COMPILATION
11. General purpose of radiation imaging system have a minimum total filtration of?
a. 2mm Al
b. 5mm Pb
c. 2.5mm Rb
d. 2mmRb
12. To compensate for loss of film density when a grid used the mAs should be?
a. Decrease
b. mAs should be multipled by 4
c. Increase
d. mAs should be divided by 4
13. A stationary grid is?
Answer: fix in the bucky
14. to test for poor screen film contact is?
a. open the cassette and look
b. place a sheet of fine wire mesh inside the cassette a film and make exposure
c. x ray a lot of paper clips
d. place a sheet of fine wire mesh and exposure together with paper clips
15. How to check x-ray field alignment?
a. look in the collimator mirror
b. open close shutter rapidy
c. place the cassette adjacent to the light field measuring tools and expose
d. place metal markers on the face of cassette and expose
16. Coincidence of x ray and light field
a. 15mm inside the light field
b. 8mm inside the light field
c. 10mm outside the light field
d. 3mm outside the light field
17. Fixing time in manual processing
b. twice the clearing time
18. Automatic processor developer temperature
c. 35
19. Manual processing film should be
a. agitated every 30 seconds in developer
20. When transferring the film from the rinse to fixer, film should be drained
a. rinse
b. fixer
c. doesn’t matter
d. drip drying
21. When the lid has accidentally left off of box of unexposed film in white light
a. throw away all the films
b. put the lid back on and do nothing
c. process film and inspect
d. developed exposed film and inspect
22. The test to developed optimum time
a. develop test strings for different time and compare
b. devlelop test strips for the same length of time
c. develop test strips for the same length of time
d. develop an exposed film and inspect
23. Developer temperature should be checked
a. only when film densities look different
b. once a week
c. annually
d. daily
26. To test alignment of xray beam to grid the ff. test should be done?
a. Sheet of perforated zinc
b. Lead strip and series of 1cm holes on it
c. 8peso coins
d. All of the above
27. Film screen contact test tool must be?
a. Plastic sheet
b. Perforated aluminum sheet
c. Fine mesh metal
d. Perforated lead sheet
28. The indicated level of tomographic section and actual level of section must corresponds to
plus minus __ mm
a. 2
b. 5
c. 10
d. 15
29. Requires sheet of lead to be placed on cassette and exposed.
a. point speed function (PSF)
b line speed function (LSF)
c. Edge speed function (ESF)
d. modulation transfer function (MTF)
32. Is a graph that is more accurate and easily to obtain than the PSF graph?
a. PSF
b. LSF
c. ESF
d. MTF
33. Any variation of kVp in the panel and measured kVp must be plus minus __%?
a. 2
b. 5
c. 10
d. 20
34. Which type of x ray generator has the least amount of voltage ripple?
a. Single phase - half wave rectified
b. Single phase- full wave rectified
c. 3phase- 6pulse
d. 3phase- 12 pulse
35. The 8 or 9 penny test is used to evaluate
a. Image receptor- radiation field alignment
b. X-ray beam- central ray congruency
c. X-ray field rad congruency
d. SID indicator
36. Collimator should be assessed?
a. Semi annual
b. Weekly
c. Monthly
d. Never
37. QC evaluation performance of camera system without collimator/
a. Extrinsic
b. Intrinsic
c. Phantom
d. deadtime
38. Required darkroom ventilation?
a. 6-8 air changes
b. 8-10 " "
c. 10-12 " "
d. 12-15
39. Required kVp for knee?
a. Daily
b. Weekly
c. Monthly
d. Annually (ANSWER)
40. Spatial resolution can be improved by increasing
a. FOV
b. matrix
c.
d. slice thickness
41. Filtration should be evaluated
a. daily
b. weekly
c. monthly
d. annually
42. The 2 most importance QC procedure in scintillation camera?
43. When the films are stored in high temperatureplace, it will produce?
*fog
44. If the film is stored in 10’C, how long can it be utilized
a. 3 months
b. 2 months
c. 12 months
d. 6 months
45. If the film is stored in 15.6’C, how long can it be utilized?
a. 3 months
b. 2 months
c. 12 months
d. 6 months
66. a radiographic is made using 300mA 0.01s 75kVp resulting unsharpness motion. Otherwise
statisfactory technical, the radiograph repeated using 86mA 500kV, What is the new exposure
time?
a. 0.12
b. 0.03
c. 0.06
d. 0.01
67. The stepper the straight line portion of characteristic curve for a particular.
i. Slower the film
ii. Higher the film
iii. Greater exposure
a) 1 only
b) 2&3 only
c) 1&3
d) All of the above
68. A change from 200 speed screen would require what change in?
a. mAs should be increased by 15
b. mAs should be increased by 30
c.mAs should be doubled
d. mAs should be haved
69. a particular mAs regardless of the combination of mA & time will reproduce the same
radiographic density?
a. line focus principle
b. reciprocity law
c. inverse square law
d.lack of a ______ of energy