircraft in farmland
(2013).10.1089/nat.2012.0386
information acquisition[J]. Journal of agriculture
[17]. Weaver J. B., Rauwerdink A. M., and
mechanization research,2013(2):1-4
Hansen E. W., Med. Phys. 36, 1822
[2].Guifa Cheng. Study on the development of the
(2009).10.1118/1.3106342
principal and the effect of coagulation
[18]. Zhang X., Reeves D. B., Perreard I. M.,
analyzer[J].Medical Information (Late
Kett W. C., Griswold K. E., Gimi B., and Weaver
journal),2011,24(8):317-318
J. B., Biosens. Bioelectron. 50, 441
[3]. LuWang. Four rotor unmanned Ariel vehicle
(2013).10.1016/j.bios.2013.06.049
control technology research[D]. Harbin Engineering
[19]. Tu L., Jing Y., Li Y., and Wang J.-P., Appl.
University.2012:21-35
Phys. Lett. 98, 213702
[4]Zhenfang Zhao. Design of the embedded blood
(2011).10.1063/1.3595273
coagulation detection system[D].WuHan University
[19]. Pankhurst Q. A., Connolly J., Jones S. K.,
of technology.2012:16-42.
and Dobson J., J. Phys. D. Appl. Phys. 36, R167
[5].Zhenfang Liu, Yaping Hu,Wenchao Zhang.
(2003).10.1088/0022-3727/36/13/201
Improve the instrument with LED-type excitation
[20]. Tasset D. M., Kubik M. F., and Steiner
light source stability measures[J]. Life Sciences
W., J. Mol. Biol. 272, 688
Instruments,2016,4(6):29-23
(1997).10.1006/jmbi.1997.1275
[6]. Rauwerdink A. M. and Weaver J. B., Appl. Phys.
[21]. Weaver J. B., Rauwerdink A. M., Sullivan
Lett. 96, 1 (2010).10.1063/1.3291063
C. R., and Baker I., Med. Phys. 35, 1988
[7]. Crawley J. T. B., Zanardelli S., Chion C. K. N. K., and
(2008).10.1118/1.2962600
Lane D. A., J. Thromb. Haemost. 5, 95
(2007).10.1111/j.1538-7836.2007.02500.x

© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 3
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 04 | Apr 2019 www.irjet.net p-ISSN: 2395-0072

[22]. Bock L. C., Griffin L. C., Latham J. A., Vermaas E.


H., and Toole J. J., Nature 355, 564
(1992).10.1038/355564a0
[23]. Weaver J. B., Zhang X., Kuehlert E., Toraya-
Brown S., Reeves D. B., Perreard I. M., and Fiering
S., Nanotechnology 24, 325502
(2013).10.1088/0957-4484/24/32/325502
[24]. Perreard I. M., Reeves D. B., Zhang X., Kuehlert
E., Forauer E. R., and Weaver J. B., Phys. Med. Biol.59,
1109 (2014).10.1088/0031-9155/59/5/1109
[25]. Ramström S., Blood Coagul. Fibrinolysis 16, 447
(2005).10.1097/01.mbc.0000178827.52242.89s