Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Clinical Biomechanics
journal homepage: www.elsevier.com/locate/clinbiomech
a
School of Biological Science and Medical Engineering, Beihang University, Beijing 100083, China
b
Naton Institute of Medical Technology, Beijing 100095, China
c
Beijing Engineering Laboratory of Functional Medical Materials and Devices, Beijing Naton Technology Group Co. Ltd., Beijing 100082, China
d
Beijing Medical Implant Engineering Research Center, Beijing Naton Technology Group Co. Ltd., Beijing 100082, China
e
Beijing Advanced Innovation Center for Biomedical Engineering, Beihang University, Beijing 100191, China
Keywords: Background: The wear properties of knee implants need to be thoroughly evaluated prior to clinical use to ensure
ISO 14243-1 implant longevity. ISO 14243-1:2009 and ASTM F3141-17 are the two standards typically used for evaluating
ASTM F3141 wear, with the ISO standard being more common; ASTM F3141-17 was first released in 2015. The aim of this
Tibial wear study is to compare differences between these two standards in terms of wearing on a knee prosthesis.
Finite element analysis
Methods: Using finite element analysis based on Archard's law, this study evaluated anterior-posterior and in-
Knee kinematics
ternal-external motion, contact area, contact force, contact stress, volumetric wear rate, wear depth, and wear
distribution on the knee prosthesis.
Findings: The results show that simulations performed according to ASTM F3141 produced knee kinematics that
were more similar to human gait. The maximum wear depth occurred on the medial side of the tibia. However,
the region of peak contact stress did not always correspond with the region of the maximum wear depth, in-
dicating that considering the contact stress alone is not sufficient for evaluating wear as the sliding distance also
plays an important role. The resulting wear region from the ASTM F3141 simulation was smaller but deeper than
the wear region from the simulation per ISO 14243-1. However, the volumetric wear rates were very similar,
with 13.48–55.26 mm3/million for ASTM F3141 and 13.64–54.9 mm3/million for ISO 14243-1.
Interpretation: The resulting rate of wear is almost identical between ISO 14243-1 and ASTM F3141. However,
there are differences in wear contours and wear depth.
1. Introduction 2018; Endolab, 2018; FDA, 2003; ISO, 2009). ISO 14243-1 is also re-
commended by FDA guidance documents for evaluating knee pros-
Total knee arthroplasty (TKA) is a common surgical treatment for theses, with the requirement that if the thickness of the tibial bearing
degenerated knees to alleviate patient pain and increase knee function. (polyethelene) is below 6 mm, a wear test should be performed to de-
Polyethylene wear and wear-related complications, such as implant monstrate that the tibial bearing can survive 10 million cycles under
loosening, are still a leading cause of revision (Harman et al., 2010; normal physiological loading (walking) (FDA, 2003; ISO, 2009). An-
Liang et al., 2018; Pinol et al., 2014; Sharkey et al., 2014). Implant other option is to comply with ASTM F3141-17 which has similar input
retrieval studies are the most effective way of directly analyzing wear rules for knee wear simulators under walking conditions.
on the knee joint, but such studies require samples from suitable pa- A reliable standard for wear performance is crucial for the devel-
tients with a detailed assessment of individual physiological and pa- opment of knee prostheses as it predicts the lifetime of the tibial
thological conditions. In addition, it is often time consuming and costly bearing. There are two control modes for wear testing during simulated
(Gebert de Uhlenbrock et al., 2012; Huang et al., 1999; Huang et al., gait: displacement control according to ISO 14243-3, and load control
2002; Lu et al., 2010). according to ISO 14243-1 and ASTM F3141 (ASTM, 2017; ISO, 2002;
Wear simulators are often used to evaluate the wear properties of ISO, 2009; ISO, 2014). The inputs for displacement control are anterior-
knee implants prior to clinical use, with ISO 14243-1:2009 being the posterior (AP) displacement and internal-external (IE) angle, and the
most commonly used standard governing the test methods (Accutek, inputs for load control are AP load and tibial rotation (TR) torque. This
⁎
Corresponding author at: School of Biological Science and Medical Engineering, Beihang University, No.37, XueYuan Road, HaiDian District, Beijing, China.
E-mail address: ckcheng2009@gmail.com (C.-K. Cheng).
https://doi.org/10.1016/j.clinbiomech.2019.02.008
Received 10 September 2018; Accepted 13 February 2019
0268-0033/ © 2019 Elsevier Ltd. All rights reserved.
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
study focuses on load control according to ISO 14243-1 and ASTM of wearing of knee prostheses. The aim of this study is to quantitatively
F3141, as there is only a load control mode defined in the ASTM compare differences between ISO 14243-1 and ASTM F3141 in terms of
standard. Given its longer history, ISO 14243-1 is more common and wearing of knee prostheses. The hyposthesis is that there are differences
established in industry than ASTM F3141. ASTM F3141 was first re- in wear rates because of different input conditions, in particular the AP
leased in 2015 and then revised in 2017. ISO 14243-1 is still re- load as shown in Fig. 1.
commended for use by FDA guidance documents as well as a number of
multinational research organizations (Accutek, 2018; Endolab, 2018;
2. Methods
FDA, 2003). ASTM F3141 is based on studies by Bergmann et al.
(Bergmann et al., 2014; Bergmann and Graichen, 2014), while ISO
Finite element analysis (FEA) models of the femoral component and
14243-1 is based on decades of data (Bergmann et al., 2014). However,
tibial insert were developed in Abaqus® 2017 (Abaqus, SIMULIA, East
the differences between ISO 14243-1 and ASTM F3141 are still unclear
Providence, USA) using 3D scans of a PFC implant (Size two, PFC,
in terms of wearing of knee prostheses. Comparing the current revision
Depuy Synthes, Salt Lake, Utah, USA) (Figs. 2 & 3). Geometrical dif-
of ISO 14243-1:2009 with the previous 2002 revision, the definition of
ferences in the articular surface between the model and physical im-
soft tissue constraints has changed, but the definition of input curves
plant were controlled to be < 0.1 mm. The tibial insert was modeled as
remains the same (ISO, 2002; ISO, 2009). The definition of input curves
a deformable polyethylene (Gur 1050) material. The contact region of
in ASTM F3141 has not changed from its initial publication through to
the insert was meshed using hexahedron 8-noded elements C3D8R, and
its revision in 2017 (ASTM, 2017).
tetrahedron 10-noded elements C3D10M were used for all other regions
There are four inputs to the knee simulator required by ISO 14243-
(Fig. 3). Based on convergence testing (< 2% difference) of maximum
1: 2009 and ASTM F3141-17; flexion, axial load, internal-external (IE)
contact stress, an element size of around 1 mm was chosen for the
torque and anterior-posterior (AP) load. The flexion, axial load and IE
contact region and 1.6 mm was chosen for all other regions. The fe-
torque are similar between the two standards, but the definition of AP
moral component was modeled as a rigid body to save computational
load is different. For illustration, variations in these four inputs against
time. The coefficient of friction between the tibia and femur was 0.04
gait cycles are plotted in Fig. 1 for both standards.
(Grupp et al., 2009).
In a healthy knee joint, when the knee flexes between 5° and 60°, the
The boundary conditions were set according to the ISO and ASTM
tibia moves anteriorly. This is due to the fact that knee flexion and
standards (ISO, 2009; ASTM, 2017). The axial loading axis was offset to
femoral rollback occur simultaneously (Bergmann et al., 2014;
the medial side by a distance 0.07 times the width of tibia. Three of the
Dimitriou et al., 2016; Iwaki et al., 2000; Reinschmidt et al., 1997). As
loading curves (axial load, AP load and IE torque, Fig. 1B, C & D) were
seen in Fig. 1A, knee flexion noticeably increases in the range of 40% to
applied to the tibia and the flexion angle curve (Fig. 1A) was applied to
70% of the gait cycle. ASTM F3141 correctly gives the AP load for the
the femur. The tibia insert was constrained in flexion but free to move
tibia in the anterior direction while ISO 14243-1 states a posterior di-
in all other directions; medial-lateral, AP, superior-inferior, valgus
rection.
varus and rotation. The femoral component was only permitted to move
To the best of our knowledge, no studies have been conducted so far
in flexion during the gait simulation. The flexion axis was defined as a
to evaluate differences between ISO 14243-1 and ASTM F3141 in terms
line connecting the lateral femoral center and medial femoral center
Fig. 1. Four inputs from ISO 14243-1: 2009 and ASTM F3141:2017. (A) Flexion angle; (B) axial load; (C) internal-external (IE) torque; (D) anterior-posterior (AP)
load.
35
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
Fig. 2. Geometrical differences on the articular surface between the model and physical implant were controlled to be < 0.1 mm.
Fig. 3. FEA model; (A) AP and IE constraints, (B) wear region of tibial insert.
36
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
Fig. 4. Curves for Load-Displacement and Rotation-Torque in the AP and IE directions, respectively.
Results are presented for AP translation, IE rotation, contact area, 3.3. Volumetric wear rate, depth and distribution
contact force, contact stress, volumetric wear rate, wear depth and wear
distribution. The predicted wear rate, volumetric wear and wear depth are listed
in Table 2. For both K values, the wear rate and volumetric wear were
3.1. AP and IE kinematics similar for ASTM F3141 and ISO 14243-1. However there was a great
difference in the calculated wear depth between the two standards,
Results for AP translation and IE rotation according to ISO 14243-1 possibly due to the different knee kinematics and contact stresses. Also,
and ASTM F3141 are shown in Fig. 5. As detailed previously in Fig. 1A, increasing the K value from 2.64 ∗ 10−7 to 10.656 ∗ 10−7 resulted in an
there was an obvious increase in knee flexion between 40% and 70% of almost four-fold increase in volumetric wear rate and wear depth.
the gait cycle. Referring to the ASTM plots in Fig. 5, between 40% and Fig. 9 shows the wear depth on the tibial insert for both standards.
70% of the gait cycle, there was a corresponding increase in AP The wear region is smaller but deeper for the ASTM standard, while the
translation (Fig. 5A), signifying femoral rollback, and also a noticeable region is more dispersed for the ISO model. The maximum depth is
increase in internal rotation of the tibia (Fig. 5B). However, for the ISO present on the medial side of the insert for both models.
37
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
Fig. 6. Contact areas and contact forces from ASTM and ISO standards. (A) Contact areas; (B) contact forces.
38
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
Fig. 8. The respective locations of the peaks of maximum contact stress: A, B, C, and D.
Table 2
Predicted wear rate, volumetric wear and wear depth for ASTM and ISO standards.
ASTM F3141 ISO 14243-1
−7 3 3
K = 2.64 ∗ 10 mm /Nm Wear rate (mm /million) 13.48 13.64
Volumetric wear (mm3) 67.4 68.2
Wear depth (mm) 1.137 0.538
K = 10.656 ∗ 10–7 mm3/Nm Wear rate (mm3/million) 55.26 54.9
Volumetric wear (mm3) 276.3 274.5
Wear depth (mm) 4.606 2.171
human gait. For both ISO 14243-1 and ASTM F3141, the trends of maximum wear depth.
contact area, total contact forces, and contact stress curves were similar
to the axial load inputs. The location of maximum depth occurred on Declarations of interest
the medial side of the tibia for both standards, and the region of peak
contact stress did not always correspond with the region of the None.
39
X.-H. Wang, et al. Clinical Biomechanics 63 (2019) 34–40
CRediT authorship contribution statement Influence of time in-situ and implant type on fixation strength of cemented tibial
trays - a post mortem retrieval analysis. Clin. Biomech. 27 (9), 929–935.
Grupp, T.M., Kaddick, C., Schwiesau, J., Maas, A., Stulberg, S.D., 2009. Fixed and mobile
Xiao-Hong Wang: Conceptualization, Data curation, Formal bearing total knee arthroplasty—influence on wear generation, corresponding wear
analysis, Investigation, Methodology, Project administration, areas, knee kinematics and particle composition. Clin. Biomech. 24 (2), 210–217.
Software, Validation, Supervision, Visualization, Writing - original Harman, M.K., Schmitt, S., Rössing, S., Banks, S.A., Sharf, H.P., Viceconti, M., Hodge,
W.A., 2010. Polyethylene damage and deformation on fixed-bearing, non-conforming
draft, Writing - review & editing. Hui Li: Methodology, Software, unicondylar knee replacements corresponding to progressive changes in alignment
Validation, Writing - review & editing. Xiang Dong: Methodology, and fixation. Clin. Biomech. 25 (6), 570–575.
Software, Validation, Writing - review & editing. Feng Zhao: Huang, C.H., Yang, C.Y., Cheng, C.K., 1999. Fracture of the femoral component associated
with polyethylene wear and osteolysis after total knee arthroplasty. J. Arthroplast. 14
Methodology, Software, Validation, Writing - review & editing. (3), 375–379.
Cheng-Kung Cheng: Conceptualization, Formal analysis, Huang, C.H., Ho, F.Y., Ma, H.M., Yang, C.T., Liau, J.J., Kao, H.C., Young, T.H., Cheng,
Investigation, Methodology, Project administration, Software, C.K., 2002. Particle size and morphology of UHMWPE wear debris in failed total knee
arthroplasties—a comparison between mobile bearing and fixed bearing knees. J.
Validation, Supervision, Visualization, Writing - original draft,
Orthop. Res. 20 (5), 1038–1041.
Writing - review & editing, Funding acquisition, Resources. ISO 14243-1, 2002. Implants for Surgery. Wear of Total Knee-joint Prostheses. Part 1:
Loading and Displacement Parameters for Wear-testing Machines With Load Control
Acknowledgement and Corresponding Environmental Conditions for Test.
ISO 14243-1, 2009. Implants for Surgery. Wear of Total Knee-joint Prostheses— Part 1:
Loading and Displacement Parameters for Wear-testing Machines With Load Control
This study was supported by a Grant from the Chinese National and Corresponding Environmental Conditions for Test.
Science and Technology Program (National Key Research and ISO 14243-3, 2014. Implants for Surgery — Wear of Total Knee-joint Prostheses — Part 3:
Loading and Displacement Parameters for Wear-testing Machines With Displacement
Development Plan) (Grant number: 210YBXM2016110002), and a Control and Corresponding Environmental Conditions for Test.
grant from National Natural Science Foundation of China (Grant Iwaki, H., Pinskerova, V., Freeman, M.A., 2000. Tibiofemoral movement 1: the shapes
number: 81572180). The authors would like to thank Mr. Colin and relative movements of the femur and tibia in the unloaded cadaver knee. J. Bone
Joint Surg. (Br.) 82 (8), 1189–1195.
McClean for his assistance with proofreading this manuscript, and Prof. Knight, L.A., Pal, S., Coleman, J.C., Bronson, F., Haider, H., Levine, D.L., Taylor, M.,
Chuan Li at the National Yang Ming University for his support and Rullkoetter, P.J., 2007. Comparison of long-term numerical and experimental total
advice through numerous communications. knee replacement wear during simulated gait loading. J. Biomech. 40 (7), 140–146.
Liang, H., Bae, J.K., Park, C.H., Kim, K.I., Bae, D.K., Song, S.J., 2018. Comparison of mode
of failure between primary and revision total knee arthroplasties. Orthop. Traumatol.
Appendix A. Supplementary data Surg. Res. 104 (2), 171–176.
Liau, J.J., Cheng, C.K., Huang, C.H., Lo, W.H., 2002. The effect of malalignment on
stresses in polyethylene component of total knee prostheses—a finite element ana-
Supplementary data to this article can be found online at https://
lysis. Clin. Biomech. 17 (2), 140–146.
doi.org/10.1016/j.clinbiomech.2019.02.008. Lu, Y.C., Huang, C.H., Chang, T.K., Ho, F.Y., Cheng, C.K., Huang, C.H., 2010. Wear-
pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total
References knee prostheses. J. Bone Joint Surg. (Br.) 92 (4), 500–507.
McEwen, H.M., Fisher, J., Goldsmith, A.A., Auger, D.D., Hardaker, C., Stone, M.H., 2001.
Wear of fixed bearing and rotating platform mobile bearing knees subjected to high
Accutek testing laboratory, 2018. https://www.element.com/more-sectors/medical- levels of internal and external tibial rotation. J. Mater. Sci. Mater. Med. 12 (10–12),
device/wear-testing-of-medical-devices. 1049–1052.
Archard, J.F., 1953. Contact and rubbing of flat surfaces. J. Appl. Phys. 24 (8), 981–988. Morra, E.A., Postak, P.D., 1999. The influence of mobile bearing knee geometry on the
ASTM F3141, 2017. Standard Guide for Total Knee Replacement Loading Profiles. wear of UHMWPE tibial inserts II: a finite element study. In: Annual meeting of the
Barnett, P.I., Fisher, J., Auger, D.D., Stone, M.H., Ingham, E., 2001 Oct–Dec. Comparison American Academy of Orthopedic Surgeons (AAOS). Orthopedic Research
of wear in a total knee replacement under different kinematic conditions. J. Mater. Laboratories. http://orl-inc.com.
Sci. Mater. Med. 12 (10−12), 1039–1042. Pal, S., Haider, H., Laz, P.J., Knight, L.A., Rullkoetter, P.J., 2008. Probabilistic compu-
Barnett, P.I., McEwen, H.M., Auger, D.D., Stone, M.H., Ingham, E., Fisher, J., 2002. tational modeling of total knee replacement wear. Wear 264, 701–707.
Investigation of wear of knee prostheses in a new displacement/force-controlled si- Pinol, I., Torres, A., Gil, G., Prats, E., Puig-Verdier, L., Hinarejos, P., 2014. Polyethylene
mulator. Proc. Inst. Mech. Eng. H 216 (1), 51–61. particles in joint fluid and osteolysis in revision total knee arthroplasty. Knee. 21 (2),
Bergmann, G., Graichen, F., 2014. Loading of orthopeadic implants. In: Ortho Load, cited 402–405.
2014. http://www.orthoload.com/. Reinschmidt, C., Bogert, A.J.V.D., Lundberg, A., Nigg, B.M., Murphy, N., 1997.
Bergmann, G., Bender, A., Graichen, F., Dymke, J., Rohlmann, A., Trepczynski, A., Heller, Tibiofemoral and tibiocalcaneal motion during walking: external vs skeletalmarkers.
M.O., Kutzner, I., 2014. Standardized loads acting in knee implants. PLoS One 9 (1), Gait Posture 6 (2), 98–109.
e86035. Sharkey, P.F., Lichstein, P.M., Shen, C., Tokarski, A.T., Parvizi, J., 2014. Why are total
Dimitriou, D., Tsai, T.Y., Park, K.K., Hosseini, A., Kwon, Y.M., Rubash, H.E., Li, G., 2016. knee arthroplasties failing today—has anything changed after 10 years? J.
Weight-bearing condyle motion of the knee before and after cruciate-retaining TKA: Arthroplast. 29 (9), 1774–1778.
in-vivo surgical transepicondylar axis and geometric center axis analyses. J. Biomech. Simpson, D.J., Gray, H., D'Lima, D., Murray, D.W., Gill, H.S., 2008. The effect of bearing
49 (9), 1891–1898. congruency, thickness and alignment on the stresses in unicompartmental knee re-
Ding, W.Y., Ma, S.Q., Zhou, X.C., Yang, Y.J., Sun, Y., Zhao, F., Fan, Y.B., 2018. Wear placements. Clin. Biomech. 23 (9), 1148–1157.
simulation of tibiofemoral joint surface of total knee prosthesis with high conformity. Viceconti, M., Olsen, S., Burton, K., 2005. Extracting clinically relevant data from finite
J. Med. Biomechanics 33 (3), 7–13. element simulations. Clin. Biomech. 20, 451–454.
Endolab implant testing laboratory, 2018. https://www.endolab.org/simulator-knee- Wang, C., Zhao, F., Ding, W.Y., Sun, Y., Fan, Y.B., 2017. Finite element study on total
implants.asp. knee prosthesis wear during stair ascent. J. Med. Biomechanics 32 (2), 109–114.
FDA, 2003. Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial Willing, R., Kim, Y., 2009. Three dimensional shape optimization of total knee replace-
and Femorotibial Metal/Polymer Porous-coated Uncemented Prostheses; Guidance ments for reduced wear. Struct. Multidiscip. Optim. 38, 405–414.
for Industry and FDA.
Gebert de Uhlenbrock, A., Püschel, V., Püschel, K., Morlock, M.M., Bishop, N.E., 2012.
40