"The final acceptance of a measurement system should not come down to a single set of indices.
The performance of the measurement system should also be reviewed using graphical analyses."
For measurement systems whose purpose is to analyze a process, general rules of thumb are:
Acceptability Criteria for %R&R (% of Total Variation and % of Tolerance)
ü Under 10% - acceptable system - Under 5% preferred
ü 10% to 30% - may be acceptable depending on importance of the application, cost of the system, cost of repair, etc
ü Over 30% - considered not acceptable - every effort should be made to improve the system
Based on the AIAG Measurement Systems Analysis Reference Manual -- Call 810-358-3570 to order.
12
R&R Plot - by Parts
10
0
A B C A B C A B C A B C A B C A B C A B C A B C A B C A B C
1 1 1 2 2 2 3 3 3 4 4 4 5 5 5 6 6 6 7 7 7 8 8 8 9 9 9 0 0 0
12
R&R Plot - by Appraisers
10
0
A A A A A A A A A A B B B B B B B B B B C C C C C C C C C C
1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0
10
0
A A A A A A A A A A B B B B B B B B B B C C C C C C C C C C
1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0
12
Part by Appraiser Plot
10
Average
0
1 2 3 4 5 6 7 8 9 10
PART
A B C