Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
S E P T E M B E R 2 010 (2 of 3)
Basic Principles of
Phacoemulsification
and Fluid Dynamics
Uday Devgan, MD, FACS, FRCS (Glasg)
D. Michael Colvard, MD, FACS; Thomas L. Beardsley, MD, Editors for Cataract Surgery Barry S. Seibel, MD
James C. Bobrow, MD, Basic and Clinical Science, Course Faculty, Section 11 Donald N. Serafano, MD
Edward K. Isbey III, MD, Practicing Ophthalmologists Advisory Committee for Education
Platform tion rate of the f luid from the eye. There may be a linear
control of vacuum and f low, so that the top of position 2
All phacoemulsification (phaco) platforms share the provides less vacuum or f low than the middle or bottom
same basic structure and concepts. The phaco machine range of the pedal. This is similar to the gas pedal in a
aims to balance f luidics within the eye, while delivering car, where the car’s throttle opens more as the gas pedal
ultrasonic energy and vacuum in order to emulsify and is further depressed. The vacuum and aspiration levels
aspirate the cataract through a small incision. created during surgery draw the f luid out of the eye and
The main functions of the phaco machine are to pro- into a waste f luid collection via the outf low tubing.
vide irrigation into the eye; to create vacuum/aspiration
to remove the cataract; and to deliver ultrasound energy Foot Position 3: Ultrasound Energy
in order to emulsify the nucleus. These functions corre-
Position 3 controls the delivery of ultrasound energy.
spond to the machine’s 3 foot pedal positions (Figure 1).
There is linear control of the ultrasound energy level so
that further pedal depression results in more ultrasound
Foot Position 1: Irrigation
energy. Note that if the pedal is in position 3, the sur-
The irrigation function of the phacoemulsification geon is already engaging the full function of both posi-
machine provides a source of f luid infusion into the eye tions 1 and 2. The irrigation is on, and the vacuum and
during the surgery. Depressing the foot pedal to posi- aspiration level is at its highest preset level. Ultrasound
tion 1 turns on the infusion. There is no linear control energy should only be applied once the tip of the phaco
of the infusion—the infusion is either turned on or probe is in contact with part of the cataract.
turned off. The height of the infusion bottle determines
the relative infusion pressure and f low rate during the
surgery. To keep the eye inf lated during surgery, the
surgeon needs to make sure that the f luid inf low rate Basic Concepts
is at least as great as the f luid outf low rate, because
any excess inf low will pressurize the eye and leak from of Fluidics
Due to the small volume of the anterior and posterior
chambers, the control of f luidics during phacoemulsifi-
cation surgery is critical. By keeping a constant infusion
pressure and by carefully controlling the outf low, the
surgeon can ensure that the eye stays inf lated and stable
during surgery. If the surgeon allows outf low to exceed
inf low, even for a fraction of a second, the chamber is at
risk of collapse.
Modulating Phacoemulsification
Fluid Flow: Poiseuille’s Equation
The basic equation that governs all f luid f low during
phacoemulsification surgery is Poiseuille’s equation:
Figure 1 Position of the phaco system foot pedal controls
the 3 basic functions of the device. Position 1 delivers irriga- F = ΔP π r 4/8 η L
tion alone, position 2 provides irrigation plus aspiration, and
In this equation, F = f low, ΔP = pressure gradient,
position 3 adds phaco energy to irrigation and aspiration.
r = radius of the tube, η = viscosity of fluid, and L = length
Peristaltic and Venturi rate is variable and is determined by the vacuum level
created; the surgeon cannot directly set it.
Fluid Pumps
Two types of vacuum pumps are used in phacoemulsifi-
cation platforms: peristaltic and venturi. Although they
work in different ways, each has its advantages.
Fluidic Settings peristaltic f low rate determines the speed at which the
fragments are attracted to the phaco tip, with a f low rate
The function of the phacoemulsification f luidics is to of 20 cc/min being very slow and 50 cc/min being very
balance the inf low and outf low of f luid in order to main- fast. During phaco chop, the surgeon can use the same
tain the working space, bring cataract material to the vacuum and f low rate settings for the entire nucleus
phaco tip, and prevent collapse of the eye. Optimizing removal procedure.
the phaco f luidic settings is crucial to maximizing effi- For divide-and-conquer phaco, there are 2 distinct
ciency and safety of phacoemulsification surgery. stages of nucleus removal: sculpting of the nucleus and
In order to optimize the phaco f luidic settings, it is quadrant removal. Different f luidic settings are required
important to match the machine parameters to the sur- for each. For grooving and sculpting of the nucleus, the
geon’s phaco technique. With a typical peristaltic phaco work is being done by the ultrasonic energy and thus the
machine platform, the most common type in the United f low and vacuum settings are quite low—just enough to
States, only a few parameters are adjustable. These aspirate the nuclear material removed from each for-
include needle size, bottle height, f low rate, and maxi- ward stroke of the phaco probe. A vacuum level of less
mum vacuum level. than 100 mmHg and a f low rate of less than 30 cc/min
The selection of phaco needle size is of great impor- are sufficient for this purpose. For quadrant removal, a
tance. The most common sizes are the smaller-bore moderate amount of holding power is required to bring
0.9‑mm needle and the larger-bore 1.1-mm needle. If the each quadrant into the phaco tip. Using a higher vac-
surgeon’s preference is a quicker procedure with rapid uum level of 200 to 300 mmHg and a f low rate of 30 to
nucleus removal, then the larger 1.1‑mm needle size 50 cc/min, depending on the needle size, is typically suf-
is desirable because it will give a significantly greater ficient for this purpose.
f low rate. If the surgeon prefers a slower but more con- With knowledge of the concepts behind the variables,
trolled procedure, then the smaller-bore 0.9-mm needle it is easy to tailor the f luidic settings to the surgeon
is more suited to that technique. The holding power of and technique. Understanding the concepts behind the
the phaco probe on to the cataract nucleus is related phaco f luidic settings is instrumental in optimizing the
to the vacuum level used as well as the surface area of parameters for increasing the efficiency and safety of the
occlusion. Using a larger phaco needle allows for lower surgeon’s phaco technique.
Figure 9 If the duty cycle is held constant, simply increas- Figure 10 The total phaco energy delivered can be reduced
ing the pulses per second (PPS) does not decrease the total significantly by altering the “on-off” duty cycle.
phaco energy delivered.
Clinicians’ Corner provides additional viewpoints on 1. Discuss the relative advantages and disadvantages
of venturi-based and peristaltic phacoemulsifica-
the subject covered in this issue of Focal Points. Con- tion (phaco) systems as they pertain to a surgeon’s
level of experience.
sultants have been invited by the Editorial Review
Dr. Seibel: In one perspective, vacuum-based pumps,
Board to respond to questions posed by the Acade- such as a venturi system, are simpler than f low-based
systems, such as peristaltic pumps, in that they have but
my’s Practicing Ophthalmologists Advisory Committee a single f luidic parameter for the surgeon to adjust. The
surgeon commands only vacuum (mmHg) in a venturi
for Education. While the advisory committee reviews pump and clinically produces f low to the extent that
the aspiration port is open, and grip to the extent that it
the modules, consultants respond without reading the is occluded. The f low pump, by comparison, has at least
2 f luidic parameters (aspiration outf low in cc/minute
module or one another’s responses. – Ed. and vacuum limit in mmHg) as well as additional f low
modulations that may alter rates of pump speed change
with varying degrees of occlusion or vacuum load. This
relative difference in f luidic complexity would seem to
tip the scales in favor of a vacuum pump for less experi-
enced surgeons. However, vacuum pumps have a signif-
icant disadvantage of producing potentially dangerous
f low rates of 60 cc per minute or higher with typical
commanded vacuums of modern phaco techniques (250
to 450 mmHg). For this reason, a peristaltic pump may be
preferred by a less experienced surgeon (or a more expe-
rienced surgeon dealing with compromised anatomy) for
the ability to set a slower, gentler f low rate.
followability include increasing f low rate (f low pump), Torsional movement allows emulsification to take place
increasing vacuum limit (f low pump), or increasing com- when the tip moves in each cutting direction, or 100%
manded vacuum level (vacuum pump). of the time, whereas longitudinal ultrasonic cuts only in
the forward stroke, or only 50% of the time.
Dr. Serafano: “Followability” is the surgeon’s term for
The advantage of torsional over transverse is that the
how readily the nuclear fragments are attracted to the
tip can be controlled separately in the x and y axis as
ultrasound tip. In peristaltic systems, this can be opti-
2 distinct motions (ie, longitudinal and torsional). The
mized by adjusting the AFR setting. AFR is the rate at
limitation of the transverse motion is that its movement
which f luid is removed from the anterior chamber. It
is along 1 plane only (cannot mix the x with the y axis
is expressed in milliliters per minute. Irrigation inf low
movements of the tip).
must be equal to, and potentially greater than, the sum
of the AFR plus incisional outf low to maintain a steady
5. Discuss settings for patients with a relatively soft
IOP and therefore a stable anterior chamber. In coax-
cataract, such as those with primarily a posterior
ial phaco, inf low entering through the infusion sleeve
subcapsular cataract in a younger patient.
ports creates a f luid current that helps bring nuclear
fragments into the phaco tip if the AFR is balanced for Dr. Seibel: Using phacodynamic principles (Phacodynam-
the surgeon’s technique. The peristaltic system allows ics, 4th ed, Thorofare, NJ: Slack; 2005), the surgeon ascer-
for the direct adjustments of the AFR, whereas most ven- tains the clinical task for each parameter for a given type
turi systems cannot accomplish this. Some venturi sys- of cataract at a given moment in surgery. Ultrasound
tems have introduced “pitch and yaw” to the foot pedal functions to mechanically disrupt lens material that
in order to allow for more control of vacuum and power. would otherwise be too rigid to deform sufficiently for
aspiration under a given vacuum load alone. Therefore,
4. Does either torsional or transverse movement offer in a soft cataract, only minimal amounts of ultrasound
advantage over the other for cataract surgery? energy are required in order to moderately reduce the
amount of f luidics required; 15% to 20% might be a
Dr. Seibel: Controversy exists concerning a blanket supe-
starting point, realizing that there is no industry stan-
riority of one modality over another, and whether either
dardization for phaco power. Vacuum is titrated to allow
is absolutely superior to traditional longitudinal ultra-
deformation and aspiration of the material. In these
sound that is modulated with hyperpulse. Using ultra-
cases, a value of about 150 mmHg is typically sufficient,
highspeed videography, Teruyuki Miyoshi elegantly
assuming augmentation by mild levels of ultrasound.
illustrated the enhanced proximal followability afforded
Flow functions to pull nuclear material into the tip and
by hyperpulsed longitudinal phaco. The best statement
aspirate out of the eye, and for softer nuclei, settings of
for this question would be the recommendation to avoid
20 cc to 25 cc per minute will often suffice. Given these
traditional longitudinal continuous ultrasound in virtu-
relatively moderate f luidic settings, a bottle height of
ally all applications in favor of ellipsoidal, torsional, or
95 cm should provide adequate chamber stability. Such
longitudinal ultrasound with hyperpulse.
settings are starting points, with adjustments made
Dr. Serafano: There are currently 3 movements avail- according to phacodynamic principles based on surgeon
able with phaco tips. First is the traditional longitudinal observations.
movement along one line (x axis) that all machines have
Dr. Serafano: Soft cataracts require less ultrasonic power.
available. Second is torsional movement, which is “side
This allows for “vacuum-assisted phaco.” Once hydrodis-
to side” (y axis). Third is transverse movement, which
section and hydrodelineation have been completed, vac-
closely resembles the longitudinal line motion, but it is
uum levels and aspiration f low rates can be increased
delivered at an angle to the tip shaft (x and z axis).
from normal levels. In order to maintain a steady IOP,
Torsional and transverse are meant to improve effi-
bottle height may have to be increased to balance the
ciency by eliminating or minimizing the repulsive force
increased AFR. Divide and quadrant management or
inherent in the forward stroke of longitudinal phaco.
6. Discuss the advantages and disadvantages and fluid 7. What adjustments to the surgeon’s usual setting
dynamics for microincisional phacoemulsification. should be made in patients with pre-existing cor-
neal disease such as Fuchs corneal dystrophy?
Dr. Seibel: Advantages include a lower amplitude of
induced incisional astigmatism, as well as a more stable Dr. Seibel: Parameters should be adjusted to mini-
anterior chamber due to less incisional leakage, all other mize further trauma to already compromised corneal
things being equal (eg, rigidity of the infusion sleeve). endothelial cells. Ultrasound usage can be minimized
Originally, microincisional phaco involved a bimanual through more efficient modulations such as hyperpulse,
approach that had the disadvantage of requiring a modi- and by using more efficient occlusion type methods such
fication of technique and at least some learning curve. as chopping, and of course ultrasound should not be
The smaller incisions (typically 1.4 mm wide) required engaged when the tip is anterior to the iris plane. With
smaller-gauge instruments (20 to 23 gauge), and these regard to f luidics, anterior chamber turbulence should
often had difficulty in supplying sufficient infusion be minimized by decreasing the f low rate (either directly
inf low to mitigate against postocclusion surge with with a f low pump or indirectly with a vacuum pump)
higher vacuum levels. Newer coaxial microincisional and by minimizing chamber f luctuations by reducing
systems essentially eliminate the learning curve associ- postocclusion surges, by either raising bottle height,
ated with bimanual methodology but increase incisions reducing vacuum parameter, or dynamically reducing
up to 1.8 to 2.2 mm, depending on the platform, reduc- the commanded vacuum level (with dual linear pedal
ing induced astigmatism (as compared to traditional control) just prior to an occlusion break. Attempts can
19‑gauge phaco) to around 0.25 diopters. Adequate cham- also be made to keep the irrigation ports oriented in the
ber stability is achieved in these systems by restricting iris plane so as not to direct f luid f low with correspond-
outf low, such as with small-bore aspiration line tubing ing shear force directly against the corneal endothelium.
in the Intrepid system (Alcon), or a vacuum surge sup-
Dr. Serafano: Protection of the endothelium is the goal
pressor used in the Stellaris (Bausch & Lomb) platform.
in this situation. This is best achieved by low f low and
Dr. Serafano: I would like to define microincisional phaco low turbulence. Decreasing these parameters will assist
as a phaco incision of 2.4 mm or less. Since the total area in the retention of OVD and therefore endothelial protec-
for f luid exchange is reduced with smaller incisions, irri- tion. Low f low means to lower the bottle height, lower
gation may be reduced depending on the tip and sleeve aspiration f low rate and lower vacuum. Consider replac-
chosen by the surgeon. Therefore, a knowledgeable and ing OVD as needed and try to remove as much nucleus
matching selection of the aspiration and infusion com- as possible below the iris plane.
ponents is mandatory. Depending on component selec-
tion, to compensate for decreased irrigation, AFR and 8. Are there any adjustments to be made in a patient
vacuum levels may have to be reduced to balance f luidics with eyes that have long or extremely long axial
and prevent surge and IOP f luctuations. Some microin- lengths?
cisional ultrasound tips have been modified by slightly
Dr. Seibel: These patients have a greater likelihood of
f laring the port of the tip to maintain holding power and
having weak zonules and/or an excessively deep ante-
decreasing the outer diameter of the shaft to allow for
rior chamber, setting these patients up for lens-iris dia-
more irrigation to f low between the tip and sleeve into
phragm retropulsion syndrome in which the complete
the anterior chamber. The main advantages of micro-
pupil perimeter seals against the anterior lens capsule.
coaxial phacoemulsification are the improved patient
Surgeons would historically have to compensate by low-
outcomes, which include lower postoperative surgically
ering the infusion bottle, which would compromise
surge control in an already unbalanced anterior cham- with large f luctuations in anterior chamber depth is an
ber. Robert Cionni reported on a novel technique to dis- eye having an IOP of 5 mmHg or less after glaucoma fil-
rupt this syndrome and restore normal anterior chamber ter surgery. In eyes with excessive filters, the anterior
depth and architecture by mechanically separating the chamber will not react predictably with normal phaco
iris from the lens, with the phaco/irrigation/aspiration f luidics. In all of these situations, lower the initial f low
tip (in pedal position 1) or a second instrument. These parameters in order to decrease anterior chamber f luctu-
longer eyes often have large anterior segments, and sur- ations. After the initial entry into the anterior chamber
geons must be particularly vigilant about appropriate and establishment of f low, the parameters may be slowly
sizing of the capsulorrhexis; direct measuring is advised, increased to find the proper f luidic balance.
such as with the Rhexis Ruler (MST, Redmond, WA).