GS 12A01A02-01E
n General
The model FLXA21® 2-Wire Analyzer, one model
of FLEXA® series, offers single or dual sensor
measurement. The modular-designed analyzer offers 4
kinds of measurements – pH/ORP (oxidation-reduction
potential), contacting conductivity (SC), inductive
conductivity (ISC) or dissolved oxygen (DO) – with the
respective sensor module.
For dual sensor measurement, the combination of
two same type sensor inputs – pH/ORP and pH/ORP
(analog sensor only), SC and SC, and DO and DO –
are available with two sensor modules. Dual sensor
measurement offers additional functionalities; calculated
data function and redundant system.
Variety of calculated data from two measuring
parameters is selectable for each measurement.
On the redundant system built on two measuring
parameters of two sensor inputs, main output
parameter is automatically switched over to the
second sensor output in case of the main sensor’s
failure condition.
Addition to conventional analog pH/ORP sensors, the
analyzer FLXA21 can be connected to Yokogawa’s
digital sensor, FU20F pH/ORP SENCOM Sensor.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
3
2-3. Inductive Conductivity (ISC) ■ Output Range
■ Input Specification DO concentration:
Compatible with the Yokogawa inductive conductivity mg/l (ppm):
ISC40 series with integrated temperature sensor: min.: 1 mg/l (ppm)
NTC30k or Pt1000. max.: 50 mg/l (ppm)
ppb:
■ Input Range min.: 1 ppb
Conductivity: 0 to 2000 mS/cm at 25 ºC reference max.: 9999 ppb
temperature. % saturation:
Temperature: -20 to 140 ºC min.: 10 %
Cable length: max.: 600 %
max. 60 meters total length of fixed sensor Temperature:
cable + WF10(J) extension cable. min. span 25 ºC
Influence of cable can be adjusted max. span 170 ºC
by doing an AIR CAL with the cable
connected to a dry cell. ■ Performance (Accuracy)
(The specifications are expressed with simulated
■ Output Range inputs.)
Conductivity: Performance in ppm mode:
min. span: 100 µS/cm Linearity: ±0.05 ppm or ±0.8% F.S., whichever is
max. span: 2000 mS/cm (max 90% zero greater
suppression) Repeatability: ±0.05 ppm or ±0.8% F.S., whichever
Temperature: is greater
min. span 25 ºC Accuracy: ±0.05 ppm or ±0.8% F.S., whichever is
max. span 160 ºC greater
■ Performance (Accuracy) Performance in ppb mode:
(The specifications are expressed with simulated Linearity: ±1 ppb or ±0.8% F.S., whichever is
inputs.) greater
(Output span is 0-100 µS/cm or more) Repeatability: ±1 ppb or ±0.8% F.S., whichever is
Conductivity: greater
Linearity: ±(0.4 %F.S. + 0.3 µS/cm) Accuracy: ±1 ppb or ±0.8% F.S., whichever is
Repeatability: ±(0.4 %F.S. + 0.3 µS/cm) greater
Temperature: ±0.3 ºC Temperature
Step response: 90 % (< 2 decades) in 8 seconds Linearity: ±0.3 ºC
Note: "F.S." means maximum setting value of analyzer Repeatability: ±0.1 ºC
output. Accuracy: ±0.3 ºC
2-4. Dissolved Oxygen (DO) Note: "F.S." means maximum setting value of analyzer
output.
■ Input Specification
The FLXA21 accepts output from membrane covered 2-5. pH/Oxidation-reduction Potential (pH/ORP)
Dissolved Oxygen sensors. These sensors can with digital sensor, FU20F pH/ORP SENCOM
be Galvanic type, where the sensor generates its Sensor
own driving voltage or Polarographic type, where ■ Input Specification
the sensor uses external driving voltage from the Bi-directional digital communication (RS-485)
converter. between FU20F and FLXA21
The input range is 0 to 50 µA for Galvanic sensors ■ Input Range (depending on FU20F)
and 0 to 1 micro A for Polarographic sensors. pH: 0 to 14 pH
For temperature compensation, the FLXA21 accepts ORP: -1500 to 1500 mV
Pt1000 (DO30 sensor) and NTC22k elements rH: 0 to 100 rH
(OXYFERM and OXYGOLD sensors). Temperature: -10 to 105 ºC
■ Input Range ■ Output Range
DO30 sensor: pH: min. span 1 pH
Dissolved Oxygen: 0 to 50 mg/l (ppm) max. span 20 pH
Temperature: -20 to 150 ºC ORP: min. span 100 mV
Note: Process temperature for DO30 is 0 to 40 ºC max. span 3000 mV
Hamilton sensors: rH: min. span 2 rH
Oxyferm: max. span 100 rH
Measurement range: 10 ppb to 40 ppm Temperature: min. span 25 ºC
Temperature range: 0 to 130 ºC max. span 170 ºC
Oxygold G:
Measurement range: 2 ppb to 40 ppm
Temperature range: 0 to 130 ºC
Oxygold B:
Measurement range: 8 ppb to 40 ppm
Temperature range: 0 to 100 ºC
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
4
3. Electrical Message language:
12 (English, Chinese, Czech, French,
■ Output Signal
German, Italian, Japanese, Korean,
General: One output of 4-20 mA DC
Note: Tolerance: ±0.02 mA Polish, Portuguese, Russian and Spanish)
Bi-directional HART digital communication, One analyzer has all 12 languages.
Note: Description for a selection of language and
superimposed on mA (4-20mA) signal language names are written in English.
Output function: Note: Only English alphabet and numeric are available
Linear or Non-linear (21-step table) for a tag number, an additional description for each
Burn out function: (NAMUR 43 except ISC) value on the display screen and passwords.
Without HART/PH201G: Note: Only for message language on the screen, 12
Down: 3.6 mA languages are provided.
(signal: 3.8 to 20.5 mA for pH/ORP, SC
and DO) 4. Mechanical and others
(signal: 3.9 to 20.5 mA for ISC) ■ Housing
Up: 22mA Case: • Plastic (Polycarbonate)
With HART/PH201G: • Stainless steel without painting
Down: 3.6 mA for pH/ORP, SC and DO • Stainless steel with epoxy coating
Down: 3.9 mA for ISC • Stainless steel with urethane coating
(signal: 3.8 to 20.5 mA for pH/ORP, SC • Stainless steel + high anti-corrosion coating
and DO) Case color and finish:
(signal: 3.9 to 20.5 mA for ISC) Color: Silver gray (equivalent to Munsell
3.2PB7.4/1.2)
Up: 22mA (for plastic case, stainless steel cases
■ Power Supply with coating)
Nominal 24 V DC loop powered system Finish: Electropolishing (for stainless steel
One (1) Sensor module (1 input): case without painting)
16 to 40V DC (for pH/ORP (analog Window: Polycarbonate (flexible)
sensor), SC and DO) Window frame for stainless steel cases:
17 to 40V DC (for ISC) Polycarbonate, color: silver gray
21 to 40V DC (for pH/ORP SENCOM (equivalent to Munsell 3.2PB7.4/1.2)
sensor) Protection: IP66 (except Canada), Type 4X (except
Canada), Type 3S/4X (Canada)
Two (2) Sensor modules (2 inputs):
■ Plate
22.8 to 40V DC (for pH/ORP (analog sensor),
Main name plate: inside case cover
SC and DO) Regulation plate:
Note: When the FLXA21 is used in the multi-drop mode of
HART communication, the output signal is changed on the case outside
from 12.5 mA DC to 4 mA DC just after the power is ■ Cable and Terminal
turned on. Enough power supply for the instruments Cable size:
is to be provided. Outer diameter:
6 to 12 mm (suitable for M20 cable gland)
● Maximum Load Resistance 3.4 to 7 mm (grounding cable for plastic case)
pH/ORP (analog sensor), SC and DO: Terminal screw size: M4
Refer to the Figure 1. torque of screw up: 1.2 N•m
ISC and pH/ORP SENCOM sensor: Wire terminal:
Refer to the Figure 2. Pin terminal, ring terminal and spade
terminal can be used for analyzer’s power
■ Display supply terminals and sensor terminals.
LCD with a touch screen: For the grounding terminal on the stainless
Black/White: 213 x 160 pixels steel case, ring terminal should be used.
Contrast adjustment available on the touch screen Pin terminal: pin diameter: max. 1.9 mm
Ring and spade terminal: width: max. 7.8 mm
2-sensor measurement
1295
Load resistance (Ω)
1295
Load resistance (Ω)
V - 11.5
V - 11.5 R=
R= 0.022
0.022
1000
1000
Except SENCOM
600
600
516
Digital Communication
Digital Communication Range (HART)
Range (HART) 304
250
250
0 17 18 22.86 24.7 40
0 16 17 18 22.8 24.7 40
18.2 21 Voltage (V)
Voltage (V)
Figure 1 Supply Voltage and Load Resistance Figure 2 Supply Voltage and Load Resistance
for pH/ORP (analog sensor), SC and DO for ISC and pH/ORP SENCOM sensor
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
5
■ Cable Entry User's Manual
Plastic case: written in English
1-Sensor measurement: Safety Regulations Manual
3 holes, for European region
M20 cable gland x 3 pcs, written in 25 languages
Sleeve x 1 pc (for grounding cable line) General Specifications
2-Sensor measurement: written in English
4 holes,
M20 cable gland x 4 pcs, Technical Information
Sleeve x 1 pc (for grounding cable line) for HART Communication
Stainless steel case: written in English
3 holes, User Setting Table
M20 cable gland x 3 pcs of 5 kinds of measurement/sensor type
Close up plug x 1 pc written in English
Note: Cable gland and plug are delivered with an analyzer, ■ Regulatory Compliance
but not assembled into the analyzer. Safety: UL 61010-1
■ Mounting UL 61010-2-030
Mounting hardware (option):
CAN/CSA C22.2 No.61010-1
• Universal mounting kit (Note)
• Pipe and wall mounting hardware EMC: EN61326-1 Class A,Table 2 (For use in
• Panel mounting hardware industrial locations)
Note: This kit contains the pipe and wall mounting EN61326-2-3
hardware and the panel mounting hardware. AS/NZS CISPR11
Hood (option): Korea Electromagnetic Conformity
• Stainless steel Standard Class A 한국 전자파적합성 기준
• Stainless steel with urethane coating Installation altitude: 2000 m or less
• Stainless steel with epoxy coating
Category based on IEC 61010: I (Note 1)
■ Stainless Steel Tag Plate Pollution degree based on IEC 61010: 2 (Note 2)
When the additional code “/SCT” with a tag number Note 1: Installation category, called over-voltage category,
is specified, the tag plate on which the tag number is specifies impulse withstand voltage.
inscribed is delivered with the analyzer. Equipment with "Category I" (ex. two wire
Tag plate is hanging type. transmitter) is used for connection to circuits in
which measures are taken to limit transient over-
■ Conduit Adapter voltages to an appropriately low level.
Using optional adapter Note 2: Pollution degree indicates the degree of existence
• G1/2 (quantity: 4) of solid, liquid, gas or other inclusions which may
• 1/2NPT (quantity: 4) reduce dielectric strength. Degree 2 is the normal
• M20 x 1.5 (quantity: 4) indoor environment.
These conduit adapters are delivered with an Explosion-proof (Intrinsically safe type
analyzer, but not assembled into the analyzer. and non-incendive) (for suffix code: -EA, -ES):
■ Size of Housing Case ATEX Intrinsically safe approval
Plastic: 144 x 144 x 151 mm (L x W x D) (without Applicable standard
cable gland) Explosive Atmospheres
Stainless steel case: EN 60079-0:2012/A11: 2013 Equipment -
165 x 165 x 160 mm (L x W x D) (without General requirements
cable gland) EN 60079-11:2012 Equipment protection by
Intrinsic safety “i”
■ Weight EN 60079-26:2007 Equipment with
Approx. 1 kg (Plastic housing) equipment protection level (EPL)
Approx. 2 kg (Stainless steel housing) Ga
■ Shipping Details EN 60529:1992 Degrees of protection
Package size: provided by enclosures (IP Code)
Approx. 340 x 340 x 370 mm (L x W x H) Type of protection
■ Ambient Operating Temperature II 1G Ex ia IIC Ga
-20 to +55 ºC Group: II
Category: 1G
■ Storage Temperature
T4: for ambient temperature:–20 to 55ºC
-30 to +70 ºC
T6: for ambient temperature:–20 to 40ºC
■ Humidity (excluding SENCOM module)
10 to 90% RH at 40ºC (Non-condensing) Atmosphere pressure: 80kPa
■ Document (0.8bar) to 110kPa (1.1bar)
Following documents are delivered with an analyzer; Degree of Protection of the
Paper copy: Enclosure: IP66
Start-up Manual IECEx Intrinsically safe
written in English Applicable standard
Safety Precautions IEC 60079-0: 2011 Part 0: Equipment -
written in English General requirements
CD-ROM: IEC 60079-11: 2011 Part 11: Equipment
Start-up Manual protection by Intrinsic safety “i”
written in English
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
6
IEC 60079-26: 2006 Part 26: Equipment CSA Intrinsically safe and nonincendive approval
with equipment protection level Applicable standard
(EPL) Ga CAN/CSA C22.2 No. 94-M1991 Special
IEC 60529: 2001 Degrees of protection Purpose Enclosures
provided by enclosures (IP Code) CAN/CSA C22.2 No. 157-92
Type of protection Intrinsically Safe Equipment for
Ex ia IIC Ga Use in Hazardous Locations
T4: for ambient temperature:–20 to 55ºC C22.2 No213-M1987 Non-incendive
T6: for ambient temperature:–20 to 40ºC Electrical Equipment for Use in
(excluding SENCOM module) Class I, Division 2 Hazardous
Atmosphere pressure: 80kPa Locations
(0.8bar) to 110kPa (1.1bar) CAN/CSA-E60079-0-07 Electrical
Degree of Protection of the apparatus for explosive gas
Enclosure: IP66 atmospheres - Part 0: General
FM Intrinsically safe and nonincendive approval requirements
Applicable standard CAN/CSA-E60079-11-02 Electrical
FM-3600: 2011 Approval Standard for apparatus for explosive gas
Electric Equipment for use in atmospheres - Part 11: Intrinsic
Hazardous (Classified) Locations safety “i”
General Requirement IEC 60529:2001 Degrees of protection
FM-3610: 2010 Approval Standard for provided by enclosures (IP Code)
Intrinsically Safe Apparatus and Type of protection (C22.2)
Associated Apparatus for Use Class I, Division 1, Groups A, B, C and D
in Class I, II, and III, Division 1, (Intrinsically Safe)
Hazardous (Classified) Locations Class I, Division 2, Groups A, B, C and D
FM-3611: 2004 Nonincendive Electrical (Nonincendive)
Equipment for Use in Class I For all protection type,
and II, Division 2 and Class III, T4: for ambient temperature: -20 to 55°C
Divisions 1 and 2, Hazardous T6: for ambient temperature: -20 to 40°C
(Classified) Locations Atmosphere pressure: 80 kPa (0.8 bar) to
FM-3810: 2005 Electrical Equipment 110 kPa (1.1 bar)
for Measurement, Control and Ambient Humidity: 0 to 100% (No condensation)
Laboratory Use Degree of Protection of the Enclosure: Type 4X
NEMA 250:1991 Enclosures for Electrical Type of protection (E60079)
Equipment (1000 Volts Maximum) Ex ia IIC
ANSI/IEC 60529:2004 Degrees of T4: for ambient temperature: -20 to 55°C
protection provided by enclosures T6: for ambient temperature: -20 to 40°C
(IP Code) Atmosphere pressure: 80 kPa (0.8 bar) to
ANSI/ISA 60079-0 2009 Part 0: General 110 kPa (1.1 bar)
Requirements Ambient Humidity: 0 to 100% (No condensation)
ANSI/ISA 60079-11 2011 Part 11: Equipment Degree of Protection of the Enclosure: IP66
protection by intrinsic safety “i” NEPSI Intrinsically safe approval
Type of protection Applicable Standard
Class I, Division 1, Groups A, B, C and D GB 3836.1-2010 Explosive atmospheres-
(Intrinsically Safe) Part 1: Equipment - General
Class I, Division 2, Groups A, B, C and D requirements
(Nonincendive) GB 3836.4-2010 Explosive atmospheres-
Class I, Zone 0, in Hazardous (Classified) Part 4: Equipment protection by
Locations (Intrinsically Safe) intrinsic safety “i”
Class I, Zone 2, Group IIC, in Hazardous GB 3836.20-2010 Explosive atmospheres-
(Classified) Locations Part 20: Equipment with
(Nonincendive) equipment protection level (EPL)
AEx ia IIC Ga
For all protection type, Type of protection
T4: for ambient temperature: -20 to 55°C Ex ia IIC Ga
T6: for ambient temperature: -20 to 40°C T4: for ambient temperature: -20°C to 55°C
Atmosphere pressure: 80 kPa (0.8 bar) to T6: for ambient temperature: -20°C to 40°C
110 kPa (1.1 bar) Atmosphere pressure: 80kPa (0.8bar) to
Degree of Protection of the Enclosure: 110kPa (1.1bar)
NEMA Type 4X and IP66 Degree of Protection of the Enclosure: IP66
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
7
Electrical Parameters (Ex ia) Control Drawing (ATEX and IECEx types)
Each housing assembly (base module) and each
sensor module are respectively certificated. Hazardous Location ← → Non Hazardous Location
Input parameters of sensor module meet output
+
parameters of housing assembly. Sensor 1
Measuring module 1 Supply +
Housing Assembly Safety Barrier
Housing assembly Measuring module 2 Supply -
(Note 1)
Sensor 2 -
Supply and output circuit (terminals + and -):
Ui, Vmax = 30 V (*)
Input Ii, Imax = 100 mA
Pi, Pmax = 0.75 W Note: The measuring module on this drawing means
parameters Ci =13 nF the sensor module on this General Specifications.
Li = 0 mH
(Linear source)
Electrical data are as follows;
Measuring module input circuit (CN2 or CN3 on Supply and output circuit (Terminals Supply + and –):
Back board)
Uo Vt, Voc = 13.65 V Maximum Voltage (Ui) = 30V
Output
Io, It, Isc = 50 mA Maximum Current (Ii) = 100mA
parameters Po = 0.372 W Maximum Power (Pi) = 0.75W
Co, Ca = 80 nF Internal Capacitance (Ci) = 13nF
Lo, La = 7.7mH
Internal Inductance (Li) = 0mH
pH/ORP module, SC module, and DO module
Sensor input circuit (pH: terminals 11 through 19, SC:
terminals 11 through 16, DO: terminals 11 thourgh 18,
Ui, Vmax = 13.92 V ISC: terminals 11 through 17, SENCOM: terminals
Input Ii, Imax = 50 mA
Pi, Pmax = 0.374 W 82, 83, 84, 86 and 87):
parameters Ci = 40 nF Type of Measuring Module pH, SC and DO ISC SENCOM
Li = 2.9 mH
Maximum Voltage (Uo) 11.76 V 11.76 V 5.36 V
Sensor input circuit (pH: terminals 11 through Maximum Current (Io) 116.5 mA 60.6 mA 106.6 mA
19, SC: terminals 11 through 16, DO: terminals
11 through 18) Maximum Power (Po) 0.3424 W 0.178 W 0.1423 W
Output Uo Vt, Voc = 11.76 V External Capacitance (Co) 100 nF 100 nF 31 µF
parameters Io, It, Isc = 116.5 mA
Po = 0.3424 W External Inductance (Lo) 1.7 mH 8 mH 0.45 mH
Co, Ca = 100 nF
Lo, La = 1.7mH Note 1: In any safety barrier used, the output current
must be limited by a resistor “R” such that
Imaxout=Uz/R.
ISC module Note 2: The safety barrier shall be certified by notify body
Ui, Vmax = 13.92 V EU as ATEX.
Ii, Imax = 50 mA Note 3: When using non isolation barrier connect (*1) to IS
Input
Pi, Pmax = 0.374 W earthing system.
parameters Ci = 40 nF Note 4: Sensor 1 and Sensor 2 shall be of passive types
Li = 7.7 mH to be regarded as ‘simple apparatus’ or the ones
Sensor input circuit (terminals 11 through 17) individually certified with relevant parameters.
Uo Vt, Voc = 11.76 V Note 5: Measuring module 2 may not mounted. As for ISC
Output Io, It, Isc = 60.6 mA module and SENCOM module, only one can be
parameters Po = 0.178 W mounted.
Co, Ca = 100 nF Note 6: Measuring module is placed in an enclosure with
Lo, La = 8 mH IP20 and over.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
8
Control Drawing (FM type) Control Drawing (CSA type)
Following contents refer “DOC. No. IKE039-A12”
Intrinsically Safe Non-incendive
Class I, Division 1, Groups A, B, C, and D Group IIC, Zone 0 Class I, Division 2,
Class I, Zone 0 and 1, Group IIC Class I, Division 1 Groups A, B, C, D
T4 for Ta = 55°C, T6 for Ta = 40°C Hazardous Location ← → Non Hazardous Location
Hazardous Location ← → Non Hazardous Location 2 wire analyzer
2 wire analyzer +
Measuring module 1 Supply +
Measuring module 1 Supply + + Sensor 1
Sensor 1 Housing Assembly Safety Barrier
Housing Assembly Safety Barrier Supply - Refer to Note
(Note 1) Sensor 2 Measuring module 2 -
Measuring module 2 Supply -
Sensor 2 -
(Refer to Note 7)
(*)
(Refer to Note 7)
(*) Not use safety barrier but
Note: The measuring module on this CSA certified equipment
Note: The measuring module on this drawing means drawing means the sensor module use in Non-incendive
the sensor module on this General Specifications. on this General Specifications.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
9
3: Input voltage of the safety barrier must be less than 5. Digital Communication
250Vrms/Vdc ■ Kind of Digital Communication
4: When using non isolation barrier connect (*) to IS
earthing system.
• HART or PH201G dedicated distributor
Note: Only one kind of digital communication is available
5: pH and SC Sensor(s) are of a passive type to
for one analyzer.
be regarded as ‘simple apparatus’ same as
06ATEX0218X, 06ATEX0219, IECEx KEM ■ Output Value Parameter (HART)
06.0052X, FM3028779, 06ATEX0220X, Four value parameters (measured values) are
06ATEX0221, IECEx KEM 06.0053X or the one available for one digital communication.
individually certified with relevant parameters. • For 1-sensor measurement, these parameters are
6: ISC Sensor(s) are ISC40S of 00ATEX1067X or the measured values.
one individually certified with relevant parameters.
7: DO Sensor(s) are of a passive type to be regarded
• For 2-sensor measurement, refer to the next item.
as ‘simple apparatus’ or the one individually certified ■ Digital Communication of 2-Sensor
with relevant parameters. Measurement (HART)
8: Measuring module 2 may not mounted. As for ISC Even when two sensor modules are installed, only
module, only one can be mounted. one digital communication is available for 2-sensor
9: Installation should be in accordance with Canadian measurement.
Electrical Code Part I and Local Electrical Code.
10: Do not alter drawing without authorization from CSA.
Four value parameters can be selected from the
Note for Nonincendive Installation: followings;
1: The parameter for sensor input circuit must be taken Measured values of two sensors
into account when installed. Calculated data of 2-sensor measurement
2: Installation should be in accordance with Canadian Redundant system output
Electrical Code Part I and Local Electrical Code. ■ Specific Contact Output with dedicated
3: Do not alter drawing without authorization from CSA.
distributor, model PH201G (Style B)
Control Drawing (NEPSI types) The distributor, model PH201G, is designed to
connect with the 2-Wire Analyzer.
Hazardous Location ← → Non Hazardous Location This distributor supplies drive power to the analyzer
+
and receives simultaneously 4-20 mA DC signal from
Measuring module 1 Supply + the analyzer.
Sensor 1
Housing Assembly Safety Barrier
(Note 1) This signal is converted to 1-5 V DC signal in the
Sensor 2 Measuring module 2 Supply - - distributor.
This distributor also receives digital signals
(*) superimposed on the 4-20 mA DC signal, and
Note: The measuring module on this drawing means
provides contact outputs
the sensor module on this General Specifications. Input/Output signal:
Number of available drive/signal point: 1
Electrical data are as follows; Output signal: 1-5 V DC (2 points) (Note)
Maximum Voltage (Ui) = 30V Load resistance: 2 kΩ or less (1-5 V DC output)
Maximum Current (Ii) = 100mA Isolation system: Loop isolation type
Maximum Power (Pi) = 0.75W Note: Two output signals for one analyzer's analog output
Internal Capacitance (Ci) = 13nF are provided. Two 1-5 V DC output signals are same.
Internal Inductance (Li) = 0mH Contact output:
Note 1: The output current must be limited by a resistor “R” Contact rating:
such that Imaxout=Uz/R (linear source). 250 V AC, maximum 100 VA
Note 2: Safety barrier certified by NEPSI should be used. 220 V DC, maximum 50 VA
Note 3: When using non isolation barrier, connect (*) to IS Hold contact output:
earthing system. NC contact, normally energized
Note 4: Measuring module 2 is installed when required.
Contact closes when power is off or
When measuring inductive conductivity, only one
module can be installed. during Hold situation.
Fail contact output:
NC contact, normally energized
Contact closes when power is off or
during Fail/Warning conditions.
Wash contact output:
NO contact
Contact closes during wash cycles.
Regulatory Compliance
Korea Electromagnetic Conformity Standard Class A
한국 전자파적합성 기준
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
10
6. Model & Suffix Codes
Model Suffix code Option code Description
FLXA21 ······················································································· ···················· 2-Wire Analyzer
Power supply -D ···················· Always -D
Housing -P ················· Plastic
-S ················· Stainless steel
-U ················· Stainless steel + urethane coating
-E ················· Stainless steel + epoxy coating
-W ················· Stainless steel + high anti-corrosion coating (Note 11)
Display -D ···················· Anti-glare LCD
Type -AA ················· General purpose
-EA ················· ATEX, IECEx, FM, CSA, NEPSI (Note 5)
-ES ················· ATEX, IECEx for SENCOM sensor (Note 9)
1st input -P1 ···················· pH/ORP (Note 7)
-C1 ···················· Conductivity (SC)
-C5 ···················· Inductive conductivity (ISC)
-D1 ···················· Dissolved oxygen (DO)
-S1 ···················· pH/ORP (SENCOM sensor) (Note 8)
2nd input (Note 1) -NN ···················· Without input
-P1 ···················· pH/ORP (Note 7)
-C1 ···················· Conductivity (SC)
-D1 ···················· Dissolved oxygen (DO)
Output (Note 10) -A ···················· 4-20 mA + HART
— -N ···················· Always -N
Language set (Note 2) -LA ···················· English and 11 languages
Country (Note 3) -N ···················· Global except Japan
-J ···················· Japan
— -NN ···················· Always -NN
Option Mounting hardware /UM Universal mounting kit (Note 4)
/U Pipe and wall mounting hardware
/PM Panel mounting hardware
Hood /H6 Hood, stainless steel
/H7 Hood, stainless steel + urethane coating
/H8 Hood, stainless steel + epoxy coating
Tag plate /SCT Stainless steel tag plate
Conduit adapter /CB4 Conduit adapter (G1/2 x 4 pcs)
/CD4 Conduit adapter (1/2NPT x 4 pcs)
/CF4 Conduit adapter (M20 x 1.5 x 4 pcs)
Measurement law /K With Measurement Law certificate (Note 6)
Notes:
1: When a 2nd input is selected, only the same kind of the 1st input is available.
For example, when a 1st input is “-P1”, the 2nd input must be the same “-P1”.
The combination of ISC and ISC is not available. And, the combination of SENCOM sensor and SENCOM sensor is not
available, either.
2: These languages are message languages on the analyzer’s display.
One analyzer has English and 11 languages.
All languages are as follows; English, Chinese, Czech, French, German, Italian, Japanese, Korean, Polish, Portuguese,
Russian and Spanish.
3: When an analyzer is used in Japan, it must meet the Japanese Measurement Law.
Only SI units must be used on the analyzer and its documents in Japan.
4: The universal mounting kit contains the pipe and wall mounting hardware (/U) and the panel mounting hardware (/PM).
5: The type “-EA” is intrinsically safe type of ATEX, IECEx, FM, CSA and NEPSI, and non-incendive of FM and CSA. Temperature
classes are T4 and T6. The type “-EA” cannot be used with SENCOM sensor. For SENCOM sensor, select “-ES”.
6: The analyzer with Japanese Measurement Law certificate is available only for the following model;
FLXA21-D-[Housing code]-D-AA-P1-NN-A-N-LA-J-NN/[option code except /K]/K
Only one pH measurement with an analog sensor is certified. The output signal of 4 - 20 mA is certified. HART communication
is not certified.
7: This input is to be come from an analog pH/ORP sensor.
8: When the analyzer is connected with the digital sensor, FU20F pH/ORP SENCOM Sensor, only the following model is available;
Type: General purpose (-AA) and explosion-proof (-ES).
Explosion-proof (-ES) is available only for ATEX and IECEx
2nd input: Without input (-NN)
Option: except “/K”
9: The type “-ES” is intrinsically safe type of ATEX, IECEx for SENCOM sensor. Temperature class is T4.
10: The FLXA21 has other output types of “FOUNDATION Fieldbus” communication (suffix code: -F) and “PROFIBUS PA”
communication (suffix code: -P). Refer to GS 12A01A02-71E and GS 12A01A02-72E.
11: The housing with stainless steel + high anti-corrosion coating is available for the type “-AA” and “-ES”.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
11
n Dimensions and Mounting
Plastic Housing
151
141 Unit: mm
144 80
22.5
144
137
80
For sensor 1 cable 4-M6 depth 5
(For sensor 2 cable)
26.5 26.5 24.7
Approx. Packing
55(2.2") 49
(1.93")
G1/2 screw (/CB4), 1/2 NPT screw (/CD4)
Adapter M20x1.5 screw (/CF4)
F0204.ai
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
12
(Note) The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the panel mounting
hardware (/PM).
Panel thickness
1 to 12
+1
135
138 0
2-M5 length 35
+1
138 0
4-M6 *
195
100
185
200
3-ø10 holes
70
100
*: Tighten the four screws to a torque of 2 N•m. FB4_02.ai
Note: The wall on which the analyzer is mounted should be strong enough to bear the weight of more than 8 kg.
Pipe mounting hardware (Option code: /U, /UM)
Unit: mm
144
(205) Pipe 50A
151 (ø60.5)
4-M6 *
174
200
50
M8 U-bolt
Pipe mounting (Horizontal) Pipe mounting (Vertical)
100
*: Tighten the four screws to a torque of 2 N•m. FB4_04.ai
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
13
Stainless steel hood (Option code: /H6, /H7, /H8)
184 220 9 Unit: mm
144
57
72
144
199
(50)
(70)
(100)
FB4_05.ai
160 Unit: mm
165 150
80
32.5
165
80
137
44 22.2
137 FB4_06.ai
Approx. Packing
55(2.2") 49
(1.93")
G1/2 screw (/CB4), 1/2 NPT screw (/CD4)
Adapter M20x1.5 screw (/CF4)
F0204.ai
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
14
(Note) The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the panel mounting
hardware (/PM).
Unit: mm
Panel thickness
1 to 12
+1
134
2-M5 length 35 138 0
Plate
+1
138 0
(Only with coating) 4-M6 *
195
100
185
200
50
3-ø10 holes
70
*: Tighten the four screws to a torque of 2 N•m. 100
FB4_07.ai
Note: The wall on which the analyzer is mounted should be strong enough to bear the weight of more than 8 kg.
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
15
Pipe mounting hardware (Option code: /U, /UM)
Unit: mm
Pipe 50A 165
(214) (ø60.5)
4-M6 *
184.5
200
160
50
M8 U-bolt
Pipe mounting (Horizontal) Pipe mounting (Vertical)
100
*: Tighten the four screws to a torque of 2 N•m. FB4_09.ai
57
72
199
165
(50)
(70)
(100)
FB4_10.ai
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
16
n Wiring Diagrams
WTB10 or BA10 Case of Distributor
Terminal box *6 FLXA21 2-Wire Analyzer PH201G (Style B)
– CMN CMN
n n n n n n n n n *4 + B D H
A C F
*1 L N
Sensor
+ a c e Power supply
*3 b d f
– 20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
*2 *2 HOLD WASH
WTB10 or BA10 FAIL
(Stainless steel (Plastic housing)
Terminal box *6 Case of Distributor
housing)
SDBT
– –
+ B + H
A F
L N
Power supply
+
1 – 20 to 130V DC
2 or
(*3) 80 to 138V AC, 47 to 63Hz
*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.
*2: Connect the analyzer to ground. (Class D ground: 100 ohm or less)
The way of connecting the grounding cable varies depending on the plastic housing and stainless steel
housing.
In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside,
and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing.
Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing.
Although, on the stainless steel housing, the ground terminal symbol is (protective ground), the ground is
really functional ground.
*3: This line is connected to a distributor or 24V DC power supply.
*4: Terminal numbers for each sensor module are shown below.
*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive
conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.
*6: The terminal box may be necessary depending on the sensor cable length and the distance between the
analyzer and the sensor.
The SENCOM sensor is to be connected directly to the analyzer without a terminal box.
*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.
11 12 14 18 13 17 19 15 16
PH Module PH
11 12 13 14 15 16 NC
SC Module SC
11 12 13 17 14 16 15
ISC Module ISC
11 12 16 15 13 14 17 18 NC
DO Module DO
NC 82 83 84 NC 86 87
SENCOM Module SENCOM
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
17
n Inquiry Specifications Sheet for FLXA21 2-Wire Analyzer
Make inquiries by placing checkmarks () in the pertinent boxes and filling in the blanks.
1. General Information
Company name
Contact Person; Department;
Plant name;
Measurement location;
Purpose of use; Indication, Recording, Alarm, Control
2. Measurement Conditions
(1) Process temperature; to Normally [°C]
(2) Process pressure; to Normally [kPa]
(3) Flow rate; to Normally [l/min]
(4) Flow speed; to Normally [m/s]
(5) Slurry or contaminants; No, Yes
(6) Name of process fluid;
(7) Components of process fluid;
(8) Others;
3. Installation Site
(1) Ambient temperature; to [°C]
(2) Location; Outdoors, Indoors
(3) Others;
4. Requirements
1st Input; pH/ORP (analog sensor) Conductivity (SC) Inductive conductivity (ISC)
Dissolved oxygen (DO) pH/ORP (digital sensor, FU20F)
2nd Input; With (same as 1st Input) Without
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
18
4.2 Conductivity
1st Input
(1) Measuring range;
(2) Transmission output; 4 to 20 mA DC
-1 -1
(3) Detector/sensor; SC4AJ Two electrode system (0.02 cm ) Two electrode system (0.1 cm )
-1 -1
SC8SG Two electrode system (0.01 cm ) Two electrode system (10 cm ),
-1
Four electrode system (10 cm )
-1 -1
SC210G Two electrode system (0.05 cm ) Two electrode system (5 cm )
(4) Detector/sensor mounting method;
SC4AJ Adapter mounting, Welding socket, Welding clamp
SC8SG Screw-in, Flow-through
SC210G Screw-in, Flange, Flow-through, Screw-in with gate valve
(5) Electrode cable length; SC4AJ 3m, 5m, 10m, 20m
SC8SG 5.5m, 10m, 20m
SC210G 3m, 5m, 10m, 15m, 20m
(6) Others;
2nd Input
(1) Measuring range;
(2) Transmission output; 4 to 20 mA DC
-1 -1
(3) Detector/sensor; SC4AJ Two electrode system (0.02 cm ) Two electrode system (0.1 cm )
-1 -1
SC8SG Two electrode system (0.01 cm ) Two electrode system (10 cm ),
-1
Four electrode system (10 cm )
-1 -1
SC210G Two electrode system (0.05 cm ) Two electrode system (5 cm )
(4) Detector/sensor mounting method;
SC4AJ Adapter mounting, Welding socket, Welding clamp
SC8SG Screw-in, Flow-through
SC210G Screw-in, Flange, Flow-through, Screw-in with gate valve
(5) Electrode cable length; SC4AJ 3m, 5m, 10m, 20m
SC8SG 5.5m, 10m, 20m
SC210G 3m, 5m, 10m, 15m, 20m
(6) Others;
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00
19
4.4 Dissolved oxygen
1st Input
(1) Measuring range; 0 to 50 mg/L
(2) Transmission output; 4 to 20 mA DC
(3) System configuration selection; Electrode, Holder, Converter, Cleaning system,
Terminal box, Maintenance parts set, Calibration set
(4) Electrode cable length; 3m, 5m, 10m, 15m, 20m
(5) Type of holder; Guide pipe, Submersion, Flow-through, Suspension,
Angled floating ball, Vertical floating ball
(6) Cleaning method; No cleaning, Jet cleaning
(7) Others;
2nd Input
(1) Measuring range; 0 to 50 mg/L
(2) Transmission output; 4 to 20 mA DC
(3) System configuration selection; Electrode, Holder, Converter, Cleaning system,
Terminal box, Maintenance parts set, Calibration set
(4) Electrode cable length; 3m, 5m, 10m, 15m, 20m
(5) Type of holder; Guide pipe, Submersion, Flow-through, Suspension,
Angled floating ball, Vertical floating ball
(6) Cleaning method; No cleaning, Jet cleaning
(7) Others;
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation GS 12A01A02-01E Mar. 31, 2015-00