the multimeter 20vdc measure the white and red wire, insure signal
output normal(0-20mv) .
Introduction:
“Delete” button: delete the parameter when you input the wrong
press delete button at the left side of the panel, meanwhile press the
The “function button” under the screen is into the special submenu
When you want to set or change the parameter, please first press the
Reset total clear: press the “reset total” button under the screen ,
Main total clear: select submenu 8, input the through the panel input
The instrument has total 10 submenu, press the “menu” on the left of
the instrument , the screen display 1-5 submenu, if you want select 6-10
button and the “figure” button input the various parameter: scale
3. select submenu 4, press the “page down” button select the output
define screen, set output one for alarm1 and open status, set
output two for totalize and open status (note: I/O circuit board
After finishing all parameter input then can establish test duration.
to the last page, press the “manual duration” through the figure button
input the belt length:??.?? m(the belt length error mustn’t be more than
conveyor, when belt running stable, please press the “start duration”
button ,press the “begin start” button ,the instrument auto running, the
time on the top of the screen will automatically countdown, when the
After test duration is done, you can start calibration (zero calibration
After calibration zero, first put down the lever on the weigh sensor,
empty the belt and run the belt, when the belt is stable, select the “menu”
button into the submenu 2, select zero calibration ,press the “enter”
button on the right of the instrument into and press “start calibration”
button under the screen on the left, the instrument auto progress zero
countdown, when the time gets to the zero, zero calibration is done, the
automatically change the zero value and display the zero error under the
screen, if the error is bigger than 0. 5%, please press the “sure zero”
button confirm the new zero value and please repeat zero calibration
again until the zero error display is smaller than 0. 5%.if the zero error
Before doing span calibration please put down the weight ???KG on
the lever, running the conveyor, when the belt running stable, select the
“menu” button into the submenu 2, select span calibration ,press the
“enter” button on the right of the instrument into and press “start
calibration” button under the screen on the left, the instrument auto
progress span calibration , the time ???s on the top of screen will
automatically countdown, when the time gets to the zero, span calibration
totalize weights automatically change the span value and display the span
error under the screen, if the error is bigger than 0.5%, please press the
“sure span” button confirm the new zero value and please repeat span
calibration again until the span error display is smaller than 0.5%. if the
span error display smaller than 0.5%, please press “exit span” button exit
the span calibration. Span calibration is done please suspend the weights.
Except span calibration the weights suspend all the time ( span calibration
After span calibration finish, remove the weights and lever on the
weigh sensor and calibration zero again until the zero error smaller than
0.5%.
T/H
When the conveyor running, but the instrument not speed display in
the screen, please into the submenu 9, inspect the speed frequency,if the
the speed sensor or wire. Otherwise replace the I/O circuit board.
measure the value from the weight sensor on instrument A/D board NO 3
and NO 4 terminal .if the voltage is smaller than 0mv or bigger than
20mv, explain the weighing sensor damage. Please replace one new.
terminal 4 signal-.
Terminal 1 and 5 output about 10vdc for weigh sensor and the weigh
output 7.