SAIMO6105A INSTRUMENT OPERATION AND MAINTENANCE

First,before the belt scale instruments power on please check the

power ,insure the power was normal(110VAC)。

After energizing, please check the weighing sensor output ,through

the multimeter 20vdc measure the white and red wire, insure signal

output normal(0-20mv) .

Introduction:

“Run” button : the screen display the main screen

“Menu” button: select available submenu

“Check” button:no use

“Delete” button: delete the parameter when you input the wrong

parameter(if instruments have problem which can’t be solved ,please

press delete button at the left side of the panel, meanwhile press the

power button frequently,if you do it must be input the parameter all

and established the duration)

The “function button” under the screen is into the special submenu

When you want to set or change the parameter, please first press the

“enter” button on the right of the panel

Reset total clear: press the “reset total” button under the screen ,

according to the direction on the screen to process

Main total clear: select submenu 8, input the through the panel input

the password “0”, according to direction on the screen to process.

Tel: 61-2-97378698 Fax:61-2-97378691 Email:salesaustralia@saimogroup.com


SAiMO Saimo Electric Co., Ltd.
The new instrument should be inputting some parameters about

conveyor and belt scale.

The instrument has total 10 submenu, press the “menu” on the left of

the instrument , the screen display 1-5 submenu, if you want select 6-10

submenu ,please press “page down” button under the screen.

1. press the“menu”button ,Press “1”figure button on the right of

the instrument into the submenu 1, through the “page down”

button and the “figure” button input the various parameter: scale

capacity : ???t , (note: press the “code parameter” button , the

idler number: 1 ,weighing sensor number: 2, idler space : ???

mm , conveyor angle : 0°parameters can be set) ,weigh sensor

capacity ???KG, sensitivity : 2.000mv/v, weighing channel :“1”,

speed mode: signal, speed channel :“1”

2. Select the submenu 3, press the “page down” input the

calibration mode: weight, input the exact weight ???KG.

3. select submenu 4, press the “page down” button select the output

define screen, set output one for alarm1 and open status, set

output two for totalize and open status (note: I/O circuit board

terminal 1 is output common (24vdc),terminal 2 is channel

1…….. Terminal8 is channel 7.according to the terminal that

wire contact to set the output channel).

After finishing all parameter input then can establish test duration.

Tel: 61-2-97378698 Fax:61-2-97378691 Email:salesaustralia@saimogroup.com


SAiMO Saimo Electric Co., Ltd.
Press the “menu” into the submenu 1, press the “page down” button

to the last page, press the “manual duration” through the figure button

input the belt length:??.?? m(the belt length error mustn’t be more than

0.05m),revolution:?,running time:??s, input is done ,please start the

conveyor, when belt running stable, please press the “start duration”

button ,press the “begin start” button ,the instrument auto running, the

time on the top of the screen will automatically countdown, when the

time gets to the zero, test duration is done.

After test duration is done, you can start calibration (zero calibration

and span calibration).

After calibration zero, first put down the lever on the weigh sensor,

empty the belt and run the belt, when the belt is stable, select the “menu”

button into the submenu 2, select zero calibration ,press the “enter”

button on the right of the instrument into and press “start calibration”

button under the screen on the left, the instrument auto progress zero

calibration , the time ???s on the top of screen will automatically

countdown, when the time gets to the zero, zero calibration is done, the

instrument through compare the instrument zero totalize weights

automatically change the zero value and display the zero error under the

screen, if the error is bigger than 0. 5%, please press the “sure zero”

button confirm the new zero value and please repeat zero calibration

again until the zero error display is smaller than 0. 5%.if the zero error

Tel: 61-2-97378698 Fax:61-2-97378691 Email:salesaustralia@saimogroup.com


SAiMO Saimo Electric Co., Ltd.
display smaller than 0.5%, please press “exit zero” button exit the zero

calibration. (zero calibration has to be done at least once a month).

Now zero calibration is done, you can start span calibration.

Before doing span calibration please put down the weight ???KG on

the lever, running the conveyor, when the belt running stable, select the

“menu” button into the submenu 2, select span calibration ,press the

“enter” button on the right of the instrument into and press “start

calibration” button under the screen on the left, the instrument auto

progress span calibration , the time ???s on the top of screen will

automatically countdown, when the time gets to the zero, span calibration

is done, the instrument through compare the constant and instrument

totalize weights automatically change the span value and display the span

error under the screen, if the error is bigger than 0.5%, please press the

“sure span” button confirm the new zero value and please repeat span

calibration again until the span error display is smaller than 0.5%. if the

span error display smaller than 0.5%, please press “exit span” button exit

the span calibration. Span calibration is done please suspend the weights.

Except span calibration the weights suspend all the time ( span calibration

has to be done at least every 3 months).

After span calibration finish, remove the weights and lever on the

weigh sensor and calibration zero again until the zero error smaller than

0.5%.

Tel: 61-2-97378698 Fax:61-2-97378691 Email:salesaustralia@saimogroup.com


SAiMO Saimo Electric Co., Ltd.
Calculation formula:

Constant= ((weights/idler space)*belt length * revolution ) / 1000 =

T/H

Instantaneous flow rate=(weights/idler space)*speed =kg/s*3.6=T/H

If the instantaneous flow rate display in the screen positive or

negative all the time, please process zero calibration.

When the conveyor running, but the instrument not speed display in

the screen, please into the submenu 9, inspect the speed frequency,if the

value is 0HZ,explain no speed signal input the instrument. Please check

the speed sensor or wire. Otherwise replace the I/O circuit board.

If happen instantaneous flow rate fluctuate range is bigger, please

measure the value from the weight sensor on instrument A/D board NO 3

and NO 4 terminal .if the voltage is smaller than 0mv or bigger than

20mv, explain the weighing sensor damage. Please replace one new.

Instrument Terminal connect description:

Weigh sensor: INTO A/D board.

Terminal 1 power+(10VDC),terminal 5 power-(0V),terminal 3 signal +,

terminal 4 signal-.

Terminal 1 and 5 output about 10vdc for weigh sensor and the weigh

sensor feedback about 0-20mvdc into instrument.

Speed sensor:ON I/O BOARD

Terminal 11 com,terminal 12 power,terminal 14 speed channel 1.

Tel: 61-2-97378698 Fax:61-2-97378691 Email:salesaustralia@saimogroup.com


SAiMO Saimo Electric Co., Ltd.
Terminal 11and12 output about 24VDC for speed sensor And Speed

sensor feedback about 0-32VAC into instrument.

Alarm output: ON I/O BOARD

Terminal 1 com (24vdc), terminal 2 is output channel 1……. termianl 8 is

output 7.

Simulate output:ON ANALOG I/O BOARD

Terminal 1 1#output+, terminal 2 1#output – (4-20ma)