Sei sulla pagina 1di 3

M e d i c a l P hy s i c s a n d I n f o r m a t i c s • C l i n i c a l Pe r s p e c t i ve

Marsh and Silosky


Patient Shielding During Diagnostic Imaging

Medical Physics and Informatics


Clinical Perspective
Downloaded from www.ajronline.org by Pedro Alberto Vidal Anderson on 06/06/19 from IP address 189.24.92.105. Copyright ARRS. For personal use only; all rights reserved

Patient Shielding in Diagnostic


FOCUS ON:

Imaging: Discontinuing a
Legacy Practice
Rebecca M. Marsh1 OBJECTIVE. Patient shielding is standard practice in diagnostic imaging, despite grow-
Michael Silosky ing evidence that it provides negligible or no benefit and carries a substantial risk of increas-
ing patient dose and compromising the diagnostic efficacy of an image. The historical ratio-
Marsh RM, Silosky M nale for patient shielding is described, and the folly of its continued use is discussed.
CONCLUSION. Although change is difficult, it is incumbent on radiologic technologists,
medical physicists, and radiologists to abandon the practice of patient shielding in radiology.

atient shielding is an integral part U.S. Code of Federal Regulations has not

P of radiology. Its practice and im-


portance are so deeply ingrained
that when a group of radiologic
changed from the initial wording found in the
1976 version [9]. Patient shielding was—and
is—justified as a matter of protection from
technologists was recently asked what they hereditary risks, not as an overall reduction in
would do if their institution adopted a policy stochastic risk. Of importance, 42 years later,
to not provide patient shielding, 86% of re- no hereditary effects from radiation have ever
spondents stated that they would shield pa- been observed in humans [10].
tients anyway. (One percent of respondents
said that such a policy change would cause Patient Shielding Provides Negligible
them to quit their job [1].) This raises an im- (or No) Benefit
portant question: why do we shield patients? In addition to increased data about radiation
The assumption is that shielding improves pa- effects, any risk that may exist would be much
tient safety [2–6]. This belief is often regarded lower now than it was in the 1970s simply be-
as fact, with little consideration given to its ve- cause of the drastic decrease in the amount of
racity. However, a review of the history of pa- radiation used in radiography. In 1959, the ra-
tient shielding and the current role of patient diation dose to the testes of a 4-year-old male
shielding in radiology provides evidence that patient undergoing anteroposterior exami-
the associated risks are substantial, whereas nation of the pelvis was approximately 2.5
the benefits are negligible or nonexistent. mGy. The radiation dose to the ovaries of a
female patient undergoing the same examina-
Patient Shielding Was Intended to tion was 1.2 mGy [11]. By 2012, those doses
Keywords: gonadal shielding, patient safety, patient Alleviate Hereditary Risks had been reduced to approximately 0.06 mGy
shielding, radiation, shielding Patient shielding was first introduced into and 0.01 mGy to the testes and ovaries, re-
doi.org/10.2214/AJR.18.20508
the U.S. Code of Federal Regulations in 1976 spectively [12], for a reduction of more than
[7]. Around this time, it was recognized that 96%. Even data about fetal dose suggest that
Received August 6, 2018; accepted without revision radiation exposure from diagnostic examina- at radiation doses of less than 100 mGy, the
August 28, 2018. tions was too low to affect fertility, because risk to an embryo or fetus is either small or
1
even temporary decreases in sperm count nonexistent [13]. Multiple studies have shown
Both authors: University of Colorado School of
Medicine, 12700 E. 19th Ave, Mail Stop C278, Aurora, CO
do not occur at doses of less than 250 mGy, that fetal doses from radiographic and CT ex-
80045. Address correspondence to R. M. Marsh and because female fertility is not affected at aminations are well below this amount. In ra-
(rebecca.marsh@ucdenver.edu). ­doses of less than 3000 mGy [8]. Consequent- diography, even when the fetus is in the pri-
ly, the regulation cited only a concern regard- mary x-ray beam, the fetal dose is less than 4
AJR 2019; 212:755–757 ing hereditary risks (i.e., mutations in germ mGy [6]. During CT examination for pulmo-
0361–803X/19/2124–755
cells that may affect future generations) and nary embolism in the mother, the fetal dose is
addressed gonadal shielding only. The word- approximately 1.5 mGy; CT examinations of
© American Roentgen Ray Society ing in the April 2018 version of title 21 of the the abdomen and pelvis result in a fetal dose

AJR:212, April 2019 755


Marsh and Silosky

ranging from 15 to 20 mGy (depending on the for all examinations, including those of pe- ing FOV. Although this greatly improves the
scanner type) [14]. To our knowledge, no evi- diatric and pregnant patients [13, 15, 22–25]. consistency of image quality (signal intensity,
dence exists to indicate that a single imaging For anatomy that is within the imaging FOV, noise properties, and other factors), the con-
study poses any risk to a fetus [15, 16]. the use of patient shielding may reduce patient sequence of introducing a highly attenuating
Downloaded from www.ajronline.org by Pedro Alberto Vidal Anderson on 06/06/19 from IP address 189.24.92.105. Copyright ARRS. For personal use only; all rights reserved

Although stochastic radiation effects still dose, but this potential dose savings comes material into the imaging area is significant.
are not fully understood, the data that do ex- at the risk of inadvertently increasing patient If a lead shield, which is meant to protect the
ist are commonly misrepresented. For exam- dose or adversely affecting the diagnostic ef- patient, enters the imaging FOV, the radio-
ple, cumulative effects of radiation are used ficacy of the examination. graphic system will drastically increase the
to defend current patient shielding practices. Since gonadal shielding was introduced in tube output to try to penetrate the shield. This
It is convenient to assume that the risk to a the U.S. Code of Federal Regulations, sub- results in an increased dose to the patient and
neonate who undergoes 20 radiographic ex- stantial changes in imaging technology, from a marked degradation in image quality.
aminations in the neonatal ICU is the same x-ray generation to detection and image for- In addition to the risks posed by automatic
as if the same total amount of radiation is de- mation, have further reduced any benefit exposure control, several clinical studies have
livered during a single examination; howev- from shielding. Collimation in modern radio- shown that gonadal shields are often posi-
er, this assumption completely ignores abun- graphic systems has drastically improved; the tioned incorrectly, obscuring relevant anato-
dant evidence to the contrary. The effects of requirements for minimum amounts of beam my and increasing repeat rates [12, 28, 29]. A
varying the rate at which radiation is deliv- filtration have increased [26]. Shorter exam- study by Frantzen et al. [12] found that gonad-
ered is well documented and is even exploited ination times have expanded the applicabil- al shields were incorrectly placed for 91% of
by radiation oncologists, who use treatment ity of automatic exposure control to a wider pelvic radiographic examinations of girls and
fractionation to minimize damage to healthy range of pediatric patients [27]. The increased 66% of those of boys. Another group of inves-
tissues [17]. A thorough analysis of the lin- dynamic range of digital image receptors and tigators found that pelvic shields were mis-
ear no-threshold model is beyond the scope advances in digital image processing have al- placed in 49% of anteroposterior radiographs
of this article, but there are two highly rele- lowed further reductions in the amount of ra- and 63% of frog lateral radiographs and that
vant key points. First, epidemiologic studies diation needed to produce a diagnostic-qual- pelvic bony landmarks were obscured by
do not support the linear no-threshold model ity image. In addition, radiology has moved shielding in up to 43% of images [30]. The
at doses less than 100 mSv [18]. Second, most from a field using conventional radiographs investigators concluded that “consideration
data show that the biologic effects of radia- only to one that is heavily dominated by digi- should be given to alternative protocols or
tion delivered at low dose rates vary substan- tal image receptors, digital image processing, abandonment of this practice.” Of even more
tially from those of acute doses of radiation and automatic exposure techniques. concern is the fact that the study also found
[19]. The Biologic Effects of Ionizing Radia- that retakes were warranted in many cases
tion VII report contains a large discussion of Patient Shielding Introduces but were not performed. This forces the ra-
dose-rate dependence and includes a correc- Significant Risks diologist to fill in the gaps with information
tion factor to account for how it affects the Although these developments can reduce from previous images, negating much of the
linear no-threshold model [10]. Some groups, patient dose and improve image quality and benefit from the current examination.
including the French Académie Nationale de consistency, the efficacy of patient shield- It is important to recognize that the prac-
Médicine, even suggest that small amounts of ing must be reconsidered with these tech- tice of shielding patients is largely supported
radiation delivered at low dose rates may have nologies in mind. For example, most modern by a skewed perception of radiation risk. This
a protective effect [18, 20]. In other words, x-ray–based imaging systems have some sort is also the most difficult aspect to address
the assumption that the risk from 20 imag- of automatic exposure control. In radiography, through rational discussion. The challenges
ing studies, each with an effective dose of 10 photocells are built into the image receptor to associated with gaining public trust in health
mSv, is the same as that from a single imag- sense when the detector has received a target care are substantial and have been previous-
ing study with an effective dose of 200 mSv dose, allowing the automatic exposure control ly discussed elsewhere [31, 32]. In a discus-
is unfounded. This point is even more impor- algorithm to determine an optimal tube cur- sion of the public perception of radiation risk,
tant when one considers the very low doses rent–exposure time product. In fluoroscopy, Hendee [31] placed the onus on medical pro-
encountered in planar radiography and the automatic brightness control algorithms work fessionals to use their topical expertise to be-
risks associated with not shielding patients. as a feedback loop between the image recep- come involved in such issues: “If the sources
One must also consider the amount of pro- tor and the x-ray tube so that the tube output is of reason and wisdom in the community are
tection that shielding provides to a patient. constantly adjusted to ensure consistent image silent, only irrational and foolish voices will
This varies based on whether anatomy is lo- quality. In CT, automatic tube current modu- be heard.” He concluded by stating, “Enough
cated outside the imaging FOV (i.e., not in the lation and automatic tube voltage modulation examples of these effects exist today in our
path of the primary x-ray beam) or inside the use localizer images to determine what tube society to suggest that reasonable voices have
imaging FOV. For anatomy outside the imag- current and tube voltage are needed to create been silent long enough” [31].
ing FOV, radiation exposure results almost en- diagnostic quality images. The details regard-
tirely from internal scatter generated within a ing how these systems work are beyond the Practical Implementation of a
patient [21]. Because contact shielding cannot scope of this article, but the takeaway is that No-Shielding Practice
protect against internal scatter, shielding anat- each of these techniques depends on the sys- Discontinuing the use of patient shielding
omy outside the imaging FOV provides negli- tem optimizing image quality by adjusting the will be a significant departure from how radi-
gible protection to the patient. This holds true radiation output based on what is in the imag- ology has been practiced for decades. Although

756 AJR:212, April 2019


Patient Shielding During Diagnostic Imaging

the need to abandon this practice is clear, it is com/2018/01/15/what-would-you-do-stop-shield- 18. Tubiana M. Dose-effect relationship and estima-
incumbent on health care professionals to help ing-your-patients. Published January 15, 2018. tion of the carcinogenic effects of low doses of
patients feel confident about the care they re- Accessed July 18, 2018 ionizing radiation: the joint report of the Acadé-
ceive. This is especially important regarding 2. Gomes M, Matias A, Macedo F. Risks to the fetus mie des Sciences (Paris) and of The Académie
Downloaded from www.ajronline.org by Pedro Alberto Vidal Anderson on 06/06/19 from IP address 189.24.92.105. Copyright ARRS. For personal use only; all rights reserved

issues of radiation risk, for which misinforma- from diagnostic imaging during pregnancy: re- Nationale de Médicine. Int J Radiat Oncol Biol
tion is rampant. Consequently, how to address view and proposal of a clinical protocol. Pediatr Phys 2005; 63:317–319
patient concerns as facilities stop providing pa- Radiol 2015; 45:1916–1929 19. Brooks AL, Hoel DG, Preston RJ. The role of dose
tient shielding should be considered. 3. Pahade JK, Litmanovich D, Pedrosa I, et al. Qual- rate in radiation cancer risk: evaluating the effect
First, it may be beneficial to address patient ity initiatives: imaging pregnant patients with sus- of dose rate at the molecular, cellular and tissue
concerns before the examination begins. For pected pulmonary embolism: what the radiologist levels using key events in critical pathways fol-
example, when greeting a patient, the technol- needs to know. RadioGraphics 2009; 29:639–654 lowing exposure to low LET radiation. Int J
ogist typically introduces himself or herself, 4. Ratnapalan S, Bentur Y, Koren G. Doctor, will ­Radiat Biol 2016; 92:405–426
verifies the patient’s identity, and describes that x-ray harm my unborn child? CMAJ 2008; 20. Moghissi AA, Calderone R, Azam F, et al. Regu-
the examination to the patient. At this time, it 179:1293–1296 [Erratum in CMAJ 2009; 280:952] lating ionizing radiation based on metrics for
may be appropriate to inform the patient that 5. Sikand M, Stinchcombe S, Livesley PJ. Study on evaluation of regulatory science claims. Dose
the facility does not provide patient shielding the use of gonadal protection shields during pae- ­Response 2018; 16:1559325817749413
for imaging examinations because evidence diatric pelvic x-rays. Ann R Coll Surg Engl 2003; 21. American College of Radiology (ACR). ACR–
indicates that shielding increases the risk to 85:422–425 AAPM–SIIM–SPR practice parameter for digital
the patient and provides negligible or no ben- 6. Ursprung WM, Howe JW, Yochum TR, Kettner radiography. ACR website. www.acr.org/-/media/
efit. This gives the patient the opportunity to NW. Plain film radiography, pregnancy, and thera- ACR/Files/Practice-Parameters/Rad-Digital.pdf.
ask questions and express any concerns he or peutic abortion revisited. J Manipulative Physiol Accessed July 23, 2018
she may have. In addition, information in the Ther 2006; 29:83–87 22. Patel SJ, Reede DL, Katz DS, et al. Imaging the
form of posters or brochures can provide in- 7. 21 C.F.R. §1000.50 (1976) pregnant patient for nonobstetric conditions: algo-
formation to patients before an appointment, 8. International Commission on Radiological Pro- rithms and radiation dose considerations.
either online or in a waiting room. The con- tection (ICRP). Recommendations of the Interna- RadioGraphics 2007; 27:1705–1722
cerns of many patients may be alleviated if the tional Commission on Radiological Protection: 23. Ryckx N, Sans-Merce M, Schmidt S, et al. The use
patients know that someone is paying atten- ICRP publication 26. Ann ICRP 1977; 1(3) of out-of-plane high Z patient shielding for fetal
tion to their safety and that the lack of shield- 9. 21 C.F.R. §1000.50 (2018) dose reduction in computed tomography: litera-
ing is intentional rather than negligent. 10. National Research Council. Health risks from ex- ture review and comparison with Monte-Carlo
Last, it is important to give technologists posure to low levels of ionizing radiation: BEIR calculations of an alternative optimization tech-
discretion to provide shielding in certain cir- VII phase 2. Washington: DC: The National nique. Phys Med 2018; 48:156–161
cumstances. Patients who are extremely anx- Academies Press, 2006 24. Sadro C, Bernstein MP, Kanal KM. Imaging of
ious about the lack of shielding should be ad- 11. Bishop HA, Webber M, O’Loughlin BJ. Reducing trauma. Part 2. Abdominal trauma and pregnancy:
vised of the potential risks. If the technologist gonad irradiation in pediatric diagnosis. Calif a radiologist’s guide to doing what is best for the
still determines that shielding would provide a Med 1959; 90:20–25 mother and baby. AJR 2012; 199:1207–1219
substantial psychologic benefit to the patient, 12. Frantzen MJ, Robben S, Postma AA, et al. Gonad 25. Sidhu M, Strauss K, Connolly B, et al. Radiation
he or she should be allowed to make this pro- shielding in paediatric pelvic radiography: disad- safety in pediatric interventional radiology. Tech
fessional judgment. However, it is important vantages prevail over benefit. Insights Imaging Vasc Interv Radiol 2010; 13:158–166
to emphasize to medical staff that shielding 2012; 3:23–32 26. 21 C.F.R. § 020.30 (2017)
should be avoided whenever possible. 13. National Council on Radiation Protection and 27. National Council on Radiation Protection and
Measurements (NCRP). Preconception and pre- Measurements (NCRP). Radiation protection in
Conclusion natal radiation exposure: health effects and pro- pediatric radiology. Report 68. Bethesda, MD:
Patient shielding persists despite growing tective guidance. Report 174. Bethesda, MD: NCRP Publications, 1978
evidence that the practice should be aban- NCRP Publications, 2013 28. Fawcett SL, Barter SJ. The use of gonad shielding
doned. Although change is difficult, it is in- 14. Gilet AG, Dunkin JM, Fernandez TJ, et al. Fetal in paediatric hip and pelvis radiographs. Br J Ra-
cumbent on radiologic technologists, medi- radiation dose during gestation estimated on an diol 2009; 82:363–370
cal physicists, and radiologists to finally step anthropomorphic phantom for three generations 29. Winfeld M, Strubel N, Pinkney L, et al. Relative
up as reasonable voices on the subject. Un- of CT scanners. AJR 2011; 196:1133–1137 distribution of pertinent findings on portable neo-
til then, training programs, health care facili- 15. Colletti PM, Lee KH, Elkayam U. Cardiovascular natal abdominal radiographs: can we shield the
ties, and accreditation and regulatory bodies imaging of the pregnant patient. AJR 2013; gonads? Pediatr Radiol 2013; 43:1295–1302
will continue to encourage and engage in a 200:515–521 30. Lee MC, Lloyd J, Solomito MJ. Poor utility of go-
legacy practice that presents substantial risk 16. Wieseler KM, Bhargava P, Kanal KM, et al. Imag- nadal shielding for pediatric pelvic radiographs.
but negligible (or no) benefit to patient health. ing in pregnant patients: examination appropriate- Orthopedics 2017; 40:e623–e627
ness. RadioGraphics 2010; 30:1215–1229 31. Hendee WR. Personal and public perceptions of ra-
References 17. Arnold KM, Flynn NJ, Raben A, et al. The impact diation risks. RadioGraphics 1991; 11:1109–1119
1. Advanced Health Education Center. What would of radiation on the tumor microenvironment: ef- 32. Freudenberg LS, Beyer T. Subjective perception
you do? Stop shielding your patients? Advanced fect of dose and fractionation schedules. Cancer of radiation risk. J Nucl Med 2011; 52(suppl
Health Education Center Online website. ahecblog. Growth Metastasis 2018; 11:1179064418761639 2):29S–35S

AJR:212, April 2019 757

Potrebbero piacerti anche