Service Manual

Argus OXM C
Handheld Pulse Oximeter

EU Representative
Schiller AG
Altgasse 68, CH- 6341 Baar , Switzerland

Manufacturer
Medical Supply Co., Ltd
1647- 3 Dongwha- ri Moonmak- eub Wonjoo City Kangwon Province
TEL: 033 742 4300
FAX: 033 742 4385

Argus OXM C Service Manual v1.0


Revised Data: 2012- 01- 12

Printed in Korea
Copyright © 201
2012 Medical Supply Co.,
Co. , Ltd. All rights reserved.
Contents
Contents
Figure
Table

Notice
This is user manual of Medical Supply Co., Ltd. and protected by copy right. Thus, coping
without permission is strictly forbidden by law.

Warranty
Information of this manual can’ t be changed without permission. Medical Supply Co., Ltd is
not responsible for the problems caused by misuse or being used for other purposes.

Revision History
There is revision history in this manual. It is applied when this manual is changed. Please be
noted that the information of this manual which doesn’ t affect safety and performance of the
machine may changed without notice.

- 1- Argus OXM C Service Manual


Contents
Precautions for use ·········································································································································· 5
General safety infor matio n ·················································································· 5
Wa r n i n g s · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 5
Cautions ····································································································· 6
Product ··························································································································································· 7
Characteristics of Argus OXM C·············································································· 7
Product Appearance Screen················································································································· 8
Argus OXM C appearance ··········································································· 8
Argus OXM C Screen ···································································· 9

Product Specification ································································································································ 11


Physical specification, electrical specification, Using Condition ························································ 11
Measurement parameters············································································ 12
Compliance ··································································································13
Factory Mode ··········································································································································· 15
Summary·································································································· 15
Factory Mode Run ················································································ 15
S ys t e m I n f o . · ·· · ·· · ·· · ·· · ·· · · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · ·· · · ·· · ·· · ·· · ·· 16
D e mo M o d e · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·· · · · · · · · · · · · · 16
Q u i c k M o d e · · · · · · · · · · · · · ·· · · · · · · · · · · · · ·· · · · · · · · · ·· · · · · · · · · ·· · · · · · · · · · · · · ·· · · · · · · · · ·· · · · · · · · · ·· · · · · · · · · · · · · · 17
Firmware Download ···························································································································· 18
Summary ····························································································· 18
Equipment required ···································································· 18
How to download firmware ·································································· 18
Maintenance procedures ································································································································ 26
Summar y ······························································································· 26
Product return ·························································································· 26
Service ····································································································· 26
Regular safety inspection and functional checks ······················································ 27
P r o d u c t c l e a n i n g · · · · · · · · · · · · · · · · · · · · · · · · · · ·· · · · · · · · · · · · ·· · ·· · · · · · · · · · · · · · · · · ·· · · · · · · · · · · · ·· · ·· · · · · · · · · · · · · · 2 7
Battery maintenance ··········································································· 28
E n vi r o n me n t al p r o te ct i o n ··· ···· ··· ··· ··· ··· ···· ··· ··· ·· ···· ···· ··· ··· ··· ··· ···· ··· ··· ··· · ···· ···· ··· ··· ·· 2 8
Operation verification ································································································································· 29
Summary ·································································································· 29
E q u ip me n t r e q u ir ed ·· ··· ·· ·· ·· ··· ·· ·· ·· ·· ··· ·· ·· ·· ··· ·· ··· ·· ·· ·· ··· ·· ·· ·· ··· ·· ·· ·· ·· ··· ·· ··· ·· ·· ·· ··· ·· ·· · 29
Operation test ························································································ 29
Safety test ·················································································· 32
Disassembly guide ··································································································································· 34
Summary ······························································································ 34
Product disassembly ················································································ 34
Troubleshooting ······································································································································ 35
Summary································································································ 35
Troubleshooting ················································································· 36
Te c h n i c a l s u p p o r t · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 39

- 2- Argus OXM C Service Manual


Default setting ······································································································································ 40
Summary ························································································· 40
Equipment setting ranger and basic setting ·········································· 40
System process ··································································································································· 41
System block diagram ··································································· 41
System description ··············································································· 42

- 3- Argus OXM C Service Manual


Figure

Figure 1. Argus OXM C Product appearance · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 8


Figure 2. Graph Mode · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 9
Figure 3. Text Mode · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 9
Figure 4. Pulse Bar Mode · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 10
Figure 5. Factory Mode Password Entering · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 15
Figure 6. Factory Mode Run · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 16
Figure 7. Firmware version checking screen · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 16
Figure 8. Demo Mode setting Screen · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 17
Figure 9. System setting information screen · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 17
Figure 10. How to connect USB and communication connector · ·· · · · ·· · · · ·· · · · ·· · · ·· · · · · · ·· · · ·· · · · ·· 19
Figure 11. Communication setting - 1 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 19
Figure 12. Communication setting - 2 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 20
Figure 13. Hyper-terminal Run · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 20
Figure 14. Download Program Run · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 21
Figure 15. Preparations for download - 1 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 21
Figure 16. Preparations for download - 2 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 22
Figure 17. Firmware Download Run - 1· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 22
Figure 18. Firmware Download Run - 2· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 22
Figure 19. Firmware Download Run - 3· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 23
Figure 20. Firmware Download Run - 4· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 23
Figure 21. Drive Format · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 23
Figure 22. Argus OXM C Save File · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 24
Figure 23. Hyper-terminal Rerun · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 24
Figure 24. Re-preparation for download · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 24
Figure 25. Firmware download completed · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 25
Figure 26. Product disassembly · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 34
Figure 27. System block diagram ·· ·· · ·· ·· · ·· ·· · ·· ·· · ·· · · ·· ·· · ·· ·· · ·· ·· · ·· ·· · ·· · ·· ·· · ·· ·· ·· ·· · ·· ·· · ·· ·· · ·· 41
Figure 28. Fuse and circuit of Argus OXM C · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · 42

Table

Tab le 1 . E q uip me nt fo r Fir m war e d o wnlo ad ··· ···················································· 18


Ta b l e 2. Required test equipment ································································ 29
Ta b l e 3. Ground leakage current ···························································· 32
Ta b l e 4. External leakage current ····························································· 32
Ta b l e 5. Patient leakage current ··································································· 33
Tab le 6 . P a ti e nt le a ka g e c ur r e n t – ma i n s v o l t a ge p ar t ··· ··· ···· ···· ··· ···· ····· ···· ··· ···· ···3 3
Ta b l e 7. Main disassembling components ················································· 34
Ta b l e 8. Status message ······································································· 35
Table 9. Setting range and default setting ················································ 40
Table 10. Power supply and details ························································· 42

- 4- Argus OXM C Service Manual


Precautions for use

General safety information

This chapter explains the things people must aware to use Argus OXM C Pulse Oximeter.
Please refer to reach chapter for details.

Attention! Please make sure to read attached


attached accessory manual thoroughly before using.

‘ WARNING’
WARNING’ and ‘ CAUTION’
CAUTION’ contain information that may cause injury of human or damage
of equipment. Make sure to read this manual before using.

Warnings

WARNING: Don’ t use Argus OXM C near flammable gas like anesthetics. It
may cause explosion.

WARNING: Please make sure to disconnect SpO2 sensor when turning on or


off Argus OXM C to prevent burn.

WARNING: Don’ t put Argus OXM C equipment and accessories in corrosive


cleaning solution nor use abrasive cleaning material.

WARNING: Don’ t connect conductive part of other equipment or other


conductive material to conductive part of the equipment or accessories.

WARNING: Never connect the equipment to electrical outlet during using. The
equipment may be off accidently.

WARNING: Make sure to disconnect DC adapter when disassembling Argus


OXM C. It may cause electrical shock. Also disconnect inner battery
connector after removing case.

WARNING: Don’ t pull out power code before system power shut down
completely. You may lose stored patient measurement data or other settings.

WARNING: Be very careful to avoid electrical shock during safety test.

WARNING: Don’ t put Argus OXM C at the place until performing repair or
regular maintenance. Safety test is described in this service manual. Tests
may fail if you can’ t understand this manual.

- 5- Argus OXM C Service Manual


Cautions

CAUTION: Be careful on static electricity during operation, disassembly,


assembly and handling of components of Argus OXM C.

CAUTION:
CAUTION: Use proper screws and don’ t overtighten them when re-
assembling Argus OXM C.

CAUTION: Ferrite core is used to reduce electromagnetic wave. Don’ t


remove ferrite core when disassembling or re- assembling the equipment.
Otherwise inaccurate data can be displayed or stored due to this
electromagnetic wave.

CAUTION: Same capacity fuse should be used when replaced.

- 6- Argus OXM C Service Manual


Product
Features of Argus OXM C

This manual applies only to Argus OXM C (pulse oximeter). The following are details of the
product.

Characteristics of Argus OXM C


Mechanical characteristic
ARGUS OXM C is small and light measurement equipment of oxygen saturation. Size is
170× 88× 32 (mm) (H× W× D) and weight is about 300g (battery included). The battery is
mounted inside and the handle at the side makes easy to use for patient transfer in hospital
or ambulance.

Electrical characteristic
ARGUS OXM C can be used for around 5 hours using fully charged internal battery. ARGUS
OXM C is recharged by connecting DC adapter to the equipment. For more detail, refer to
‘ Battery Operation’ in Chapter 6 of this manual.

Screen Display
ARGUS OXM C displays patient’ s oxygen saturation and pulse measurements and status
message with letters, numbers and graphs.

- 7- Argus OXM C Service Manual


Product appearance and screen
Argus OXM C appearance
Argus OXM C screen

Argus OXM C appearance

Figure 1. Argus OXM C appearance

1. Power button 7. USB Port (for firmware update)


2. Alarm/ mute button 8. DC power port
3. UP button 9. LCD
4. DOWN button 10. SpO2 sensor connection port
5. Menu/ Enter button 11. Cradle Holder
6. Select button 12. Speaker output

- 8- Argus OXM C Service Manual


Argus OXM C Screen Constitution
Graph Mode
This is Display Mode initially set. It is used to observe the graph of oxygen saturation
signal. As show n in Figure 2, SpO2 value is displayed on the left, heart rate on the right
and PA on left bottom.

Figure 2. Graph Mode

Text Mode
SpO2 is displayed on the left, heart rate on the right and PA is displayed at the left bottom
as shown in Figure 2.

Figure 3. Text Mode

- 9- Argus OXM C Service Manual


Pulse Bar Mode
As shown in Figure 4, SpO2 is displayed on the left, heart rate on right and Pulse
Level is displayed between of the two.

Figure 4. Pulse Bar Mode

1. Logo of Medical Supply 7. PR Limit Setting


2. SpO2 value 8. PR value
3. Pulse Wave 9. SpO2 Limit Setting
4. PA display 10. Alarm status display
5. Time, Date display 11. Current status and menu display
6. Battery status display 12. Pulse Level

- 10 - Argus OXM C Service Manual


Product Specification
Physical, electrical specification, using condition
Measurement parameter
Compliance

Physical Specification
Size
Size 170× 88× 32 (mm) (H× W× D)
Weight about 300g

Electrical Specification
DC Power
Rated Voltage Input : AC 100- 240V, Output : DC 12V
Input Frequency 50/ 60 Hz
Power Consumption 2.5A

Battery
Type Li- ion rechargeable battery
Voltage/ Capacity 7.4V/ 1800mAh
Recharging 3 ~ 6 months recommended
Lifetime About 500 cycles of charge
Regulation IEC 61951- 2 2nd Ed. 2003- 04

Using Condition
Operation
Temperature 10 °C (50 °F) to 40 °C (104 °F)
Moisture
Moisture 15 % RH to 95% RH, non- condensing
Altitude - 300 m (- 985 ft) below sea level
Elevation 3,658 m (12,000 ft)

Transport and storage


Temperature –10 °C (14 °F) to 50 °C (122 °F)
Moisture 15 % RH to 95% RH, non- condensing
Altitude –610 m (–2,000 ft)
Elevation 12,192 m (40,000 ft)

Note:
Note: The device may not work properly if it is transported or stored outside the range of using
condition.

- 11 - Argus OXM C Service Manual


Measurement parameters

Oxygen saturation ( SpO2)/ Pulse Rate

Oxygen saturation ( %Saturation)


%Saturation)
Measurement range
range 0 ~ 100%
Setting Time Graph output time : Max 7 seconds
Percentage output time : Max 12 seconds
Accuracy Normal Perfusion ± 2%
Low Perfusion (PA 0.7% 이하 조건에서
) ± 3%
Average Measured 20 times average

Pulse Rate
Range 25 ~ 250 BPM
Accuracy ± 2%
Resolution
Resolution 1 BPM
Average 8 times measurement average

- 12 - Argus OXM C Service Manual


Compliance

Item Reference Title of harmonized standard

Quality Systems- Medical Devices- Particular


EN ISO 13485:2003
requirements for the application of ISO9001
EN 60601- 1:
Medical electrical equipment — Part 1: General
1990+A1
requirements for safety (IEC 60601- 1:1988)
:1993+A2:1995
Clinical Investigation of Medical Devices for
EN ISO 14155- 1:
Human Subjects - part1:General Requirement
2009
(ISO 14155- 1:2003)
Clinical Investigation of Medical Devices for
EN ISO 14155- 2:
Human Subjects – part2:Clinical investigation
2009
plans(ISO 14155- 2:2003)
Medical electrical equipment — Part 1- 1:
General requirements for safety — Collateral
EN 60601- 1-
General standard: Safety requirements for medical
1:2001
Safety electrical systems
(IEC 60601- 1- 1:2000)
Medical electrical equipment — Part 1- 4:
EN 60601- 1- 4: General requirements for safety — Collateral
1996 standard: Programmable electrical medical
systems (IEC 60601- 1- 4:1996)
Medical electrical equipment — Part 1- 6:
EN 60601- 1- 6:
General requirements for safety — Collateral
2007
standard: Usability (IEC 60601- 1- 6:2006)
Medical devices — Application of risk
EN ISO 14971:
management to medical devices (ISO
2009
14971:2007)
EN ISO10993- 1: Biological evaluation of medical devices — Part
2009 1: Evaluation and testing (ISO 10993- 1:2003)

- 13 - Argus OXM C Service Manual


Item Reference Title of harmo
harmonized standard

Medical electrical equipment — Particular


Oxygen requirements for the basic safety and essential
EN ISO 9919: 2009
saturation performance of pulse oximeter equipment for
medical use (ISO 9919:2005)
Medical electrical equipment — Part 1- 2:
General requirements for safety — Collateral
EN 60601- 1-
standard: Electromagnetic compatibility —
2:2001
Requirements and tests (IEC 60601- 1- 2:2001)
Electromagn +A1: 2006
Amendment A1:2006 to EN 60601- 1- 2:2001
etic (IEC 60601- 1- 2:2001/ A1:2004)
Compatibility
EN55011: 2007 RF Emissions Group 1, Class B Limits and
+A2: 2007 methods of measurement of radio disturbance
(CISPR 11: 2003 characteristics of industrial scientific and
+A2: 2006) medical (ISM) radiofrequency equipment
Information supplied by the manufacturer with
Labeling EN 1041:2008
medical devices
Graphical symbols for use in the labeling of
Marking EN 980: 2008
medical Devices
Pre- Shipment Test Procedures
Package ISTA 1A:
(Procedure 1A, 1994 Rev.)

- 14 - Argus OXM C Service Manual


Factory Mode
Summary
Factory mode run
Firmware version
Demo
System setting information

Summary
This Chapter explains Factory Mode in which authorized person can review/ check ARGUS
OXM C information such as Firmware Version, Demo Mode Setting, System Set Information
Check.

Note: Use only Firmware Version, System Information Check and Demo Mode Setting in this
Factory Mode.

Factory Mode Run


Enter the Factory Mode according to as follows.
1. Press Menu button ( ).
2. Move to Factory Mode Item using Up/ Down button ( / ).
3. Select required item by pressing Select Button ( ).
4. Password Enter screen appears as Figure 5.
- Select number using Up/ Down button ( / ).
- For each digit, number from 0 to 9 can be selected.
- Press Select button ( ) after selecting number to move to the next digit.
- Pressing Select button ( ) after entering all three digits, it enters Factory Mode if
password matches. Otherwise, re- enter screen is displayed.
5. It the password matches, Factory Mode will be displayed as Figure 6.
6. Select Factory Mode Item using Up/ Down button ( / ).
Items are as follows.
- 1. System Info: Firmware version check, 2. Demo Mode: Demo check,
3. Quick Mode: System Setting Information Check.
7. It enters selected item if pressing Select Button ( ).
8. It returns to Main Page of Factory Mode if pressing Menu button ( ) after
finishing checking.

Figure 5. Factory Mode Password Enter Screen


- 15 - Argus OXM C Service Manual
Figure 6. Factory Mode Run

System Info.
In this mode, it is possible to know Firmware Version of ARGUS OXM C.

1. Run Factory Mode referring to ‘ Factory Mode Run’ .


2. Select System Info by pressing Select Button ( ).
3. Check current ARGUS OXM C version as shown Figure 7.
- 4 items, O/ S Version, App Version, Alg Version, H/ W Version
4. It returns to Main Screen of Factory Mode when pressing Menu Button ( ) after
checking.

Figure 7. Firmware Version Check Screen

Demo Mode
Demo Mode sets the operation of ARGUS OXM C arbitrarily and it is to run only to check if
there is any problem.

1. Run Factory Mode referring to ‘ Factory Mode Run’ .


2. Select Demo Mode by pressing Select Button ( ).
3. If selecting Set, Demo Mode will run as shown in Figure 8. If selecting Reset, Demo
Mode will stop.
4. It returns to basic measurement screen if pressing Menu button ( ).

Note:
Note: Don’ t use above function except specific purpose (demonstration or test).

- 16 - Argus OXM C Service Manual


Figure 8. Demo Mode Setting Screen

Quick Mode
In Quick Mode, it is possible to know all current system information of ARGUS OXM C.

1. Run Factory Mode referring to ‘ Factory Mode Run’ Chapter.


2. Select Quick Mode by pressing Select Button ( ).
3. Now you can check current system setting information of ARGUS OXM C as shown in
Figure 9.
4. It returns to Main Mode screen if pressing Menu Button ( ) after checking.

Figure 9. System Information Setting Information screen

- 17 - Argus OXM C Service Manual


Firmware Download
Summary
Equipment required
How to download Firmware
Solution of the problems occurred during download

CAUTION:
CAUTION: Make sure to download the only firmware distributed by Medical Supply
through service staff. Medical Supply is not responsible for any problems occurred
from other firmware not distributed by Medical Supply.

CAUTION: Make sure to connect DC adapter when downloading the firmware.

Summary
This chapter is for the possibility of error in the firmware or reloading distributed firmware
when the firmware is renewed along with new system. Please request the most recent
version of the firmware to Medical Supply.

Equipment required
Table 1 shows equipment list required to download firmware.

Table 1. Firmware download equipment


Equipment Description
Firmware Downloading Cable USB cable 2EA
Communication sub-
sub- board Main board communication connector
Firmware Firmware to be downloaded
Personal Computer (PC) With USB port

How to download firmware


USB Driver to connect ARGUS OXM C to PC, operation program and download program
should be installed before download.

1. Turn on the PC with to download the firmware.


2. Connect communication sub- board with USB connector with main board referring to
Figure 10.
3. Turn on the main board.
4. Check if it has been done successfully by searching USB port of the PC connected the
equipment.
Note : CP210x_VCP_Win_XP_S2K3_Vista_7(USB driver) / http://www.silabs.com
Note : Sam-ba_2.10.exe(Firmware download program) / http://www.atmel.com

- 18 - Argus OXM C Service Manual


Figure 10. How to connect USB and communication connector

5. Check currently set port number by pressing System Registration – Device manager – Port
referring to left screen in below Figure 11.

6. Select Start – Program – Accessories – Communication – Hyper Terminal and press Enter
button after typing name.

Figure 11. Communication Setting


Setting - 1

- 19 - Argus OXM C Service Manual


7. If the window to be connected appears, select COM port connected with ARGUS OXM C
as ‘ Modem to be Connected’ and click Confirm (port no confirmed in the process 5).

8. If initial communication set up window appears, set as shown in Figure 11 and click
Confirm.
- bite/ second: 115200
- data bite: 8
- parity: none
- stop bite: 1
- flow control: none

9. After communication setting, connect communication sub- board and connector to main
board before downloading as shown in above Figure.

Figure 12. Communication setting - 2

10. Referring to left screen in Figure 13, remove Switch jumper and press Reset Button
then a message, ‘ RomBoot’ appears in HyperTerminal.

Figure 13.
13. HyperTerminal Run

- 20 - Argus OXM C Service Manual


11. Run installed Sam-ba_2.10 program and click Connect Button when the screen on the left
of Figure 14 appears.
- Select the connection: / usb/ ARM0
- Select your board: at91sam9rl64- ek

12. If it runs correctly, right screen of Figure 14 appears.

Figure 14.
14. Download Program Run

13. Referring to left screen of Figure 15, insert jumper into the switch. Select ‘ NandFlash’
tap as shown in right screen of Figure 15 and select ‘ Enable NandFlash’ fron Scripts
item then click ‘ Execute’ Button.

Figure 15.
15. Preparation for download - 1

- 21 - Argus OXM C Service Manual


14. As shown in the Figure 16, select ‘ Erase All’ from Scripts, click ‘ Execute’ Button, select
‘ Send Boot File’ and click ‘ Execute’ Button.

Figure 16.
16. Preparations for download - 2

15. Click Open after selecting firmware program, 'nandflash_at91sam9rlek.bin' as shown in


the Figure 17.

Figure 17.
17. Firmware download run - 1

16. Click Open File icon after entering Address as '0x020000' as shown in the Figure 18.

17. Click 'J-Boot.bin' file of firmware program as shown in the right screen of the Figure 18.

18. Click ‘ Send File’ as shown in Figure 18.

Figure 18. Firmware Download Run - 2

- 22 - Argus OXM C Service Manual


19. Enter '0x0100000' for Address and click File Open icon as shown in Figure 19.

20. Click 'NotSpO2_JBOSN.img' file of the firmware program and open the file as shown in
Figure 19. (Select ‘ All File(* .* )’ for file type).

21. Click ‘ Send File” as shown in Figure 19.

Figure 19.
19. Firmware Download Run - 3

22. If pressing ‘ Reset Button’ after terminating Sam-ba_2.10 program, message in the left
screen will appear in hyper- terminal window as shown in Figure 20. This message lasts
for 2~ 3 minutes. It goes to next stage when ‘ NO SHELL’ message appears like right
screen of Figure 20.

Figure 19.
19. Firmware Download Run - 4

23. Proceed to format after selecting ‘ My Computer’ – ‘ Portable Disk’ when ‘ NO SHELL’
message is displayed. Select ‘ FAT’ for ‘ File System’ .

Figure 21.
21. Drive Format

- 23 - Argus OXM C Service Manual


24. If format is completed, copy ‘ HP- 110_SaveFile’ of firmware program to formatted
portable disk.

Figure 22.
22. Argus OXM C Save File

25. Make ‘ RomBOOT’ state by pressing Reset Button after removing jumper of the Switch
and re- run ‘ Sam-ba_2.10’ program again as shown in Figure 23.

Figure 23.
23. Hyper Terminal Re-
Re- run

26. Insert jumper into the Switch, select ‘ NandFlash’ Tap, select ‘ Enable NandFlash’ from
Scripts item and then click ‘ Execute’ Button as shown in Figure 24.

Figure 24.
24. Re-
Re- preparations for download

- 24 - Argus OXM C Service Manual


27. Enter Address as '0x0100000' and click File Open icon as shown in Figure 25.

28. Click ‘ JBOSN.img' file of the firmware program to open as shown in Figure 25. (Select ‘ All
File(* .* )’ for file type).

29. Click ‘ Send File’ Button as shown in the Figure 25.

30. When all download is completed, terminate Sam-ba_2.10 program and turn off the power
by pressing power button of main board longer than 7 seconds.

Figure 25.
25. Firmware Download Completion

- 25 - Argus OXM C Service Manual


Maintenance Procedures
Summary
Product return
Service
Regular safety inspection and functional checks
Product cleaning
Battery maintenance
Environmental protection

WARNING: Only service personnel can disassemble the unit. Inner part can’ t be
repaired by the user directly.

WARNING: Be careful not to make CV unit, cable, connector, switch and case get
wet.

WARNING: Argus OXM C complies with Water Regulations (IPX1) of IEC 60529.

CAUTION: Please unplug the power code before cleaning.

Summary

For replacing parts, you should follow instructions and regulations on recycling or disposal.
Otherwise it may cause environmental pollution or harm to people due to improper treatment
of the battery of accessories.

Product return
If you find the device broken or accessories damaged after receiving the product, please
contact with the company or local distributor immediately. Return the product with all
accessories in the package box. If you don’ t have Argus OXM C package box, you can use
proper package box. Please pack them securely as possible.

Service
Other management is not required for Argus OXM C except battery control. Refer to this
manual if you need more information. The equipment should be checked regularly by qualified
service staff of user’ s company. If any service is required, please contact qualified service
personnel or Customer Service Department of Medical Supply Co., Ltd.

- 26 - Argus OXM C Service Manual


Regular safety inspection and functional checks
Regular safety inspection

Please check the follow every 24 months.


 Check for the damage of the exterior of the equipment.
 Check if it is easy to read the label. If not, please contact with Service Center of the
company.
 For accurate verification of the equipment, refer to Operation Verification section of the
manual.
 Please refer to Operation Verification section of this manual for detailed electrical
safety test. If it fails in electrical safety test, please contact with Customer Support
Center of the company. Do not try to repair the equipment.

Functional checks

It should be checked according to operation verification guideline every two years be


qualified person.
● It should be checked according to Verification Execution Chapter of this Service Manual
if it is damaged seen by naked eye or mechanically shocked (e.g. Drop impact).
● For regular safety test, please refer to Safety Test Chapter of this manual. It the
equipment fails in electrical safety test, please refer to Troubleshooting Chapter of this
manual.
● Check for proper fuse value and range.
- DC power part: 4A / 0451004.MRL (F1)
- Battery power part: 4A 0451004.MRL (F2)

Product Cleaning
To keep product surface clean, please follow the following.
● For cleaning surface of Argus OXM C, wipe the surface of the equipment with soft
cloths dampened with detergent available in the market.

CAUTION: Before cleaning, please unplug DC power form the equipment.

CAUTION: Be careful not to make water or detergent enter the inside of the
equipment during cleaning.

Note: Please refer to user manual attached for accessories such as SpO2 sensor.

Note: If liquid like water or detergent are spilled, clean well and dry it thoroughly before using.

- 27 - Argus OXM C Service Manual


Battery maintenance
Argus OXM C uses only Ni- MH battery. If the equipment has not been used longer than 2
months, charge the battery before using. Please refer to Operation Verification or Battery
Operation Chapter of this manual to charge the battery.

We recommend replacing the battery every two years. Please remove the battery if the
equipment is to be stored longer than 2 months. To replace or remove the battery, please
refer to Disassembly Guide Chapter of this manual.

Note: Long storage of Argus OXM C without removing battery may degrade the battery
performance. For the restoration of the normal capacity of the battery, full charge and
discharge is required. Charge the battery fully for about 120 minutes with the
equipment
equipment turned off. Medical Supply recommends replacing the battery every two
years.

CAUTION:
CAUTION If Argus OXM C has been stored longer than 2 months, please request
service person to check and remove it. It is highly recommended replacing battery
if it has not been used longer than 2 months.

CAUTION:
CAUTION If the battery seems to be damaged, has leakage or cracks, replace it
into approved battery immediately by qualified service person.

CAUTION:
CAUTION When disposing or reusing used parts after replacing battery or other
parts, please observe regulation of disposal/ recycling. They may be exploded
during incinerating.

Environmental Protection
Battery can be reused. Do not dispose the battery in the trash can.
Dispose the battery in accordance with the local lows.
Do not burn as it may be exploded.

- 28 - Argus OXM C Service Manual


Operation Verification
Summary
Equipment required
Operation test
Safety test

Summary
In this chapter, verification test done during repair or regular inspection will be explained.
All tests except battery replacement and battery test should be done before returning the
equipment. It Argus OXM C fails in the test described in this manual it should be repaired
properly before returning.

Equipment required
Table 2. Test equipment required
Equipments Description
Description
SpO2 Sensor,
Sensor, Extension Cable DOC 10 ‐
SpO2 Simulator FLUKE Biomedical Index2 or equivalent
Safety Analyzer ‐
METRON QA 90 or equivalent
Stopwatch Manual or electronic

Operation test
Battery charging and battery performance test should be done whenever the battery is
suspected as the cause of the problem. For other tests, please follow the following repair
or inspection ways described in the below. Battery has to be fully charged before battery
performance test. In this chapter, initial setting is displayed. If user set the equipment,
these settings will be displayed.

Power

1. Connect DC power to Argus OXM C using correct power code.


2. Press power button and check if the equipment is turned on.
3. Check if the equipment displays normal mode after dong self- diagnosis.
4. Check if AC power displays during charging.
5. If the equipment operates normally, remove DC power.
6. Check if Battery Status/ Battery Operation display.
7. Check if the equipment is shut down when pressing power switch longer than 3 seconds.

- 29 - Argus OXM C Service Manual


Battery Charging

1. Connect DC power to Argus OXM C using correct power code.


2. Verify if DC Power displays during battery charging.
3. If the battery is fully charged, Full Charge is displayed. If battery is being charged,
thunderbolt mark is displayed.
4. If the battery is fully charged, remove DC power and check the battery in Service Mode.

Note: Battery performance depends on the usage habit. Full charge/ discharge is required.
Note: Argus OXM C uses only Li-
Li- ion battery.

Battery discharge
discharge

1. If Li- ion battery is fully charged, remove DC power.


2. Connect the equipment with SpO2 simulator.
SpO2 simulator set up: SpO2 90%, PR 75BPM, PI 0.3%
3. Start automatic measurement.
4. Check if battery charging alarm sounds when battery is left about for 20 minutes after
operating for about 5 hours and Battery Status Indicator flashes.
5. If battery is low level, only mark displays in the screen and it shuts down
automatically when turning on the equipment after it shuts down automatically. If DC
power is connected in 5 seconds and start charging, the equipment returns to normal
mode.
6. If the test is passed, charge the battery immediately.

General operation test

Alarm and alarm release


1. Connect DC power and turn on the equipment.
2. Set the lower limit of oxygen saturation as 90% in alarm limit set menu. Refer to ‘ 8.
Setting Menu’ of use manual for detailed alarm limit setting.
3. Connect oxygen saturation (SpO2) simulator.
4. Set up the simulator as follows.
- Oxygen saturation 80%, PR 70 BPM
5. Start to measure oxygen saturation
6. Check the following
- Check if oxygen saturation and pulse rate in LCD screen are same as settings of
simulator.
- Check if alarm sounds and oxygen saturation value, oxygen saturation alarm lower
limit indicator and alarm indicator flash.

7. Press alarm release button ( ).


8. Check the following.
- if alarm is muted
- if alarm indicator changed into alarm stop sign
- if oxygen saturation value and alarm lower limit indicator continues to flash
9. Check if the things mentioned in above 6 happen again.
10. Set up oxygen saturation simulator as follows.
- Oxygen saturation 98%, PR 70 BPM
- 30 - Argus OXM C Service Manual
11. Check if the oxygen saturation of the equipment is same as the setting of the simulator.
12. Check if the alarm sound and indicator stop.
13. Remove the oxygen saturation simulator.
14. Check if oxygen saturation (SpO2) signal error displays in LCD screen of the equipment
and alarm sound.

15. Check the following after pressing alarm release button ( ).


- Alarm sound stops
- Error signal of oxygen saturation disappears from the LCD screen.

System volume control


1. Connect DC power and turn on the equipment.
2. Set the lower limit of oxygen saturation as 90% in alarm limit set menu. Refer to ‘ 8.
Setting Menu’ of use manual for detailed alarm limit setting.
3. Connect oxygen saturation (SpO2) simulator.
4. Set up the simulator as follows.
- Oxygen saturation 98%, PR 70 BPM
5. Start to measure oxygen saturation
6. Change the system volume from 0 to 9 using setting value change button ( / ) in
setting mode and check if button sound and pulse sound change.

Alarm sound control


1. Connect DC power and turn on the equipment.
2. Set the lower limit of oxygen saturation as 90% in alarm limit set menu. Refer to ‘ 8.
Setting Menu’ of use manual for detailed alarm limit setting.
3. Connect oxygen saturation (SpO2) simulator.
4. Set up the simulator as follows.
- Oxygen saturation 98%, PR 70 BPM
5. Start to measure oxygen saturation
6. Change the alarm sound from 1 to 9 using setting value change button ( / ) in
setting mode and check if button sound and pulse sound change.

Time test
1. Connect DC power and turn on the equipment.
2. Check time updates every minute in the lower right corner of LCD screen.

Measurement parameters test


test

Oxygen saturation (SpO2) test


1. Connect SpO2 simulator after turning on the power.
2. Set up the simulator as follows.
- SpO2 70 %, PR 60 BMP
3. Check SpO2 value and PR value.
- Check if pulse beat sounds.
- Check SpO2 value displays within ± 2 % tolerance.
- Check if heart rate displays within ± 2 % tolerance or ± 3BPM.
- Check if pulse size displays.
4. Check if SENSOR OFF displays on the screen when disconnecting SpO2 sensor from
the equipment.

- 31 - Argus OXM C Service Manual


Safety Test
Argus OXM C complies safety regulations of Class I, Type BF.

Protective grounding

Ground resistance test


This test is to check the integrity of ground wire of power code from AC plug. Current for
this test is 264V/ 50Hz - 60Hz/ 25A.
1. Connect AC power to the equipment and connect power code to safety tester according
to instruction of tester operation.
2. Connect ground resistance input lead to ground part of the equipment. Measured

resistance should not exceed 0.2 .

Electrical leakage

Earth leakage current


This test follows the regulation of earth leakage current of IEC 60601- 1. For this test, AC
264V/ 50Hz – 60Hz is applied. All measurements should be done with both power On and
Off.
1. Connect AC power code of DC adapter to electrical safety tester in accordance with
operational guide.
2. Do the test according to recommended method.

Table 3. Ground leakage current


Test status ㎂
Allowed leakage current ( )
Normal Condition (NC) 500
SFC Open Supply (SFC OS) 1000
Normal Condition RM (NCRM) 500
SFC Open Supply (SFC OSRM) 1000

Note: Ground leakage should not over allowed leakage current in the Table 3.
‐ ‐ ‐
Note: NC Normal Condition / SFC Single Fault Condition / RM Reverse Mains/ Line
Voltage

External leakage current


This test complies the regulations of external leakage current of IEC 60601- 1. This test is
to measure ungrounded external leakage current between external part and the ground.

Table 4. External leakage current


Test status Allowed leakage current
current ( )㎂
Normal Condition (NC) 100
SFC Open Supply (SFC OS) 500
SFC Open Earth (SFC OE) 500
Normal Condition RM(NCRM) 100
SFC Open Supply RM (SFC OSRM) 500
SFC Open Earth RM (SFC OERM) 500

Note: External leakage current should not over allowed leakage current in the Table 4.

- 32 - Argus OXM C Service Manual


Patient leakage current
This test follows Section 19, Regulations of Patient Leakage Current of IEC 60601- 1 중
Class I, Type BF. This test is to measure the patient leakage current of the patient
grounded.
1. Connect AC power code of DC adapter to electrical safety tester in accordance with
operational guide.
2. Connect test cable to SpO2 sensor connector.
3. Turn on Argus OXM C.
4. Measure patient leakage current according to recommended method.

Table 5. Patient leakage current


Test status ㎂
Allowed leakage current ( )
Normal Condition (NC) 100
SFC Open Supply (SFC OS) 500
SFC Open Earth (SFC OE) 500
Normal Condition RM(NCRM) 100
SFC Open Supply RM (SFC OSRM) 500
SFC Open Earth RM (SFC OERM) 500

Note: Patient leakage current should not be over allowed leakage


leakage current in the Table 5.
Note: Test for each cable connected to the patient is required.

Patient leakage current ( Main voltage of affected portion)


portion)
WARNING: AC main voltage is applied to the part tested. Practice carefully
to avoid the risk of electric shock.

WARNING: Don’ t touch clips or simulator connected to the patient during


test. It may cause electric shock.

This test follows Section 19, Patient Leakage Current of Class I, Type BF of IEC 60601- 1.
This test is applied to the part between the patient connected with 110% of main voltage
and grounded part. Patient leakage current is measured from the ground wire connected to
each patient.

Note: Keep patient test cable as short as possible during the leakage test.
Note: For this test, the connector
connector explained on patient leakage current is required.

1. Connect AC power code in accordance with operational guide.


2. Connect test cable to SpO2 sensor connector.
3. Turn on Argus OXM C.
4. Measure patient leakage current according to recommended way of the tester.

Table 6. Patient leakage current – Main voltage applied part


Test status ㎂
Allowed leakage current ( )
Normal polarity (SFC) 5000
Reverse polarity (SFCRM) 5000

Note: Patient leakage current should be measured for both normal direction a and
nd reverse
direction of main electrode.
Patient leakage current should not be over allowed current in the Table 6.
- 33 - Argus OXM C Service Manual
Disassembly
Summary
Product disassembly

WARNING: Disconnect Ac power and inner battery connector first when


opening or disassembling Argus OXM C to avoid electric shock.

Summary
This chapter explains how to disassembly Argus OXM C.

Product disassembly
1. Remove 4 screws and disassemble the Bottom case.
2. Remove 4 screws of the Bottom Case.
3. Remove the battery connected to the main board and speaker connecter.
4. Remove 5 screws of main board.
5. Separate Top case from main board by removing LCD connector.

Figure 26. Product disassembly

Table 7. Main component disassembled


Component Unit
Top Case 1
Bottom Case 1
Screw (3x5) 4
Screw (3x8) 5

Note: Remove inner battery first when disassembling the product.

- 34 - Argus OXM C Service Manual


Troubleshooting

Summary
Troubleshooting
Technical support

WARNING: If measured value of the equipment seems to be incorrect, you have to


check the patient with other verified method. Also make sure to check if the
equipment operates properly.

WARNING: Only qualified service personnel can disassemble and check inside of
the equipment of Argus OXM C. Use can’ t repair inside of the equipment directly.

Summary
At upper part of LCD screen of Argus OXM C, the following status messages are displayed.

Table 8. Status message


Status message Description
SpO2 HIGH LIMIT Measured value is over SpO2 HIGH LIMIT

SpO2 LOW LIMIT Measured value is not over SpO2 LOW LIMIT

PR HIGH LIMIT Measured value is over PR HIGH LIMIT

PR LOW LIMIT Measured value is not over PR LOW LIMIT

SIGNAL WEAK Measuring signal is weak

SIGNAL NOT DETECT Measurement is impossible as measuring value is too weak

NO SENSOR Sensor is not mounted at the monitor

SENSOR OFF Sensor is not attached at the patient

PULSE SEARCH It starts to measure the signal as a sensor is attached at the patient

DEMO MODE It indicates Demo Mode

- 35 - Argus OXM C Service Manual


Troubleshooting
If any problem of the system occurs during the use of Argus OXM C, please contact service
personnel of the hospital, agency or Customer Service Center of Medical Supply Co., Ltd.
Details on troubleshooting of the system of Argus OXM C is described in the service
manual.

Please respond as follows if any problem occurs during the use.

1. If the equipment is not turned on when pressing Power Button.

● It may be the problem of power fuse. Please request service personnel of Medical
Supply to exchange the power fuse.
● If Argus OXM C is operated by battery, charge the battery. Refer to ‘ Battery Operation’
Operation’
Chapter of this manual.

2. If the screen is not displayed or no sound when the equipment is turned on.

● Stop using the equipment immediately and contact service personnel of the hospital,
agency or Customer Service Center of the company.

2. It indicates that it is operated by battery despite being operated by DC power.

● Check if power code is connected correctly.


● Check if other equipment that uses DC power is turned on.
● If the equipment detects DC power error, it operates by inner battery.

- 36 - Argus OXM C Service Manual


Main power problem

- 37 - Argus OXM C Service Manual


Battery problem

Sound problem

- 38 - Argus OXM C Service Manual


Oxygen saturation (SpO2) measurement problem

Technical support
Please refer to the Service manual if you need information on technical support or other
equipments. Contact Domestic Business Division or Customer Service Center of the
company if you need Service Manual. The Service Manual contains useful information such
as inspection and repair.
When requesting service to the company, please let us know the system (Firmware) version
of the equipment.

- 39 - Argus OXM C Service Manual


Default Setting

Summary
Setting range and default setting of the equipment

Summary
Table 10 shows default setting set in the factory. You may change into required value. If
the system is terminated with changed setting, changed setting will be stored. To initialize
the settings, please refer to Menu Setting Chapter of the User Manual.
Manual

Setting range and default setting of the equipment

Table 9. Setting range and default setting


Parameter Setting range Default setting
Oxygen
Oxygen saturation Newborn Low:85 - 99 High:86 - 100 Low 85/ High 100
(SpO2) Child Low:85 – 99 High:86 - 100 Low 85/ High 100
Alarm limit (%) Adult Low:85 – 99 High:86 - 100 Low 85/ High 100
Newborn Low:25 – 245 High:20 - 250 Low100/ High 200
Pulse rate (PR)
Child Low:25 – 245 High:20 – 250 Low 75/ High 160
Alarm limit (bpm)
bpm)
Adult Low:25 - 245 High:20 – 250 Low 50/ High 140
Patient age setting Newborn, child, adult Adult
Trend data st
storage time interval 1 ~ 1800(sec) 10(sec)
Sound volume setting 0~ 9 5
Alarm mute time setting 0 ~ 1800(sec) 10(sec)
Screen display setting Graph, Normal, Pulse Bar Graph
Trend data display setting Graph Graph
Time setting 00:00 ~ 23:59 00:00

- 40 - Argus OXM C Service Manual


System Process

System Block Diagram


System description

System Block Diagram


Diagram

Figure 27. System Block Diagram

Argus OXM C is oxygen saturation measuring device which can measure oxygen saturation
and pulse rate non- invasively for adult/ infant and newborn baby. Microprocessor is
mounted to monitor and display vital signs which are measured additionally.
If the measurements are over the limit value set, it generates visual and sound alarm.
For reasonable and correct measurement, it detects the status of the equipment and
displays error code on the screen.
Measured data and error code can be stored and seen again.

Argus OXM C is operated by battery power. It is designed to charge the battery connecting
DC power.

- 41 - Argus OXM C Service Manual


System description
Input power
HP- 110 system is designed separately with module operation power and digital circuit
operation power. The power of the part to be contacted with the patients is designed
separately. If power supply of the system is unstable measured vital signs may have error and
it may cause severe problem on the trust of the equipment. To prevent this, a function to
monitor the system power is mounted from system designing stage. The device should be
able to be operated with portable battery and it should be operated at lest for 3 hours. It
should be designed to be operated with inner battery at least one month after shutting down

Table 10. Power supply and details


Power supplied
Detail
(Tolerance)
It is used for the device. It supplied the power to Power Module and
DC 12V (± 1%) provides the power required for charging/ discharging and each
module.
DC 5V(4.65V~ 5.3V) MCU driving power for charge/ discharge control of Power Module.
LED 5V(± 4%) It is applied to LED driving module of Spo2 Module.
It is the power supplied to Main Module separated from Power
DC 1.2V(± 3%) Module. It is supplied continuously after system shuts down to
guarantee Low- Power- Mode of the microprocessor in the Module.
Analog 5V (± 2%) It is used to process analogue signal of Spo2 Module.
It is used as power source of Main Module MCU and Spo2 Module
DC 3.3V (± 4%)
MCU.

Adapter power part Inner battery power part

Figure 28. Fuse and circuit diagram of Argus OXM C

Using SMPS (Switch Mode Power Supply), Argus OXM C converts 12V input power from DC
adapter to DC 5V and DC 3.3V.

- 42 - Argus OXM C Service Manual


Battery charge
Battery will be charged immediately after connecting DC adaptor as Argus OXM C has
internal battery charging circuit. Battery is fully charged in approximately 120 minutes.
Charging is stopped if battery is fully charged or charging time is over. If connecting DC
power after disconnecting, charging will be started again. During charging, the equipment
operates with power of DC adapter.

Battery
Argus OXM C uses Li- ion battery. Battery has a capacity to use the equipment for 5 hours
with no alarm condition by 7.4V/ 1.8A. It will take about 200 minutes to charge a discharged
battery fully.

MCU
By using Micro- controller, Argus OXM C manages all operation of the equipment such as
LCD Driver, Sound and Key. Also, it checks the condition of the power and system, and
notifies the user alarm and error code when problem occurs.

Key
By using 6 Keys, user can operate the function easily such as system and alarm limit
setting, data display, display mode setting and alarm stop. You can turn On/ Off the
equipment by using power switch.

Display
Argus OXM C has 3.5” TFT Color LCD. It makes you check system status with eye such as
measures, current time and battery charge status. Also it displays the equipment or
measurement through error code.

Sound
Sound
It notifies the user of current status by generating alarm or pulse sound, Key sound,
measurement completion sound and power On/ Off sound through a speaker.

- 43 - Argus OXM C Service Manual