Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
com
International Journal of Pharmaceutical Quality Assurance 2013; 4(3); 42-51
Review Article
ABSTRACT
Gas chromatography–mass spectrometry (GC-MS) is a method that combines the features of gas-liquid chromatography
and mass spectrometry to identify different substances within a test sample. Analytical methods development and validation
play important roles in the discovery, development and Manufacture of pharmaceuticals. Method development is the
process of proving that an analytical method is acceptable for use to measure the concentration of an API in a specific
compounded dosage form which allow simplified procedures to be employed to verify that an analysis procedure,
accurately and consistently will deliver a reliable measurement of an active ingredient in a compounded preparation. The
analytical method validation is essential for analytical method development and tested extensively for specificity, linearity,
accuracy, precision, range, detection limit, quantization limit, and robustness. In summary, analytical method development
and validation allows to confirm that an accurate and reliable potency measurement of a pharmaceutical preparation can
be performed.
Key words:
INTRODUCTION
Gas chromatography–mass spectrometry (GC-MS) is a instruments5. In 1967, the Finnigan Instrument
method that combines the features of gas-liquid Corporation was formed and in early 1968, delivered the
chromatography and mass spectrometry to identify first prototype quadrupole GC/MS instruments to Stanford
different substances within a test sample1. Applications of and Purdue University.
GC-MS include drug detection, fire investigation, In 1996 the high-speed GC-MS units completed analysis
environmental analysis, explosives investigation, and of fire accelerants in less than 90 seconds, whereas first-
identification of unknown samples. GC-MS can also be generation GC-MS would have required at least 16
used in airport security to detect substances in luggage or minutes. By the 2000s computerized GC/MS instruments
on human beings. Additionally, it can identify trace using quadrupole technology had become both essential to
elements in materials that were previously thought to have chemical research and one of the foremost instruments
disintegrated beyond identification2. used for organic analysis6. Today computerized GC/MS
GC-MS has been widely heralded as a "gold standard" for instruments are widely used in environmental monitoring
forensic substance identification because it is used to of water, air, and soil; in the regulation of agriculture and
perform a specific test. A specific test positively identifies food safety and in the discovery and production of
the actual presence of a particular substance in a given medicine7.
sample. A non-specific test merely indicates that a This guidance provides general recommendations for
substance falls into a category of substances. Although a bioanalytical method validation. The recommendations
non-specific test could statistically suggest the identity of can be adjusted or modified depending on the specific type
the substance, this could lead to false positive of analytical method used.
identification3. Instrumentation: The GC-MS is composed of two major
History: The use of a mass spectrometer as the detector in building blocks: the gas chromatograph and the mass
gas chromatography was developed during the 1950s after spectrometer. The gas chromatograph utilizes a capillary
being originated by James and Martin in 1952. These column which depends on the column's dimensions
sensitive devices were originally limited to laboratory (length, diameter, film thickness) as well as the phase
settings4. properties (e.g. 5% phenyl polysiloxane). The difference in
In 1964, Electronic Associates, Inc. (EAI), a leading U.S. the chemical properties between different molecules in a
supplier of analog computers, began development of a mixture will separate the molecules as the sample travels
computer controlled quadrupole mass spectrometer[2] the length of the column. The molecules are retained by the
under the direction of Robert E. Finnigan. By 1966 column and then elute (come off of) from the column at
Finnigan and collaborator Mike Uthe's EAI division had different times (called the retention time), and this allows
sold over 500 quadrupole residual gas-analyzer the mass spectrometer downstream to capture, ionize,
accelerate, deflect, and detect the ionized molecules The electrons bombard the molecules, causing the
separately. The mass spectrometer do this by breaking molecule to fragment in a characteristic and reproducible
each molecule into ionized fragments and detect the way. This "hard ionization" technique results in the
fragments8. GC-MS schematic: These two components, creation of more fragments of low mass to charge ratio
used together, allow a much finer degree of substance (m/z). Hard ionization is considered by mass
identification than used separately. It is not possible to spectrometrists as the employ of molecular electron
make an accurate identification of a particular molecule by bombardment, whereas "soft ionization" is charge by
gas chromatography or mass spectrometry alone. The mass molecular collision with an introduced gas. The molecular
spectrometry process normally requires a very pure sample fragmentation pattern is dependant upon the electron
while gas chromatography using a traditional detector (e.g. energy applied to the system, typically 70 eV (electron
Flame ionization detector) cannot differentiate between Volts)13. The use of 70 eV facilitates comparison of
multiple molecules that happen to take the same amount of generated spectra with library spectra using manufacturer-
time to travel through the column (i.e. have the same supplied software or software developed by the National
retention time), which results in two or more molecules Institute of Standards (NIST-USA).
that co-elute9. Sometimes two different molecules can also Cold electron ionization: The "hard ionization" process of
have a similar pattern of ionized fragments in a mass electron ionization can be softened by the cooling of the
spectrometer (mass spectrum). Combining the two molecules before their ionization, resulting in mass spectra
processes reduces the possibility of error. that are richer in information. In this method named cold
Purge and trap GC-MS: For the analysis of volatile electron ionization (Cold-EI) the molecules exit the GC
compounds a purge and trap (P&T) concentrator system column, mixed with added helium make up gas and expand
may be used to introduce samples. The target analytes are into vacuum through a specially designed supersonic
extracted and mixed with water and introduced into an air nozzle, forming a supersonic molecular beam (SMB).
tight chamber. An inert gas such as Nitrogen (N2) is Collisions with the make up gas at the expanding
bubbled through the water; this is known as purging. The supersonic jet reduce the internal vibrational (and
volatile compounds move into the headspace above the rotational) energy of the analyte molecules, hence reducing
water and are drawn along a pressure gradient (caused by the degree of fragmentation caused by the electrons during
the introduction of the purge gas) out of the chamber. The the ionization process14. Cold-EI mass spectra are
volatile compounds are drawn along a heated line onto a characterized by an abundant molecular ion while the usual
'trap'. The trap is a column of adsorbent material at ambient fragmentation pattern. The enhanced molecular ions
temperature that holds the compounds by returning them increase the identification probabilities of both known and
to the liquid phase. The trap is then heated and the sample unknown compounds, amplify isomer mass spectral
compounds are introduced to the GC-MS column via a effects and enable the use of isotope abundance analysis
volatiles interface, which is a split inlet system. P&T GC- for the elucidation of elemental formulae15.
MS is particularly suited to volatile organic compounds Chemical ionization: In chemical ionization a reagent gas,
(VOCs)10. typically methane or ammonia is introduced into the mass
Types of mass spectrometer detectors: The most common spectrometer. Depending on the technique (positive CI or
type of mass spectrometer (MS) associated with a gas negative CI) chosen, this reagent gas will interact with the
chromatograph (GC) is the quadrupole mass spectrometer, electrons and analyte and cause a 'soft' ionization of the
sometimes referred to by the Hewlett-Packard (now molecule. A softer ionization fragments the molecule to a
Agilent) trade name "Mass Selective Detector" (MSD). lower degree than the hard ionization of EI. One of the
Another relatively common detector is the ion trap mass main benefits of using chemical ionization is that a mass
spectrometer. Additionally one may find a magnetic sector fragment closely corresponding to the molecular weight of
mass spectrometer, however these particular instruments the analyte of interest is produced.
are expensive and bulky and not typically found in high- In positive chemical ionization (PCI) the reagent gas
throughput service laboratories. Other detectors may be interacts with the target molecule, most often with a proton
encountered such as time of flight (TOF), tandem exchange. This produces the species in relatively high
quadrupoles11. amounts16.
Ionization: After the molecules travel the length of the In negative chemical ionization (NCI) the reagent gas
column, pass through the transfer line and enter into the decreases the impact of the free electrons on the target
mass spectrometer they are ionized by various methods analyte. This decreased energy typically leaves the
with typically only one method being used at any given fragment in great supply.
time12. Once the sample is fragmented it will then be
detected, usually by an electron multiplier diode, which ANALYSIS
essentially turns the ionized mass fragment into an A mass spectrometer is typically utilized in one of two
electrical signal that is then detected. ways: full scan or selected ion monitoring (SIM). The
Electron ionization: In the electron ionization (EI) the typical GC-MS instrument is capable of performing both
molecules enter into the MS (the source is a quadrupole or functions either individually or concomitantly, depending
43
the ion trap itself in an ion trap MS) where they are on the setup of the particular instrument.
bombarded with free electrons emitted from a filament, not The primary goal of instrument analysis is to quantify an
Page
unlike the filament one would find in a standard light bulb. amount of substance. This is done by comparing the
relative concentrations among the atomic masses in the Environmental monitoring and cleanup: GC-MS is
generated spectrum. Two kinds of analysis are possible, becoming the tool of choice for tracking organic
comparative and original. Comparative analysis pollutants in the environment. The cost of GC-MS
essentially compares the given spectrum to a spectrum equipment has decreased significantly, and the
library to see if its characteristics are present for some reliability has increased at the same time, which has
sample in the library. contributed to its increased adoption in environmental
Another method of analysis measures the peaks in relation studies. There are some compounds for which GC-MS is
to one another. In this method, the tallest peak is assigned not sufficiently sensitive, including certain pesticides
100% of the value, and the other peaks being assigned and herbicides, but for most organic analysis of
proportionate values. All values above 3% are assigned17. environmental samples, including many major classes
The total mass of the unknown compound is normally of pesticides, it is very sensitive and effective.
indicated by the parent peak. The isotope pattern in the Criminal forensics: GC-MS can analyze the particles
spectrum, which is unique for elements that have many from a human body. The analysis of fire debris using
isotopes, can also be used to identify the various elements GC-MS is well established, and there is even an
present. Once a chemical formula has been matched to the established American Society for Testing Materials
spectrum, the molecular structure and bonding can be (ASTM) standard for fire debris analysis.
identified, and must be consistent with the characteristics Law enforcement: GC-MS is increasingly used for
recorded by GC-MS. Typically, this identification done detection of illegal narcotics.It is also commonly used
automatically by programs in the instrument, given a list in forensic toxicology to find drugs and/or poisons in
of the elements which could be present in the sample. biological specimens of suspects, victims, or the
A “full spectrum” analysis considers all the “peaks” within deceased.
a spectrum. Conversely, selective ion monitoring (SIM)
Sports anti-doping analysis: GC-MS is the main tool used
only monitors selected peaks associated with a specific
in sports anti-doping laboratories to test athletes' urine
substance. This is a fast and efficient analysis. When the
samples for prohibited performance-enhancing drugs,
amount of information collected about the ions in a given
for example anabolic steroids19.
gas chromatographic peak decreases, the sensitivity of the
Food, beverage and perfume analysis: Foods and
analysis increases. So, SIM analysis allows for a smaller
beverages contain numerous aromatic compounds,
quantity of a compound to be detected and measured, but
the degree of certainty about the identity of that compound some naturally present in the raw materials and some
is reduced. forming during processing. GC-MS is extensively used
Full scan MS: Full scan is useful in determining unknown for the analysis of these compounds which include
compounds in a sample. It provides more information than esters, fatty acids, alcohols, aldehydes, terpenes etc.
SIM when it comes to confirming or resolving compounds Astrochemistry: Venera 11 and 12 and Pioneer Venus
in a sample. During instrument method development it analysed the atmosphere of Venus with GC-MS.
may be common to first analyze test solutions in full scan Calibration Procedure/Performance Verification of
mode to determine the retention time and the mass Instrument And Equipment
fragment fingerprint before moving to a SIM instrument Introduction: The performance of laboratory instruments
method. and equipment may change with time, either in the short
Selected ion monitoring: In selected ion monitoring (SIM) term owing to fluctuations in the environment or, in the
certain ion fragments are entered into the instrument long term, owing to ageing of the mechanical, optical or
method and only those mass fragments are detected by the electronic components. Within a laboratory which
mass spectrometer. The advantages of SIM are that the maintains a comprehensive quality system, all aspects of
detection limit is lower since the instrument is only looking analytical work are controlled, and these potential
at a small number of fragments (e.g. three fragments) instrumental errors are controlled by carrying out regular
during each scan. More scans can take place each second18. preventative maintenance and calibration procedures.
Since only a few mass fragments of interest are being The way in which the performance of instruments and
monitored. It is relatively important to be sure that the ion equipment is to be monitored and the frequency of the
ratios of the various mass fragments are comparable to a calibration checks (calibration interval), should be
known reference standard. stipulated in Standard Operating Procedures (SOPs).
General uses: Performance verification should be based on tests which
are not specific to particular methods and which use
Identification and quantitation of volatile and
traceable calibrators and standards, thus allowing
semi volatile organic compounds in complex
equipment to be compared between laboratories.
mixtures.
Performance verification is not specifically related to
Determination of molecular weights and elemental
either screening or confirmatory methods. The calibration
compositions of unknown organic compounds in complex
of instruments and equipment (e.g., wavelength calibration
mixtures.
of an IR spectrometer, mass calibration of a GCMS) is
Structural determination of unknown organic compounds
44
exist:
Applications:
Method: Checked with a portable reference pyrometer or Mass spectrometers: Mass spectrometers are tuned and
precision thermometer, which should be placed as close as calibrated in a similar manner whether they are stand-alone
Page
possible to the oven temperature sensor. instruments or combined with chromatographic interfaces
(GC-MS and LC-MS and their multi-sector derivations). these drugs, reports of new toxicities (resulting in their
Differences arise between quadrupole and magnetic sector withdrawal from the market), development of patient
instruments, especially if the latter are capable of high resistance and introduction of better drugs by competitors.
resolution23. Under these conditions, standards and analytical
Most bench-top instruments are controlled directly by a procedures for these drugs may not be available in the
computer data system, and tuning and calibration are pharmacopoeias26. It becomes necessary, therefore to
carried out automatically. Warnings are generated by the develop newer analytical methods for such drugs.
data system if the instrument fails to achieve the pre-set Basic criteria for new method development of drug
performance characteristics, often mandating operator analysis:
intervention, for example to clean the source. The drug may not be official in any pharmacopoeias.
Parameters to be calibrated: Source tuning and mass A proper analytical procedure for the drug may not be
calibration. available in the literature due to patent regulations.
Method: A calibration compound such as Analytical methods may not be available for the drug in
perfluorokerosene (PFK) or heptacosafluorotributylamine the form of a formulation due to the interference caused by
(perfluorotributylamine) is introduced to the spectrometer the formulation excipients.
using a direct inlet device. The source is tuned using Analytical methods for the quantification of the drug in
selected fragment ions to give optimum sensitivity and biological fluids may not be available.
peak shape, and obtain peak ratios (for example, of m/z 69, Analytical methods for a drug in combination with other
219 and 264 and 502 in the perfluorotributylamine drugs may not be available.
spectrum) usually determined by the manufacturer. The existing analytical procedures may require expensive
Spectra are recorded and compared with the reference reagents and solvents. It may also involve cumbersome
spectrum with respect to mass assignments and relative extraction and separation procedures and these may not be
peak intensities. reliable.
Calibration interval: Daily or immediately prior to use.
Daily (Continuing) GC/MS Calibration ANALYTICAL METHOD VALIDATION
1. Prior to start of sample analysis; inject 50ng of the Analytical Method Validation is “the collection and
bromofluorobenzene tuning compound. Insure that all evaluation of data, from the process design stage
BFB tune criteria are met according to Table throughout production, which establishes scientific
2. The initial calibration curve for each compound of evidence that a process is capable of consistently
interest must be checked and verified once every 12 hours delivering quality products”.
of analysis time. This is accomplished by analyzing a Validation is an act of proving that any procedure, process,
calibration standard at the mid-point concentration for the equipment, material, activity or system performs as
working range of the GC/MS by checking the SPCC and expected under given set of conditions and also give the
CCC compounds. required accuracy, precision, sensitivity, ruggedness, etc.
3. System Performance Check Compounds (SPCCs): A When extended to an analytical procedure, depending
system performance check must be made every 12 hours. upon the application, it means that a method works
If the SPCC criteria are met, a comparison of response reproducibly, when carried out by same or different
factors is made for all compounds. If the minimum persons, in same or different laboratories, using different
response factors are not met, the system must be evaluated, reagents, different equipments, etc28.
and corrective action must be taken before sample analysis Advantages of analytical method validation: The biggest
may begin.
advantage of method validation is that it builds a
4. Calibration Check Compounds (CCCs): After the
degree of confidence, not only for the developer but
system performance check is met, CCCs listed in section.
also to the user. Although the validation exercise may
5. Above are used to check the validity of the initial
appear costly and time consuming, it results
calibration. If the % Difference (%D) for each CCC is less
inexpensive, eliminates frustrating repetitions and
than or equal to 20%, the initial calibration is considered
leads to better time management in the end. Minor
valid and may be used to quantitate sample data. Calculate
changes in the conditions such as reagent supplier or
the percent difference using:
grade, analytical setup are unavoidable due to obvious
% Difference= RFi-RFc × 100 reasons but the method validation absorbs the shock of
RFi such conditions and pays for more than invested on the
process.
ANALYTICAL METHOD DEVELOPMENT Validation parameters:
The number of drugs introduced into the market is The various parameters according to the ICH Guidelines
increasing every year. These drugs may be either new as follows29
entities or partial structural modification of the existing The fundamental parameters for this validation include
one, very often there is a time lag from the date of (1) accuracy,
46
introduction of a drug into the market to the date of its (2) precision,
inclusion in pharmacopoeias. This happens because of the (3) selectivity,
Page
samples).The mean value should be within ±15% of the methods, however, testing six independent matrices for
theoretical value, except at LLOQ, where it should not interference may not be important. In the case of LC-MS
Page
deviate by more than ±20%. The precision around the and LC-MS-MS-based procedures, matrix effects should
mean value should not exceed 15% of the CV, except for be investigated to ensure that precision, selectivity, and
sensitivity will not be compromised. Method selectivity Evidence of purity and identity of drug standards,
should be evaluated during method development and metabolite standards, and internal standards used during
throughout method validation and can continue throughout routine analyses
application of the method to actual study samples. Summary tables containing information on sample
Acceptance/rejection criteria for spiked, matrix-based processing and storage. Tables should include sample
calibration standards and validation QC samples should be identification, collection dates, storage prior to shipment,
based on the nominal (theoretical) concentration of information on shipment batch, and storage prior to
analytes. Specific criteria can be set up in advance and analysis. Information should include dates, times, sample
achieved for accuracy and precision over the range of the condition, and any deviation from protocols.
standards, if so desired. Summary tables of analytical runs of clinical or preclinical
validation documentation/protocol: The validity of an samples. Information should include assay run
analytical method should be established and verified by identification, date and time of analysis, assay method,
laboratory studies, and documentation of successful analysts, start and stop times, duration, significant
completion of such studies should be provided in the assay equipment and material changes, and any potential issues
validation report. General and specific SOPs and good or deviation from the established method.
record keeping are an essential part of a validated Equations used for back-calculation of results.
analytical method. The data generated for bioanalytical Tables of calibration curve data used in analyzing samples
method establishment and the QCs should be documented and calibration curve summary data.
and available for data audit and inspection. Documentation Deviations from the analysis protocol or SOP, with reasons
for submission to the Agency should include and justifications for the deviations 19.
(1) summary information, D. Other Information: Other information applicable to both
(2) method development and establishment, method development and establishment and/or to routine
(3) bioanalytical reports of the application of any methods sample analysis could include:
to routine sample analysis, and Lists of abbreviations and any additional codes used,
(4) other information applicable to method development including sample condition codes, integration codes, and
and establishment and/or to routine sample analysis30. reporting codes.
A. Summary Information: Summary information should
Reference lists and legible copies of any references.
include:
SOPs or protocols covering the following areas:
Summary table of validation reports, including analytical
Calibration standard acceptance or rejection criteria
method validation, partial revalidation, and cross-
Calibration curve acceptance or rejection criteria
validation reports. The table should be in chronological
sequence, and include assay method identification code, Quality control sample and assay run acceptance or
type of assay, and the reason for the new method or rejection criteria
additional validation (e.g., to lower the limit of Acceptance criteria for reported values when all unknown
quantitation). samples are assayed in duplicate
Summary table with a list, by protocol, of assay methods Sample code designations, including clinical or preclinical
used. The protocol number, protocol title, assay type, assay sample codes and bioassay sample code
method identification code, and bioanalytic report code Assignment of clinical or preclinical samples to assay
should be provided. batches
B. Documentation for Method Establishment: Documentation Sample collection, processing, and storage
for method development and establishment should Repeat analyses of samples
include: Reintegration of samples
An operational description of the analytical method. Protocol Guidance: The following provides
Evidence of purity and identity of drug standards, guidelines/tools that should be used to define method
metabolite standards, and internal standards used in performance31:
validation experiments. General Protocol: Prepare and analyze method blanks,
A description of stability studies and supporting data. matrix blanks, reference materials (if available) and matrix
A description of experiments conducted to determine spikes (using matrix blanks if available) of known
accuracy, precision, recovery,selectivity, limit of concentration as generally described .Accuracy or bias and
quantification, calibration curve (equations and weighting precision are calculated from these results. Data will also
functions used, if any), and relevant data obtained from be used to evaluate matrix effects and
these studies. ruggedness/robustness of the method resulting from
Documentation of intra- and inter-assay precision and changes in the sample matrix.
accuracy The following general validation tools should be used to
generate method performance characteristics as described
Any deviations from SOPs, protocols, or GLPs (if
in the Performance Characteristics section below.
applicable), and justifications for deviations
Blanks: Use of various types of blanks enables assessment
49
Reference materials and certified reference materials: The purpose of documenting the procedure covered by the
use of known reference materials (when available and SOP. SOPs shall be located in an accessible place(s) and
applicable) can be incorporated to assess the accuracy or copies shall be available to all personnel needing them to
bias of the method, as well as for obtaining information on perform their duties. Once prepared, these SOPS would
interferences32. only have to be included in QA project plans by reference,
Matrix Blank: A substance that closely matches the after being subjected to prior review and approval. This
samples being analyzed with regard to matrix components. section will be quite flexible and will include all the lab's
Matrix blanks are used to establish background level SOPs for all its' routine operating and analytical
(presence or absence) of analyte(s) and to verify that procedures. The following is a list of possible SOPs and
sample matrix and equipment used does not interfere with should not be considered to be all inclusive35:
or affect analytical signal. 1. Evidentiary SOPs
Matrix Spikes (Laboratory Fortified Matrix): Recovery2. Sample Receipt and Storage
determinations can be estimated from fortification or3. Sample Preparation
spiking with a known amount of analyte and calculation of4. Glassware Cleaning
spike recoveries. (Note: spike recovery may not be truly5. Calibration (Balances)
representative of recovery from naturally incurred6. Analytical Procedures (for each analytical System)
analytes.) Matrix effects can also be assessed with these7. Maintenance Activities (for each Analytical System)
samples. Accuracy or bias and precision are calculated8. Analytical Standards
from these results. The data can also be used to evaluate9. Data Reduction Procedures
robustness of the method resulting from changes in the10. Documentation Policy/Procedures
sample matrix. 11. Data Validation/ Self Inspection Procedures
Incurred Samples: Samples that contain (not laboratory12. Data Management and Handling
fortified) the analyte(s) of interest (if available) may also13. Quality Assurance and Quality Control
be used to evaluate precision and bias (if analyte GC/MS Operating Conditions
concentration(s) are reliably known). Analyte recovery can Preconcentrator: The following are typical cryogenic and
also be evaluated through successive extractions of the adsorbent preconcentrator analytical conditions which,
sample and/or comparison to another analytical procedure however, depend on the specific combination of solid
with known bias. sorbent and must be selected carefully by the operator36.
Sample Collection Conditions
STANDARD OPERATING PROCEDURE Cryogenic Trap Adsorbent Trap
Standard Operating Procedures (SOPs) are to be available Set point - 150EC Set point 27EC
for all routinely used sampling or analytical laboratory Sample volume - up to 100 mL Sample volume -
methods. The Laboratory must maintain a log of all SOPs up to 1,000 mL
in use and must maintain a file of all revisions of SOPs Carrier gas purge flow - none Carrier gas purge flow -
used in the past. A current list of all SOPs and revision selectable
number and date must be appended to the Laboratory QA Desorption Conditions: Cryogenic Trap Adsorbent Trap
Plan. All such methods shall be documented in detail. Desorb Temperature - 120EC
Generally, simply citing a published method is not Desorb Flow Rate - 3 mL/min He
adequate for a SOP. Desorb Time - <60 sec
Published methods rarely have all the procedural details, The adsorbent trap conditions depend on the specific solid
and those that do generally have to be modified for the adsorbents chosen.
applications or facilities at hand. Suggested references for Trap Reconditioning Conditions: Cryogenic Trap
the format of SOPs are included in the reference section of Adsorbent Trap
this document (6, 8). These SOPs shall be prepared in Initial bakeout 120EC (24 hrs) Initial bakeout Variable (24
document control format34. hrs) .After each run 120EC (5 min), After each run
As a minimum the following items should be included: Variable (5 min).
* Title Page GC/MS System: Optimize GC conditions for compound
* Scope and Applications separation and sensitivity. Baseline separation of benzene
* Definitions and carbon tetrachloride on a 100% methyl polysiloxane
* Procedures stationary phase is an indication of acceptable
* QC Limits chromatographic performance.
* Corrective action Procedures, Including Procedures for The following are the recommended gas chromatographic
Secondary Review analytical conditions when using a 50-meter by 0.3-mm
* Documentation Description and Example Forms I.D., 1 μm film thickness fused silica column with
* Miscellaneous Notes and Precautions refocusing on the column.
* References Item Condition
At times certain SOPs may not cover all the above Carrier Gas : Helium
50
and elements should be developed to properly address the Initial Hold Time : 2 min
Ramp Rate : 8E C/min 6. Stein, SE; Scott DR (1994). "Optimization and testing
Final Temperature : 200EC of mass spectral library search algorithms for
Final Hold Time : Until all target compounds elute. compound identification". J Am Soc Mass Spectrom
5(9): 859–866.
SUMMARY 7. Amirav, A.; Gordin, A. Poliak, M. Alon, T. and
The efficient development and validation of analytical Fialkov, A. B. (2008). "Gas Chromatography Mass
methods are critical elements in the development of Spectrometry with Supersonic Molecular Beams".
pharmaceuticals. Success in these areas can be attributed Journal of Mass Spectrometry 43: 141–163.
to several important factors, which in turn will contribute 8. Alon, T.; Amirav, A. (2006). "Isotope Abundance
to regulatory compliance. Experience is one of these Analysis Method and Software for Improved Sample
factors-both the experience level of the individual Identification with the Supersonic GC-MS". Rapid
scientists and the collective experience level of the Communications in Mass Spectrometry 20: 2579–
development and validation department. A strong 2588.
monitoring and training program is another important 9. Robert P., Dr Adams (2007). Identification of
factor for ensuring successful methods development and Essential Oil Components By Gas
validation. Companies must maintain an appropriate level Chromatography/Mass Spectrometry. Allured Pub
of expertise in this important dimension of developing safe Corp. ISBN 1-932633-21-9.
and effective drugs. 10. Adlard, E. R.; Handley, Alan J. (2001). Gas
Gas chromatography-mass spectrometry is the single most chromatographic techniques and applications.
important tool for the identification and quantitation of London: Sheffield Academic. ISBN 0-8493-0521-7.
volatile and semivolatile organic compounds in complex 11. Eugene F. Barry; Grob, Robert Lee (2004). Modern
mixtures. GC-MS is widely used for the quantitation of practice of gas chromatography. New York: Wiley-
pollutants in drinking and wastewater and also used for the Interscience. ISBN 0-471-22983-0.
quantitation of drugs and their metabolites in blood and 12. Eiceman, G.A. (2000). Gas Chromatography. In R.A.
urine. For the residual solvent detection GC-MS is used Meyers (Ed.), Encyclopedia of Analytical Chemistry:
when compared to LC-MS. This is important for samples Applications, Theory, and Instrumentation, pp. 10627.
that have been in the field, held in storage for months, and Chichester: Wiley. ISBN 0-471-97670-9
for highly weathered samples and the most typical use of 13. Giannelli, Paul C. and Imwinkelried, Edward J.
GC-MS in the petrochemical industry it is for process (1999). Drug Identification: Gas Chromatography. In
troubleshooting. Scientific Evidence 2, pp. 362. Charlottesville: Lexis
Law Publishing. ISBN 0-327-04985-5.
REFERENCES 14. McEwen, Charles N.; Kitson, Fulton G.; Larsen,
1. D.R. Jenke, “Chromatographic Method Validation: A Barbara Seliger (1996). Gas chromatography and
review of Current Practices and Procedures. I. General mass spectrometry: a practical guide. Boston:
Concepts and Guidelines”, J. Liq. Chrom. and Academic Press. ISBN 0-12-483385-3.
Rel.Technol., vol. 19 (1996), pp. 719-736. SWGDRG, 15. Niessen, W. M. A. (2001). Current practice of gas
Quality Assurance/General Practices chromatography–mass spectrometry. New York,
Recommendations, 2008. N.Y: Marcel Dekker. ISBN 0-8247-0473-8.
2. A.G. Rowley, Evaluating Uncertainty for 16. Weber, Armin; Maurer, Hans W.; Pfleger, Karl
Laboratories, A Practical Handbook (version 1.1, (2007). Mass Spectral and GC Data of Drugs,
2001). Poisons, Pesticides, Pollutants and Their Metabolites.
3. International Organization for Standardization, ISO Weinheim: Wiley-VCH. ISBN 3-527-31538-1.
9000:2000 Quality management systems— 17. FDA CDER Guidance for Industry Analytical
Fundamentals and vocabulary. Procedures and Methods Validation. Available at:
4. International Organization for http://www.fda.gov/cder/guidance/2396dft.pdf.
Standardization/International Electrotechnical 18. FDA CDER Reviewer Guidance Validation of
Commission, ISO/IEC 17025:2005 General Chromatographic Methods. Available at:
Requirements for Competence of Testing and http://www.fda.gov/cder/guidance/cmc3_rev.pdf.
Calibration Laboratories, paragraphs 5.5-5.6. 19. Grob, RL. Barry, EF. 2004. Modern Practice of Gas
5. David M. Bliesner, Validating Chromatographic Chromatography: Fourth Edition. Wiley and Sons,
Methods, (John Wiley and Sons, 2006, p. 72). Inc: New Jersey.
51
Page