Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
3/2019
Pursuant to article 99, para 8 of the Law on medicines and medical devices
(Official Gazette of the RS, No. 105/17),
RULEBOOK
on vigilance
I INTRODUCTORY PROVISIONS
2. Definitions
Article 2
Terms used in the Law on medical devices (hereinafter: the Law) and
this Rulebook shall have the following meaning:
9) Field Safety Notice (FSN) is a notice for customers, i.e. users sent
from by the manufacturer, or an authorized representative of the
manufacturer in relation to the Field Safety Corrective Action;
10)damage is a physical injury or damage to the health of people,
animals or damage to property or the environment;
11)without delay is the method of urgent action, that is, the treatment
whose postponement cannot be justified;
(1) events that are of significant and unexpected nature such that they
become alarming as a potential public health hazard, e.g. human
immunodeficiency virus (HIV) or Creutzfeldt-Jacob Disease (CJD).
These concerns may be identified by either the Ministry, the Agency
or the manufacturer, or the authorized representative of the
manufacturer,
II VIGILANCE SYSTEM
Article 3
The manufacturer, i.e., the authorized representative of the manufacturer
investigates the incident and the Agency supervises the investigation and
undertakes measures necessary to amend or supplement the measures
taken by the manufacturer, i.e., the authorized representative of the
manufacturer. Depending on the results of the survey, the Agency
provides all the information needed to prevent incidents or to limit their
consequences after the placing of medical devices on the market and into
service (hereinafter: the vigilance system), in accordance with the Law.
Corrective actions include but are not limited to: medical device recall,
issuance of the Field Safety Notice, additional supervision, modification of
a medical device in use, modification of the medical device design or a
part of the manufacturing process, modification of labelling or user
instruction manual.
Article 4
Article 5
Article 6
(3) advice given to users relating to a change in the stated shelf life of
medical devices for in vitro fertilization and assisted reproduction
technology (for inst. the manufacturer informs the users of an
error on the labelling of their device which indicates a shelf life
longer than the validated shelf life for the product);
Article 7
The manufacturer, i.e., the authorized representative of the manufacturer
shall report to the Agency or Ministry each and every incident and Field
Safety Corrective Action if the reporting criteria foreseen by this Rulebook
are met.
Article 8
In case of medical devices which do not generally come into contact with
patients (most diagnostic devices, in vitro diagnostic and medical devices
for in vitro fertilization and assisted reproduction technology), if a direct
harm to the patient is difficult to prove (unless the device itself causes
deterioration in state of health) the harm is considered to be indirect –
the result of an act or omission based on incorrect results obtained
through the use of a diagnostic device, in vitro diagnostic medical device
or the consequence of the treatment of cells (e.g. gametes and embryos
in the case of medical devices for in vitro medical fertilization and
assisted reproduction technology) or organs outside of the human body
that will later be transferred to a patient. Indirect damage can also be
caused by a software considered to be a medical device. Incidents and
events leading to indirect harm must be reported to both the Agency and
the Ministry.
In case of the error made by either the user or a third person, the
labelling and the user instruction manual shall be closely examined for
inadequacy, especially in connection with devices intended for self-testing
where the medical decision can be taken solely by the patient.
Inadequacies in the information provided by the manufacturer that led or
could have led to harm to the user, patient or a third party shall be
reported in adherence with this Rulebook.
4. Agency’s Activities
article 9
The Agency shall post, on its webpage, all the information necessary to
prevent future incidents or limit the consequences thereof, depending on
the investigation conducted by the manufacturer, i.e., the authorized
representative of the manufacturer.
The Agency shall be responsible for raising awareness of the
organizations and individuals included in the health protection system in
charge of purchasing medical devices about the importance of their
cooperation as well as for ensuring the first step actions within the
vigilance system.
For the purpose of increasing the efficiency of the vigilance system, the
Agency shall encourage incident reporting by the users and professionals
included in the distribution, supply or putting medical devices into use.
This includes organizations and individuals responsible for ensuring calibration
and maintenance of the medical devices. Such reports are to be submitted to the
manufacturer, i.e., the authorized representative of the manufacturer as well as
to the Agency.
Article 10
Article 11
The manufacturer, i.e., the authorized representative of the
manufacturer shall submit the initial report on the incident to the
Agency, recording and evaluation purposes. Each and every initial
report shall lead to a final report unless they are combined into a single
report, however not all incident reports will lead to a corrective action.
As a general principle, there should be a pre-disposition to report rather than
not to report in case of doubt on the reportability of an incident..
Article 12
Any event which meets all of the basic reporting criteria listed below is
considered as an incident and must be reported to the Agency:
(2) for in vitro diagnostic medical devices, where there is a risk that an
erroneous result may cause indirect damage to the patient,
consequentially leading to life-threatening illness of the tested individual
or to the individual’s offspring or to death or severe disability to the
individual or fetus being tested, or to the individual’s offspring, all false
positive or false negative results shall be considered to be incidents. For
all other IVDs, false positive or false negative results falling outside the
declared performance of the test shall be considered as events.
(1) the
opinion of the healthcare professionals based on available
evidence,
This judgement may be difficult when there are multiple devices and
drugs involved in the event that has led on the incident. In that case,
it should be assumed that the device may have caused or contributed
to the incident and the manufacturer should err on the side of
caution;
3) the event led or might have led to one of the following outcomes:
Article 13
Article 14
Incidents specified in the Field Safety Notice that occur after the
manufacturer, i.e., the authorized representative of the manufacturer has
issued the Field Safety Notice and conducted the Field Safety Corrective
Action are not to be reported individually. The manufacturer, i.e., the
authorized representative of the manufacturer and the Agency shall
agree upon the frequency and content of the Periodic Safety Report (for
inst. the manufacturer issued a Periodic Safety Report and implemented
a Field Safety Corrective Action for a coronary stent that migrated due
inadequate inflation associated with the balloon mechanism. Subsequent
examples of stent migration were summarized in quarterly reports
concerning the Field Safety Corrective Action and individual incidents did
not have to be reported).
Article 15
Article 16
Medical device deficiencies that are always detected (that could not go
undetected) by the user prior to use shall not be reported under the
vigilance system, regardless of the provisions printed in the
manufacturer’s instruction manual (for inst., the packaging of a sterile
single use device is labelled with the caution 'do not use if the packaging
is opened or damaged'. Prior to use, obvious damage to the packaging
was observed, and the device was not used; intravenous
administration set tip protector has fallen off the set during distribution
resulting in a non-sterile fluid pathway. The intravenous administration
set was not used; a vaginal speculum has multiple fractures. Upon
activating the handle, the device fell apart. The device was not used; in
an in vitro diagnostic test kit, a bottle labelled lyophilized is found to be
fluid, this is discovered by the USER prior to use).
In cases referred to in para 1 of this Article, the user shall notify the
manufacturer, i.e., the authorized representative of the manufacturer
accordingly prior to the use of the medical device.
Article 17
Article 18
When the only cause for the event was that the device exceeded its
service life or shelf-life as specified by the manufacturer and the failure
mode is not unusual, the incident does not need to be reported (e.g.,
loss of sensing after a pacemaker has reached end of life. Elective
replacement indicator has shown up in due time according to device
specification. Surgical explantation of pacemaker required; insufficient
contact of the defibrillator pads to the patient was observed. The patient
could not be defibrillated due to insufficient contact to the chest. The
shelf life of the pads was labelled but exceeded; a patient is admitted to
hospital with hypoglycemia based on an incorrect insulin dosage following
a blood glucose result. The investigation found that the test strip was
used beyond the expiry date specified by the manufacturer).
Article 19
Events which did not lead to a serious deterioration in state of health or
death, because a design feature protected against a fault becoming a
hazard (in accordance with relevant standards or documented design
inputs) shall not be subject to reporting. As a precondition, there must be
no danger for the patient to justify not reporting. If an alarm system is
used, the concept of this system should be generally acknowledged for
that type of product (e.g. an infusion pump stops, due to a malfunction,
but gives an appropriate alarm in compliance with relevant standards and
there was no injury to the patient; microprocessor-controlled radiant
warmers malfunction and provide an audible appropriate alarm in
compliance with relevant standards and there was no deterioration in
state of health of the patient; during radiation treatment, the automatic
exposure control is engaged. Treatment stops. Although patient receives
less than optimal dose, patient is not exposed to excess radiation; a
laboratory analyzer stops during analysis due to a malfunction of the sample
pipetting module, but the appropriate error message was provided for the user.
An intervention by the user or an immediate remote intervention by
manufacturer allowed the analyzer to resume the analysis, resulting in correct
results).
Expected and foreseeable adverse effects
Article 20
Expected and foreseeable adverse effects are those that are well-known
in medical, scientific or technological sense and are clearly identified
during clinical trials or clinical practice and listed on the label or in the
manufacturer’s user manual.
Article 21
The incident for which the risk of a death or serious deterioration in state
of health is quantified and found to be negligibly low shall not be subject
to reporting unless it resulted in death or serious deterioration in state of
health, provided the risk has been described and documented as
acceptable within a comprehensive risk assessment (for inst., the
pacemaker manufacturer placed on the market identifies a software error
and quantifies the probability of serious deterioration of health within the
set environment as negligible and no adverse effects upon the patient’s
health were recorded).
Article 22
Article 23
The manufacturer, i.e., the authorized representative of the
manufacturer shall evaluate all events, potential errors in use and
abnormal uses as well as other claims pertaining to the medical device.
Such an evaluation shall be governed by risk management, engineering,
usability, design validation and corrective and preventive action
processes.
Article 24
Abnormal use
Article 26
Article 27
Data on incident stated in each and every report the manufacturer, i.e.,
the authorized representative of the manufacturer submits to the Agency
and used for the initial, follow-up and the final report shall be submitted
as a report by manufacturer, i.e., the authorized representative of the
manufacturer on the incident given as a template in Annex 4 attached to
this Rulebook as an integral part thereof.
If the initial report is submitted orally (over the phone, for inst.), the
manufacturer, i.e., the authorized representative of the manufacturer
shall, without delay file a written report as well.
The initial, follow-up and final reports shall be filed using a template given
as Annex 3 to this Rulebook.
Article 28
As soon as it becomes aware of an event and of the fact that the medical
device caused or contributed to the event, the manufacturer, i.e., the
authorized representative of the manufacturer shall proceed with
determining whether or not it was an incident.
The manufacturer, i.e., the authorized representative of the manufacturer
shall report to the Agency each and every serious public health threat
without delay, however within two calendar days at the latest from the
day of becoming aware of the threat.
Article 30
Article 31
The manufacturer, i.e., the authorized representative of the manufacturer
may consult with the user on a particular, reported incident before a
report is made to the Agency.
Article 32
The Agency may take any further action it deems appropriate, consulting
with the manufacturer, i.e., the authorized representative of the
manufacturer.
Follow-Up Report
Article 33
Article 34
If the Agency performs the investigation then the manufacturer, i.e., the
authorized representative of the manufacturer shall be informed of the
result.
In assessing the need for the Field Safety Corrective Measure , the
manufacturer, i.e., the authorized representative of the manufacturer
shall use the methodology described in the harmonized Risk
Management standard EN ISO 14971.
Article 36
1) Exact name of the medical device, along with the group of generic
medical devices it belongs to;
8) a request to pass the Field Safety Notice to all those who need to be
aware of it within the organization and to maintain awareness over
an appropriate defined period;
9) a request for the details of any affected devices that have been
transferred to other organizations and a copy of the Field Safety
Notice to be passed on to the organization to which the device has
been transferred;
10) affected medical devices and serial / lot / batch number range
Article 37
(4) timelines;
5) a request to pass the Field Safety Notice to all those who need to be
aware of it within the organization and to maintain awareness over
an appropriate defined period;
8) Confirmation that the Agency has been advised of the Field Safety
Corrective Action;
An acknowledgment form for the receiver of the Field Safety Notice might
also be included especially useful for Manufacturer’s control purposes.
Article 38
Article 39
A report which appears to meet the criteria of article 12 of this Rulebook,
received by the Agency from a user reporting system or other source,
shall be copied by the Agency to the manufacturer, i.e., the authorized
representative of the manufacturer without delay or translation. In
doing so, patient confidentiality should be maintained pursuant to the
Law.
In the case referred to in para 1 of this article, the manufacturer, i.e., the
authorized representative of the manufacturer shall act in line with article
11-37 of this Rulebook.
Article 40
1) Acceptability of the risk, taking into account criteria such as: causality,
technical/other cause, probability of occurrence of the problem,
frequency of use, detectability, probability of occurrence of harm,
severity of harm, intended purpose and benefit of the product,
requirements of harmonized European, i.e., Serbian standards, the
Medical Device safety principles foreseen by the law and by-law
governing the basic requirements pertaining to medical devices,
potential users, affected populations, etc.;
Article 41
The Agency normally monitors the investigation being carried out by the
manufacturer, i.e., the authorized representative of the manufacturer
and it may intervene at any time in consultation with the manufacturer
where practicable.
2) users;
Article 42
For drug device combination products regulated under the medical device
regulations, pursuant to the Law, the Agency receiving the incident report
should establish a link with any other relevant bodies in the country and
abroad, if required .
1) no further action;
When the manufacturer has informed the Agency in advance of the start
of an FSCA this information should be held confidential by the Agency
until the information becomes public.
Article 44
The Agency shall place the final report by the manufacturer or the
authorized representative of the manufacturer on file and make any
other observations necessary. The files investigation may then be
endorsed as "complete”.
Article 45
Even though the Body in charge of conformity assessments does not play
a key operational role in the Medical Device Vigilance System, the overall
performance of the Medical Device Vigilance System is supported by the
Body’s activity in the following areas:
Article 47
If a serious adverse event occurs in the course of a clinical trial, the
sponsor, i.e., its authorized representative (hereinafter: sponsor) is
obliged to immediately notify the Agency of:
Article 48
The sponsor shall report to the Agency any serious adverse event that
may lead to immediate danger of death, serious injury or illness, thus
requiring urgent corrective actions for the benefit of other respondents,
users or third persons resulting in a new finding in connection to such a
serious adverse event, within two calendar days from the day the
sponsor becomes aware of the event or the finding that is subject to
reporting.
The sponsor shall report to the Agency any other events referred to in
article 47 para 1 of this Rulebook within seven calendar days from the day
the sponsor becomes aware of the event or the finding that is subject to
reporting.
Article 49
Article
50
On the day this Rulebook enters into force the Rulebook on the method
of reporting, gathering and monitoring adverse reactions to medical
devices (Official Gazette of the RS No. 64/11 and 21/14) shall cease to
apply.
Article
51
This Rulebook shall enter int force on the day it is published in the
“Official Gazette of the Republic of Serbia”.
No. 110-00-450/2018-06
In Belgrade, 14 January 2019
Minister,
Dr. Zlatibor Lončar, in person