Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
10 – margin of error
Beer-Lambert’s Law
Approximate Color of Light Color of Light - determines the concentration of the absorbing species
Wavelength (nm) Absorbed Reflected in the analyte (substance of interest) solution
400-435 violet green-yellow - states that absorbance is directly related to the
435-500 blue yellow concentration if the measurements are made at a fixed
wavelength in a cell of constant path which is usually 1
500-570 green red
cm.
570-600 yellow blue - absorbance of a material sample is directly
600-630 orange green-blue proportional to its thickness
630-700 red green may be expressed by:
The complementary color of the light you see is the actual 𝐴
color. 𝐴 = Ɛ𝑐𝑙 or 𝑐 = Ɛ𝑙
Spectral Absorbance Curve
A = measured absorbance (no unit)
Ɛ = molar extinction coefficient (M-1cm-1)
c = concentration in molarity
l = path length (usually 1cm); the length of solution the light
passes through
𝐴𝑠 𝐴𝑢 𝐴𝑢𝐶𝑠
𝐶𝑠
= 𝐶𝑢
or 𝐶𝑢 = 𝐴𝑠
Where:
Cu = concentration of the unknown solution
Cs = concentration of the standard solution
Parts Au = absorbance of the unknown solution
1. White Light Source As = absorbance of the known solution
2. Monochromator – disperse light into its compartment
wavelengths Calibration or Standard curve
3. Sample Compartment – holds the transparent tube or - preferable to construct when a large number of
cell containing the sample samples are being run
o Cuvette – holds the sample - obtained by measuring the absorbance of a series of
- Fill the cuvette ¾ full standard solutions whose concentration is in the range
- Quartz cuvette has 3mL max of the solution being measured
capacity - each standard solution must be prepared in exactly the
- Has a clear and opaque side same way as the solution being measured
Clear side nakatapat sa *picture of a graph*
optical path because light
pass through this side Graphical Method preparation
Opaqe side ang pwede 1. compute the concentration of the stock solution
lang mahawakan 2. make the proper dilution of the stock solution to make
4. Detector – measure the amount of light transmitted standard solutions of known concentration
5. Read-out device – display the reading 3. Read the absorbance of the standard solutions
4. Plot the absorbance (y-axis) and the concentration of
the standard solutions (x-axis) IMPORTANT POINTS TO CONSIDER IN UV-VIS ANALYSIS
5. Determine the best-fit line 1. Errors in UV-VIS analysis are sometimes due to the
6. This may now be used to determine the concentration inadequate mixing of reagents. Solutions should be
of the sample solution well-mixed prior to absorbance reading
2. The color formation in some assays takes place largely
SPECTROPHOTOMETERS during the first few minutes after mixing. Enough time
Spectronic Genesys 5 should be given for the complete color development
How To Use: of the solution before the absorbance is measured
1. make sure that no samples are in the optical path and 3. Assays should be performed in duplicate or triplicate,
that the door of the samplec compartment is closed and the individual values, not the means, should be
completely plotted. This procedure allows the experimenter to
2. plug the instrument into a 220-volt- outlet omit erroneous values from the calibration curve
3. press the power switch located at the lower left side 4. The formation of bubbles, turbidity, fingerprint, or
of the instrument condensation on or inside the cuvette should be
4. let the instrument warm up for about 5 minutes. While avoided because it diminishes the accuracy of the
warming up, it also runs a diagnostic test on its readings
internal parts 5. The “lines of best fit” should be drawn through the
5. when the main menu appears, press 1 for absorbance data points. It is not necessarily the line that passes
or transmittance measurement through the origin and the other data point
6. press GO TO WL to select desired wavelength 6. The Beer-Lambert’s Law is most accurate between the
7. place the blank reagent in the cell holder B and the absorbance of 0.05 to 0.70. Above 70, the measured
solutions in the succeeding cell holders absorbance underestimates the real absorbance.
o the clear side of the cuvette should face the Below 0.05, the absorbance of the instrument is not
light beam and should be free from accurate
fingerprints 7. The calibration curve should never be diluted so that
8. use cell or to move the sample into the light its absorbance falls within the calibration curve
beam
9. take the reading APPLICATIONS OF UV-VIS SPECTOPHOTOMETRY
10. switch off the instrument then unplug it Reaction kinetics
Quantitation
Multiskan GO pKa determination
- measures a UV/IS absorbance from 200 to 1000nm DNA structural information
on 96- or 384-well microplates
- can incubate up to 45oC and can shake the The Micropipettor
microplate - Used to transfer small amounts (<1 mL) of liquids
- microplates allow up to 96 or 384 smaller sample - Used in conjunction with disposable plastic tips
volumes (in microliter) to be read in one run Proper Pipetting Techniques
- can also be used with a cuvette *picture*
How To Use 1. Never exceed the upper or lower limits of the pipettor
1. Press the START/ON key, the instrument will perform 2. Set the desired volume by turning the rings clockwise to
self-diagnostics increase volume or counterclockwise to decrease
2. When the screen flashes “Self-diagnostics passed”, volume
the machine is ready to use 3. Place the tip on the discharge end of the pipettor
3. Press the IN/OUT key to bring out the plat carrier 4. The plunger will stop at two different stopping points:
4. Press the microplate containing the samples to be i. The level of depression that will result in the
read properly on the plate carrier desired volume of solution
5. Pres the IN/OUY key again to close the plate carrier ii. Used for the complete discharging of solutions
6. Enter the parameters using the keypad from the tip
7. Press START/ON key to start the instrument o The second stop should not be reached when
8. Save the result with the appropriate file name. The drawing liquid into the pipettor
file may also be exported using a USB 5. Depress the plunger until the first stopping point and
9. Press the IN/OUT key to draw out the carrier and insert the tip into the solution
remove the microplate 6. Carefully and slowly release the plunger
10. Switch off the instrument by pressing and holding the 7. Discharge the solution into the appropriate container
STOP/OFF key for a few seconds by depressing the plunger to the first stop, wait one
second, and then continue pressing as far as it will go
to discharge the entire volume solution
TWO-FOLD/ SERIAL DILUTION 8. Remove the tip by pressing down on the ejector button
- A stepwise dilution method of a substance in solution
- Reduces the concentration of the substance by a factor Notes:
of two, resulting in a geometric progression (eg. ½, ¼,
1/8…) of the concentration in a logarithmic fashion Never point a pipettor up
*picture* When withdrawing liquids, always release the
*shake the cuvette for a homogenous mixture and change the tip plunger slowly
of the micropipettor for every container Be sure to use the proper size tip for each pipettor
Always use a new tip for each different liquid
Use the correct pipettor for the volume that is to
be dispensed
Solutions
- Homogenous mixture
- Composed of solute and solvent
Percentage Concentration
- “by the hundred” or “in a hundred”
- Amount of solute in a solution
- Types of percentage solutions:
o %w/w, %w/v, %v/v