REAGENTS AND SAMPLES TRAY
• Positions for unrefrigerated racks 3
• Positions for racks 3 Capacity of refrigerated zone 30 reagent
bottles (20 or 50mL)
• Capacity of the sample racks 24
• Maximum number of samples 72
• ø13 mm, ø15 mm sample tubes (max. height 100 mm), ø13 mm
paediatric well
• Capacity of the reagent racks 10
• Maximum number of reagents 50 (plus 3 auxiliary fixed
positions)
• 20 ml and 50 ml reagent bottles
© BRII 2014
Possible configurations
© BRII 2014
• Cooling system
• The analyser is equipped with a second switch that
independently controls switching the refrigerator on and off.
• New Icons
© BRII 2014
Cycle time 24 s. (150 prep/h) 15 s (240 prep/h)
© BRII 2014
Sytem Liquid aprox 2.4 ml per prep. aprox 5 ml per prep.
Consumption
System liquid container 3000 ml 2700 ml
volume
Waste container volume 3000 ml 2700 ml
Waste and water level Waste and water level
control per capacitance control per load cell
detection
© BRII 2014
Scale Capacitance detection
© BRII 2014
A25Qty. A15Qty.
Electrovalve 3 1
Membrane Pump 2 1
© BRII 2014
Thermostated Probe
Peltier
© BRII 2014
Mechanical elements : Operating Arm
The only difference is in the rollers. has a line rollers and has
cylindrical rollers.
• No maintenance is needed (just remove dust)
• NEVER add oil in the carriage
Used in Z-axis
and X- up axis
Used in Y-axis
and X- down
axis
© BRII 2014
Peltiers connection A15 vs A25
© BRII 2014
Optical group differences
LAMP OF 12V/20W
5 READINGS
/SECOND
LAMP OF 6V/10W
1.25 READINGS
/SECOND
© BRII 2014
Reaction Rotor group differences:
Adjustment
screws
© BRII 2014
UNPACKING
© BRII 2014
© BRII 2014
• List of contents
© BRII 2014
Accessories
© BRII 2014
• Size 1080 x 695 x 510 mm (42.5”x27.4”x20.1”)
(length x depth x height)
• Weight 70 kg (154 lb)
o Avoid near sources of
Receive direct, intense or artificial light
Electromagnetic radiation (motors, centrifuge)
Heat sources
Leave a minimum space of 10cm between
the wall and the back of the analyzer
Surface must be rigid and robust
Front right leg is adjustable
© BRII 2014
© BRII 2014
• Power requirements
Power 300 VA
UPS 1500VA
© BRII 2014
User/Service Installation
© BRII 2014
• Recommended PC Minimum requirements
2Gb RAM
© BRII 2014
• User Software/Service Installation
© BRII 2014
Go to Control Panel, Click the Power Options.
Select High performance on the Power plan
Click change plan settings
o Dim the display Never
o Turn off the display Never
o Put the computer to sleep Never
Click the change advanced power setings,
Modify the following:
o Hard Disk
Turn off hard disk after Never
o USB settings
USB selective suspend settings Disabled
Click apply then ok.
© BRII 2014
Click Start -> Programs -> A25 Folder
Right click the A25 icon, Select Send to -> Desktop (create shortcut)
© BRII 2014
User Software
© BRII 2014
© BRII 2014
• New Icons
System conditioner:
Button that indicates to the analyser to perform a cycle of
conditioning of the fluid system. This conditioning consists of
performing a cycle of priming with air, another cycle with
washing solution and a final cycle with system liquid.
© BRII 2014
Incorporation of a reagent refrigeration system
© BRII 2014
Parts
© BRII 2014
Some tricks of the design:
1.- The shape of the hole prevent any
drop condensation to fall in to the
reagent bottle
2.- There are 3 drainage hole that send
the water to the metal sink of the peltier
radiator.
3.- Folded by neoprene to prevent any
water condensation externally.
4.- The reagent racks can be removed
with the reagent bottles together.
5.- On the bottom of the refrigerator
there’s one seal to prevent any leakage
of water condensation around the cover
6.- There’s on magnetic sensor for the
cover detection (brown-red cables) and
one led indicating the refrigerator status
(black-red cables)
© BRII 2014
Refrigerator removal:
© BRII 2014
RACK BOARD – FRIDGE CONTROL
- Schema of the voltage monitoring in the Peltier:
Peltier
ADC
Peltier
© BRII 2014
POWER MODULE FOR FRIDGE
This Power Board is the model HWS100-12/A of the manufacture Lambda
(TDK), one of the best brand in the world of power supply modules. The same
brand is used to supply the power to the full Analyser A25 v2.
© BRII 2014
CPU BOARD
The main changes are also visual. The only change that we can not do it are
some connectors to respect the compatibility with the previous version of A25.
© BRII 2014
Connector for Needle
Auto-centring by
Photodetector: In the
new analyser the old
conductive plate for
needle auto-centring
has been changed by 2
optical photodetectors,
for better robustness and
precision of the arm in
XY position.
© BRII 2014
Hinges
© BRII 2014
MAINTENANCE