Sei sulla pagina 1di 41

Gas Chromatography

Troubleshooting and Reference Guide


Version 1.0, August 2005

Copyright 2005 MSP Kofel


GC Troubleshooting and Reference Guide
Troubl es hoo t ing Re f e re n ce
Troubleshooting Tools........................... 1 What Is a Capillary Column? ................ 7
Eight Problem Categories ...................1 Stationary Phase Considerations ............. 8
Baseline Disturbances .........................1 Bonded and Cross-Linked
Spiking .......................................... 1 Stationary Phases ..................................... 8
Noise ................................................1 Column Length .................................. 9
Wander ......................................... 2 Column Diameter ....................................9
Drift (Upward/Downward) ............. 2 Film Thickness ..................................... 10
Offset............................................ 2 Phase Ratio (ß) ......................................11
Irregular Peak Shapes or Sizes ...........2 Capacity............................................. 11
Reduced Peak Sizes .......................3 Temperature Limits .............................. 12
Tailing Peaks ...................................4 Column Bleed .................................... 12
Rounded or Flat-Topped Peaks .......4 Chemical Compatibilities ...................... 13
Split Peaks .......................................4 Column Storage ................................. 13
Negative Peaks ...............................4 Selecting Capillary Columns ................ 14
Retention Time Shifts ........................ 4 Column Installation Tips....................... 15
Loss of Separation or Resolution .........4 Carrier Gas ........................................... 16
Quantitation Difficulties ...................5 Makeup Gas ...................................... 17
Rapid Column Deterioration ............... 5 Capillary GC Injectors........................... 18
Ghost Peaks ........................................5 Injection Techniques ............................ 19
Broad Solvent Front ........................... 5 Split Injection ................................. 19
Troubleshooting Tools........................... 6 Splitless Injection ............................. 20
On-Column Injection ...................... 21
Megabore® Direct Injection .............. 21
Injector Liners ....................................... 22
Split Injector Liners ........................... 22
Splitless Injector Liners ..................... 22
Megabore® Injector Liners ................ 22
Septa ............................................. 24
Guard Columns/Retention Gaps .......... 24
Unions, Glass Press-Fit ............................ 24
Traps .................................................... 25
Column Contamination ....................... 26
Performance Chromatogram
Definitions .......................................... 26

Column Test Standards ........................ 28


Troubleshoo t ing GC
The gas chromatograph an d Column Test Mixture or Baseline Disturbances
capillary column function as a
Reference Sample (Figure 1) see page 2 for figure
complete system and not as two
These are used to diagnose select
individual parts. A problem or defi- Spiking:
system and column problems. They are
ciency in any part of the system 1. Particulate matter passing
useful to compare current system
usually will result in some type of through the detector.
performance to past performance.
chromatographic difficulty. The same Solution: Clean the detector per the
problem can be caused by a number instruction manual.
of different system deficiencies. A Checkout Column
logical and controlled roubleshooting This is a column that is not used for 2. Loose connections on cables or
procedure will quickly and accu- samples. The performance and circuit boards (usually random
rately identify the source of the quality is known so that evaluation of spiking).
prob- lem. This will result in the the system can be made. It helps to Solution: Clean and repair the
fastest, easiest and most complete verify or eliminate the previous electrical connections as needed.
solution to the problem. column as the source of a problem.
Noise:
Troubleshooting is a skill that be- Instrument Manuals 1. Contaminated injector and / or
comes easier with practice. Someone These are not a last resort. The column.
equipped with the right tools and a manuals are a good source of Solution: Clean the injector.
rudimentary understanding of cap- illary troubleshooting information special to Solvent rinse the column (pg 26).
column gas chromatography, can a particular model of gas chro-
matograph. Performance 2. The column is inserted into the
identify, locate and correct prob- lems
specifica- tions are often contained flame of an FID, NPD or FPD.
with minimal amount of effort.
in the Solution: Reinstall the column.

Troubleshooting Tools manuals. 3. Air leak when using an ECD


or TCD.
Flow meter
Solution: Find and repair the
A digital or manual model with Eight Problem leak.
a range of 10 to 500 mL / min is
suitable.
Categories 4. Incorrect combustion gases or
Most performance problems can be flow rates when using an FID,
New Syringe placed within one of eight NPD or FPD.
A working syringe that has not been areas. These are baseline distur- Solution: Check and reset the
gases at their proper values.
used for samples should be avail- bances, irregular peak shapes or
able. Some problems may actually be sizes, retention time shifts, loss of 5. Physical defect in the detector.
syringe or autosampler related. separation or resolution,quan- Solution: Clean or replace parts as
necessary.
titation difficulties, rapid column
Methane or Another deterioration, ghost peaks and broad 6. Defective detector board.
Nonretained Compound solvent fronts. It is not uncommon to Solution: Consult the instruc-
A non-retained compound is used to have more than one of these prob- tion manual or contact the GC
set and verify carrier gas flow manufacturer.
lems occurring at the same time.
and to check out injector operation
Sometimes, it is difficult to
and setup.
deter-
mine the actual nature of the
New Septa, Ferrules and
prob- lem. This makes a logical and
Injector Liners systematic approach to problem
These are used to replace parts that solving very important.
eventually become defective, worn out
or dirty.
It is important to realize that the
following comments and recom-
Leak Detector
mendations are generalizations
Electronic models are recommended.
Liquid leak detection fluids are and simplifications. Every possible
satisfactory, but care has to be problem or correction cannot be
exercised to avoid possible contami- covered, nor can every detail be
nation problems. mentioned. The p age w here addi-
tional information can be found is
shown in parentheses following each
solution.
Phone 031 972 3152 © 2005 MSP 1
GC Troubleshoo t ing
Baseline Baseline Disturbances

Disturbances (Continued)

Wander:
1. Contaminated carrier gas if using
isothermal conditions.
Solution: Change the carrier gas or 63,.,1*
SPIKING 12,6(
NOISE :$1'(5
2))
WANDER
use (change) carrier gas impu- rity
traps (pg 25).
2. Contaminated gas chromatograph.
Solution: Clean the injector
and / or gas lines. Solvent rinse the
column (pg 26).
3. Poor control of the carrier gas or
'5,)7 2)) 6(7
OFFSET
detector gas flows. DRIFT:$1'(5 '5,)7 SLIPPERY
Solution: Clean, repair or WHEN WET
change the flow controller. '

4. Poor thermal control of the detector.


Solution: Consult the instruction Figure 1
manual or contact the GC
manufacturer. Offset:
1. Injector or column Irregular Peak
Drift (Upward): contamina- tion. Shapes or Sizes
1. GC or column contamination. Solution: Clean the injector.
Solution: Clean the injector. Solvent rinse the column (pg 26). (Figure 2) See page 3 for figure
Solvent rinse the column (pg 26). 2. Column is inserted into the flame No Peaks:
2. Damaged stationary phase. of an FID, NPD or FPD. 1. Plugged syringe.
Solution: Replace the column. Solution: Reinstall the column. Solution: Clean the syringe or use
Determine the cause of the 3. Contaminated carrier or a new syringe.
dam- detector gases.
age (oxygen, thermal or chemical) to 2. Broken column.
Solution: Change the gases or
prevent future problems (pg 15). install (change) impurity traps Solution: Replace or reinstall the
(pg 25). column.
Drift (Downward): 4. Contaminated detector. 3. Injecting the sample into the
1. Incomplete conditioning of the Solution: Clean the detector. wrong injector.
column. Solution: Use the correct
5. Malfunctioning or improperly set
Solution: Condition the column injector or move the column
recording device.
until a stable baseline is obtained (pg to the correct injector.
Solution: Check the recorder
15).
settings. Consult the instruc- 4. Column installed into the wrong
2. Unequilibrated detector. tion manual, or contact the detector.
Solution: Allow the detector manufacturer.
enough time to equilibrate. Solution: Reinstall the column
into the correct detector.
5. Integrator or recording device is
connected to the wrong detector
or not connected at all.
Solution: Connect the
integra- tor to the correct
detector.
6. Detector gases improperly set
or not on.
Solution: Check and reset the
detector gases.

2 © 2005 MSP FAX 031 971 4643


Troubleshoo t ing GC
Irregular Peak Shapes Irregular Peak Shapes and Sizes

or Sizes (Continued)
7. Very low or no carrier gas flow.
Solution: Immediately lower
the column temperature to
35-40C. Measure and verify the
carrier gas flow rate (pg 17). TAILING
7$,/,1* FRONTING
)5217,1* SPLIT
63/,7
Check for leaks.

All Peaks Reduced in Size:


1. Partially plugged syringe.
Solution: Clean the syringe or use
a new syringe.
2. Change in the injection tech- REDUCED
5('8&(' ROUNDED
5281'(' NEGATIVE
1(*$7,9(
nique.
Solution: Check the injection '
technique and verify that it is the
same as before. Figure 2

3. Large leak in the injector (usually


accompanied by poor peak higher boiling solvent (pg 20). Select Peaks Reduced in Size:
shapes). 10. High background signal caused 1. Column and / or liner activity or
Solution: Find and repair the by contamination, excessive contamination, if the reduction
leak. column bleed (damage) or or loss is for active compounds
4. Split ratio is too high. autozero problem. (e.g., amines, carboxylic acids,
alcohols, diols).
Solution: Lower the split ratio (pg Solution: Clean the GC. Sol-
Solution: Clean or replace the
19). vent rinse the column (pg 26).
injector liner (pg 22-23). Solvent
5. Too short of a purge activation Replace the bleeding column rinse or replace the column (pg 26).
time for splitless injections. ( pg 12-13). Check the autozero
2. Leak in the injector, if the reduc-
Solution: Increase the purge function and setting.
tion or loss is the most volatile
activation time (pg 20). 11. Improperly operated detectors. compounds.
6. Very high septum purge flow. Solution: Consult the instruc- Solution: Find and repair the
tion manual for the proper gas leak.
Solution: Decrease the septum
purge flow (pg 18). flows and type and operating 3. Too high of an initial column
guidelines. temperature for splitless or
7. Too low of an injector tempera- on-column injections.
ture (especially for high 12. Impurities in the detector gas.
Solution: Decrease the initial
molecu- lar weight or low Solution: Use impurity traps
column temperature or use a
volatility an d / or replace the contami- higher boiling solvent (pg 20).
compounds). nated gas (pg 25).
4. Mixed sample solvents for split-
Solution: Increase the injector 13. Detector-compound mismatch. less or on-column injections.
temperature (pg 18). Solution: Make sure that the Solution: Use a single solvent
8. Column temperature is not detector will respond to the for sample injection (pg 20).
hot enough. compounds being analyzed. 5. Decomposition or error in the
Solution: Increase the column 14. Excessive attenuated integrator sample.
temperature or the upper tem- signal. Solution: Check and verify the
perature value of the column Solution: Check an d verify the sample integrity and
concentration.
temperature program (pg 12). attenuation settings.
9. Initial temperature of the column 15. Sample concentration or
is too high for splitless or on- integrity problems.
column injections. Solution: Check the sampleís
Solution: Decrease the initial concentration or stability.
column temperature or use a

Phone 031 972 3152 © 2005 MSP 3


GC Troubleshoo t ing
Irregular Peak Shapes 11. Solution: Use a pH-modified Retention Time Shifts
or Sizes (Continued) stationary phase. Derivatize the 1. Different column temperature.
compounds. Some peaks will Solution: Check and verify the
Tailing Peaks: always exhibit some tailing. column temperature or tempera-
ture program.
1. Active injector liner or column.
Rounded or Flat-Topped Peaks: 2. Different carrier gas flow rate or
Solution: Clean or replace
liner (pg 22-23). Replace the 1. Overloaded detector. linear velocity.
column if it is damaged. Solution: Decrease the amount of Solution: Check and verify the
sample reaching the detector. carrier gas flow rate or linear
2. Contaminated injector liner or velocity (pg 16-17).
column. 2. Exceeding the range of the
Solution: Clean or replace the integra- tor or recording device 3. Leak in the injector, especially the
injector liner (pg 22-23). Solvent (especially for computer systems). septum.
rinse the column (pg 26). Solution: Reset the range or Solution: Find and repair the
attenuation levels on the recorder. leak. Change the septum.
3. Dead volume caused by a poorly
installed column, liner or union. 4. Contaminated column.
Solution: Check and verify the Split Peaks: Solution: Solvent rinse the
installation of each fitting. Re- column (pg 26).
1. Poor injection technique (jerky
install the column, if necessary. or erratic). 5. Change in the sample solvent.
4. Poorly cut column end . Solution: Change injection Solution: Use the same solvent
Solution: Recut and reinstall the tech- nique (smooth and steady for all samples and standards.
column (pg 15). plunger depression).
5. Polarity mismatch of the 2. Poorly installed column in the in-
station- ary phase, solute or jector. Loss of Separation or
solvent. Solution: Recut the column end (pg Resolution
Solution: Change to a solvent or 15) and reinstall in the injector.
phase that have a better po- larity 1. Contaminated column.
3. Column temperature fluctuations. Solution: Solvent rinse the
match (pg 8). Solution: Check the oven column (pg 26).
6. Cold spot in the flow path. temperature or contact the GC
Solution: Check the flow path of manufacturer. 2. Damaged stationary phase.
the sample for possible cold Solution: Replace the column.
4. Coelution of two or more Excessive bleed should be evident
spots or zones. compounds. also (pg 12-13).
7. Solid debris in the liner or Solution: Check for any changes in
column. the operational parameters. 3. Different column temperature,
Solution: Clean or replace the Contamination or a change in the carrier flow rate or column.
liner (pg 22-23). Cut the ends of the sample will introduce additional Solution: Check and verify
column until the debris is compounds to the injected sample. temperature programs, flow rates
removed (pg 15). Check for these possibilities. and column identity.

8. Poor injection technique (usu - 5. Mixed sample solvent for splitless 4. Large changes in the sample
ally too slow of an injection). or on-column injections. concentration.
Solution: Change injection Solution: Use a single solvent for Solution: Adjust or compensate for
technique. sample injections (pg 20). the concentration change.

9. Too low of a split ratio . 5. Improper injector operation.


Solution: Increase the split Negative Peaks: Solution: Check the
ratio (pg 19). tempera- ture, split ratio, purge
1. All peaks are negative .
time and type of liner (pg 18-
10. Overloading on a PLOT column.
10. Solution: Check the polarity of the
23). Also check for leaks.
Solution: Decrease the recorder connections.
amount of sample reaching the 2. Select peaks on a TCD.
column. Solution: Compoun d has greater
Some compounds such as thermal conductivity than the car- rier
alcoholic amines, primary gas; a negative peak is expected in this
case.
and secondary amines, and
carboxylic acids tail on most 3. After a positive peak on an ECD .
columns. Solution: Dirty or old ECD cell .
Clean or replace the ECD.

4 © 2005 MSP FAX 031 971 4643


Troubleshoo t ing GC
Quantitation Rapid Column Ghost Peaks
Difficulties Deterioration 1. Contamination of the injector
or column.
1. Injection technique. 1. Exposure of the column to air
Solution: Clean the injector
Solution: Use a consistent (oxygen) at elevated tempera-
and liner (pg 22-23). Solvent rinse the
injection technique. tures.
column (pg 26).
2. Split discrimination. Solution: Find and repair any leaks
(pg 1). Check the quality of the 2. Septum bleed.
Solution: Use a consistent injec- Solution: Use a higher temper-
tion technique (volume, injector impurity traps and carrier gas (pg
25). ature septum. Lower the injector
temperature and split ratio) (pg 18-
temperature. Condition septum
19). 2. Exceeding the upper temperature
before use (pg 18).
3. Using a different purge activation limit of the column for prolonged
periods. 3. Previous run terminated too
time for splitless injection.
Solution: Replace the column. Do soon.
Solution: Use a consistent purge Solution: Use a higher
activation time (pg 20). not exceed the upper tem-
perature limits (pg 12). temper- ature to elute all of the
4. Baseline disturbances. sample components. Prolong the
3. Chemical damage.
Solution: See the section on run
Solution: Do not inject
baseline disturbances time to allow the complete elu-
inor- ganic acids or bases (pg
(pg 1-2). tion of the sample.
13).
5. Improper integrator or recorder
4. Contamination of the column
settings.
with high molecular weight ma-
Solution: Check and verify the Broad Solvent Front
terials.
integrator and recorder settings.
Solution: Use a sample 1. Poorly installed column.
6. Inconsistent detector gas flows preparation technique to remove the Solution: Recut (pg 15-16) and
or temperatures. problem contaminants. Use a guard reinstall the column.
Solution: Check and verify column (pg 24, 26).
detector operation. 2. Leak in the injector.
5. Column breakage. Solution: Find and repair the
7. Column or liner activity Solution: Avoid abrading or leak.
(adsorption). scratching the column. Avoid
Solution: Clean or replace the 3. Too low of a split ratio.
sharp turns or bends in the
injector liner (pg 22-23). Solvent Solution: Use a higher split
tubing (pg 14-15).
rinse or replace the column. ratio (pg 19).
4. Too low of an injector
tempera- ture.
Solution: Use a higher injector
temperature (pg 18).
5. Too long of a purge activation
time for splitless injections.
Solution: Use a shorter purge
activation time (pg 20).
6. Large injection volume.
Solution: Decrease the
injection size.
7. Low column temperatures and
high boiling solvent.
Solution: Use a higher initial
column temperature or a lower
boiling solvent (pg 20).
8. High column temperatures and
low boiling solvent.
Solution: Use a lower initial
column temperature or a higher
boiling solvent (pg 20).

Phone 031 972 3152 © 2005 MSP 5


GC Troubleshoo t ing
Technical Support
MSP offers a variety of products Septa and Ferrules MSP employs skilled scientists
that assist with troubleshooting. MSP offers a complete line of whose first priority is to answer
Please contact us to get the current silicone septa and ferrules. Over- your technical questions. These
cataolog. used septa and ferrules are prone to scientists offer analytical consult
leaks, which can cause column bleed ing and assist you in selecting
due by allowing oxygen to be columns and accessories. No
introduced. Particulates from matter which produts you are
ADM 1000 the overused septa and ferrules can using, theyëre here to help you
Model Flowmeter also cause problems when they with your chromatography
Flow rates are critical to efficient contaminate the liner. questions.
GC operation. Make sure flow
rates are correct by using the J&W
model ADM series of flowmeters.
They are based on ìacoustic
displacementî technology. No
bubbles, messy liquids or breaking
glassware to deal with. Ideal for
field or laboratory use. These flow-
meters are compatible with all
noncorrosive gases. A computer-
optimized calibration
Septa
incorporat- ing a NIST calibrated
flow
standard ensures the highest avail- able
accuracy, making ISO9000 and GLP
compliance that much easier.

Ferrules

4 mm Splitless Liner
Pyrolyzed compounds can build
up on liner walls. This buildup
causes clogging and sample ad-
ADM 1000 Model Flow meter sorption, which can result in a
nonrepresentative chromatogram.

Hamilton Cemented
Needle
Be sure to have a clean, working
syringe. Problems can sometimes
be traced to the autosamplers.
J&W offers a complete line of
4 mm Splitless Liner
Hamilton Syringes.

Hamilton Cemented Needle

6 © 2005 MSP FAX 031 971 4643


Re f e rence GC
Capillary Column
Upon first inspection, fused silica Polyimide Coating The most common capillary
capillary columns appear to be sta- tionary phases are silicone
quite simple. Further investigation Immediately after the drawing
poly- mers (Figure 2). The type
reveals that capillary columns are process, the outer surface of the
and
actually complex, highly sophisti- cated tubing is coated with polyimide. This
amount of substitution on the
devices. Considerable tech- nological polyimide coating serves two
polysiloxane backbone
knowledge, attention to detail and functions. First, it fills any flaws in the
distin-
refined techniques are required to tubing. Second, it provides a
guishes each phase and its
produce capillary strong, waterproof barrier. Both
proper- ties. The phase
columns of the highest quality. functions add to the strength and
description refers to the amount
Capillary columns are much more durability of the tubing. Any
an d type of substi-
than just tubes containing a dam- age to the polyimide
tution on the polysiloxane back-
polymer. coating will result in a weak
bone. For example, a
point and is a
(5%-phenyl)-methyl phase has two
potential for tubing breakage. The
What Is a Capillary Column? phenyl groups bonded to 2.5%, by
color of the polyimide often varies
number, of the silicon atoms; the
A capillary column is composed of between columns. Color differ-
remaining 97.5% of the silicon
three parts (Figure 1): ences will have no effect on col-
atoms have methyl groups bonded to
1. Fused silica tubing umn performance or durability
them.
2. Polyimide coating because the polyimide coating is
3. Stationary phase on the outer surface of the column.
Column performance is strictly a
function of the deactivation of the
fused silica tubing and the quality of [
HO -- CH2 - CH2 - O -- H ] n
the stationary phase coated
onto its inner walls.
Figure 3

Stationary Phase
The stationary phase is a polymer Another widely used stationary
that is coated onto the inner wall of phase is polyethylene glycol
the fused silica tubing. The Æ
(Figure 3). Carbowax 20M is one of
thickness, uniformity and chemi- cal the most widely used polyethylene
nature of the stationary phase are glycols to be used as a gas chro-
(Not to Scale)
extremely important. It is the matographic phase. The major dis-
stationary phase that has the great est advantage to polyethylene glycol
Figure 1 influence on the separations phases is their high susceptibility to
Fused Silica Tubing obtained. structural damage by oxygen at
elevated temperatures. Damage
The fused silica used to manufac- occurs at lower temperatures and
ture capillary columns is synthetic lower oxygen levels than most
quartz typically containing less polysiloxane stationary phases. The
than 1 ppm metallic impurities. R high polarity and unique separa-
Blanks (preforms) of fused silica tion characteristics of polyethylene
are drawn through a furnace at a
carefully metered rate. Laser mi-
crometers are used to ensure a
[ O Si ] n glycol stationary phases are useful;
thus, the liabilities are tolerated.
constant tube diameter. As part of R A newer class of capillary column
the column manufacturing pro- contains a gas-solid adsorption
cess, the inner surface of the tub-
R = CH3 methyl type of stationary phase. These
ing is purified and deactivated.
CH2CH2CH2CN cyanopropyl columns are often called porous
This process is used to minimize
CH2CH2CF3 trifluoropropyl layer open tubular or PLOT col-
chemical activity (unwanted umns. PLOT columns contain a
inter- actions between the tubing layer of solid particles coated onto the
and inner walls of the fused silica
the injected sample) and to create a tubing. Instead of a gas-liquid par-
phenyl
chemically uniform surface for titioning process between the in- jected
the stationary phase. sample and stationary phase, a gas-
solid adsorption process
Figure 2 occurs. Examples of PLOT
stationary phases include polysty-
rene, aluminium oxide and
molecu- lar sieve.
Phone 031 972 3152 © 2005 MSP 7
GC Re f e rence

Stationary Phase Selectivity


ability related interactions. In-
Stationary Phase As for polarity, stationary phase
creased retention occurs for
Considerations selectivity is determined by its
solutes that are more polarizable. For
structure. Stationary ph ase selectivi-
Within a constant set of operating methyl- and phenyl-
ty is not completely understood,
conditions, it is the structure of the substituted polysiloxanes, it will be
nor can it be easily explained or
stationary phase that determines the only significant interaction.
characterized. Using a severe
the relative retention (elution Solubility of the solute in the
sim- plification and condensation,
order) of the compounds. Focusing stationary phase will affect
selectivity can be thought of as the
only on the column, the stationary retention. The more soluble a
ability of the stationary phase to
phase determines the relative solute is in the stationary phase,
differentiate between two com-
amount of time required for two the greater its retention.
pounds by virtue of a difference in
compounds to travel through the Polyeth- ylene glycols and
their chemical and / or physical
column. The stationary phase cyano- p ro pyl- substitute d
properties. From the perspective of
ìretardsî the progress of the polysiloxanes have
a stationary phase, if there is a dis-
com- pounds moving through the strong dipole and hydrogen
cernible difference in the properties
col- um n. If any two compou n ds bond- ing characteristics.
of two compounds, the amount of
take Trifluoro-
interaction between the compounds
the same amount of time to migrate propyl-substituted polysiloxanes
and the phase will be different. If
through the column, these two will have a moderate dipole
there is a significant difference in
compounds will not be separated char- acteristic. As previously
the interactions, one compound will be
(i.e., they co-elute). If any two com- stated,
retained to a greater extent and
pounds take a different amount of because of the inexactness of these
separation will occur. If there are no
time, these two compounds will be characteristics, predictions and
discernible differences, coelution
separated. In other words, the sta- precise explanations of solute
will occur. The compounds may
tionary phase retains one com- separations are very difficult.
have different structures or
pound to a greater extent than the
properties, but if a particular sta-
other.
tionary phase cannot distinguish
between the compound differences, Bonded and Cross-
coelution will occur.
Stationary Phase Polarity Linked Stationary
Columns are often selected on the Stationary phase and solute factors Phases
basis of their polarity. Polarity is a such as polarizability, solubility, The first capillary columns had
bulk property of the stationary magnitude of dipoles and hydrogen stationary phase coated onto the
phase and is determined by the bonding behavior will influence inner tubing walls without any
structure of the polymer. Station- selectivity. In many cases, more type of chemical attachment. The
ary phase polarity does not have a than one factor will be significant, stationary phase was easy to dis-
direct influence on the separations thus there will be multiple rupt or damage with solvents, heat or
obtained. Polarity will have an effect selectivity influences. contaminants. Removal of a
on a variety of column char- Unfortunate- ly, most compound short piece of tubing at the front of the
acteristics. Some of the most im- characteristics, such as the column was often necessary to return
portant characteristics are column strength of hydrogen column performance after
lifetime, temperature limits, bleed bonding or dipoles, are not readily phase disruption had occurred.
levels and sample capacity. It is available or easily determined. This The advent of bonded and
the selectivity of the stationary makes it very difficult to accurately crosslinked phases substantially
phase that directly influences the predict and explain the separations increased the stability and lifetime of
separations. Synonymous use of obtained for a column and set of capillary columns. The station-
polarity and selectivity is not compounds. However, some ary phase is bonded to the inner
accurate but is very common. gener- alizations can be made. All surface of the fused silica tubing by
stationary phases will have polariz- means of covalent bonds.
Crosslinking is the joining of the
individual strands of the polymer.
Unlike nonbonded phases, bonded and
crosslinked phases can be
solvent rinsed if they become con-
taminated, and they also exhibit
better thermal and solvent
stability.
8 © 2005 MSP FAX 031 971 4643
Re f e rence GC
Column Length Column Diameter
pounds, especially if they are not The internal diameter will have a
The effects of column length on a very similar in structure, polarity direct impact on the efficiency,
separation become less important as or volatility. Shorter columns are retention characteristics and
column length increases. Resolution also useful for screening analyses. sample capacity of a column.
is a function of the square root of Most analyses are performed with Smaller diameter columns are
column length. This means that intermediate column lengths (20-30 more efficient than larger
dou- bling the resolution between meters). Usually, 60 meter or diam-
two longer columns are necessary only for eter columns. For two columns of
peaks without changing any other extremely complex samples equivalent phase, film thickness,
column dimension or operational and special applications. Longer length and quality, the smaller
parameter, requires a fourfold in- columns will exhibit higher bleed diameter column will provide
crease in column length (e.g., 30 than a corresponding shorter col- better resolution of the peaks.
meters increased to 120 meters). To umn because of the proportionally Increased resolution is especially
halve the resolution via length alone will greater amount of stationary phase. beneficial when there are closely
require a reduction in length of 75% eluting sample components.
(e.g., 30 meters reduced to 7.5 meters). Increased retention will be ob-
A large portion of the col- tained with longer columns. As column diameter decreases,
umn length can be lost before Longer analysis times will result, the retention of a given solute will
resolu- tion (separation) is reduced especially for isothermal tempera- increase providing no other
significantly. In a practical sense, ture conditions. For a temperature changes to the chromatographic
removing 1 meter from a 30 meter program situation, the extra analy- sis system have been made. This
column will decrease the resolution by time can be somewhat reduced by inverse relationship is approxi-
only 1.7%. using a faster ramp rate. mately linear in nature. Figure 4
illustrates the difference in
Shorter column lengths are intended for reten- tion attributable to column
samples containing a diam- eter. Bleed increases
relatively small number of com- slightly as column diameter
increases.
Larger diameter columns have
greater sample capacities.

Effect of Column Diameter on Retention


Effect of column diameter on retention
Column: DB-5 Carrier: Helium at 40 cm/sec 1. Decane
a. 30 m x 0.25 mm I.D., 0.25 µm Oven: 105 C 2. 1-Octanol
J&W P/N: 122-5032 Injector: Split 1:100, 250 C 3. 2,6-Dimethylphenol
1 µL of column test mixture 4. 2,6-Dimethylaniline
b. 30 m x 0.32 mm I.D., 0.25 µm Detector: FID, 300 C 5. Naphthalene
J&W P/N: 123-5032 Nitrogen makeup gas 6. 1-Decanol
at 30 mL/min 7. Tridecane
8. Methyl decanoate

1
Tridecane k = 6.3 Tridecane k = 5.7
a. b.

Figure 4
Phone 031 972 3152 © 2005 MSP 9
GC Re f e rence

Film Thickness increased retention results in no thick of a stationary phase is used.


resolution improvement, and a loss Figure 5 illustrates the effect of
Film thickness will primarily in resolution may actually occur. film thickness on solute retention.
affect the retentive character and Increasing the film thick- Sample capacity increases dra-
capacity of a column. Increasing ness to improve separation will be matically with increasing film
film thickness will cause a sub- effective only for poorly retained thickness. Column bleed is much
stantial increase in the retention solutes. higher for thick film columns.
of a solute. Thick film columns
are used primarily for the separa- Thin film columns are useful for the
tion of extremely volatile solutes analysis of low volatility or high
without the use of cryogenic cool- ing. boiling samples. Thick film
For fast eluting solutes, the columns will excessively retain the
increased retention results in sample. Unnecessarily long analysis
improved resolution. For medium to times or high column
slow eluting solutes, the temperatures will result when too

Effect of Film Thickness on Retention

Column: DB-5 1. Decane


a. J&W P/N: 123-5032 2. 1-Octanol
30 m x 0.32 mm I.D., 0.25 µm 3. 2,6-Dimethylphenol
b. J&W P/N: 123-5033 4. 2,6-Dimethylaniline
30 m x 0.32 mm I.D.,1.0 µm 5. Naphthalene
Carrier: Helium at 40 cm/sec 6. 1-Decanol
Oven: 105 C 7. Tridecane
Injector: Split 1:100, 250 C 8. Methyl decanoate
1 µL of column test mixture
Detector: FID; 300C
Nitrogen makeup gas at 30 mL/min

a. Tridecane k = 5.7

b.
Tridecane k = 20.5

Figure 5

10 © 2005 MSP FAX 031 971 4643


Re f e rence GC
Phase Ratio Under these conditions, solute Capacity
retention (k) is inversely
The combined effect of column The capacity of a column is
propor- tional with column
diameter and film thickness on defined as the maximum amount of
diameter (ra- dius, r) and
retention can be understood by sample that can be injected into a
proportional to film thickness
comparing the phase ratio ( ) of column before significant peak
(d f ). Solute retention
different columns with the same distortion occurs. Capacity is
increases with decreasing column
stationary phase. diameter; solute retention de- directly related to film thickness,
creases with decreasing film column diameter and stationary
Equation 1: thickness. For the same solute and the phase polarity. Increased capacity
same column temperature, the relative results as film thickness and
r column diameter increase. Table 1
magnitudes of the phase
KC = kß = k lists the capacity ranges for a
2df ratio for two columns will allow
the prediction of which column variety of column sizes. Notice the
will be more retentive for a given increase in sample capacity with a
where
solute. The column with the decrease in the phase ratio.
r = column radius ( µm)
d f = film thickness ( µm) smaller phase ratio will be more
retentive for a given solute than a The more soluble a solute is in the
k = partition rati o
column with a larger phase ratio. stationary phase, the greater the
KC = distribution constan t
Table 1 lists the phase ratios for a column capacity for the solute. For
variety of column diameters and film example, a polar stationary phase
The distribution constant (KC) is (e.g., DB-WAX) will have a higher
thicknesses.
the ratio of the concentration of capacity for a polar solute (e.g.,
the solute in the stationary phase methanol) than a nonpolar solute
and mobile phase (c s / cm). The (e.g., hexane).
distribution constant remains the
same for column temperature, Exceeding the column capacity or
stationary phase and solute. ''overloading" is indicated by peak
broadening or asymmetry. Usually,
Diameter Film Phase Ratio Capacity * overloading is evident as a peak
(mm) (µm) (ß) (ng per with a leading edge (fronting or
component) sharkfin shaped). For gas-solid
phases (e.g., GS-Q and GS-Alu-
0.10 **0.18 250 40-50
0.20 225 55-70 mina), an overloaded peak will
0.30 150 80-100 appear to be tailing. Injector
0.40 113 125-150 problems can also give peak
0.25 0.10 625 40-50
shapes similar to those of over-
0.15 417 60-70 loaded peaks.
**0.25 250 100-150
0.50 125 200-250
1.0 63 350-400

0.32 0.10 800 60-70


0.15 533 70-80
**0.25 320 150-200
0.50 160 250-300
1.0 80 400-450
3.0 27 1200-1500
5.0 16 2000-2500

0.53 0.15 883 80-90


**1.0 133 1000-1200
1.5 88 1400-1600
3.0 44 3000-3500
5.0 27 5000-6000
Table 1

* Capacity for solutes that are well matched with the stationary phase. Amounts are for each component (single
peak) in the sample and not the amount of the entire sample. Capacity is defined as the amount where peak
broadening by 10% at half-height occurs. If the inner diameter you are using is not listed, please call MSP
for additional information.
** Standard.

Phone 031 972 3152 © 2005 MSP 11


GC Re f e rence

Temperature Limits time, use any column at the lowest The bleed trace is a record of the
reasonable temperature. There is a baseline rise during a temperature
All stationary phases have upper direct relationship between in- program. The rise is caused by the
and lower temperature limits creased column lifetime and increased normal degradation of
which define the temperature lower operating temperatures. the stationary phase at higher col-
range over which the column can Leaving the column oven at lower umn temperatures. For a blank run
be safely used. The lower tem- temperatures when the column is (no injection), there are several
perature limit is the point where not in use will extend its lifetime. important characteristics for the
the properties of the stationary bleed trace:
phase are not conducive to chro- 1. The baseline rise starts about
matography. At or below the lower Column Bleed 30-40°C below the isothermal
temperature limit, the column will Column bleed is defined as the upper temperature limit of the
exhibit poor separation and peak normal background signal caused by column.
shape problems. Permanent dam- the elution of stationary phase 2. The baseline remains relatively
age will not result if the column is (polymer) degradation products. level before the rise begins and
exposed to temperatures below These degradation products are after the maximum temperatur e
the lower limit. Heating the col- always present and are not neces- is reached and held.
umn above the lower temperature sarily a sign of a damaged column. 3. There are no discrete peak s
limit will restore its performance. present.
Every column, regardless of the
source or quality, will exhibit som e
Exceeding the upper temperature column bleed. The amount of nor- The presence of discrete peaks in a
limit will result in accelerated mal column bleed will increase blank run indicates that the inlet o r
degradation of the stationary front portion of the column is con -
with increasing film thickness,
phase. Upper temperature limits taminated. Stationary phase degra -
column diameter and length. Polar
are usually given as two numbers. dation is a continuous process an d
phases will usually exhibit slightly
The first or lower temperature is not an isolated, one-time event .
higher bleed levels than nonpolar A single point introduction o f
called the isothermal limit. This is phases. A bleed trace (normal back-
the temperatu re at w hich the col- ground) is shown in Figure 6. Ap- compound(s) into the column i s
umn can be maintained for indefi- proximately 10 ng of tetradecane required to obtain peaks. Sinc e
nite periods of time. The second or was injected into the column to use stationary phase degradation is a
higher number is called the as a scale. continuous process and peak gener -
temperature program limit. The
column should not be used at this Normal Bleed Profile
temperature for prolonged periods
(>10 min). The upper temperature
limits are not precise thresholds, Column: DB-5
but temperatures where column 30 m x 0.25 mm I.D., 0.25 µm
J&W P/N: 122-5032
lifetime is substantially decreased. Carrier: Helium at 40 cm/sec
Exceeding the upper temperature for Oven: 100-320C at 20/min
320C for 10 min
short periods of time will not result in Injector: Split 1:100, 250C
instant phase damage; Detector: FID, 300C
Nitrogen makeup gas at
however, the rate of column deg- 30mL/min
radation increases with increasing
(Peak represents approximately 10 ng of tetradecane)
temperatures.

Slight thermal damage is normally


evident as decreased column life-
time. As thermal damage in-
creases, peak tailing and excessive
column bleed may be seen. Very
rapid and substantial thermal
d am age occu rs w henever the col- 21 min
umn is exposed to elevated tem-
peratures without carrier gas flow.
Polar phases are much more sus-
Figure 6
ceptible to thermal damage than
nonpolar phases. To prolong life-

12 © 2005 MSP FAX 031 971 4643


Re f e rence GC
ation is the result of an isolated event normally during the higher tem- Column Storage
(momentary introduction of sample perature portion of the program,
into the column), discrete peaks in a creates the appearances of column Capillary columns should be
blank run cannot be due to a bleed. The presence of discrete stored so that abrasion of the
degrading stationary phase. peaks or humps in the blank run tubing is avoided. The ends
The presence of peaks in a blank indicates that there are residues should be sealed with a GC
run is a sure sign of contamina- contaminating the GC system. septa and the column returned to
tion. Contaminated columns can be its original box. Upon re-installa-
cleaned by solvent rinsing. tion, the column ends need to be
The actual magnitude of the trimmed to ensure that a small
base- line offset at higher piece of septa has not become
temperatures will be dependent lodged in the column.
Chemical
upon the sensi- tivity levels used.
Column bleed Compatibilities
becomes significant for trace level Bonded and cross-linked
analysis requiring high sensitivity phases are not damaged by
or for very sensitive detectors. injecting water or organic
Selective detectors may exhibit solvents. Poor results may be
higher bleed levels for certain evident with certain solvents
phases. For example, the baseline depending upon phase / solvent
offset for a cyanopropyl-phenyl polarity mismatches; however, no
containing phase will be greater with damage will occur. The only sub-
a nitrogen - phosphorus de- stances that will damage a sta-
tector (NPD) than a flame tionary phase are strong,
ioniza- tion detector (FID) due to inorganic acids (HCl, H 2SO4,
the H3PO4, HNO 3, etc.) and bases
presence of nitrogen in the station- (KOH, NaOH, NH 4OH, etc.).
ary phase degradation products. The Organic acids and bases will not
NPD is several orders of mag- nitude damage the
more sensitive to nitrogen than a FID. stationary phase unless they are
allowed to reside in the column
There are two reasons for exces- for prolonged periods. At low Dual Columns
sive column bleed or behavior ppm concentrations, HCl and
that appears to be column bleed: NH 4OH in the presence of water
damage to the stationary phase or will do minimal or no damage to the
contamination. Exposure of the phase, providing the column
stationary phase to oxygen at temperature exceeds 100 °C during
elevated temperatures will result the analysis. Non-aqueous
in phase damage. The damage is samples containing low levels of
magnified and accelerated at HCl or NH 4OH will not cause
higher temperatures. Oxygen ex- significant phase damage.
posure usually occurs when a leak is
present in the gas lines or injec-
tor. Thermal damage from over- GC Colu m n
heating the column or heating the T ubing Dia m e t ers
column without carrier gas flow
will also result in excessive bleed Column J&W Column Ferrule
due to phase damage. Chemical I.D. (mm) Nomenclature O.D. * (mm) I.D. (mm)
damage to the phase from the 0.05 Microbore 0.363 0.4
injection of inorganic acids or 0.10 Microbore 0.363 0.4
bases will cause a column to bleed 0.18 Minibore 0.340 0.4
0.20 Narrowbore 0.340 0.4
excessively. Contamination of the 0.25 Narrowbore 0.350 0.4
injector an d / or column by 0.32 Widebore 0.430 0.5
semivolatile residues introduced 0.45 Hi Speed Megabore 0.673 0.8
with the injected samples is an- 0.53 Megabore 0.673 0.8
other common source of excessive * Nominal outer diameter specifications
baseline rise. The elution of these
residues during later GC runs,

Phone 031 972 3152 © 2005 MSP 13


GC Re f e rence

Selecting Capillar y Length


Columns Diameter 1. A 30 meter column is suitable
The following guidelines and 1. Use a 0.25 mm I.D. column for for most applications.
generalizations may be helpfu l split and splitless injections
when attempting to selec t when sample overloading is not 2. Use a 15 meter column for
a problem. High column effi- simple samples (less than 10
columns.
ciencies are realized with these components) or for sample
Stationary Phase small diameter columns. screening purposes.
1. Use a stationary phase with 2. Use 0.32 mm I.D. colum ns for 3. Use a 60 meter or longer col-
polarity closely matching that splitless and on-column injec- umn for very complex
of the solutes (e.g., nonpola r samples or for situations re-
tions, especially when injecting
phase for nonpolar solutes). quiring the highest possible
large amounts of sample.
2. Use the least polar phase that number of theoretical plates.
3. Use 0.53 mm I.D. (Megabore) These long columns are usu
will provide satisfactory sepa-
columns as replacements for ally limited to use for the
ration; nonpolar phases ex- packed columns or for many
hibit superior lifetimes ove r analysis of complex samples
purge and trap applications . such as petroleum products,
polar phases. PCB congeners,
3. 4. Use 0.45 mm I.D. columns when dioxins, etc.
For general purpose use, the
best column is a DB-1 or ease of use like Megabore is
DB-5 . desired but greater column
efficiency is needed.
4. For solutes with dipoles or
5. Use 0.18 mm I.D. columns for
hydrogen bonding capabili-
GC / MS systems with low
ties, use a cyanopropyl (DB-
pumping capacities or when Other sources of column
1301, DB-1701, DB-225 and
very high column efficiencies selection information:
DB-23) or Carbowax (DB-WAX
are needed.
and DB-FFAP) stationary 1. Chromatograms in
phase. Do not forget to the catalog.
consider polarity and Film Thickness
temperature characteristics. 1. Use a standard film thickness 2. Literature references.
column for most applications.
5. For light hydrocarbons or 3. MSP
inert gases, use a PLOT 2. Use thin film columns for high Support: 031
column (GS-Q, GS-Alumina, boiling solutes such as petro- 972 31 52
or GS-Molesieve). leum waxes, triglycerides, ste -
roids, etc.
6. If possible, avoid using phases
containing the specific ele- 3. Use thick film columns for very
ment corresponding with volatile solutes such as gases,
element specific detectors low boiling solvents and
(e.g., cyanopropyl phases with purgeables.
NPDs and trifluoropropy l
phases with ECDs) .

7. The wid est range of selectivi-


ties using the smallest number
of columns will be DB-1 or
DB-5, DB-1701, DB-17 and
DB-WAX. More than 90% of
all analyses can be adequately
performed with these five
columns.

14 © 2005 MSP FAX 031 971 4643


Re f e rence GC
Column Installation age at these abrasion points.
Tips Column Placement
Column Conditioning
The column should be placed in the
Cutting the Column GC oven so that the tubing is not After a column has been properly
The importance of a properly cut subjected to tight bends or installed, it will need to be condi-
column is often under appreciated. turns. Bending places a high tioned. Heating a column without
A poorly cut column will have chips of amount of stress on the tubing. carrier gas flow will result in seri-
polyimide or fused silica ex- These areas of stress are more ous and permanent damage of the
posed to the solutes and the flow likely to break especially if the stationary phase.
path of the carrier gas. This may polyimide coating has been
result in tailing, split or broadened abraded or scratched in these Before the oven is heated for col- umn
peaks due to poor sample introduc- areas. The avoidance of unneces- conditioning, it is absolutely
tion into the column. Active com- sary bending stress is critical with necessary to verify that there is
pounds may also experience some larger diameter columns. Large carrier gas flow through the col- umn
adsorptive losses. diameter tubing cannot tolerate and there are no leaks. The most
sharp turns or bends as well as common sites for leaks are the
To properly cut a column, a carbide or small diameter tubing. Megabore septum and injector fittings. These
diamond tipped pencil or column (0.53 mm I.D.) columns should not be leaks can often be found with an
cleaving tool is needed. Lightly scribe wound on cages less than electronic leak detector. Caution
the outer surface of the
seven inches in diameter, or must be exercised when using a
column at the desired location. Very
premature breakage may occur. It is liquid leak detector such
little force is necessary since only ®
important that any points of as Snoop . If the leak is large,
the thin coating of polyimide needs to ®

be cut. Attempting to cut through the possible contact of the tubing with small amounts of the Snoop may
fused silica will result in an sharp edges such as column iden- aspirate into the leak and contami- nate
uneven and unsatisfactory cut. After tification tags, oven walls, column the injector, detector or col-
scribing the polyimide, grasp the hangers, etc. are eliminated. The umn. Using a nonretained
column on each side of the scribe. forced air currents inside the GC compound to verify the injector
By pulling along the tubing length and oven will cause movement of the
bending away from the scribe point, a column, and these contact points
clean cut will be obtained. This may will abrade the column. This will
seem awkward at first, eventually lead to column break-
but after several attempts, the pro- cess
will become quite routine. The end of
the column should be
examined with a 10-20X magnifier to
verify that a clean and straight cut has
been obtained.

It is important to cut the column


ends after the column nut and fer-
rule have been placed onto the
column. Very small pieces of ferrule
material may enter the column upon
placement of the ferrule over the
column end. If present, these ferrule
scrapings will cause peak broaden-
ing or tailing, loss of resolution or
solute losses due to adsorption. In
severe cases, blockage of the
column by the solid ferrule particles Column Installation Tools.
may occur.

Phone 031 972 3152 © 2005 MSP 15


GC Re f e rence

Column Installation De t ec t o r Vola t ile Co m pounds


Tips (Continued)
FID/TCD methane, butane
setup and operation is ECD methylene chloride * , other halogenated methanes*
recom- men ded. PID ethylene, acetylene
NPD acetonitrile *
After verifying a leak-free system, the MS butane, air, halogenated methanes*
column can be safely condi-
Table 2
tioned. Heat the column to its
isothermal upper temperature * Do not inject a liquid (neat) sample directly into the column, or overloading will occur. Make a headspace
injection by placing a small volume of the appropriate solvent in a septum capped vial. Shake the vial, then
limit or a temperature at least 20- insert the syringe needle into the headspace above the liquid. Pull up 1-3 µ L of the headspace vapors and
30°C above the highest oven tem- inject.
perature that will be used. Do not
exceed the upper temperature
limit of the column. Carrier Gas calculated.
Equation 2: Average linear
After 30-60 minutes at the condi- After conditioning the column, the velocity
tioning temperature, a steady (flat) carrier gas has to be accurately set. L
baseline should be obtained. If the The carrier gas flow rate is depen- µ (cm/sec) =
dent on column temperature. It is tM
baseline is unstable or excessively
important to set the carrier gas at where
high after 60-90 minutes of condi-
the same column temperature for a L = column length (cm)
tioning, do not continue to condi- tion
given analysis. Significant changes in tM = retention time of a
the column. There is either a leak in
the resolution can occur with nonretained peak
the system or the GC is
small changes in the carrier gas (sec)
contaminated. If there is a leak,
prolonged heating of the column flow rate. For convenience, the
will result in serious damage to carrier gas is often set at the initial Equation 3: Average volumetric
the stationary phase. The injector temperature of the temperature flow rate
may have residues from previ- program. r2L
ously injected samples. In many F (mL/min) =
For capillary columns, carrier gas tM
cases, the installation of a new
column into a contaminated injec- flow is best expressed as an
aver- age linear velocity, (µ, cm / where
tor will result in a contaminated
sec) r = column radius (cm)
column.
instead of a volumetric flow rate (F, L = column length (cm)
mL / min). The average linear ve- tM = retention time of a
Saturated gas traps, poor quality
locity can be thought of as the nonretained peak
carrier gas or a dirty detector will
average rate at which a nonretained (min)
contribute to an elevated or erratic
background signal. Be sure that compound travels through the
column or the "speed" of the carrier The nonretained peak should be
the integrator or recording device is
gas. very sharp and symmetrical. Any
not set at one of its most sensi- tive
peak broadening or tailing is un-
levels. Normal, but small,
Due to the low flow rates used with acceptable. Leaks in the injector, a
deviations in the detector signal
capillary columns, a bubble flow- poorly installed column or a lack
will appear to be very large at
meter will not provide an accurate and of sufficient makeup gas flow will
these sensitive settings.
reproducible measure of the carrier all cause tailing or broad,
gas flow rate nor can the nonretained peaks. Any problems
linear velocity be directly deter- will have to be corrected before
mined. The linear velocity is proceeding.
calcu- lated by injecting a highly
volatile compound that is not The effect of the carrier gas linear
retained by velocity on column efficiency is
the column. The recommended best described using van Deemter
compounds for various detectors are curves (Figure 7). The curves illus-
listed in Table 2. From the re- trate that there is an optimal linear
tention time of the nonretained velocity (µopt) that provides the
peak, the average linear velocity highest efficiency. This point is
can be calculated using Equation 2. where the curve reaches the small-
Using Equation 3, the average
volumetric flow rate can also be
16 © 2005 MSP FAX 031 971 4643
Re f e rence GC
Carrier Gas (Continued) van Deemter Curves provides the highest efficiency.
est value of H min (the point of Recommended linear velocities
greatest column efficiency). The 1.2
N2 for helium and hydrogen are
best resolution will be obtained 1.0
listed in Table 3.
when using a linear velocity that

H (mm)
0.8
generates the highest efficiency He
Use the highest purity carrier gas
for a column. The curves also 0.6
for maximum column life. The
show that using a linear velocity 0.4
H2
use of impurity traps (water and
that is too low or high will result 0.2 oxygen) on the gas lines is
in a rapid loss of column highly recommended to extend
efficiency. 10 20 30 40 50 60 70 80 90
column lifetime and to improve
Average Linear Velocity (cm/sec)
D517
detector sensitivities. The
Usually a linear velocity that is Figure 7 slightly higher cost of high
greater than the value purity gases will be offset by
correspond- ing to the minimum greatest column efficiency, but the
longer column and trap life.
minimum in the van Deemter
in the van Table 4 lists recommended
curves occurs over a very narrow
Deemter curve is used. A value minimum purities for carrier and
range and at a low linear velocity.
1.5-2 times the µopt (called the detector gases.
OPGV or optimal practical gas Substantial analysis speed must be
sacrificed for optimal resolution
velocity) is the point where maxi-
mum column efficiency per unit when using nitrogen. For helium
and hydrogen, high linear veloci-
time is obtained. Setting the linear
ties can be used to reduce analysis
velocity at a higher value also
compensates for the decrease in
times without sacrificing a large Makeup Gas
linear velocity with increasing amount of efficiency. The faster Most commercially available GC
flow rates also sweep the injector detectors require 30-40 mL / min
column temperature as encoun-
tered when using a temperature faster, which improves the sample total gas flows for best sensitivity
program. introduction process. Helium, and and peak shape. Carrier flows for
especially hydrogen, provide the capillary columns range from less
best resolution when the analytes than 1 mL / min to over 10 mL / min.
With helium and hydrogen as
elute over a wide temperature These flow rates are well below the
carrier gases, the minimum in the van
Deemter curves occurs over a much range. Nitrogen is not recom- range where most detectors will
broader range and at higher linear mended as a carrier gas for capil- exhibit optimal performance. To
velocities than with nitro- gen. Using lary columns even though it supplement the carrier gas flows,
nitrogen provides the makeup gas is added at the column exit
to obtain a total gas flow of 30-
Recommended Linear Velocities for 30 Meter Columns
40 mL / min into the detector. The
makeup gas can be the same as the
Column Linear Velocity Flow Rate carrier gas or a different gas de-
Diameter (cm/sec) (mL/min) pending on the type of detector
(mm)
being used. The makeup gas is
independent of the flame gases in
He H2 He H2
combustion type detectors (e.g., FID,
0.18 30-45 45-60 0.5-0.7 0.7-0.9
NPD and FPD). For some de-
0.25 30-45 45-60 0.9-1.3 1.3-1.8
tectors, such as the ECD, the
0.32 30-45 45-60 1.4-2.2 2.2-2.9
moder- ating or auxiliary gas may
0.45 30-45 45-60 2.9-4.3 4.3-5.7
act as the makeup gas. Exact
0.53 30-45 45-60 4.0-6.0 6.0-7.9
recommenda-
tions for makeup gas flows and
Table 3 type can be found in the instruction
Recommended Minimum Gas Purities manual for the GC being used.
hydrogen 99.99% carrier
99.95% FID
helium 99.995% carrier
nitrogen 99.999% carrier and ECD moderating gas
99.95% makeup
Table 4

Phone 031 972 3152 © 2005 MSP 17


GC Re f e rence

active compounds may occur will be smaller than the more vola-
Capillary GC Injectors when they come in contact with tile compounds. The longer the
the metal surfaces. sample spends in the heated injec-
There are two goals of sample in- tor, the less severe the discrimina-
troduction. One is to introduce the Backflash problems can be tion.
sample into the column such that it minimized by:
1. Using a septum purge with Septum Purge
will occupy the shortest length of split / splitless injectors.
column. The shorter the sample Most split / splitless capillary injec-
band is at the beginning of the 2. Using small injection volumes. tors have a septum purge function.
process, the sharper and more The septum purge minimizes the
narrow the peaks will be on the 3. Using large volume injector amount of septum bleed materials that
chromatogram. The end result is liners. may contaminate the GC sys-
more sensitivity and better 4. Using the optimal injector tem. The septum purge gas sweeps the
resolu- tion. The second goal is to temperature. bottom face of the septum and carries
have the contaminants out
the composition of the sample Injector Temperatures through the septum purge vent. The
introduced into the column be as septum purge flow is usually be-
The injector temperature should be
close to the composition of the tween 0.5 and 5 mL / min. High or
just hot enough to insure "in- stant"
sample before the injection. There higher than optimum septum purge
vaporization of the entire
should be no sample degradation or flows may result in the loss of some
sample without degrading any of the
adsorptive losses occuring or caused of the more volatile sample compo-
sample components. If the
by the injector. nents. The septum purge function is
injector temperature is too low,
not essential for good chromato-
carry over problems, incomplete
Backflash graphic results; however, any
sample vaporization or broad
septum bleed problems may be
With the exception of on-column peaks (especially the solvent front)
minimized.
injection, all injectors utilize va- will be obtained. If the injector
porization to introduce the sample temperature is too high, excessive
into the capillary column. The backflash or sample degradation may
injected sample is rapidly occur. Using an injector tem- perature
vapor- ized in the heated above the upper tempera- ture limit of
injector, and a gas (the carrier the column stationary phase will not
gas) flowing damage the col-
through the injector carries the umn. For most samples, 250 °C is a
sample into the column. One good injector temperature. Some
prob- lem that vaporization experimentation may be
injection warranted to obtain the smallest
techniques have is backflash. amount of backflash and the maxi-
Upon sample vaporization, the mum amount of sample
gaseous sample will expand to fill the vaporization.
injector liner volume.
Backflash is when the vaporized Inlet Discrimination
sample expands beyond the capac- Upon injection, the less volatile
ity of the liner volume and into the sample components will not va-
injector body. Since the sample porize as rapidly as the more
now occupies a larger volume, it volatile sample components. Im-
takes longer for the sample to be mediately following injection, the
carried into the column. A large vaporized sample has a greater
and tailing solvent front is ob- proportion of the more volatile
tained. If the vaporized sample compounds than the less volatile
comes in contact with cold spots like compounds. Thus, more of the
the septum and gas inlets of the volatile compounds are intro-
injector, small amounts of the
duced into the column. This effect is
sample may condense. This con- called discrimination. The peaks for
densation can result in carry over the less volatile compounds
problems (ghost peaks) on subse-
quent injections. The metal
injector is not inert and loss of
18 © 2005 MSP FAX 031 971 4643
Re f e rence GC
Injection Techniques Split Ratio: The amount of sample Applications involving highly
entering the column will be depen- concentrated samples or very
There are four major capillary dent on the carrier gas flows into small diameter columns may re-
injection techniques - split, splitless, on- the column and out of the split vent. quire the use of higher split ratios.
column and Megabore direct. By measuring the column flow and the
Note: Split ratio can be directly measured
Nearly every standard capillary split vent flow, the amount of using ADM flowmeters.
injector is capable of split and sample "splitting" that occurs can be
splitless injections. On-column calculated. This value is called the Equation 4: Split ratio
injections require a dedicated split ratio. The split ratio is nor-
capillary on-column injector. mally reported with the column
split vent flow
Megabore injections utilize flow rate normalized to 1. The split column flow
= split ratio
packed column injectors that have ratio is determined using Equation
been converted to a Megabore 4. The split ratio can be used to Example:
injector. estimate the amount of sample Column flow = 2 mL/mi n
entering the column. A split ratio of Split vent flow = 100 mL/mi n
Split Injection 1:50 would indicate that one part of the
Split injection is very simple and sample enters into the column
and 50 parts are discarded out of 100
the most common of the capillary Split ratio =
2
= 50
injection techniques. The highest the split vent. Therefore, 1 / 51 of the
resolution and system efficiencies total sample theoretically makes it into Therefore, the split ratio is 1:50.
are obtained with split injections. the column. Typical split ratios range
Split injections are used for highly from 1:10 to 1:100.
concentrated samples with typical
per component concentrations of 0.1-
10 µg /µ L. Injection volumes of 1-2
µL are normally used, but vol- umes
up to 5 µL can be used with- out
significant problems.

Split injection is a vaporization


technique (see Figure 8). The
sample is vaporized upon injection
and rapidly mixed with carrier gas. A
small amount of the carrier gas enters
the column, and a much
larger amount leaves the injector
via the split vent. Since the
vapor- ized sample is mixed with
the
carrier gas, only a small amount of
the injected sample actually enters
the column. The total gas flow
through the injector at the moment of
injection is quite high (the sum of the
column and split vent
flows). The sample is rapidly
swept into the column which ac-
counts for the high efficiency of
split injections. This also accounts
for the severe discrimination ob-
tained with split injections. The
less volatile compounds do not Figure 8
have sufficient time to fully
vapor- ize before they are
discarded via
the split vent.
Phone 031 972 3152 © 2005 MSP 19
GC Re f e rence

Injection Techniques
Splitless Injection
Splitless injections are used for
trace level analyses or when the per
component amounts are no more
than approximately 200 ng. Splitless
injections are slightly more
complex than split injections and are
subject to several restrictions and
conditions.

The injected sample is vaporized


and carried into the column by the
carrier gas. At the moment of injec-
tion, the flow through the injector is the
same (1-2 mL / min) as the col-
umn flow (see Figure 9, purge off).
Fifteen to ninety (15-90) seconds
after the mo ment of in jection, addi-
tional carrier gas flow is introduced
into the injector (see Figure 9, purge
on). This extra gas purges the injec-
tor of any remaining sample that has
not entered the column. The
time at which the extra gas flow is Figure 9
introduced is called the purge acti-
vation time (or purge on).
lower boiling solvent may remedy this
is not refocused as required for situation.
Solvent Effect and Cold Trapping:
With splitless injections, the sample is good peak shapes. The earlier
eluting peaks will suffer greater Inlet discrimination is less severe for
introduced into the column at a
peak shape degradation than the this technique than for split
much slower rate than for split in-
later eluting peaks. If this occurs, injections. If the sample solvent and
jections. To avoid extremely broad
either a lower initial column the column stationary phase
peaks, the sample needs to
tem- perature or a higher boiling polarities are very different, poor
be refocused before starting the
sol- peak shapes may occur. The
chromatographic process. One re-
vent will have to be used. If the polar- ity mismatch between the
quirement of most splitless injec-
sample components boil at 150 °C or phase
tions is that the initial temperature
above the initial column tem- and solvent causes this problem. The
of the column is at least 10 °C below
perature, the solvent effect does peak shapes can be improved by
the boiling point of the sample sol-
not have to occur for good peak installing a retention gap
vent. When the vaporized solvent
shapes. These high boiling (guard column).
leaves the injector and enters the
cooler column, the solvent rapidly com- pounds will cold trap in
condenses at the front of the col- the col- um n an d refocus into a
umn. A solvent film forms, and this short
film will trap and refocus the sample band without the aid of the
sample. This is called the solvent solvent effect.
effect. Starting at too high of a col-
umn temperature can result in Other Considerations for Splitless
broad and malformed peaks. The Injections: Injection sizes are
solvent film cannot form, and the usually limited to 2 µL or less for
sa mple is introduced into the col- splitless injectors. Large injection
umn over a long time (equal to the volumes will normally result in
purge activation time). The sample broader peaks. Another limitation of
splitless injections is peaks that elute
before the solvent front will be
malformed. Changing to a
20 © 2005 MSP FAX 031 971 4643
Re f e rence GC
Injection Techniques and the column is returned to the converted packed column instru-
(Continued) heated oven to start the ments or for situations where large
chromato- graphic process. This sample capacities are needed.
On-column Injection injector can be added to a GC Megabore injections are well
On-column injection is not a without dis- suit- ed for trace level analyses
vapor- ization technique. The abling existing injectors. also.
sample is Highly concentrated samples may
deposited directly into the column Retention gaps (guard columns) have to be diluted to avoid over-
with a syringe. On-column injec- are almost always required for on- loading the column. Small diame-
tion provides the optimum in capil- column injections. Substantially ter capillary columns (0.32 mm I.D. or
lary column performance by improved peak shapes will result less) are not compatible with
eliminating discrimination and when a retention gap is used. The Megabore injectors.
degradation effects that can result retention gap will also protect the
from using a vaporization tech- analytical column. Since the entire A Megabore injector consists of a
nique. On-column injections are sample is deposited directly into the glass injector liner that is held in
particularly well suited for high column, potentially large place with a metal injector fitting.
boiling compounds like petroleum amounts of contaminants or other The ferrule in the fitting seals and
waxes, triglycerides and thermally column damaging materials enter the holds the liner in place. The liner
labile compounds. column. These materials will deposit reduces the volume of the packed
in the retention gap in- column injector and provides an inert
On-column injection requires the stead of the column and longer environment for the sample to
solvent effect or cold trapping to column life and performance will vaporize (see the section on
obtain acceptable peak shapes. result. Megabore Direct Injector Liners,
Some on-column injectors use a page 250). The sample is injected and
secondary cooling function to Megabore Direct Injection rapidly vaporized in the liner (see
eliminate the need to cool the Figure 10). The carrier gas
Packed column injectors that have
entire column down to the appro- sweeps all of the vaporized
been converted to accept Mega-
priate temperature for the solvent sam- ple into
bore capillary columns (0.45 and
effect. Only a small portion of the the Megabore column. The gas
0.53 mm I.D.) are very simple and
front of the column is cooled at the flow through the injector is high (4-
relatively trouble free. Megabore
moment of injection. 10 mL / min), thus the sample is
columns can tolerate large injec-
rapidly introduced into the column.
tion volumes (5-6 µL) and high
One on-column injector design There is no splitting of
sample concentrations (1-10 µg).
forces air around a short length of the the sample or solvent / temperature
Megabore systems are used for
front of the column, which re- requirements.
condenses the sample in this
region. After the injection, the air is
turned off and this region of col-
umn will rapidly equilibrate to the
higher oven temperature. The
sample will now begin the
chro- matographic process.
Usually a
savings in time will result since the
column oven does not have to be
cooled to low temperatures before each
analysis.

Another injection design uses a


telescoping injector that allows the
front of the column to be pulled out of
the heated GC oven and into the
ambient air. The surrounding air
cools this portion of the column so
that the injection can be made. Figure 10
After injection, the unit is collapsed
Phone 031 972 3152 © 2005 MSP 21
GC Re f e rence

Injector Port Liners Splitless Injector Liners


Direct Flash Vaporization
Injector liners provide an inert envi- Splitless injector liners are nor- Liners:
ronment where the sample can mally straight tubes without any The direct flash vaporization liner
vaporize and be properly introduced into flow disruption devices. Any has a restriction at the top of the
the column. Liner design will flow disruption in the injector liner and another
vary depending on the type of will usually cause peak restriction several centimeters
injection technique being used. broadening. For this reason, it is below the upper restriction. The
Liners that are dirty, broken, poorly not recommended to pack a sample is injected into the
installed or incorrectly selected will splitless liner with silylated chamber formed by the two
often contribute to inferior glass wool. Sometimes there is a restrictions. When the syringe
chromatographic results. restriction at the bottom of the needle is inserted into the liner for
liner. This is to keep the column injection, the needle blocks most
centered in the liner and away of the upper restriction.
Split Injector Liners
from the liner walls. Small This prevents the vaporized
The split injector liner must have a volume liners will result in sample from escaping from the
large enough volume to accommo- greater inlet efficiency since the top of the liner. The lower
date the expansion of the vaporized sample is transferred into the restriction is tapered so that the
sample. However, the volume must be column over a shorter period of Megabore column will become
small enough for the gas flow to quickly time. The smaller volume liners are lightly wedged in the taper. The
sweep the vaporized sample into the more subject to backflash polyimide column coating will
column. Liners with too problems. For small injection compress and cause a leak-free seal
small of a volume will be subject to volumes (<1 µL), the 2 mm I.D. between the column and
severe backflash problems. Larger splitless liner is recommended; for the liner. Carrier gas (or
liner volumes become less important at larger injection sample) will not escape around the
high split ratios since the high volumes, the 4 mm liner is rec- column. Backflash is greatly
carrier gas flows rapidly sweep the ommended to minimize reduced, thus the solvent front is
injector. Flow disruption within the backflash narrow and without signifi- cant
liner will insure problems. tailing.
thorough mixing of the sample, thus
minimizing discrimination problems. Megabore Injector Liners Direct flash vaporization liners
Various types of flow disruption can be packed with silylated glass
The type of direct injection liner
liners are available. The greatest wool providing there is
used will greatly impact the
discrimination is obtained with no top restriction. This
quality of Megabore
straight bore liners while the in- restriction can be removed by
chromatog- raphy. There are
verted cup liners discriminate the cutting the liner, but the
three general
least. The hourglass shaped liners purchase of liners without a top
types of direct injection liners:
provide good sample mixing, and restrictor is recommended. The
the straight tube, the direct flash
they are easier to clean than the glass wool should be lightly
vaporization liner and the hot
inverted cup liners. packed so unnecessary peak
on-column.
broadening will not occur. Do
Packing a split injection liner with not use regular laboratory
Straight Tube Liners: Using a
silanized glass wool is another way grade glass wool.
straight tube type of Megabore
to insure flow disruption. The packed
liner is not recommended. It will
liner has a higher thermal mass It is essential that the column
typically give a very broad and
which will aid in the rapid end is cleanly cut and checked
tailing solvent front which may
volatiliza- tion of the less volatile with
interfere with some of the early
components. Additionally, the glass a magnifying lens. The column is
eluting peaks of interest. Due to the
wool acts as a filter to help to trap inserted into the liner until it fits
lack of any type of
some of the snugly in the restriction.
restrictions, the vaporized
nonvolatile materials in the injected Too much force will crush the
sample can readily backflash out of
sample. The glass wool should be column end, and too little force
the liner. The severity of
lightly packed so unnecessary peak will not seal the column in the
solvent front tailing will be more
broadening will not occur. Do not use restriction. If the column is not
pronounced for large injection
regular laboratory grade glass wool. sealed properly,
volumes, volatile sample sol-
vents and solutes, low carrier gas the solvent peak will tail exces-
flows and excessively hot sively. Also, any debris in the
injectors. injection liner taper (small

22 © 2005 MSP FAX 031 971 4643


Re f e rence GC
pieces of septa, polyimide from gauge) can now be inserted temperatures are lower than
the last installation, etc.) will directly into the Megabore those used for vaporization tech-
prevent a proper seal. column. It is very important that niques, thermally labile samples can
the syringe needle is straight also be chromatographed
Hot On-column Liners: The hot and that the end has no burrs or with the on-column Megabore
on-column injection port liner hooks. Injection volumes are liner. A direct flash vaporization
has a single tapered restriction at the limited to 0.5 µL or less. liner system will be more
top of the liner. The suitable and better in greater
Megabore column will seal in Hot on-column Megabore injec- than 95% of all Megabore appli-
this region in the same manner as tions are suitable for high cations. Hot on-column Mega-
for the direct flash boiling samples that are difficult bore liners should be used only
vaporization liner. The needle of by standard vaporization when direct flash vaporization
a standard GC syringe (a 26 techniques. Since the injector liners are not suitable.

Phone 031 972 3152 © 2005 MSP 23


GC Re f e rence

Injector Septa Unions Useful tips for achieving a gas tight


seal are as follows: The ends of the
High temperature septa are Glass Press-fit Unions
tubing to be joined should be cut
recom- mended to minimize Universal glass press-fit unions very cleanly with a column cutting
injector and column replace stainless steel, Vespel, and tool. Examination under a 10-20X
contamination by septa bleed resin methods of joining two pieces of magnifier is advisable. Cleaning
materials. It is recom- fused silica tubing. The glass the ends of the tubing by swabbing
mended that septa should be re- unions form a gas tight seal by a with methanol and rinsing the
placed after 30-50 manual
minor compression of the polyim- ide union with methanol can help en-
injections or 75-100 autosampler
coating on the column (Figure sure a seal. Making sure there is no
injections. Leaky septa will allow air
12). There is no need for wrenches, carrier gas flowing, the column
to enter the column which will
glue or ferrules. Since these unions are ends are pressed firmly into the
rapidly damage the column. Septa
will have to be replaced more fre- very light weight and have union. A visible ring of compressed
quently when using large diameter very little thermal mass, they track polyimide inside the union and a gentle
needles or needles with burrs or oven temperature and eliminate tug on the connection
hooks. Over tightening the septum cap the possibility of a cold or hot spot in should confirm that the seal has
will reduce septa life also. The column the sample path. Because the glass been formed. If difficulties arise in
temperature should be unions have a similar expan- sion making a good seal, try another glass
reduced to 35 °C or less when coefficient to fused silica union. In some cases, there may be
changing the septa. Reduced col- tubing, they expand and contract some variability in the
umn lifetime will result if air is together and can be used over a wide taper of the union.
introduced into the carrier gas temperature range (-60 °C to 350°C).
(column) at elevated temperatures. With proper installation,
they exhibit no solvent tailing or
adsorption of active compounds.
Guard Columns Universal glass unions can be used to
Guard Columns/Retention Gaps
connect fused silica tubing with outer
diameters of 0.4-0.8 mm.
A guard column or retention gap is
a short length of deactivated
fused silica tubing (0.5-1.0 meter)
installed between the inlet and the
analytical column (Figure 11). The
tubing serves two functions, one
as a guard column and the other
as a retention gap.

As a guard column, it catches non-


volatile residues which prevent
them from reaching the analytical
column. As a retention gap, the
tubing allows larger amounts of
solvents with polarities unlike that of
the stationary phase to be in-
jected without adversely affecting
peak shapes.

Glass or stainless steel unions


may be used to join the analytical
column to the guard column.

Figure 12 Figure 11

24 © 2005 MSP FAX 031 971 4643


Re f e rence GC
Traps Generally, the polyethylene glycol purity helium instead of nitrogen.
based phases (i.e., DB-WAX, These traps will need periodic
Traps are strongly recommended, DB-FFAP and Carbowax) are replacement. The use of an indi-
even if high purity carrier gases cating oxygen trap is recom-
readily oxidized, especially at
are used. Trace amounts of con- elevated temperatures. mended to determine when the high
taminants can cause baseline Polysiloxane phases can also be capacity oxygen trap needs
irregularities, detector noise and replacing.
irreversibly damaged via oxida- tion
possible column damage. Con- at high temperatures, but
taminated detector gases can lead to much more slowly than polyethyl- Glass Indicating Oxygen Traps
artificially high baseline ene glycol based phases. Oxidized
readings and poor performance. A glass indicating oxygen trap
phases exhibit poor chromato- should be placed downstream of the
graphic performance and higher high capacity oxygen trap as a means
Individual traps are designed to than normal bleed. Oxygen has a
remove moisture, oxygen, hydro- of indicating when to re-
deleterious effect on many GC place the high capacity non-indi-
carbons and other contaminants detectors: it degrades the perfor-
from the gas supply. Traps are cating trap. This will prevent
mance of electrolytic conductivity premature disposal of the high
offered with either 1 / 4" or 1 / 8" detectors (ELCD), reduces fila-
Æ
Swagelok fittings. The trap sys- capacity oxygen trap. The indicat- ing
ment lifetime in a thermal conduc- material undergoes a vivid color
tem recommended is a moisture tivity detector (TCD) and reduces
trap, high capacity oxygen trap, and change as it is depleted. The high
the linearity and sensitivity of capacity oxygen trap must be replaced
indicating oxygen trap. The
electron capture detectors (ECD). immediately upon the
moisture trap should be first in
Both the column and detector first indication of color change in
line because moisture will quickly benefit when oxygen is removed
deactivate the oxygen traps. The high the indicating trap. (It is not un-
from the carrier gas supply. usual to partially expire the indi-
capacity oxygen trap should
be installed second followed by an cating trap on each end upon
Oxygen contamination can usually be installation). Fully expired oxygen
indicating oxygen trap. An expired attributed to small leaks in the gas
trap should be replaced immediately. traps will recontaminate the gas
lines, septum leaks or low with previously trapped materials.
purity carrier gas. Oxygen traps Unnecessary frequent maintenance of
Moisture Traps only address the last problem. the GC system and the
A moisture trap should be in- possibility of reduced column
stalled in the carrier and detector gas lifetime will be avoided by the
High Capacity Oxygen Traps
line(s). It will remove trace timely changing of the high capac-
levels of water, some nonperma- The most efficient oxygen traps
ity oxygen trap. The indicating
nent gases and light hydrocarbons. are 99% efficient at reducing oxy-
oxygen trap will rapidly expire if used
Two types of absorbents are avail- gen in carrier gases such as he-
as the solitary oxygen trap in the
able. Both contain blue indicating lium, nitrogen, hydrogen, argon,
system or for prolonged peri- ods
Molecular Sieve 4A that turns tan argon-methane, or CO 2 at 99.999% after the high capacity oxygen trap
at 20% humidity. Molecular Sieve 5A purity. A 20 to 40 fold reduction in
has expired.
will trap water, some non-per- manent oxygen in the gas is often ob-
gases and small hydrocar- bons tained. These traps
(<n-C4H10). Molecular Sieve 13X will effectively remove oxygen Help me ! I'm
trapped and I
will trap larger hydrocarbons in from argon-methane can't get out
addition to most of the impurities mixtures (used with
trapped by the Molecular Sieve 5A ECDs) without
material. disturbing the
ratios of these gases.
Oxygen Traps
Metal bodied traps
Trace levels of oxygen in the car- avoid the signal noise
rier gas can dramatically shorten the associated with any plastic
life of a gas chromatographic bodied trap. These traps are filled
column. under a heated flow of ultra high '

'

Phone 031 972 3152 © 2005 MSP 25


GC Re f e rence

Column Performance
Columns can be easily rinsed Chromatogram
Contamination using a rinse kit (see Figure 13).
One of the most common causes of Before rinsing the column, break Definitions
off one-half meter from the injec- Partition Ratio (k)
column performance degradation
is contamination. Usually these tor end. Place the detector end of The partition ratio (k) is a measure of
contaminants originate from the the column into the rinse kit. Fill how much time a solute spends in the
injected samples. Semivolatile or the vial with the appropriate sol- stationary phase relative to
nonvolatile sample components vent. Apply 10-15 psi pressure to the time it spends in the mobile
accumulate in the injector and force the solvent through the col- phase (carrier gas). All solutes
column after repeated injections of the umn. For columns with 0.18-0.32 spend the same amount of time in
sample. Dirty samples will mm I.D., use 4-5 mL of each sol- the mobile phase, thus the partition
contaminate the GC system at a vent. For larger I.D. columns, use ratio is directly related to retention
faster rate than clean samples. Loss 8-10 mL of each solvent. Each caused by the stationary phase. It is a
of resolution or separation, solvent should stay in the column unitless number and can be calcu-
peak shape problems and / or base- for at least 10 min utes. The p revi- lated using Equation 5. The parti- tion
line disturbances (artificial bleed) are ous solvent does not have to ratio is a better measure of retention
common symptoms of column com- pletely vacate the column than the actual solute retention time.
contamination. Active compounds before rinsing with the next The partition ratiois inversely
such as carboxylic acids, amines, solvent. proportional to column
phenols and diols are particularly temperatures. This means that re-
affected by contamination. After the last solvent has left the tention increases as the column
column, allow the pressurizing gas temperature is decreased and reten-
Solvent rinsing a capillary column to flow through the column for 5-10 tion decreases as the column tem-
will remove most contaminants minutes. Install the col- umn into the perature is increased.
and restore column performance. injector of the GC
A column must have a bonded and and let carrier gas flow through the Equation 5: Partition Ratio (k)
cross-linked stationary phase to column for 5-10 minutes. In- stall
be solvent rinsed. the column in the detector and
check for leaks. Heat the column tR - tM t ’R
A good general column rinsing at k= =
tM tM
can be accomplished by using 2-3°C / min u ntil the normal con di-
methanol, methylene chloride, and tioning temperature is reached.
where
hexane in series. Other Condition the column as usual. t R = solute retention time
solvent choices will work, but t M = retention time of a non
they should meet the following retained compound
criteria: tíR = adjusted retention time

Include a polar and nonpolar


solvent. Start with the most
polar and finish with the least
polar solvent.
Include the injection solvent if Capillary
Column Column
possible. Hanger
Each successive solvent must be Special
Connector
1/16'' Flexible
miscible with the previous one. Teflon Line
& Ferrule
If aqueous-based or extracted
Flexible Teflon
samples were injected Tubing
(biological extracts, soils,
Special Adapter
waste water, etc.), begin with To Regulated
Pressure Source
water followed by methanol. Cap

If using an ECD, avoid Capillary


halogenated solvents as the final Column

rinse solvent, or if using an Beaker for


Collecting Vial
NPD, avoid acetonitrile. Solvent D 051

Figure 13

26 © 2005 MSP FAX 031 971 4643


Re f e rence GC
Performance theoretical plates per meter and Retention Indices (I)
greater efficiency. High efficiency
Chromatogram columns will have small values of
Retention Index (I) is a measure of
Definitions the relative retention of a solute
H. van Deemter curves are usually compared to normal alkanes (hy-
(Continued ) reported using h values. Equation 7 drocarbons) at a given temperature
Column Efficiency or Number o f can be used to calculate H. (isothermal) for a particular sta-
Theoretical Plates (N )
tionary phase. Retention indices
Column efficiency is the relationship Equation 7: Height equivalent to a
normalize instrument variables so
between solute retention time and theoretical plate (H)
that retention data can be com-
the amount of band broadening L pared for different chromato-
(peak width increase). Symmetrical H= graphic systems. For example, a
peaks with the smallest width are N solute with a retention index of
desired. Narrow peaks can be close where 1250 elutes between n-C12 (I=1200)
together and still be resolved. Broad- L = column length (mm) and n-C13 (I=1300) under the same
er peaks the same distance apart will N = number of theoretical test conditions. Retention indices
not be as well resolved and may plates
are useful when comparing relative
even coelute. Column efficiency is elution orders of various solutes for a
expressed as the number of theoreti- cal Coating Efficiency
given column and conditions.
plates (N). The number of theo- retical Coating efficiency is a comparison Also, retention indices are good for
plates can be calculated using Equation between actual column efficiency comparing the retention or selec- tive
6. The higher the number of theoretical and its theoretical maximum effi- behavior of two columns. Re-
plates, the higher the columnís ciency (Equation 8). The property tention indices can be calculated
efficiency and its potential to resolve actually compared is peak width. using Equation 9.
two closely eluting sol- Coating efficiencies of less than
utes. The number of theoretical 100% are departures from the Equation 9: Retention indices (I)
plates per meter is the usual method theo- retical behavior of a ìperfectî
col-
for reporting efficiency. Decreasing umn (i.e., broader peaks). There are log t' R(x)- log t' R(y)
column diameter will increase the I = 100y + 100(z-y)
some unavoidable measurement log t R' - log t ' R
(z) (y)
number of theoretical plates, thus errors, thus it is possible to obtain
increasing column efficiency. In coating efficiencies slightly greater
general, thinner film columns will than 100%. Coating efficiency can where
have slightly higher efficiencies be thought of as a measure of the t'R = adjusted retention time
than a corresponding thicker film uniformity of the stationary phase. x = solute of interest
column. A more uniform phase results in a y = normal alkane with y
column with higher efficiencies. number of carbon
Equation 6: Number of theoretical Coating efficiency is not the amount atoms eluting before
plates (N) of column coated with phase or the solute x
amount of column coated as the z = normal alkane with z
number of carbon
prescribed film thickness. Typical
tR atoms eluting after
coating efficiencies are 85-100% for solute x
N = 5.545 non-polar phases and 60-80% for
wh z-y = difference in carbon
polar phases. number between the
two normal alkanes
where Equation 8: Coating efficiency
N = number of theoretical
plates
tR = retention time Htheoretical
CE% = x 100
wh = peak width at half Hactual
height (in units of
time)

Another measure of column effi-


ciency is the height equivalent to a
theoretical plate (H). The shorter
each theoretical plate, the more
that can be placed into a length of
column. This translates into more

Phone 031 972 3152 © 2005 MSP 27


GC Re f e rence

Column Test Standards high molecular weight materials), or column and cleaning the injector.
an extremely damaged station- ary
One of the best means to evaluate the phase. Gas flow problems Acids and Bases
performance of a capillary include dead volume, leaks in the
column is by analyzing a Usually substituted phenols (acid) and
injector, improper installation of
proper- ly designed test anilines (base) are used to
the column, broken or improperly
standard. Test measure a columnís behavior
installed injector liner, backflash and
standards contain compounds that toward acidic or basic com-
poorly cut column ends.
have a range of functional groups. By pounds, respectively. The presence of
Straight chain hydrocarbons are
evaluating the shape, size and peak tailing for either peak
used to calculate retention indices.
retention of the various test stan- indicates that there is reversible
They may also be used to deter-
dard components, a substantial adsorption and that the column (or
mine the number of theoretical
amount of information is obtained perhaps the injector liner) is
plates or coating efficiency.
about the column and possibly the exhibiting acidic or basic charac-
chromatographic system. Every J&W teristics. The acid peak will tail if the
Alcohols
column is tested using a spe- cially column is too basic, and the
designed test standard. The actual Hydroxyl groups easily interact base peak will tail if the column is
chromatogram is included with every with any material or species in the too acidic. Columns should exhib-
column (see Figure 14 on page 275). carrier gas flow path that can hy- it a ìneutralî character if they are
It is important to save this test drogen bond. Ideally the only to be applicable for a wide range
chromatogram for future reference. interactions occurring is with the of analyses. The height of the acid
The column performance stationary phase. A tailing alcohol and base peaks are also compared
chromatogram provides peak often indicates column to the height of a hydrocarbon
additional information about the activ- ity. Silanol groups present peak. Hydrocarbons are not sus-
column such as temperature limits, in the column or injector liners ceptible to adsorption, thus they
actual dimensions and serial are a often serve as reference peaks. The
numbers. source of this activity. An oxidized ratio of the peak height of the acid
stationary phase will also cause an d base to a hy d rocarbon is cal-
Hydrocarbons tailing alcohol peaks. In most cases, culated. A reduction in the height
stationary phase damage and the ratio indicates that the column is
Hydrocarbon peaks are the stan- resulting active sites are irreversibly adsorbing the corre-
dard to which all other peaks are permanent and cannot be reversed sponding acid or base.
compared. Due to the lack of func- without substantial effort. Alcohol
tionality, hydrocarbons can only peak tailing can also be caused by Others
interact with the stationary phase. contamination. Sample residues
Hydrocarbon peaks should be Polynuclear aromatic hydrocar-
from previous injections or other
sharp and symmetrical. Mal- bons (PAHs) or fatty acid methyl
contaminants will interact with the
formed hydrocarbon peaks are esters (FAMEs) are often included in
alcohols resulting in tailing peaks.
due to gas flow problems, poor test mixtures. They may be
These contaminants can be re-
injection technique, column used to calculate the number of
moved by solvent rinsing the
con- tamination (especially theoretical plates, coating
solid or efficiencies, retention ( k) or
retention indices.

28 © 2005 MSP FAX 031 971 4643

Potrebbero piacerti anche