Micro-ohmmeters
User’s Manual
WWW.MEGGER.COM
User’s Manual
Contents
1 Introduction
.............................................................. 6
General...............................................................6
Function description............................................6
2 Safety
.............................................................. 8
General...............................................................8
Symbols on the instrument....................................... 8
Safety instructions...............................................8
3 Control panels
............................................................ 10
Control panel – MOM200A..............................10
Control panel – MOM600A..............................11
4 Operating instructions
............................................................ 12
External measuring instrument............................... 13
Troubleshooting...................................................... 13
5 Application examples
............................................................ 14
Measuring resistance in a breaker.....................14
Measuring resistance at bus-bar joints...............15
Expanded measuring range...............................16
Application example............................................... 16
6 Specifications
............................................................ 18
Specifications MOM200A.................................18
Specifications MOM600A.................................18
1 Introduction
2 Safety
Important
3 Control panels
1 2 2
11
4
12
5
13
6 14
15
7
8
16
9
10
1 2 3
10
4
11
5 12
13
7 14
8
15
4 Operating instructions
Troubleshooting
Fault Cause Remedy
Display shows Current range set on If possible, select ano-
only a 1 at instrument has been ther test current.
left. exceeded. Expand the measuring
range using the method
described in section
"Expanded measuring
range".
Poor contact at ends Check the sensing cable
of sensing cables. connections.
Thermal Overload. Wait until the tempe-
cut-out has rature drops and the
tripped. light-emitting diode
goes out.
Mains mini- Overload or fault in Reset the miniature
ature circuit instrument. circuit breaker. If it trips
breaker has again, contact service
tripped personnel.
5 Application examples
Fig. 2
Important
The sensing cables must be connected in-
side the current cables. Otherwise the test
data may be incorrect. See fig. 2.
Fig. 3
Application example
Important
Read the manual and comply with the
Safety instructions before using the mic-
roomhmmeter.
6 Specifications
Subject to change without notice. Printed matter No. ZP-BB04E Doc. BB0141nE V06 2015