Sei sulla pagina 1di 48

No.

2Y730-502E*C

MAINTENANCE MANUAL
FOR
DIAGNOSTIC ULTRASOUND SYSTEM
/
MODELS SSA-680A/SSA-660A
(2Y730-502E*C)

© TOSHIBA MEDICAL SYSTEMS CORPORATION 2004-2009


ALL RIGHTS RESERVED
No. 2Y730-502E*C

IMPORTANT!

1. No part of this manual may be copied or reprinted, in whole or in part,


without written permission.

2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
No. 2Y730-502E*C

REVISION RECORD

REV. DATE REASON PAGE SER. DOC.


(MM/YY) /AUTHOR CHANGED No. PRODUCT.

INI. 10/'04 Mr. Takahashi ------- TM-WP7

*A 05/'06 Support of V5.0 Mr. Nakai P. 5-4 TM-WP2

*B 08/'07 Addition of SSA-680A, P. 1-1, 1-3,


Support of SSA-660A V6.0 1-4, 5-4 to 5-6,
6-3 to 6-9, 8-3,
Mr. Nakai A-1

*C 03/'09 Change of check procedure Mr. Nakai P. 2-2

3
No. 2Y730-502E*C

Safety Precautions

1. Meaning of Signal Words

In this manual, the signal words DANGER, WARNING, and CAUTION are used regarding
safety and other important instructions. The signal words and their meanings are defined as follows.
Please understand their meanings clearly before reading this manual.

Signal word Meaning

DANGER Indicates an imminently hazardous situation which, if not avoided, will


result in death or serious injury.

WARNING Indicates a potentially hazardous situation which, if not avoided, could


result in death or serious injury.

CAUTION Indicates a potentially hazardous situation which, if not avoided, may


result in minor or moderate injury.

CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in
property damage.

2. Meaning of Safety Symbols

Symbol Description
Type-B applied part
* Type B when Type-B applied part is connected.
The PCG sensor and pulse sensor that can be connected
to this system are Type-B applied parts.
Type-BF applied part
* Type BF when Type-BF applied part is connected.
All ultrasound transducers, ECG cables, and respiration
sensors that can be connected to this system are Type-BF
applied parts.
"Attention" (Refer to the operation manual.)

4
No. 2Y730-502E*C

3. Safety Precautions

Please observe the following precautions to ensure the safety of service engineers as well as operators
when using this system.

DANGER: Do not use flammable gases such as anesthetics, or flammable liquids such
as ethanol, near this product, because there is danger of explosion.
If flammable or combustible gas is present in the room where the system is
installed, ventilate the room by opening the windows or using other
appropriate measures before turning the system power ON/OFF.
If the system power is turned ON/OFF in a room containing flammable or
combustible gas, an explosion may occur as a result of sparking at contacts in
a relay circuit or circuit breaker.

WARNING: 1. Information concerning the terminal to which the potential equalization


conductor is to be connected ( ):

To use another medical device in combination with this system, an


equipotential wire for connecting to an equipotential bus must be
supplied. For more information, contact your TOSHIBA representative.

* Be sure to connect the potential-equalization lead wire before inserting


the equipment power plug into the receptacle. Also, be sure to remove
the equipment power plug from the receptacle before disconnecting
the wire to avoid electrical shock.

2. Information concerning the functional earth terminal


( ):

Use of the functional earth terminal as protective earth is not allowed in


Europe. The system should be connected according to local
requirements.

3. Do not connect this system to outlets with the same circuit breakers and
fuses that control current to devices such as life-support systems. If this
system malfunctions and generates an overcurrent, or when there is an
instantaneous current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.

4. Do not connect to the system transducers other than those specified by


TOSHIBA, to prevent accidents such as fire.

5. Do not subject the transducers to knocks. Use of defective transducers


may cause an electric shock.

5
No. 2Y730-502E*C

CAUTION: 1. Malfunctions due to radio waves


(1) Use of radio-wave-emitting devices in the proximity of this kind of
medical electronic system may interfere with its operation. Do not bring
or use devices which generate radio waves, such as cellular telephones,
transceivers, and radio controlled toys, in the room where the system is
installed.

(2) If a user brings a device which generates radio waves near the system,
they must be instructed to immediately turn OFF the device. This is
necessary to ensure the proper operation of the system.

2. Be sure to install the system on a level floor and lock the casters. If this is
not done, the system may move, injuring the service personnel.

3. Move the system forward or backward only. If the system is moved to the left
or right, it may fall, causing injury.

4. To prevent electric shock, do not connect the peripheral units (video printer,
VCR, etc.) to an external outlet. Peripheral units should be connected to the
service outlet of the system. For the connection procedures, contact your
TOSHIBA representative.

5. The service outlet of the main unit is intended solely for recommended
peripheral units. Do not connect devices other than recommended
peripheral units to the service outlet. Connecting devices other than those
recommended may cause the consumption current to exceed the power
capacity of the main unit, possibly resulting in malfunction.

6. Do not place any objects on top of the monitor. They may fall, causing
injury.

7. Do not allow fluids such as water to contact the system or peripheral


devices. Electric shock may result.

8. Do not handle the system with wet or moist hands. Electric shock may
result.

9. The monitor is a heavy unit. Two persons are required when mounting or
removing the monitor. If a single person performs this work, he or she may
get injured.

10. Be sure to turn OFF the power breaker, disconnect the mains plug from the
outlet, and wait at least 30 seconds before removing the covers of the main
unit and performing internal wiring or other work.

11. When discarding any part of this system, follow all applicable local
regulations.

12. Wear protective gloves to ensure safety and prevent infection when
performing unit replacement, board replacement, internal wiring, or cleaning.

13. Wait at least 30 seconds after turning OFF the power breaker and
disconnecting the mains plug from the outlet before replacing the battery.
Otherwise, an electric shock may result.

6
No. 2Y730-502E*C

CONTENTS

Page

Safety Precautions --------------------------------------------------------------------------------------------------------------- 4

1. OUTLINE -------------------------------------------------------------------------------------------------------------------- 9

1.1 Introduction--------------------------------------------------------------------------------------------------------- 9

1.2 Tools and Measuring Devices Required -------------------------------------------------------------------- 9

1.3 Inspection Schedule---------------------------------------------------------------------------------------------12

1.4 Check Before Inspection ---------------------------------------------------------------------------------------12

1.5 Inspection Results -----------------------------------------------------------------------------------------------13

1.6 Concerning Safety Checks ------------------------------------------------------------------------------------13

1.7 About the Inspection Notification Message----------------------------------------------------------------13

2. SYSTEM SELF-DIAGNOSIS ------------------------------------------------------------------------------------------14

2.1 Diagnostics --------------------------------------------------------------------------------------------------------15

3. CHECK OF FUNCTIONS-----------------------------------------------------------------------------------------------16

3.1 Display of the Time and Date ---------------------------------------------------------------------------------17

3.2 Operation of the Trackball -------------------------------------------------------------------------------------17

3.3 Operation of the Fan --------------------------------------------------------------------------------------------17

4. CHECK OF MECHANICAL SAFETY --------------------------------------------------------------------------------18

4.1 Casters -------------------------------------------------------------------------------------------------------------19

4.2 Caster Attachment Sections ----------------------------------------------------------------------------------19

4.3 Monitor Mounting Mechanism --------------------------------------------------------------------------------19

4.4 Operating Panel --------------------------------------------------------------------------------------------------19

4.5 Mounting Mechanism for Peripheral Devices -------------------------------------------------------------19

4.6 Other Mechanical Parts ----------------------------------------------------------------------------------------19

4.7 External Appearance and Connections of the Cables --------------------------------------------------19

4.8 External Appearance of the Transducer -------------------------------------------------------------------19

5. CLEANING -----------------------------------------------------------------------------------------------------------------20

5.1 Interior of the System/Fan -------------------------------------------------------------------------------------21

5.2 Filter-----------------------------------------------------------------------------------------------------------------21

7
No. 2Y730-502E*C

CONTENTS - continued

Page

5.3 Covers/Panels ----------------------------------------------------------------------------------------------------23

5.4 Touch Panel-------------------------------------------------------------------------------------------------------23

5.5 Monitor -------------------------------------------------------------------------------------------------------------23

5.6 Transducer/Transducer Holder/Gel Holder----------------------------------------------------------------25

5.7 Video Printer (Option)-------------------------------------------------------------------------------------------25

5.8 MO Drive (Option) -----------------------------------------------------------------------------------------------25

5.9 CD-R Drive --------------------------------------------------------------------------------------------------------25

5.10 Cables/Cords -----------------------------------------------------------------------------------------------------25

6. CHECK OF ELECTRICAL SAFETY ---------------------------------------------------------------------------------26

6.1 Protective Conductor Resistance ----------------------------------------------------------------------------28

6.2 Ground Line Leakage Current --------------------------------------------------------------------------------32

6.3 Enclosure Leakage Current -----------------------------------------------------------------------------------32

6.4 Patient Leakage Current I -------------------------------------------------------------------------------------32

6.5 Patient Leakage Current III------------------------------------------------------------------------------------33

6.6 Patient Auxiliary Current ---------------------------------------------------------------------------------------34

7. CHECK OF 2D MODE AND RECORDING OF IMAGES -------------------------------------------------------35

7.1 Lateral Resolution/Axial Resolution -------------------------------------------------------------------------36

7.2 Penetration --------------------------------------------------------------------------------------------------------36

7.3 Speckle Characteristics ----------------------------------------------------------------------------------------36

7.4 Recording of Images --------------------------------------------------------------------------------------------36

8. REPLACEMENT OF PARTS ------------------------------------------------------------------------------------------37

8.1 Replacing the Battery -------------------------------------------------------------------------------------------38

8.2 Charging the Battery --------------------------------------------------------------------------------------------40

8.3 Replacing the Battery for the ATX Board ------------------------------------------------------------------40

9. VCR RECORDING -------------------------------------------------------------------------------------------------------45

PERIODIC INSPECTION CHECK SHEET --------------------------------------------------------------------------------47

8
No. 2Y730-502E*C

1. OUTLINE

1.1 Introduction

This manual describes the maintenance and inspection procedures for the diagnostic ultrasound
system SSA-680A/SSA-660A.

1.2 Tools and Measuring Devices Required

1.2.1 Phantoms and measuring devices

Product name Type Manufacturer Specifications/standards


Phantom for checking 000021 CARDIFF Specified type only allowed
axial and lateral resolution

1.2.2 Documents

Product name Type Specifications/standards


Maintenance manual SSA-680A, SSA-660A: 2Y730-502E
Service manual SSA-680A, SSA-660A (V6.00 or later):
2D730-519EN
SSA-660A (V5.00 or earlier):
2D730-502E
SSA-680A, SSA-660A (V6.00 or later):
2D730-520EN
SSA-660A (V5.00 or earlier):
2D730-503E

9
No. 2Y730-502E*C

1.2.3 Tools, measuring devices, and consumables for the diagnostic ultrasound system

Tool/measuring device Q'ty Remarks


Testing device for electrical safety check 1 set Toshiba-specified type
Ground circuit tester 1 set Toshiba-specified type
Resin container or stainless container 1 Large enough to accommodate two
transducer heads
Vinyl bags 1 Approx. 2 m × 2 m (0.3 mm thick)
Rubber gloves 1 pair
Vacuum cleaner 1
Blower (600W 0 to 16000 mm/min) 1 Makita 4014NV or equivalent
Cleaner
Brush 1 As small as a toothbrush
Air spray or blower brush 1
MO drive head cleaner 1 Fujitsu: Product No. 0240470,
Drawing No. CA90002-C980

Consumable parts Q'ty Remarks


Aluminum foil 1
Physiological saline solution (0.85 to 0.95% The saline container should be filled.

saline solution)
Glass cleaner 1 For cleaning the CRT monitor
Soft cotton cloth or lens cleaning paper 1 For cleaning the LCD surface
Mild detergent For cleaning the LCD monitor cover
and touch panel
Neutral detergent
Rag

CAUTION: 1. Do not use hydrocarbon glass cleaner or cleaner for the OA equipment to clean
the monitor. This substance may cause deterioration in the monitor.

2. Do not use solvents (such as paint thinner or benzine) or abrasive cleansers for
cleaning the unit. These substances may cause deterioration in the monitor.

10
No. 2Y730-502E*C

1.2.4 Standard set of tools, consumables

Tool Type Manufacturer Specifications/standards


Screwdriver (Not specified)
(various sizes and types)
Stubby Phillips screwdriver (large) (Not specified)
Wire nipper (Not specified)
Needle nose pliers (Not specified)
Diagonal pliers (Not specified)
Soldering iron, stand (Not specified)
Desoldering tool (Not specified)
Wrench set (Not specified)
Adjustable wrench (Not specified)
Tweezers (Not specified)
Hammer (Not specified)
Metric Allen wrench set (Not specified)
File (various types) (Not specified)
Cutter (Not specified)

Consumables Type Manufacturer Specifications/standards


Insulation tape (Not specified)
Hishilite tube, Empire tube (Not specified)
(various types)
Wiring material (Not specified)
(various types)
Screws, nuts, washers (Not specified)
(various types)
Solder (Not specified)
Paper file (various types) (Not specified)
Ethanol for cleaning and disinfection (Not specified)

11
No. 2Y730-502E*C

1.3 Inspection Schedule

The prescribed inspection schedule for the first year after installation is shown below.

Inspection schedule
Model Remarks
12 months after installation
Basically the system should be inspected every
SSA-680A/
c year. Clean it depending on its condition or the
SSA-660A
system environment.

The contents of contract-based inspections made subsequent to the first year after installation
differ depending on the maintenance contract. The relevant service company must determine the
contents and intervals of such inspections.

1.4 Check Before Inspection

1.4.1 Checking the system conditions

Check the items below with the user or in the log book to understand the system conditions.

(1) The occurrence of anything out of the ordinary while the system is in operation

(2) Abnormalities occurring intermittently

(3) History of problems

(4) Requests from the user

1.4.2 Checking the operating environment

Check the temperature and humidity in the examination room. Measurements related to electric
safety are particularly sensitive to the humidity. These measurements are also sensitive to the
measurement location and measurement circuit.

If the insulation characteristics deteriorate due to system changes over time or system
malfunctions, the measurement fluctuations are likely to increase with the increase in humidity.

1.4.3 Checking the system settings

Check the preset conditions of the system, brightness and contrast of the monitor, and options
including the transducer. Make a hard copy of the preset conditions used normally and file it in
the log book.

1.4.4 Checking the system version

Check the system version in the service menu.

1.4.5 Checking the system operation

Check the system operation.

12
No. 2Y730-502E*C

1.5 Inspection Results

If any abnormalities are found in the inspection, instruct the user to stop using the system and to
request repair.

1.6 Concerning Safety Checks

The items shown in section 4 "CHECK OF MECHANICAL SAFETY", those shown in section 6
"CHECK OF ELECTRICAL SAFETY", and the item 5.2 "Filter" in section 5 "CLEANING" are
applicable.

1.7 About the Inspection Notification Message

The following message is displayed when the period of operation (factory setting: 600 hours) set in
the system has elapsed. The interval for display of this message can be changed. For details,
refer to the service manual.

The maintenance period coming.


Please contact your service representative.

13
No. 2Y730-502E*C

2. SYSTEM SELF-DIAGNOSIS

• Inspection procedure

Perform system self-diagnosis using Test Tools in the service menu.

• Flowchart of the inspection process

Diagnostics tests

14
No. 2Y730-502E*C

Corrective actions and


Item Code Inspection contents Procedures Tools
parts to be replaced
Self-diagnosis A01 2.1 Diagnostics (1) Start up the system with no transducers connected.
Display the service menu and select [Test Tools] →
[Diagnostics].
(2) Select "TI", "TX", "RX", and "RC" in the [Test Tools -
Diagnostics] window and click [Execute].
(3) Confirm that all items in the Result column are "OK".

15
No. 2Y730-502E*C

3. CHECK OF FUNCTIONS

• Flowchart of the work process

Display the time and date.

Operate the trackball.

Operate the fan.

16
No. 2Y730-502E*C

Corrective actions and


Item Code Inspection contents Procedures Tools
parts to be replaced
Functions B01 3.1 Display of the Time (1) Confirm that the time and date displayed on the
and Date monitor are correct.
(2) If they are incorrect, correct them.
B02 3.2 Operation of the (1) Set "Measurement" to ON.
Trackball
(2) Operate the measurement caliper vertically and
horizontally using the trackball. Confirm that the
caliper moves smoothly.
B03 3.3 Operation of the (1) Confirm that the fan installed in the main unit rotates
Fan when the main unit power is turned ON.
(2) Confirm that the fan makes no abnormal sounds
when it rotates.

17
No. 2Y730-502E*C

4. CHECK OF MECHANICAL SAFETY

• Evaluation

Evaluation is performed by visual and operational checks.

• Flowchart of check procedures

Check the casters.

Check caster attachment sections.

Check the monitor mounting mechanism.

Check the operating panel.

Check the peripheral device mounting mechanisms.

Check the other mechanical parts.

Check the external appearance and connections of each cable.

Check the external appearance of the transducer.

If any of the check results are unacceptable, the system is abnormal. Stop using the system and
take appropriate measures.

18
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Mechanical C01 4.1 Casters Confirm that the four casters are straight and free from excess
safety play. Also, confirm that the caster lock/unlock mechanism
functions correctly.
C02 4.2 Caster Attachment Sections When the system is used at a site in which the system is moved
frequently or moved over a route with many differences in level,
confirm that the caster attachment sections of the chassis are
not distorted or cracked.
* If the system cannot be moved smoothly in the straight
direction, there is possibility of an abnormality such as
distortion or cracks in casters or caster attachment sections.
C03 4.3 Monitor Mounting Mechanism (1) Confirm that the monitor is not tilted and does not show
excessive play.
(2) Confirm that the monitor can be moved and swiveled
normally.
C04 4.4 Operating Panel Confirm that the panel is not tilted and does not show excessive
play.
C05 4.5 Mounting Mechanism for Confirm that the peripheral devices are mounted correctly.
Peripheral Devices
C06 4.6 Other Mechanical Parts Confirm that the suspension mechanism and the external
panels are free from abnormalities.
C07 4.7 External Appearance and (1) Confirm that the cables do not have any cracks or damage.
Connections of the Cables
(2) Confirm that the connector for each cable is fixed securely
in the unit.
(3) Confirm that the cables are grounded correctly.
C08 4.8 External Appearance of the Confirm that the transducers do not have any cracks, peeling,
Transducer looseness of parts, or damage.
If a TEE transducer is used, perform the insulation resistance
test. (Refer to the safety kit operation manual supplied with the
transducer.)

19
No. 2Y730-502E*C

5. CLEANING

• Flowchart of cleaning process

Disconnect the power cable from the outlet.

Remove the accessories.

Remove the covers of the main unit and monitor.

Clean the interior of the system and the fan.

Clean the filter.

Clean the covers and panels.

Clean the monitor.

Clean the transducer, transducer holder, and gel holder.

Clean the peripheral devices and accessories.

Clean the power supply cord and the cables.

20
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Cleaning D01 5.1 Interior of the (1) Disconnect the power plug from the outlet. Phillips screwdriver
System/Fan
(2) Remove accessories such as the transducer, transducer holder,
printer, and VCR from the main unit.
(3) Remove the monitor cover.
(4) Remove the covers of the main unit so that the PWBs and the interior
of the power supply are visible. Do not remove any PWBs unless it is
abnormal.
(5) Cover the main unit with a big vinyl bag. Vinyl bag
Size 2 × 2 m
(6) Put the nozzle of the vacuum cleaner in the vinyl bag and turn it ON to Vacuum cleaner
vacuum dust.
(7) On the vinyl bag, make an opening as large as can allow the nozzle of Blower
the blower to be inserted and blow out dust on the monitor, fan, PWBs, Makita 4014NV 600W
rack, power supply, etc. from top to bottom. 0 to 16000 mm/min
(8) Remove the vinyl bag gently, taking care not to spread the dust on the
inner surface of the bag.
(9) Vacuum all remaining dust with the vacuum cleaner. Vacuum cleaner
D02 5.2 Filter (1) Cleaning the air filters of the rack section
(a) Remove the air filters from the cover.
(b) Remove dust from the mesh grids using a vacuum cleaner at a
location away from the system.
(c) Insert the air filters fully into the slots in the cover.

21
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Cleaning D02 5.2 Filter (2) Cleaning the air filters of the power supply unit
(a) Remove the Velcro tapes and remove the air filters from the unit.
(b) Remove dust from the mesh grids using a vacuum cleaner at a
location away from the system.
(c) Reinstall the air filters at their original locations and secure them
with the Velcro tapes.

Slot

Air filter of the rack section

Velcro tape

Slot

Air filter of the power


supply unit

CAUTION: Do not start the system with the filters removed.

22
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Cleaning D03 5.3 Covers/Panels Wipe the covers and panels gently using a piece of soft dry cloth.
If necessary, spray neutral detergent onto a piece of soft, finely woven cloth
and use it to clean the covers and panels. Use cotton swabs to clean areas
around the switches and keys on the main panel.
D04 5.4 Touch Panel Clean the surface sheet using a commercially available eyeglass cleaning
cloth (non-woven rayon cloth etc.) or a soft dry cloth. If necessary, use a
piece of soft cloth moistened with a small amount of ethyl alcohol.
D05 5.5 Monitor (1) CRT Monitor
Wipe the surface of the monitor and the cover.
Spray glass cleaner onto a piece of soft, finely woven cloth and use it
to clean the monitor gently.
(2) LCD Monitor
Monitor : Use a soft cotton cloth or lens cleaning paper to clean the
LCD surface.
Cover : Clean the monitor cover using a soft cloth moistened with
mild detergent.

23
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

CAUTION: 1. Do not hit or apply strong pressure to the LCD


surface during cleaning. Doing so may
damage the monitor.

2. Never use a stiff or hard cloth to clean the LCD


monitor.
Doing so may damage the monitor cover or
LCD surface.

3. If stains on the LCD surface are difficult to


remove, wipe it gently with a soft cloth slightly
moistened with water. Detergents or chemical
agents should not be used because they may
damage the monitor.

4. Do not attempt to clean the LCD monitor with


organic solvents such as benzine, paint
thinner, and alcohol or abrasive cleaners.
Such agents may damage the monitor cover or
LCD surface.
Also, do not use mild detergent to clean the
LCD surface.

5. Do not allow liquid droplets to remain on the


LCD surface.
The droplets may cause small stains when they
dry or may enter the LCD surface, possibly
resulting in failure.

24
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Cleaning D06 5.6 Transducer/ Clean the transducers according to the operation manual supplied with Toothbrush-size brush
Transducer Holder/ each transducer. Blower brush or LPG
Gel Holder Clean the transducer holder and gel holder with a piece of soft dry cloth. gas spray
If necessary, use a piece of soft cloth moistened with water. Phillips screwdriver
D07 5.7 Video Printer Clean the thermal head of the printer according to the operation manual for
(Option) the printer.
D08 5.8 MO Drive Clean the MO drive using the MO drive head cleaner.
(Option)
D09 5.9 CD-R Drive Clean the CD-R drive according to the operation manual provided with the
CD-R unit.
D10 5.10 Cables/Cords Wipe off all dirt from the main unit power supply cord and the signal cables Neutral detergent
using neutral detergent.

25
No. 2Y730-502E*C

6. CHECK OF ELECTRICAL SAFETY

• Check procedures

For the measurements below, use the Toshiba-specified measuring devices such as the testing
device for electrical safety check, type ESA-1. For details of the items below, refer to the
maintenance and inspection manual (2Y549-903E).

• Flowchart of check procedures

Measure the protective conductor resistance.

Measure the ground line leakage current.

Measure the enclosure leakage current.

Measure patient leakage current I.

Measure patient leakage current III.

Measure the patient auxiliary current.

If any of the check results are unacceptable, the electrical safety of the system is not ensured. Stop
using the system and take appropriate measures.
If the check result for patient leakage current III does not meet the standard, replace the sensors.

26
No. 2Y730-502E*C

• Measurement procedure for items 6.2 to 6.6

1. Measure the ground line leakage current, enclosure leakage current, patient leakage currents
I and III, and the patient auxiliary current with normal power input polarity A (normal condition
A).
Patient leakage current III must be measured with each of the applied voltage polarities A' and
B'.

2. Turn the protective grounding OFF using the testing device (single fault condition A), and
measure the enclosure leakage current, patient leakage current I, and the patient auxiliary
current.

3. Turn OFF the power switch and the breaker of the system, and turn OFF power output polarity
A using the testing device. Turn ON the protective grounding. Then, turn ON the breaker
(single fault condition B) and measure the ground line leakage current, enclosure leakage
current, patient leakage current I, and the patient auxiliary current.

4. Turn OFF power output polarity B using the testing device (single fault condition C). Measure
the ground line leakage current, enclosure leakage current, patient leakage current I, and the
patient auxiliary current.

5. Turn the system breaker OFF. Turn ON the breaker and power switch of the system (normal
condition B) with power input polarity B of the testing device. Measure the ground line leakage
current, enclosure leakage current, patient leakage currents I and III, and the patient auxiliary
current.
Patient leakage current III must be measured with each of the applied voltage polarities A' and
B'.

6. Turn the protective grounding OFF using the testing device (single fault condition D), and
measure the enclosure leakage current, patient leakage current I, and the patient auxiliary
current.

Protective grounding
Power input polarity Power conducting wires
wire
Normal condition A A Connected Connected
Normal condition B B Connected Connected
Single fault condition A A Connected Broken
Single fault condition B A Single wire broken Connected
Single fault condition C B Single wire broken Connected
Single fault condition D B Connected Broken

27
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Electrical safety E01 6.1 Protective Perform measurements while the system is not in operation.
Conductor
Resistance
From the transducer hanger mounting screw to the ground pin of the 3-pin 0.2 Ω or less
power plug of the system.

Measurement points for the protective conductor resistance check (SSA-680A)

Protective ground circuit tester

28
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Measurement points for the protective conductor resistance check (SSA-660A)

Protective ground circuit tester

29
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Mesurement points for the ground line leakage current check (SSA-680A)
Testing devise for the electrical safty check
Electrical outlet
ECG terminal
A
B

C1

B1,C1,E1,F1
G1,G2.
G1 E1
G2 F1

G1
G2

30
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Mesurement points for the ground line leakage current check (SSA-660A)
Testing devise for the electrical safty check
Electrical outlet
ECG terminal
A
B

C1

B1,C1,E1,F1
G1,G2.
G1 E1
G2 F1

G1
G2

31
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Electrical safety E02 6.2 Ground Line Perform measurements while the system is in operation. Normal:
Leakage Current 0.5 mA or less
(1) Connect the 3-pin power plug of the system to the outlet on the testing
device for the electric safety check (referred to hereinafter as testing Single fault:
device) (B1). 1 mA or less
E03 6.3 Enclosure Leakage Perform measurements while the system is in operation. Normal:
Current 0.1 mA or less
(1) Connect the transducer hanger mounting screw to the testing device
with the supplied cable and connect the Single fault:
3-pin power plug of the system to the outlet on the testing device (C1). 0.5 mA or less
E04 6.4 Patient Leakage Perform the following measurements while the system is in operation. Normal:
Current I AC 0.1 mA or less
(1) Wrap the transducer with aluminum foil and place it in a container filled DC 0.01 mA or less
with physiological saline solution. Connect the aluminum foil to
measuring point A on the testing device with the supplied cable and Single fault:
connect the 3-pin power plug of the system to the outlet on the testing AC 0.5 mA or less
device (E1). DC 0.05 mA or less

CAUTION: There must be some physiological saline solution


between the transducer and the aluminum foil.

(2) Wrap the electrode and sensors listed below separately with aluminum
foil.
Connect each aluminum foil to measurement point A on the testing
device with the supplied cable, and connect the 3-pin power plug of
the system to the outlet on the testing device.
(a) ECG electrode
(b) PCG sensor
(c) Pulse wave sensor
(d) Respiration sensor

32
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Electrical safety E05 6.5 Patient Leakage Perform the following measurements while the system is in operation. Single fault:
Current III 5 mA or less
(1) Wrap the transducer with aluminum foil and place it in a container filled
with physiological saline solution. Connect the aluminum foil to
measuring point A on the testing device with the supplied cable and
connect the 3-pin power plug of the system to the outlet on the testing
device (F1).
(2) Wrap the electrode and sensors listed below separately with aluminum
foil.
Connect each aluminum foil to measurement point A on the testing
device with the supplied cable, and connect the 3-pin power plug of
the system to the outlet on the testing device.
(a) ECG electrode
(b) PCG sensor
(c) Pulse wave sensor
(d) Respiration sensor

CAUTION: An AC voltage that is 10% greater than the line voltage is applied to the
electrodes and sensors. Follow the instructions below to prevent an
electric shock.
(1) Do not perform measurements with moist hands.
(2) Do not allow other persons to come close to the sensors during
measurement.
(3) Connect the testing device securely with aluminum foil. The
aluminum foil should enclose the electrode or sensor.

33
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Electrical safety E05 6.5 Patient Leakage CAUTION: There must be some physiological saline solution between the transducer and the
Current III aluminum foil.

E06 6.6 Patient Auxiliary Perform the following measurements while the system is in operation. Normal:
Current AC 0.1 mA or less
DC 0.01 mA or less
(1) Wrap the transducer with aluminum foil and connect the aluminum foil
to measuring point A on the testing device with the supplied cable. Single fault:
In addition, wrap each electrode of the ECG with aluminum foil and AC 0.5 mA or less
further wrap them together with aluminum foil. Then, connect the DC 0.05 mA or less
aluminum foil to measuring point B on the testing device with the
supplied cable (G1).
Connect the 3-pin power plug of the system to the outlet on the testing
device.
(2) Wrap the electrodes of the ECG with aluminum foil and connect one
electrode to measuring point A and the other to measuring point B on
the testing device (G2).
Connect the 3-pin power plug of the system to the outlet on the testing
device.

34
No. 2Y730-502E*C

7. CHECK OF 2D MODE AND RECORDING OF IMAGES

• System settings

Use the factory settings for items other than those described in this manual. If any settings are
changed for special reasons, record the settings as supplementary information.

• Recording and filing images

Print images and file them together with the recorded data.

• Flowchart of check procedures

Check the lateral resolution and axial resolution.

Check the penetration.

Check the speckle characteristics.

Record images and check them.

* Perform the inspections shown above for all transducers used by the user.

35
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Checking 2D F01 7.1 Lateral Resolution/Axial (1) Spread ultrasound gel on the phantom and apply the
mode and Resolution transducer to it.
recording
(2) Apply the transducer to the phantom so that the resolution
images
of the wire target is optimal.
(3) When the optimal image is obtained, freeze the image and
record it. (In DUAL-display mode, print out the image
together with the penetration image, which is obtained as
described below.)
F02 7.2 Penetration (1) Spread ultrasound gel on the phantom and apply the
transducer to it.
(2) Adjust the GAIN and STC so that the speckle for the
parenchyma is displayed at the deepest position.
(3) Measure the depth at the border between the noise and
parenchyma and record an image on which the measured
depth is displayed.
F03 7.3 Speckle Characteristics Evaluate the changes over time by periodically recording the
images as described above.
F04 7.4 Recording of Images File the images printed according to the procedure described
above.

36
No. 2Y730-502E*C

8. REPLACEMENT OF PARTS

Replace the periodically replaced parts.

Periodically replaced part Replacement interval


Battery 1 year
Battery for the ATX board 1 year

CAUTION: Wait at least 30 seconds after turning OFF the power breaker and
disconnecting the mains plug from the outlet before replacing the battery.
Otherwise, an electric shock may result.

37
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Replacement of G01 8.1 Replacing the Battery CAUTION: Do not perform this work in the standby status.
parts (Replacement interval: 1 year)
(1) Open the printer tray and remove one M4 screw <A>.
(2) Remove one cap <A> and remove one M4 screw <B>.
(3) Remove one cap <B> and remove one M4 screw <C>.
(4) Remove the caps <C> and <D> (one each) and loosen two
M4 screws <D>.
(5) Slide the left side cover toward the front and open it from
the rear to remove it.
(6) Remove the connector that is connected to the battery and
remove the battery.

38
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Replacement of G01 8.1 Replacing the Battery


parts
Screw <A>
Battery
(SBPP4-10SCR)

Cap <C>

Printer tray

Operation
check LED
Cap <D>

Screw <D>
(SBPP4-10SCR) Cap <A>
Screw <B>
Left side cover (SBPP4-12SCR)

Cap <B> Screw <C>


(SBPP4-10SCR)

39
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Replacement of G02 8.2 Charging the Battery (1) Connect the power plug to the outlet.
parts
(2) Turn ON the circuit breaker.
(a) Charging starts automatically.
(b) The operation check LED (segment LED) blinks.
(c) Charging stops after 2 hours.
(d) To charge the battery fully, turn the circuit breaker
OFF and then ON again.
(3) After charging has been completed, confirm that the
operation check LED (segment LED) is lit.

NOTE: If the breaker is set to OFF, the battery discharges


and recharging is required.

G03 8.3 Replacing the Battery for the (1) Remove the CPU assy.
ATX Board
CAUTION: Do not perform this work in the standby status.

(a) Shut down the system.


If the system is ON : Shut down the system.
If the system is OFF : Start the system and then,
after it has started up
normally, shut it down.
(b) Turn OFF the breaker of the system.
(c) Remove the USB cable if required and then remove
the CPU assy.

40
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Replacement of G03 8.3 Replacing the Battery for the (2) Replace the battery.
parts ATX Board
(a) Remove the battery near the IDE connector on the
ATX board (refer to the figure below).
(b) Replace the battery with the new one.

Battery (CR2032)

41
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Replacement of G03 8.3 Replacing the Battery for the (3) Remove the HDD unit.
parts ATX Board
CAUTION: Set the date and time on the BIOS setting
screen before reinstalling the HDD unit.
If the system is started without setting the date
and time on the BIOS setting screen, the initial
setting time on the CPU board is set as the
system time.

(4) Start the system.

(a) Install the CPU board in the system.


Connect the USB cable if it was removed.
(b) Turn ON the breaker of the system.
(c) Press the power switch. (The power supply may be
turned ON again after it is turned OFF.)
(d) Press the [Delete] key while the logon screen is
displayed. The BIOS setting screen is displayed.

42
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools

Replacement of G03 8.3 Replacing the Battery for the (5) Perform BIOS setting.
parts ATX Board
(a) Check the date displayed on the main page and
change it if necessary.
"Date": Enter the date in the order month, day, year.
(The day of the week is automatically determined.)
"Time": Enter the time in the order hours, minutes,
seconds (24-hour time format).
<Change procedure>
• Selecting the page
Use the left and right arrow keys [←] and [→] to
display the page.
• Selecting the main item
Use the up and down arrow keys [↑] and [↓] to
select "Date" or "Time".
• Selecting the specific item in the main item
Press the [Enter] key to select each item in the
order month, day, year or in the order hours,
minutes, seconds.
• Changing the value for the item (day, year, hours,
minutes, seconds)
Press [PageUp]/[PageDown] to increase/decrease
the value and then press the [Enter] key.
It is also possible to set the value by entering it
directly from the keyboard and then pressing the
[Enter] key.
• Changing the value for the item (month)
Press [PageUp]/[PageDown] to change the name of
the month (for example, Jan. → Feb. → Mar.).
After selecting the month name, press the [Enter]
key.

43
No. 2Y730-502E*C

Item Code Inspection contents Procedures Tools


Replacement of G03 8.3 Replacing the Battery for the (b) Load the default settings.
parts ATX Board
<1> Press the right arrow key [→] to select the EXIT
page.
<2> Press the down arrow key [↓] to select "Load
Optimized Default". Press the [Enter] key.
<3> A confirmation message is displayed. Enter "y"
(yes) and then press the [Enter] key.
<4> When loading is completed, the previous screen
is displayed again.
(c) Save the settings.
<1> Select "Save & Exit Setup" on the EXIT page
and then press the [Enter] key.
<2> A confirmation message is displayed. Enter "y"
(yes) and press the [Enter] key.
<3> The system restarts automatically.
<4> Confirm that the message "OS not found" is
displayed and then terminate the system by
holding down POWER for at least 5 seconds.

(6) Reinstall the HDD unit.


(7) Start the system again. Confirm that the system starts
normally and then shut it down.

44
No. 2Y730-502E*C

9. VCR RECORDING

Perform the following checks to ensure that the VCR is ready for recording and playback
(Code: H01).

<<Check procedure>>

The vertical and horizontal scales of a stored image are measured using the manual calibration function.

(1) Start the system and connect a body-surface transducer such as the PVT-375BT to the system.
Display a 2D-single image and set the DEPTH to 20 cm. Record the image on the VCR.

(2) Play back the recorded image. Using the manual calibration function, set a calibration marker 100
mm in length using the vertical scale as the reference and enter the length of the calibration marker
in the box on the right of [2D Distance]. Measurement becomes possible.

45
No. 2Y730-502E*C

(3) Measure the distance corresponding to 10 graduations on the horizontal scale and confirm that the
error is within ±5% of the distance set on the vertical scale (95 mm to 105 mm).

(4) If the error is more than ±5%, set the calibration marker again carefully. If the error is still greater
than ±5%, check with the user whether the error is acceptable for the intended use. If the error is not
acceptable, instruct the user not to use the results.

46
No. 2Y730-502E*C

PERIODIC INSPECTION CHECK SHEET

Item No. Inspection contents Result


System self-diagnosis A01 Diagnostics † OK † NG
Check of functions B01 Display of the time and date † OK † NG
B02 Operation of the trackball † OK † NG
B03 Operation of the fan † OK † NG
Mechanical safety C01 Casters † OK † NG
C02 Caster attachment sections † OK † NG
C03 Monitor mounting mechanism † OK † NG
C04 Operating panel † OK † NG
C05 Mounting mechanism for peripheral devices † OK † NG
C06 Other mechanical parts † OK † NG
C07 External appearance and connections of the cables † OK † NG
C08 External appearance of the transducer † OK † NG
Cleaning D01 Interior of the system/fan † OK † NG
D02 Filter † OK † NG
D03 Covers/panels † OK † NG
D04 Touch panel † OK † NG
D05 Monitor † OK † NG
D06 Transducer/transducer holder † OK † NG
D07 Video printer † OK † NG
D08 MO drive † OK † NG
D09 CD-R drive † OK † NG
D10 Cables/cords † OK † NG
Electrical safety E01 Protective conductor resistance † OK † NG
E02 Ground line leakage current † OK † NG
E03 Enclosure leakage current † OK † NG
E04 Patient leakage current I † OK † NG
E05 Patient leakage current III † OK † NG
E06 Patient auxiliary current † OK † NG
Checking 2D mode and F01 Lateral resolution/axial resolution † OK † NG
recording images
F02 Penetration † OK † NG
F03 Speckle characteristics † OK † NG
F04 Recording of images † OK † NG
Replacement of parts G01 Replacing the battery † OK † NG
G02 Charging the battery † OK † NG
G03 Replacing the battery for the ATX board † OK † NG
VCR recording H01 VCR recording † OK † NG

47 E
LEGAL MANUFACTURER

1385, SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN 324-8550, JAPAN

"Made for Life" is a trademark of Toshiba Medical Systems Corporation.

Potrebbero piacerti anche