Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User Manual
Canada
A DIVISION OF AMETEK PROCESS & ANALYTICAL INSTRUMENTS
PN 903-8726, Rev G Western Research
Offices
For other offices not listed here, visit us at www.ametekpi.com.
USA – HEADQUARTERS FRANCE
150 Freeport Road AMETEK – APIF
Pittsburgh, PA 15238, USA Rond point de l’epine des champs
Tel: 412-828-9040 Buroplus Bat D
Toll Free: 800-537-6044 78990 Elancourt, France
Fax: 412-826-0399 Tel: 33-1-30-68-89-20
USA – Delaware Fax: 33-1-30-68-89-29
455 Corporate Boulevard CHINA
Newark, DE 19702, USA AMETEK Commercial Enterprise (Shanghai) Co. Ltd.
Tel: 302-456-4400 (Main) Part A First Floor, 460 North
800-537-6044 (Service) Fute Road
800-222-6789 (Ordering) Waigaoqiao Free Trade Zone
Fax: 302-456-4444 Shanghai, 200131, China
USA – Texas Tel: 86-21-5868-5111
4903 West Sam Houston Parkway North Fax: 86-28-5866-0969
Suite A-400 Beijing Branch
Houston, TX 77041, USA Tel: 86-10-8526-2111
Tel: 713-466-4900 Fax: 86-10-8526-2141
Toll Free: 1-800-634-8990 Chengdu Branch
Fax: 713-849-1924 Tel: 86-28-8675-8111
CANADA Fax: 86-28-8675-8141
AMETEK Canada Guangzhou Branch
2876 Sunridge Way N.E. Tel: 86-20-8363-4768
Calgary, AB, T1Y 7H9, Canada Fax: 86-20-8363-3701
Tel: 403-235-8400 MIDDLE EAST – Dubai
Toll Free: 800-661-9198 P.O. Box 17067
Fax: 403-248-3550 Jebel Ali Free Zone
INDIA Dubai, UAE
AMETEK Instruments India Pvt. Ltd. Tel: 971-4-881-2052
1st Floor, Prestige Featherlite Tech Park Fax: 971-4-881-2053
Plot 148, EPIP Phase II SINGAPORE
Whitefield, Bengaluru – 560066, Karnataka, India AMETEK Singapore Pte. Ltd.
Tel: 91-80-6782-3200 No. 43, Changi South
Fax: 91-80-6782-3232 Avenue 2, #04-01
GERMANY 486164 Singapore
AMETEK GmbH Tel: 65-6486-2388
Rudolf-Diesel Strasse 16 Fax: 65-6481-6588
D-40670 Meerbusch, Germany
Tel: 49-2159-9136-0
Fax: 49-2159-9136-39
© 2012–2018 AMETEK Canada, A Division of AMETEK Process & Analytical Instruments Printed in Canada
This manual is a guide for the use of the Model 9900 Analyzer. Data herein has been verified and validated and is believed adequate for the intended use
of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and
suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability. This publication is not a license to
operate under, or a recommendation to infringe upon, any process patents.
Contents | iii
Customer Inputs........................................................................................................ 2-5
Supply Voltage.......................................................................................................2-6
Power Consumption..............................................................................................2-6
Measuring Cell Construction.................................................................................. 2-6
Sample Gas Requirements....................................................................................... 2-6
Sample System Pressure Limits............................................................................... 2-6
Flow Measurement (Optional)................................................................................ 2-7
Sample Gas Flow Rate.............................................................................................. 2-7
Zero Fluid................................................................................................................... 2-7
Storage Conditions.................................................................................................... 2-7
Operating Conditions............................................................................................... 2-7
Ambient Operating Temperature..........................................................................2-7
Humidity...............................................................................................................2-7
Maximum Altitude...............................................................................................2-8
Ingress Protection..................................................................................................2-8
Physical Dimensions................................................................................................. 2-8
Electrical Classifications............................................................................................ 2-8
Analyzer Certificates and Marking............................................................................... 2-9
CSA Certificate, Hazardous Locations................................................................... 2-9
CSA Certificate, Ordinary Locations.................................................................... 2-11
ATEX Certificate....................................................................................................... 2-13
IECEx Certificate...................................................................................................... 2-15
Analyzer Marking.................................................................................................... 2-19
Contents | v
CHAPTER 5 CALIBRATION.................................................................................................. 5-1
Zero and Span Calibration.............................................................................................. 5-2
Manually Zero-Calibrate the Analyzer and Optional O2 Sensor, and
View the Results........................................................................................................ 5-3
Span-Calibrate the Analyzer and View the Results............................................. 5-6
Setting Up Auto-Zero Parameters and Viewing Live Results – Optional...... 5-10
Viewing the Auto Zero Offset (Drift) Results..................................................... 5-16
Calibrate the Analog Outputs............................................................................... 5-17
APPENDIX A – OPTIONS........................................................................................................A-1
Paramagnetic O2 Sensor Concentration Measurement............................................ A-1
Contents | vii
Safety Notes
WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical instructions
as follows:
! An operating procedure which, if not strictly observed, may result in personal injury or envi-
ronmental contamination.
WARNING
! An operating procedure which, if not strictly observed, may result in damage to the equipment.
CAUTION
Electrical Safety
High voltages are present in the analyzer housings. Always shut down power source(s) before
performing maintenance or troubleshooting. Only a qualified electrician should make electrical
connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety
protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if
instrument is operated from an improperly grounded power source.
All personnel involved with the installation, start-up, operation, maintenance, service, or trou-
bleshooting of the analyzer must review and follow these Warnings and Cautions.
Warnings
Review and follow these Warnings to avoid personal injury or environmental contamination.
! All covers/doors must be tightly fastened with all screws and the Measuring Cell/Optical
Bench Assemblies must remain in place while the circuits are live.
WARNING
If installed in a hazardous classified area and it is necessary to remove/open the analyzer’s cov-
ers/doors while the circuits are live, first monitor the area for flammable gases and proceed only
when the area is safe.
! When the Analyzer Enclosure’s cover is off, take appropriate precautions to avoid electrical
shock. Hazardous voltages are present inside.
WARNING
! Always disconnect main AC power and/or alternate power sources (if used, i.e., for relay con-
tacts) to the analyzer before opening any covers/doors on the analyzer and before removing any
WARNING
components from the analyzer.
! All electrical connections, adjustments, or servicing of the analyzer should be performed only
by properly trained and qualified personnel.
WARNING
All electrical connections, materials, and methods (plus all safety policies and procedures) must
be made in compliance with local wiring regulations and electrical codes for the hazardous
area, as specified by the Owner Company, local electrical-inspection authority, and National/
EU regulations.
! Because ultraviolet radiation can harm your eyes, never view the light emanating through the
end window of the source lamp. If the source lamp must be viewed while energized, wear safety
WARNING
glasses that block the ultraviolet radiation.
Contents | ix
Cautions
Review and follow these Cautions to avoid damaging the equipment.
! If it becomes necessary to handle any of the electronic circuit boards, do not subject the boards
to static discharge. The ideal solution is a static-safe work area. Since such areas typically are
CAUTION
not available at field installation sites, the use of a wrist strap connected directly to a ground
is recommended. If a wrist strap is not available, you should at the very least touch the metal
chassis (to ground yourself) before handling or touching the boards.
! The electronic circuit boards and other static-sensitive components should be stored and trans-
ported in static-shielding carriers or packages.
CAUTION
! For electric-shock protection, the analyzer must be operated from a grounded power source that
has a securely connected protective-ground contact.
CAUTION
! When handling the source lamp, it is very important not to touch the window of the lamp be-
cause residual oils from the fingers will absorb ultraviolet light. The window is the flat surface
CAUTION
at the end of the narrow glass tube. The lamp assembly is fragile and should be handled with
care.
Environmental Information
This AMETEK product contains materials that can be reclaimed and recycled. In some cases the
product may contain materials known to be hazardous to the environment or human health.
In order to prevent the release of harmful substances into the environment and to conserve
our natural resources, AMETEK recommends that you arrange to recycle this product when it
reaches its “end of life.”
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a munici-
pal waste system (residential trash). The Wheelie Bin marking on this product is a reminder to
dispose of the product properly after it has completed its useful life and been removed from
service. Metals, plastics, and other components are recyclable and you can do your part by doing
one of the following steps:
• When the equipment is ready to be disposed of, take it to your local or region-
al waste collection administration for recycling.
• In some cases, your “end of life” product may be traded in for credit towards
the purchase of new AMETEK instruments. Contact your dealer to see if this
program is available in your area.
• If you need further assistance in recycling your AMETEK product, contact our
office listed in this manual.
Contents | xi
Electromagnetic Compatibility (EMC)
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designated to provide reasonable protec-
NOTE
tion against harmful interference when equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful interference to radio communica-
tions. Operation of this equipment in a residential area is likely to cause harmful interference in
which case the user will be required to correct the interference at own expense.
! Read and follow the recommendations in this section to avoid performance variations or dam-
age to the internal circuits of this equipment when installed in harsh electrical environments.
CAUTION
The various configurations of the Model 9900 Analyzer should not produce, or fall victim to, electromag-
netic disturbances as specified in the European Union’s EMC Directive. Strict compliance to the EMC Di-
rective requires that certain installation techniques and wiring practices are used to prevent or minimize
erratic behavior of the Analyzer or its electronic neighbors. Below are examples of the techniques and
wiring practices to be followed.
In meeting the EMC requirements, the various Analyzer configurations described in this manual rely heavily
on the use of metallic shielded cables used to connect to the customer’s equipment and power. Foil and braid
shielded I/O and DC power cables are recommended for use in otherwise unprotected situations. In addition,
hard conduit, flexible conduit, and armor around non-shielded wiring also provides excellent control of radio
frequency disturbances. However, use of these shielding techniques is effective only when the shielding ele-
ment is connected to the equipment chassis/earth ground at both ends of the cable run. This may cause ground
loop problems in some cases. These should be treated on a case-by-case basis. Disconnecting one shield ground
may not provide sufficient protection depending on the electronic environment. Connecting one shield
ground via a 0.1 microfarad ceramic capacitor is a technique allowing high frequency shield bonding while
avoiding the AC-ground metal connection. In the case of shielded cables the drain wire or braid connection
must be kept short. A two-inch connection distance between the shield’s end and the nearest grounded chassis
point, ground bar or terminal is highly recommended. An even greater degree of shield performance can be
achieved by using metallic glands for shielded cable entry into metal Electronics Enclosures. Expose enough of
the braid/foil/drain where it passes through the gland so that the shield materials can be wrapped backwards
onto the cable jacket and captured inside the gland, and tightened up against the metal interior.
Inductive loads connected to the low voltage “Alarm Contacts” are not recommended. However, if this
becomes a necessity, follow proper techniques and wiring practices. Install an appropriate transient volt-
age suppression device (low voltage MOV, “Transzorb,” or R/C) as close as possible to the inductive device
to reduce the generation of transients. Do not run this type of signal wiring along with other I/O or DC in
the same-shielded cable. Inductive load wiring must be separated from other circuits in conduit by using
an additional cable shield on the offending cable.
In general, for optimum protection against high frequency transients and other disturbances, do not allow
installation of this Analyzer where its unshielded I/O and DC circuits are physically mixed with AC mains
or any other circuit that could induce transients into the Analyzer or the overall system. Examples of elec-
trical events and devices known for the generation of harmful electromagnetic disturbances include mo-
tors, capacitor bank switching, storm related transients, RF welding equipment, static, and walkie-talkies.
! Explosion Hazard – Substitution of Components May Impair Suitability for Class I, Division 2.
WARNING
! Explosion Hazard – Do Not Disconnect Equipment Unless Power Has Been Switched Off or
the Area is Known to be Non-Hazardous.
WARNING
All input and output wiring must be in accordance with wiring methods authorized for the area
classification by the authority having jurisdiction. The need for sealing glands must be considered
for connections made directly to the enclosure when the area is pressurized or when the sample
is toxic or flammable in order to avoid the possibility of transporting gas to another area by way
of these wireways. Although the probability of leaks of the sample system is very low, a fitting left
un-tightened during maintenance could allow hazardous sample gas to flow into the analyzer
enclosure and the wiring area. Tests have indicated that with the sample system completely discon-
nected at the rear panel, all covers/doors installed/closed and on a table at atmospheric pressure,
with an inlet pressure of 13 psi applied to both inlet and vent ports, the pressure in the enclosure
will rise to 0.5 mBar (0.2” WC). This is very low pressure and normally would not pose any hazard,
but the installer must consider that if the analyzer is installed in a sealed (Type 4 or IP65) enclosure,
pressure at the wireway ends could build up to a level as high as the sample system inlet.
Contents | xiii
EU Declaration of Conformity
EU Declaration of Conformity
Manufacturer’s Name: AMETEK Canada LP
A Division of AMETEK Process & Analytical Instruments
(ISO 9001:2008 Registered)
Canada LP
A DIVISION OF AMETEK PROCESS & ANALYTICAL INSTRUMENTS
Western Research
The object of the declaration described [herein] is in conformity with the relevant Union harmonization
legislation (Directive 2014/34/EU):
EN 60079-0:2012 + A11 General requirements
EN 60079-15:2010 Equipment Protection by type of protection ‘n’
Certificate Number: DEKRA 15ATEX0118 X
Notified Body: DEKRA Certification B.V. 0344
Meander 1051, 6825 MJ Arnhem
The Netherlands
____________________________
Randy Meads
Quality Assurance Manager
Calgary, Alberta, Canada
June 28, 2017
Canada LP
A DIVISION OF AMETEK PROCESS & ANALYTICAL INSTRUMENTS
Western Research
Page 2 of 2
Contents | xv
Warranty and Claims
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications
which shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship un-
der normal use and service will be repaired or replaced (at our option) by us free of charge, provided that written
notice of such defect is received by us within twelve (12) months from date of shipment of portable analyzers or
within eighteen (18) months from date of shipment or twelve (12) months from date of installation of permanent
equipment, whichever period is shorter. All equipment requiring repair or replacement under the warranty shall be
returned to us at our factory, or at such other location as we may designate, transportation prepaid. Such returned
equipment shall be examined by us and if it is found to be defective as a result of defective materials or workman-
ship, it shall be repaired or replaced as aforesaid. Our obligation does not include the cost of furnishing any labor
in connection with the installation of such repaired or replaced equipment or parts thereof, nor does it include the
responsibility or cost of transportation. In addition, instead of repairing or replacing the equipment returned to us
as aforesaid, we may, at our option, take back the defective equipment, and refund in full settlement the purchase
price thereof paid by Buyer.
Process photometric analyzers, process moisture analyzers, and sample systems are warranted to perform the in-
tended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample
stream composition data, process conditions, and electrical area classification prior to order acknowledgment. The
photometric light sources are warranted for ninety (90) days from date of shipment. Resale items warranty is limited
to the transferable portion of the original equipment manufacturer’s warranty to AMETEK. If you are returning
equipment from outside Canada, a statement should appear on the documentation accompanying the equipment
being returned declaring that the goods being returned for repair are Canadian goods, the name of the firm who
purchased the goods, and the shipment date.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after
leaving our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect,
abuse or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but
shall not be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications
shall be construed to include improper or inadequate protection against shock, vibration, high or low temperature,
overpressure, excess voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible
medium, unless the equipment is specifically designed for such service, or exposure to any other service or environ-
ment of greater severity than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original
purchaser from us. Should the Buyer’s technical staff require the on-site assistance of AMETEK’s agents or employ-
ees for service calls covered by this warranty clause, the Buyer shall pay travel time plus actual travel and living
expenses.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANT-
ABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LI-
ABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER
PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES,
CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL
PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR
OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE.
WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTIES
OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT, WHETHER
SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT
AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CON-
NECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.
This chapter discusses the features and summarizes the operation of the
Model 9900RM (Rack Mount) Analyzer. It also provides an overview of the
sample system, electronics, and other analyzer sub-systems.
Theory of Operation
The Model 9900RM Analyzer measures the absorbance of ultraviolet radia-
tion (light) by a gas or liquid species. In the standard motorless filter (single-
species) configuration the analyzer measures at two discrete wavelengths,
while in the optional motorized filter wheel (multi-species or multi-range)
configuration it uses a maximum of six discrete wavelengths. In gases, the
concentrations of the species absorbing the light are determined through
application of the ideal gas law in concert with the laws of Bouguer, Beer,
and Lambert law, extended by using a system of linear equations. The
measurements can be viewed from the Human-Machine Interface (HMI),
on a computer via a serial or Ethernet connection, or can be tied to ana-
log output channels for viewing on a remote Distributed Control System
(DCS), Data Acquisition System (DAS), or Programmable Logic Controller
(PLC).
The Beer-Lambert law, in combination with the ideal gas situation, states
that:
Overview | 1-1
The Implementation
The analyzer is comprised of two ultraviolet light sources, a Filterblock
that houses two (motorless filter, or “single-species”, version) to six (mo-
torized filter wheel – multi-species or multi-range version) wavelength
selective filters, a Beamsplitter, front surfaced mirrors, a Measuring Cell,
and two matched Photodetectors (Figure 1-1).
Figure 1-1.
Analyzer schematic.
The Bench Controller board also handles customer inputs and outputs,
which includes the optional touchscreen color Human-Machine Interface
(HMI), processing of detector data to generate customer outputs, conver-
sion of auxiliary data input signals, and communication with the DCS,
DAS, or PLC. See Figure 1-1.
The voltage outputs are selectable to either 0–5 V or 1–5 V. Each of the
contact inputs can be configured to initiate an action (e.g., Zero-Calibra-
tion or Span-Calibration) or to indicate a sample system state (e.g., on
sample stream or on calibration gas). The status relay can be set up to be
triggered on a combination of analyzer detectable Fault or Warning condi-
tion, calibration in progress state, or bench setup in progress.
Overview | 1-3
Optical Bench
Model 9900 Analyzer (Multi-Species/Motorized Filter Wheel Version)
In this configuration up to six filters are used in a rotating filter wheel
(Figure 1-1). To achieve measurements at discrete wavelengths, the lo-
cations of the filters in the filter wheel and the placement of the source
lamps are chosen so that, at any given time, the Photodetectors are ex-
posed to light from only one lamp and filter combination (i.e., a discrete
wavelength). With two source lamps and six filters, twelve discrete wave-
lengths can, theoretically, be produced. In practice, up to six are used.
When absorbers are present in the sample gas, the amount of light re-
ceived by the Measure Photodetector is reduced further. The magnitude
of the reduction is dependent upon the concentrations and absorbance
constants of the absorbers. Thus, for each wavelength, the difference in
the amounts of light received by the two Photodetectors (corrected for the
optical-path factors) is used to determine the absorbance of the sample
gas. Then, the concentrations of the individual absorbers of interest are
calculated.
Zero gas and Span gas also come in through the Sample Inlet, switched
from sample gas by manual or automated valves. The sample system can
also contain pressure sensors/indicators, additional sensors (O2, Cell Pres-
sure, Flow Rate), a particulate filter, sample dryer, and other application-
dependent components. Information about optional components that
make up the analyzer or its sample system are included in Appendix A
– Options, or in the Supplemental Information section of this manual, in the
form of Manual Supplements.
Sample Flow
The sample gas is drawn from the sample stream through a Sample
Probe (by AMETEK or by customer) and through the Sample Line to the
Measuring Cell where it is analyzed and returned through the Vent Line
to the dispersal location, either the sample stream, a low pressure flare
header, or a vent pipe to atmosphere.
CUSTOMER ANALYZER
Figure 1-2.
Analyzer plumbing
diagram.
Overview | 1-5
Analyzer Electronics and Sample System Features
The Model 9900 Analyzer has been designed to comply with electrical
classification requirements for General Purpose and Division 2/Zone 2
areas specific to this analyzer’s intended application. Depending on the
application, some of these features may not be available for your analyzer.
Status Relay
One dry contact (normally open) status relays. The relay is configu-
rable to indicate any of the following: a Fault or Warning condition,
calibration in progress state, or bench setup in progress..
Contact Inputs
Up to two contact inputs. The inputs are configurable to trigger one of
these actions: Zero-Calibration, Span-Calibration, Bench Auto-Setup,
Track-Hold, or an on- oroff-process stream.
ModbusTCP-over-Ethernet:
One RJ-45 Ethernet port is for 10/100 Base-T (IEEE802.3) network con-
nection, using TCP/IP protocol, which allows for remote communica-
tion between the analyzer and the customer DCS, DAS, or PLC. It
supports Modbus TCP protocol.
Advanced Diagnostics
Preventive maintenance and diagnostics are made easy with improved:
• Lamp setup and optimization.
• Lamp, optical system, and I/O diagnostics available via the HMI.
• LEDs on the circuit boards to indicate the status of key system func-
tions such as relay status and heater duty cycles. Information about
the LEDs on the Bench Controller board and Customer I/O board,
and their operation descriptions and current states, are included
with each board’s layout drawing in Appendix C.
• Test points on boards to facilitate functional operation checks.
Information about Bench Controller board test points, their pa-
rameters, and their values are included with the Bench Controller
board’s layout drawing in Appendix C.
Optional Sensors
Support for external sensors for measuringO2 and sample flow.
Overview | 1-7
Working in This Manual
While working in this manual, icons in the margins represent various
kinds of information that serve as reminders or extra information about
the topic.
Example:
Human-Machine Interface (HMI) Navigation icon:
Indicates the menu structure to follow to view a menu, sub-menu, or analyzer
information while working from the HMI. In the example at left, the Diagnostic
DIAGUV Bench Status
menu is first displayed, followed by the UV Bench Status diagnostic screen. From
Status Code1< > this screen, press the button to view the Status Code <1, 2, or 3> screen.
Pressing the button will display the second of two Status Code screens.
Example:
Refer to “Replacing the
Reminder icon:
These reminders indicate related information about the topic, or certain actions
Source Lamps” in Chapter 6 that are necessary before continuing with the current procedure. See example at
for information about replacing left.
the lamps.
Overview | 1-9
This page intentionally left blank.
The Standard Range is defined as the full-scale range that falls within the
minimum and maximum allowable for a given Measuring Cell. The speci-
fications apply only to these ranges, are typical, and in some cases are
application-specific. Calculations of minimum and maximum are based on
760 mmHg atmospheric pressure.
IMPORTANT
The specifications listed in this chapter apply to both the Single-Spe-
NOTE
cies (Motorless Filter) and Multiple-Species (Motorized Filter Wheel)
versions of the Model 9900 Analyzer, except where noted.
Specifications | 2-1
Methodology
Multiple wavelength, high resolution, non-dispersive ultraviolet.
Optional O2 Sensor:
Hummingbird (Servomex) Paramagnetic Pm1158 (CSA-approved for inte-
grating into the G.P Div 2/Zone 2 Analyzers package only).
Standard Ranges
The Standard or Customer Range is defined as the full-scale ranges that
fall within the minimum and maximum allowable for a given Measuring
Cell. The specifications apply only to these ranges, which are typical and
in some cases application-specific. Calculations of minimum and maxi-
mum are based on 760 mmHg atmospheric pressure.
Number of UV Species
Single-Species Analyzer or Multi-Range (Motorless Filter): 1
Response Time
Response time excludes the external sample system.
90 % in 30 seconds or less
Accuracy
Accuracy is determined by comparing analyzer response to a known stan-
dard gas after a calibration has been performed.
SO2, H2S, NO, NO2, NOx: Better than 1.0 % full-scale of standard ranges.
O2: ± 0.1 % O2
Repeatability
SO2, H2S, NO, NO2, NOx: ± 0.5 % full-scale of standard ranges.
O2: ± 0.02 % O2
Linearity
Per ANSI/ISA 51.1–1979 (R1993):
SO2: < ± 1.0 % full-scale of reading
H2S, NO, NO2, NOx: < ± 1.5 % full-scale of reading
O2: ± 0.1 % O2
Stability
Noise (applicable to SO2, H2S, NO, NO2, NOx measurement) at constant
ambient temperature:
Specifications | 2-3
Cross-Talk
Multi-Species (Motorized Filter Wheel) Analyzer Only
± 1.0 % NO2 full-scale on to SO2.
± 1.0 % SO2 full-scale on to NOx.
Temperature Drift
Specification based on using a temperature-controlled Optical Bench.
• One (standard) RS-232 Serial port, isolated (used for Optical Bench
firmware upgrade).
Customer Outputs
• Four Analog Outputs, isolated and configurable individually as
0–20 mADC, 4–20 mADC, 0–5 VDC, or 1–5 VDC.
Analog and alarm relay contact outputs are available on the Cus-
tomer I/O board.
NOTE
Customer Inputs
Two dry contact inputs, isolated as a group:
Voltage: 5 VDC, unregulated
Current: 0.5 mADC approximately
Specifications | 2-5
Electrical Requirements
Supply Voltage
Power Consumption
Typically < 200 W maximum start-up (from a cold start) with average
power depending on ambient temperature.
300 W maximum.
Air background
0.0–10.0 L/minute (0.0–21.2 SCFH) ±10 %Accuracy
Zero Fluid
Others fluids may be used, depending on the application.
NOTE
Gas Applications
UHP N2 or Instrument Air for non-flammable streams.
Instrument Air Quality:
As per ANSI/ISA S7.0.01 (1996) Quality Standard for Instrument Air.
Liquid Applications
Distilled water of at least pharmaceutical grade.
Storage Conditions
-20 °C to +60 °C (-4 °F to +140 °F), limited by Pro-face HMI unit.
Operating Conditions
Humidity
Specifications | 2-7
Maximum Altitude
2000 m
Ingress Protection
The enclosure provides ingress protection IP20 per IEC 60950. The cus-
tomer is required to provide additional ingress protection to IP54 when
installing in Division 2 or Zone 2 classified hazardous areas.
Physical Dimensions
Height Enclosure only: 173.99 mm (6.9")
Front Panel: 177.00 mm (7.0")
Width Enclosure only: 447.48 mm (17.6")
Front Panel: 482.00 mm (19.0")
Depth Enclosure only: 559.56 mm (22.0")
Enclosure with Handles: 602.56 mm (23.7")
Weight 18.2 kg (40 lb) maximum, with all options. (Weight
may vary slightly, depending on options included.)
Electrical Classifications
The Model 9900RM Analyzer is certified with regard to electrical safety for
Pollution Degree 2 (normally nonconductive environmental pollution
occurs with occasional condensation) and Installation Category II (local
level transients, less than those found at power distribution level). Mains
supply voltage fluctuations are not to exceed 10 % of the nominal supply
voltage.
Other approvals and certifications for the standard General Purpose ana-
lyzer include:
CEC/NEC: Class I Division 2 Group ABCD, T4
Ex nA IIC T4
Class I Zone 2 AEx nA IIC T4
IECEx: Certificate No.: IECEx CSA 15.0011X
Ex nA IIC T4 Gc
ATEX: Certificate No.: DEKRA 15ATEX0118 X
II 3G Ex nA IIC T4 Gc
EMC: IEC/EN 61326
FCC (47 CFR 15)
Industry Canada (ICES-0003)
Additional: CE Marking: LVD – 2014/35/EU
RoHS: Restriction of Hazardous Substances Directive:
2011/65/EU (RoHS 2)
Certificate of Compliance
Certificate: 2568583 Master Contract: 164369
Jay McVeigh
Issued by: Jay McVeigh
PRODUCTS
CLASS 2258 02 - PROCESS CONTROL EQUIPMENT - For Hazardous Locations
CLASS 2258 82 - PROCESS CONTROL EQUIPMENT - For Hazardous Locations -
Certified to US Standards
Model 9900RM (Rack Mount) Analyzer; Rated 120/240Vac, 50/60Hz, 300W max
Notes:
Specifications | 2-9
Certificate: 2568583 Master Contract: 164369
• 9900RM shall be housed in an enclosure with a minimum IP54 rating or location providing equivalent
protection.
• Product bearing the Main Board 301-2681 Revision D and Assembled Board 100-2380 Revision E have been
successfully evaluated to Standards 142 and 508 only
• Product bearing the Main Board 301-2681 Revision E and Assembled Board 100-2380 Revision F have been
successfully evaluated to Standard 61010
APPLICABLE REQUIREMENTS
CAN/CSA- C22.2 No 0-10 - General Requirements � Canadian Electrical Code Part II.
CAN/CSA-C22.2 No 0.4-04 - Bonding of Electrical Equipment
C22.2 NO. 142-M1987 - Process Control Equipment
C22.2 NO. 213-M1987 - Non Incendive Electrical Equipment for Use in Class 1 Division 2 Hazardous
Locations Industrial Products
CAN/CSA-C22.2 NO. 61010-1-12 - Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use - Part 1: General Requirements
UL 60079-0’ 5th Edition - Electrical apparatus for explosive gas atmospheres - Part 0: General Requirements
UL 60079-15: 4th Ed. - Explosive atmospheres - Part 15: Equipment protection by type of protection ‘n”
Certificate of Compliance
Certificate: 2554141 Master Contract: 164369
Alan McLaughlin
Issued by: Alan McLaughlin, P.Eng.
PRODUCTS
CLASS 2252 06 - PROCESS CONTROL EQUIPMENT
CLASS 2252 86 - PROCESS CONTROL EQUIPMENT - Certified to US Standards
Process Spectrometer, Model 9900 RM, Rack mount Rated 120V, 50/60Hz, 240V, 50/60Hz 300W max, Type 1,
IP20.
Notes:
• The above model is cord or permanently connected, Equipment Class I, Pollution Degree 2, Installation
Category II.
• Mode of operation: Continuous
• Environmental Conditions: 0°C to 50°C, 2000 m max, 95% rH;
Specifications | 2-11
Certificate: 2554141 Master Contract: 164369
APPLICABLE REQUIREMENTS
CSA Standards:
CAN/CSA-C22.2 No. 61010-1-12 - Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
UL Standards:
UL Std. No. 61010-1 (3rd Edition) - Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use - Part 1: General Requirements
Specifications | 2-13
2-14 | Model 9900RM Analyzer
IECEx Certificate
IECEx Certificate
of Conformity
INTERNATIONAL ELECTROTECHNICAL COMMISSION
IEC Certification Scheme for Explosive Atmospheres
for rules and details of the IECEx Scheme visit www.iecex.com
Type of Protection: Ex nA
Marking:
Ex nA IIC T4 Gc
120/240Vac, 50/60Hz, 300W max.
Tamb: -20°C to +50°C or 0°C to +50°C *
Note:
* - Tamb = 0°C to +50°C is only for analyzer with HMI Display or Oxygen Sensor.
Signature:
(for printed version)
Date:
Specifications | 2-15
IECEx Certificate
of Conformity
Certificate No: IECEx CSA 15.0011X Issue No: 0
CSA Group
178 Rexdale Boulevard
Toronto, Ontario M9W IR3
Canada
and
1707 - 94th Street
Edmonton, AB T6N 1E6
and
8503 East Pleasant Valley Road,
Independence, Ohio, USA
44131-5516
Canada
Additional Manufacturing
location(s):
This certificate is issued as verification that a sample(s), representative of production, was assessed and tested and found to comply with the
IEC Standard list below and that the manufacturer's quality system, relating to the Ex products covered by this certificate, was assessed and
found to comply with the IECEx Quality system requirements. This certificate is granted subject to the conditions as set out in IECEx
Scheme Rules, IECEx 02 and Operational Documents as amended.
STANDARDS:
The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents, was
found to comply with the following standards:
This Certificate does not indicate compliance with electrical safety and performance requirements other than those expressly included in the
Test Report:
CA/CSA/ExTR15.0009/00
NL/DEK/QAR12.0037/01
Specifications | 2-17
IECEx Certificate
of Conformity
Certificate No: IECEx CSA 15.0011X Issue No: 0
Schedule
EQUIPMENT:
Equipment and systems covered by this certificate are as follows:
The AMETEK Model 9900 RM (Rack Mount) Analyzer is an UV process spectrometer that is primarily used for measuring
concentrations of gases or liquids in the range from ppm to percent. It is used primarily in the chemical industries. In the standard fixed
filter configuration, the analyzer measures at two discrete wavelengths, while in the optional motorized filter wheel configuration it uses
a maximum of six discrete wavelengths. The product is enclosed in a 19-inch rack-mountable enclosure made of 2mm thick aluminum
with optional an approved third party Human Machine Interface display on the front panel. The analyzer is comprised of two UV light
sources, a Filterblock that houses two (motorless filter, or “single-species”, version) to six (motorized filter wheel – multi-species or
multi-range version) wavelength selective filters, a Beamsplitter, front surfaced mirrors, a Measuring Cell, and two matched
Photodetectors.
The equipment shall be installed in an enclosure that provides a degree of protection not less than IP 54 in accordance with IEC
60079-15 and IEC 60079-0, and accessible only with use of a tool or key.
The equipment shall only be used in an area of not more than pollution degree 2, as defined in IEC 60664-1.
Transient protection shall be provided that is set at a level not exceeding 140% of the peak rated voltage value at the supply terminals
to the equipment.
AMETEK Canada LP
2876 Sunridge Way N.E., Calgary, AB T1Y 7H9 CANADA
Phone: 403-235-8400 Fax: 403-248-3550 301-3069-F
Toll free (in North America): 1-800-661-9198
NOTES:
IF AN O2 SENSOR AND DISPLAY ARE NOT INSTALLED,
TEMPERARURE RATING IS: Ta -20°C TO 50°C.
Specifications | 2-19
This page intentionally left blank.
• How to unpack and inspect the analyzer for damage before installa-
tion.
• How to install the mechanical components and gas lines, and perform
a leak check on all of the fittings and connections before powering up
the analyzer.
Pre-Installation Requirements
Failure to comply with these storage conditions will void your warranty.
Upon receiving the analyzer system, remove it from the shipping con-
tainer and check the exterior of the analyzer for damage. Remove the
top cover from the Analyzer Enclosure and check all of the components
within the enclosure to ensure they are secure and there is no physical
damage.
• Plugs (supplied by AMETEK) for each unused Cable Entry Port (120 V
analyzers have ½" NPT cable entries; 240 V analyzers have M20 cable
entries). Customer/end user required to supply/install cable glands
that meet the requirements of local wiring regulations and electrical
codes.
• A supply of ¼" non-rigid tubing, such as Teflon, for the Sample and
Vent gas lines. Length will vary, depending on the distance between
the analyzer and the sample extraction and dispersal points. Ap-
propriate ferrules for the tubing material must be used. Supplied/in-
stalled by the customer/end user.
All external connections to the analyzer are made through the Rear Panel.
When mounting the analyzer into a rack, allow sufficient room behind the
analyzer for the Sample Line and Vent Line and for the cable or conduit
carrying wiring to the input/output signal terminal strips inside the ana-
lyzer.
When chassis slides are used the Sample and Vent Line tubes, Ethernet
cable, and signal cable(s) must be flexible and long enough to allow re-
moval of the analyzer from the rack.
The AC Power cable is routed into the Analyzer Enclosure through the
Cable Entry Ports at the analyzer Rear Panel and terminated at the AC
Power Distribution/MOV board.
The valve configuration and the flow control associated with the sample,
Zero, and Calibration gases are to be supplied by the customer.
Sample Prerequisites
The following are typical conditions which the sample gas must meet:
• The maximum allowable pressure is the lesser of 275 kPag (40 psig);
35 kPag (5 psig) if the optional Paramagnetic O2 Sensor is installed;
or the maximum pressure rating of any optional pressure measuring
device that is installed.
• The typical sample flow rate should be 0.5–5.0 L/min (1.0–10.0 scfh).
The volume of the sample system will affect the response time of the over-
all system.
The analyzer is supplied with Sample Line Inlet and Vent Line Outlet con-
nection fittings, which are mounted on the analyzer Rear Panel (Figure
3-3). Unless otherwise specified, the sample system fittings are ¼" Swa-
gelok® connectors.
Install the Zero gas line at the same time as you make the Sample and
Vent Line connections to the analyzer.
1. Route the Sample Line from the analyzer to the gas sample probe on
the stack.
Route the Vent Line (if used) from the analyzer to the gas dispersal
location (vent to atmosphere).
Ensure there are no loops, sags, or other traps in the Sample and Vent
Lines. Provide support where needed.
2. Make the connection for the Sample Line at the analyzer and at the
gas sample probe.
Make the connection for the Vent Line at the analyzer and at the dis-
persal location (vent pipe to atmosphere).
The connector required for connecting the Zero gas to the analyzer is
a 1/8" Swagelok® tube fitting.
NOTE
Install the Zero gas line at the same time as you make the Sample and
Vent Line connections to the analyzer.
• The Zero gas must be dry and must not contain any components
which absorb ultraviolet light (radiation) at the wavelengths em-
ployed by the analyzer, nor contain any components which might
react with components in the sample gas mixture. Nitrogen (UHP) or
Instrument Air (as per ANSI/ISA S7.0.01 (1996) Quality Standard for
Instrument Air) that meets these requirements are suitable for use as a
Zero gas for the analyzer.
• A supply of 1/8" non-rigid tubing, such as Teflon, for the Zero gas line.
Length will vary, depending on the distance between the analyzer
and the Zero gas cylinder. Appropriate ferrules for the tubing material
must be used. Supplied/installed by the customer/end user.
• Adjust the Zero gas pressure to <5 PSIG at 2 SCFH prior to initiating
an Auto-Zero.
1. Route the Zero gas line from the analyzer to the Zero gas cylinder.
2. Connect the Zero gas line to the Zero gas cylinder and to the Zero gas
inlet on the Rear Panel of the analyzer (see Figure 3-3).
If using the USB connection, confirm the USB Flash Drive or cable
is secured in place (Figure 3-4).
The RS-485 terminals on the Customer I/O board can be used for remote
communication with a Distributed Control System (DCS), Data Acquisi-
tion System (DAS), or Programmable Logic Controller (PLC) using Modi-
con Modbus® RTU protocol. The Modbus® RTU protocol can be used with
the analyzer operating as a slave device, responding to Modbus queries
within one second, and providing full access to all registers.
When used via the analyzer HMI, this option provides read access to data
and status registers. The system is limited to calibration command write.
When used via the optional Modbus gateway, this option provides full
access to the registers, and the ModbusRTU-over-RS485 becomes unavail-
able.
Connect the Ethernet cable between the RJ-45 Ethernet port on the back
of the analyzer (see Figures 3-3) and the network.
The USB Flash Drive comes inserted in the analyzer HMI’s USB port on
the bottom of the HMI (inside the analyzer enclosure). This procedure
describes how to safely remove and replace the USB Flash Drive when
upgrading the HMI-Display firmware, or in the event of a USB Flash
Drive failure. Otherwise, removing the USB drive is not recommended.
To remove/install a USB Flash Drive in the USB port on the HMI (use of
Locking Clip required for Division 2/Zone 2) – Figures 3-4.1 and 3-4.2:
1. Power down the analyzer and then remove the screws that secure the
top cover on the analyzer electronics enclosure.
2. Press the yellow “PRESS TO OPEN” button on the top of the HMI to
disengage the Communication module from its HMI Display. Careful-
ly back out the Communication module from its HMI Display, taking
care not to drop it or pull the wiring harnesses from the module.
3. Turn the Communication module slightly to allow access to the USB
Drive (and its Locking Clip – Division 2/Zone 2). (Division 2/Zone 2:
remove the Locking Clip.) Pull the USB Drive out of its port. Note the
orientation of the text on the USB Drive.
4. Obtain the new USB Drive (note the orientation) and insert it in the
USB port. (Division 2/Zone 2: Replace the Locking Clip.)
Figure 3-4.1.
USB Port location, inside
front cover of analyzer.
Figure 3-4.2.
USB Locking Clip.
1. Remove the Rear Access Plate to access the Customer I/O board.
Route the wiring through one of the Cable Entry Ports and then to the
Customer I/O board.
Connect the conduit or cable gland to the Cable Entry Ports, and then
terminate the wiring at the Customer I/O board using the supplied
connectors (blank connectors inserted in the Customer I/O board).
TB-A
ANALYZER CALIBRATION STATUS
TB3
DRY CONTACT INPUTS
TB2
ANALOG OUTPUTS
HMI
TB1
RS-485 COMMUNICATION
TB7
RELAY CONTACTS
TB6
DRY CONTACT INPUTS
Figure 3-6.
Customer Connections.
Connect the supplied power cord to the Power Entry Switch on the ana-
lyzer Rear Panel and to an appropriate power source.
Division 2/Zone 2 analyzers have a Gland Plate on the analyzer Rear Panel
(see Figure 3-3). These analyzers do not have the Power Entry Switch
Plate (including AC Power Switch and AC Plug Connector) on the analyz-
er Rear Panel; therefore, an external power-disconnect switch (supplied
by customer/end user) is required to turn the analyzer power On/Off.
See “Division 2/Zone 2 Analyzers” under “Tools, Equipment and Supplies
Required for Installation” earlier in this chapter for more information.
2. Remove the M20 or ½" plug from the AC Power Gland Plate.
For safety reasons during maintenance, this switch allows the main
AC power to be disconnected from the analyzer prior to performing
service on the analyzer.
OPTICAL BENCH
BOARD
TB-A DIODE
TERMINAL
BLOCK
"TB-A"
PMT SUPPLY
BOARD
PMT SIGNAL
BOARD
Figure 3-7.1.
Analyzer Wiring
Diagram, Sheet 1 of 2.
Figure 3-7.2.
Analyzer Wiring
Diagram, Sheet 2 of 2.
1. Connect Zero Gas to the Sample Inlet fitting and plug the Sample
Outlet.
2. Adjust the Zero Gas pressure within the Measuring Cell approximate-
ly 65 kPa (10 psi) above the expected Measuring Cell operating pres-
sure.
! Do not exceed 100 kPag (15 psig) if the Paramagnetic Oxygen Sensor
is installed.
CAUTION
6. The analyzer leak check is complete. After turning off the Zero Gas
and slowly relieving the pressure, disconnect the Zero Gas and re-
move any plugs or caps installed for the leak check.
! Hazardous Locations
Do not cycle the analyzer power on if there is a hazardous gas atmo-
WARNING
sphere present, unless a power-disconnect switch (approved for the
Hazardous Area) has been installed for use with the analyzer.
Before applying power to the analyzer, test the area around the ana-
lyzer for hazardous gases and proceed only when the area is found to
be safe.
If this screen does not appear within approximately 30 seconds, cycle the
analyzer power Off then On again. If this screen still does not appear,
check the AC power connections to the analyzer, and the 24 VDC Power
Connector on the HMI unit (from the Power Supply). If there is still no
response from the analyzer, contact the local distributor or factory.
Figure 3-9.
HMI-Display,
failure to establish
communications with
the analyzer’s Bench
Controller board.
4. If using the DCS RS-485 port for Modbus® communications, check the
settings:
b. From the home screen press the CONFIG button to view the
The Level2 password
is required to make changes
Configuration sub-menu. Press the UV Bench Modbus button to
view the UV Bench Modbus Communication screen.
to the UV Bench Modbus
Communication screen.
For details about the parameters on this screen, refer to “Viewing and
Changing UV Optical Bench Modbus Communication Parameters”
NOTE
in Chapter 4.
Figure 3-11.
UV Bench
Communication
Configuration screen.
! Since the HMI uses the same Modbus Node Address to communicate
with the UV Bench, changing the UV Bench Modbus Node Address
CAUTION
will necessitate changing the HMI Modbus communication setup
in order to reestablish communication between the HMI and the UV
Bench. Refer to “Changing the Modbus Node Address” in Chapter 4
for information about how to change the HMI Modbus communica-
tion settings.
• To make changes to the DCS Port Baud Rate, DCS Port Par-
ity, or DCS Port Stop Bits settings, tap the drop-down list and
then tap a selection from the list. Press the blank ( ) button
beside the drop-down list to change it to a check mark ( ),
which enables the parameter.
5. Set the HMI Date and Time to local time. For information about how
to change the HMI Date and Time, refer to “Viewing and Changing
HMI Date\Time” in Chapter 4.
8. From the home screen, view the species and parameters that have
been selected for display at the factory.
After the calibration, note the time the analyzer takes to display a
stable concentration value. The response time should be approximate-
ly 30 seconds or less (excluding the external sample system).
Problem Encountered
The normal HMI start-up sequence of the message “Connecting to UV
bench...” appearing on the HMI-Display followed by the home screen ap-
pearing do not appear immediately after applying AC power.
Corrective Action
Take all necessary precautions and check the AC wiring to ensure it is prop-
erly connected.
Problem Encountered
During the normal HMI start-up sequence (“Connecting to UV bench...”
message on the HMI) the error message “UV BENCH NOT CONNECTED
TO HMI UNIT...” appears on the HMI instead of the home screen appear-
ing.
Corrective Action
Take all necessary precautions and ensure the connectors on all cables be-
tween the HMI and the analyzer are properly connected, and check these
cables for damage. Replace the cables if damaged.
Problem Encountered
Flashing DIAG button (RED or ORANGE) on the home screen.
Corrective Action
Check the UV Bench Status diagnostic screen to view current status condi-
Status
DIAGUV Bench tions.
Certain status conditions will be active until the analyzer has warmed up to
normal operating temperature, at which time they should clear automatically.
Refer to “Analyzer Diagnostic Status Codes, Trigger Conditions, and Cor-
rective Action” in Chapter 6 for complete status condition descriptions and
corrective action to take to correct the status conditions.
2. Connect (or turn on) the Zero Gas to the analyzer and perform a
manual Zero.
3. Connect (or turn on) the Span Gas to the analyzer and perform a
manual Span.
IMPORTANT
To help you determine in the future if there are problems with your
NOTE
analyzer, it is important to observe and record initial readings (when
the analyzer sample system is clean) of the PMT (photomultiplier
tube) signals for each filter and the sample response time.
Also record the date the analyzer was commissioned for operation
to help you determine lamp age, which can be helpful if diagnosing
problems with the analyzer. Whenever lamps are replaced, record the
date for future reference.
The data – which the HMI stores for the last 12 hours, at 1-minute inter-
vals – can also be saved to a USB Flash Drive.
Observe and record the signals every month to monitor signal strength.
Make the first recording after commissioning the analyzer for operation.
1. Manually Zero the analyzer. Refer to “Manual Zero and Span Calibra-
tion” in Chapter 5 for information about how to perform a manual
zero.
2. After the Zero is complete, return the gas flow to Sample Gas. View
the home screen and observe the output concentration. Record the
time it takes the analyzer to display the first concentration reading.
• How to work from the touchscreen color HMI and navigate through
the various screens.
Before working from the HMI, read the entire section titled
“Introduction to the Human-Machine Interface (HMI)” to learn
NOTE
how to navigate through the screens to view data, to learn how
to make changes to analyzer settings, and to learn how to initi-
ate analyzer functions.
Sensor Results
Analysis Results
Analyzer Communications
Status Indicators
Figure 4-1.
HMI layout, HOME
screen.
The HOME screen areas listed in Figure 4-1 are described under
“Working From the HOME Screen” later in this chapter.
NOTE
as either Fault or Warning This is an analyzer Status Code indicator: the analyzer is operating
conditions, refer to “Relay with a Fault status (RED = fault).
Output Setup” in this chapter.
Selection is disabled. Press this button to enable an option; the button
will change to to indicate the selection is enabled.
When the button text or symbol is greyed out the button is not available for
the functions on that screen. See examples below.
NOTE
(Example) CONFIG
•
HMI Navigation:
While working from the HMI, use these navigational aids to quickly access
Main Menu screens, Sub-Menu screens, or analyzer information from the
UV AnalysisSample Cell HOME screen. See example at left.
Conditions
• Live values on the screens are updated every two seconds. When
switching to a different screen, it may take up to two seconds for a
new set of values to be displayed.
• If the display is left at a screen other than the home screen for 30 min-
utes without activity the display will switch back to the home screen.
1. Tap the top-left corner of the display screen area, followed by the
top-right corner of the screen (within one second) as illustrated in
Figure 4-2. The brightness/contrast control adjustment bar will ap-
pear at the bottom of the screen. This can be done from any screen.
2. Press the ‘+’ key to increase the brightness level or the ‘-’ key to
decrease the brightness level.
3. After the display has been adjusted to the desired level, press Can-
cel ( ) to close the adjustment bar.
Figure 4-2.
Adjusting the HMI
display brightness level. Brightness / Contrast Control Bar
2. Use the numeric keypad to enter the desired value (including deci-
mal, if required) and then press the ENT key.
(Example) CONFIG
Temperature Control
Numeric field
Figure 4-3.1.
Changing values in a
numeric field, example.
& Unit
Figure 4-3.2.
Changing values in
an alphanumeric field,
example.
1. Tap the drop-down list and then tap a selection from the list.
Drop-down list
(Example) CONFIG
UV Bench Modbus
Figure 4-4.
Choosing a setting
from a drop-down list,
example.
Figure 4-5.
Enter Password keypad.
To change Modbus Node Address that the HMI uses to communicate with
the UV Bench:
1. Tap the top-right corner of the display screen area, followed by the
bottom-left corner of the screen (within one second). The system
menu bar will appear at the bottom of the screen.
7. Tap the Slave Address numeric field and enter the new Modbus Node
Address of the UV Bench.
8. After changing DCS Modbus serial port settings, press the Back but-
ton twice to return to the Peripheral Settings menu.
10. Press the Exit button to exit the offline mode and reboot the HMI unit.
! Hazardous Locations
Do not remove the Analyzer Enclosure cover, do not cycle the ana-
WARNING
lyzer power On or Off, and do not attempt to connect or disconnect
the USB Flash Drive (or cable) to or from the USB port on the HMI if
there is a hazardous gas atmosphere present.
Before connecting a USB Flash Drive (or cable) to the USB port on
the HMI, test the area around the analyzer for hazardous gases and
proceed only when the area is found to be safe.
1. Unzip the Pro-face HMI software programming file onto the root
directory of a USB Flash Drive.
2. With the HMI unit powered down, remove the factory-installed USB
Flash Drive from the HMI’s USB port on the bottom of the HMI and
insert the USB Flash Drive (with the firmware upgrade files) into the
HMI’s USB port.
10. The process of copying from USB Flash Drive to the HMI unit takes
a few minutes. When the process finishes the message “Successfully
Completed” will be displayed.
11. Press the Back button repeatedly to return to the Memory Loader
Tools screen.
12. Press the Exit button to exit the Memory Loader Tools screen.
13. When prompted to reboot the system, press the No button. The HMI
unit will need to be rebooted before operation, but this will be done
upon power up after replacing the factory-installed USB Flash Drive.
! Before continuing with Step 14, power down the analyzer/HMI unit
and then disconnect the HMI’s Communication module from its HMI
WARNING
Display to allow access to the USB Flash Drive.
14. Remove the USB Flash Drive, reinstall the factory-installed USB Flash
Drive, and replace the top cover on the analyzer electronics enclosure.
15. Power up the analyzer/HMI unit. The system will reboot with the new
HMI firmware programming file.
Figure 4-6.
shipped with the analyzer.
A laminated copy of
the S9900_HMI Menu Map is
Restore
Selection
Photometric
PMT Lamp PMT\Motor
Status Trigger HMI NOx
Signals Ops Ops Sensor Setup Factors
Filter
Bench Ops 1 Coefficients
Bench Ops 2 StatusCode1 Linear PSWD
UV Analysis Sensor Trigger Setup Coefficients Setup
Trans\Abs (1/2) Absorbance
Results Signals
Offsets
StatusCode2 StatusCode1 Species
Trigger Setup Trigger Setup Name & Unit
System System Data (1/2) (2/2) Analysis
Info Status Trending Matrix
through the sub-menus of each of the three main menus.
StatusCode2 StatusCode3
Trigger Setup Trigger Setup
Hardware Wavelength Species UV Analysis UV Bench Sensor (2/2) (1/2) Row1 Row5
Presence Valid Valid Results Signals Signals
StatusCode3
S Standard menu screen
Row2 Row6
UV Analysis UV Bench Trigger Setup
Sensor Signal
StatusCode1 StatusCode2 StatusCode3 Results Signal (2/2) S Optional menu screen
Trend Graphs
(1/2) (1/2) (1/2) Trend Graphs Trend Graphs Row3 Row7
Use the Model 9900 Analyzer (HMI) Menu Map to help you navigate
The home screen displays the process concentration values (e.g., SO2 and
O2) and the status of integral analyzer/system information. This screen
also contains the Main Menu buttons that allow you to access other
screens to view analyzer data, configure analyzer settings, and perform
analyzer functions.
Sensor Results
Analysis Results
Analyzer Communications
Status Indicators
Figure 4-7.
HOME screen.
CellP
Displays the real-time pressure (mmHg) of the sample gas in the
Measuring Cell. (Optional pressure transducer must be installed.)
Flow
Displays the real-time flow rate (litres per minute) being measured
by the (optional) Flow Sensor. (Optional Flow Sensor must be
installed.)
O2–%
Displays the real-time concentration (%) being measured by the
O2 Sensor. (Optional O2 Sensor must be installed.)
O2–T
Displays the real-time temperature (°C) being measured by the O2
Sensor. (Optional O2 Sensor must be installed.)
Initiation via dry contact is only available when using the Cus-
CONFIGCustomer
I/OContact Input Setup
tomer I/O board. View the Contact Input Setup screen to view the
input assigned for calibration initiation. Refer to the section “Con-
tact Input Setup” in this chapter for more information about this
screen.
Logging to USB
When this indicator light is BLUE, it means the operator has initi-
ated Data Trend logging from the analyzer to the USB Flash Drive.
Figure 4-8.
Configuration menu.
To access the Password Setup screen, press the PSWD Setup button. See
“Entering Passwords to Change Analyzer Parameter Settings” earlier in
this chapter for more information. Detailed descriptions of each sub-menu
– and the parameters under each – are listed in the following pages.
The UV Bench Setup screen allows you to initiate an Auto Setup, align
the source lamps, and view the results of the Auto Setup, plus view cer-
tain Auto Setup related values.
Setup
CONFIGUV Bench
Figure 4-9.
UV Bench Setup
Configuration sub-
menu.
Lamp1 Base
Lamp2 Base
This is the lamp base current (mA) for Lamps 1 and 2. This is the “sim-
mer” current when the lamp is not being pulsed. AMETEK does not
recommend adjusting this value from the factory default setting.
The Customer I/O setup sub-menu allows you to view and change the
settings of the analyzer’s inputs and outputs. From this screen, press the
button for the inputs or outputs you wish to change and make the chang-
es. Each setup option is discussed in the following pages.
CONFIGCustomer I/O
Figure 4-10.1.
Customer I/O
Configuration sub-
menu.
The Analog Output Setup screen allows you to view and change the ana-
log outputs that are displayed, and modify the output settings.
CONFIGCustomer
I/OAnalog Output Setup
Figure 4-10.2.1.
Analog Output Setup
Configuration screen.
Tap the Assignment field for the analog channel, to display a numeric
keypad. Enter the Assignment Code and press ENT to accept the
change. Repeat this step to configure other output channels.
5. To clamp the output to the selected Range, press the clamp out-
put button to change it to , which indicates it is enabled.
6. After the changes have been made, press to save the new values,
or press to refresh the changes.
The Analog Output Assignment Code sub-menu allows you to view the
codes for all analyzer signals that are available to be assigned to an analog
output channel. These codes are the corresponding Modbus Register Ad-
dress numbers for the analyzer signals.
CONFIGCustomer
I/OAnalog Output Setup
Figure 4-10.2.2.
Analog Output Assignment
Code Configuration
sub-menu.
CONFIGCustomer
I/OAnalog Output Setup
Concentration
Figure 4-10.2.3.
Concentration assignment
codes.
Photometric
This screen displays the assignment codes for Photometric measure-
ments of Absorbance (Abs) and Transmittance (Trans).
CONFIGCustomer
I/OAnalog Output Setup
Photometric
Figure 4-10.2.4.
Photometric assignment
codes.
CONFIGCustomer
I/OAnalog Output Setup
Sensor
Figure 4-10.2.5.
Sensor assignment
codes.
CONFIGCustomer
I/OAnalog Output Setup
Bench Ops:1
Figure 4-10.2.6.
Optical Bench
Operations assignment
codes, screen 1 of 2.
CONFIGCustomer
I/OAnalog Output Setup
Bench Ops:2
Figure 4-10.2.7.
Optical Bench
Operations assignment
codes, screen 2 of 2.
The Relay Output Setup setup screen allows you to view and change the
status indicated by the relay output.
After changes have been made, press to save the new values, or press
to refresh the changes.
CONFIGCustomer
I/ORelay Output Setup
Figure 4-10.3.1.
Relay Output
Configuration screen.
fault status
The relay will be triggered if any system fault status condition is
triggered within the analyzer. The analyzer requires service. The
relay can only be reset by correction of the condition causing the
fault status.
warning status
The relay will be triggered if any warning status condition is trig-
gered within the analyzer. The analyzer requires attention. The
relay will reset automatically upon correction of the condition
causing the warning status.
The different Status Code Trigger Setup screens are shown in the
following pages. After changes have been made, press to save the
new values, or press to refresh the changes.
Each Status Code Trigger Setup screen lists the error conditions de-
tectable by the analyzer. Each detectable error condition can be in-
cluded ( ) or excluded ( ) in triggering the Warning (ORANGE
light) or Fault (RED light) status.
For example, to enable the “PMT data not received” status condition
(Figure 4-14.3.3) to trigger the Fault status, press the (disabled) but-
ton to change it to (enabled) and tap the ORANGE light to change
it to RED.
CONFIGCustomer
I/ORelay Output Setup
Status Trigger Setup
Figure 4-10.3.2.
Status Trigger Setup
sub-menu.
Figure 4-10.3.3.
Status Code 1 Trigger
Setup screen (1 of 2).
CONFIGCustomer
I/ORelay Output Setup
Status Trigger SetupStatus
Code 1 (2/2)
Figure 4-14.3.4.
Status Code 1 Trigger
Setup screen (2 of 2).
CONFIGCustomer
I/ORelay Output Setup
Status Trigger SetupStatus
Code 2 (1/2)
Figure 4-10.3.5.
Status Code 2 Trigger
Setup screen (1 of 2).
Figure 4-10.3.6.
Status Code 2 Trigger
Setup screen (2 of 2).
CONFIGCustomer
I/ORelay Output Setup
Status Trigger SetupStatus
Code 3 (1/2)
Figure 4-10.3.7.
Status Code 3 Trigger
Setup screen (1 of 2).
CONFIGCustomer
I/ORelay Output Setup
Status Trigger SetupStatus
Code 3 (2/2)
Figure 4-10.3.8.
Status Code 3 Trigger
Setup screen (2 of 2).
The Analog Output Calibration setup screen allows you to calibrate the
analog outputs. For details about how to calibrate the analog outputs,
refer to the section “Calibrate the Analog Outputs” in Chapter 5.
CONFIGCustomer
I/OAnalog Output
Calibration
Figure 4-10.4.
Analog Output
Calibration
Configuration screen.
The Contact Input Setup screen displays the action assigned to each con-
tact input. When the contact input closes, the analyzer will perform the
action assigned. After changes have been made, press to save the new
values, or press to refresh the changes.
CONFIGCustomer
I/OContact Input Setup
Figure 4-10.5.
Contact Input Setup
screen.
The Temperature Control setup screen displays the current control set-
tings for each temperature-controlled zone. From this screen, you can
change the settings for each zone. The zones displayed on this screen cor-
respond with the type of optional (installed) temperature sensors, and are
controlled by the analyzer.
If required, make changes to the values and press to save the new
values, or press to refresh the changes.
CONFIGTemperature
Control
Figure 4-11.
Temperature Control
Configuration screen.
Setpoint (C)
This is the control operating setpoint (°C) for each temperature-control
zone. Enter “0” to disable a temperature-control zone.
Kp (C)
This is the proportional gain (°C) for each temperature-control zone.
AMETEK does not recommend adjusting this value from the factory
default setting.
Ti (sec)
This is the integral time (seconds) for each temperature-control zone.
AMETEK does not recommend adjusting this value from the factory
default setting.
2. Change the Setpoint value and press ENT to save the change. Repeat
this step for other temperature-control zones and other parameters on
this screen.
Setup
CONFIGSensor
Figure 4-12.
Sensor Setup
Configuration screen.
O2 Concentration (%)
Displays the O2 Concentration Low Scale and High Scale values.
(Optional O2 Sensor must be installed.)
1. Tap the setpoint field for the value you wish to change. A numeric key-
pad will be displayed.
2. Change the Low Scale and/or High Scale values for Cell Pressure,
Flow Rate, and/or O2 Concentration. After the changes have been
made, press to save the new values, or press to refresh the
changes.
The Track-Hold setup screen displays the Hold Time (sec) for which the
concentration value for each species will be held constant during a zero-/
span-calibration. It retains this value for the duration of the zero-/span-
calibration, at which time it will resume outputting live results. See Figure
4-13.2 for two examples of Track-Hold Timing diagrams.
CONFIGTrack-Hold
Figure 4-13.1.
Track-Hold
Configuration screen.
1. Tap the Hold Time field for the value you wish to change. A numeric
keypad will be displayed.
2. Enter a new value (seconds) and press to save the new values, or
press to refresh the changes.
The UV Analysis setup screen allows you to view and set up Measuring
(Sample) Cell conditions, current algorithm selections, filter coefficient
values, linear coefficient values, absorbance offsets, species name and
unit, and analysis matrix.
CONFIGUV Analysis
Figure 4-14.1.
UV Analysis
Configuration sub-
menu.
! The settings for your analyzer have been configured at the factory to
meet specified customer requirements. Changing the factory-set con-
CAUTION
figuration could cause the analyzer to operate incorrectly.
CONFIGUV
AnalysisSample Cell
Conditions
Figure 4-14.2.
Sample Cell Conditions
Configuration screen.
Compressibility Factor
In addition to pressure and temperature compensations of the concen-
trations according to the ideal gas law, it is possible to enter an ad-
ditional parameter (Z) to account for the compressibility factor of the
sample gas. One value of Z is used for all species. The default value
for this parameter is ‘1’, but can be any value between 0.5 and 1.5.
AMETEK does not recommend adjusting this value from the factory
default setting. The correction is applied to the concentrations in the
following manner:
<Conc>corr. = <Conc>uncorr. / Z
For the Cell Pressure Compensation, change the switch to the dynamic
or static position.
The Algorithm Selection screen is for use by service personnel during di-
agnosis, and can be disregarded by users. Do not alter the selections made
on this screen, unless instructed to do so by AMETEK. The button
indicates the algorithm is enabled.
! The settings for your analyzer have been configured at the factory to
meet specified customer requirements. Changing the factory-set con-
CAUTION
figuration could cause the analyzer to operate incorrectly.
CONFIGUV
AnalysisAlgorithm Selection
Figure 4-14.3.
Algorithm Selection
Configuration screen.
NDR Compensation
Select this to enable the analyzer to perform the Neutral Drift com-
pensation.
Linearization-1
Select this to enable the analyzer to perform for linearization-1 calcula-
tion.
Linearization-2
Select this to enable the analyzer to perform for linearization-2 calcula-
tion.
SO2 Extended-Range
Select this to enable the analyzer to measure extended-range (higher
than typical range) SO2. [This function is available only for analyzers
with a motorized filter wheel.]
SO2 Crosstalk
Select this to enable the analyzer to perform SO2 Crosstalk Correction.
[This function is available only for analyzers with a motorized filter
wheel.]
MeSH Correction
Select this to enable the analyzer to perform methyl mercaptan
(MeSH) measurement corrections. [This function is available only for
analyzers configured to measure MeSH.]
Hot Application
Select this to enable the analyzer to perform additional calculations in
a hot, wet gas sample. [This function is available only for analyzers
with an optional Oven.]
Liquid Application
Select this to enable the analyzer to perform additional calculations for
a liquid sample.
The Filter Coefficients setup screen allows you to view and change the
adaptive filtering parameters, Sigma and Minimum Gain, for Absorbance
Filter and Neutral Drift Filter.
! Do not change the values from the factory default settings unless
advised to do so by AMETEK.
CAUTION
CONFIGUV
AnalysisFilter Coeff.
Figure 4-14.4.
Filter Coefficients
Configuration screen.
! Do not change the values from the factory default settings unless
advised to do so by AMETEK.
CAUTION
CONFIGUV
AnalysisLinear Coeff.
Figure 4-14.5.
Linear Coefficients
Configuration screen.
The Absorbance Offsets setup screen displays the absorbance offset factor
for each lamp\filter combination being used.
! Do not change the values from the factory default settings unless
advised to do so by AMETEK.
CAUTION
CONFIGUV
AnalysisAbsorbance Offsets
Figure 4-14.6.
Absorbance Offsets
Configuration screen.
The Species Name & Unit setup screen displays the species (name) be-
ing measured by the analyzer, the unit of measurement for each species
(PPMV, %, etc.), and the conversion factor (e.g., 1 = PPMV (1 x 104 = %))
for each species.
! Do not change the values from the factory default settings unless
advised to do so by AMETEK.
CAUTION
CONFIGUV
AnalysisSpecies Name &
Unit
Figure 4-14.7.
Species Name & Unit
Configuration screen.
2. Use the keypad to enter/change the information and then press the
ENT key.
CONFIGUV
AnalysisAnalysis Matrix
Figure 4-14.8.1.
Analysis Matrix
Configuration
sub-menu.
Each Matrix screen displays the value for each lamp\filter combination, for
that species, in the calculation matrix. To view information for other spe-
cies, return to the Analysis Matrix sub-menu and select another species
(row).
! Do not change the values from the factory default settings unless
advised to do so by AMETEK.
CAUTION
CONFIGUV
AnalysisAnalysis Matrix
Row 1<species>
Figure 4-14.8.2.
Matrix Row 1 <species>
Configuration screen.
For more information about changing the node address, refer to “Chang-
ing the Modbus Node Address” in this chapter and to Step 4 under “Pow-
ering Up the Analyzer” in Chapter 3.
CONFIGUV Bench
Modbus
Figure 4-15.
UV Bench Modbus
Communication
Configuration screen.
! The HMI uses the same Modbus Node Address to communicate with
the analyzer. If the Node Address is changed, change the HMI system
CAUTION
setting to use the new Node Address.
The HMI Date\Time screen allows you to view and change the HMI date
Level2 password is
required to make changes to
and time which is necessary for time-stamping the analyzer data log.
the HMI Date\Time screen.
The date and time must be changed to that of the installation location on
startup; otherwise, the date/time of the factory location will be in effect.
Time
CONFIGHMI Date\
Figure 4-16.
HMI Date\Time
Configuration screen.
CONFIGSave\
Restore
Figure 4-17.1.
Save\Restore
Configuration screen.
Use the SAVE to NV memory button to save the current analyzer configu-
ration settings in nonvolatile memory that can later be used to restore the
settings. When you press this button, the analyzer immediately begins
backing up its current configuration settings. The message “Saving...” ap-
pears on the screen until the file saving process is complete.
Figure 4-17.2.
Saving configuration
settings screen.
Figure 4-17.3.
Restoring configuration
settings screen.
Calibrating the analyzer can be done directly from the Distributed Con-
trol System (DCS), Data Acquisition System (DAS), or Programmable
Logic Controller (PLC), or using a combination of DCS, DAS, or PLC and
Human-Machine Interface (HMI) settings. This chapter discusses how to
manually Zero- or Span-calibrate, and how to configure and enable the
Auto Zero function, from the analyzer HMI. This chapter also discusses
how to view the results of Zero and Span calibrations. Finally, this chapter
discusses how to calibrate the analog outputs.
This manual does not discuss how to enable the appropriate flow control
settings using the DCS, DAS, or PLC system, prior to performing the cali-
bration procedures.
Calibration | 5-1
Zero and Span Calibration
From the Calibration menu, you can access various screens to configure
Zero and Span calibration parameters, initiate Manual Zero and Span cali-
brations, set up and enable the optional Auto-Zero calibration, and view
the Zero and Span calibration results.
The Model 9900RM Analyzer can support Manual calibration (typical ap-
plications) and Auto-Zero calibrations (optional). For Manual calibrations,
the Zero or Span calibration gas must be introduced manually into the
Measuring Cell. For Auto-Zero calibrations, an optional solenoid is inte-
grated within the analyzer sample system, and the Auto-Zero function
can be configured to control the solenoid, which will automatically intro-
duce Zero gas through the analyzer sample system and its Measuring Cell
according to predetermined settings.
Figure 5-1.
Calibration menu.
The CAL Averaging Duration (along with the Zero Gas Flushing
timer) is part of the overall Countdown timer seen on the calibration
screens.
Using the DCS, DAS, or PLC system, ensure the sample gas flow
control is set to Zero flow, and that the appropriate zero gas has been
NOTE
introduced into the analyzer sample system.
1. Disconnect the Sample Gas line from the Sample Gas Inlet port on the
back of the analyzer and connect the Zero gas tube to the Inlet port
(see Figure 5-4).
CAL
CALManual Zero-
Figure 5-2.1.
Manual Zero-
Calibration screen.
Calibration | 5-3
3. When the live concentration values that are displayed are stable and
near zero, press to start the Zero-Calibration. The Countdown
timer will begin to count down from the value of its setting. The Zero-
Calibration is automatically adjusted based on the average reading dur-
ing the CAL Averaging Duration timer period. The adjustment applies
to the option(s) selected.
4. After the Zero-Calibration is complete you can view the results. Press
to return to the Calibration menu and then press the ZERO-CAL
Results button to view the ZERO-CAL RESULTS screen. This screen
displays the results for the Zero-Calibration. This screen displays the
Zero Transmittance results for each filter and the results of the species
being measured by the optional, installed sensor (e.g., O2 sensor).
CALZero-CAL
Results
Figure 5-2.2.
Manual Zero-Calibration
Results screen.
Zero Transmittance
Displays the recorded PMT signal transmittance value for each
lamp/filter combination, with Zero Gas in the sample system. It is
adjusted automatically whenever the analyzer is zero-calibrated.
Record this value after the Zero for long-term diagnostic purposes.
CALZero-CAL
ResultsZero-CAL Offsets
Figure 5-2.3.
Manual Zero-Calibration
Offsets screen.
5. Bleed down the Zero gas and disconnect it from the analyzer.
Reconnect and tighten the Sample Gas line to the Sample Gas Inlet
port on the back of the analyzer.
Calibration | 5-5
Span-Calibrate the Analyzer and View the Results
Using the DCS, DAS, or PLC system, ensure the sample gas flow
control is set to Span Flow, and that the appropriate Span gas has
NOTE
been introduced into the analyzer sample system.
1. Disconnect the Sample Gas line from the Sample Gas Inlet port on the
back of the analyzer and connect the Span gas tube to the Inlet port
(see Figure 5-4).
CAL
CALManual Span-
Figure 5-3.1.
Manual Span-
Calibration screen.
3. In the field for the species that will be spanned, enter the Span Gas
Value and then press the ENT button. Press to send the new Span
Gas Value to the Bench.
Allow Span gas to flow and adjust the Span gas flow rate.
5. After the Span-Calibration is complete you can view the results. Press
to return to the Calibration menu and then press the Span-CAL
Results button to view the SPAN-CAL RESULTS screen. This screen
displays the Span Factor results of each UV species and the results of
the species being measured by the optional, installed sensor (e.g., O2
sensor).
CALSpan-CAL
Results
Figure 5-3.2.
Manual Span-Calibration
Results screen.
Calibration | 5-7
Span-CAL Offsets
Press this button to view the SPAN-CAL OFFSETS screen, which
displays the Span-Calibration offset of each species. This is the
difference between the measured concentration and the calibration
concentration, before the corresponding Span Factor is adjusted at
the end of a Span-Calibration. The Span-CAL Time displays the
Date and Time at which the Span-Calibration was performed for
each species.
CALSpan-CAL
ResultsSpan-CAL Offsets
Figure 5-3.3.
Manual Span-Calibration
Offsets screen.
6. Bleed down the Span gas and disconnect it from the analyzer.
Reconnect and tighten the Sample gas line to the Sample Gas Inlet
port on the back of the analyzer.
Calibration | 5-9
Setting Up Auto-Zero Parameters and Viewing Live
Results – Optional
Before initiating an Auto-Zero, ensure the Zero gas pressure has been
set to <5 PSIG @ 2 SCFH.
NOTE
CALAuto-CAL
Figure 5-5.1.
Auto-Calibration sub-
menu, Auto-Zero idle.
Auto Zero is idle and the Zero Solenoid Control is under automatic
analyzer control:
In this state, the Auto Zero, Auto Zero Setup, and Auto Zero
Results buttons are not greyed out (white text), and the Zero
Solenoid Control Auto button is GREEN. Under these conditions,
the Auto-Zero can be started manually by pressing this button.
The ZSolCtrl status indicator on the HOME screen will also be on
(GREEN).
CALAuto-CALAuto
Zero Setup
Figure 5-5.2.
Auto-Zero Setup
screen.
Calibration | 5-11
3. To set up the Auto-Zero parameters, press the Auto Zero Setup button
to view the AUTO ZERO SETUP screen.
If the interval entered is not divisible into 24, the interval is round-
ed up to the nearest interval divisible into 24. For example, ‘5’ will
be rounded up to ‘6’, ‘7’ to ‘8’, ‘10’ to ‘12’.
This value is also used as the duration of which the system will
flow Sample gas, as part of the Auto-Zero sequence, prior to the
analyzer resuming normal sampling and operation.
The Zero Gas Flushing timer (along with the CAL Averaging
Duration timer) is part of the overall Countdown timer seen on the
calibration screens.
This parameter is used for the Auto-Zero sequence and for manu-
al Auto-Zeroes.
a. Tap the First Auto Zero at hour numeric field to display a numeric
keypad, enter the hour value, and tap the ENT key to save the
change. Repeat this for the minute numeric field. The hour and
minute must be entered separately.
c. Tap the Zero Gas Flushing numeric field display a numeric key-
pad. Enter the number of seconds (minimum 60) the system will
flush Zero gas through the sample system before signal averaging
is performed. Tap the ENT key to save the change.
Calibration | 5-13
4. If required, a manual Auto Zero can be performed. From the AUTO-
CAL sub-menu, ensure the conditions are met to allow a manual Auto-
Zero to be started, and press the Auto Zero button to view the Auto-
Zero screen. From this screen, press the button to manually initiate
an Auto-Zero.
Zero
CALAuto-CALAuto
Figure 5-5.3.
Auto-Zero screen.
Auto Zero Progress Indicator Lamp
Species results
Live results (in ppm) are displayed in the left portion of this
screen, including O2 results if an optional O2 Sensor has been
installed.
Current Time
Current analyzer clock time.
Time To Next
Amount of time until the next Auto-Zero will begin.
Next Auto-Zero
Scheduled clock time of the next Auto-Zero.
For example, the Zero Gas Flushing timer is set at “60” seconds
and the CAL Averaging Duration timer is set at “45” seconds. The
Auto-Zero sequence will be:
b. The actual Zero (or averaging) will occur for 45 seconds (CAL
Averaging Duration timer setting = 45).
GO and X buttons
Press the button to manually initiate an Auto-Zero. When an
Auto-Zero is in progress, either manually initiated or scheduled
(auto), the button can be used to abort the Auto-Zero. If these
buttons are not displayed, it indicates an Auto-Zero cannot be ini-
tiated manually because the Zero solenoid is not under Automatic
analyzer control or an Auto-Zero is already in progress.
Calibration | 5-15
Viewing the Auto Zero Offset (Drift) Results
After the Auto-Zero is complete, you can view the Zero Offset (Drift)
results. To view the Auto Zero Offset results, press the Auto Zero Results
button from the Auto-Calibration sub-menu to view the AUTO ZERO
RESULTS screen.
CALAuto-CALAuto
Zero Results
Figure 5-5.4.
Auto Zero Results
(offsets) screen.
CONFIGCustomer
I/OAnalog Output
Calibration
Figure 5-6.
Analog Output
Calibration screen.
When entering input ranges, the values must be within the 0–20 mA
and 0–5 V ranges for the respective current and voltage outputs.
NOTE
Calibration | 5-17
3. Enable ( ) the 25%-span check operation and then press to
send the selected operation to the analyzer.
Use a volt meter to measure the voltage of each analog output channel
to verify that it is measuring 25 % of full-scale (e.g., for a 1–5 V output
it is measured to be 2 V).
Use a volt meter to measure the voltage of each analog output channel
to verify that it is measuring 50 % of full-scale (e.g., for a 1–5 V output
it is measured to be 3 V).
Use a volt meter to measure the voltage of each analog output channel
to verify that it is measuring 75 % of full-scale (e.g., for a 1–5 V output
it is measured to be 4 V).
Safety Considerations
Before performing any maintenance, service, or troubleshooting on the
analyzer, review and follow all safety information under “Personnel
NOTE
and Equipment Safety Information” following the Table of Contents
near the beginning of this manual. This information describes proce-
dures to follow to avoid personal injury and/or damage to the equip-
ment. All regulatory agency and personnel safety procedures for
your jurisdiction must be followed.
Preventive Maintenance
To reduce the occurrence of problems with the analyzer, AMETEK recom-
mends that you follow the Analyzer Preventive Maintenance Schedule.
Since most analyzer problems originate within the sample system, the
primary objective of this schedule is proper care of the sample system.
Frequency Task
Daily Check for Fault or Warning Status Conditions
The DIAG button on the HOME screen is the first indication from the
Human-Machine Interface (HMI) that there are active status condi-
tions. If status conditions are triggered, this button begins flashing and
Status
DIAGUV Bench
changes color (ORANGE=Warning status; RED=FAULT status). If so,
view the UV Bench Status diagnostic screen.
From other screens, the button (with ORANGE or RED indicator)
also serves to indicate there are active status conditions.
Review “Troubleshooting and Diagnostics” in this chapter for information
about the status conditions and the suggested corrective actions to take
to clear the conditions.
Before working on the sample system, manually zero the sample system,
isolate it (blocked it in) from the sample stream, and disconnect analyz-
er power. This procedure is necessary prior to changing out any ana-
lyzer components or replacing any replaceable parts (as part of regular
preventive maintenance), or when performing leaks checks following the
replacement of the Zero Gas supply or other adjustments to any of the
connection points in the analyzer’s sample system.
If the sample response time is longer than normal, it is most likely due to
plugging somewhere in the sample system. The portion of the sample sys-
tem with the highest risk of plugging is the sample inlet and the section
of the sample inlet most prone to plugging is at the sample gas extraction
point. One way to confirm that a plug is in the sample inlet is to close the
Vent Valve’s isolation valve and Zero the analyzer.
After confirming that the plug is somewhere in the sample inlet, take
all appropriate safety precautions and dismantle the sample system,
NOTE
starting at the sample gas extraction point.
Plugs in the sample outlet (or vent side) of the sample system are rare.
Plugs in the vent system cause the aspirator drive air to flow back into the
Measuring Cell and, essentially, simulate a Zero Gas Flow or Zero condi-
tion.
The o-rings in the Measuring Cell should be replaced every year to avoid
a leak in the sample system. If the sample system operates above atmo-
spheric pressure, sample gas can leak from the Measuring Cell.
While the Measuring Cell is disassembled, clean its chambers and parts.
Refer to Figure 6-1 and Figure 6-2 for assistance in locating and identifying
the hardware components.
• If the analyzer responds slowly to zero when the sample system is free
of restrictions, or
2. After the Zero is complete, close the isolation valve on the Sample
Probe and the Vent Valve (or cap the Sample and Vent Lines) to isolate
the analyzer from the sample stream.
4. Remove the Analyzer Enclosure cover and remove the Optical Bench/
Measuring Cell Assembly from the analyzer. To do this (see connector
“Item” locators in Figure 6-3):
If the analyzer has the optional Flow Sensor, disconnect its cable
connector (Cable 5) from J306 (Item 5) on the Bench Controller
board.
If the analyzer has the optional 5 VDC Fan (Model 9900 Wall Mount
Analyzer only), disconnect its cable connector from J607 (Item 9)
on the Bench Controller board.
Before proceeding with the next steps, note the position of the sample
tube fitting elbows. This will be helpful when replacing these parts
NOTE
later.
1
5
4 8
2
3
Figure 6-3.
Cable connection points, Bench
Controller board, Optical Bench
removal/installation.
Loosen and back out – but do not remove – the (2) outer screws
on the blue Optical Bench Mounting Bracket (near the front panel
of the analyzer) until they are loose enough to slide the Optical
Bench/Measuring Cell Assembly toward the back of the analyzer.
! Do Not lift the Optical Bench Assembly by the printed circuit board
or lamp assembly.
CAUTION
! If the analyzer uses the optional Cell Temperature Sensor, the Mea-
suring Cell will be hot. Allow it to cool to allow safe handling or use
WARNING
insulated gloves when removing the Optical Bench/Measuring Cell
Assembly.
5. Hold the assembly with the Measuring Cell end up and remove the
(3) screws that secure the Measuring Cell to the Optical Bench Assem-
bly. Remove the Cell from the Optical Bench Assembly and place it on
a soft, non-abrasive cloth. Take care to ensure the Cell Window does
not drop out when these parts are separated.
Carefully remove the Cell Window from the Measuring Cell by grasp-
ing it by its outer edge. Avoid touching the Window surfaces.
Inspect the Window for any chips, cracks, or scratches. Place the Win-
dow on the cloth – and set it aside in a safe location – to avoid damag-
ing it.
Use an o-ring removal tool to remove the o-ring from the Cell. Take
care not to scratch or damage the o-ring groove.
6. While holding the Measuring Cell vertically so that the Reflector Block
end is up, remove the three screws that secure it to the Measuring
Cell. Remove the Reflector Block and place it on the cloth. Take care
to ensure the Cell Window does not drop out when these parts are
separated.
Carefully remove the Cell Window from the Measuring Cell by grasp-
ing it by its outer edge. Avoid touching the Window surfaces.
9. Inspect the interior of the Cell Body for foreign material. If necessary,
use a cotton swab to clean it. If a more thorough cleaning is required,
use only cleaning materials that are suitable for use with the sample
species and the Cell.
Clean the fittings using the same method as the Cell Body.
11. Starting on the Reflector Block end of the Cell, insert (1) new o-ring
(Part No. 100-2009) in the Reflector Block and (1) in the Cell Body.
Place the Window (Part No. 300-0281, handle by its outer edges only)
in the Cell Body. Replace the Reflector Block on the Cell, with the
“peak line” of the Reflector Block at 90° to a line through the fittings.
Secure the parts together using the (3) screws. Tighten the screws
evenly.
12. Invert the Measuring Cell and insert (1) new o-ring (Part No. 100-2009)
in the Cell Body.
With the Measuring Cell held vertically, grasp the edge of the Window
and carefully place it in the o-ring groove in the Cell.
13. Align the two larger-diameter holes in the Measuring Cell with the
two larger-diameter holes in the Optical Bench Assembly. Ensure
that the Cell Window does not slide out of position when the Cell is
initially placed against the Optical Bench Assembly. Secure the Cell
to the Optical Bench assembly using the (3) screws removed earlier.
Tighten the screws evenly.
! Do Not lift the Optical Bench Assembly by the printed circuit board
or lamp assembly.
CAUTION
Gently slide the assembly toward the front of the analyzer until
the mounting bracket screws are engaged in the smaller slots of
the Optical Bench Mounting Bracket. Ensure the Measuring Cell is
properly seated against the Cell Clamp.
Replace the top part of the blue Cell Clamp and fasten it with its
(2) screws.
If the analyzer has the optional 5 VDC Fan (Model 9900 Wall Mount
Analyzer only), connect its cable connector to J607 (Item 9) on the
Bench Controller board.
15. Perform a leak check on the fittings to ensure there are no leaks in the
sample system. Refer to “Pressure Leak Check” in Chapter 3 for details
on performing a leak check.
After you are sure there are no leaks, return to this procedure and
continue with the remaining steps.
16. Replace the Analyzer Enclosure cover and secure it with its screws.
18. Manually Zero and Span the analyzer as described under “Manual
Zero and Span Calibration” in Chapter 5.
19. Open the isolation valve on the Sample Probe and the Vent Valve to
allow the flow of sample gas through the sample system.
20. Adjust the flow rate to the normal operating flow rate.
The typical life span of the source lamps is approximately 9–12 months of
continuous operation. However, a source lamp can begin to exhibit signs
of degradation after 5–6 months.
DIAGLamp Ops
Figure 6-4.
Lamp Operations
diagnostic screen.
• The “ALC high error” or “ALC low error” (Automatic Lamp Control)
status conditions are set and displayed on the Human-Machine Inter-
face (HMI).
DIAGUV Bench
StatusStatus Code2
Figure 6-5.
Status Code 2 (2 of 2)
screen.
Refer to Figures 6-1 and 6-5 for assistance in locating and identifying the
hardware components encountered while replacing the source lamps.
2. After the Zero is complete, close the isolation valve on the Sample
Probe and the Vent Valve (or cap the Sample and Vent Lines) to isolate
the analyzer from the sample stream.
5. Loosen the Lamp Retainer Clip screw on each lamp and move each
clip away from each Lamp Socket. Loosen the Lamp Clamping Screw.
Remove the Lamp Socket from each lamp being replaced.
6. Remove the faulty lamp by rotating and sliding it away from the
Lamp Block Assembly.
! Do not rotate or pull on the base of the source lamp because this may
cause it to separate from the glass envelope. Grasp the glass envelope
CAUTION
when rotating or pulling the source lamp.
7. Install the new lamp(s). Make sure that the narrow end of each lamp is
pushed completely into the holder in the Lamp Block Assembly.
Replace the Lamp Retainer Clips on the Lamp Sockets and tighten the
screws.
9. Lightly tighten the Lamp Clamping Screw to secure the lamps. Do not
over tighten; the spring should not be collapsed. The lamp must be
allowed to move so it can be adjusted later in this procedure.
Figure 6-6.
Lamp maintenance diagram,
motorized Filter Wheel (multi-species)
Optical Bench, 100-2421-1B.
11. Manually Zero the analyzer for 2–3 minutes and continue to flow Zero
gas until the end of this procedure.
The Auto Setup function optimizes PMT (photomultiplier tube) high volt-
age control levels and the lamp current-pulse levels to maximize the PMT
signal strength for each valid and controlled wavelength. Always initiate
an Auto Setup after any lamps, optical filters, or PMTs have been installed
or replaced.
a. From the UV Bench Setup screen, press the Auto Setup button to
The Level1 password is
required to make changes to
display the Auto Setup screen (Figure 6-6.2). Press to start the
the UV Bench Setup screen. Auto Setup. The filters used for this application will be displayed
on the Auto Setup screen. After the Auto Setup has progressed
through all eight Setup Steps (approximately three minutes), the
message “completed” is displayed to the right of the setup steps.
Figure 6-7.1.
UV Bench Setup
screen.
CONFIGUV Bench
SetupAuto Setup
Figure 6-7.2.
Auto Setup screen.
b. After the Auto Setup is complete, view the Auto Setup Results
screen and check the PMT HV Msr and Ref supply voltage values
for both PMTs. Record these setpoints (which are adjusted during
the Auto Setup).
Also from the Auto Setup Result screen, check the Base and
Filterx (where ‘x’ = Filters 1 through 6) readings for the lamp\
filter combinations being used. Record these setpoints (which
are adjusted during the Auto Setup). Note and record the filter/
lamp combination with the highest mA value. This is the weakest
wavelength which will be used in the lamp alignment procedure.
Filters that aren’t used for the specific lamp will display a value that
corresponds with the Base reading. For example, if the unused filters
NOTE
are Filters 2, 4, 6 , the displayed values for these filters will read
“0.250,” the same as the Base reading.
a. View the Align Lamps screen and note the Filter with the weakest
signal level.
CONFIGUV Bench
SetupAuto Setup Results
Figure 6-7.3.
Auto Setup Results
screen.
CONFIGUV Bench
SetupAlign Lamps
Figure 6-7.4.
Align Lamps screen
(Multi-Species Analyzers
shown).
d. Loosen the Locking Screw, located between the two source lamps.
While viewing the Align Lamp screen, further maximize the signal
by adjusting the Lamp (1 or 2) Adjusting Screw for the lamp that is
being aligned. Again, use slow, small movements, pausing each turn
to view the new value, which is updated at 2-second intervals.
14. After the signal for both lamps has been maximized, retighten the
Locking Screw.
15. Tighten the Lamp Clamping Screw only enough to secure the lamps.
Do not over-tighten these screws; the springs should not be col-
lapsed.
16. Allow the analyzer to operate for 15 minutes and then initiate another
Auto Setup, which automatically turns on the ALC function. The Auto
Setup is complete after it has progressed through all eight steps and
the message “completed” is displayed on the Auto Setup screen.
From the Auto Setup Results screen (Figure 6-6.3) record the values
for each filter/lamp combination for long-term diagnostics.
17. Replace the Analyzer Enclosure cover and secure it with its screws.
18. Manually Zero and Span the analyzer as described under “Manual
Zero and Span Calibration” in Chapter 5.
20. Open the isolation valve on the Sample Probe and the Vent Valve to
allow the flow of sample gas through the sample system.
21. Adjust the sample gas flow rate to the normal operating flow rate.
When do the Filter Wheel Drive Belt and Bearings need to be replaced?
• Replace these parts every two years, as indicated in the “Analyzer Pre-
ventive Maintenance Schedule,” earlier in this chapter.
• If there are recurring status conditions that suggest problems with the
Filter Wheel or other related parts in the Filter Assembly, replace these
parts sooner.
Refer to Figures 6-1 and 6-8 for assistance in locating and identifying the
hardware components encountered while replacing the Drive Belt and
Bearings in the Optical Bench.
2. After the Zero is complete, close the isolation valve on the Sample
Probe and the Vent Valve (or cap the Sample and Vent Lines) to isolate
the analyzer from the sample stream.
4. Remove the Analyzer Enclosure cover and remove the Optical Bench/
Measuring Cell Assembly from the analyzer. To do this (see connector
“Item” locators in Figure 6-3):
If the analyzer has the optional Flow Sensor, disconnect its cable
connector (Cable 5) from J306 (Item 5) on the Bench Controller
board.
If the analyzer has the optional 5 VDC Fan (Model 9900 Wall Mount
Analyzer only), disconnect its cable connector from J607 (Item 9)
on the Bench Controller board.
Before proceeding with the next steps, note the position of the sample
tube fitting elbows. This will be helpful when replacing these parts
NOTE
later.
b. Disconnect the Sample and Vent Line connection fittings from the
Cell.
Figure 6-8.
Optical Bench Filter
Maintenance diagram, multi-
species (motorized Optical
Bench) analyzers, 100-2421-1A.
Loosen and back out – but do not remove – the two outer screws
on the blue Optical Bench Mounting Bracket (near the front panel
of the analyzer) until they are loose enough to slide the Optical
Bench/Measuring Cell Assembly toward the back of the analyzer.
! Do Not lift the Optical Bench Assembly by the printed circuit board
or lamp assembly.
CAUTION
! If the analyzer uses the optional Cell Temperature Sensor, the Mea-
suring Cell will be hot. Allow it to cool to allow safe handling or use
WARNING
insulated gloves when removing the Optical Bench/Measuring Cell
Assembly.
5. Remove the four M4 x 25 screws that secure the Filter Wheel Cover to
the Filter Wheel Housing. Separate the two halves of the Filter Wheel
Assembly.
6. Remove the Filter Wheel Drive Belt by rolling it outward and away
from the groove of the Pulley.
a. Remove the M4 x 10 flat head screw that secures the Filter Wheel
in place on the Filter Wheel Shaft.
Obtain two new Bearings (Part No. 300-9437) and place one in
each side of the Filter Wheel.
d. Replace the Filter Wheel on the Filter Wheel Shaft followed by the
Bushing, Wave Spring, and Retaining Washer.
9. Align the two halves of the Filter Wheel Assembly and secure them
using the four M4 x 25 screws.
! Do Not lift the Optical Bench Assembly by the printed circuit board
or lamp assembly.
CAUTION
Gently slide the assembly toward the front of the analyzer until
the mounting bracket screws are engaged in the smaller slots of
the Optical Bench Mounting Bracket. Ensure the Measuring Cell is
properly seated against the Cell Clamp.
Replace the top part of the blue Cell Clamp and fasten it with its
two screws.
If the analyzer has the optional Flow Sensor, reconnect its cable
connector (Cable 5) to J306 (Item 5) on the Bench Controller
board.
If the analyzer has the optional 5 VDC Fan (Model 9900 Wall Mount
Analyzer only), connect its cable connector to J607 (Item 9) on the
Bench Controller board.
11. Perform a leak check on the fittings to ensure there are no leaks in the
sample system. Refer to “Pressure Leak Check” in Chapter 3 for details
on performing a leak check.
After you are sure there are no leaks, return to this procedure and
continue with the remaining steps.
12. Replace the Analyzer Enclosure cover and secure it with its screws.
14. Manually Zero and Span the analyzer as described under “Manual
Zero and Span Calibration” in Chapter 5.
15. Open the isolation valve on the Sample Probe and the Vent Valve to
allow the flow of sample gas through the sample system.
16. Adjust the sample gas flow rate to the normal operating flow rate.
This section describes how the analyzer detects and displays Status code
conditions associated with its operation, what the conditions mean, and
action to take to clear the conditions.
Figure 6-9.
Diagnostic menu.
This section also discusses how to view diagnostic data from a variety of
analyzer diagnostic screens, and how to view current Status code condi-
tions from the HMI. Generally, most troubleshooting and diagnostics can
be done while working from the HMI.
The main Diagnostic menu also displays Customer Digital I/O indica-
tors that display the current state of up to five Status Relays and up to four
Contact Inputs on the Customer I/O board:
LED on
Relay Indicator State Customer I/O board
Status Relay 1 D1
Status Relay 2 White=Relay energized D6
Black=Relay de-energized
Status Relay 3 D7
Note: Relays 4,5 not available for
Status Relay 4 M9900RM Analyzers. D8
Status Relay 5 D9
Contact 1 White=Contact Closed D2
Contact 2 Black=Contact Open D3
Contact 3 Note: Contacts 3,4 not available for D4
M9900RM Analyzers.
Contact 4 D5
Note: The indicator lights on this screen correspond with the LEDs on the Customer I/O board. See
“Customer I/O Board” in Appendix C for more information.
When status conditions are detected, you can view the conditions from
the UV Bench Status diagnostic screen. Methods to notify you of detected
Warning or Fault status conditions include:
• From the home screen, the DIAG button can alert you to status con-
ditions:
- If this button is WHITE, the status is normal; no Warning or Fault
conditions have been detected.
- If this button is ORANGE and flashing, a Warning condition has
been detected. The analyzer requires attention.
- If this button is RED and flashing, a Fault condition has been de-
tected. The analyzer requires service.
• While working from screens that have the (home) button, this
button can alert you to status conditions:
The PMT Signals diagnostic screen displays the real-time value of the
Photomultiplier Tube (PMT) Measure (Msr) and Reference (Ref) voltage (V),
for each lamp\filter combination. Only values for the lamp\filter combina-
tions being used will be displayed.
DIAGPMT Signals
Figure 6-10.
PMT Signals diagnostic
screen.
The Lamp Operations diagnostic screen displays the real-time lamp Base
current (mA)/Base voltage (V) readings for Lamps 1 and 2 and the pulse
current (mA) and voltage (V) for each lamp\filter combination. The ALC
indicator light is also displayed here (GREEN indicates an Auto Lamp Con-
trol function has been enabled; OFF indicates the ALC is not enabled). Only
values for the lamp\filter combinations being used will be displayed.
DIAGLamp Ops
Figure 6-11.
Lamp Operations
diagnostic screen.
DIAGPMT\Motor Ops
Figure 6-12.
PMT\Motor Operations
diagnostic screen.
Measure-PMT HV
Displays the high voltage control of the Measure PMT (photomultiplier
tube). The unit is volts (V).
Reference-PMT HV
Displays the high voltage control of the Reference PMT (photomulti-
plier tube). The unit is volts (V).
Motor RPM
Displays the real-time reading of the number of Filter Wheel Motor
revolutions per minute (RPM). Note that the motor is geared to the
wheel at about an 8:1 ratio (i.e., the motor speed will be about eight
times the wheel speed). Nominal motor speed is 2600 RPM. (Multi-
Species or Multi-Range analyzers only.)
DIAGTrans\Abs
Figure 6-13.
Transmittance\
Absorbance diagnostic
screen.
Trans.
Displays the normalized transmittance value, which is the ratio of the
Measure PMT signal to the Reference PMT signal normalized to the
zero-transmittance of the corresponding wavelength (lamp-filter com-
bination), displayed for each lamp\filter combination. It is an interme-
diate value used in the calculation of absorbance and concentration.
Abs.
Displays the absorbance for each lamp\filter combination.
Each species is listed along with its current Track-Hold value. The Track-
Hold concentration result is the last measured concentration value for
each species that will be held constant during a Zero or Span calibra-
tion. For more information about Track-Hold settings, refer to the section
“Viewing and Changing Track-Hold Parameters” Chapter 4.
DIAGUV Analysis
Results
Figure 6-14.
UV Analysis Results
diagnostic screen.
The Sensor Signals diagnostic screen displays the real-time readings for
each operational sensor installed on the analyzer. Note that this is the
same information that is displayed on the home screen.
DIAGSensor Signals
Figure 6-15.
Sensor Signals
diagnostic screen.
UV Bench Temp
Displays the real-time temperature (°C) of the Optical Bench, as mea-
sured by the Bench Temperature Sensor. The value to the right of the
temperature is the heater duty cycle (%).
Cell Temp
Displays the real-time temperature (°C) of the Measuring Cell, as
measured by the Cell Temperature Sensor. The value to the right of
the temperature is the heater duty cycle (%). (Optional Measuring Cell
Temperature Sensor must be installed.)
Cell Pres
Displays the real-time pressure (mmHg) of the sample gas in the Mea-
suring Cell. (Optional pressure transducer must be installed.)
O2 Conc
Displays the real-time concentration (%) being measured by the O2
Sensor. (Optional O2 Sensor must be installed.)
O2 Temp
Displays the real-time temperature (°C) being measured by the O2
Sensor. (Optional O2 Sensor must be installed.)
DIAGSystem Info
Figure 6-16.1.
System Information
diagnostic screen.
Serial Number
This is the serial number of the Optical Bench installed on this ana-
lyzer, or of the analyzer itself.
Firmware Version
This is the firmware version of the Bench Controller board installed
on this analyzer.
Hardware Presence
This is optional hardware installed on this analyzer. Press to view
the Hardware Installed pop-up window.
If a box is GREEN, the corresponding hardware is installed on the
analyzer.
If a box is BLACK, the hardware is not installed.
Figure 6-16.2.
Hardware Installed
pop-up window.
Figure 6-16.3.
Wavelength Valid
diagnostic pop-up
window.
UV Species Valid
This indicates if a UV species is being measured and thus valid. Press
the UV Species Valid button to view the UV Species Valid pop-up
window (Figure 6-16.4), which expands the Species Valid code into
individual species (e.g., SO2, NO, NO2, NOx, H2S). Press again to
hide the pop-up window.
A ‘1’ below a species indicates that species is being measured and thus
valid.
A ‘0’ below a species indicates that species is not being measured and
thus not valid.
Figure 6-16.4.
UV Species Valid
diagnostic pop-up
window.
This screen also allows you to log data trending information stored in the
HMI directly to a USB Flash Drive using two different methods, for further
analysis on a computer (these methods are described later in this section):
! Hazardous Locations
Do not cycle the analyzer power on if there is a hazardous gas
WARNING
atmosphere present, unless a power-disconnect switch (approved
for the Hazardous Area) has been installed for use with the ana-
lyzer.
Figure 6-17.1.
Data Trending
diagnostic screen.
• The USB DRIVE indicators alert you to the current status of the USB
Flash Drive, either:
- inserted: When this light is ORANGE, it indicates there is a USB
Flash Drive inserted into the USB port on the HMI. When OFF, it
indicates there is no USB Flash Drive connected or the data trans-
fer is complete.
or
- writing: When this light is ORANGE, it indicates the HMI is
transferring data to the USB Flash Drive. Do not remove the USB
Flash Drive from the HMI while data is being written to it. When
OFF, it indicates there is no data being transferred to the USB Flash
Drive or the data transfer is complete.
The Data Trending screen (Figure 6-17.1) is used to initiate the transfer of
analyzer information to a USB Flash Drive.
The HMI stores the latest 12 hours of trend data. This procedure al-
lows for a one-time transfer of the data stored on the HMI to the USB
NOTE
Flash Drive.
Figure 6-17.2.
USB drive contents.
• Four directories for four trend data groups:
o SAMP01: UV analysis results including UV1 to UV9 Concentration.
o SAMP02: Sensor measurements including BoardTemp, BenchTemp, CellTemp, CellPres, FlowRate, O2% and O2Temp.
o SAMP03: UV bench signal group A including PMT Signals and Lamp Base and Pulse Current,
o SAMP03: UV bench signal group B including PMT HV, Lamp HV and Motor and Wheel RPM.
To do a one-time transfer of trend data from the HMI to the USB Flash
Drive:
1. From the Data Trending screen, check to ensure the “inserted” light is
on (ORANGE).
! Before removing the USB Flash Drive in Step 2, power down the
analyzer/HMI unit and then disconnect the HMI’s Communication
WARNING
module from its HMI Display to allow access to the USB Flash Drive.
The data transfer will take several seconds. When both “inserted” and
“writing” lights turn off, the USB Flash Drive can be safely removed. Fig-
ure 6-17.2 shows the contents of the USB Flash Drive, and describes how
the different directories are used.
To start continuous logging of trend data from the HMI to the USB Flash
Drive:
2. Press the LOGGING TO USB switch to turn it ON. This will initiate
data logging and the transfer of trend data continuously to the USB Flash
Drive.
1. Press the LOGGING TO USB switch to turn it OFF. This will stop data
logging and the transfer of trend data to the USB Flash Drive.
Press the UV Analysis Results button to view trend graphs of the UV spe-
cies being analyzed and to configure the concentration range you wish to
view. From the UV Analysis Result Trend Graphs screen, change the high
and low Y-AXIS concentration ranges you wish to view, then press one of
the buttons to view the corresponding group of UV species in graphi-
cal format. See Figure 6-17.6 for an example of a Trend Graph screen.
DIAGData Trending
UV Analysis Results
Figure 6-17.3.
UV Analysis Result
Trend Graphs
diagnostic screen.
Press the UV Bench Signal button to view trend graphs of signals derived
from the Optical Bench and to configure the signal range you wish to
view. From the UV Bench Signal Trend Graphs screen, change the high
and low Y-AXIS range of values you wish to view, then press one of the
buttons to view the information in graphical format. See Figure 6-17.6
for an example of a Trend Graph screen.
DIAGData Trending
UV Bench Signals
Figure 6-17.4.
UV Bench Signal
Trend Graphs
diagnostic screen.
Press the Sensor Signals button to view trend graphs of the signals from
each of the optional (installed) sensors and to configure the signal range
you wish to view. From the Sensor Signal Trend Graphs screen, change
the high and low Y-AXIS range of values you wish to view, then press one
of the buttons to view the information in graphical format. See Figure
6-17.6 for an example of a Trend Graph screen. If a sensor is not installed
or enabled, the button will not be displayed.
DIAGData Trending
Sensor Signals
Figure 6-17.5.
Sensor Signal Trend
Graphs diagnostic
screen.
When you press a graph ( ) button from one of the Trend Graph
setup screens, the data you selected is sent to a graph. Figure 6-17.6 is an
example of a Trend Graph.
Valid or installed parameters will have a white background
DIAGData Trending
<<data trend option>>
Figure 6-17.6.
Example, Data Trend
graph diagnostic
screen.
While working from a Trend Graph, the following navigation buttons are
available:
Press To
Scroll backward; view trend data before the current data point (only if more
than a full screen of data has been collected).
Scroll forward; view trend data after the current data point.
The , , , and
Graph has been paused.
buttons are active only if the Trend
NOTE
If Warning and/or Fault conditions are triggered by the analyzer, further in-
vestigation into the cause of the condition and correction of the problem(s) is
required to ensure the analyzer operates problem-free. This section describes
how to determine which status codes (Warning or Fault) are active, possible
trigger conditions, and steps to take to correct or reset the conditions.
If Status Code conditions have been detected, the sum of their respective
primary codes is displayed to the right of the Status Code1/2/3. To view
the condition(s) that are active, press the corresponding button. De-
pending on which button is pressed, the corresponding Status Code (1, 2,
or 3) screen will be displayed. To view more status conditions, press .
To return to the previous screen, press .
DIAGSystem Status
Figure 6-18.1.
System Status
diagnostic screen.
Run Time
This is the amount of time the Bench Controller board has been per-
forming controls since the analyzer was last powered up or reset.
The Status Code (1, 2, or 3) Screens and Diagnostics sections in the follow-
ing pages list all of the analyzer status conditions. These status conditions
are described using the following information:
Description
Describes the status code condition.
Trigger Condition
Describes the possible causes that can trigger the status condition.
Corrective Action
Describes the conditions that must be met and the action that must be
taken to correct the conditions that triggered the condition.
These status code conditions are derived from the Optical Bench. After
performing each Corrective Action procedure, check the Status Code
screen to see if the condition has been reset.
For location of cables referenced in the Status Code 1 Diagnostic chart, refer
to the “Analyzer Wiring Diagram” in Appendix C.
NOTE
DIAGUV Bench
StatusStatus Code1
Figure 6-18.2.
Status Code1 diagnostic
screen (1 of 2),
UV Bench Status.
DIAGUV Bench
StatusStatus Code1
Figure 6-18.3.
Status Code1 diagnostic
screen (2 of 2),
UV Bench Status.
Trigger Condition:
• The cable (Cable 3) between the Bench Controller board and the PMT Supply
board may be disconnected, may not be properly installed, or may be faulty.
Corrective Action:
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the PMT Supply Board Ribbon Cable (Cable 3) between the Bench Con-
troller board and the PMT Supply board for proper connections and inspect it for
damage (cuts, nicks, burn marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the boards.
• Take appropriate safety precautions, isolate the analyzer sample system, and power
down the analyzer to make the following check:
- Remove the PMT Block cover (see Figure 6-5 for location) and check the connec-
tions between the PMT Supply board and the PMT Signal board. Ensure all pins
on both connectors are properly seated in the sockets.
WARNING:
High voltage exists on the PMT Supply board. Do not touch the connec-
tor or pins.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the PMT Data Communication Circuit on the Bench Controller
board, or problems with the PMT Signal board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• The cable (Cable 3) between the Bench Controller board and the PMT HV Supply
board may not be properly installed, or may be faulty.
Corrective Action:
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the PMT Supply Board Ribbon Cable (Cable 3) between the Bench Con-
troller board and the PMT Supply board for proper connections and inspect it for
damage (cuts, nicks, burn marks, etc.)
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the boards.
• Take appropriate safety precautions, isolate the analyzer sample system, and power
down the analyzer to make the following check:
- Remove the PMT Block cover (see Figure 6-5 for location) and check the connec-
tions between the PMT Supply board and the PMT Signal board. Ensure all pins
on both connectors are properly seated in the sockets.
WARNING:
High voltage exists on the PMT Supply board. Do not touch the connec-
tor or pins.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the PMT Data Communication Circuit on the Bench Controller
board, or problems with the PMT Signal board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• The status condition may be triggered upon analyzer power-up while the Bench is
warming up to its temperature control setpoint, but will clear after the Bench has
properly warmed up to its normal operating temperature.
• Low Measure-PMT signal level may be caused by concentration level exceeding the
maximum expected range for a given Measuring Cell length. The status will clear
itself once the high concentration occurrence from the process stream ends.
• Problem with the UV source lamp(s) (e.g., lamps may be nearing the end of their
operational lives), or faulty socket and wiring.
Corrective Action:
• From the HMI, view the PMT Signals screen to check the PMT signal for each
DIAGPMT Signals
valid wavelength to see which lamp\filter combination(s) cause the condition. If the
condition is triggered by wavelengths from the same lamp, that lamp may need to be
replaced. If the condition is triggered by signals from the same PMT (i.e., Ref PMT or
Msr PMT), that PMT may be faulty.
• From the HMI, view the Auto Setup Results screen to check if the ALC switch is
CONFIGUV Bench
SetupAuto Setup Results
OFF. If necessary, turn it ON.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the lamps to ensure they are firing. This is evident if an orange flicker or
glow can be seen in the lamps.
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sockets
are properly connected. To verify lamp types and locations, check the “9900_
Signed Final QC document”.
- Check the lamps to ensure they are not damaged. Replace them if necessary. Re-
fer to “Replacing the Source Lamps” in this chapter for information about the lamps
and replacing them.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the boards.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the PMT Signal Monitoring Circuit and PMT Control Circuit or the
Lamp Current Signal Monitoring Circuit and Lamp Current Control Circuit on the
Bench Controller board. It can also indicate problems with the PMT Signal board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
Corrective Action:
• From the HMI, view the PMT\Motor Operations screen and check the Measure-
DIAGPMT\Motor Ops
PMT HV or Reference-PMT HV readings to see if they are outside their setpoints. If
so, the PMT Supply board may be faulty.
• Initiate another Auto Setup to try to clear the problems. Refer to the section “Source
Lamps Replacement” in this chapter for information about how to do this.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the boards.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the PMT Signal Monitoring Circuit and PMT Control Circuit on the
Bench Controller board. It can also indicate problems with the PMTs or the PMT
Supply board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• Problem with Lamp1 or Lamp2 (e.g., lamps may be nearing the end of their opera-
tional lives, or faulty socket and wiring).
Corrective Action:
- View the Lamp Operations screen to check the measured Lamp1 Base/Lamp2
DIAGLamp Ops
Base and Lamp1/Filter and Lamp2/Filter combination pulse currents and
compare them to the values on the Auto Setup Results screen to see if they are
CONFIGUV Bench
SetupAuto Setup Results
outside of their pulse control levels.
- Initiate another Auto Setup to try to clear the problems. Refer to the section
“Source Lamps Replacement” in this chapter for information about how to do this.
• Determine the age of the lamps to see if the lamps are approaching the end of their
useful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the corresponding Lamp Socket Cable to ensure its socket is properly con-
nected at the lamp and at the Bench Controller board (J201). Inspect the cable
for damage (cuts, nicks, burn marks, etc.).
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sockets
are properly connected. To verify, check the “9900_Signed Final QC” document for
lamp types and locations.
- Check the lamp that caused the condition to ensure it is not damaged. Replace it if
necessary. Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle the
analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Signal Monitoring Circuit or the Lamp Current Con-
trol Circuit on the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• Problem with Lamp1 or Lamp2 (e.g., lamps may be nearing the end of their opera-
tional lives, or faulty socket and wiring).
Corrective Action:
- View the Auto Setup Results screen to check Lamp1 Base/Lamp2 Base current
CONFIGUV Bench
SetupAuto Setup Results
readings to see if they are outside their setpoints.
- View the Lamp Operations screen to check Lamp1 Base/Lamp2 Base HV read-
DIAGLamp Ops
ings to check if they are within 15 % of the maximum lamp HV value and to check
the measured Lamp1/Lamp2 Pulse HV readings to see if they are within 10 % of
the maximum.
- Initiate another Auto Setup to try to clear the problems. Refer to the section
“Source Lamps Replacement” in this chapter for information about how to do this.
• Determine the age of the lamps to see if the lamps are approaching the end of their
useful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the corresponding Lamp Socket Cable to ensure its socket is properly con-
nected at the lamp and at the Bench Controller board (J201). Inspect the cable
for damage (cuts, nicks, burn marks, etc.).
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly), and that the Lamp Sockets
are connected properly. To verify, check the “9900_Signed Final QC” document for
lamp types and locations.
- Check the lamp that caused the condition to ensure it is not damaged. Replace it if
necessary. Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Signal Monitoring Circuit on the Bench Controller
board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
This status code
condition applies only to
• The Filter Wheel Opto-interruptor cable or Filter Wheel Motor cable may be discon-
nected, may not be installed properly, or may be faulty.
Multi-Species (Motorized
Optical Bench) or Multi-Range
• Problem with the Filter Wheel Opto-interruptor.
Analyzers.
• Problem with the Filter Wheel drive belt or bearings.
Corrective Action:
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure and:
- Check the frequency of the Motor at TP309 on the Bench Controller board. The
frequency should be between 120 to 140 Hz (typical). If it is outside this range, the
Motor may be faulty.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle the
analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
• If cycling the AC power Off-On does not reset the condition, cycle the power Off
again, disassemble the Filter Assembly (part of Optical Bench Assembly), and check
for:
If these parts are damaged, replace them. Refer to the section “Filter Wheel Mainte-
nance – Motorized Optical Bench Analyzers” in this chapter.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Opto-interruptor, or the Opto-interruptor Signal Conditioning Cir-
cuit or Motor Power Supply Circuit on the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
This status code
condition applies only to
• The Filter Wheel Motor cable may be disconnected, may not be installed properly, or
may be faulty.
Multi-Species (Motorized
Optical Bench) or Multi-Range • Problem with the Filter Wheel Motor.
Analyzers.
• Problem with the Bench Controller board.
Corrective Action:
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the Filter Wheel Motor cable (at J303 on Bench Controller board) to
ensure it is properly connected and inspect it for damage (cuts, nicks, burn marks,
etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle the
analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
- The Filter Wheel Motor or its cable appears damaged – order a new part and
replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Motor Speed Pulse Signal Conditioning Circuit or Motor Power
Supply Circuit on the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
This status code
condition applies only to
• The Filter Wheel Opto-interruptor cable or Filter Wheel Motor cable may not be
installed properly, or may be faulty.
Multi-Species (Motorized
Optical Bench) or Multi-Range • Problem with the Filter Wheel Opto-interruptor.
Analyzers.
• Problem with the Filter Wheel drive belt or bearings.
Corrective Action:
• View the PMT\Motor Operations screen and check if the Filter Wheel RPM reading
is outside its nominal speed range.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
• If cycling the AC power Off-On does not reset the condition, cycle the power Off
again, disassemble the Filter Assembly (part of Optical Bench Assembly), and check
for:
If the Filter Wheel drive belt is stretched or the other parts are damaged or worn,
replace them. Refer to “Filter Wheel Maintenance – Motorized Optical Bench Analyz-
ers” in this chapter.
- The cable or other internal parts appear damaged – order new parts and replace
them.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Opto-interruptor, or the Opto-interruptor Signal Conditioning Cir-
cuit on the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
This status code
condition applies only to
• The Filter Wheel Motor cable may not be installed properly, or may be faulty.
Multi-Species (Motorized • Problem with Filter Wheel Motor.
Optical Bench) or Multi-Range
Analyzers. • Problem with the Bench Controller board.
Corrective Action:
• View the PMT\Motor Operations screen and check if the Motor RPM reading is
outside its nominal speed range.
• Check the Filter Wheel Motor cable (at J303 on Bench Controller board) to ensure it
is properly connected and inspect it for damage (cuts, nicks, burn marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any problems
with the Bench Controller board.
- The cable appears damaged or the Motor is not operating properly – order a new
part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Motor Speed Pulse Signal Conditioning Circuit or Motor Power
Supply Circuit on the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
These status code conditions are derived from the Optical Bench and the
Customer I/O board. After performing each Corrective Action procedure,
check the Status Code screen to see if the condition has been reset.
For location of cables referenced in the Status Code 2 Diagnostic chart, refer
to the “Analyzer Wiring Diagram” in Appendix C.
NOTE
DIAGUV Bench
StatusStatus Code2
Figure 6-18.4.
Status Code2 diagnostic
screen (1 of 2),
UV Bench Status.
DIAGUV Bench
StatusStatus Code2
Figure 6-18.5.
Status Code2 diagnostic
screen (2 of 2),
UV Bench Status.
Trigger Condition:
• The communication cable (Cable 2) between the Bench Controller board and the
Customer I/O board may be disconnected, may not be properly connected, or may
be faulty.
Corrective Action:
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check the cable between the Bench Controller board (J602) and the Customer
I/O board (J1) for proper connection and inspect it for damage (cuts, nicks, burn
marks, etc.).
- Check the power supply cables to the Bench Controller board (cable from AC
Power Distribution/MOV board, connects to J604; cable from 24 VDC Power
Supply, connects to J101) for proper connections and inspect them for damage
(cuts, nicks, burn marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the boards.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board or the Customer I/O board firmware,
or a problem with the Communication Circuit on the Bench Controller board or
the Customer I/O board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Corrective Action:
• If the cause is due to a process upset, the status will clear itself once the process
upset occurrence ends.
• From the HMI, view the Analog Output Setup screen and:
- Check the numbers (data source assignment) in the Assignment code field for
the problematic analog output channels. Compare these numbers to the codes in
the Analog Output Assignment Code screens to ensure they correspond. For
more information about the Assignment codes, refer to the section “Analog Output
Setup” in Chapter 4.
- Check the low Scale and/or high Scale output value for each output and com-
pare them to their corresponding values on the “9900_CFG Report” document. If
necessary, adjust the analog output range.
• Check the sample system for plugs or leaks, as described under “Detecting a Plug
in the Sample System” in this chapter.
- There has been a process upset, it has been corrected, and the condition still has
not reset.
- The concentration(s) of the species in the process stream appear to be too high
or too low for the Measuring Cell installed in the analyzer.
Trigger Condition:
• Weak PMT Signal(s) due to process gas or span gas in the Measuring Cell, or Cell
contamination, when an Auto Setup has been initiated.
• Problem with the UV lamp(s) (e.g., lamps may be nearing the end of their opera-
tional lives, or faulty socket and wiring).
Corrective Action:
• Check to ensure Zero gas is flowing through the Measuring Cell prior to initiating an
Auto Setup.
• Determine the age of the lamps to see if they are approaching the end of their use-
ful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
- View the PMT\Motor Operations screen and check the voltage being supplied to
DIAGPMT\Motor Ops
each PMT to see if one of the PMT high voltage controls is not within 2 % of the
minimum (below 310 V) or maximum (above 790 V) values.
- View the Lamp Operations screen and check the measured Lamp1/Filter and
DIAGLamp Ops
Lamp2/Filter combination pulse Base current readings to ensure the unused
wavelengths (lamp-filter combinations) are set to the Base (mA) values.
- View the UV Bench Setup screen and check the Target PMT Signal and
Setup
CONFIGUV Bench
Lamp1/2 Base (ma) values to ensure they have not been adjusted from the
factory settings. Verify this by comparing them to the values on the “9900_CFG
Report” document.
- View the UV Bench Info screen and check to ensure the Wavelength Valid and
DIAGUV Bench Info
UV Species Valid values have not been changed from the values on the “9900_
CFG Report” document.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check both Lamp Socket cables to ensure they are properly connected at the
lamps and at the Bench Controller board (J201). Inspect the cables for damage
(cuts, nicks, burn marks, etc.).
Check the Lamp Sockets to ensure they have not been switched in the Optical
Bench (i.e., Lamp 1 socket is connected to Lamp 1 – Lamp 1 is closest to the
Bench Controller board).
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sock-
ets are properly connected. To verify, check the “9900_Signed Final QC” docu-
ment for lamp types and locations.
- Check both lamps to ensure they are not damaged. Replace them if necessary.
Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the boards.
• If the reset doesn’t correct the problem, check the Measuring Cell Windows to
ensure they are clean and not damaged. While the Cell is removed, check for leaks
from the Measuring Cell into the Reflector Block and Lamp Block Assembly. Clean
and/or replace any damaged parts. Replace all o-rings. This procedure is discussed
under “Measuring Cell Preventive Maintenance” in this chapter.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Signal Monitoring Circuit, Lamp Current Control
Circuit, PMT Signal Monitoring Circuit, or PMT Control Circuit on the Bench
Controller board. It can also indicate problems with PMT Signal board or PMT
Supply board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
ALC setpoint error One of the ALC setpoints is outside 2 % to 98 % of the ADC range (i.e., below 0.2 V or
above 9.8 V).
Trigger Condition:
• Problem with Bench Auto Setup function (e.g., Auto Setup not completed success-
fully or incorrect Auto Setup parameters).
Corrective Action:
• Determine the age of the lamps to see if they are approaching the end of their use-
ful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
- Initiate another Auto Setup. Refer to the section “Source Lamps Replacement,”
earlier in this chapter, for information about how to do this.
- View the UV Bench Setup screen and check the Target PMT Signal and
Setup
CONFIGUV Bench
Lamp1/2 Base (ma) values to ensure they have not been adjusted from the
factory settings. Verify this by comparing them to the values on the “9900_CFG
Report” document.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- Check both Lamp Socket cables to ensure they are properly connected at the
lamps and at the Bench Controller board (J201). Inspect the cables for damage
(cuts, nicks, burn marks, etc.).
Check the Lamp Sockets to ensure they have not been switched in the Optical
Bench (i.e., Lamp 1 socket is connected to Lamp 1 – Lamp 1 is closest to the
Bench Controller board).
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sock-
ets are properly connected. To verify, check the “9900_Signed Final QC” docu-
ment for lamp types and locations.
- Check both lamps to ensure they are not damaged. Replace them if necessary.
Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the boards.
• If the reset doesn’t correct the problem, check the Measuring Cell Windows to
ensure they are clean and not damaged. While the Cell is removed, check for leaks
from the Measuring Cell into the Reflector Block and Lamp Block Assembly. Clean
and/or replace any damaged parts. Replace all o-rings. This procedure is discussed
under “Measuring Cell Preventive Maintenance” in this chapter.
- The cable appears damaged – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Signal Monitoring Circuit, Lamp Current Control
Circuit, PMT Signal Monitoring Circuit, or PMT Control Circuit on the Bench
Controller board. It can also indicate problems with PMT Signal board or PMT
Supply board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• Problem with the UV lamp(s) (e.g., incorrect replacement lamp, lamps may be near-
ing the end of their operational lives, or faulty socket and wiring).
Corrective Action:
• From the HMI, view the Lamp Operations screen to check the Lamp1 Pulse mA
DIAGLamp Ops
and Lamp2 Pulse mA readings to see which lamp-filter combinations have reached
at or below the base lamp current level.
• Determine the age of the lamps to see if the lamps are approaching the end of their
useful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sock-
ets are properly connected. To verify, check the “9900_Signed Final QC” docu-
ment for lamp types and locations.
- Check the Lamp Sockets to ensure they have not been switched in the Optical
Bench (i.e., Lamp 1 socket is connected to Lamp 1 – Lamp 1 is closest to the
Bench Controller board).
Check both Lamp Socket cables to ensure they are properly connected at the
lamps and at the Bench Controller board (J201). Inspect the cables for damage
(cuts, nicks, burn marks, etc.).
- Check the lamps to ensure they are not damaged. Replace them if necessary.
Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the boards.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Control Circuit or PMT Control Circuit on the
Bench Controller board, or a problem with the PMT Signal board or PMT Sup-
ply board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• Problem with the UV lamp(s) (e.g., incorrect replacement lamp, lamps may be near-
ing the end of their operational lives, or faulty socket and wiring).
Corrective Action:
• From the HMI, view the Lamp Operations screen to check the Lamp1 Pulse mA
DIAGLamp Ops
and Lamp2 Pulse mA readings to see which lamp-filter combinations have reached
the maximum lamp current level.
• Determine the age of the lamps to see if the lamps are approaching the end of their
useful lives (approximately 270 days). The lamps may need to be replaced. Refer to
“Replacing the Source Lamps” in this chapter for details.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and:
- If the analyzer uses two different types of lamps, check the lamps to ensure they
are in the correct location in the Lamp Block Assembly, and that the Lamp Sock-
ets are properly connected. To verify, check the “9900_Signed Final QC” docu-
ment for lamp types and locations.
- Check the Lamp Sockets to ensure they have not been switched in the Optical
Bench (i.e., Lamp 1 socket is connected to Lamp 1 – Lamp 1 is closest to the
Bench Controller board).
Check both Lamp Socket cables to ensure they are properly connected at the
lamps and at the Bench Controller board (J201). Inspect the cables for damage
(cuts, nicks, burn marks, etc.).
- Check the lamps to ensure they are not damaged. Replace them if necessary.
Refer to “Replacing the Source Lamps” in this chapter.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the boards.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Lamp Current Control Circuit or PMT Control Circuit on the
Bench Controller board, or a problem with the PMT Signal board or PMT Sup-
ply board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
These status code conditions are derived from the standard and optional
sensors (if installed) and the Optical Bench. After performing each Correc-
tive Action procedure, check the Status Code screen to see if the condition
has been reset.
For location of cables referenced in the Status Code 3 Diagnostic chart, refer
to the “Analyzer Wiring Diagram” in Appendix C.
NOTE
DIAGUV Bench
StatusStatus Code3
Figure 6-18.6.
Status Code3 diagnostic
screen (1 of 2),
UV Bench Status.
DIAGUV Bench
StatusStatus Code3
Figure 6-18.7
Status Code3 diagnostic
screen (2 of 2),
UV Bench Status.
Corrective Action:
• From the HMI, view the Sensor Setup screen and check the O2 Concentration
Setup
CONFIGSensor
High/Low Scale to verify they are within acceptable levels. Compare the readings
to the values listed on the “9900_CFG Report” document.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and check the O2 Sensor cable (Cable 1) to ensure it is properly connected at
the Bench Controller board (J307). Inspect the cable for damage (cuts, nicks, burn
marks, etc.).
• Test the O2 Sensor for proper operation and replace it if necessary. Refer to the O2
Sensor manual shipped with the analyzer for information on how to verify proper
sensor operation.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer; this may also clear any prob-
lems with the Bench Controller board.
- The cable appears damaged or the O2 Sensor does not appear to be operating
properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Oxygen Sensor Signal Conditioning Circuit on the Bench Con-
troller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
This status code
condition applies only to
• Problem with sample flow (e.g., plugging, or faulty regulator).
analyzers with the optional
• Problem with Flow Sensor scaling parameters.
Flow Sensor installed.
• Problem with the Flow Sensor cable.
Corrective Action:
- View the Sensor Signals diagnostic screen and check the Flow Rate reading to
DIAGSensor Signals
verify there is a flow of sample gas. If there is no or low flow, it may indicate a plug
in the sample system. Refer also to “Detecting a Plug in the Sample System” in
this chapter. Unplug the lines if necessary.
- View the Sensor Setup screen and check the Flow Rate Low Scale/High Scale
Setup
CONFIGSensor
parameters. Compare the readings to the value listed on the “9900_CFG Report”
document and adjust the scaling parameters if necessary.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and check the Flow Sensor cable (Cable 5) to ensure it is properly connected
at the Bench Controller board (J306). Inspect the cable for damage (cuts, nicks,
burn marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
- Test the Flow Sensor for proper operation and replace it if necessary. Refer to the
Flow Sensor documentation shipped with the analyzer for information on how to
verify proper sensor operation.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer; this may also clear any problems with the Bench Controller
board.
- The cable appears damaged or the Flow Sensor does not appear to be operating
properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Flow Sensor Signal Conditioning Circuit on the Bench Control-
ler board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Corrective Action:
- View the Sample Cell Conditions screen and check the Nominal Cell Pressure
CONFIGUV
AnalysisSample Cell
value to verify it agrees with the Pressure value listed on the “9900_CFG Report”
document, or that of the process pressure. If necessary, change the Nominal Cell
Pressure value.
Conditions
Setup
CONFIGSensor
- View the Sensor Setup screen and check the Flow Rate reading to verify there is
a flow of sample gas. If there is no or low flow, it may indicate a plug in the sample
system. Refer also to “Detecting a Plug in the Sample System” in this chapter.
Unplug the lines if necessary.
Also, check the Cell Pressure Low Scale/High Scale parameters. Compare the
readings to the value listed on the “9900_CFG Report” document and adjust the
scaling parameters if necessary.
• Take appropriate safety precautions, remove the top cover from the analyzer Enclo-
sure, and check the Pressure Transducer cable to ensure it is properly connected at
the Bench Controller board (J601). Inspect the cable for damage (cuts, nicks, burn
marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
- Test the Cell Pressure Transducer for proper operation. Use a multimeter to mea-
sure the resistance of the pressure transmitter to test for an open circuit.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer; this may also clear any problems with the boards.
- The cable appears damaged or the Pressure Transducer does not appear to be
operating properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Flow Sensor Signal Conditioning Circuit on the Bench Control-
ler board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
For systems with a Measuring Cell Heater installed, this status indicates that the Cell
temperature is 3 Celsius degrees below the Cell temperature control setpoint.
Trigger Condition:
• This status condition may be triggered upon analyzer power-up while the Cell is
warming up to its operating temperature control setpoint, but will clear after the Cell
has properly warmed up to its normal operating temperature.
• For systems with a Measuring Cell Heater, the Cell temperature Setpoint has not
been entered or is incorrect.
• Problem with the Cell Temperature Sensor cable or heater (systems with a Measur-
ing Cell heater is installed).
Corrective Action:
Also, compare the Nominal Cell Temperature value to the Cell Temp value
listed on the “9900_CFG Report” document. If the Nominal Cell Temperature is
different, change it to match the value on the “9900_CFG Report” document, or to
the normal temperature expected at installation.
Also, compare the Cell Setpoint temperature to the Cell Temp Setpoint value
listed on the “9900_CFG Report” document. If the Cell Setpoint temperature is
different, change it to match the value on the “9900_CFG Report” document, or to
the normal temperature expected at installation.
• Ensure the Analyzer Enclosure cover is properly secured with its screws to stop
heat loss.
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and
check the Cell Temperature Sensor cable to ensure it is properly connected at the
Bench Controller board (J605). Inspect the cable for damage (cuts, nicks, burn
marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
- Test the Cell Temperature Sensor for proper operation. Use a multimeter to mea-
sure the resistance of the temperature sensor to test for an open circuit.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer and may clear any problems with the Bench Controller board.
- The cable appears damaged or the Cell Temperature Sensor does not appear to
be operating properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Cell Temperature Sensor Signal Conditioning Circuit on the
Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
• This status condition may be triggered upon analyzer power-up while the Bench
is warming up to its operating temperature control setpoint, but will clear after the
Bench has properly warmed up to its normal operating temperature.
Corrective Action:
- If the analyzer has been operating under normal conditions, view the BenchT
CONFIGTemperature
Control
reading (from the HOME screen). Then, view the Temperature Control screen
and check the Bench Setpoint temperature to verify if the BenchT temperature
reading is 3 Celsius degrees below Bench Setpoint.
Also, compare the Bench Setpoint temperature to the Bench Temp Setpoint
value listed on the “9900_CFG Report” document. If the Bench Setpoint tem-
perature is different, change it to match the value on the “9900_CFG Report”
document.
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and:
- Check the Bench Temperature Heater Circuit wiring to ensure it is properly con-
nected at the Bench Controller board (J604) and inspect it for damage (cuts,
nicks, burn marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
Notes:
- The PMT Block will have to be removed to access the Overtemp
Switch and Temperature Transmitter.
- Refer to the “Analyzer Internal Component Layout, All Options”
and “Optical Bench Heater, Overtemp Switch, and RTD Locations”
drawings in Appendix C for locations of Optical Bench Overtemp
Switch and Temperature Transmitter.
- Check the Optical Bench Overtemp Switch. If the plunger has popped out, press it
in to reset it.
Warning:
Disconnect the AC power before attempting to access or reset the
Overtemp Switch.
- Test the Bench Temperature Sensor and Heater for proper operation. Use a multi-
meter to measure the resistance of the Temperature Sensor and the Heater to test
for an open circuit.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer and may also clear any problems with the Bench Controller
board.
- The Optical Bench Overtemp Switch, Temperature Sensor, or Heater (or their wir-
ing), appears damaged – order new parts.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board firmware, or a problem with the Bench
Temperature Sensor Signal Measurement Circuit or Bench Heater Control Circuit.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Bench board The Bench Controller board temperature is below 5 °C. The Bench Controller
temperature low board temperature is typically considered to be the air temperature within the analyzer
Enclosure in most cases.
Trigger Condition:
Corrective Action:
• Ensure the Analyzer Enclosure cover is closed and secured with its screws to stop
heat loss.
• Check the ambient temperature around the analyzer and adjust it if necessary.
For systems with a Measuring Cell Heater installed, this status indicates that the Cell
temperature is 3 Celsius degrees above the Cell temperature control setpoint.
Trigger Condition:
• Problem with the Measuring Cell Heater (systems with a Measuring Cell Heater
only).
Corrective Action:
• Ensure the Analyzer Enclosure cover is closed and secured with its screws.
Also, compare the Cell Setpoint temperature to the Cell Temp Setpoint value
listed on the “9900_CFG Report” document. If the Cell Setpoint temperature is
different, change it to match the value on the “9900_CFG Report” document, or to
the normal temperature expected at installation.
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and
check the Cell Temperature Sensor cable to ensure it is properly connected at the
Bench Controller board (J605). Inspect the cable for damage (cuts, nicks, burn
marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
- Test the Cell Temperature Sensor for proper operation. Use a multimeter to mea-
sure the resistance of the temperature sensor to test for an open circuit.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer and may clear any problems with the Bench Controller board.
- The cable appears damaged or the Cell Temperature Sensor does not appear to
be operating properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Cell Temperature Sensor Signal Measuring Circuit on the Bench
Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Bench temperature The Bench temperature is 3 Celsius degrees above the Bench temperature control
high setpoint.
Trigger Condition:
Corrective Action:
- If the analyzer has been operating under normal conditions, view the BenchT
CONFIGTemperature
Control
reading (from the HOME screen). Then, view the Temperature Control screen
and check the Bench Setpoint temperature to verify if the BenchT temperature
reading is 3 Celsius degrees above Bench Setpoint.
Also, compare the Bench Setpoint temperature to the Bench Temp Setpoint
value listed on the “9900_CFG Report” document. If the Bench Setpoint tem-
perature is different, change it to match the value on the “9900_CFG Report”
document.
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and:
- Check the Bench Temperature Heater Circuit wiring to ensure it is properly con-
nected at the Bench Controller board (J604) and inspect it for damage (cuts,
nicks, burn marks, etc.).
Notes:
- The PMT Block will have to be removed to access the Overtemp
Switch and Temperature Transmitter.
- Refer to the “Analyzer Internal Component Layout, All Options”
and “Optical Bench Heater, Overtemp Switch, and RTD Locations”
drawings in Appendix C for locations of Optical Bench Overtemp
Switch and Temperature Transmitter.
- Check the Optical Bench Overtemp Switch. If the plunger has popped out, press it
in to reset it.
Warning:
Disconnect the AC power before attempting to access or reset the
Overtemp Switch.
Allow the interior of the analyzer to cool down enough to ensure safe
handling of components within the Electronics Enclosure.
- Test the Bench Temperature Sensor and Heater for proper operation. Use a multi-
meter to measure the resistance of the Temperature Sensor and the Heater to test
for an open circuit.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer; this may also clear any problems with the Bench Controller
board.
- The Optical Bench Overtemp Switch, Temperature Sensor, or Heater (or their wir-
ing), appears damaged – order new parts.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board firmware, or a problem with the Bench
Temperature Sensor Signal Measurement Circuit or Bench Heater Control Circuit.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
Corrective Action:
• Check the ambient temperature around the analyzer and adjust it if necessary.
- The Optical Bench Overtemp Switch, Temperature Sensor, or Heater (or their wir-
ing), appears damaged – order new parts.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board firmware, or a problem with the Bench
Temperature Sensor Signal Measurement Circuit or Bench Heater Control Circuit.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Cell temperature The Bench Controller board has detected a problem with the Measuring Cell
sensor error Temperature Sensor.
Trigger Condition:
Corrective Action:
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and
check the Cell Temperature Sensor cable to ensure it is properly connected at the
Bench Controller board (J605). Inspect the cable for damage (cuts, nicks, burn
marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
- Test the Cell Temperature Sensor for a short or open circuit. Replace the Tem-
perature Sensor, if necessary.
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer and may clear any problems with the Bench Controller board.
- The cable appears damaged or the Cell Temperature Sensor does not appear to
be operating properly – order a new part and replace it.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Trigger Condition:
Corrective Action:
• Take appropriate safety precautions, remove the Analyzer Enclosure cover, and
check the Bench Temperature Sensor to ensure it is properly connected at the
Bench Controller board (J603) and inspect the wiring for damage (cuts, nicks, burn
marks, etc.).
• Take appropriate safety precautions, isolate the analyzer sample system, power
down the analyzer, and:
After this check has been made, cycle the analyzer AC power back On; this will also
reset the analyzer and may clear any problems with the Bench Controller board.
- The Bench Temperature Sensor or its wiring appears damaged – order new parts.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board firmware.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Bench board The Bench Controller board has detected a problem with the Bench Controller
temperature sensor board temperature sensor.
error
Trigger Condition:
Corrective Action:
• Check the ambient temperature around the analyzer and adjust it if necessary.
• Take appropriate safety precautions, isolate the analyzer sample system, and cycle
the analyzer AC power Off-On to reset the analyzer. This may clear any problems
with the Bench Controller board.
- Cycling the AC power Off-On does not reset the condition. This can indicate a
problem with the Bench Controller board firmware.
- You need to verify operation or to order replacement parts, or you require replace-
ment instructions.
Technical Support
AMETEK Western Research is committed to providing you the best tech-
nical support in the industry. If you need service or application assistance,
contact your local or nearest AMETEK Service Centre or the AMETEK
factory at (403) 235-8400 or 1-800-661-9198 or contact your local AMETEK
Western Research representative. For office locations and contact infor-
mation, refer to the Offices page near the beginning of this manual, or
visit us at www.ametekpi.com.
Before contacting AMETEK with questions regarding the installation, op-
eration, maintenance, or troubleshooting of the analyzer, carefully review
the contents of this manual. If you are unable to find an explanation for
your problem in this manual, please gather the following information
prior to contacting AMETEK:
• Billing and shipping address, and telephone number.
• Model number.
• Serial number.
• Purchase order number.
• AMETEK part number for the specific component/part you are enquir-
ing about, if known.
• Detailed information about the problem(s), such as operating charac-
teristics, status conditions, etc.
IMPORTANT
NOTE
Before returning parts or equipment, you must get a Return Material
Authorization (RMA) number from your local or nearest AMETEK
Service Centre, or the factory. This will ensure your equipment is
serviced and returned to you in a prompt and efficient manner.
MINUTES OR HOURS,
WE’RE THERE FOR YOU
ANYTIME / ANYWHERE
For drawings that illustrate the location of all spare parts in the analyzer,
see “Changing Out Replaceable Parts” in Chapter 6.
Refer to “Replacing the Note 1: Source Lamp (Optical Bench) Spare Parts
In the event of natural degeneration, damage, or other failure of a
Source Lamps” in Chapter 6 NOTE
for information about replacing Source Lamp, AMETEK recommends changing out both Source
the lamps. Lamps to ensure optimal operation and to minimize additional costly
service and downtime.
Refer to “Measuring
Cell Preventive Maintenance”
Note 2: Measuring Cell Spare Parts
The Measuring Cell parts (Cell Windows and o-rings) in the “Optical
NOTE
in Chapter 6 for information Bench/Measuring Cell Spare Parts” list in this chapter apply only to
about changing out parts in standard Measuring Cells (2.5 cm to 40 cm path lengths). If the Cell
the Measuring Cell.
is a different path length – or the application is for a non-standard –
refer to an accompanying Manual Supplement or Custom Spare Parts
List for spare parts ordering numbers. These documents can be found
in the “Supplemental Information” folder of the analyzer Documenta-
tion Package CD shipped with the analyzer.
Notes: For lamp types not in this list, see Note 1, previous page.
Cl2
300-8707 Copper (Cu) Source Lamp 2 4
(Optical Bench, Lamp 1, Lamp 2)
Certain cables are used only if the analyzer has been configured with cer-
tain optional equipment.
301-1559 Customer I/O Cable, Ribbon (to Bench Controller board), Cable 2 1
Type C, 16 Conductor, 24"
301-2107 PMT Supply Board Ribbon Cable (to Bench Controller board), Cable 3 1
1.27 MM, 34 Conductor, 14"
301-2196 Optical Sensor (Opto-Coupler) Cable (to Bench Controller board), Cable 4 1
(Optional: Used only if analyzer is Multi-Species.)
301-2942 Ethernet Cable (HMI/Display Unit to Ethernet Cable on Rear Panel), Cable 7 1
301-2099 RS-485 Cable (Ethernet Module to Ethernet Cable on Rear Panel), Cable 9 1
301-2164 Lamp Socket Assembly (Source Lamps on Optical Bench to Bench Controller board) 1
Refer to the Analyzer Wiring Diagram (Sheets 1 and 2 ) in Chapter 3 for the interconnect locations of these cables.
120 V
240 V
Description (Location*) Part No. Fuse Type Part No. Fuse Type
F5 — — — —
Spare
* These fuses (F1, F2, F3, F4, F5) are located on the AC Power Distribution/MOV Board (100-2450) – see Appendix C. For
the location of this board, refer to the “Analyzer Internal Component Layout, All Options” diagram under “Installing the
Mechanical Components” in Chapter 3.
For the location of these parts in the analyzer, refer to the “Analyzer Internal Component Layout, All Options” diagram
under “Installing the Mechanical Components” in Chapter 3.
For the location of these boards in the analyzer, refer to the “Analyzer Internal Component Layout, All Options” diagram
under “Installing the Mechanical Components” in Chapter 3.
D/A digital-to-analog
DAS Data Acquisition System
DCS Distributed Control System
DP differential pressure
Motor Filter Wheel Motor, used to rotate Filter Wheel (multi-species analyzers only).
Motorized Filter Analyzer, or Motorized Optical Bench
Refers to Model 9900 Analyzers that measure multiple species, therefore requir-
ing a motorized filter wheel (or, motorized Optical Bench).
Motorless Filter Analyzer, or Motorless Optical Bench
Refers to Model 9900 Analyzers that measure a single species, therefore requir-
ing a fixed filter (a Filter Wheel is not used; therefore a motor is not required).
Multiple Species Analyzer (or Version)
See Motorized Filter Analyzer.
SD SD Memory Card
Single Species Analyzer (or Version)
See Motorless Filter Analyzer.
SO2 SO2 Low Range (when low and high range SO2 are being measured; or when
standard range SO2 is being measured)
SO2:HR SO2 High Range
SSR solid state relay
T-H (Track-Hold) This is the Hold Time (sec) for which the concentration value for
each species will be held constant while a zero or span calibration is being per-
formed on the analyzer.
This Appendix is reserved for optional equipment. For systems with op-
tional equipment, some information about certain options is included in
the main body of this manual, or in Manual Supplements that can be found
in the “Supplemental Information” folder of the Analyzer Documentation
Package CD, or in the Supplemental Information section of this manual.
Additional information about optional equipment will be included here.
Reducing the sample flow rate may reduce excessive noise on the out-
put signal when the paramagnetic O2 Sensor is used.
NOTE
Figure A-1.
Paramagnetic O2 Sensor
diagram.
Overview
One RS-485 port (RS-485:1), accessed via J501 of the UV9000 Bench Con-
troller board (100-2380), is reserved for communicating with the HMI. The
UV Bench-HMI communication link operates at 38.4 kbps, 8 data bits, 1
stop bit, and no parity. The HMI can be integrated in a Series-9900 Ana-
lyzer system or be located remotely.
The second RS-485 port (RS-485:2) can be accessed either at J502 of the
UV9000 Bench Controller board or at TB1 of the Customer I/O board
(100-2380). The port is available to communicate with a Modbus master
device (e.g., PLC and PC), or through the customer communication sys-
tem (e.g., DCS). The default communication settings are 38.4 kbps, 8 data
bits, 1 stop bit, and no parity.
Modbus Registers
Refer to the section “Analyzer Modbus Register Map” for the definitions
and descriptions of the Modbus registers available on the UV9000 analysis
subsystem. There are 439 data items that are grouped into categories:
• Analysis data (live UV concentration results)
• Photometric measurement data (absorbances and transmittances)
• Sensor measurement data (temperature, pressure, flow rate and O2
measurement results)
• Calibration status, command, configuration and results
• UV bench setup status, command, configuration and results
• UV bench operations configuration and diagnostic
• System identification and operations command and status
• Communication configuration
• Temperature control status and configuration
• Sensor measurement configuration
• Customer input/output state and configuration
• Analysis operations and configuration
• Track-hold analysis data (concentration results) and configuration
Result and status registers have read-only access; configuration and most
command registers are read-write accessible. Result and status register
values are updated once per second.
If full-function access to the Modbus registers is not required and the op-
tional HMI is installed as part of the analyzer system, Modbus-TCP/Ether-
net support with reduced functionality is provided via the HMI. Through
the HMI’s Ethernet port, all read-only registers and these command and
operation state registers are accessible:
REG# Data
(40nnn) Description \ Definition Unit Type Access
1 UV Species1 concentration * FLOAT RO
Note: Standard UV species is low-range SO2.
3 UV Species2 concentration * FLOAT RO
Note: Standard UV species is NO2.
5 UV Species3 concentration * FLOAT RO
Note: Standard UV species is NO.
7 UV Species4 concentration * FLOAT RO
Note: Standard UV species is high-range SO2.
9 UV Species5 concentration * FLOAT RO
Note: Standard UV species is extended-range SO2.
11 UV Species6 concentration * FLOAT RO
Note: Standard UV species is H2S.
13 UV Species7 concentration * FLOAT RO
Note: Standard UV species is NDr2.
15 UV Species8 concentration * FLOAT RO
Note: Standard UV species is NDr1.
17 Aggregated UV species concentration * FLOAT RO
Note:
1) When NOx measurement is enabled in the Algorithm bitmask (REG#321.b3),
the aggregated UV species is NOx, which is the sum of UV Species2 (NO2) and
Species3 (NO) concentrations.
2) When NOx measurement is not enabled in the Algorithm bitmask, the UV species
to be aggregated is defined by the Aggregated UV species inclusion bitmask
(REG#761) and the Species Factors (REG#802-816).
19 Lamp1-Filter1 absorbance - FLOAT RO
21 Lamp1-Filter2 absorbance - FLOAT RO
23 Lamp1-Filter3 absorbance - FLOAT RO
25 Lamp1-Filter4 absorbance - FLOAT RO
27 Lamp1-Filter5 absorbance - FLOAT RO
29 Lamp1-Filter6 absorbance - FLOAT RO
31 Lamp2-Filter1 absorbance - FLOAT RO
33 Lamp2-Filter2 absorbance - FLOAT RO
35 Lamp2-Filter3 absorbance - FLOAT RO
37 Lamp2-Filter4 absorbance - FLOAT RO
39 Lamp2-Filter5 absorbance - FLOAT RO
41 Lamp2-Filter6 absorbance - FLOAT RO
43 Lamp1-Filter1 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter1 zero-transmittance.
45 Lamp1-Filter2 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter2 zero-transmittance.
47 Lamp1-Filter3 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter3 zero-transmittance.
49 Lamp1-Filter4 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter4 zero-transmittance.
51 Lamp1-Filter5 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter5 zero-transmittance.
53 Lamp1-Filter6 transmittance - FLOAT RO
Note: Normalized to Lamp1-Filter6 zero-transmittance.
nication with the analyzer is made to the HMI’s Ethernet port. When the status bit
is raised, all other status bits in StatusCode1 (REG#245), StatusCode2 (REG#246)
and StatusCode3 (REG#247) are undertermined.
248 System hardware presence bitmask - UINT16 RO/RW
b0: cell temperature sensor type
0 = PNP sensor selected; 1 = RTD probe selected
b1: customer IO board
0 = board not detected; 1 = board detected
b2: sample system interface board
0 = board not detected; 1 = board detected
b3…b7 = reserved
b8: sample conditioning columns
0 = not installed; 1 = installed
b9: oven heater RTD
0 = not installed; 1 = installed
b10...b11: O2 sensor type
0 = Hummingbird Pm1158
1 = Figaro KE-25
2 = Hummingbird Paracube-Delta
3 = reserved
b12…b14 = reserved
b15: without UV bench
0 = false; 1 = true
Note: 1) Lower 8 bits are read-only as the presence of the corresponding hardware
is detectable by the analyzer firmware.
2) Upper 8 bits are read-\write-able to inform the analyzer firmware of the corre-
sponding hardware being installed.
3) Only O2 sensor types ‘0’ and ‘1’ are currently available.
4) Bits 8-9 and 12-14 are not applicable.
249 Modbus slave node address - UINT8 RW
Note: Default for node address for UV9900 is 101.
250 DCS Modbus port baud rate - UINT8 RW
0 = 38400 (default setting)
1 = 19200
2 = 9600
3 = 4800
Note:
Change to DCS Modbus port baud rate, parity or stop bits takes effect on next reset.
(DCS Modbus port is J502 of Bench Controller board or TB1 of Customer I/O board.)
251 DCS Modbus port parity - UINT8 RW
0 = none (default setting)
1 = odd
2 = even
This Appendix includes most of the drawings that are included in the
main body of this manual, plus additional drawings not included in the
main body.
TB-A
ANALYZER CALIBRATION STATUS
TB3
DRY CONTACT INPUTS
TB2
ANALOG OUTPUTS
HMI
TB1
RS-485 COMMUNICATION
TB7
RELAY CONTACTS
TB6
DRY CONTACT INPUTS
USB2
24 VDC Display
USB1
Power Connector
The Ethernet Port on the bottom of the HMI is shown for information
purposes only. The customer Ethernet connection is on the analyzer
NOTE
Rear Panel. See drawing “Analyzer Rear Panel Layouts (GP, Div 2/
Zone 2) (100-2421-1F)” in this appendix.
OPTICAL BENCH
OVERTEMP SWITCH
(301-3329)
TEMP TRANSISTOR
(100-2436)
(THREADED INTO
UNDERSIDE OF
PMT PLATE)
SUPPLEMENTAL INFORMATION
This section consists of documents that are not part of the main manual,
but which describe and illustrate installation, operation, layout, and mainte-
nance procedures for non-standard or optional equipment – and derivative
analyzer models – that make up your analyzer and its sample system.
If you order a hard copy of the analyzer User Manual, this information will
be included in this section. If you receive only the standard Analyzer Docu-
mentation Package CD (all files in PDF format), this information is included
in the Supplemental Information folder on the CD. “Supplemental Infor-
mation – Where Can I Find It?” in Chapter 1 describes these documents.
• Manual Supplements
• 9900_CFG Report