Sei sulla pagina 1di 66

Chapter 9: Mammography

Slide set of 66 slides based on the chapter authored by


M.J. Jaffe and A.D.A. Maidment
of the IAEA publication (ISBN 978-92-0-131010-1):

Diagnostic Radiology Physics:


A Handbook for Teachers and Students

Objective:
To familiarize the student with the requirements and principles of
imaging of the breast using X ray mammography.
Slide set prepared
by C P Lawinski (BSc, MSc, MPhil),
London, UK

IAEA
International Atomic Energy Agency
CHAPTER 9. TABLE OF CONTENTS

9.1. Introduction
9.2. Radiological requirements for mammography
9.3. X ray equipment
9.4. Image receptors
9.5. Display of mammograms
9.6. Breast tomosynthesis
9.7. Breast CT
9.8. Computer-aided diagnosis
9.9. Stereotactic biopsy systems
9.10. Radiation dose
Bibliography
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 2
9.1. INTRODUCTION

Breast cancer is a major killer of women


Over 1.38 million women diagnosed with breast cancer
internationally in 2008
• More than 458,000 deaths
Causes not currently known
Mortality can be significantly reduced if disease is detected
at an early stage

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 3
9.2. RADIOLOGICAL REQUIREMENTS FOR MAMMOGRAPHY

Mammography is a radiographic (X ray) procedure


optimized for examination of the breast
Highly effective means of detecting early-stage breast
cancer
Mammography is used both for
• Investigating symptomatic patients (diagnostic mammography)
• Screening of asymptomatic women (selected age groups)
A typical mammographic screening examination consists
of one or more commonly two views of each breast
• Cranio-caudal (CC) and medial-lateral oblique (MLO)
Other uses
• Pre-surgical localisation and guidance of biopsies.
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 4
9.2. RADIOLOGICAL REQUIREMENTS FOR MAMMOGRAPHY

A medial-lateral oblique
(MLO) mammogram is
shown
• The pectoralis muscle is
visualized
• Characteristic benign
calcifications are seen

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 5
9.2. RADIOLOGICAL REQUIREMENTS FOR MAMMOGRAPHY

Breast tissues
intrinsically lack subject
contrast
Low-energy X ray
spectrum required
Emphasises
compositional
differences of the
breast Dependence of the linear X ray attenuation coefficient ( µ) on
X ray energy.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 6
9.2. RADIOLOGICAL REQUIREMENTS FOR MAMMOGRAPHY

Sufficient spatial resolution


• Details possibly as fine as 50 µm must be adequately visualised
Adequate contrast in image
• Low-energy X ray spectra
Broad dynamic range
• Required due to composition of the breast and age-dependent
changes in the breast
Lowest absorbed dose compatible with adequate
diagnostic image quality

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 7
9.3. X RAY EQUIPMENT

Specialised gantry to
accommodate the Schematic of a
breast mammography
imaging system.
• Rotation and vertical
movement
Specialised beam
geometry
• Improves visualisation
of chest wall edge
System geometry for
image acquisition
showing a) correct
alignment and b)
missed tissue
associated with
incorrect alignment.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 8
9.3. X RAY EQUIPMENT
9.3.1. Tubes, filters and spectra

X ray generator
• High frequency
• Near constant potential
waveform
X ray tube
• Rotating anode
• Dual focus 0.3/0.1 mm
• Beryllium exit window (low
attenuation)
FID (focus image distance)
generally in the range 60 to The geometry of an X ray tube. The perpendicular
line abuts the chest wall. The reference axis on a
65 cm particular system will be specified by the
manufacturer.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 9
9.3. X RAY EQUIPMENT
9.3.1. Tubes, filters and spectra

X ray spectrum should provide a range of energies that


give an appropriate compromise between radiation dose
and image quality for the tissues under examination
X ray spectrum determined by target material, filter
material, and tube voltage (kV)
For screen-film mammography optimum beam energy lies
between 18 and 23 keV depending on breast thickness
and composition
• Characteristic X rays from molybdenum and rhodium are suitable
Higher energies may be more optimal for digital
mammography

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 10
9.3. X RAY EQUIPMENT
9.3.1. Tubes, filters and spectra

Metallic filters used in mammography


Molybdenum (Mo) filter (30 to 35 µm thick) commonly
employed with Mo anode
Filter acts as energy window
• Greater attenuation of X rays at low energies and at energies above
the K-absorption edge of Mo at 20 keV
• Mo characteristic X rays from the target and X rays of similar energy
produced by bremsstrahlung pass through the filter
• Resultant spectrum enriched with X rays in the range 17 to 20 keV
Higher energies are desirable for imaging thick, dense
breasts
• Use of Mo/Rh (molybdenum/rhodium) and Rh/Rh target/filter
combinations

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 11
9.3. X RAY EQUIPMENT
9.3.1. Tubes, filters and spectra

Examples of mammographic X ray spectra.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 12
9.3. X RAY EQUIPMENT
9.3.2. Compression

Compression should be firm but not painful


Reasons for applying compression
• Reduces superposition of tissues
• Decreases ratio of scattered to transmitted radiation reaching the
image receptor
• Decreases the distance from any plane within the breast to the
image receptor reducing geometric unsharpness
• Compressed breast provides lower overall attenuation allowing
radiation dose to be reduced
• Compressed breast provides more uniform attenuation over the
image reducing the exposure range which must be recorded
• Provides a clamping action reducing anatomical motion during the
exposure reducing image unsharpness

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 13
9.3. X RAY EQUIPMENT
9.3.3. Grids

Scattered radiation reduces image quality


Use of grid significantly decreases ratio of scattered to
transmitted radiation reaching the image receptor
Focused linear grids (integral part of the system)
Grid moves during exposure to blur the image of the
grid septa
• Motion must be uniform and of sufficient amplitude to avoid non-
uniformities in the image
Bucky factor (increase in dose due to use of grid) can
be as large as 2 to 3
• Justified by improvement in image quality

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 14
9.3. X RAY EQUIPMENT
9.3.4. Automatic exposure control

All modern mammography units are equipped with


automatic exposure control (AEC)
Essential in order to provide the optimum dose to the
image receptor
• Target optical density for screen-film mammography
• Target SNR (signal-to-noise ratio) or preferably SDNR (signal-
difference-to- noise ratio) for digital mammography

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 15
9.3. X RAY EQUIPMENT
9.3.4. Automatic exposure control

For screen-film mammography and cassette-based digital


systems AEC sensor located behind image receptor to
avoid a shadow on the image
Sensor terminates exposure when pre-set amount of
radiation received by the image receptor
Location of sensor adjustable
• Can be positioned below appropriate region of the breast
AEC generally microprocessor controlled
• Correction for reciprocity law failure of the film
• Automatic selection of exposure parameters (kV, filter, target)
depending on breast thickness and composition
• Sensing of breast thickness (compression device) and attenuation (short
(typically <100 ms) X ray pre-exposure)

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 16
9.3. X RAY EQUIPMENT
9.3.4. Automatic exposure control

For digital mammography the digital detector can act as


the AEC sensor
Pre-exposure concept typically used
• Entire low-dose image captured by the digital detector
• Image analysed to determine the overall SDNR or minimum SDNR
over small (~1cm2) region-of-interest (roi) in the image
• Target, filter and kV selected automatically to give desired SDNR
when main exposure performed
• As digital detectors can be operated at a wide range of input dose
levels, possible to optimize imaging according to a priority of
SDNR, low dose or a combination.
Development in this area is on-going
• Location of the edges or critical areas of breast identified
automatically
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 17
9.3. X RAY EQUIPMENT
9.3.5. Magnification mammography

Magnification mammography can be used to improve the


diagnostic quality of the image
Breast supported above the image receptor
• Focus object distance reduced
• Object to image receptor distance increased
• Magnification results
Benefits of magnification mammography
• Increased SNR
• Improved spatial resolution
• Dose-efficient scatter rejection.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 18
9.3. X RAY EQUIPMENT
9.3.5. Magnification mammography

Main benefit of magnification is to increase the size of the


projected anatomic structures compared to the granularity
of the image
• SNR in the image is improved
• Improvement can be valuable, particularly for the visualization of
fine calcifications and spiculations
Magnification in digital mammography
• Film grain noise eliminated
• Limiting spatial resolution of detector lower than that provided by
the screen-film image receptor
• Benefits of magnification may be different in nature
• Increase in projected size of anatomical features does improve the
effective resolution of the detector, which in some cases is a
limiting factor
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 19
9.3. X RAY EQUIPMENT
9.3.5. Magnification mammography

Spatial resolution in magnification mammography is limited


by focal spot size
• Use of a small spot (typically 0.1 mm) is critical
As the breast is closer to the X ray source in magnification
mammography
• Dose to breast increases (compared to contact mammography)
• Air gap between the breast and image receptor provides some
scatter rejection
• Anti-scatter grids not employed for magnification (partially offsets
increase in dose)

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 20
9.3. X RAY EQUIPMENT
9.3.5. Magnification mammography

A suspicious region is visible in the lower aspect of the mammogram (left). A magnified
image of this region obtained with focal compression shows an obvious mass (right).

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 21
9.4. IMAGE RECEPTORS

Screen-film mammography
Digital mammography
• Area detectors
• Indirect detectors
• Direct detectors
• Photo-stimulable phosphors (computed radiography or CR)
• Scanning detectors
• Photon counting detectors

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 22
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Single back intensifying screens used with single emulsion


radiographic film enclosed in a light-proof cassette
High resolution fluorescent intensifying screen
• Absorbs X rays
• Converts the pattern of X rays into an optical image
Two sizes of film typically available
• 18 cm x 24 cm and 24 cm x 30 cm
Customary to use the smallest possible size which
ensures complete coverage of the breast
• Superior breast positioning and compression
Women with large breasts, may require multiple films to
image the breast fully

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 23
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Screen and film are arranged in the cassette as shown


(next slide)
X rays must pass through cover of the cassette and the
film to impinge upon the screen
• As absorption is exponential, a larger fraction of the X rays are
absorbed and converted to light near the entrance surface of the
screen
• By minimizing the distance that the light must travel before being
collected by the film, blurring due to lateral spreading is reduced
• Spatial resolution is improved
• To further discriminate against light quanta travelling along long
oblique paths, the phosphor material may be treated with a dye
which absorbs much of this light, giving rise to a sharper image

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 24
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Configuration for a mammographic screen-film image


receptor. A single-emulsion radiographic film is held in close
contact with a fluorescent screen in a light-proof cassette.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 25
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Typical phosphor used for screen-film mammography is


gadolinium oxysulphide (Gd2O2S:Tb)
Phosphor material is dense
• Quantum detection efficiency (fraction of incident X rays which
interact with the screen) is reasonably high
• Approximately 60% for a typical screen thickness and X ray spectrum
Conversion efficiency (fraction of absorbed X ray energy
converted to light) exceeds 10% (high for a phosphor)
Photographic film emulsion for mammography is matched
to be sensitive to:
• Spectrum of light emitted from the particular phosphor screen
• Range of X ray fluence exiting the breast

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 26
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Important to examine the overall characteristics of the


screen and film combination rather than those of the
individual components
Compression of the breast reduces overall range of X ray
fluence exiting the breast
• Allows films with high gradient to be used
• Enhances contrast between subtly varying soft-tissue structures
In addition, mammography film has a high Dmax (4.0 to 4.8
OD)
• Maximizes exposure latitude over which the high gradient exists
• Important near the periphery of the breast (skin edge) where its
thickness decreases rapidly
• Nevertheless, some regions of the mammogram will generally be
under or overexposed, i.e. rendered with sub-optimal contrast.
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 27
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Characteristic curve of a film emulsion used for mammography.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 28
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Mammography film is processed in an automatic processor


similar to that used for general radiography
Important that the development temperature, time and rate
of replenishment of the developer chemistry are compatible
with the type of film emulsion used and are designed to
maintain good contrast of the film
Daily quality assurance is required in mammography to
ensure on-going optimal performance
In screen-film mammography, the film must act as the
image acquisition detector as well as a storage and display
device

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 29
9.4. IMAGE RECEPTORS
9.4.1. Screen-film mammography

Several factors associated with screen-film mammography


can limit the ability to display the finest or most subtle details
Sigmoidal shape of the characteristic curve results in limited
latitude (range of X ray exposures over which the film display
gradient is significant)
• If a tumour is located in a more lucent or more opaque region of the
breast, contrast displayed may be inadequate due to limited gradient
of the film
• Particularly of concern in patients whose breasts contain large
amounts of fibroglandular tissue (dense breasts)
Effect of fixed pattern noise due to the granularity of the
phosphor screen and film emulsion
• Can impair the detectability of microcalcifications and other fine
structures
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 30
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Digital mammography introduced commercially in 2000


• Able to overcome many of the technical limitations of screen-film
mammography
In digital mammography, image acquisition, processing,
display, and storage are performed independently, allowing
optimisation of each
Acquisition performed with low-noise X ray detectors with
wide dynamic range
As the image is stored digitally:
• It can be displayed with contrast independent of the detector
properties
• Image processing techniques that are found to be useful can be
applied prior to image display
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 31
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Challenges in creating a digital mammography system


with improved performance are mainly related to the X ray
detector and the display device.
The detector should have the following characteristics:
• Efficient absorption of the incident radiation beam
• Linear or logarithmic response over a wide range of incident
radiation intensity
• Low intrinsic noise and little-to-no fixed-pattern noise to ensure that
images are X ray quantum noise limited
• Limiting spatial resolution of the order of 5 to10 cycles/mm (50
to100 µm sampling)
• Can provide at least an 18x24 cm and preferably a 24x30 cm field
size
• Can image immediately adjacent to the chest wall
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 32
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Two main approaches in detector development


Area detectors
• Entire image is acquired simultaneously
• Fast image acquisition
• Can be used with conventional design of mammography X ray unit
equipped with a grid to reduce scatter
Scanning detectors
• Image is obtained by scanning the X ray beam and detector across
the breast
• Mechanically complex and longer acquisition times
• Use relatively simple detector(s)
• Good intrinsic scatter rejection

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 33
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Typical area detectors are based of an amorphous silicon


thin-film transistor (TFT) panel
• Contains a rectangular matrix of 2000 to 3000 columns by 3000 to
4000 rows of detector elements (dels)
• Each del is connected to electrical lines running along each row
and column by a TFT switch
• This array is covered by a phosphor or a photoconductor X ray
detector

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 34
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

In “indirect” detectors each del includes both a light-


sensitive photodiode and a TFT switch
The array is covered with a phosphor layer
• Typically thallium-activated CsI
X rays transmitted by the breast are absorbed by the
phosphor and light produced is converted in the
photodiode to charge which is stored in its capacitance
After the X ray exposure
• Readout signals sent sequentially along the lines for each row
activate corresponding switches
• The charge is transferred down the columns to readout amplifiers
and multiplexers and digitized to form the image

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 35
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

This readout system allows the signals from all of the dels
to be read in a fraction of a second
• Fast image display
The needle-like phosphor crystals of CsI behave
somewhat like fibre-optics
• Conduct the light to the photodiodes with less lateral spread than
would occur with granular phosphors
• Allows the thickness of the phosphor to be increased relative to a
granular phosphor to improve the quantum detection efficiency of
the detector without excessive loss of spatial resolution

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 36
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

In “direct” detectors a similar readout strategy is used but the


phosphor is replaced with an X ray absorber composed of
amorphous selenium which is a photoconductor
The energy of the absorbed X rays causes the liberation of
electron-hole pairs in the selenium
The charged particles are drawn to the opposite faces of the
detector by an externally applied electric field
To collect the signal an array of electrode pads (rather than
photodiodes) forms the dels
Unlike the phosphor-based detectors, the electric field can
be tailored to collect the charge with minimal lateral spread

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 37
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

This allows the use of a relatively thick detector to achieve


excellent quantum detection efficiency without significant
reduction in resolution at near normal incidence
Other materials in which X ray energy is directly converted
to charge are under development
These materials include:
• Lead iodide, zinc cadmium telluride and thallium bromide
• The higher atomic number of these materials allow the thickness of
the X ray converter to be reduced
• This can mitigate against the degradation of the MTF due to the
oblique incidence of the X rays

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 38
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Computed radiography (CR) systems can be used for


mammography
Employs a photo-stimulable phosphor (PSP) plate housed
in a light-proof cassette
• When exposed to X rays, electrons in the crystalline material are
excited and subsequently captured by traps in the phosphor
• After exposure the plate is placed in a reader device and scanned
with a laser beam
• The energy of the laser light stimulates the traps to release the
electrons
• The transition of these electrons through energy levels in the
phosphor crystal results in the emission of light
• The light is collected by a photomultiplier tube, the signal digitised
and attributed to a particular pixel in the image
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 39
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

For Mammography CR systems the resolution of the


image is determined by:
• Size of the scanning laser beam
• Underlying scatter of the readout laser light in the phosphor
• Distance between sample measurements

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 40
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Mammography CR systems differ from the general


radiography CR systems in several key areas
• Mammography CR system is designed for higher spatial resolution
• Uses a thinner phosphor material and is scanned with finer
sampling pitch (typically 50 µm)
However, the result is less signal per pixel
To overcome this limitation various innovations have been
developed to improve light coupling and reduce readout
noise, including:
• Dual-sided readout of the phosphor plates
• Needle-like phosphors which permit the use of thicker detectors
having superior quantum detection efficiency

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 41
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Detector systems discussed so far acquire the image by


integrating the signal from a number of X ray quanta
absorbed in the detector and digitizing this signal
The image noise from these systems depends on:
• Poisson X ray quantum fluctuations associated with X ray
absorption
• Additional noise sources associated with the production of the
converted electronic signal
• These noise sources can arise from:
• Fluctuation in the amount of light produced in a phosphor in response to
absorption of an X ray of a particular energy or from the X ray spectrum itself

As an alternative it is possible to count the number of


interacting quanta directly, thereby avoiding these
additional noise sources
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 42
9.4. IMAGE RECEPTORS
9.4.2. Digital mammography

Typically quantum-counting detectors employ a geometry in


which the X ray beam is collimated into a slot or multi-slit
format and scanned across the breast to acquire the image
The detector can be based on:
• Solid-state approach (electron-hole pairs are produced in a material
such as crystalline silicon)
• Pressurized gas (the signal is in the form of ions formed in the gas)
• Collection of charge signal plus amplification produces a pulse for each
interacting X ray quantum and pulses are counted to create the signal
As the beam is collimated to irradiate only part of the breast at
a time the system has:
• Good intrinsic scatter rejection without the need for a grid
• Increased dose efficiency
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 43
9.5. DISPLAY OF MAMMOGRAMS

Film mammograms
• Specially designed transillumination devices
Digital mammograms
• Computer displays and workstations

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 44
9.5. DISPLAY OF MAMMOGRAMS
9.5.1. Display of film mammograms

Specially designed transillumination devices are available


for reading film mammograms
• The luminance levels must be appropriate for reading
mammograms and sufficient to illuminate areas of interest
(luminance of at least 3000 cd m-2)
• The illuminator surface should provide diffused light of uniform
brightness
Optimal viewing conditions essential for reviewing screen-
film mammograms
• Allows visualization of as much of the information recorded in the
mammogram as possible
Mammograms should be interpreted under conditions that
provide good visibility, comfort, and minimal fatigue

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 45
9.5. DISPLAY OF MAMMOGRAMS
9.5.1. Display of film mammograms

Contrast sensitivity of the eye (ability to distinguish small


differences in luminance) is greatest when surroundings
are of about the same brightness as the area of interest
Therefore to optimally see detail in a mammogram
• Glare should be reduced to a minimum
• Surface reflections should be avoided
• Ambient light level should be subdued and approximately that
reaching the eye through the mammogram
Important to have a variable brightness, high output light
source to view high optical density (dark) areas on the film
mammogram

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 46
9.5. DISPLAY OF MAMMOGRAMS
9.5.2. Display of digital mammograms

Display system plays a major role in influencing overall


performance of digital mammography
• Image quality presented to the film reader
• Ease of image interpretation
Images viewed on a computer (“softcopy”) display
• Cathode ray tube (CRT) or more commonly LCD monitor
Typical reporting workstation has matched pair of high
quality, high resolution monitors (normally 5 megapixel (MP))
• A 5 MP monitor is capable of displaying only a single mammogram at
full resolution
• Wide range of image manipulation tools available
• Brightness, contrast, zoom, , roam, scroll, etc.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 47
9.5. DISPLAY OF MAMMOGRAMS
9.5.2. Display of digital mammograms

“Hardcopy” images use a laser printer to produce a


printout of the digital image on transparent film
Image brightness and contrast usually adjusted before
printing out the image making use of the controls provided
at the acquisition workstation
“Hardcopy” images do not allowing control of image
processing operations during viewing
Therefore, it is strongly recommended that digital
mammography images are displayed and reviewed using
a high quality “softcopy” device

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 48
9.6. BREAST TOMOSYNTHESIS

In projection radiography tissue superposition can result in


a masking effect
Breast tomosynthesis can provide reconstructed planar
images of sections of the breast
• Can aid in reducing the masking effect
Breast tomosynthesis generally based on modified digital
mammography systems
• Planar digital imaging and tomosynthesis
• Tomosynthesis only

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 49
9.6. BREAST TOMOSYNTHESIS

X ray tube pivots about a point


Breast platform remains stationary
Detector usually stationary but may also move
Series of low-dose projection images (typically 9 to 25)
acquired over a limited range of angles (±7° to ±30°)
X ray spectrum
• Higher energy typically employed (e.g.W/Al)
X ray tube movement
• Continuous exposure or series of discrete exposures (“step and
shoot”)
Total acquisition time must be minimized
• Possible image degradation due to patient motion
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 50
9.6. BREAST TOMOSYNTHESIS

Planar cross-sectional images are reconstructed from the


projections using filtered back projection or an iterative
reconstruction algorithm
Spatial resolution of tomosynthesis is anisotropic
• Highest resolution in-plane
• Relatively poor resolution between planes
Reconstructed voxels are generally non-isotropic
• Pixel size approximately equal to the size of the del
• Reconstructed slice spacing is typically 1 mm
Because of the limited range of acquisition angles
• Projection data do not form a complete set
• Reconstructed image is not a true 3D representation of the breast
• Possibility of artefacts in the images
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 51
9.7. BREAST CT

Dedicated breast CT systems have been developed using


cone-beam geometry and a flat-panel X- ray detector
• Data for all of the CT slices acquired simultaneously
• Rapid image acquisition
• Pixel dimensions substantially larger than for digital mammography
or tomosynthesis
• Large number of projections
• To keep doses at an acceptable level images are generally acquired at a higher
tube voltage (50 to 80 kV)
• Very low dose per projection can result in noisy images
Current designs provide a dedicated prone imaging table
Breast CT can be performed without the need to compress
the breast.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 52
9.8. COMPUTER-AIDED DIAGNOSIS

Computer-aided diagnosis or computer-aided detection


(CAD) systems are designed to assist the film reader in
detecting breast cancers
Computer system with sophisticated pattern recognition
software
• Natural adjunct to digital mammography
• Screen-film mammograms must be digitised (scanned)
Interpretive aid used during image review
Identifies “suspicious” features and alerts image reader
• Does not replace the image reader
CAD algorithms must be trained using sets of
mammograms for which the presence or absence of
cancers is known
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 53
9.8. COMPUTER-AIDED DIAGNOSIS

Results of CAD are conveyed to the film reader by means


of an image annotated to show the computer detections
• Different symbols for different lesions
Main use in screening mammography
• Double reading has been shown to increase the cancer detection
rate
• CAD has the potential to be a cost-effective alternative to double
reading
• CAD algorithm could be used to simulate the second film reader
• CAD has the potential to:
• Reduce the number of missed cancers
• Reduce the variability between film readers
• Improve the consistency and productivity of a single film reader

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 54
9.8. COMPUTER-AIDED DIAGNOSIS

Recent research in CAD considers the combination of


information from multiple images
• Different views from the same examination
• Same view from a previous examination
• This approach more closely mimics how a film reader reads a case
• May improve the performance of a CAD scheme
CAD may be particularly valuable in 3D breast imaging
• Large amount of image data to be considered
• CAD may be useful for automatic detection of microcalcifications
• Film reader can focus attention on more sophisticated
interpretation tasks

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 55
9.9. STEREOTACTIC BIOPSY SYSTEMS

Stereotactic procedures are used to investigate suspicious


mammographic or clinical findings without the need for
surgical (excisional) biopsies
• Reduced patient risk, discomfort and cost
In stereotactic biopsies, the gantry of a mammography unit
has the facility to allow a pair of angulated views of the
breast (typically at ± 15° from normal incidence) to be
obtained
From measurements obtained from these images, the
three-dimensional location of a suspicious lesion is
determined and a needle equipped with a spring-loaded
cutting device can be accurately placed in the breast to
obtain tissue samples
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 56
9.9. STEREOTACTIC BIOPSY SYSTEMS

The geometry for stereotactic breast biopsy is shown. The X ray tube is rotated about the
breast to produce two views. The Z-depth of an object can be determined by the lateral (X)
displacement observed between the two views.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 57
9.9. STEREOTACTIC BIOPSY SYSTEMS

These systems may use small-format (e.g. 5 cm x 5 cm)


digital detectors or full-field digital detectors
Stereotactic procedures can be performed on a
conventional mammography unit with a stereotactic
attachment with the patient standing
Dedicated stereotactic units where the patient lies prone
on a table with the breast pendant through an aperture in
the table top into the imaging region are also available

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 58
9.10. RADIATION DOSE

Three dosimetric quantities used in mammography


• Incident air kerma (IAK), Ki
• Entrance surface air kerma Ke
• Mean glandular dose (MGD, mean dose to the glandular tissue of
the breast), DG
MGD is the primary quantity of interest related to the risk
of radiation induced cancer in breast imaging
MGD is calculated using factors obtained experimentally or
by Monte Carlo radiation transport calculations
IAK (measured) converted to MGD for a breast of specific
composition and size
• Conversion coefficients are tabulated in various publications
(including IAEA Technical Report TRS-457)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 59
9.10. RADIATION DOSE

Dose in mammography depends on the size and


composition of the breast as well as the imaging device
and exposure settings selected
In screen-film mammography goal is to maintain a target
value of optical density
K increases as the thickness and/or density of the breast
e

increase resulting in a corresponding increase in MGD


Increase in beam energy (tube voltage, target/material
combination) will mitigate against some of dose increase
• Image contrast will be reduced and at some point this will become
unacceptable

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 60
9.10. RADIATION DOSE

In digital mammography goal is to maintain a target SDNR


at the detector
K increases as the thickness and/or density of the breast
e

increase resulting in a corresponding increase in MGD


Increase in beam energy (tube voltage, target/material
combination) will mitigate against some of dose increase
On a digital system where contrast can be adjusted during
image display, an acceptable compromise can be
achieved at a higher energy than with screen-film imaging
This allows the advantage of a greater relative decrease of
dose compared to film for large and/or dense breasts

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 61
9.10. RADIATION DOSE

There is a risk of cancer induction associated with the


radiation doses received in mammography.
BEIR VII report critically examined data on doses and
increased cancer incidence to develop a radiation risk
model for breast cancer
Model can be useful in predicting the lifetime risk following
a single mammographic examination or from multiple
exposures at different ages as would occur in periodic
screening
Risk for a woman at age 60 from a dose to the breasts of
0.004 Gy previously received from mammograms at age
45 is predicted to be 7.9 x 10-7

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 62
9.10. RADIATION DOSE

For a screening regimen that consists of annual


mammography examinations from ages 40 to 55 and
biennial examination thereafter until age 74 (i.e. 25
screenings) with a dose of 3.7 mGy to both breasts, it is
estimated that in 100,000 women 86 radiation-induced
cancers will be caused resulting in 11 deaths and a loss of
136 women-years of life.
In the same group, earlier detection through screening
would save 500 lives or 10,670 women-years, resulting in
a benefit-to-risk ratio of 47 (in lives) or 78 ( in women-
years). If the same diagnostic accuracy could be achieved
at reduced radiation dose, the benefit-to risk would
become even higher
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 63
BIBLIOGRAPHY

AMERICAN COLLEGE OF RADIOLOGY, Stereotactic


breast biopsy quality control manual, American College of
Radiology, Reston, VA (1999).
AMERICAN COLLEGE OF RADIOLOGY, Mammography
quality control manual, American College of Radiology,
Reston, VA (1999).
BICK, U., DIEKMANN, F., Digital mammography, Springer,
Heidelberg, Germany (2010).
EUROPEAN COMMISSION, European Guideline for
Quality Assurance in Mammography Screening, Office for
Official Publications of the European Communities Rep.
V4.0 Luxembourg (2006). http://www.euref.org.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 64
BIBLIOGRAPHY

PRESTON, D.L., et al., Radiation effects on breast cancer


risk: a pooled analysis of eight cohorts, Radiat Res 158 2
(2002) 220-35.
http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retriev
e&db=PubMed&dopt=Citation&list_uids=12105993.
FERLAY, J., et al., "Cancer Incidence and Mortality
Worldwide in 2008: IARC CancerBase No. 10",
GLOBOCAN 2008 (Proc. Conf. Lyon, France, 2008),
International Agency for Research on Cancer, World
Health Organization, http://globocan.iarc.fr.

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 65
BIBLIOGRAPHY

INTERNATIONAL ATOMIC ENERGY AGENCY, Quality


Assurance Programme for Screen-Film Mammography,
Human Health Series, 2, IAEA Vienna (2009).
http://wwwpub.iaea.org/MTCD/publications/PDF/Pub1381_
web.pdf.
INTERNATIONAL ATOMIC ENERGY AGENCY, Quality
Assurance Programme for Digital Mammography, Human
Health Series 17, IAEA Vienna (2011).
INTERNATIONAL COMMISSION ON RADIATION UNITS
AND MEASUREMENTS, Mammography - Assessment of
image quality, ICRU Rep. 82, Journal of the ICRU Vol. 9
No. 2 Bethesda, MD (2009).

IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 9, 66

Potrebbero piacerti anche